Document o16w4JJz3G8eG62OYgrbmGv8

THE ANALYSIS OF POLYCHLORINATED BIPHENYLS IN ' TRANSFORMER FLUID AND HASTE OILS o ft 1. Scope 1.1 This Is a gas chromatography (SC) method applicable to the deter* mlnatlon of coamerclal mixtures of polychlorinated biphenyls (PCBs) In transformer fluids and certain other hydrocarbon based waste oils. The method can be used to analyze waste oils for individual PCB Isomers or complex mixtures of chlorinated biphenyls from pgpp> chloroblphenyl through decachloroblphenyl only if they have been previously Identified by other methods^ ^ or by knowledge of sample history. ' 1.2 The detection limits are dependent upon the complexity of the sample matrix and the ability of the analyst to properly maintain the analytical system. Using a carefully optimized Instrument, this method has been shown to be useful for the detersrlnation of commercial PCB mixtures spiked Into transformer fluid over a range of $.0 to SOO mg/Kg. Based upon a statistical calculation at 5 mg/Kg for a simple oil matrix, the method detection limit for Aroclors 1221, 1242, 1254, and 1260 Is 1 mg/Kg. The method U. S. Environmental Protection Agency Office of Research and Development Environmental Monitoring and Support Laboratory Cincinnati, Ohio 45268 February 1981 HONS 015542 detection Halt (l) 1$ defined the minimum concentration of substance that can be naasurad and reported with 99t confldance that the value Is abova zero. 1.3 This Method Is restricted to use by or under the supervision of analysts experienced In the use of gas chrautography and In the Interpretation of gas chroutograu. Prior to sample analysis, each analyst must demonstrate the ability to generate acceptable results with this uthod by following the procedures described In Section 10.2. 2. Sumeary % 2.1 The sample Is diluted on a welght/volume basis so that the contan tratIons of each PC8 Isomer Is within capability of the gas jjhror utographlc (SC) system (0.01 to 10 ng/uC). ; `. 2.2 The diluted sample Is then Injected Into a gas chroutograph for separation of the PCS isomers. Measurement is accomplished with a halogen specific detector which maximizes baseline stability and minimizes Interferences norully encountered with other detectors. The electron capture detector (ECO) can norully be substituted for the halogen specific detector, when samples contain dlchloro- threugh decachloroblphenyl 1 sours (Aroclors 1232, 1242, 1016, 124S, 1254, 1260, 1262 and 1268) or when the sample matrix does not Interfere with the PCBs. Several cleanup techniques are provided * for samples containing Interferences. A uss spectrometer oper ating In the selected Ion monitoring mode of data aqulsltlon may also be used as the 6C detector,*when PCI levels are sufficiently . high and the PCS m/e ranges are free from Interference, Interfer- MONS 015543 nets aay occur In som waste oil samples even after exhaustive cleanup. 2.3 The concentrations of the PC8s are calculated on a ng/Kg basis, using coanerclal mixtures of PCSs as standards. The analysis time, not Including data reduction. Is approximately 35 mln/sanple. 3. Interferences 3.1 Qualitative mlsldentlflcatlons are always a potential problem { K analysis. The use of a halogen specific detector and the amplyet's skill In recognizing chromatographic patterns of coamarclal bC|. mixtures minimizes this possibility. . 3.2 Whenever samples are analyzed that do not provide chromatographic patterns which are nearly Identical to the standards prepared from commercial PCSs, the analyst must confirm the presence of PCSs by analysis after column cleanup, analysis on dissimilar gas chromato graphic columns or by gas chromatography/nass spectrometry (GC/MS). 