V Working Document for Committee Use Only Hot for Publication, Except as Expressly Approved by the Chairman of the Committee or the Managing Director of the Society Draft No. 6 6/18/75 J. Lichtenberg TENTATIVE METHOD OF TEST FOR POLYCHLORINATED BIPHENYLS (PCB) IN WATER1 ASTM DESIGNATION-T 1, Scone and Application 1.1 This method covers the determination of certain polychlorinated biphenyls (PCB's) Including: Aroclors2 1221, 1232, 1242, 1248, 1254, 1260, and 1016. 1.2 The method Is an extension of the ASTM Proposed Tentative Method of the Test for Organochlorine Pesticides in Water, D 3086-72^T, It is designed so that determination of both the PCB's and the organochlorine pesticides may be made on the same sample. 2. Applicable Documents 2.1 ASTM Standards: 3 D510 Sampling Water 3 D1129 Definitions of Terms Relating to Water D1192 Specifications for Equipment for Sampling Water and Steam^ 3 D1193 Specification for Reagent Water 3 D1496 Sampling Homogenous Waste Water E260 Recommend Practice for General Gas Chromatography Procedures* * This method is under the jurisdiction of the ASTM Committee D-19 on Water 2 Trade name of the Monsanto Co., St. Louis, MO 3 Appears In this publication a 1974 Annual Book of ASTM Standards, Part 42 MGNS 047456 E355 Recommended Practice for Gas Chromatography Terms and 4 Relationships. 3. Summary 3.1 The pen's and the organochlorine pesticides are co-extracted by liquid-liquid extraction end, insofar as possible, the two classes of compounds separated from one another prior to gas chromato graphic determination. A combination of the standard Florisil 5 column cleanup procedure and a silica gel mlcrocolumn separation procedure (1,2) are employed. Identification is made from gas chromatographic patterns obtained through the use of two or more unlike columns. Detection and measurement is accomplished using an electron capture, microcoulometric, or electrolytic conduct ivity detector. Techniques for confirming qualitative identi fication are suggested. The detection limit is approximately 100 ng/1 for the PCS mixtures (Arodors) listed in 1.1 when analyzing 1-llter of sample. 4. Significance 4.1 The extensive end widespread use of PCB's has resulted in their presence in all parts of the environment. Like the organochlorine pesticides, the PCB's are very persistent. While they are generally less toxic than the organochlorine pesticides, they do have lethal effects on mammals, birdB, fish, and other aquatic animals. Thus, we must identify and quantitate the PCB's present in the environment. Because of their cumulative nature and level of occurrence, the method must be capable of measuring quantities less than 1 jig/1 of the PCB's in water.* *Plorisll TM of, and available fron Floridin Co., 2 Gateway Center, Pittsburg, PA 15222 MONS 047457 2 5. Definitions 5.1 For definitions of terms used in these methods refer to ASTM j definitions D1129, Terms Relating to Water and ASTM Recommended Practices E-355 for Gas Chromatography Terms and Relationships^. 6. Interferences - 6.1 Certain phthalate esters, organophosphorus pesticides, and elemental sulfur will lnterfer when using electron capture for detection. 6.2 Organochlorlne pesticides and other halogenated compounds constitute interferences in the determination of PCB's. Most of these are separated by the method described below. However, certain compounds, if present in the sample, will occur with the ?CB*s. Included are sulfur, heptachlor, aldrln, DDE, chlordane, mirex, and to some extent o,pf-DDT and p,p'-DDT. 7. Apparatus . 7.1 Kuderna-Danish (K-D) Glassware 7.1.1 Snyder Columns - three ball (macro) 7.1.2 Evaporative flasks - 500 ml 7.1.3 Receiver Ampuls - 10 ml, graduated 7.1.4 Ampul Stoppers 7.2 Chromatographic Column - Chromaflex^ (400 mm long X 19 mm ID) with coarse fritted plate on bottom and Teflon stopcock; 250 ml reservoir bulb at top of column with flared out funnel shape at top of bulb - a special order (Kontes-42540-9011). 6 Chromaflex, trademark of Kontes Glass Co., Vineland, NJ 08360 3 HONS 047458 7.3 Chromatographic Column - borosillcate glass (approximately 400 tnm long X 20 mm ID) with a coarse fritted plate. 7.4 Micro Column borosillcate glass - constructed according to Figure 1. 