Document vBJgNKQdoq5bKkwRDryVZmek6

.-.~4 NIOSH MANUAL of ANALYTICAL METHODS U.S. DEPARTMENT OF HEALTH, EDUCATION, AND WELFARE Public Health Service Center for Disease Control National Institute for Occupational Safety and Health 000281 refill PLAINTIFF'S EXHIBITvS CEL-292^ ORGANIC SOLVENTS IN AIR Physical and Chemical Analysis Branch Analytical Method Analyte Organic Solvents (See Table 1) Matrix: Air Procedure: Adsorption on charcoal desorption with carbon disulfide, GC Date Issued: 9/15/72 Date Revised: 7/15/74 Method No: Range: P&CAH 127 For the specific compound, refer to Tables I&II Precision 10.5X RSD Classification: See Table 1 1. Principle of the Method 1.1 A known volume of air le drawn through a charcoal tube to trap the organic vapors present. 1.2 The charcoal in the tube Is transferred to a small, graduated test tube and desorbed with carbon disulfide. 1.3 An aliquot of the desorbed sample Is injected into a gas chromato graph. 1.4 The area of the resulting peak is determined and compared with areas obtained from Che Injection of standards. 2. Range and Sensitivity The lower limit in mg/sample for the specific compound at 16 x 1 attenuation on a gas chromatograph fitted with a 10:1 splitter Is -hown in Table 1. This value can be lowered by reducing the attenuation or by eliminating the 10:1 splitter. 3. Interferences 3.1 When the amount of water In the air Is so great that condensation actually occurs In Che tube, organic vapors will not be trapped. Preliminary experiments Indicate that high humidity severely decreases the breakthrough volume. 3.2 When two or more solvents are known or suspected to be present la che air, such Information Including their suspected Identities, should be transmitted with the sample; since with differences la polarity, one may displace another from the charcoal. 127-1 3.3 It oust be emphasized that any coapound which has the ease retention tiae as the specific coapound under study at the operating conditions described In this method Is an Inter ference. Hence, retention tiae data on a single eolurm, or even on a number of coluans, cannot be considered as proof of chemical identity. For this reason It Is important that a sample of the bulk solvent(a) be submitted at the same time so that ldentlty(les) can be established by other means. 3.4 If the possibility of interference exists, separation conditions (column packing, temperatures, etc.) must be changed to circum vent the problem. 4. Precision and Accuracy 4.1 The mean relative standard deviation of the analytical method la 8Z. (Ref. 11.4). 4.2 The mean relative standard deviation of the analytical method plus field sampling using an approved personal sampling pump la 10Z (Ref. 11.4). Part of the error associated with the method lo related to uncertainties In the sample volume collected. If a more powerful vacuum pump with associated gas-volume integrating equipment Is used, sampling precision can be Improved. 4.3 The accuracy of the overall sampling and analytical method la 10Z (NIOSH's unpublished data) when the personal sampling pump la calibrated with a charcoal tube In the line. 5. Advantages and Disadvantages of the Method 5.1 The sampling device is small, portable, and Involves no liquids. Interferences are minimal, and most of those which do occur can be eliminated by altering chromatographic conditions. The tubes are analyzed by means of a quick. Instrumental method. The method can also be used for the simultaneous analysis of two or more solvents suspected to be present in the same sample by simply changing gas chromatographic conditions from isothermal to a temperatureprogrananed mode of operation. 5.2 One disadvantage of the method la that the amount of sample which can be taken is limited by the number of milligrams that the tube will hold before overloading. When the sample value obtained for the backup section of the charcoal trap exceeds 25Z of that found on the front section, rh* possibility of somplc loss exists. During sample storage the more volatile compounds will migrate throughout the tube until equilibrium is reached (33Z of the sample on the backup section). 