TO: Chief Chemists (See Distribution;
FROM:
J. W. Born
Brecksviile R&D Center, D/8506
11/18/82
SUBJECT:
Report of Results of the "One-Time'' Vinyl Chloride Cross-Check
Summary
RETENTION CODE 801-001 PROJECT REPORT PERMANENT RECORD COPY
During the past two-year cross-check study of vinyl chloride analyses at BFG plants, the BRDC Environmental Laboratory provided the samples for analysis. The Environmental Lab cleaned the Bendix Flasher tubes (PMCC's) and then loaded each with approximately 5 or 10 micrograms of vinyl chloride monomer (VCM). The recoveries of VCM averaged above 90 percent with 95 percent confidence limits of 35 percent or better each time. The question arose lately whether results would be as good if we used Flasher tubes which each plant cleaned. We made this study to answer the question.
The average percent recovery was essentially the same for both sets of cleaned tubes. The BRDC-cleaned tubes gave 91 percent average recovery of VCM. The plant-cleaned tubes averaged 92 percent recovery. However, the 95 percent confidence limits were 22 percent for the BRDC-cleaned tubes compared with 34 percent for the plant-cleaned tubes. Thus, the accuracy of analysis averaged the same. But the precision of measurement was much better for the samples in the BRDC-cleaned Flasher tubes.
Experiment
The Environmental Lab cleaned five Bendix Flasher tubes (PMCC's) for each of the 13 BFG plants named in Table I. The tubes were cleaned by thermal desorp tion, one by one, in the Bendix Flasher. We repeated the cleaning of each tube until no significant VCM peak appeared in the subsequent GC chromatogram. We then loaded each Flasher tube (PMCC) with approximately 5 micrograms of VCM. The loadings involved flowing a standard commerical 10.0 ppm VCM-air mixture through each tube for 4.0 minutes at about 49 cc per minute. We mailed five Flasher tubes to each plant for GC analysis. The plant labs reported the analytical results shown in Appendices A and B and Table I.
The Environmental Lab also asked eight of the labs to clean five Flasher tubes per plant in the usual way and send them to us. We loaded them with approximately the same amount of VCM and in the same way as before without further cleaning. The five loaded tubes were returned to the plant lab for analysis in each case. Appendix B and Table I give the results.
The Plaquemine plant customarily uses SKC charcoal tubes instead of Flasher tubes. We loaded five SKC tubes for the Plaquemine plant and five more for the Brecksviile Research & Development Center. They were loaded just as the Flasher tubes were. The two labs analyzed the samples. Their results appear in Table
THE B. F. GOODRICH CO.
FEB 1 0 1983
RESEARCH LIBRARY
092200
TABLE I
A Summary of the Analytical Results of the Cross-Check
Identification of Plant______
Percent Recovery of VCM
BRDC-Cleaned Tubes
Plant-Cleaned Tubes
Avon Lake General Chemical Plant Avon Lake Technical Center Brecksville R&D Center Calvert City Deer Park Henry La Porte Long Beach Louisville Niagara Falls Pedricktown Plaquemine Shawinigan
79.2 88.3 115.0 98.4 93.4 89.6
93.1 93.7 72.0 99.1 91.2 81.2 162.0
(a = (a = (a = (a = (a * (ct =
(a = (a = (a = (a = (a = (a = (ct =
8.5) 5.9) 9.2) 10.8)
1.9) 10.4)
18.2) 3.7) 3.7) 1.8)
1.5) 8.5) ll.l)a
67.2 (ct = 6.2)
--
--
124.0 94.3 89.2
108.0 86.8 80.9
(CT = 18.3) (ct = 14.6) (CT = 5.8) (ct = 15.8) (ct = 1.7) (ct = 2.3)
--
--
--
*
Overall Results:
91.2 (ct = 11.0)
92.5 (ct = 17.2)
a This value is a statistical "outlier" according to the Table of Critical Values for T(One-Sided Test). As such, this value was omitted in computing the overall result.