7.5 Capillary pipets disposable (5-3/4 in.) with rubber bulb. 7.6 Low Pressure regulator (0 to 5 PS1G) with low-flow needle valve 7.7 Beaker - 100 ml 7.8 Micro syringe - 10 ill 7.9 Separatory funnel - 2000 ml with TFE-fluorocarbon stopcock 7.10 Centrifuge tubes - Borosillcate glass, calibrated (15 ml) 8. Reagents and Materials 8.1 Purity of Reagents - Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is Intended that all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available.^ Other grades may be used, provided It Is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination. 8.2 Purity of Water - Unless otherwise indicated, reference to water should be understood to mean reagent water conforming to ASTM Specifications D1193 for Reagent Water, Type II. ^ Reagent Chemicals, American Chemical Society Specifications. American Chemical Soc., Washington, D.C. Por suggestions on the testing of reagents not listed by the American Chemical Society, see "Reagent Chemicals and Standards", by Joseph Rosin, D. Van Noatrand Co. Inc. Mew York, N.Y. end the United States Pharmacopeia." 4 047^59 HONS 8.3 Anti-static Solution - STATNUL 8.4 PCB Standards^: Aroclore 1221, 1232, 1242, 1248, 1234, 1260 and 1016 from Monsanto Co., St. Louis, MO or commerlcal sources. Arodor standard solutions. Environmental Research Laboratory, EPA, GA 30601. 8.5 Diethyl Ether - Pesticide quality, redistilled in glass, if necessary, and contain 2% (v/v) ethanol. 8.5.1 Must be free of peroxides by following test: to 10 ml of ether in glass-stoppered cylinger previously rinsed with ether, add one ml of freshly prepared 10% KI solution. Shake and let stand one minute. No yellow color should be observed in either layer. As alternative, use JM Quant*0 test strips or equivalent. 8.5.2 Decompose either peroxides by adding 40 g of a solution of 30% (w/v) ferrous sulfate solution per liter of solvent. CAUTION! Reaction may be vigorous if the solvent contains a high concentration of peroxides. 8.5.3 Distill deperoxldized ether in glass and add 2% (v/v) ethanol. g Stanul, TM of and available from Daystrom Inc., Weston Instrument Division, Newark, N.J. 95212 A The proportions of individual PCB Isomers may vary from one lot to another and from one manufacturer to another. EM Quant test strips, trade mark of and available from 01 Laboratories, Inc., 500 Executive Blvd., Elmsford, N.Y. 10523 5 HONS 047460 8*6 Floriail - PR grade (60-100 mesh); purchase activated at 1250 F and store in the dark in glass containers with glass stoppers or foil-lined screw caps. Before use, activate each batch overnight at 130 C In foil-covered glass -container. Determine lauric-acid value (See Appendix I). 8.7 Ferrous Sulfate Solution (30%). Dissolve 30 g of ferrous sulfate (FeSO^) in water and dilute to 100 ml. 8.8 Gas Chromatographic Materials 8.8.1 Refer to ASTM D3086 Section 7 8.8.2 Tubing - Borosillcate glass (180 cm long X 4 mm ID) 8.8.3 Glass Wool - Silanized 8.8.4 Solid Support - Gas Chrom Q*1 (100-120 mesh) 8.8.5 liquid Phases - Expressed as weight percent coated on solid support. 8.8.5.1 SE-30 or OV-1, 3% 8.8.5.2 0V-17, 1.5% + QF-1, 1.95% 8.9 Class Wool - Hexane extracted 8.10 n-Hexane - Pesticide quality (HOT MIXED HEXANES) 8.11 Mixed Solvents - Pesticide quality 8.11.1 Ethyl Ether - Benzene Mixture (0.5%) - Take 1/2 volume of ethyl ether and mix with benzene to 100 volumes. 8.11.2 Ethyl Ether - Petroleum Ether Mixture (6%) - Take 6 ' volumes of ethyl ether and mix with petroleum ether to 100 volumes. 11 Ges Chrom Q, TM of end available from Applied Science Laboratories, State College, PA 16801 6 HONS 8.11.3 Ethyl Ether - Petroleum Ether Mixture (15%) - Take 15 volumes of ethyl ether and mix with petroleum ether to 100 volumes. 8.11.4 Ethyl Ether - Petroleum Ether Mixture (50%) - Take 50 volumes of ethyl ether and mix with petroleum ether to 100 volumes. 8.11.5 Methylene Chloride - Hexane Mixture (15%) - Take 15 volumes of methylene chloride and mix with n~hexane to 100 volumes. 