Document p24gm47NYEEvwpmJBQ7LQ74k7
Regulatory needs Exchange of a PFAS containing manufacturing device (e.g. tube, filter ...) with a PFAS free alternative
Introduction
Many medical devices may be affected by the restriction of PFAS in different ways. Even if there are no substantiated data available yet, it can be expected that a high number, if not even most, of the medical devices on the market are affected.
Flow-chart
no 7a new constituents identified 8b. Possible without further tests
1. Availability of alternative on the market
2. Technical feasibility (lab)
3. Manufacturing feasibility (production, 3 batches)
4. Detailed verification of performance
5. short-term stability test
6. start of long-term stability stest
7. Chemical characterization of the product
yes 8. Evaluation of biocompatibility
yes
at any time
9. Testing in-vitro and/or in-vivo
10. Biocompatibility assessment
11. Complete Risk assessment (incl. favorable long-term stability)
12. Check of all Safety and performance requirements
13. possible involvement of notified body
14. Declaration of conformity
2a not feasable 3a not feasable 4a not fulfilled 5a not fulfilled 6a not fulfilled
8a. not feasable
9a not fulfilled 10a not fulfilled 11a not fulfilled 12a not fulfilled 13a not accepted
Discussion
Step Discussion
Duration (month)
Min Max Comments
1
Could be 6.5y or 13.5 y after
publication of the Restriction
depending on the derogation
2
First check of feasibility
1
4
Depending on the complexity of the
replacement and possibly the
number of affected products
2a At practically any point it could be
found that no alternative is
feasible (the requirements are
not fulfilled)
3
Check of feasibility in the 2
9
Depending on complexity of the
manufacturing environment.
replacement, the availability of the
Usually, 3 batches must be
manufacturing devices (e.g. use of
manufactured
the devices for other products as
well) and possibly the number of
affected products
4
Even if the basic performance was (1) (3)
Depending on the complexity of the
checked in step 2 a detailed
replacement and possibly the
performance check of products
number of affected products -
out of manufacturing process is
included in the timing for step 5
required by MDR
5
Can be done in parallel with 4; 3
6
3 months after the start of the
however at least 3 months of
storage (in case of multiple products
stability are usually needed for a
this timeline is defined by the last
first evaluation
storage start)
6
Depends on the shelf-life of the (9) (24) If the shelf life exceeds the timeline
product. Usually medical devices
for all other further steps, they are
need a shelf-life of 18 - 48
determining the timeline
months to be commercialized
(favorable test results usually for
9 - 24 months at elevated
temperatures).
Long-term
stability tests can be started in
parallel to step 4 and 5
7
Requirement of ISO 10993-1 and 1
6
Depending on the complexity of the
10993-18
product, the needs for preliminary
tests, development of extraction and
analytical methods, availability of lab
capacity at specialized contract labs
7a If no new constituents are
identified in the product only the
steps 11 - 14 are further relevant
8
Based on the new constituents a 1
3
Depending on the complexity of the
first biocompatibility evaluation
replacement and possibly the
should be done to verify the need
number of affected products
of testing and develop a test plan
(ISO 10993-series)
8b If no further tests are needed
step 9 is obsolete
9
Finalization of test plans, 3
18
Depending on the tests required
preparation of test specimens,
based on the type of medical device,
clarification of test procedures
the concerns arising from new
with contract labs, testing and
constituents (see ISO 10993-series),
finalization of test reports by
the number of products affected by
contract labs (ISO 10993-series; for respiratory pathways devices instead of ISO 10993series the requirements of ISO 18562-series apply
the exchange, and the capacities of contract labs
10 Based
on
chemical 1
2
In case of an external toxicologist,
characterization, data regarding
tolerable intake of constituents
and test results a biocompatibility
assessment is prepared by a
trained person (e.g., toxicologist)
11 Risk assessment is done based on 1
2
ISO 14971 by a group of product
and safety experts
12 Final check of all safety and 1
2
performance requirements of
MDR Annex I by a group of
regulatory and product specialists
13 In some cases (e.g., medical 2
6
devices class III) the full technical
documentation (manufacturing
and quality documentation,
documents mentioned above in
the flowchart) must be verified
first by the notified body (MDR)
14 If no involvement of the notified 1
2
body is needed, the Declaration
of Conformity can be prepared by
the
manufacturer
after
compilation of the technical
documentation
some discussion is needed in front, during and after development of the assessment
Depending on the complexity of the replacement and possibly the number of affected products Depending on the complexity of the replacement and possibly the number of affected products
Depending on the complexity of the replacement and possibly the number of affected products and the number of dossiers submitted at the same time to the notified body by different manufacturers.