The first group of 75 samples was sent on October 1, 1982. We mailed the second group of 40 samples on October 15, 1982. The last plant report reached the Environmental Lab on November 5, 1982. The dates on which the labs analyzed the samples and the elapsed times appear in Appendix C.
Results
NIOSH advises that analyses of gas concentrations in air should have 95 percent confidence limits of 35 percent of the true value between 0.5 TLV and 1.0 TLV and 50 percent below 0.5 TLV. For VCM in air, 0.5 ppm equals 1.28 mg per cubic meter (at 25C and 760 mm Hg). Six hours' sampling at 10.0 cc per minute gives a sample volume of 0.0036 m3. On that basis 100 percent recovery at 0.5 ppm would give a 4.6 pg VCM sample. thus, measurements of 4.6 pg to 9.6 pg VCM per charcoal tube should have 95 percent confidence limits of 35 percent of the true value. In this study we define "true value" as the weight of VCM loaded per tube. Measurements below 4.6 pg per tube should have 95 percent confidence limits of 50 percent of the true value. In this study the 95 percent confidence limits are 2 CT (twice the standard deviation for the set of measurements). Only two sets of measurements in Table I exceed the 95 percent confidence limits of 35 percent of the true value. The Calvert City value of 149 percent recovery
(Appendix B) accounts for the one case. In the Shawinigan case all five values
,
1
i
TABLE II
A Two-Plant VCM Cross-Check Using SKC Charcoal Tubes
Identification of Plant Sample
Plaquemine:
1 2 3 4 5
Average:
Weight (Mg)
of VCM per SKC Tube
Recovered
Loaded
2.92 3.05 2.74
3.12 2.12
5.30 5.39 5.53 5.69 5.30
Percent Recoverv
55.1 56.6 49.5 54.8 40.0a
51.2 (o = 6.8)
Brecksville R&D Center:
1 2 3 4 5
Average:
b
5.68 8.04 6.51 7.53
4.84 4.42
5.12 5.16 6.14
--
128. 157.a 126. 128.
134.0 (a = 15.8)
a Although this value appears to be quite different from the other values in the set, it is not an "outlier'1 according to the Table of Critical Values for T(One-Sided Test).
The sample was spilled during preparation for analysis.
in the set exceed the +35 percent limit. While a single value beyond the limits in a set is not disturbing, having all five values beyond calls for investigating the cause. Figure 1 presents a graphic comparison of the results in Table I.
The overall average VCM recoveries were 91.2 percent for the BRDC-cleaned Flasher tube samples and 92.5 percent for the plant-cleaned Flasher tube samples. See Table I. The overall standard deviations were 11.0 percent in the former study and 17.2 percent in the latter study. These standard deviations represent 95 percent confidence limits of 22 percent and 34 percent the true values, respectively. Thus, in both studies the overall results were in compliance.
The results of the SKC tube study were both surprising and disappointing. The BRDC average recovery of 134 percent barely complied with the NIOSH limit of 35 percent of the true value for 95 percent confidence. See Table II. Again, an investigation of the cause for the high BRDC values is indicated. All five of the Plaquemine values did not meet the NIOSH requirement.
There are at least four possible causes for recoveries very much above 100 percent. One is inaccurate loading of the sample tubes. A second is incomplete cleaning of the Flasher tubes before loading. A third is poor integration of the area under the VCM peak in the GC curve. And a fourth possible cause is assign* meat of the wrong retention time to the VCM adsorption peak.
The data in Appendix A and Table I indicate that the VCM loading of the tubes was accurate. Also, the tubes were loaded in the order shown in Appendix A. But the selection of the five samples for each plant was spread throughout the entire series at random. Thus, if there had been inaccuracies in loading the Shawinigan Flasher tubes, the data in Appendix B would have shown numerous other very high percent recoveries.
Incomplete cleaning of the Flasher tubes can cause one or both of two problems in analysis. Tailing of adjacent GC peaks can add to the net area under the VCM peak. Also, incomplete cleaning of the Flasher tube can lead to inter* ference of contaminant GC peaks with the VCM peak.
Third, a contaminant peak can merge with the VCM peak to give poor resolu* tion and poor integration.