3.3 During the development and testing of this method, certain analyti cal parameters and equipment designs were found to affect the validity of the analytical results. Proper use of the method requires that such parameters or designs must be used as specified. These Items are Identified In the text by the word 'must.* Anyone wishing to deviate from the method In areas so Identified, must demonstrate that the deviation dees net affect the validity of the data. Alternative test procedure approval must be obtained through the EMSL-CI Equivalency Program^. An experienced mmlyst may make modifications to parameters or equipment Identified by the HONS 015544 term "recq--ndid". Each time such modifications ara and* to tha athad, tha analyst aust rapaat tha procedure In Section 10.2. In this case, formal approval Is not required, but the documented data from 10.2 aust be on file as part of tha overall quality assurance program. 3.4 Samples diluted 100:1 analysed by electron capture GC, consistently produce results that are 10 to 20X lower than the true value (Section 12). This is due to quenching of the detector response by high boiling hydrocarbons coeluting with the PCBs. The degree of error Is Matrix dependant and Is not predictable for simples eaT*'* _ , .it unknown origin. As the PCS .concentration approaches 20 parent of a control level, e.g., SO mg/kg, the analyst must routinely analyze a duplicate spiked sample to determine the actual reCovoagr. Spike the duplicate or diluted sample at 2 times the electron " capture observed value and reanalyze according to Section 10.2. Correct the results accordingly. 4. Apparatus 4.1 6as Chromatograph - The gas chroawtograph should be temperature prograsemble, equipped for on-column Injection, and capable of accepting 1/4 Inch 00 glass columns. 4.2 Gas Chromatographic Detector 4.2.1 A halogen specific detector 1$ used to eliminate Interfer ences causing mlsldentlflcatlons or false-positive values due to non-organohalidos which co--only coelute with the PDCs. a. Electrolytic conductivity detector - the Hall Model 70O-A HONS 0155V5 (HQ), available from Tracor, has been found to provide the sensitivity and stability naadad for tha currant PCI Regulations. b. Other halogen spaclfle datectors. Including oldar mode) electrolytic conductivity datactors and mlcrocoulcmetrlc titration, my ba usad. Howavar, the stability, sensi tivity, and response time of these datectors my raise the nathod detection limit and adversely affect peak resolution. Each system must ba shown to bo operating within requirements of tha PCS regulations by collecting single laboratory accuracy and precision data and mafNmd \. detection limits on simple spiked samples as descrffcpd.ln Section 10.2. *' 4.2.2 Semi-specific detectors such as ECO my be substituted whom sample chromatographic patterns closely match those of the standards. Acid cleanup (Section 8.1) or Florlsll slurry cleanup (Section 8.7) should be Incorporated routinely when the ECO Is used. See Section 3.4 for additional quality control procedures for ECO. 4.2.3 Quantitative 6C/HS techniques can be used and the recom mended approach Is selected Ion monitoring (SIN). The SC/NS data system must have a program that supports this method of data acquisition. The program must ba capable of monitoring a minimum of eight ions, and It Is desirable far the system ' to have the ability to change the Ions monitored as a function of time. For PCI measurements several sets of HONS 015546 Ions nay be used depending on the objectives of the study end the date systeei capabilities. The alternatives are as follows: Single Ions for high sensitivity 154, 188, 222, 25C, 292, 326, 360, 394 Short nass ranges which my give enhanced sensi tivity depending on the data systan capabilities 154-156, 188-192, 222-226, 256-260, 290-295, 322-328, 356-364, 390-398 Single Ions that give decreased sensitivity but are selective for levels of chlorination (1). 