CONOCO INC. CONOCO CHEMICALS LAKE CHARLES CHEMICAL PLANT LABORATORY
'Li
IC-PlUA-q.
Benzene in Air Analysis for Benzene from Personnel Dosimetry Charcoal Badges
Safety Precautions
Both benzene and carbon disulfide are extremely toxic chemicals. A11 transfers of these materials must be made in a fume hood with the blower running. Rubber gloves must be worn whenever any contact with benzene or carbon disulfide is possible. Badge caps or plugs must be tight enough not to leak CS2 fumes.
Explanation: Note 1
Procedure
X. From a calibrated solvent dispenser add 1.5 ml ( .05 ml) carbon disulfide containing 0.05 mg/ml undecane (Cn Paraffin, 33%) to each badge and cao tightly. Allow 10 minutes for desorption while swirling occasionally. (Note 2) After desorption attach short piece of tubing to side part on badge and pour contents into an auto sampler vial and cap.
2. The new Perkin-Elmer Sigma I chromatograph on your left when facing the console has been designated and set up to run Benzene and certain other personnel dosimetry analyses. This GC has a self-contained computer. Following is the step by step procedure for setting up the GC system for the first analysis in a series.
A. The 20* 1/8" 20% FFAP column is to be connected to the upper injector and the lower (B) detector. On top of the GC: Be sure B detector button is depressed and the to A flame is off. GC must be set to Range 10 and Flows to 20. On`the Keyboard: Push in the exact order listed.
1. Void 2, 2 3. Enter 4. Set-up 5, 2 (Refers to GC) 6. Enter 7. 1 (Refers to Detector) 8. Enter
9. 211 (Refers to Method)
10. Enter 11. Enter
12. Enter 13. Sample Number (Your own sample #) 14. Enter 15. Sample Name (From tag by Safety Dept
16. Enter
Hen oooox 1929
LC-PCDA-4
2
f
The 6C Is now set up to run. Sample can be injected when light goes on in button by injection port. Samples may also be run using Auto
The above procedure will only provide a printout. In order to have a trace (chromatogram) the plotter may be linked to the GC as in "B" below, or a strip chart recorder may be connected to the chromatograph.
6, On the Keyboard: Push in the exact order listed.
1. Plot 2. Void 3. Plot 4. 2 (Refers to GC) 5. Enter
6. 1 (Refers to Detector) 7. Enter 8. 0 9. Enter
Both the GC and recorder are now ready to go. Inject 1 microliter into the lower injection port on the GC and push the lighted button just below to start the computer and recorder.
In order to run another sample in the series, only Section A above has to be repeated. (Assuming no one uses the instrument between your runs).
In order to rerun the same sample nothing has to be done except wait for the start button to relight, indicating the system is ready. Simply inject again and push the start botton.
3. Results will be printed out in mg (milligrams) benzene and logged on a sheet designated by the Control Section. All traces and printouts are to be turned into the Control Chemist for forwarding to the Safety Department. No calculations are required by the Analyst. If no area is detected for benzene, report<0.0025 mg.
Notes:
1. Charcoal badges will be tagged with a sample number. A blank badge will be submitted with each series of samples and should be handled and logged the same as samples.
2. Carbon disulfide containing 0.05 mg (milligram)/ml undecane: Fill 1 100 ml volumetric flask equipped with a glass stopper to the mark with reagent grade carbon disulfide (CS?). With a 10 microliter syringe add exactly 6.75 microliters of high purity 99%+) undecane (C^ paraffin). Stopper and nix by shaking vigorously for at least 1 minute. (Caution: Must be done in a fume hood with the blower on while wearing rubber gloves.
3. Reagents:
1. Carbon disulfide (CS2) Reagent Grade 2. N-Undecane (Cn Paraffin) 99% or Better Purity
Revised: 2-14r80 by DDP Approved by
MOD 000011930
LC-PCDA-4
Note 5
To verify Benzene Personal Dosimetry Method
1. Prepare carbon disulfide containing 0.05 mg/ml undecane as explained in Note 2 of LC-PCDA-4 (Benzene in air) method.
2. Prepare a stock solution of 5 mg/ml benzene by weighing 0.5000 grams of benzene into 100 ml Volumetric flask. Dilute to the mark with desorbing solution described in #1 above.