127-2 000283 5.3 Furthermore, the precision of the nethod Is Halted by the reproducibility of the pressure drop scross the tubes. This drop vill sffeet the flow rate and cause the volume to be imprecise, because the puop is usually calibrated for one tube only. 6. Apparatus 6.1 An approved and calibrated personal-sampling pump for personal samples. For an area sample any vacuum pump whose flow can be determined accurately at 1 liter per minute or leas. 6.2 Charcoal tubes: glass tube with both ends flame sealed, 7 cm long with a 6-mm 0.0. and a 4-sn Z.D., containing 2 sections of 20/40 mesh activated charcoal separated by a 2-am portion of urethane foam. The activated charcoal is prepared from coconut shells and is fired at 600*C prior to packing. The absorbing section contains 100 mg of charcoal, the backup section 50 mg. A 3-bbb portion of urethane foam is placed between the outlet end of the tube and the backup section. A plug of silylated glass wool is placed in frontof the absorbing section. The pressure drop across the tube must be less thnn one inch of mercury at a flow rate of 1 1pm. 6.3 Cas chromatograph equipped with a flame ionization detector. 6.4 Column (20 ft x 1/8 in) with 101 FFAP stationary phase on 80/100 mesh, acid-washed DMCS Chromosorb W solid support. Other columns capable of performing the required separations may be used. 6.5 A mechanical or electronic integrator or a recorder and some method for determining peak area. 6.6 Class stoppered micro tubes. The 2.5-ml graduated microccntrifuge tubes arc recommended. 6.7 Hamilton syringes: 10 wl, and convenient sizes for making standards. 6.8 Plpets: 0.5 nt delivery plpets or 1.0 ml type graduated in 0.1 ml increments. 6.9 Volumetric flasks: 10 mi or convenient sizes for making standard solutions. 7. Reagents 7.1 Speetroquality carbon disulfide (Matheson Coleman and Bell) 127-3 000284 7.2 Sample of the specific compound under study, preferably chromatoquality grade. 7.3 Bureau of Mines Grade A helium. 7.4 Prepurified hydrogen. 7.5 Filtered compressed air. Procedure 6.1 Cleaning of Equipment. All glassware used for the laboratory analysis should be detergent washed and thoroughly rinsed with Cap water and distilled water. 8.2 Calibration of Personal Pumps. Each personal pump must be calibrated with a representative charcoal tube in the line. This will minimize errors associated with uncertainties in the sample volume collected. 8.3 Collection and Shipping of Samples 8.3.1 Immediately before sampling, the ends of the tube should be broken to provide an opening at least one-half the Internal diameter of the tube (2mm). 8.3.2 The smaller section of charcoal is used as a back-up and should be positioned nearest the sampling pump. 8.3.3 The charcoal tube should be vertical during sampling. 8.3.4 Air being sampled should not be passed through any hose or tubing before entering the charcoal tube. L 3.5 The flow, time, and/or volume must be measured as accurately as possible. The sample should be taken at a flow rate of 1 tpm or less to attain the total sample volume required. The minimum and maximum sample volumes that should be collected for each solvent are shown in Table 1. The minimum volume quoted must be collected if the desired sensitivity is to be achieved. 8.3.6 The temperature and pressure of the atmosphere being sampled should be measured and recorded. 8.3.7 The charcoal tubes should be capped with the supplied plastic caps immediately after sampling. Under no circumstances should rubber caps be used. 127-4 8.3.8 One Cube ehould be handled In Che same Banner ae Che ample Cube (break, aeal, and Cransporc), excepc chac no air la aampled Chrough chle Cube. Thla Cube ehould be labeled aa a blank. 8.3.9 Capped cubea ehould be packed tightly before chey are ahlpped Co minimize cube breakage during chipping. 