190, 224, 260, 294, 330, 362, 394 The data systan oust have the capability of Integrating the abundances of the selected Ions between specified limits, and relating Integrated abundances to concentra tions using the calibration procedures described In this method. 4.3 Sas Chromatographic Colions 4.3.1 The 6C columns and conditions listed below are rnco--ndad for the analysis of PCS mixtures In oil. if these columns and conditions are not adequate, the analyst may vary the coltmn parwMters to Improve separations. The columns and conditions selected must be capable of adequately resolving the PCBs In the various Aroclor mixtures so that each Aroclor Is Identifiable through Isomer pattern recognition. See Figures 1 through 6 to establish this. To properly uae the calculation procedure described in Section 11.5, the analyst must use the methyl silicone column described In HONS 015547 4.3.2. Capillary colons and their associated specialized Injection techniques are acceptable alternatives to the MCO--uded packed colons. Due to problem associated with the use of capillary colons the analyst mist demonstrate that the entire systaai will produce acceptable results by performing the operations described In Section 10.2. 4.3.2 Recommended primary analytical colon: Glass, 1/4 Inch 0.0. (2 me 1.0.) 6 ft (180 cm) long, packed with 6as-Chro* q (100/120 mesh) coated with 3X 0V-1. Carrier gas: 40 to 60 nL/mln (helium, nitrogen or 101 : methane in argon). v o 3. Temperature Program: 120 C isothermal for 2 minutes, t. - y '* 6/m1nute to 220C and hold until all compounds elute.> * ' Figure 6 shows a chromatogram of the PCB locator mixture (see Section 5.8) analyzed under these conditions. Each PCS peak has been Identified by assigning the same relative retention times determined In the Isothermal runs Figures 1 through 6. Isothermal Operation: Aroclor 1221, 1232, or C1-] through C1^ Isomers 140 to 150C Aroclor 1016, 1242, 1254, 1260, 1262, 1268, or Cl3 through C110 Isomers 170 to 200C 4.3.3 Recommended confirmatory column: Glass tubing 1/4* 0.0., 2 Rm 1.0. 6 ft (180 cm) long, packed with Gas-Chram Q 100/120 mesh coated with 1.5X OV-17 + 1.95X 0V-210. MOMS 015548 Carrier gas: 40 to 60 mL/mlnute (hellim, nitrogen or 10ft methane In argon). Column temperatures: Aroclor 1221, 1232, or Clj through Cl^ 1somers-.170 to 180C. ' Aroclor 1016, 1242, 1254, 1260, 1262, 1268, or Clj through C11Q 1somers-200C. 4.4 Volimetrlc flasks - 10- 100- 200- and 250-aiL 4.5 Plpets - 0.10 mL, 1.0 ml, and 5.0 ml, Mohr delivery (for viscous oils cut off tip of plpat). 4.6 Micro syrlngas - 10.0 id. ' : 4.7 Sample containers - 20 ml or larger screw-cap bottles with Itflon faced cap liners. (Aluminum foil cap liners can be used for non corrosive samples). 4.8 Chromatographic column - Chronaflex, 400 no long x 19 ne ID (Kontes K-420540-9011 or equivalent). 4.9 6e1 Permeation Chromatograph - 6PC Autoprep 1002 or equivalent, available from Analytical Bio Chemistry Laboratories, Inc. 4.10 Balance - Analytical, capable of accurately weighing 99 i .0001 grams. 4.11 Kudarna-Oanlsh (K-0) Apparatus 4.10.1 Concentrator tube-10 ml, graduated (Kontes K-570050-102S or equivalent). Calibration must be checked. 6round glass stopper (size 19/22 Joint) Is used to prevent evaporation of solvent. 4.11.2 Evaporative flask - 500 mL (Kontes K-57001-0500 or HONS 015549 equivalent). Attach to concantrator tub* with springs (Kontes K-W275Q-0012). 