3. Add 1.50 ml desorbing solution described in #1 to as many Sigma I sample vials as needed and cap.
4. To simulate in a 10 liter air sample:
0.030 mq 10 liters
= 1 ppm
Add 6 microliters stock solution from #2 to vial containing 1.5 ml D.S.
0.015 mq = 0.5 ppm 10 liters
n 0.003 mq
10 liters
* 0.1 ppm
Add 3 microliters stock solution from #2 to vial containing 1.5 ml D:5.
Add 0.6 microliters stock solution from #2 to vial containing 1.5 ml D.S.
D.S. - Desorbing Solution from #1
6-20-78 DOH
MCt> 0000ll93i
LC-PCDA-U
3
c
4. GC Parameters
GC North P & E Sigma I
Method:
#11
Conditions: Injection Port Temp 250C
Detector Temp
275c
Carrier Flow
30 cc/min
Initial Oven Temp 50C 9 min. Hold
Program Rate Range
5c/min to l60C 10
Column:
20* x 1/8" 10# FFAP on Chrom. W-AW-DMCS 80/100 Mesh
Issued: 9/9/77 'by DDP Revised: 2/l/lS by LBD & DDH
Approved by
r*
,nQ
.vY
V,'
--* v* v
j-jU'v-< m
i v
o.-*-* * >.v***.. ** < v'.*' *"
'
*' - .!* ^ *?' w-<
CONTINENTAL OIL COMPANY
CONOCO CHEMICALS Lake Claries Chemical Plant Laboratory
LC-PC
Methyl Chloride in Air (Personnel Dosimrtry via Charcoal Tubes and Bendix Flasher)
GC Parameters IP-203, Option 39, S Conditions as listed on Instrument Carrier Flow 50 cc/min Column - 10T, l/8" Poropak QS, 80/100 Mesh Order of Elution - See Reference Trace
Flasher Parameters Temperature 275C Check for leaks using an empty charcoal tube in flasher.
Calibration of Instrument Syringe inject 0.25 cc pure methyl chloride frcsn a LCCP shipment into GC s times to obtain an average Area, record. Instrument sensitivity (range 10" be such that an attentuation of 128, or more, is required to keep methyl c peak on scale. "Methyl Chloride should be handled in a hood".
Procedure 1. Place Flasher mode switch in by-pass position. 2. Insert charcoal tube (ringed end down) and allow to heat for one minut
(Be sure angel hair is not protruding outside tube ends as this contri to leaks). 3. Place Flasher mode switch in Inject Position, start recorder and compu* Record area of methyl chloride peak. 4. Analyze both Primary (2-rings) and Secondary (3-rings) tubes.
V3 o,0.0
rr1(2i3 4 ;r
LC-Pd>-10
2
Calculation:
Area of Sample Area of Standard
ml Methyl Chloride
Reporting Results:
Report ml Methyl Chloride on 1. Primary Tube 2. Secondary Tube 3, Total (Primary + Secondary)
O
Issued: 9/15/77 by DDP Approved by
i
\
MOD 000011934
v J*-*+*'#***>'fj*a*,'*+**;*!+**.> *' ^->-"-4 * - ^ - *'*'
CONOCO INC. CONOCO CHEMICALS LAKE CHARLES CHEMICAL PLANT LABORATORY
ANALYSIS OF METHYL CHLORIDE
Safety Precautions
Methylene chloride is highly toxic by inhalation, and a strong eye irritant. Avoid contact with eyes, skin and clothing. All transfers must be made in a fume hood with the blower running.
Procedure
i
1. The Analyzer 2 is used.
A. The FFAP column is to be connected to the upper injector and the lower
detector (B).
j
On top of the GC:
Be sure detector B button is depressed and H2 to A flame is off.
On the Keyboard:
Push in the exact order listed.
1. Void 2. 2 3. Enter 4. Set-up
5. 2 (refers to GC) 6. Enter 7. 1 (refers to detector) 8. Enter
9. 250 (method no.) 10. Enter
11. Enter 12. Enter 13. Sample no. (your sample no. 14. Enter
15. Enter
The above procedure will only provide a printout. In order to have a trace (chromatogram) the recorder must be turned on using 0.25 in/min.
2
2. Results will be printed out in mg (milligrams) of methyl chloride and logged on a sheet designated by the Control Section. All traces and printouts are to be forwarded to the Control Chemist.