8.12 Petroleum Ether - Pesticide quality, 30-60 C boll 8.13 Potassium Iodide Solution (10%). Dissolve 10 g of potassium iodide (KI) in water and dilute to 100 ml. 8.14 Silica gel - Davison code 930-08-08-226 (60/200 mesh)12 6.15 Sodium Sulfate - Granular, anhydrous, conditioned for 4 hours 9 400 C. 9. Extraction of Ssmole 9.1 Transfer 1 liter of sample to a 2-liter separatory funnel equipped with a TFE-Fluorocarbon stopcock. Extract the sample three times each with three 60 ml solutions of methylene chloride-hexane 15X (v/v). If a larger volume of sample is to be analyzed, l.e., 3 liters, extract three times with 150 ml the mixed solvent. ' 9.2 Dry the combined extracts by pouring through a 4-inch column of anhydrous sodium sulfate, (previously rinsed with hexane) contained in a 20 mm ID chromatographic column, and concentrate to 6-10 ml in a Kuderna-Danish evaporator. Qualitatively analyze the sample 12 ' Available from Fisher Scientific, 711 Forbes Ave., Pittsburgh, PA 15219 7 HONS 047462 by gas chromatography, using the 3% Ov-l column described in Figures 3 to 5, with an electron capture detector. From the response obtained decide: a* If it is obvious that only organochlorine pesticides are present. b. If it is obvious that only PCB's are present. c. If there is a combination of a and b, d. If the response is too complex to determine a, b, or c. If no response, concentrate to 1.0 ml or less and repeat the analysis looking for a, b, c, and d. Trace quantities of PCB's are often masked by background which usually occurs in the samples. If detection limits below 100 ng/1 are required, continue as in 10. below, even though the presence of PCB peaks is not apparent in the chromatogram. 9.3 If condition a. exists, determine the organochlorine pesticides. 9.4 if condition b exists, PCB's only are present, no further separation or clean-up is necessary, proceed to 10.2.2 or 10.2.3 below. 9.5 If condition c exists, compare peaks obtained from sample to those of standard Arodors and make a judgment as to which Aroclor standard or combination of standards best represents the PCB's present. To separate the PCB's from the organochlorine pesticides, continue as outlined in 9.7. . 9.6 If condition d exists, apply standard Florisil column procedure (9.7). 9.7 Florisil Column Procedure 9.7.1 Adust the sample extract volume to 10 ml with petroleum ether. 8 MQNS 047463 9.7.2 Place a charge of activated Florisll (weight determined by lauric-acid value, see Appendix X) in a Chromaflex column. After settling the Florisll by tapping the column, add about one-half inch layer of anhydrous granular sodium sulfate to th,,e top. 9.7.3 Pre-elute the column, after cooling, with 50-60 ml of petroleum ether. Discard the eluate and just prior to exposure of the sulfate layer to air; quantitatively transfer the sample extract into the column by decanta tion and subsequent petroleum ether washings. Adjust the eluation rate to about 5 ml per minute and, separately, collect up to three eluates in 500 ml K-D flasks equipped with 10 ml ampuls. (See Eluste Composition below). Perform the first elution with 200 ml of 6% ethyl ether in petroleum ether, and the second elution with 200 ml of 15% ethyl ether in petroleum ether. Perform the last elution with 200 ml of 50% ethyl ether petroleum ether. Eluete Composition - By using an equivalent quantity of any batch of Florisll as determined by its lauric-acid value, the PCB'a and pesticides will be separated into the eluates indicated below: 6% Eluate PCB'a Aldrln BHC Chlordane DDD DDE DDT Heptachlor Heptachlor Epoxide Lindane Methoxychlor Nirex Pentachloro- nltrobenzene Strobane Toxaphene Trifluralin 9 HONS 047.64 15% Eluate Endosulfan I Endrln Dleldrin Dichloran 50% Eluate Endosulfan II Captan Certain thlophosphate pesticides will occur in each of the above fractions. For additional information regarding eluate composition, refer to the FDA Pesticide Analytical Manual (9). 