Fourth, aging can cause migration of the adsorbed VCM throughout the char coal bed of the Flasher tube. A significant change in VCM retention time can result. Failure to recognize the time shift can lead to selection of the wrong peak. Refrigeration of the Flasher tubes until analysis retards the VCM migra tion.
Conclusions The loading of all the Flasher tubes with VCM was accurate. Except for only the Shawinigan results, all percent recoveries and standard deviations met the NIOSJI requirements. The percent recoveries in this "one-time" study are consis tent with the quarterly results of the two-year VCM cross-checks. With the exception of Calvert City, all plants cleaned their Flasher tubes well before the VCM loading occurred. The Shawinigan plant needs to determine why its analyses gave such consistently high results. The Niagara Falls lab deserves praise for its 99.1 percent average recovery with a standard deviation of only 1.8 percent. Those results show that it can be done.
<> *.
i
Recommendations
1. All BFG plants should periodically, at least once a year, make sure that their analyses of Flasher tube samples meet NIOSH requirements.
2. The Shavinigan lab should compare its analytical procedure with the Niagara Falls procedure in detail to improve its performance.
3. The BRDC lab should study the Niagara Falls procedure in an effort to avoid high percent VCM recoveries.
JWB/ms j
Distribution:
M. J. Mannino, ALGC R. D. Hardesty, ALTC R. B. Whitehead, BRDC E. E. Atkins, Calvert City Cliff Miller, Deer Park Vickie Marquis, Henry Archie Hager, La Porte Ron Smith, Long Beach E. D. Hancock, Louisville R. F. Woodruff, Niagara Falls D. P. O'Keefe, Pedricktown R. J. Bigner, Plaquemine J. P. Gregoire, Shavinigan
Senior R&D Chemist
cc
E. B. Kat2enneyer, Jr. J. D. Fannin J. M. Whitney P. M. Zakriski gRjysD^Files;
i)
to >-#
I t s o f th e "O n e -T im e " VCM C ro s s -C h e c k
0 Q
a:
CQ
cn> .
"O a2 C O
O
041)
_n
q
Cl
_uc g
UJ >
CQ
-qa> c q o
O <<ln>
-Q
u
0
U02
ea
Ll. Lt-
XX
-OC 02
CD
-qc 02
CQ
4u u
c 0k.
<rt
3 3--us
o> OC
02 OS
in
01 _Q ,3 i-= O
tn 0*LO.
*n
--U
3
<0 02 QC
0o0
O
E<(s*qm- 3s) ajj|M3ijitiPid
lck>a^x!>o<;<xx>c^(5aRnx3)>fS) ao aisy i||us^3-ig
UObrui/^tJLjg
s
auiujsn DO!d
X
UMO^OJ `P*d
su*y oao J0N
\ xs
S3
EEX
1.
SI
||iAs|no-i
KX x
XX\ EXE
N
LjODSQ i?UO"J
s xxs x xxsxxxxS 9^at ^ d-j
XES X
S5 X XXS S3 ,iudH
X XT \
X3 XX X L2S -ID^ .L99Q
XESS X \ v EE EE
S3N >
VP MAP3
oai*y a)iLMs>fi3^g
u < |D3IUU39J_
XX *?fd
NJSW 3JLU3J3 |DaBur)
4-
o c<
o o
0o0
VoO
o *r
o
WDA 4 /j8A0:,ay }u3:uad
UOA^
joAy
o u <> ** ti