NOTE: Since the desorption solution may contain a small amount of methyl chloride, it may be necessary to subtract this value (obtained from blank) from milligrams reported by instrument.
3. GC Parameters
GO: Analyzer 2 Method 250
Conditions:
Injection Port Temp Detector Temp Carrier Flow
Initial Oven Temp Program Rate Range
250C &75C
30 cc/min 50&C, 8 min
20OC/min to 180C
10
Column: 20' x 1/8" 103S FFAP on Chrom. W-AW-DMCS 80/100 mesh
Sample Size:
1-5 ul
Recorder:
0.25 in/min, ATTN-4
Auto-Sampler Use in Sampling Methyl Chloride for Analysis
Safety Precautions:
Methylene chloride is highly toxic by inhalation and a strong eye irritant. Avoid contact with eyes, skin and clothing.
All transfers should be made in a fume hood with the blower running.
1. Take a 10 ml serum vial which contains the charcoal to be desorbed. Using a 5 ml gas tight syringe, add 4 ml of desorption solution. Gently, shake solution. Allow solution to set for 10 minutes before proceeding to Step 2.
2. Using a different 5 ml gas tight syringe, withdraw about 1 ml of liquid from 10 ml serum vial in Step 1. Prior to removal from the vial, the syringe must be closed. This is accomplished by holding the barrel firm and turning the tip in a clockwise direction two full revolutions. Take care to align dot on tip with dot on barrel, NOTE l MCD 000011936
NOTES:
1. This syringe must be rinsed in a small vial of desorption solution between each sample. This is to minimize cross-contamination.
3
C
3- Remove the syringe from the serum vial. Inject into a capped auto-sampler vial until about 3/4 full. Auto sampler vials caps should be tight enough not to turn.
4. Make a list as you proceed, starting with Lab #1 and correlating this number to the date, sample no., and whether the vial contains primary or secondary sample.
EXAMPLE:
Lab #1 Lab #2
Date
1/1/79 1/1/79
Sample #
MC-700 MC-101
Primary Secondary
Included in the auto-sampler tray should be at least one vial of desorption solution (neat) to be used as a blank.
5. Load the circular auto-sampler rack in numerical order with the Lab # for each vial corresponding to numbers inscribed around the upper section of the rack.
Make two copies of the list prepared in Step 4 hnd keep the original in the work area to attach to the data printouts from the Sigma 1 printer.
Before releasing this data, it must be checked by the Shift Supervisor/or Control Group Leader.
4
6. Make sure the auto-sampler is turned off (on the control unit). Turn the mode selector (on the control unit) to EXTERNAL. After checking the injector unit alignment, insert the sample rack into the injector unit and tighten with the big Teflon screw. Slide the rack holder into the unit and lock in place by means of a screw on the right side rear of the injector unit.
7. Adjust blank knob on pressure gauge (labeled 7/20) to a reading of 1-2 pounds. Turn to right to increase and to left to decrease.
8. Follow the procedures in Step I of the specific method for the analysis involved to set up the Sigma I GC and recorder.
9. Only after everything is ready, turn the switch on the auto-sampler control unit to the ON position. The unit should then process all the samples in the sample rack. There must be no empty spaces between samples in the rack that are to be run. Always leave position #30 (upper rack section #) empty (no vial) or the instrument will continue to cycle.
3*7
/->*-- >
4
Preparation of Desorption Solution for Methyl Chloride Analysis
Caution: Methylene Chloride is highly toxic by inhalation; moderately by injestion and skin absorption. Strong eye irritant.
1. Weigh a 100 ml volumetric flask with stopper on a 4 place balance. Record weight.
2. Weigh 1 g + 0.01 of n-octane into the flask (replacing stopper to record weight).
3. Dilute to 100 ml mark with methylene chloride-that is methyl chloride free. Record weight. This is Solution "A",
4. Calculation:
a. Subtract wt. in Step 2 from wt. in Step 1.
Step 1 weight - Step 2 weight
Gram of n-octane
b. Divide wt. of n-octane by wt. in Step 3.
n-octane _______________________________________ ______ _ f (Step 3 wt. - Step 1 wvT) or wt* of total solution = Tac
5. Weigh a second 100 ml volumetric flask with stopper on a 4 place balance. Record weight
6. Calculate the amount of Solution "A" to add to the volumetric flask by th's equation:
\=
.01540 factor (Step 4b)
= number of grams (Sol. "A") to add to volumetric flask (B).
