COLUMN CHROMATOGRAPHIC SEPARATION OF POLYCHLORINATED BIPHENYLS FROM CHLORINATED HYDROCARBON PESTICIDES, AND SUBSEQUENT GAS CHROMATOGRAPHIC QUANTITATION IN TERMS OF DERIVATIVES
by O. N. Berg P. L. Diosady G.A.V. Rees
Division of Laboratories Ontario Water Resources Commission
J.H. Neil, Director
MENS 059415
Summary. A novel liquid chromatographic technique, based on charcoal, has been developed for the quantitative separation of PCB's from organo chlorine pesticides.
A gas chromatographic method for the identification and quantitation of PCB's in terms of a well-defined derivative has also been developed. It is based on exhaustive chlorination of PCB's to decachlorobiphenyl. The catalytic dechlorina tion of PCB's to bicyclohexyl is also discussed.
Polychlorinated biphenyls (PCB's) are universal physio logically active pollutants (1,2,3). The analytical (and bio-) chemistry of these materials is poorly developed (2). Although identification of these com pounds can be achieved by GC/MS techniques, the quantitation problem has not yet been solved. The analytical difficulties have been reviewed (4,5). This paper describes a column chromatographic technique for the separation of PCB's from organochlorine pesticides and their subsequent quantitation in terms of welldefined derivatives - bicyclohexyl and decachlorobiphenyl.
This system of analysis is in routine operation in our laboratory. It has been applied to a variety of samples, mainly from the aqueous environment. For example, PCB's (roughly of the Aroclor 1254 variety) have been isolated and quantitated in samples from ducks and seagulls, porpoise blubber, sewage sediments and bottom fauna in ^general. A water sample from Lake Michigan, at the inlet to a water purification plant supplying a major city was found to contain 275 ppt PCB's (determined as decachlorobiphenyl but expressed in terms of Aroclor 1254).
MQhS 059476
Separation of PCB'b from DDT Group of Pesticides
For the quantitative determination of PCB's by gas chromatography. It Is necessary to separate them from Other chlorinated pesticides. More or less successful column chromatographic separation procedures have been published] Reynolds-Florisil (4), Holden and Marsdenalumina and silica gel (6), Armour and Burke - silicagel and Celite (7). For proper separations, the adsorbents have to be partially deactivated with water. In our handB, these techniques have not been too successful. In more general terms, our past experience in other areas of column chromatography has indicated that it is extremely difficult to reproduce the activity of the adsorbents from day to day and from one laboratory to the other. Recently, Zitko (5) discussed these and other difficulties in some detail. In particular, he indicated that the commercial pesticide-grade hexane from various suppliers was contaminated with benzene. The separation of Aroclor 1254 from organochlorine pes ticides on silicic acid was affected markedly by the addition of small (5-20 mls/1) quantities of benzene. In addition, he pointed out that several batches of com mercial silica gel were contaminated with PCB's of the Aroclor 1254 type. Although the PCB's could be removed from the contaminated silica by washing with acetone, the treatment irreversibly changed the chromatographic properties of the adsorbent, and the required activities could not be restored.
In view of these difficulties, we developed a separation technique based on activated charcoal. It was based on the observation that PCB's once adsorbed onto charcoal could not be recovered by hot chloroform extraction (8). We discovered, however, that they could be removed quan titatively with cold benzene. By trial and error, it was found that the DDT group of pesticides and a number Of others* could be eluted from the charcoal column with an ether-acetone mixture, while the PCB's may later be recovered with benzene. The choice of charcoal is critical. Of the several commercially available mater ials, only one is found to be suitable - Fisher cocoanut charcoal for "decolorizing". There is some contribution of EC materials from the adsorbent. These can be removed effectively by Florisil cleanup of the eluates. Oily or fatty extracts of up to about 100 mg of oil do hot change the separations markedly. After the charcoal separations, the oils are also removed by Florisil cleanup.
*Lindane, Aldrin, Dieldrin, Endrin, Heptachlor and Heptachlorepoxide
MOMS 059477
It should be noted that the standard eluotropic solvent series used In chromatographic separations is reversed on charcoal (9). Thus, benzene is the strongest solvent in the series and n-hexane is next to it. Consequently, the sample should be added in the minimum amount of acetone or 25* acetone in ethyl ether. Wet packing of the column gives more reliable results.