FIGURE .1 A v e ra g e
004
TI
APPENDIX A
Analytical Results for VCM Samples in BRDC-Clcaned Tubes
Identification of Samples
Weight of VCM
Cus) per Flasher Tube
Loaded
Recovered
Percent Recovery
C-l C-2 C-3 C-4 C-5 C-6 C-7 C-8 C-9 C-10 C-ll C-12 C-14 C-16 C-17 C-19 C-21 C-23 C-24 C-25 C-26 C-27
028 030 031 032 C-3S 036 038 039 041 042 043 045 047 049 050 051 052
053 054
055 057 059
5,16 4,92 5.91 4.69 5.59 5.48 5.96 5.04
5.59 5.25 5.69 6.02 6.27 4.88 4.62
5.25 5.49 5.74 5.74
5.39 4.88 4.88
5.25 4.39 4.88 5.00 5.12 5.59 5.00 6.21 6.08 5.44 4.88 4.69 5.39 5.49 5.39 5.12 5.30
5.39 5.08 5.34 5.74 5.30
4.96 4.74 5.20 5.60 6.15 3.71 5.11 1.87 4.00 4.50 3.81 5.66 7.26 3.79 4.91 4.44 4.78 4.89 5.73 5.03 2.63 3.48
4.55 4.12 4.83 4.65 4.65 4.57 3.48 6.20 4.14
6.83 4.45 3.98 5.68
5.35 4.48 4.77 5.37 5.90 4.82 4.96 3.96 5.83
96.1
96.3 88.0 119. 110. 67.7a
85.7 37. la 71.6 85.7 67.0 94.0 116. 77.7 106. 84.6 87.1 85.2 99.8 93.3 53.9a
71.3 86.7 93.8 99.0
93.0 90.8 81.8 69.6 99.8 68.1 126. 91.2 84.9 105. 97.4 83.1 93.2 101.
10994.9 92.9 69-0
110.
APPENDIX A (Continued)
Identification of Samples
Weight of VCM
(ur) ter Tlasher Tube
Loaded
Recovered
C-60
C-61 C-62 C-64 C-68 C-70 C-72 C-74
C-75 C-78 C-79 C-80 C-81 C-83 C-87 C-89 C-100 C-101 C-102 C-103 C-104 C-146
4.92 5.00 4.88 5.08 5.49 5.91 5.91 5.49 6.14 6.40 5.69 5.30 5.00 4.84 4.76 5.59 5.85 5.96 4.92 5.30
5.49 6.02
4.85 4.83 4.49 4.21 6.11 4.30 5.45 8.16 9.36 5.53 4.48 8.78 8.69 5.71 8.10 4.40 5.39 5.52 4.48 4.21 3.84 5.37
Percent Recovery
98.6 96.6 92.0 82.9 111. 72.8 92.2 149.0a
152. 86.4 78.7
166. 174. 118. 170.
78.7 92.1 92.6 91.1 79.4a
69.9 89.2
This value is a statistical "outlier according to the Table of Critical Values for T(One-Sided Test).
0
u
1 %i
GOO
APPENDIX B Results of Analyses of VCM Samples Collected in PMCC's
A.) Results for BRDC-Cleaned PMCC's:
Identification of Plant
Weight (jJg)
of VCM per PMCC
Recovered
Loaded
Avon Lake General Chemical Avon Lake Technical Center Calvert City
5.11 4.78 3.48 4.48 3.96
1.874 4.89 4.57 4.45 4.82
3.71a 4.91 4.55 5.90 4.49
5.96 5,49 4.88 5.39 5.74
5.04 5.74 5.59 4.88 5.08
5.48 4.62 5.25 5.39 4.88
Deer Park
4.96 5.66
4.65 4.77 4.96
5.16 6.02 5.12 5.12 5.34
Henry
4.50 4.44 2.63a
5.68 4.21
5.25 5.25 4.88 5.39 5.08
La Porte
6.15 3.81 4.12 3.98
5.83
5.59 5.69 4.39 4.69 5.30
Long Beach
5.20 5.03 4.65 5.35 4.83
5.91 5.39 5.00
5.49 5.00
Percent Recovery
85.7 87.1 71.3 83.1 69.0
37. la 85.2 81.8 91.2 94.9
67.7a 106.
86.7 109.
92.0
96.1 94.0 90.8 93.2 92.9
85.7 84.5 53.93 105. 82.9
110. 67.0 93.8 84.9
110.0
88.0 93.3 93.0 97.4 96.6
G bazzan?.