7. Weigh in exactly this amount of Solution "A" to the nearest 0.005 g (replacing stopper to record wt,).
8. Dilute to exactly 100 ml with methylene chloride.
Calculation:
(wt. from Ste_7,)iiactor fron Step. 46j------ 1000 mg/g = j< 9
9. Transfer to a septum capped 4 oz bottle.
MCD ooooil938
5
O . Selection of sample size to be injected
1. Set-up injector for a "dummy" injection.
a. Unscrew thumbscrew which holds sample tray base plate in position (right rear corner).
b. Pull out the plate which holds the sample tray.
c. Place an empty vial (no cap, no septum) in Position 1 (read lower row of numbers). Rotate the tray until the number 1 on the upper ring of numbers is aligned with the arrow on the base plate. This insures that vial No. 1 is in position under the filter probe, ready for sampling.
d. Make sure the sample filler probe is all the way up, and that the needle and syringe are out of and away from the GC injection port,
e. Turn rotary selector switch on programmer to manual.
f. Turnpower switch "on".
g. Turn volume adjustment knob (smaller nut on the end of the syringe) --. clockwise as far as it will go, (Approximately 10 turns) Do Not f J exert excess force to turn the knob.
h. Push and hold the lighted "Inject" button until the light goes out. This will start a sample cycle, which you will interrupt to set sample size.
i. As the cycle proceeds, watch volume adjustment knob as it moves out approximately one inch. After a few seconds the knob will spring back in. Immediately turn off the power switch.
j. Observe the plunger tip of the syringe. It will be located at some volume calibration mark on the scale of the syringe barrel.
To reduce the volume to be injected, turn the volume adjustment screw counter-clockwise until the plunger tip aligns with the desired volum.e mark. The unit is now ready to operate.
k. To enlarge the sample injection volume, start at Step e and proceed through step i,
l. Repeat Step e through Step i until desire volume is obtained. tfCD 0000 U939
r
Issued: 7-3-7.9L|?y W or.r^Appraved. -by
*'r***,'*J.i**. /***.-**>*
CONTINENTAL OIL COMPANY CONOCO CHEMICALS
LAKE CHARLES CHEMICAL PLANT LABORATORY
Methanol in Air
Analysis for Methanol frcm Personal Dosimetry Silici. Gel Tubes
Safety Precautions
Reasonable care should be exercised when handling methanol and 2-butanol which have TLV*s of 200 ppn and 150 ppn in air respectively.
Explanation Note 1
Procedure
1. Using a 2.5 cc gas tight syringe equipped with a 21 ga 1 1/2 inch needle, accurately + 0.05 ml add 1.5 ml (1.5 cc) deionized water containing 3.0 mg/m.1 2-butanol to each vial and cap tightly. Allow 4 hours for desorption while swirling occasionally. Note 2
2. The new Perkin-Elmer Sigma I chromatograph on your left when facing the console has been designated and set up to run methanol and certain other personal dosimetry analyses. This GC has a self-contained computer.
The step-by-step procedure for setting up the GC and its computer system is as follows for setting up the first analysis in a series. Note: If the auto-sampler is to be used see the special procedure.
A. The OV-1 Column is to be connected to upper injection port and the upper (A) detector.
On Top of the GC: Be sure that "A" detector button is depressed and the H2 to "B" detector is "off" and H2 to "A" detector is "on". On Front of the GC: Range 100
3. On the Keyboard: Push designated keys in the exact order listed.
1. Void 2. 1 3. Enter 4. Set-up 5- 1 (Refers to GC) 6. Enter
7- 1 8. Enter
9. 12 (Refers to Method) 10. Enter 11. Enter 12. Enter 13* Sample Number (Your Own Spl #) 14. Enter 15. Sample Name (From Tag by Safety Department)
16. Enter
15-30 second delay*wait till light rocs off
The GC is now set-up to run. Sample can be injected when light goes on in i-.-
...... by. injection port. ^ .. ...
. .... .
-
The above procedure will only provide a printout. In order to have a trace (chromatogram) the plotter must be linked to the GC as follows:
................ ' ............