8.3.10 Samples of Che suspecced aolvenc(a) ehould be eubalcced Co Che laboratory In conCainera furnished by NIOSH for auch purpose. These liquid bulk samples should not be transported In Che same container as Che samples or blank Cube. If possible, a bulk air sample (at least 501 air dravn through tube) should be shipped for qualitative Identification purposes. ' 6.4 Analysis of Samples 6.4.1 Preparation of Samples. In preprstlon for analysis, each charcoal Cube Is Bcored vlch a file In fronC of che first section of charcoal and broken open. The glass wool Is removed and discarded. The charcoal In Che flrsc (larger) section Is transferred co a small stoppered test cube. The separating section of foam Is removed and discarded; che second section is transferred to another test tube. These tvo sections are analyzed separately. 8.4.2 Desorption of Samples. Prior to analysis, one-half ml of carbon disulfide is pipetted Into each test cube. (All work with carbon disulfide should be performed in a hood because of Its high toxicity.) Tests indicate chat desorption Is complete in 30 minutes if the sample is stirred occasionally during this period. The use of graduated glass-stoppered, mlcrocencrlfuge tubes Is rccomended so that one can observe any apparent change in volume during the desorption process. Carbon disulfide is a very volatile solvent, so volume changes can occur during the desorption process depending onthe surrounding temperature. The initial volume occupied by the charcoal plus the 0.5 mt CS, should be noted and corres ponding volume adjustments should be made whenever necessary just before CC analysis. 8.4.3 CC Conditions. The typical operating conditions for the gas chromatograph are: 1. 85 ec/nln. (70 psig) hell:;;.; carrier gas flow. 2. 65 cc/mln. (24 pslg) hydrogen gas flow co detector. 3. 500 cc/mln. (50 pslg) air flow to detector. 4. 200*C Injector temperature. 127-5 5. 200*C manifold temperature (detector) 6. Isothermal oven or column temperature - refer to Table 1 for specific compounds. 8.4.4 Injection. The first step in the analysis is the injection of the sample into the gas chromatograph. To eliminate difficulties arising from blowback or distillation within the syringe needle, one should employ the solvent flush injection technique. The 10 ut syringe Is first flushed with solvent several times to wet the barrel and plunger. Three microlltcrs of solvent are drawn into the syringe to increase the accuracy and reproducibility of the injected sample volume. The needle is removed from the solvent, and the plunger is pulled back about 0.2 ul to separate the solvent flush from the sample with a pocket of air to be used as a marker. The needle is then Immersed in the sample, and a 5-ui aliquot is withdrawn, taking into consideration the volume of the needle, since the sample in the needle will be completely injected. After the needle is removed from the sample and prior to injection, the plunger is pulled back a short distance to minimize evap oration cf the sample from the tip of the needle. Duplicate Injections of each sample and standard should-be made. No more than a 32 difference in area is to be expected. 6.4.5 Measurement of area. The area of the sample peak is measured by an electronic integrator or some other suitable form of area measurement, and preliminary results are read from a standard curve prepared as discussed below. 8.5 Determination of Desorption Efficiency 8.5.1 Importance of determination. The desorption efficiency of a particular compound can vary from one laboratory to another and also from one batch of charcoal to another. Thus, it is necessary to determine at least once the percentage of the specific compound that is removed in the desorption process for a given compound, provided the sane batch of charcoal is used. The Physical and Chemical Analysis Branch of NIOSH has found that the desorption efficiencies for the compounds in Table 1 are between 81Z and 1002 and vary with each batch of charcoal. 