APPENDIX B (Continued)
Identification of Plant
Weight (pg)
of VCM per PMCC
Recovered
Loaded
Louisville
4.00 3.79 3.48 4.14 4.30
5.59 4.88 5.00 6.08 5.91
Niagara Falls
4.74
5.73 6.20 5.37 4.85
4.92 5.74
6.21 5.30 4.92
Pedricktovn
5.39 5.52 4.48 4.21a
5.37
5.85 5.96 4.92 5.30 6.02
Plaquemine
5.45 5.53 4.48 4.40 3.84
5.91 6.40 5.69 5.59 5.49
Shawinigan
8.16 9.36 8.78 8.69 8.10
5.49 6.14 5.30 5.00 4.76
Brecksville R & D Center
5.60 7.26 4.83 6.83 6.11 5.71
4.60 6.27 4.88 5.44 5.49 4.84
Percent Recovery
71.6 77.7 69.6 68.1 72.8
96.3 99.8 99.8 101. 98.6
92.1 92.6 91.1 79.4a 89.2
92.2 86.4 78.7 78.7 69.9
149. 152. 166. 174. 170.
119. 116.
99.0 126. 111. 118.
o f-
APPENDIX B (Continued)
B.) Results for Plant-Cleaned PMCC's:
Identification of Sample
Weight (yg)
of VCM per PMCC
Recovered
Loaded
ALGCP
1 2 3 4 5
Calvert City I 2 3 4 5
5.25 5.12 5.12 5.59 5.91
4.95 5.49 5.74 6.14 6.47
4.08 3.39 3.22 3.73 3.69
5.22 8.14
7.41 7.86
Deer Park
1 2 3 4
5
4.88 4.88 4.92 5.25 5.69
5.73 4.60 4.01 4.28 5.53
Henry
1 2 3 4 5
5.74 5.34 5.64 5.30 5-00
4.70
4.85 4.77 4.90 4.81
La Porte
1 2 3 4 5
5.00
4.73 5.25 5.25 5.34
5.19 5.11 5.41 7.04a 4.87
Long Beach
1 2
3 4 5
5.69 5.54 5.44 5.34 5.34
4.82 4.78 4.71 4.64 4.78
Louisville
1 2 3 4
5
6.01 4.95 5.04 5.21
5.25
4.78 4.03 3.97 3.58a 4.41
Percent Recovery
77.7 66.2 62.9 66.7 62.4
105149t
121. 121.
117. 94.3 81.5 81.5 97.2
81.9 90.8 84.6 92.5 96.2
104. 108. 103. 134.a
91.2
84.7 86.3 86.6 86.9 89.5
79.5 81.4 78.8 68.7a 84.0
f}
O
c.;
.>t
,
j
O
h*
APPENDIX B (Continued)
Weight (Mg)
of VCM per PMCC
Identification of Sample
Recovered
Percent Loaded
Pedricktovn
1 2 3 4 5
5.80 5.64 5.54 5.39 5.85
5.23 5.28 5.34
5.03 4.83
Recovery
90.2 93.6 96.4 85.4 82.6
a This value is a statistical "outlier" according to the Table of Critical Values for T(One-Sided Test).
The sample was lost during the thermal flashing step.
APPENDIX C
Elapsed Time Between 1the VCM Loading and Analysis of the Samples
Identification of Plant
Avon Lake General Chemical Plant Avon Lake Technical Center Brecksville R&D Center Calvert City Deer Park Henry La Porte
Long Beach Louisville Niagara Falls Pedriektown Plaquemine Shawinigan
Avon Lake General Chemical Plant
Calvert City
Deer Park
Henry
La Porte
Long Beach
.
Louisville
Pedriektown
Dates on Which
the Samples Were--
Loaded
Analyzed
09/28/82 09/28/82
10/09/82 10/05/82 10/05/82 10/26/82 10/12/82 10/13/82 10/19/82 11/03/82 10/13/82 10/19/82 10/13/82 10/20/82 10/21/82
10/14/82 10/14/82
10/19/82 10/27/82 10/27/82 10/19/82 10/19/82 11/03/82 10/21/82 10/18/82
Elapsed Time (Days)
11 7 7
28 14 15 21 36 15 21 15 22 23
5 13 13
5 5 20 7 4
8
i*