B. On the Keyboard: Push in the exact order listed
1. Plot 2. Void 3. Plot 4. 1 (Refers to GC)
5. Enter
6. 1 (Refers to Detector)
7. Enter 8. 0 9* Enter
(
Both the GC and the recorder are now ready to run. Inject 1.5 microliters into the upper injection port on the GC and push the lighted button just below to start the computer and recorder.
In order to run another sample in the series, only section "A" above has to be repeated. (Assuming no one used the instrument between your runs).
In order to rerun the same sample nothing has to be done except wait for the start button to light up indicating the system is ready. Then inject again and push the start button.
3* Results will be printed out in mg (milligrams) methanol and logged on a sheet designated by the Control Section. All traces and printouts are to be turned in to the Control Chemist for forwarding to the Safety Department. No calculations are required by the Analyst.
Notes:
r
1. One dram vials containing silica gel from tubes used in personal dosimetry sampling will be prepared and tagged by the Safety Department. Two vials will be submitted for each tube (sample). One vial will contain silica gel from the primary section of the tube and will be designated by the letter "PM on the tag (Example: Me-U-P) and the other vial will contain silica gel from the secondary section and be designated by the letter "SM (Example Me-4-s) The contents of each vial will be analyzed exactly the same way.
2. Deionized water containing 3 mg/ml 2-butanol: Weigh exactly 0.3000 grams + 0.001 gram 2-butanol into a 100 ml volumetric flask equipped with a ground glass stopper. Fill carefully to the mark with deionized water. Stopper and shake vigorously for 30 seconds three times over a 10 minute period.
3- Reagents: 2-butanol 99-0+J Deionized water (Lab supply)
4. GC Parameters GC: North P & E Sigma I H# 327 Method: #12 Detector Temp: 275C
Injection Port Temp: 250C Carrier Flow: 16 cc/min Initial Oven Temp: 40C - 4 min Hold Program Rate: 10c/min to 170C Range: 100
Yv 0.0^ S$>
Column 12* by l/8" 3% OV 1 on Chromsorb HP GO/IOO Mesh ^-r
Issued: 3/15/78 bv DDK ____ Approved by ...
*
l Note 5
To verify method for Methanol
LC-PCDA-15
1. Prepare deionized water containing 3 mg/ml 2-butanol desorbing solution as described in Note 2.
2. Prepare a stock solution of 130 mg/ml methanol by weighing 13.000 grams (4 place balance) methanol into a 100 ml Volumetric flask and diluting to the mark with desorbing solution from #1 above.
3. Add 1.50 ml desorbing solution described in #1 above to as many Sigma I sample vials as needed and cap.
4. To simulate as in a 10 liter air sample.
1.3 mg 10 liters
100 ppm
Add 10 microliters stock solution from #2 to vial containing 1.50 ml D.S.
0.65 mg 10 liters
50 ppm
Add 5 microliters stock solution from #2 to vial containing 1.50 ml D.S.
-
D.S. - Desorbing Solution from #1
6-20-78 DH
/o<&
&
C0NT3 RENTAL OIL COMPANY CCNOCO CHEMICAIS
TAKE CHARLES CHEMICAL PLANT IABORATORY
Ethanol, Butanol, and Hexanol in Air
Analysis for CgOH, C4OH and C60H from Personal Dosimetry Charcoal Tubes
Safety Precautions:
Carbon disulfide is an extremely toxic chemical. All transfers of material containing carbon disulfide must be made in a fume hood with the blower running. Rubber gloves must be worn whenever any contact with carbon disulfide is possible.
Explanation: Note 1
Procedure:
1. Using a special 2.5 cc syringe equipped with a 21 ga. 1 1/2 needle, accurately (+ 0.05 nil) add 1.5 ml (1.5 cc) carbon disulfide containing l.M by wt., 2 butanol and 1.52 mg/ml N dodecane (Cx2 paraffin, 99$+) to each vial and cap tightly. Allow 30 minutes for desorption while swirling occasionally. Note 2
2. The new Perkin-Elmer Sigma I Chromatograph (IP-327), on your left when facing the console, has been designated and set up to run ethanol, butar.cl, and hexanol as well as certain other personal dosimetry analyses. This GC has a self-contained computer. Following is the step by step procedure for setting up the GC sys .em for the first analysis in a series.