9.7.4 Concentrate the eluates to 6-10 ml In the K-D evaporator In a hot water bath. 9.7.5 To further separate the PCBfs from organochlorlne pesticides continue on to 10, below with the 6% eluate. Analyze the 15% eluate and the 50% eluate for pesticides. Silica Cel Mlcrocoluan Separation Procedure 10.1 Activation for Silica Gel 10.1.1 Place about 20 gms of silica gel in a 100 ml beaker. Activate at 180 C for approximately 16 hours. Transfer the activated silica gel to a 100 ml glass stoppered bottle. When cool, cover with about 35 cc of diethyl ether-benzene 0.5% (v/v). Keep bottle well sealed. If silica gel collects on the ground glass surfaces, wash off with the above solvent before resealing. Always maintain an excess of the mixed solvent in bottle (approximately 1/2" above silica gel). Silica gel can be effectively stored In this manner for several days. 10.2 Preparation of the Chromatographic Column 10.2.1 Pack the lower 2 mm ID Section of the micro column with 10 MONS 047465 glass wool. Permanently mark the column 120 mm above the glass wool. Using a clean rubber bulb from a disposable pipet seal the lower end of the micro column. Fill the micro column with the ether-benzene solution to the bottom of the 10/30 joint (Figure 1). Using a disposable capillary pipet, transfer several aliquots of the silica gel-slurry into the micro column. After approximately 1 cm of silica gel collects in the bottom of the micro column, remove the rubber bulb seal, tap the column to Insure the silica gel settles uniformly. Carefully pack the column until the silica gel la within + 2 mm of the 120 mm mark. Be sure that there are no air bubbles in the column. Add about 10 mm of sodium sulfate to the top of the silica gel. Under low humidity conditions, the silica gel may coat the sides of the column and not settle properly. This can be minimized by wiping the outside of the column with an anti-static solution (8.3). 10.2.2 Deactivation of the Silica Gel 10.2.2.1 Fill the micro column to the base of the 10/30 joint with the ether-benzene solution, assemble reservoir (using spring clamps) and fill with approximately 15 ml ether benzene. Attach the air pressure device (using Bpring clamps) and adjust the column exit flow to approximately 1 ml/mln with the air pressure control. Release the air pressure and detach reservoir just as . the last of the solvent enters the sodium 11 HONS 047466 sulfate. Fill the column with n-hexane to the base of the 10/30 fitting. Evaporate all residual benzene from the reservoir, assemble the column and fill with 3 ml of n-hexane. Apply air pressure and readjust solution flow to 1 ml/min. Release the air pressure and remove the reservoir just as the n-hexane enters the sodium sulfate. The column is now ready for use. 10.2.2.2 Flpet a 1.0 ml aliquot of the concentrated sample extract (previously reduced to a total volume of 2.0 ml) on to the column. As the last of the sample passes into the sodium sulfate layer, rinse down the Internal wall of the column twice with 0.25 ml of n-hexane. Then assemble the upper section of the column. As the last of the n-hexane rinse reaches the surface of the sodium sulfate, add enough n-hexane (volume predetermined, see 10.3 below) to just elute all of the PCB'b present in the sample. Apply air pressure until the effluent flow is 1 ml/min. Collect the desired volume of eluate in an accurately calibrated ampul. As the last of the n-hexane reaches the surface of the sodium sulfate, release the air pressure and change the collection ampul. 12 MONS 047*67 10.2.2.3 Fill the column with ether-benzene; again apply air pressure and adjust flow to 1 ml/rain. Collect the eluate until all of the organochlorine pesticides of interest have been eluted (volume predetermined, see 10.3 below). 10.3 Determination of Elution Volumes 10.3.1 The elution volumes for the PCB's and the pesticides depend upon a number of factors which are difficult to control. These include variation in: a. mesh size of the silica gel b. adsorption properties of the silica gel c. polar contaminants present in the eluting solvent d. polar materials present in the sample and sample solvent e. the dimensions of the micro columns Therefore, the optimum elution volume must be experi mentally determined each time a factor is changed. To determine the elution volumes, add standard mixtures of Aroclors and pesticides to the column and serially collect 1 ml elution volumes. Analyze the individual eluates by gas chromatography and determine the cut-off volume for n-hexane and for ether-benzene. Figure 2 shows the elution patterns of the various PCB components and of the pesticides. Using this information prepare the proper standard mixtures required for analysis of the n-hexane and ether-benzene. MOWS 047468 13 10.3.2 In determining