Document zzxoB4mRY8KXqK141996omzq0
DOW CHEMICAL U.S.A.
April 12, 1984
LOUISIANA DIVISION P O BOX 150
PLACtUEMINe, LOUISIANA 7O7&A~0T6O SOA 389-8000
Hugo Andricain - Major Manager
P. J. Prejean - Railroad Ellis George - Tank Car Cleaning Charlie Halphen- Industrial Hygiene
Dannie Kennedy - Industrial Hygiene Elayne Hyatt - Safety
To keep our Beckman 6750 Analyzer at Tank Car Cleaning operating at peak efficiency, with a minimum of maintenance costs, the following is recommended and can be done by our personnel:
1. Inspection of analyzer filters, vacuum pump, sample lines and gas bottles using a daily check list (attached).
2. Re-align and secure all field sample points under Tank Car shed.
3. Number field sample points for easy location.
4. Install filters at all field sample points to minimize problems with water, steam, etc. clogging sample lines. These filters are on hand.
The daily inspection check list and location of sample points are attached.
Additionally, regular calibration of the analyzer by the Instrument Dept, is necessary, probably on a weekly basis. do not have the expertise to do this.
We
I am available to instruct personnel or assist asnieeded.
J. L. Moreau, Jr.
Industrial Hygiene RaiIroad/Marine
DO A 108615 CONFIDENTIAL
DAILY AREA ANALYZER CHECKLIST
DATE
NAME
1. INLET AIR FILTER CHECKS (Open bleed valve to blow out water vapor.) COMMENTS
Filter #1 __ __________ Filter #Z ____________
____________________________________ ____________________________________
Filter #3 _____________
____________________________________
Filter #4 _____________
____________________________________
2. VACUUM PUMP OPERATION/FILTERS
(Oil vapor Filter #5 No bleed valve)
Filter #6
COMMENTS
3. ROTOMETER OPERATION (Check ball position on all rotometers. If ball is seated, sample line is plugged with water. Refer to chart, locate corresponding field sample point and remove filter. Disconnect sample point tubing at outside of analyzer housing. Connect air line tubing to sample tubing. Open air needle valve and blow back for approximately 5 minutes. Check field sample point to insure that water has been blown out. Close needle valve on air sample line. Replace filter at field sample point. Disconnect air line tubing and store in analyzer housing. Reconnect sample tubing. Check for ball position on rotometer.) COMMENTS Rotometer, Sample Point #1 Rotometer, Sample Point #2 Rotometer, Sample Point #3 Rotometer, Sample Point #4 Rotometer, Sample Point #5 Rotometer, Sample Point #6 Rotometer, Sample Point #7 Rotometer, Sample Point #8 Rotometer, Sample Point #9 Rotometer, Sample Point #10
Revised: 3/25/85 LGS/lrj
DO A 108616 CONFIDENTIAL
DAILY AREA ANALYZER CHECKLIST PAGE 2
4. CARRIER GAS AND FUEL GAS BOTTLE PRESSURES (Change bottle if 4000# gauge reads less than 200#.)
COMMENTS
Zero Hydrogen - on stream____________
Zero Hydrogen - spare
_________ _
Zero Nitrogen - on stream _______ __
Zero Nitrogen - spare
__________
Zero Air Zero Air
- on stream _________
- spare
__________
______ ______________ ;
NOTE:
Filter elements should be changed when excessively oily and wet.
Revised: 3/25/85 LGS/lrj
D0 A 108617 CONFIDENTIAL
DAILY AREA ANALYZER CHECKLIST PAGE 2
4. CARRIER GAS AND FUEL GAS BOTTLE PRESSURES (Change bottle if 4000# gauge reads less than 200#,,)
COMMENTS
Zero Hydrogen - on stream_________ ___________________________
Zero Hydrogen - spare
__________
Zero Helium Zero Helium
- on stream
- spare
__________
_____________ __________
Zero Air Zero Air
- on stream
- spare
__________ ___________________________
NOTE:
Filter elements should be changed when excessively oily and wet.
Revised: 10/29/34 LGS/lrj