Analytical Chemistry Method 71-35
(Consolidated Revision of Analytical Chemistry Methods 69-13 and 70-1)
ANALYSIS OF ENVIRONMENTAL MATERIALS FOR POLYCHLORINATED BIPHENYLS
SCOPE
This methodology was developed for the determination of the amount and type of polychlorinated biphenyls (PCB's) in environmental materials. Absolute confirmation of PCB structures is not obtained with this method. Structure proof can be obtained using additional techniques such as mass spectrosMtry to further identify the GC fractions.
PRINCIPLE
The PCB's present in environmental and laboratory exposed samples are isolated from the bulk of the sample matrix by solvent extraction. Interfering matrix components and endogenous residues are then removed from the sample by a suitable combination of clean up procedures and the amount and type of PCB's present determined by electron capture gas chromatography (EC/GC).
REAGENTS Hexane
Nanograde, Mallinckrodt Chemical Works, Catalog No. 4159.
N,N-Dimethyl Formamide
Certified, Fisher Scientific Company, Catalog No. D-119. (Saturated in a separatory funnel with hexane.)
Ethyl Ether
Anhydrous, AR grade, Mallinckrodt Chemical Works, Catalog No. 0848.
Acetonitrile
Nanograde, Mallinckrodt Chemical Works, Catalog No. 2442.
Sodium Sulfate
Anhydrous, granular: AR grade, Mallinckrodt Chemical Works, Catalog No. 8042. Heat at 400*C for one hour prior to use.
Alumina Adsorption
(For chromatographic analysis) 80/200 mesh, Fisher Scientific Co., Catalog No. A540. Heat at 400*C for a minimum period of 4 hours and deactivate with 5t (w/w) distilled water.
Alumina column preparation: Fill a chromato graphic column with hexane up to the point where the reservoir joins the column and push a glass wool plug to the bottom with a glass rod. In a 50 ml beaker measure 35 ml of deactivated alumina (7&tg), and pour this
MOMS 077682
REAGEHTS (Cont'd) Alumina Adsorption (Cont'd)
Distilled Water
Sulfuric Acid Potassium Oxalate Potassium Hydroxide Ethanol 2.58 (w/v) Alcoholic Potassium Hydroxide 51 (w/v) Aqueous Potassium Oxalate PCI Standards
Method 71-35 Page 2
slowly into the column. Tap or vibrate the column to settle the alumina and top the alumina with 2-3 cm of anhydrous sodium sul fate. Wash the column with 50-100 ad of hexane prior to the addition of the sample.
Extracted with hexane to remove hexane soluble electron capturing impurities.
Analytical Reagent Grade, SG 1.84.
Analytical Reagent Grade.
Analytical Reagent Grade.
Formula 2B.
Dissolve >tl2.5 grams of AR grade KDH in 500 ml of ethanol.
Dissolve v5 grams of AR grade potassium
oxalate in 100 ml of distilled water.
Production Grade - Aroclor 1221, Aroclor 1242, Aroelor 1248, Aroclor 1254 and Aroclor 1280. -
APPARATOS
1. Separatory funnels equipped with ground glass stoppers and Teflon stopcocks: 125, 250, 500, 1000 and 2000 ml capacities.
2. Xundarna-Danish Evaporative Concentrators, 500 ml capacity equipped with 3-ball Snyder columns and graduated 5 ml capacity vials: Ace Glassware Company, Catalog Ho. 6707.
3. Chromatographic columns, glass, 10* X 20 am (OD) with a 5* X 50 ms (OD) reservoir at the top, equipped with Teflon stopcocks.
4. Sintered glass filter funnels, 600 ml capacity, 90M.
5. Plat bottomed boiling flasks, 125 aO. capacity: Ace Glassware Company Catalog Ho.' 6896, Code - 04.
6. Liebig Condenser, 200 am in length: Ace Glassware Company, Catalog Ho. 5915, Code - 12.
7. Hot plates. Corning PC-100: fisher Scientific Company.
8. Water bath, Thelco, Precision Scientific, Model Ho. 84, Fisher Scientific Company.
MCNS 077483
Method 71-35 Page 3
APPARATUS (Cont'd)
9. Reciprocating variable speed shaker, Eberbach Corporation, Flatter Scientific Company.
10. 10 til Hamilton Syringes, Catalog No. 701 RN.
11. 32 or. all glass mortars and pestles.