the volume of hexane required to elute the PCB's, the sample volume (1 ml) and the volume of n~hexane (1/2 ml) used to rinse the column wall must be considered. Thus, if It is determined that a 10.0 ml elution volume is required to elute the PCB's, the additional volume of hexane to be added should be 6.5 ml. 10.3.3 Figure 2 shows that as the average chlorine content of a PCB mixture decreases the solvent volume for complete elution Increases. Qualitative determination (10.2) indicates which Aroclor standard(s) best represent the PCB's present and provides the basis for selection of the ideal elution volume. This helps to minimise the quantity of organochlorlne pesticides which will elute along with the low percent chlorine PCB's and insures the most efficient separation possible for accurate analysis. 10.3.4 For critical analysis where the PCB's and pesticides are not separated completely the column should be accurately calibrated according to 10.3.1 determining the percent of material of Interest eluting In each fraction, then flush the column with an additional 15 ml of ether-benzene (0.5X) solution followed by 5 ml of n-hexane, then use the eame column for the sample separation. Using this technique one can accurately predict the amount (X) of materials In each micro column fraction. 14 MONS 047*69 11. Quantitative Determination 11.1 Measure the volume of solvent containing the PCB'e and Inject 1 to 5 til into the gas chromatograph. (Conditions are listed in Figures 3 to 8). If necessary, adjust the volume of solvent to give linear response to the electron capture detector (detection limit approximately 0.1 ng). The mlcrocoulometrlc or the electrolytic detector may be employed to improve specificity for samples having higher concentrations of FCB's (detection limit approximately 50 ng). 11*2 Calculations 10.2.1 Since polychlorinated biphenyls occur in the environment in varying degrees of complexity, it is impossible to prescribe a simple method for quantitative determination. They may occur: a. As the unchanged commercial product, e.g., Aroclor 1242. b* As a combination of unchanged commercial products, e.g., Aroclors 1242 4- 1260. c. Aa metabolized or biodegraded products of the original commercial product or products. d. As a combination of a, b, and c. 11.2.2 For the least complicated situation, a, compare quantitatlve Aroclor reference standards (e.g., 1242, 1260) to the unknown. Measure and sum the areas of the unknown and the reference Aroclor and calculate the result as follows: MtJNS 04 7*70 15 Microgram/liter fAl fB1 fVtl fNl [Vi] [Vs] (Formal* I) Am ng of Standard Injected ne of CStfaannd^aarrdJ DPaeoaLk- AA mreaaos *mmm>*' B of Sample Peak Areas Vi - Volume of sample Injected (p.1) Vt - Volume of extract (ill) Vs - Volume of water sample extracted (ml) N 2 when micro column used 1 when micro column not used 11.2.3 For complex situations (b> c and d) the most practical calibration method is described in (3) below. Small variations In components between different Aroclor batches make it necessary to obtain samples of several specific Aroclors. These reference Aroclors can be obtained from Dr. Ronald Webb, Southeast Environmental Research Laboratory, EPA, Athens Georgia 3Q601. The procedure is as follows: 1) Using the OV-1 column described in Figures 3 to 5, chromatograph a known quantity of each Aroclor reference standard. Also chromatograph a sample of p,p'_DDE. Suggested concentration of each standard la 0.1 ng/iil for the Aroclora and 0.02 ng/ixl for pjp'-DDE. 2) Determine the relative retention time (RRT) of each PCB peak in the resulting chromatograms using p,p'-DDE as 100. See Figures 3 through 5. 16 HONS 047*71 KRT * RT X 100 (Formula XI) RRT m Relative retention time RT m Retention time of peak of Interest RT Retention time of p.p'-DDE DDE - Retention time Is measured as that distance In mm between the first appearance of the solvent peak and the maximum response for each compound. 