One particular run gave the following recoveries: (using 90 mis of 25* acetone in diethyl ether followed by 60 mis of benzene).
p,p'-DDE p,p'-DDD p,p'-DDT 0,p'-DDT PCB-Aroclor 1254
Per cent
91 92 92 94 90
The recoveries are quite consistent. The PCB's are recovered with very little alteration to overall GC peak
profile. Under our experimental conditions, about 1.1 q of charcoal, the system can handle about 5 ug PCB's and about 0.5 ug each of the DDT's.
Identification and Quantitation of PCB's in Terms of Derivatives
The identification and quantitation of PCB's by gas chromatography is a difficult matter. Although a cer tain amount of information can be obtained by the trial and error matching of chromatographic traces with those of commercial PCB's and measuring of prominent peak heights or total areas, the quantitative results are very uncertain.
In particular, it is difficult to determine PCB residues in biological samples. Thus, for example, Mulhern et al. found that GLC analyses of Japanese quail and eggs from mallards fed Aroclor 1254 gave peak profiles that did not match those in Aroclor 1254 (10). In view of this, it would be desirable to base the quantitation of PCB's on a single.well-defined derivative. Our experi mental work has indicated that PCB's can be converted quantitatively into bicyclohexyl or to decachlorobiphenyl. Outlines of our techniques have been disclosed (11, 12).
HONS 059*78
*r
.
I
>
,
. , f 1 | f | t { j ' |
! s > j
,
.
j
i
,
j
i } i ; j '
i
;
------------------------------
Catalytic Dechlorination of PCB's
Under suitable conditions, the PCB's can be dechlorinated quantitatively with hydrogen over Pt or Pd catalysts. The reaction product is a single compound bicyclohexyl. Our experimental procedure followed essentially the guidelines established by Thompson et al, in the early sixties (13) . The major reason for following Thompson's procedure was the discovery that fairly large, about 2.5 mis, quantities of hexane were necessary to recover the reaction product quantitatively from the reactor. Also, occasionally, with lowactivity cataylsts, the conversion was not complete. In this case, the products could be recycled. A somewhat different approach was taken by Gunther et al., who used carbon-skeleton chromatography (14), The conversions were however not quantitative. Depending on the conditions, the PCB's were dechlorinated to cyclohexylbenzene, biphenyl and bicyclohexyl.
During the early stages of our work, this technique was used extensively for the identification and quantitation of PCB's. At the 10 ug levels, the conversions were good, better than 95%. The conversions were somewhat better with the palladium rather than platinum catalysts. There were occasional problems with the reproducibility of catalyst activities. The main dis advantage of this procedure, however, was that the hydrocarbon had to be determined with GC/F1D instrumen tation at a fairly low level of sensitivity. When it was realized that PCB's could be perchlorinated on a pre parative scale to another well-defined derivative decachlorobiphenyl that could be determined with GC/EC Instrumentation (15, 16), attempts were made to adapt this macro reaction to analytical problems.
Perchlorlnatlon of PCB's
The treatment of PCB's with antimony pentachloride under elevated temperatures and anhydrous conditions gives very good yields of decachlorobiphenyl. On the preparative scale (0.5 g), the yields are better than 90%, while on the analytical scale, 1 ug, the yields are somewhat lower, but consistent at 85 i 5%. The latter procedure includes a Florisil cleanup for GC/EC deter Sninations.
MOMS 059479 4
The preparative reaction is best carried out in a Teflonlined screw-cap vial. The temperature and time do not appear to be critical. Treatment for about 5 hours at 170OC will produce a complete conversion. For analytical preparation, it was found necessary to carry out the conversion in sealed glass tubes. On occasion, the tubes were left overnight at the reaction temperature. No by products were formed. The hydrolysis of the unreacted antimony pentachloride was carried out with 20% hydro chloric acid in order to suppress the precipitation of antimony oxychloride and to avoid possible entrainment of product. The gas chromatographic behaviour of decachlorobiphenyl is elementary. Since no separation from other materials is involved, the determination can be carried out with a short column. In our laboratory, the analysis is carried out with a 2 ft. x 1/8* stainless steel column. With the column temperature at 215C, the retention time for decachlorobiphenyl is about 9 minutes (Figure 1). The Aroclor 1254 is eluted in about 3 minutes. The decachlorobiphenyl peak is well removed from those of common chlorinated pesticides and those of Aroclor 1254 and Aroclor 1260. This was confirmed recently by Zitko et al. (17). The relative retention time (six foot column, SE-30) in terms of p,p'-DDEl, for decachlorobiphenyl was 8.20, while for two octachlorobiphenyls they were about 2.5. It was also noted that the relative detector response in terms of p,p*-DDEl for decachlorobiphenyl was 1.6.