A On top of the GC: Be sure B detector button is depressed.
On the Keyboard: Push keys in the exact order listed.
On Front of GC:
Range switch 100 depressed
1. Void
9. 26 (Refers to Method)
2. 1
10. Enter
3. Enter 4. Set up
11. Enter 12. Enter
5. 1 (Refers to GC)
13. Sample Number (Your Own Spl. #)
6. Enter
14. Enter
7. 1 (Refers to Detector) 15. Sample Name and/or Number (From
8. Enter
16. Enter
The GC is now set-up to run. Sample can be injected when light car-on in button by the injection port.
The above procedure will only provide a printout. In order to hav* 1 trace (chromatogram) the plotter must be linked to the GC as folio-4
2
B On the Keyboard: Push keys in the exact order listed.
1. Plot
2. Void
3. Plot 4. 1 (Refers to GC)
5. Enter
6. 1 (Refers to Detector)
7. Enter 8. 0 9. Enter
Both the GC and recorder are now ready to run. Inject 2 ul microliter sample into the lower injection port on the GC and push the lighted button just below to start the computer and recorder.
In order to run another sample in the series (analysis for the same compounds), only section A above has to be repeated. (Assuming no one uses the instrument between your runs)
In order to rerun the same sample, nothing has to be done except wait for the start button to relight, indicating the system is ready. Simply inject again and push the start button.
3. Results will be printed out in mg (milligrams) C2OH, Ci+OH, and/or C60H and are to be logged on a sheet designated by the Control Section. All traces and printouts are to be turned in to the Control Chemist for forwarding to the Safety Department. No calculations are required by the Analyst.
Notes:
1. Explanation - One dram vials containing charcoal from tubes used in personal dosimetry sampling will be prepared and tagged by the Safety Department. Two vials will be submitted for each sample tube. One vial will contain charcoal from the primary section of the tuba and will be designated by the letter "p" on the tag (Example: Alc-2-I) and the other vial will contain charcoal from the secondary section and be designated by the letter "S" (Example: Alc-2-S). The contents of each vial will be analyzed exactly the same way. Blanks (unused tube's charcoal) will be submitted with each series of samples and will also be analyzed and logged like the samples.
2. Carbon disulfide desorbent preparation. Fresh carbon disulfide desorbent must be prepared monthly or with each series of samples (which are scheduled several weeks apart). For C2OH, C4OH, and COH determinations prepare desorbent as follows: Weigh exactly 1 gram + .05 2-butanol into a glass stopp 100 ml volumetric flask. Then weigh in exactly 0.1523 + 0.001 grams dodecane (99i+ NC12 paraffin) on a 4 place balance. Fill the flask carefully to the mark with reagent grade (approved by Control Section as being acceptable for use) carbon disulfide. Stopper and mix by shaking vigorously for at least 1 minute.
w
o,0O.O\
&
.V..
*
3, (a) Carbon disulfide (CS^) - Reagent Grade O (each lot needs to be approved by Control Section for use)
(b) N-dodecane (C12 paraffin) - 99% or better purity
(c) 2 Butanol - Reagent Grade
U* GC Parameters (conditions)
GC: North P&E Sigma I IP# 327 Method; #26 Injection Port Temp: 250C Detector Temp: 275C Carrier Flow: 30 cc/min Initial Oven Temp: 65C Program Rate: 5C/min Max Temp: 220/C Column 20' x 1/8" 10# FFAP on Chrom. WAW-DMCS 80/100 mesh
3
O
Issued: 2-15-78 by LBD & DDH Approved by fVy\
( <.* * j-
MCD 000011945
r~ 1X
ry
i Note 5
To verify method for n-Butanol
IC-PCDA-2Q
1. Prepare carbon disulfide desorbent solution containing \% 2-butanol and 1.52 mg/ml n dodecare as explained- in Note 2.
2. Prepare a stock solution of 75 mg/ml n-butanol by weighing 7.5000 grams (4 place balance) n-butanol into 100 ml Volumetric flask. Dilute to the mark with desorbing solution described in #1 above and mix thoroughly.
3. Add 1.50 ml desorbing solution described in #1 above to as many Sigma I sample vials as needed and cap.
4. To simulate as in a 10 liter air sample.
0.75 mg 10 liters
25 ppm
Add 10 microliters stock solution from #2 to vial containing 1.5 ml D.S.