3) To calibrate the Instrument for each PCB, measure the area of each peak. Using Tables 1 through 6 obtain the proper mean weight factor then determine the response factor, ng/mm^. Area height X peak width at 1/2 height (Formula III) (ng ) (mean weight percent) ng/mnr - *100 100 (Area) (Formula IV) ng of Aroclor Standard Injected mean weight percent * obtained from Tables 1 through 6 4) Calculate the RRT value and the area for each PCB peak in the sample chromatogram. Compare the sample . chromatogram to those obtained for each reference Aroclor standard. If It Is apparent that the PCB peaks present are due to only one Aroclor then calculate the concentration of each PCB using the following formula. 17 HONS 047472 2 ng PCB - ng/nrn X Area (Formula V) 2 where Area (mm ) of sample peak 2 ng/nsn - Response factor for the peak measured. Then add all the nanograms of PCB's present in the volume injected to get the total number of nanograms of PCB's present in the sample. Use the following formula to calculate the concentration of PCB's in the sample: Micrograms/liter * [Eng] IVIN] (Formula VI) Tvp Tv^ volume of water extracted (ml) - volume of extract (^1) Ti Z>g - volume of sample injected dll) * sum of all the PCB's in nanograms for that Aroclor Identified N 2 when microcolumn used 1 when microcolumn not used The value can then be reported as: Mlcrograms/llter PCB's, and for samples containing more than one PCB use Figure 10 chromatogram divisional flow chart to assign a proper response factor to each peak and also identify the "most likely" Aroclors present. Calculate the ng PCB for each PCB present (Formula V), then sum them up according to the divisional flow chart using Formula VI to calculate and report the concentration of the various Aroclors present in the sample. 18 MOMS 047473 12. Confirmatory Techniques 12.1 Unequivocal Identification of PCB's can be made by gas chromato graphy-mass spectrometry (GC-MS) if present in sufficient concentration (approximately 20 ng/y.1 in the final extract). The methods described by Bonelli (4), Eichelberger, et al. (5), Goerlitz (6), and Goerlitz and Law (7) are useful for this purpose. When GC-MS is not avail able, separate GC analysis using both non-polar (Figures 3 to 5) and polar columns (Figures 6 to 9) will give added confidence in the qualitative determination. The use of specific halogen detectors, such as the mlcrocoulometrlc and electrolytic conductiv ity, will eliminate non-halogen interferences and further support the Identification. The concentration of PCB's required is about 10 ng/(il in the final extract. 12.2 ASTM method D 3304-74 (8) for PCB's which incorporates a two-step chemical treatment - saponification with alcoholic potassium hydroxide followed by sulfuric acld-effectively eliminates many interferences while the FCB'a are retained Intact. This procedure D 3304-74 may be used for analyses of industrial effluents when the determination of pesticides is not required and when the sensitivity of the method is adequate to meet the need. 13. Precision and Accuracy 13.1 Polychlorinated Biphenyl Recovery and Precision Data Concentration Material Range ng/1 Number of Samples Recovery Single Operator Average X Precision S0X Aroclor 1242 5000 7 104 Aroclor 1016 1000 27 99 10.2 6.9 19 HONS 047474 APPENDIX I * 14. Standardization of Flori3il Column bv Weight Adjustment Based on Adsorption of Laurie Acid. 14.1 A rapid method for determining adsorptive capacity of FlorlBil la based on adsorption of lauric acid from hexane solution (9) (10). An excess of lauric acid is used and the amount not adsorbed is measured by alkali titration. Weight of lauric acid adsorbed la used to calculate, by simple proportion, equivalent quantities of Florlsll for batches having different adsorptive capacities. 14.2 Apparatus 14.2.1 Buret -- 25 ml with 1/10 ml graduations. 14,2*2 Erlenmeyer flasks --* 125 ml narrow mouth and 25 ml, glass stoppered. 14.3.3 Plpet -- 10 and 20 ml transfer 14.3.4 Volumetric flasks -- 500 ml 14.3 Reagents and Solvents 14.3.1 Alcohol, ethyl -- USP or absolute, neutralized to phenolphthaleln end point. 14.3.2 Hexane -- Distilled from all glass apparatus. 14.3*3 Lauric acid -- Purified, CP 14.3.4 Lauric acid solution -- Transfer 10.000 g lauric add to 500 ml volumetric flask, dissolve in hexane, and dilute to 500 ml (1 ml * 20 mg). 