12. 8 X 12 X 2' (2-1/2 qt.) Pyrax baking dishes.
13. U.S. Standard Sieve, No. 30, Fisher Scientific Company.
14. Polytron Tissue Hcstogenizer, PT 10-20-3500, Brinkman Instruments Co.
15. Grinding Hills, Catalog No. 8-450, and Grinding Unit, Catalog No. 8-455, Fisher scientific Company.
16. Usual laboratory glassware.
PROCEDURES
Sampling
It is to be assumed that a rather wide variety of sampling techniques were employed in collecting the samples submitted to us for analysis. In general the procedures used were probably selected for ease of adaptation to the local situation. For this reason, a brief description of the menner in which we treat the samples submitted will usually precede the sample preparation and extraction sections.
Sample Preparation and Extraction
Water
Where possible, the entire water sample, including the container in which it was collected, is extracted with hexane. With larger samples, where this is not physically possible, the containers are simply agitated and a 250 ad portion used for analysis.
Extraction of Water Samples
1. Extraction of water samples - after agitating, transfer the entire ; aqueous sample or a 250 ml aliquot into a graduated glass cylinder.
Record the volume of the sample and quantitatively transfer it to a separatory funnel with distilled water.
2. Rinse the sample container and graduated cylinder with two 50 ml portions of hexane and add each to the separatory funnel.
3. Stopper the separatory funnel and shake vigorously for at least 1 minuts. Allow the layers to separate and transfer the lower aqueous phase to a second separatory funnel.
MOAS 077464
Method 71-35 Page 4
PROCEDURES (Cont'd)
4. Extract the water sample a second time with a 50 ml portion of haxana. After the layers have separated, add the first hewans
extract to the second separatory funnel and transfer the aqueous layer to the original separatory funnel.
5. Repeat the extraction with a third 50 ml portion of hexane. Discard the aqueous layer and combine the hexane extracts.
6. Filter the combined extracts through a 4" funnel plugged with glass wool which is covered with sodium sulfate. Collect the filtrate in a Kundarna-Danish evaporative concentrator, add a small boiling chip, put the Snyder column in place, and reduce the hexane volume to less than 5 ml by heating the apparatus in a 80-90*C water bath. (CAUTIONt SOLVENT VAPORS MUST BE VENTED TO A HOOD.)
Sediment and Soil
Any excess water is decanted and the entire sediment or soil sample transferred to a glass baking dish to air dry at room temperature. The dried material is transferred from the dish into a mortar and pestle and ground. The ground sediment is sieved, remixed, and a 250g portion taken for analysis.
Extraction of Sediment and Soil Samples
1. Neigh 2S0g (to the neareet O.Olg) into a 16 os. narrow neck screw . cap (aluminum foil liner) glass bottle.
2. Moisten the soil with water (tlO oil) and add 150 ml of acetonitrile. Cap the bottle tightly, and mechanically shake for a minimum period of 1 hour.
3. Quantitatively transfer the acetonitrile extract into a 600 ml sintered glass filter funnel containing a 1/4* layer of anhydrous sodium sulfate. Collect the filtrate in a 600 ml beaker (vacuum filtration may be necessary).
4. After the acetonitrile has completely drained into the beaker, wash the bottle twice with 50 ml portions of acetonitrile, adding each wash to the funnel after the previous has completely percolated through the sediment.
5. Quantitatively tranefer the extract to a Kundarna-Danish evaporative ooncantrator, add a small boiling chip, put the Snyder column in place, and reduce the solvent volume to less than 5 ml by heating the apparatus in a 80-90*C water bath. (CAUTIONt SOLVENT VAPORS MUST EE VENTED INTO A HOOP.)
6. Carefully evaporate the acetonitrile extracts to dryness with the aid of a gentle stream of dry nitrogen. Re-dissolve the residue in maximum of 5 ml of nanograda hexane.
RONS 077485
Method 71-35 Page 5
PROCEDURES (Cont'd)
Whole Animal. Poultry and Fish and Individual Tiaaue (Muaela Liver, Fat. Kidney, Heart, Spleen. Lungs, rat, Etc.) Sample*
All individual and composite tissue samples exceeding 75 grams (25 grasw for liver) must initially be homogenised by passing through a meat grinder. The tissue mass is re-ground, with mixing between grindings, until the tissue appears homogeneous. A 50-60 gram portion of the ground tissue (25-30 grams for liver) is transferred to a tared 500 ml Erlanmeyer flask for extraction.