0.375 mg 10 liters
12.5 ppm
Add 5 microliters stock solution from #2 to vial containing 1.5 ml D.S.
D.S. - Desorbing Solution from #1
6-20-78 DDH
CK -vQ Oo
CONTINENTAL OIL COMPANY CONOCO CHEMICALS
LAKE CHARLES CHEMICAL PLANT LA3ORATORY
Ethylene Oxide in Air
s
Analysis for Ethylene Oxide from Personal Dosimetry Charcoal Tubes
Safety Precautions
Both carbon disulfide and ethylene oxide must be considered extremely toxic -chemicals. All transfers of these chemicals must be made in a rune hoed with
the blower running. Rubber gloves must be worn whenever any contact with ethylene oxide or carbon disulfide is possible.
I. Sample Preparation:
(See Note 1 for Sample Explanation)
1. Put one dram sample vials in a shallow ice bath for at least 10 minutes.
2. Keeping the vial on ice, add 3.0 t 0.05 ml desorbing solution (carbon
disulfide containing 0.0930 mg/ul 2 methyl pentane) to each vial and cap tigntly immediately. Note 2
. .3* After 30 to 45 minutes transfer enough of the solution from each vial to .an auto sampler vial to fill 2/3 to 3/4 full. Cap each one tightly as soon as it is filled.
4. Hold vials in ice bath until they are run or until they are put in the auto sampler rack just before running.
II. Set up the P&E Sigma I GC (IP 237) as follows:
1. Make sure the Poropak QS column is hooked up to "A" detector (top right connection in oven)
2. If auto-sampler is to be used make sure the Poropak QS column is hooked up to "A" injection port (top left connection in oven)
3. Set both carrier flows on 20 cc/min.
4. Switch "3" detector hydrogen valve lever to "off" position.
5. Switch "A" detector hydregen valve to the "on" position.
.6. Depress "A" detector button.
7. Check "A" detector flame to be sure it is lit. 0. Depress range 10 switch cm front of CC
*9
#
2
III. Set up the Sigma I Computer by pushing keyboard buttons as follows:
1. Void
2. 1 3. Enter
4. Set up 5. I (Refers to GC) 6. Enter 7. 1 (Refers to Detector)
8. Enter
9. 43 (Refers to Method) 10. Enter 11. Enter 12. Enter 13. Sample # (Your Own Sample ?)
14. Enter 15. Sample Name (From Tag by Safety Dept.)
16. Enter (10-20 second delay till light goes out)
The GC system is now set up to run. Sample may be injected when the light by the injection port comes on.
IV. The above procedure will only provide a printout. In order to have a trace (chromatogram), the plotter must be linked to the GC as follows:
On the keyboard: Push keys in the exact order listed:
1. Plot 2. Void
3. Plot 4. 1 (Refers to GC) 5. Enter
6.
. 7. 8. 9.
1 (Refers to Detector)
Enter
0
Enter
0o
Both GC and Recorder are now ready to run. Inject a 2 microliter sarol^ into the correct injection port (the one the correct column is connects with) and press the lighted button just below the injection ports.
*0
To run another sample, wait until the button below the inection ports is relighted-inject sample and press button. Section III, above,only -2; to be repeated if you wish to change the name of the sample.
V. Results will be printed in mg (milligrams) ethylene oxide and are to logged on a sheet designated by the control section. All traces and printouts are to be turned in to the Control Chemist for fon-Mrdir.g t: the Safety Department. No calculations are required by the Analyst.
Notes:
1. Explanation of the Procedure - One dram vials containing charcoal fr tubes used in personal dosimetry sampling will be prepared and tag-"': by the Safety Department. Two vials will :*? submitted for each
tube. One vial will contain charcoal frr: rhe primary section <*; and will be designated by the letter "P" on the tag (Example EO-2-r
and the other vial will contain charcoal from the secondary sortiebt designated by the letter "S" (Example E0-2-S). The contents of
vial will be analyzed exactly the sane way. Blanks (charcoal fro*Tubes) will be submitted with each series of samples and will also
---- - * - -
J 1 ^ A J .1 4 t. <.