8.5.2 Procedure for determining desorption efficiency. Activated charcoal equivalent to the amount in the first section of the sampling tube (100 mg) is measured into a 5cm, 4-tma I.D. glass Cube, flame-sealed at one end (similar to cooaerclally available culture tubes). ihis charcoal must be from the same batch as that used in obtaining the samples and can be obtained from unused charcoal tubes. The open end is capped 127-6 % with ParafiIn. A known anount of the eonpound la injected directly into the activated charcoal with a microliter syringe, and the tube ia capped with more Parafllm. The amount injected la uaually equivalent to that preaent in a 10-liter aample at a concentration equal to the federal etandard. At leaat five tubea are prepared in thla manner and allowed to atand for at leaat overnight to assure complete abosrptlon of the specific compound onto the charcoal. These five tubes are referred to aa the samples. A parallel blank tube should be treated in the same manner except that no sample is added to it. The sample and blank tubes are desorbed and analyzed in exactly the same manner aa the sampling tube described in Section 8.3. IVo or three standards are prepared by injecting the same volume of compound into 0.5 mt of CS2 with the same syringe used in the preparation of the sample. These are analyzed with the samples. The desorption efficiency equals the difference between the average peak area of the samples and the peak area of the blank divided by the average peak area of the standards, or Area aample - Area blank desorption efficiency - Area standard 9. Calibration and Standards It is convenient to express concentration of standards in terms of mg/0.5 mt CS2 because samples arc desorbed in this amount of CSj. To minimize error due to the volatility of carbon disulfide, one can inject 20 tines the weight into 10 mt of CS2. For example, to prepare a 0.3 mg/ 0.5 mt standard, one would inject 6.0 mg into exactly 10 mi of CSj in a glass-stoppered flask. The density of the specific compound is used to convert 6.0 mp, into mlcrollters for easy measurement with a nicrolitcr syringe. A scries of standards, varying in concentration over the range of interest, is prepared and analyzed under the same CC conditions and during the sane time period as the unknown samples. Curves are estab lished by plotting concentration in mg/0.5mt versus peak area. NOTE: Since no internal standard is used in the method, standard solutions must be analyzed at the 6ame time i!:Jt the sample analysis is done. This will minimize the effect of known day-to-day variations and variations during the same day of the FID response. 127-7 000288 fF 10. Calculations 10.1 The weight. In ag, corresponding to esch peek area is read froa the standard curve Cor the particular coapound. No voluae corrections are needed, because the standard curve is based on ag/0.5 ml CSj and the voluae of sample injected is identical to the voluae of the standards injected. 10.2 Corrections for the blank suit be made for each saaple. Correct mg " mg# - mgfe where: mg8 " 8 found in front section of sample tube mg^ " ag found in front section of blank tube A similar procedure is followed for the backup sections. 10.3 The corrected aaounts present in the front and backup sections of the same saaple tube are added to determine the total measured amount In the sample. 10.4 This total weight is divided by the determined desorption efficiency to obtain the total ag per saaple. 10.5 The volume of air sampled is converted to standard conditions of of 25*C and 760 mm Hg. P 298 v.-v* 760 T+273 where: Vs " volume of air in liters at 25*C and 760 on Hg V volume of air in liters as measured P Barometric pressure in mm Hg T " Temperature of air in degree centigrade 10.6 The concentration of the organic solvent In the air sampled can be expressed in mg per m3, which is numerically equal to pg per liter of air mg/m3 pg/1 total mg (Section 10.4) x 1000 (uc/mg) 10.7 Another method of expressing concentration is ppm, defined aaplof compounds per liter of air pom pt of compound/Vt ppm ui of compound 24.45 MV where: 24.45 " aolar voluae at 25*0 and 760 mm Hg MV molecular weight of the compound (Table 1) 127-8 t&3d30 000289 II. Reference* 11.1 White, L.D., D.C. Taylor, P.A. Mauer, and R.E. Kupel, "A Convenient Optimised Method for the Analysis of Selected Solvent Vapors In the Induatrlal Atmosphere," Amer. 2nd, tty?. Assoc. 