14.3.5 Phenolphthaleln Indicator -- Dissolve 1 g in alcohol and dilute to 100 ml. 20 MONS 0.7.75 14.3.6 Sodium hydroxide -- Dissolve 20 g NaOH (pellets, reagent grade) in water and dilute to 500 ml (IN). Dilute 25 ml IN NaOH to 500 ml with water (0.05N). Standardize as follows: Weigh 100-200 mg lauric acid into 125 ml Erlenmeyer flask. Add 50- ml neutralized ethyl alcohol and 3 drops phenophthalein Indicator; titrate to permanent end point. Calculate mg lauric acld/ml 0,05 N NaOH (about 10 mg/ml). 14.4 Procedure 14.4.1 Transfer 2.000 g of activated Florisil to 25 ml glass stoppered Erlenmeyer flasks. Cover loosely with aluminum foil and heat overnight at 130 0. Stopper, cool to room temperature, add 20.0 ml lauric acid solution (400 mg), stopper, and shake occasionally for 15 min. Let adsorbent settle and plpet 10.0 ml of supernantant into 125 ml Erlenmeyer flask. Avoid inclusion of any Florisil. 14.4.2 Add 50 ml neutral alcohol and 3 drops Indicator solution; titrate with 0.05 NaOH to a permanent end point. 14.5 Calculation of Lauric Acid Value and Adjustment of Column Weight 14.5.1 Calculate amount of lauric acid adsorbed on Florisil as follows: Lauric Acid value mg lauric add/g Florisil 200 (ml required for titration X mg lauric acid/ml 0.05 N NaOH) 14.5.2 To obtain an equivalent quantity .of any batch of Florisil, divide 110 by lauric acid value for that batch and multiply by 20 g. Verify proper elution of pesticides by 13.6 21 MQNS 047476 14.6 Test for Proper Elution Pattern and Recovery of Pesticides: Prepare a test mixture containing aldrln, heptachlor epoxide, p,p'-DDE dieldrin, parathlon and malathion. Dieldrin and parathion should elute in the 15% eluate; all but a trace of malathion in the 50% eluate and- the others in the 6% eluate. 22 MONS 047477 REFERENCES 1. Leoni, V., "The Separation of Fifty Pesticides and Related Compounds and Polychlorinated Biphenyls into Pour Groups by Silica Gel Microcolumn Chromatography", Journal of Chromato graphy. 62, 63 (1971). ' 2. McClure, V. E., "Precisely Deactivated Adsorbents Applied to the Separation of Chlorinated Hydrocarbons", Journal of Chromatogranhv. 70. 168 (1972). 3. Webb, R. G. and McCall, A. C., "Quantitative FCB Standards for Electron Capture Ga9 Chromatography", Journal of Chromatographic Science, 11, 366 (1966), 4. Bonelll, E. J., "Gas Chromatographic/Mass Spectrometer Techniques for Determination of Interferences in Pesticide Analysis", Analytical Chemistry. 40, 603 (1972). 5. Eichelberger, J. W., et al., "Analysis of the Polychlorinated Biphenyl Problem - Application of Gas Chromatography-Mass Spectrometry with Computer Controlled Repetitive Data Aqulsltlon from Selected Specific Ions", Analytical Chemistry. 46, 277 (1974). 6. Goerlltz, D. F., Private Communication, February, 1972. * 7. Goerlltz, D. F. and Law, L. M., "Chlorinated Naphthalenes in Pesticide Analysis", Bulletin of Environmental Contamination and Toxicology. 2 243 (1972). 8. D 3304 - Standard Method of Test for Analysis of Environmental Materials - Polychlorinated Biphenyls, Annual Book of Standards, Part 40, 1974. 23 M0NS 047478 9. "Pesticide Analytical Manual", U. S. Dept, of Health, Education and Welfare, Food and Drug Administration, Washington, D. C. 10. Mills, P. A., "Variation of Florisil Activity: Simple Method for Measuring Capacity and its Use in Standardizing Florisil Columns", Journal of the Association of Official Analytical Chemists. 51, 29 (1968). 24 HONS 047-79 RRTa . Table 1 Composition of Aroclor 1221 (3) . Mean . Weight Percent . * Relative Std. Dev.b Humber of Chlorines . 11 14 16 19 . 21 28 .32 P7 140 31.819.3 ,, 10.1 2.8 20.8 5.4 1.7 Total . 93.3 15.8 1. " . . 9.1 1. . 9.7 '2 . \ ' 9.7 2 : 9.3 2 ' 13.9 .... 2*1 85* : . 3j 15* . 30.i - A 10* . 3j 90* 48.8 3 , 3 `' - ' _ . . ' detention time relative to p,p*-DDE=100. Measured from first appearance of solvent. Overlapping peaks that are quantitated as one peak are bracketed. ' - bstandard deviation of seventeen results as a percentage of the mean of the results. ' CFrorct GC-MS data. Peaks containing mixtures of isomers of different chlorine numbers are bracketed. . . , ' ; 047480 hons RHT* ' Tab-le 2 Composition of Aroclor-1232 (3] Kean Weight Percent Relative Std. Dev.b . Number of Chlorines0 " - 11 14 16 ("20 121 28 32 37 . 40 47 .54 58 70 78 16.2 -9.9 7.1 17.8 9.6 3.9 6.8 6.4 4.2 3.4 * ' 2^6 4.6 .* 1.7 ' 3.4 2.5 6.8 ' 2.4 . ' , 3.4 : 4.7 . . 