^ 'I f\
O
;J
3
Z. Carbon Disulfide OgsorntiorT Solution Preparation - Fresh carbon disulfiie desorptlorrsolution mustBe prepared weekly or with each series of sanoles
(which are scheduled several weeks apart). Sinrra method internal standard values must be corrected for each new batch cf solution. Note 5
Prepare as t-oflows:
1. Tare an unsealed 10 ml sealable sample vial along with the cap and seal to be used.
2. By pipette add 5 nl carbon disulfide to above vial; seal and reweigh. (in heed
3* Using a gastight syringe add 0.25 .01 cc 2-methyl pentane through septum cap to above vial and reweigh.
4. Calculate mg/gram 2 methyl pentane as
wt. of 2 methyl pentane x 1000 wt. of CS + wt. of 2 methyl pentane
mg/gr 2 methyl pentane
5. Calculate the amount of stock solution above needed to prepare ICO ml desorptionsolution with a concentration of 0.09 mg/ml 2 methyl pentane.
100 x 0,09 a 9 mg
Sample Calculation:
If the stock solution contains 6.5 grams total and 0,2200 grams 2 methyl nxane:
Example:
0.2200 x 1000
675(5(50
33.5 mg/gr 2 methyl pentane
MCD 000011949
Wt. of stock solution required is
-iJia-- = .2690 grams 33.5 mg
Note: density of solution is 1.24
a. To a 100 ml volumetric flas.k add stopper and weigh.
30 ml carbon disulfide in the hoed,
b. Using a gastight syringe add 0.22 cc stock solution from the sealed vial from 3 above (in the hood) to the ICO ml volumetric flask from "a" above,
stopper and reweigh.
c. Calculate actual mg/ml as in desorption solution as follows:
(Wt. of Stock Solution)x (mg/gr 2 methyl pc-n 1 -
mg/ml 2 methyl pentane
in* stock solution m--
)
j
LC-PCDA- *3 o
3. To verify the method. i
'
' a. Prepare carbon disulfide desorption solution containing 2 methyl pentane per procedure in Note 2, .
4
0.09 mg/ml
b. Prepare a'stock solution of ethylene oxide In carbon disulfide as follows:
1. Weigh a sealable 10 ml sample vial with a seal and septum on an analytical balance.
2. Add 5 ml carbon disulfide desorption solution (containing 0.09 mg/ml 2 methyl pentane) to the vial, seal securely and reweigh.
3. Using a gastight syringe inject 0.25 cc ethylene oxide liquid through the septum into the vial and reweigh.
Calculate the mg/gr ethylene oxide in the solution as follows:
mg/gr. E.O.
Weight- of E.O.in gr. x 1000
______
Weight of .0. + Weight of Desorption Solution
Each microliter of this solution will then contain
1.24 x mg/qr E.0.s mg E.O.^ul loSo
4. Determine the range of E.O. concentration to be validated.
5. Determine the number of microliters of the solution needed to contain the desired weight of E.O. to be validated by the following.
ul of Sol. = Weight of E.O. desired mg/yjl E.O. in solution-from' 3 above
6. Add the volume ( in fit ) determined (in 5 above) to 3 ml ( 0.05) desorption solution in a sealed vial. Caution: Syringe must be chilled solution will flash out during transfer.
7. Follow procedure outlined in Sections I, II, III, and IV for GC determination
4. GC Parameters
GC: IP '237 (North P&E Sigma I) Method: #43 Column: 7' x 1/8" Poropak QS 80/100 mesh Conditions: Carrier Flow: 20 cc/min
Injection Port Tc-mp: 100C
Detector Temp: 275C Initial Oven Temp: 110 6 rrm Hold Program Rater 60C/min to 200C Range: 10 Method: 43
A?.o & ooo
LC-PCDA- 2> 0 5 r
5. Mew internal standard values must be entered in Sigma method 43 each time fresh desorption solution is made because the volatility of the reagents make it almost impossible to weigh in exact amounts of 2 methyl pentane. The values must be changed in both the standard weight.line and in the data table or results will be in error.
6. Desorption efficiencies of approximately 60% in the range of interest were obtained by this method. A short study using dry ice to cool desorption solution, one dram vials containing charcoal, and the sigma auto sampler vials indicated recoveries of over 80% in the 0.2 to 0.9 mg range
o
Issued: 10-30-78 by DDH Approved by:
LJ