Samples weighing less than 75 grams (25 grams for liver) are cut into small chunks (1-2 cm3) and the entire sample transferred to a tared 500 ml Erlenmeyer flask for extraction.
Extraction of Tiaaue Samples
1. Weigh the tared Erlenmeyer flask containing the tissue to be extracted and record sample weight. Add tlOOg of anhydrous sodium sulfate and ''ISO ml of hexane to the flask. Insert the Polytron tissue hcoogeniier probe and blend, while rotating the flask, until the mixture is well homogenized (''-60-90 seconds).
2. Carefully decant the superoate (hexane extract) through a 4* funnel plugged with glass wool and covered with a layer of anhydrous sodium sulfate, collecting the supernate in a 500 ml Kundema-Danish evaporative concentrator or a 250 ml volumetric flask.
3. Re-extract the homogenate with a second portion of hexane, and again decant the supernate into the filtering funnel.
4. Wash the Polytron* shaft with several small portions of hexane and transfer the homogenate and washings into the filtering funnel with the aid of several small portions of hexane. Press the homogenate with a spatula to remove as much hexane as possible and complete the filtration by washing the funnel and tissue cake with several small portions of hexane.
5. (a) If the extracts are collected in a volumetric flask, dilute to volume with hexane and mix.
(b) Transfer a suitable aliquot to a tared 50 ml beaker and evaporate the solvent with the aid of heat and a gentle stream
, of clean dry nitrogen. Re-weigh the beaker and record the lipid weight.
*The Polytron tissue homogenizer is cleaned between each sample by operating it with the shaft immersed in distilled water followed by nanograde acetone and hexane.
MOMS 077466
Method 71-35 Page 6
PROCroURES (Cont'd)
6. (a) If tha extracts are collected in a Kundarna-Danlsh evaporative concentrator, add a email boiling chip, put
the Snyder column in place and reduce tha solvent volume to less than 5 ml by heating the apparatus in a 80-90*C water bath. (CAPTIOHl SOLVBMT VAPORS HOST BE VTOTED HTTP A BOOB.)
(b) After cooling, remove tha S ml graduated tube, dilute to volume with hexane, cap and mix thoroughly. The entire concentrate or an aliquot is used to determine the lipid weight in manner described above.
Mi11c
1. Weigh *50 grams of whole milk in a 150 ml beaker and quantitatively transfer it to a 500 ml separatory funnel with the aid of several portions of water.
2. Md 5 nl of the 5t aqueous potassium oxalate solution and swirl to mix.
3. Add 100 ml of diethyl ether, shake for several minutes, add 50 ml of hexane and shake for several more minutes.
4. Allow the mixture to stand for 10-15 minutes and discard the lower aqueous layer.
5. Filter the remaining solvent layer through a 4* funnel plugged with glass wool covered with anhydrous sodium sulfate into a tared 400 ml beaker.
6. Rvaporate the solvent, allow the beaker to cool, re-weigh, and record the lipid weight.
Bangle Clean Pp
DMF/Bexane Partition. Chemical Treatment
1. Place an appropriately sised aliquot of the hexane extracts, containing a maximum of 3.5g of fat and not exceeding 25 ml, in a 500 ml separatory funnel. If the aliquot taken is some volume less
" than 25 ml, add enough additional hexane to yield a final volume of 25 ml.
2. Add 50 ml of OMF (hexane saturated) to the separatory funnel, stopper the funnel, and shake vigorously for at least 1 minute. Allow the layers to separate and transfer the lower DMT layer to a second separatory funnel.3
3. Repeat the extraction of tha hexane layer with two mare 50 ml portions of DMF (hexane saturated) collecting each in turn in the second separatory funnel. Discard the remaining hexane; lays*MCMS 077487
Method 71-35 Page 7
PROCgDPRJES (Cont'd)
4. Add 100 ml of distilled water to the combined DMF extracts, swirl to mix, without excessive agitation, allowing the hexane to layer out.
5. Transfer the lower aqueous-DMF layer to another 500 ml separatory funnel.
C. Extract the aqueous DKF sample with a 50 ml portion of hexane. After the layers have separated, add the first hexane extract to the second separatory funnel and transfer the aqueous-DMF layer to the original separatory funnel.