Depending on the complexity of the replacement and possibly the number of affected products
Conclusion
For the case of replacement of a PFAS-containing manufacturing device with an PFAS-free alternative, no fix timeline can be established. A minimum and a maximum amount of time for such a process from the regulatory perspective is given below. For a better understanding, a rough, schematic project plan is added at the end of the document
Those timelines are realistic if: - all steps lead directly to a favorable result - there is only one or a very low number of medical devices affected by the replacement - there are no multiple changes of manufacturing devices that would complicate the process - no new constituents are identified by the chemical characterization - no preliminary involvement of the notified body is needed
an absolute minimum of about 12 -14 months is needed. If a longer shelf life as 18 month is required by customers, this minimum period will increase up to 27 months.
Those timelines are not realistic if: - many medical devices are affected by the replacement - new constituents are identified by the chemical characterization - chemical characterization is complex - there's a requirement of long-term biological testing (e.g. subchronic toxicity (9 months invivo) or implantation (min. 9 months in vivo)
Then the time period needed to fulfill the regulatory needs will increase to about 50 months (more than 4 years).
In this calculation some aspects are not considered:
- if the replacement(s) fails due to unfavorable results at any time point, new replacements must be found. In a worst-case scenario this can more than double the time needed.
- if many manufacturers need at the same time chemical characterization and/or biological testing the bottleneck could be the capacity of the contract labs, leading to delays of some months to 1 year or longer.
- if many manufacturers need to involve the notified bodies at the same time for verification of such replacements in the technical documentation, the number of notified bodies may become a bottleneck as well (this fact already led to the extension of the time frame for implementation of MDR). Additional time up to some months or a year would be possible.
- in some cases, especially (but not limited to this) if a PFAS-constituent of the medical device is replaced, additionally to the regulatory steps shown in the flowchart above, a clinical trial could be necessary. In these cases, additionally to the given timeline additionally 2 - 5 years would be needed for the replacement, increasing the time needed to 36 - 115 month (3 years - nearly 10 years).
Based on these considerations the derogation timeline of 13.5 years for medical devices is considered very short, especially considering that the replacement would be available probably at the end of the derogation period for the manufacturing devices (probably 6.5 years, in some cases possibly 13.5 years). For this reason, for highly regulated products like medical devices a longer derogation timeline is required. We propose an additional derogation time of 5 - 10 years or at least an additional derogation time of 3 - 5 years for medical devices of class I, IIA and IIB and 5 - 10 years for medical devices of class III or devices with replacement of a PFAS-constituent. Even this will not cover fully the absolute worst cases that may occur. Only by such a longer derogation time the supply of vital medical devices to the healthcare system can be ensured.
Schematic project plan for replacement of a PFAS-containing manufacturing device
Availability of replacement ( < 6,5 y or< 13,5 y after entering in force )
Step Y M
1 2 3 4 5 6 7
1
2
3
4
5
6
7
8
9
10
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120
Explanation absolute best case best case possibly needed (best case) worst case possibly needed (worst case)
combined duration best case combined duration best case (with clinical trials) combined duration worst case combined duration wordt case (with clinical trials)
8
9
10
11
12
13
14
7b + 11-14 + Clin.
Trial
Note: In case of the replacement of a PFAS-containing constituent the time-line could be even somewhat longer.