31:225 (1970). 11.2 Toung, D.M. and A.D. Crowell, Physical Adsorption of Gaaea, Buttervortha, London, 1962, pp. 137-146. 11.3 Federal Register, 37 (#202), 22139-22142 (October 18, 1972). 11.4 NIOSH Contract HSM-99-72-98, Scott Research Laboratories, Inc., "Collaborative Testing of Activated Charcoal Sampling Tubea for Seven Organic Solvents," pp. 4-22, 4-27 (1973). 127-9 ERS ASSOCIATED WITH P&CAB A3ALYTICAL METHOD WO. 127 may d is p la c e more 4-* u0 9 41 H n poH 41 V *o4 pH M o 9 eO 0 /"l a N 4 w 0u 0 044 40*4 V U C0W *04 40* 044 0 tj 00 JUG ap 0 0 0 40 404 P"V 0 *40440 0 tj 0a >0 H0 pH en C0 4001 pHH)H ^0 41 ^00V) s c (/> > "O O9c o *04 i ^< >a o3 HU3 4ce4 jus > .e * gH. o C *oo1 tt O C m41 a 41 js H u C <8 C4 u ^*H a *; IOft IMou m 404 * 0 0a 00 4i H oC O> Cl JZ 40 W 9 0 O 44 Cl > 041 hW 0 C 00 0C-H4 0^ h6H *09 HH W 44 0 pH 5 =* g.44 0 oH 4J *0o 6w 0 W M >oo 0 -C 3*4 44 o o u 0 0 0 a O 9 "O A0J 0 0 co *4 44 40U-* r-- e 0 U **044 0 0 jz 400 c9 oO. TJ U e e 0 pH 44 H 0 V 9 0* 0 00 44 e p9H 0 u c o o N ou * 44 0 p9OH >s 44 hH po4 a 9 r0H p4 44 0 iH 0 o 0 C o 44 0 44 w4 pH e >o u e 4s 00 0 9 wO -e u 44 *0H W 0 > 0 X JZ > 00 0 *4 o J= r0H O 44 44 0 .C* a0 *94 440>o 0 0 OCu9O U9f* 00u0 .wC0 *0 00- aao.. O J4aOaC4. 40ui 0a* UeH 00 se 4>-4 om um wo 0 o 9 /-s ^ 0 0 000291 a TABLE XX CHEMICALS WHICH HAVE CREATE* THAN 80Z DESORPTION EFFICIENCY BUT HAVE NOT BEEN THOROUGHLY TESTED BY NIOSH Class E (Proposed) Acrylonitrile Allyl glycidyl ether n-Amyl acetate 2-Butoxyethanol n-Butyl acetate n-Butyl alcohol n-Butylglycidyl ether Chlorobenzene Cyclohexane Cyclohexanone o-Dlehlorobenzene p-Diehlorobenzene Diethyl ether N.N-Dlmethyl aniline Eplchlorohydrin 2-Ethoxylethyl acetate Ethyl acetate Ethylbenzene Ethyl butyl ketone Fufural Heptane . Hexane Isoamyl acetate Xsobutyl acetate Xsobutyl alcohol Xsoctane Isophorone Isopropyl acetate Isopropyl glycidyl ether 26-Lutldine Methyl acetate Methyl acrylate Methyl n-butyl ketone Methyl ethyl ketone Methyl isobutyl ketone Methyl methacrylate a-Methyl styrene .p-Methyl styrene n-Octane 3-Octanone Pentane 2-Pcntanone a-pinene n-Propyl acetate 1,1,2,2-Tetrachloroethane Tetrahydrofuran Trichlorotrifluorocthane (Freon 113) Recommended Sample Size 101 l i l 127-11 000292 Che mer 1 Plant Method A -1 -30 October, 19 7 3 Page 1 of 8 FORMALDEHYDE IN AIR BY CHROMOTROPIC ACID Scope: This method describes a procedure for the determination of formalde hyde in air. Compounds yielding formaldehyde under color develop ment conditions will interfere with the chromotropic acid test; they are 1) paraformaldehyde, 2) trioxane, 3) Celcon, and 4) sugars (PE and TMP). These compounds will generally be present as particulates and can be eliminated by placing a filter on the sample inlet. * Acetone, acrolein, acetaldehyde, diacetone alcohol, methanol and methylal^ can interfere if present in sufficient concentration (Table I). The method is sensitive to 375 micrograms per cubic meter of air (0. 3 ppm) based on a sample collection rate of 1 SCFH. It takes advantage of the fact that formaldehyde forms a purple color with 1,8-dihydroxy-3, 6-naphthalene disulfonic acid (chromotropic acid) in the presence of sulfuric acid. The absorbance is measured at 580 my against a reagent blank. Equipment: 1. Mine Safety Appliances Company Permissible Portable Pump Model C or equivalent. 2. 30 ml gas scrub bottles with spargers 3. 100 and 25 ml volumetric flasks 4. 1 and 5 ml Mohr Pipette 5. 2,3,5 and 10 ml volumetric pipettes .6 10 and 50 ml burettes 7. Beckman Model DB-C Spectrophotometer, 1 cm silica cells .8 Ice bath 9. Boiling water bath 10. Millipore Field Monitors (0. 8 u pore size paper): Aerosol type Cat. No. MAW Go 37Ao Reagents: 1. lOOu g/ml Stock Solution - Enough fine paraformaldehyde powder to as nay 100 mg of formaldehyde (HCHO), 10 gr.tms of sodium bisulfite, dissolve and dilute to one liter with water. Stable for at least two months. 000293 C h e m c e 1 Plant Method A-1-30 October, 1 9 7 3 Page 2 of 8 2. 