7. Re-extract the aqueous-DMF with two more 50 ml portions of hexane. Discard the aqueous-DMF layer and combine the hexane extracts.
0. Filter the combined extracts through a 4* funnel plugged with glass wool which is covered with sodium sulfate. Collect the filtrate in a Kunderna-Danish evaporative concentrator, add a small boiling chip, put the Snyder column in place, and reduce the hexane volume to less than 5 ml by heating the apparatus in a 80-90*C water bath. (CADTIOMi BOLVEMT VAPORS MOST BE VESTED TO A HOOD.)
Saponification with Alcoholic Potassium Hydroxide and Extraction with
1. Quantitatively transfer the concentrated extracts to a 125 ml extraction flask with the aid of several small portions of solvent.
2. After cooling, remove the 5 ml graduated tube and transfer the con centrate of extracts to a 125 ml extraction flask with the aid of several small portions of solvent.
3. Evaporate the extract just to dryness with a gentle streem of dry filtered air and add 25 ml of 2.5t alcoholic potassium hydroxide.
4. Add a boiling chip, put a water condenser in place, and allow the solution to reflux for 45 minutes.
5. After cooling, transfer the solution to a 250 ml separatory funnel with the aid of 25 ml of distilled water.
t. Rinse the extraction flask with 25 ml of hexane and add it to the separatory funnel.
7. Stopper the seperatory funnel and shake vigorously for at least 1 minute. Allow the layers to separate and transfer the lower aqueous phase to a second separatory funnel.
HONS 077488
Nthod 71-35 Page 8
procedures (cont'd)
8. extract tha saponification solution with a second 25 ml portion
of hexane. After the layers have separated, add the first hexane
extract to the second separatory funnel and transfer the aqueous
alcohol layer to the original separatory funnel.
>
9. Repeat the extraction with a third 25 ml portion of hexane. Discard the saponification solution and cosfoina the hexane extracts.
10. CAREFULLY add 25 ml of concentrated sulfuric acid, in small portions, to the hexane extracts.
11. Stopper the separatory funnel and shake vigorously for at least one minute. Allow the layers to separata.and discard the lower aqueous -scid layer. Repeat this step until the acid layer is colorless.
12. Wash the hexane with 25 ml portion of water. Discard tha water wash.
13. Filter the hexane extract through a 4* funnel plugged with glass wool which is covered with a layer of sodium sulfate. Collect the filtrate in a xunderna-Danish evaporative concentrator, add a small bailing chip, put the Snyder column in place, and reduce the hexane volume to less than 5 ml by heating the apparatus in a 80-90*C water bath. (CAPTIOH: SOLVEHT VAPORS MUST BE VHMTBD TO A HOOD.)
Llguid/Solid Chromatographic Clean-Up With Alumins
1. After preparation of the alumina column, described in the 'Reagents* section, quantitatively transfer the hexane extracts (5 ml or less) on to the column with the aid of several 5 ml portions of hexane (reducing the solvent level to the adsorbant level after each addition of solvent).
2. Carefully add 125 ml of hexane to the column reservoir and collect the total eluent in a Kunderna-Danish evaporative concentrator.
3. Add a small boiling chip to tha Kunderna-Danish evaporative concen trator, put the Snyder aoluam in place, and reduce the hexane volume to less than 5 ml by heating the apparatus in a S0-90*C water bath. (CAPTIOH: SOLVEHT VAPORS MUST BE VENTED TO A HOOD.)
Electron Capture Gas Chromatography
Instrument(s):
Hewlett-Packard 5750
Research Chromatograph and/or 402 High Efficiency Gas Chromatograph
Detector: High Temperature Ni63 Electron Capture Cell
Column: Glass, Spiral and/or *U*, 2M x4mm, 4t XE-60 on 80/100 Nash Chramosorb W, HP, AW-DMCS
KQNS 0 7 7465
Method 71-35 p*a 9
Column Temperature:
170*C-220*C (specific temperatures employed
for each Aroclor are shown on attached chromatograms)
Detector Temperature: 250*-300*C Injection Port: 220*C Pulaa: 50 nsec
Flow Rates
Carrier: '60 ml/min He
Purge: ^120 ml/min ab/CH, (10* AR)
The elaotron capture gas chromatographic system can be affectively esqployed to guide the isolation, separation and clean up procedures. Sample extracts should be checked for the presence of PCB's and/or matrix and other chlorinated hydrocarbon interferences by injecting
icroliter portions of the extracts at various points in the extraction and concentration schasie. In this manner, it can be determined if the sample must be concentrated or diluted and what, if any, confirmatory and/or clean-up procedures are needed. The approaches routinely employed in our laboratory are shown schematically in Figure I.