4.11.3 Snyder column Thro*-ball macro (Kontas K503000-0121 or aquivalant). - $. Reagents and Materials 5.1 Raagant safety precautions 5.1.1 The toxicity or carcinogenicity of each reagent used In this method has not been precisely defined; however, each cfcaml' cal compound should be treated as a potential health hae^Aj t from this viewpoint, exposure to these chemicals mus^be- reduced to the lowest possible level by whatever meant'' available. The laboratory Is responsible for snlntalitlny * current awareness file of OSHA regulations regarding the safe handling of the chemical sspeclfled In this method. A reference file of material data handling sheets should also be made available to all personnel Involved In the chemical analysis. 5.1.2 Polychlorinated biphenyls have been tentatively classified as known or suspected, human or mameallan carcinogens. Prlmery standards of these toxic compounds should be pre pared In a hood. 5.1.3 Diethyl ether should be monitored regularly to determine the peroxide content. Under no circumstances should diethyl ether be used with a peroxide content In excess of 50 ppm as an explosion could result. Peroxide test strips manufac tured by EM Laboratories (available from Scientific Products MOhS 015550 Co.. Cat. No. P1126-8 and other suppliers) aro i oi i--mlail for this tost. Procedures for renoval of peroxides froo diethyl ether are included In the Instructions supplied with the peroxide test kit. 5.2 Hexane (nixed hexanes). Isooctane, acetonitrile, nethylene chloride, cyclohexane, and diethyl ether - Pesticide grade. 5.3 Co linen packings (recomeended) 5.3.1 Gas Chrom Q (100/120 nesh) coated with 3X OV-1. 5.3.2 6as Chron Q (100/120 nesh) coated with 1.55 OV-17 1.955 0V-210. . 5.4 Standards >' . 5.4.1 Aroclors 1221, 1232, 1242, 1016, 1254, 1260, 1262, 12$*. *a Prlnary dilutions of various Aroclors In Isooctane will M available from U.S. EPA, EMSL-Clnclnnatl quality Assurance Branch, 26 H. St. Clair Street, Cincinnati, Ohio 45268. 5.4.2 2-Chloroblphanyl, 3>ch1orob1pheny1, and detach1orob1phenyl 5.4.3 Pure Individual PCBs as Identified In the saaple by ness spectroaatry or Indicated by retention data. 5.4.4 Alusrtna (Fisher AS40 or equivalent). 5.4.5 Silica gel (Davison Grade 950 or equivalent). 5.4.6 Flarlsll (PR grade or equivalent). 5.4.7 Sulfuric acid A.C.S. 5.4.8 quality Control Check Staple, Certified Sanple ef PCIs in oil available froa U.S. EPA, EMSL-CInclimatl quality Assurance Branch, 26 M. St. Clair Street, Cincinnati, Ohio 45268. HONS 015551 S.S Standard Stock Solutions - Prepart primary dilutions of each of the* Aroclers or Individual PCBs by weighing approximately 0.01 g of naterlaT within 0.0001 g. Olssolve and dilute to 10.0 ml with Isooctane or hexane. Calculate the concentration in ug/uL. Store the primary dilutions at 4C In 10 to IS ml narrow-mouth, screwcap bottles with Teflon cap liners. Primary dilutions are stable Indefinitely If the seals are maintained. The validity of primary and secondary dilutions must be monitored on a quarterly basis by analyzing EMSL-CI Quality Control Check Sanples. S.S Working Standards - Prepare working standards similar In PCS eg*, position and concentration to the samples by mixing and dllu^lmi' the Individual standard stock solutions. Dilute the m1xture*Jte volume with pesticide quality hexane. Calculate the concentratlegi In ng/uL as the Individual Aroclors or as the Individual PCBs. " Store dilutions at 4C In 10- to 15-mL narrow-mouth, screw-cap bottles with Teflon cap liners. These secondary dilutions can be stored Indefinitely If the seals are maintained. 5.7 Laboratory control standard (LCS) - Prepare a LCS by spiking a PCS-free oil, typical of the matrix normally analyzed, at 50.0 mg/Kg (l.e. transformer oil) with a PC8 mixture typical of those normally found in the samples (l.e. Aroclor 1260 at 50.0 mg/Kg). 5.8 PCS Locator Mixture - Prepare a PCS locator mixture containing 0.1 ng/uL of 2-chloroblphenyl, 0.1 ng/uL 3-chloroblphenyl, 0.5 ng/ui. Aroclor 1242, 0.5 ng/uL Aroclor 1260, and 0.2 ng/uL Aroclor 1268 In hexane (0.1 ng/uL of decachloroblphenyl can be substituted for Aroclor 1268). Use the chromatogram generated by the PCI locator HONS 015552 mixture to hplp Identify the rotontlon times of the various PCI 1 inman coMonty found In coammrclal PCI Mixtures. 