EXPERIMENTAL LC Separation of PCB's from DDT Group Pesticides
Materials and Reagents
Column - 6 mm ID x 140 mm Pyrex glass tube with a 25 ml reservoir attached.
Charcoal - Fisher No. 5-690 charcoal, activated, 50-200 mesh, (Fisher Scientific Co., no substitutes have been found). The absorbent is boiled with acetone on a steambath. It is then cooled, and the solvent is removed by suction. This procedure is repeated. The
t filter cake in the funnel is washed with some cold acetone. The charcoal is air dried and stored at 135C.
Ottawa Sand - acid washed.
Glass Wool - solvent extracted.
~
Eluting Solvent A - 25 acetone in ethyl ether (anhy drous)
MOMS 059480
Eluting Solvent B - benzene
Procedure
The tube is plugged with a small wad of solvent-washed glass wool. About 25 mm of acid-washed Ottawa sand is then added; this amount of sand is necessary to retain the fines from the charcoal. A solvent (acetone) slurry
of the adsorbent is then added until the column of the charcoal is 90 mm long (1.1 gm dry weight). The sample is injected directly onto the top of the column. The DDT group pesticides are eluted with 90 mis of Solvent A and the PCB's with 60 mis of Solvent B. Heavily con taminated eluates might require cleanup on Florisil. For testing the column efficiency, the following loadings are suitable:
PCB - Aroclor 1254 p,p'-DDE p,p'-DDD 0,p*-DDT p,p'-DDT
- 1 ug - 0.5 ug - 1.0 ug - 1.0 ug - 1.0 ug
Catalytic Dechlorination of PCB'b
Catalyst
5t by weight Pt or Pd as metal on 60-80 mesh porous glass beads (Applied Science Laboratories, State College, Pa.). An aqueous solution of chloroplatinic acid or palladium chloride is evaporated gently in con tact with the glass beads. The catalyst is reduced in situ.
Reactor and Procedure
12 x 0.25 inch stainless steel tubing fitted with a modified Swagelok tee; one arm of the tee is fitted
with a silicon rubber septum, the other one is inlet for hydrogen gas. The other end of the reactor is fitted with a 2 ft. length of 1/8 inch Teflon spaghetti tubing.
The coiled tubing is cooled in ice water. The products are collected in a cooled 15 ml centrifuge tube. The reactor is kept at 180C. The PCB is injected in
hexane. After a few minutes, 5-6 o.5 ml quantities of
\ hexane are injected in order to purge the reaction
products. The conversion is about 95% at the 10 ug level with on-catalyst injection.
HONS 059481
Perchlorinatlon of PCB
Macro Synthesis
A 0.5 g quantity of PCB (Aroclor 1254) was reacted with 3 mis of SbCIs in a Teflon capped glass tube. After 8 hours at 150C the tube was cooled, opened and the con
tents treated with 20% hydrochloric acid (to prevent the precipitation of antimony oxychloride). The aqueous phase was extracted with warm benzene. The extract was shaken with aqueous bicarbonate solution, water and finally dried with anhydrous sodium sulphate. Evapora tion gave a 95% yield of crude decachlorobiphenyl. The material was crystallized from a 50:50 mixture of benzene and pet. ether, M.P.304C.
The white crystalline material gave one peak on GC. IR spectrum was obtained with Perkin-Elmer 221 spectro photometer (Figure 2). It was sublimed in vacuum for elemental analysis.
C12C110
Calc. 28.88 71.12
Found 28.99 70.78
The molecular weight of the material was obtained by mass spectrometry.
Micro Synthesis
The PCB, one microgram or less, is treated with 0.2 mis of SbCIs in a sealed glass tube.
The PCB in a small quantity of a suitable solvent is transferred into a Carius tube, B an OC, 6 mm ID, 160 mm long. The solvent is removed by gentle vacuum - thin Teflon tubing inserted into the glass tube. A quantity of 0.2 mis of SbClj is added and the
tube is sealed. After a minimum of three hours at 160-170C, the tube is opened and the contents are
worked up according to instructions given above for macro scale. With careful attention to the details, the ' recovery is about 85 1 5%.