Absolute confirmation of the presence of PCB's in a sample is obtained by GC/Mass analysis of the residue. Alternate chemical and physical confirmatory techniques not routinely employed in this laboratory are listed in the reference section.
Quantitation
Quantitative determinations employing the electron capture detector are non-stoichicmetric measurements made by comparing peak heighta or areas for known concentrations with those for unknown compositions. Four variations of the peak height or area quantification procedures have been employed.
Case I
EC gas chromatogram of PCB unknown unchanged with respect to standard PCB with no evidence of interferences.
Casa II
EC gas chromatogram of PCB unknown altered with respect to standard PCB with no evidence of interference.
Case III EC gas chromatogram of PCB unknown unchanged with respect to standard PCB with evidence of interference.
Casa IV
BC gae chromatogram of PCB unknown altered with respect to standard PCB with evidence of interference.
The PCB level in a Case I sample can be determined by coaqparison of the height of the major peak in the unknown with a calibration curve pre pared by plotting the height of the major peak in the electron capture chromatogram of the corresponding standard vs the number of ng of the standard injected. With a Case II sample because of the alteration, the maximum PCB level is estimated by comparison of the total area of the electron capture envelope of the unknown to a calibration curve
prepared by plotting the total area of the electron.capture envelope of the most similar standard vs the number of ng Of*the standard
HONS 077490
SAMPLE MATRIX
OPERATIONS
o
71-35 HONS 0 7 7 4 9 1
Method 71-35 Page 11
PROCEDURES (Cont'd)
Injected. Ceae III and XV ample* are Initially aubjected to the chemical clean-up procedure followed by chromatography a* outlined on alumina. If the interference* are renewed by thia treatment, the PCB level* are calculated for Caae III and IV in the aame manner aa Caae I and II, reapectively. If dominant interference(a) ia (are) atill preaent, then the height of a different analytical peak, free from obvioua interference, ia employed to calculate the PCB level in
Caae III. with Caae rv aamplea, the maximum PCB level ia eatimatad aa
in Caae II after correcting the total area for that of the interfering peak(a).
In all caae*, the reaponae of the electron capture muat be linear for quantitative analyaia.
Detection Limit*
Sample Matrix Mater
Tiaauea Milk
Sample Size
250 ml 25-50g
50g
Pinal Extract Volume 5 ml
5 ml
5 ml
Injection Volume 5 ill 5 Ml
5 Ml
Detection* Lim 2 ppb
10-20 ppb 20 ppb
`Greater aenaitivity can be achieved aa needed by incfeaaing
the aample aize, decree*ing the final extract volume, or
increaaing the injection volume.
.
DISCUSSION
Extraction
The extraction of PCB'a from water employing hexane ae the extractant waa found to be quantitative and to be aufficiently aimple and rapid for uee aa a routine procedure.
The evaluation of thia method waa baaed on apiking water aamplea with etandard acetone aolution* of PCB'a. The apiking method conaiatad of adding the PCB'a in acetone (25-50 ul) to 500 ml of tap water in a 32 oz. narrow neck acrew cap jar. After thoroughly mixing, duplicate 225-250 ml aliquot* were taken and aubjected to the propoaed eample preparation and work up aa outlined. The reaulta were quantified by preparing a calibration curve uaing atandard hexane aolutiona of the PCB'a uaed to apike the water aamplea. The total area of the electron
capture chromatogram waa uaed to conatruct the calibration plot.
The average recovery and deviation achieved aubatantiated the appli
cability of the method for the- quantitative recovery and analyaia of
PCB'a from water at the ppb-ppm level.
-
MCkS C7?<>S2
Method 71-35
& pg 12
DISCUSSIOH (Cont'd)
m KB recovery experiments from spiked sediment and aoil samples have been performed. Instead, several of tha residual aolids representative of soma of tha types of sediment or soil analysed were re-axtracted with haxana/acatona <40/(0) in a soxhlat extractor to t#st for tha efficiency of tha acetonitrile extraction atop. Tha hexane, after isolation by dilution with distilled water was then carried through the purification steps. Recoveries by soxhlat extrac tion have indicated that the acetonitrile extraction of KB's was essentially quantitative in the cases checked.
