INDUSTRIAL HYGIENE FUME HOOD EVALUATION AT PROMIX
On December 30, 1986, the laboratory hood at Promix was evaluated for face air velocity using the Alnor Model 8525 Thermo Anemometer. Laboratory hoods are classified according to their hood face velocities with the sash at an 18-inch opening. Laboratory hoods are grouped into four classifications:
Class A (high toxicity use) Carcinogenic, toxic, highly flammable.
125 - 150 ft/min average hood face velocity. 100 ft/min minimum reading.
Class B (normal lab use) Toxic or flammable
100 ft/min average face velocity. 80 ft/min minimum reading.
Class C (low toxicity use) Non-toxic or non-hazardous compound.
75 ft/min average face velocity. 50 ft/min minimum reading.
Class D (Inadequate)
The lab hood was determined to be Class B, but just barely so. The filter should be checked to see if it is plugged and the belts should be checked to make sure they are not slipping.
William E. Ledford Industrial Hygiene