Document wv6ejY2ezomgZpEwgj8re2y4
POLYPROPYLENE
13R9A
ti
I FLINT HILLS resources
DESCRIPTION: Random copolymerfor injection molding.
FEATURES:
Produced without animal derived components.
STERILIZATION: Gamma radiation, ETO, E-beam and autoclave.
APPLICATIONS: Medical device components, labware, closures.
RESIN PROPERTIES
Melt Flow Rate Density Melting Temperature Range
TYPICAL VALUE
(SI)
9 g/10 min
0.90 g/cm3 145-152 C
TYPICAL VALUE
(ENGLISH)
293-306 F
ASTM TEST METHOD
D 1238 D 1505 D 3418
MECHANICAL PROPERTIES Tensile
Yield Strength Yield Elongation
Flexural Modulus Tangent
Deflection Temperature @ 66 psi (.455 MPa) Rockwell Hardness Notched Izod @ 23C Gardner Impact @ 23C
Feb. 17
28 MPa 9.0 %
966 MPa 90 C
60 J/m >36.2 J
4000 psi 9.0 %
140 kpsi 194 F 90 R 1.2 ft-lb/in >320 in-lb
D 638
D 790 D 648 D 785 D 256 D 5420
Regulatory
21 CFR 177.1520(a)(3)(i) and (c)3.1a Drug Master File listed
USP <88> Class VI, USP<85>, & ISO 10993-5
Notice regarding medical applications: This Flint Hills product meets certain requirements for use in medical applications. It is the responsibility of the medical device or pharmaceutical manufacturer to determine that this Flint Hills product is safe, lawful and technically suitable for the intended use. Flint Hills encourages its customers to review their application with a Flint Hills technical representative to ensure that this product is not used in ways for which it was not intended. Flint Hills makes no warranties (express or implied), promises or guarantees concerning the suitability of this product for use in any given medical application.
118 Huntsman Way, Longview, TX 75602 Tel: 1
The data and information represented herein refer to typical values obtained in our laboratories by the methods or apparatuses indicated, and should be so considered. Since processing variables are a major factor in product performance, this information should serve only as a guide. Since customers' testing conditions are outside our control, the reproducibility of our data in a customer's testing facility is not guaranteed. Customer should confirm results under its testing conditions. There is no implied warranty of merchantability or fitness for a particular purpose. Establishing satisfactory performance of the product for the intended application is the customer's sole responsibility. No warranty is given concerning the existence or non-existence of any patents claiming any pertinent subject matter presented herein. The Company assumes no obligation, express or implied, or liability for use of or reliance on the information and data presented. FHR disclaims all product warranties expressed or implied, including warranties of fitness for particular purpose or of merchantability. Further, this product is not intended for use in the manufacture of any form of implanted medical or surgical device.