Based on tha assumption that fat or fatty organs generally yield the lowest recoveries a series of experiments were conducted with hexane solutions of chicken fat fortified with known anounts of Aroclor. Tha spiked fat samples were cleaned up (DMF/Hexane partition and alumina chromatography) and the amount of KB determined by EC/GC. All recoveries fell in the range of 98-102*> for this reason recovery experiments were not carried out with each individual tissue. While a mere realistic evaluation of the overall recoveries might produce recoveries in the range of 70-901, it was felt that this would be more than sufficient for our purposes, especially in light of the quantitation preblens which exist.
Sample Concentration
Concentration of sample extracts is necessary, prior to clean up by chromatographic or chemical means, to reduce sample sisa and increase spnsltivity. The preferred method of concentrating allows minimum loss through volatilisation or chemical decoatposition and requires a time. The three methods of solvent volume reduction most coasonly used age evaporation by exposure to a stream of air, evaporation employing a> Kunderna-Danish evaporative concentrator equipped with a Snyder cplumn, and evaporation under reduced praesure. We have used all three techniques and have not encountered any significant lossea from vola tilisation or chemical alternation. However, the Kunderna-Danish evaporative concentrator and the stream of air were employed because of the ease of uae.
Column Adsorption Chromatography and Chemical Clean-Up
Silica gel, Plorisil and Alumina deactivated with 0, 1.0, 1.5, 2.0 end SB water were investigated as adsorbents for the elimination of inter ferences. Alumina (SB water) was found to be effective and more reproducible than either silica gel or Plorisil. The activity of alumina varies with age and lot, therefore, SB water was added to the alumina, after heating for a minimum of 4 hours at 400*C, to insure a reproducible activity.
Saponification and aubaequent extraction of the sample with sulfuric acid is an affective way to remove a number of chlorinated hydrocarbon interferences as well as other, matrix interferences, weverthaless, several pesticides survive the chamioal clean-up procedures (e.g. sjtrobane, toxaphene, DDK, chlordane, etc.) demanding careful inter pretation of the EC chromatograms or alternate chemical and separation procedures referenced in the appendix. KB's are no*affected.
f
HONS 077493
Method 71-35 Page 13
DXSCUSSXOM (Cont'd)
Electron Captura Gas chromatography
Columns
Column performance Is the key to effective gas chromatographic analysis and as such the choice of column materials is particularly important. Ideally, the support employed should be inert, mechanically strong, and of high surface area. For these reasons, Chroraosorb w, HP, AW-DMCS was used in all of our work.
A variety of polar and non-polar liquid phases were investigated. The following columns were found to provide adequate separation, etc., for use in PCS analysis by electron captures 4t (w/w) DC-200, SF-96, OV-17, SE-30, SE-54, XE-60, Apieson I>, and 6% QF-1. DC-200 and XE-60 or OP-1 have been found to be the isost suitable of these liquid phases.
Another important consideration when working with an extremely sensitive detector and consequently low levels of materials is column conditioning. With polar phases such as XE-60 and QF-1, we have found that operating a new column overnight at a temperature 25-50'C higher than to be used during analysis results in a more stable column. A no-flow conditioning technique is employed to condition non-polar columns. The column is purged with carrier gas, heated for 30 minutes at an elevated tesiperature without carrier flow and then cooled to roam temperature. At the end of this cycle, the carrier flow is resumed and the conditioning is com pleted as in the case of the polar liquid phase. Two precautionsi During conditioning, the column should not be connected to the detector and one should not exceed the maximum safe temperature of the liquid phase.
Since all liquid substrates bleed to one degree or another and columns eventually degrade, we characterize all new columns with two column performance indicators - the number of theoretical plates (N) and a tailing factor (T). p,p'-DDT is smployed to cheek these parameters because it is known to degrade on "poor* columns. In this manner, we can determine if the performance of a new column is satisfactory and when the column performance begins to fall off. We consider a column good if the number of theoretical plates per foot is on the order of 400-500 with tailing factors of 1.0-1.3. Calculation of these parameters is shown in the appendix. Additionally, there should be no significant extraneous peaks upon injection of a pure p,p-DDT standard.