6. Smwla Collection nd Hind11 no 6.1 Staple containers should have a volume of 20 art. or more and .have Teflon lined screw caps. 6.2 Stifle Bottle Preparation 6.2.1 Hash all sample bottles and seals in detergent solution. Rinse with tap water and then distilled water. Allow the bottles and seals to drain dry In a contaminant free area. Then rinse seals with pesticide grade hexane and allow to air dry. ' : 6.2.2 Heat sample bottles to 400?C for 15 to 20 min or rinse j. V . with pesticide grade acetone or hexane and allow to a<r,dfy 6.2.3 Store the clean bottles Inverted or sealed until use. ' 6.2.4 Prior to reuse, rinse the bottles and seals 3 times with hexane, air dry, and then proceed to 6.2.1. 6.3 Sample Preservation - The samples should be stored In a cool, dry, dark area until analysis. Storage times In excess of 4 weeks are not recomwnded for unknown or undefined sample matrices. 6.4 Sample Collection 6.4.1 Fill a large container (l.e. 500-mL beaker) from a represen tative area of the sample source (If practical, mix the sample source prior to sampling). 6.4.2 Collect a minimum of two 20-mL samples (Field Sample 1 (FS1) and Field Sample 2 (FS2)) approximately BOS full from the sampling container. HONS 015553 6.4.3 Repeat 6.4.1 and 6.4.2 If there Is a need to monitor sampling precision as described In Section 10.6. 7.Dilution and Analysis 7.1 The approximate PCB concentration of the sample may be determined by X-ray fluorescence, mlcrocoulometry, density measurements, or by analyzing a very dilute mixture of the sample (10,000:1) according to 7.4. 7.2 For samples In the 0 to 100 mg/Kg range, dilute 100:1 In hexane. 7.2.1 Plpet 1.0 mL of sample Into a 100 ml volumetric flask using a 1.0-mL Mohr plpet (for viscous samples, cut the capillary tip off the plpet). Ollute to volume with hexane. Stopper and mix. 7.2.2 Using the same plpet as In 7.2.1, deliver 1.0-mL of sample Into a tared 10-mL beaker weighed to .001 g. Reweigh the beaker to t .001 g to determine the weight of sample used In 7.2.1. 7.2.3 As an alternative to 7.2.1 and 7,2.2, weigh approximately 1 gram to 1.001 g of sample In a volumetric flask and dilute to 100-ri. In hexane. 7.2.4 Analyze the diluted sample according to 7.4 or store the diluted sample In a narrow-mouth bottle with a Teflon lined screw-cap. 7.3 For samples above 100 mg/Kg In concentration, dilute 1000:1 In hexane. 7.3.1 Plpet 0.10 mL of sample Into a 100-mL volumetric flask using a 0.10 mL Mohr plpet. Ollute to volume with hexane, stopper and mix. , HONS 015554 7.3.2 Using the sane pi pet as In 7.3.1. deliver 0.10 art. of SMple Into a tared 10 id beaker to t .0001 g. Reweigh the beaker to determine the weight of sample used In 7.3.1. 7.3.3 As an alternative to 7.3.1 and 7.3.2, weigh approximately 0.1 gram to .0001 g of sample and In a 100 mL volumetric flask dilute to volume with hexane. 7.3.4 Analyze the diluted sample according to 7.4 or store the diluted sample In a narrow-mouth bottle with a Teflon-lined screw-cap. 7.3.5 If the concentration of PCBs Is still too high for the chromatographic system, prepare secondary dilutions from 7.3.1 or 7.3.3 until acceptable levels are obtained. 7.4 Analyze the sample by Injecting the hexane mixture Into the gas chromatograph using auto Injectors or the solvent flush techn1que(3). Note: When semi-specific detectors are used, cleanup techniques (See Section 4.2.2) should be routinely Incorporated Into the analysis scheme prior to Injection. 7.4.1 Recommended Injection volumes: Halogen specific detector - 4 to 5 ul, electron capture detector-2 to 3 ul. Smaller volumes may be Injected when auto Injectors are used If the resulting method detection limits are acceptable. 