Gas Chromatographic Conditions for the Determination of:
1.Bicyclohexyl
Instrument - Varian 1500 FI Detector
'
Column - 8 ft. x 1/8" stainless steel, 10% DC-710 on Chromosorb W
HONS 059482
Temperature - 1. Column 90C for 2.5 minutes then programmed at 10C per minute
2. Injector 220C
3. Detector 220C
Gas Flow - Air 200 mls/min. Hydrogen 20 mls/min. Nitrogen 25 mls/min.
Range - 1
Attenuation - x8
Recorder - 1 mv full scale
Under these conditions, an injection of 50 ng of bicyclohexyl gives a chart response of 45 units. The retention time is 9.0 minutes.
2. Decachlorobiphenyl
Instrument - Varian 1500f Tritium Detector
Column - 2 ft. x 1/8" stainless steel 5% Se-30 on Chromosorb W
Temperature - 1. Column 215>C
2. Injector 220C
3. Detector 215C
Gas Flow - Nitrogen at 50 mls/min.
Range - 1
Attenuation - x8
Recorder - 1 mv full scale
The peak is Gaussian in shape. The retention time is approximately 9 minutes (Figure 1).
ACKNOWLEDGEMENTS
The skillfull technical assistance of Mr. J. Osborne is
gratefully acknowledged. Monsanto Company kindly
supplied a number of Aroclor samples.
MGNS 059483
REFERENCES
1. RISEBROUGH, R.W., RIECHE, P., PEAKALL, D.B.,
HERMAN, S.G., and KIRVEN, M.N., Nature 220 109B
(1968).
.
2. PEAKALL, D.B., and LINCER, J.L., BloScience 20 958 (1970).
3. GUSTAFSON, C.G., Envir. Sci. Technol. 4 814 (1970).
4. REYNOLDS, L.M., Bull, of Environ. Contam. & Toxicol. 4, 128 (1969).
5. BITKO, V., 54th Canadian Chemical Conference, Halifax, May 1971.
6. HOLDEN, A.V., and MARSDEN, K., J. Chromatog 44, 481, (1969).
7. ARMOUR, J.A., and BURKE, J.A., J. Assn. Offic. Anal. Chem. 53 761 (1970).
.8 Newsletter 18, January 1971, United States Environ mental Protection Agency, Hater Quality Office, Analytical Quality Control Laboratory, Cincinnati, Ohio.
9. ALM, R.S., Acta Chem. Scand. 6 1186 (1952).
10.
MULHERN, B.M., CROMARTIE, E., REICHEL, H.L. and BELISLE, A.A., J. Assn. Offic. Anal. Chem. 54 548 (1971).
11.
BERG, O.H., and REES, G.A.V., Identification and
Ouantitation of Polychlorinated Biphenyls by Catalytic Dechlorination., 2nd Seminar on Pesticide Residue Analysis (Eastern Canada), May 22, 1970, Ottawa, Ontario.
12. *
BERG, O.W., DIOSADY, P.L., and REES, G.A.V., Notes on Column Chromatographic Separation of PCB's from Chlorinated Hydrocarbon Pesticides, and Subsequent Gas Chromatographic Quantitation of PCB's in terms of Decachlorobiphenyl, with EC Detector, 3rd Seminar on Pesticide Residue Analysis (Eastern Canada), April 29, 1971, Montreal, Quebec.
13. THOMPSON, C.J., COLEMAN, H.J., WARD, C.C., and
RALL, H.T., Anal. Chem. 34 154 (1962).
.
MCNS 059484
14. ASAI, R.I., GUNTHER, F.A., WESTLAKE, W.E., and IWATA, Y.J., Agr. Food Chem. 19 396 (1971).
15.
HUTZINGER, 0., National Research Council of Canada, Atlantic Regional Laboratory, Halifax, N.S., Personal Correspondence.
16. HUTZINGER, 0., SAFE, S., and ZITKO, V. , Bull, of Environ. Contain. 6 Toxicol. 6 209 (1971).
17. ZITKO, V., HUTZINGER, 0., and SAFE, S., Bull, of Environ. Contain. 6 Toxicol. 6 160 (1971).
V HONS 059485
1.
i .1
FIGURE 1. gc/ec chromatogram of a gull tissue sample.
I
1 I
A - Aroclor 1254
B - Fraction from charcoal column
C - Fraction after chlorination and Florisil clean-up.