Other chromatographic conditions that can be adjusted are column tempera ture and flow rates. Although resolution of a mixture increaaee with decreasing temperature, a tesiperature should be chosen that allows the elution of all components within a convenient time period. The flow rates are optimum for our instrument, column and detector system and, of course, should be adjusted if better results can be achieved.
Two gas chromatographic systems have been used for PCB analysis - HP Model 402 and 5750. We find that any system of instrument and column suitable for chlorinated pesticides is satisfactpJtt for PCB analysis.
MOWS 077494
Method 71-35 Page 14
DISCUSSI01I (Cont'd)
The bulk of analyse* in our laboratories was carried out using the systasu outlined. The use of the high temperature Ni*3 electron
capture cell is highly recommended. The ability to operate at higher temperatures prevents maintenance problems due to contamination froai high boiling components.
COHTAMIHATICM
.
In determining PCB's by electron capture gas chromatography, laboratory sources of contamination can be a major problem. The samples and extracts should never be allowed to come in contact with materials other than glass. Teflon or metal. Laboratory glassware should be
thoroughly washed with hot, soapy water, rinsed with distilled water, asetona and then hexane. All equipment should also be rinsed again with hexane just prior to use and blanks should be frequently carried through all steps of the procedures to insure against the possibility of contamination.
db
Monsanto Company Organic Chemicals Division Applied Sciences Section St. Louis, Missouri
10/71 - N. J. Litschgi, W. M. Mess, E. S. Tucker
*
MGNS 077495
MOMS 077496
COLUMN PERFORMANCE INDICATORS
Calculating Column efficiency 1. Theoretical Plate*, N
N 16<*/y)2
2. TaiUng, T T a/2b
HONS 077497
REFERENCES ALTERNATE CHEMICAL/PHYSICAL SEPARATION AMD
IDENTIFICATION PROCEDURES
1. "Method for Separating Polychlorinated Biphenyls Iron DDT and Its Analogs", J. A. Armour and J. A. Burke, J. of the AQAC, 53, (), 1970.
2. "Column Chromatographic Separation of Polychlorinated Biphenyls from Chlorinated Hydrocarbon Pesticides, and Subsequent Gas Chromatographic Quantitation in Terms of Derivatives", 0. W. Berg, P. L. Diosady and G. A. V. Haas, Division of Laboratories, Ontario Water Resources Commission. J. H. Neil, Director - private communication, submitted for publication.
3. "Behavior of Chlorinated Naphthalenes in Analytical Methods for Organochlorine Pesticides and Polychlorinated Biphenyls", J. A. Armour and J. A. Burke, J. of the AOAC, 54 <1), 1971.
4. "Structural Identification of Polychlorinated Biphenyls in Coamarcial Mixtures by Gas-Liquid Chromatography, Nuclear Magnetic Resonance and Mass Spectrometry", D. Sissions and D. Welti, J. Chromatogr., 60, 15-32 (1971).
HONS 077498
Figure I
Figure II Figure ill Figure IV Figure V Figure VI Figure VII
Figure VIII
Figure IX
Figure X
Figure XI
Figuree XII - XIX
FIGURES
Composite electron capture chromatograms demonstrating overlap regions for technical Aroclor products.
Aroclor 1221 electron capture reference chromatogram at analysis conditions.
Aroclor 1242 electron capture reference chromatogram at analysis conditions.
Aroclor 1248 electron capture reference chromatogram at analysis conditions.
Aroclor 1254 electron capture reference chromatogram at analysis conditions.
Aroclor 1280 electron capture reference chromatogram at analysis conditions.
Support Coated Open Tubular flame ionisation chromatogram - demonstrating the complexity of the isomer distribution of Aroclor 1242.
Typical alteration of Aroclor 1242 observed after amtabolism and/or excretion by Albino rats.
Typical alteration of Aroclor 1254 observed after metabolism and/or excretion by Albino rats.
Typical alteration of Aroclor 1260 observed after metabolism and/or excretion by Albino rats.
Typical alteration of Aroclor 1242 after amtabolism and/or excretion by White Leghorn chickens.
Electron capture chromatograms of scam commercial pesticide mixtures before and after treatment on XE-60 and DC-200.