7.5 If the resulting chromatogram shows evidence of column flooding or nonlinear responses due to excess sample, further dilute the sample ` according to 7.3.5. 7.6 Determine whether or not PCBs are present In the sample by HONS Q1SS55 comparing the sample chromatogram to that of the PCB locator mixture Section 5.8. 7.6.1 If a series of peaks In the sample match some of the reten tion times of PCBs In the PCB locater mixture, attempt to Identify the source by comparing chromatograms of each standard prepared from commercial mixtures of PCBs (See Section 5.6). Proceed to Section 11.4 if the source of PCBs Is Identified. 7.6.2 If the sample contains a complex mixture of PCBs, proceed to Section 11.5. 7.6.3 If a 1000:1 or higher dilution ratio sample was analyzed and no measurable PCB peaks were detected, analyze an aliquot of sample diluted to 100:1. 7.6.4 If several PCB Interference problems are encountered or If PCB ratios do not match standards, proceed to Section 8, use alternate columns, or use GC/MS to verify whether or not the nonrepresentative patterns are due to PCBs. B. Cleanup - Several tested cleanup techniques are described. Depending upon the complexity of the sample, one or all of the techniques may be required to resolve the PCBs from Interferences. 8.1 Acid Cleanup 8.1.1 Place 5.0 mL of concentrated sulfuric acid Into a 40-nt narrow-mouth screw-cap bottle. Add 10.0 mL of the diluted sample. Seal the bottle with a Teflon-lined screw-cap and shake for 1 minute. 8.1.2 Allow the phases to separate, transfer the sample (upper HONS 015556 phase) to a clean narrow-mouth screw-cap bottle. Seal with a Teflon-lined cap. 8.1.3 Analyze according to Section 7.4. .- 8.1.4 If the sample Is highly contaminated, a second or third acid cleanup may be employed. NOTE: This cleanup technique was tested over a period of about S months using both electron capture and electrolytic conductivity detectors. Care was taken to exclude any samples that formed an emulsion with the acid. The sample was withdrawn well above the sample-acid Interface. Under these conditions no adverse effects associated with column performance and detector sensitivity to PCBs were noted. This cleanup technique could adversely affect the chromato graphic column performance for future acid degradable samples. 8.2 Florlsll Column Cleanup 8.2.1 Variances between batches of Florlsll may affect the elution volume of the various PCBs. For this reason, the volune of solvent required to completely elute all of the PCBs must be verified by the analyst. The weight of Florlsll can then be adjusted accordingly. 8.2.2 Place a 20.0 g charge of Florlsll, activated at 130C, Into a Chromaflex column. Settle the Florlsll by tapping the column. Add about 1 cm of anhydrous sodium sulfate to the top of the Florlsll. Pre-elute the column with 70 to 80 nri. of hexane. Just before the exposure of the sodium HONS 015557 sulfate layer to air, stop the flow. Discard the eluate. 8.2.3 Add 2.0 ml of the undiluted sample to the column with a 2- ml Mohr plpet. (For viscous samples, cut the capillary tip off the plpet.} Add 225 ml of hexane to the column. Carefully wash down the Inner wall of the column with a small amount of the hexane prior to adding the total volume. Collect and discard the first 25.0 ml. 8.2.4 Collect exactly 200 ml of hexane eluate In a 200-mL volu metric flask. All of the PCBs must be In this fraction. 8.2.5 Using the same plpet as In 8.2.2, deliver 2.0 ml of sample Into a tared 10 ml beaker weighed to t 0.001 g. Reweigh the beaker to determine the vfclght of the sample diluted to 200 ml. 8.2.B Analyze the sample according to Section 7.4. 8.3 Alumina Column Cleanup 8.3.1 Adjust the activity of the alumina by heating to 200C for 2 to 4 hours. When cool, add 3% water (wt:wt) and mix until uniform. Store In a tightly sealed bottle. Allow the deactivated alumina to equilibrate at least 1/2 hr before use. Reactivate weekly. 