0.5% Chromotropic Acid (salt) solution - Dissolve 0.5 gram of 4, 5-dihydroxy-2, 7-naphthalene disulfonic acid or I, 8-dihydroxy-3, 6-naphthalene disulfonic acid (disodium salt) in 100 ml of water. Keep refrigerated, stable for two weeks. 3. Reagent grade concentrated sulfuric acid. 4. 1% aqueous sodium bisulfite. Standard Curve Preparation: 1. Label nine 100 ml volumetric flasks 0. 1, 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, and 10. 0 ug/ml respectively. 2. With pipettes add 0. 1, 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, and 10. 0 ml of the 100 ug/ml HCHO Stock Solution respectively, to the 100 ml volumetric flasks. 3. Dilute to the mark with 1% aqueous bisulfite solution, stopper and mix well. Save these standard solutions. 4. Label ten 25 ml volumetric flasks blank, 0.1, 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7. 0 and 10. 0 yg/ml respectively. 5. Pipette 5 ml aliquots of the standard solutions to their respective 25 ml volumetries. Use 5 ml of bisulfite solution for the blank. 6. From a 1 ml pipette add 0. 5 ml of the chromotropic salt solution to the 25 ml volumetric flasks. Swirl to mix. 7. From a 50 ml burette add 12. 5 ml of concentrated sulfuric acid while swirling the 25 ml flask in an ice bath. Remove the flask from the bath and stopper. Swirl again to ensure complete mix ing. 8. Place the 25 ml flasks in a boiling water bath for 15 minutes. Time required for the water to resume boiling should be in cluded in the '5 minutes.9 9. Cool the 25 ml flasks in an ice bath; dilute to volume with water and mix. After cooling to room temperature add more water if necessary. 000294 Che me el Plant Method A -1 - 30 October, 19 7 3 Page 3 of 8 .10 Record the absorbance at 580 mp against the blank using 1 cm silica cells. .11 Plot absorbances versus pg/ml as in Figures I and II. Procedure: 1. To each of two gas scrub bottles add 10 ml of bisulfite solution. Close openings until ready to sample. .2 Calibrate the Mine Safety pump to deliver 1 SCFH with impingers attached. 3. At sampling site set up the collector as shown in Figure III. Note: If paraformaldehyde, trioxane, Celcon and sugars (PE and TMP) are suspected to be present as particulates place the filter assembly before the sample unit. 4. Start airflow and continue scrubbing for at least 15 minutes. If the HCHO concentration is low, a longer scrub time will be required. 5. After sampling is complete disconnect the assembly and close the openings of the scrub bottles. .6 Pipette a 5 ml aliquot of the solution from the scrub bottle into a 25 ml volumetric flask and proceed as in Steps 6-10 under Standard Curve Preparation. 7. From the standard curve determine the concentration in pg/ml. 8. Calculate the ppm HCHO in Air from the following formula. ppm HCHO*=----- 3-------------------------------V.&/M1 X-IQ-M1---------- ------ ---------------------lft3/hr x Scrub Time in hrs x 0. 02595 M^/ft-* x 1293 g/K ppm HCHO x 1293 = pg HCHO/cubic meter of air < Adjusted to 2 5*C 000295 Chemcel Plant Method A-l-30 October, 1973 Page 4 of 8 Refercncea: 1. Sander*, F. D., "The Determination of Formaldehyde in Air", FDS2-71, April 2, 1971, Chemcel Lab File 261.0. 2. Walker, J. F., Formaldehydc. Reinhold, 1964 pp 470-1. 000296 VJ0JO TABLE I Chemcel Plant Method A -1 - 30 October, 1973 Page 5 of 8 Effect of Interference on Formaldehyde Analyses 3 ppm HCHO (HCHO Std. ) 3 ppm DAA + HCHO Sid. 3 ppm DMK + HCHO Std. 3 ppm AcH + HCHO Std. 3 ppm Acrolein + HCHO Std. 3 ppm Meal + HCHO Std. 3 ppm MeOH f HCHO Std. A mixture of 3 ppm each of HCHO, DAA, DMK, AcH, Acrolein, Meal, and MeOH ppm as HCHO 3.00 3. 00 % Increase as HCHO ... 0. 0 3. 05 1.7 3.20 6. 7 3. 70 23. 3 3. 85 5-00 28. 3 66. 7 1 3. 65 20. 7 1. Scrubbing efficiency of MeOH in 1% bisulfite solution is negligible. 000297 ,L.I J iu 08S 1* aauq40Bqv C a lib ra tio n C urve fo r C h ro m o tro p ic A c id and F orm aldehyde M g/m l of F o rm a ld e h yd e 000298 oi i 00 o ro sO d ITi d o o <*VJ d o C a lib ra te n C urve fo r C h ro m o tro p ic A c id and F orm aldehyde No. u g /m l of Form aldehydi F orm aldehyde A ir S am pler Che me el Plant Method A-1-30 October, 19 7 3 Page 8 of 8 000300