HGMS 077499
1
HONS 0 7 7 5 0 0
FIGURE II
AROCLOR 11
*F 402 Rl-63 IC 1/4"R6' 4Z XR-60 on 80/100 _ . Chroaooorb V HF Col. Toap. - J70*c IaJ. Toap. - 220*C Dot. Toap. - 230*C Carrier Coo - go 60 nl/aln. Forgo Coo - 10X CH4/Argon 120 al/aln. Fnloo Interval - 30 peec **. Injected - S Ml td. Cone. - 1.48 Z 10_ o/al tango - 10 Attn. - 4
'ni/fes
T
T
FIGURE III
AROCLOR 1242
HAP 402 Hl-63 (C
1/4"X' 4X XE-60 on 80/100
Cbroaooorb H HP .
Col. Toap. - 190*C
InJ. Toap. - -220*C
Dot. Toap. - 2S0*C
Carrlor Goo - Ho 60 al/aln.
Fargo Goo - 10Z CH4/Argon 120 al/aln.
Pnloo Interval - 50 liooe
Aat. Injected - 5 ill
ltd. Cone. - 1.03 X 10~* g/al
Kongo - 10
Attn. - 8
.
% -So
-4o
3o
-----------------------------1--------1---------1---------1-----------1---------1-----------1---------1--
minutes 7 < S 4 3 2 / o
HONS 072502
FIGURE IV
AROCLOR 1248
MRP 402 Mi-63 EC 1/4*X61 44 XE-60 on 80/X00
Chromosorb H HP Col. Temp. - 190*C Inj. tamp. - 220*C Dot. Tamp. - 2S0*C Carriar Gas - Ho 60 ml/min. Purge Gas - 10* CH4/Argon 120 ml/min. Pulsa Interval - SO usee Amt. Injected - 5 ul Std. Cone. - 1.31 X 10~ g/ml Range - 10 Attn. - 4
<d>
% So
I "I
I I1 I 1 I
I
In " I I I
I II
Hi a n. U to 1 8 7 6 4- 4 3 / o
HONS 077503
FIGURE V
AROCLOR 1234
RAF 402 Rl-63 BC
1/4"X6' 41 XB-60 on <0/100
Cbroaoeorb V RP
Col. Teap. -> 203*C
Inj. Teap. - 220*C
Dot. Teap. - 2S0*C
Carrier Caa - Ha 60 al/aln -
Purge Caa - 102 CR4/Argon 120 al/aln.
Pulse Interval - 30 ysec
Aat. Injected - 3 yl
Std. Cone. - 1.02 X 10*6 g/nl
Range - 10
Attn. - t
'
z
*Vo
-3o
-So Y/o
-i-------1----------1-----------t----------1----i-----r-----r------r-----1------t ft // /o f 6 7 < ^ V 3 2
r/
RONS 077504
FIGURE VI
ABOCIOR 1260 E6P 402 N1-63 EC l/4nX6' 4Z XE-60 on 60/100
Chromooorb W HP Col. Trap. - 220*C InJ. Toap. - 220*C Dot. Taap. - 250*C Corrior Co* - Bo 60 al/ala Purgo Goo - 10Z CB5/Argon 120 al/aln. Pulso Intorvol - 30 |iooc Aat. Injoctod - 3 ill Std. Cone. - 0.98 X 10~6 g/al Bongo - 10
. .jitatifms M0NS 077505
CHROMATOGRAPHIC CONDITIONS '
Instruaent! Perkin-Blaer Model 800
Type Detectori Flaae Ionisation
PIGPBB VII
Coluan:
100' X 0.02* I.D. FFAP S.C.O.T. (Support Coated Open Tubular) Capillary (Perkin-Elaer Part No. 008-0804 for each 50* length)
Coluan Teaperature: 230*C
Detector Block Teaperaturei 285*C Injection Port Teaperaturei 330*C Saaple Voluae Injected! 0.002 ml
. Split 30 to 1
Flow Rates
Coluan Beliuat 5 al/ain (7 psig)
Auxiliary Heliua: 50 al/ain Hydrogen: 50 al/ain Air: 400 al/ain
12.
LuS xi <XootnL A m J e r
HONS 077506
FIGURE VIII
HOMS 0 7 7 5 0 7
MONS C 7 7508
FIGURE X /*/
/ffaseJe. z*"hvcJ~
/frecJer /2"j/ five*/
-
i i --i--------- i " '"i"-- r 1 i --i------- r
/ // fc f 6
7A
4
FIGURE XI RONS 077510
'I
HONS 0 7 7 5 1 3
\
MONS 077S 14
SEZ
KONS 077515
HONS 0 7 7 5 1 6
HONS 0 7 7 5 1 7