2.5 2.7 4.1 3.4 " 3.7 3.1 . 7.5 1 1 2 2 2 ' 2"J 3j . 3 ' 3 3 . 40* 60* .. ; - . 4 . . . 31 33* " 4j 67* ' 4 41 90* . 5J 10* . 4 Total 94.2 ^Retention time relative to p,p'-DDE=100. Measured from first appearance of solvent. Overlapping'peaks that are quantitated as one peak axe bracketed. . ' . . "Standard deviation of four results as a mean of the results. ' cprom GC-MS data. Peaks containing mixtures of isomers of different chlorine numbers are bracketed. .. MONS 047481 i RRTa Y 'Table 3 Composition q Aroclor 1242 '** (3) Mean Weight Percent Relative Std. Dev." Number of . Chlorines^ 11 * 16 . 21 28 32 . ' 37 . 40 47 54 . 58 70 78 84 98 104 125 146 . `' Total ' 1.1 35.7 2.9 4;2 11.3 . 3.0 11.0 * 5.0 i 2 2 21 25% 6.1 11.5 3J 75% . 4.7 '3 . 5.7 -3 11.1 ' . . . 8.8 , 6.8 . 6.2 4.3 2.9 . -3 ' 4 " ' ' .31 33% . 5.6 3.3 . .4j 67% / 4 . aa.3 .2.8 41 905 5J 10% 3.6 4.2 ' 4 2.7 9.7 5 . ' 1.5 2.3 1.6 ** 1.0 9.4 16.4 20.4 19.9 5 *5 . 51 85% 6J 15% 51 75% 6J 25% 98.5 *. . : ' *Retention time relative to p,p'~DDE=100. Measured from first appearance of solvent. ' . "Standard deviation of six results as a percentaqe of ' the mean of the results. cFrom GC-MS data. Peaks containing mixtures of isomers of different chlorine numbers are bracketed. ' . ' MONS 04 7482 Composition of Airoclor 1248 (3) KUTa Mean Weight Percent'- Relative Std. Number of Chlorines0 21 28 32 37 40 47 54 58 70 .78 84 98 104 ` 112 125 . 146 Total 1.2 5.2 3.2 8.3 8.3 15.6 . 9.7 . 9.3 19.0 ' 6.6 4.9 3.2 3*3 1.2 ` 2,6 1.5 . 103.1 23.9 3.3 - 3.8 3.6 : *J 3 3.9 . . 31 85% * 1.1 . : 4j 15% 6.0 : * 5.8........... - 3l- 10% `' 4J 90% 1.4 2-7 . 2.6 ' 3.2 . ` 3.6 . 6.6 . 41 80% . 5J 20% -4 5 '5 ; . 41 10% 5j 90* . 5'9 10. d - ' 51 90% 6J 10% - 5] 85% 6j 15% . '* - . '* "Retention time relative to p,p'-DDE=100 ^ first appearance of solvent. . ` Measured fro `'Standard deviation of six results as a percent*TM ,,* the mean of the results. . . . en*sage of ^front GC-MS. data. Peaks containing nixturp*: -- * of different chlorine. numbers are bracketed. . f isomers mons 047483 Table 5- Composition of Aroclor 1254(3) Mean Weight RRTa . Percent * Relative Std. Dev." Number of Chlorines^ *. 47 54 58 70 84 98 104 125 -*1'46 160 174 '203 232 6.2 2.9 1.4 . 13.2 - 3.7 2.6 2.8 2.7 17.3 7.5 13.6 15.0 ' 1.9 5.3 3.8 2.4 10 :-4 2.7 1.3 . 8.4 1.8 1.0 . 8.4 5.5 18.6 26.1 4. 4 ' . . : 41 25% 5j 75% .5 . . ' 5 : . 5 51 70% ' 6j 30% 51 30% 6J 70% 6 .' 6 6. 7. Total 100.0 . _ detention time relative to p,p'~DDE=100. Measured from first appearance.of solvent. '. ^Standard deviation of six results' as a percentage of.the mean, of the results. . ` * ' . . eFrom GC-MS data. Peaks containing mixtures of isomers are bracketed. HONS 047444 ' RR'JC3 70 . 84 T 98 U04 . 117 125 146 ' 160 174 203 f232 1244 280 332 372 448 528 Composition of Aroclor 1260 (3) Mean Weight Percent 2.7 4.7 .. . . 3.8 3.3 12.3 V 14.1 4.9 . 12.4 . 9.3 ' ' 9.8 11.0 4.2 4.0 .6 1,5 . Relative Std. Dev.b Number of Chlorines0 ' ' 6.3 ' '5 1.6 ' 5 513.5 *60, ' 6J 40, 6.7 ' 6 3.3 .. * 2T 15, 6. 85%. ;. 3.6 - 6 2.2 ' 6`150, 2.7 . 4.0 7. 50% : , . 6 61 10* ' ^ 90% e r 3.4 6 10% 7 90% . 2.4 7 5.0 '7 8.6 8 25.3' - 8 10.2 8 Total 98.6 ' .* ^Retention time relative to p,p'-DDE=100. Measured from first appearance of solvent. Overlapping peaks that are guantitatcd as one peak are bracketed. ' "Standard deviation of six results as a mean of the ' ' results. . . ' . '_ cFrom GC-MS data. Peaks containing mixtures of isomers of different chlorine numbers are bracketed. ^Composition determined at the center of peak 104. . Composxtion determined at the center of peak 232. HONS 047485 HONS 047486 SULFUR HEPTACHLOR 1 MIR EX ' ODE ALORIN 1 TECHNICAL CHLOROANE | 1 OP* a PP' DOT r-CHLOR DANE. SNOW d) S .Figure 4. Column: 3% OV-J, Carrier Gas: Hitrogen at 60 ml/min, . Column Temperature: 170 C, Betector: Electron Captura. MONS 047489 Figure 5. Column: 3% OV-1. Carrier Gas: nitrogen at GO ml/min. / Column Temperature: 170 C, Detector: Electron Capture MONS 047490 \ AROCLOR 1242 : INJECT RETENTION TIME III MINUTES ' . .' ' Figure B. Column: 1.5% OY-17 + 1.95% QF-1. Carrier Gas: nitrogen at 60 ml/min. Column Temperature: 200 C, Detector: Electron Capture. MONS IHIECT HONS 047492 MONS 0 9 7 4 9 3 MONS 047494