8.3.2 Variances between batches of alumina may affect the elution volume of the various PCBs. For this reason, the volume of solvent required to completely elute all of the PCBs must be verified by the analyst. The weight of Alumina can then be adjusted accordingly. 8.3.3 Place a 50.0 g charge of alumina Into a Chromaflex column. HONS 015558 Settle the alumina by tapping. Add about 1 cm of anhydrous sodium sulfate. Pre-elute the column with 70 to 80 ml of hexane. Just before exposure of the sodium sulfate layer to air, stop the flow. Discard the eluate. . 8.3.4 Add 2.5 mL of the undiluted sample to the column with a 5-mL Mohr plpet. (For viscous samples, cut the capillary end off the plpet.) Add 300 mL of hexane to the column. Carefully wash down the Inner walls of the column with a small volume of hexane prior to adding the total volume. Collect and discard the 0 to 50 mL fraction. 8.3.5 Collect exactly 250 mL of the hexane in a 250-mL volumetric flask. All of the PCBs must be In this fraction. 8.3.6 Using the same plpet as In 8.3.4, deliver 2.5 mL of sample Into a tared 10-nL beaker 0.001 g. Reweigh the beaker to determine weight of sample diluted to 250 mL. 8.3.6 Analyze the sample according to Section 7.4. 8.4 Silica Gel Column Cleanup. 8.4.1 Activate silica gel at 135C overnight. 8.4.2 Variances between batches of silica gel may affect the elution volume of the various PCBs. For this reason, the volume of solvent required to completely elute all of the PCBs must be verified by the analyst. The weight of silica gel can then be adjusted accordingly. 8.4.3 Place a 25 g charge of activated silica gel Into a Chroma- flex column. Settle the silica gel by tapping the column. Add about 1 cm of anhydrous sodium sulfate to the top of the silica gel. - HONS 015559 8.4.4 Preelute the column with about 70 to 80 mL of hexane. 01scard the eluate. Just before the exposure of the sodium sulfate layer to air, stop the flow. 8.4.5 Add 2.0 mL of the undiluted sample to the column with* 2-mL Mohr plpet. (For viscous samples, cut the capillary tip off the plpet.) 8.4.6 Wash down the inner wall of the column with 5 mL of hexane. 8.4.6 Elute the PCBs with 195 mL of 10S dlethylether In hexane V:V. 8.4.8 Collect exactly 200 mL of the eluate In a 200-mL volumetric flask. All of the PCBs must be In this fraction. 8.4.9 Using the same plpet as In 8.4.5, deliver 2.0 mL of sample Into a tared 10-mL beaker (t 0.001 g). Reweigh to determine the weight of sample diluted to 200 mL. 8.4.10 Analyze the sample according to Section 7.4. 8.5 Gel Permeation Cleanup 8.5.1 Set up and calibrate the gel permeation chromatograph with an SX-3 column according to the Instruction manual. Use 15X methylene chloride In cyclohexane (V:V) as the mobll phase. 8.5.2 Place 1.0 mL of sample Into a 100-mL volumetric flask, using a 1-mL Mohr plpet. (For viscous samples, cut the capillary tip off the plpet.) 8.5.3 Dilute the sample to volume, using 15% methylene chloride In cyclohexane (V:V). 8.5.4 Using the same plpet as In 8.5.2 deliver 1.0 mL of sample Into a tared 10-mL beaker t 0.001 g. Reweigh the beaker 0.001 g to determine the weight of sample used In 8.5.2. HONS 015560 8.5.5 As .an alternative to 8.5.2 and 8.5.3 weigh approximately 1 gram 0.001 g of sample and dilute to 100.0 ml in 15* methylene chloride In cyclohexane (V:V). 8.5.6 Inject 5.0 ml of the diluted sample Into the Instrument. Collect the fraction containing the Cl^ through C1-,Q PC8s (see operator's manual) In a K-0 flask equipped with a 10-ral ampul. 8.5.7 Concentrate the 8.5.4 fraction down to less than 5 ml. Using evaporative concentration techniques. 8.5.8 Ollute to 5.0 ml with hexane, then analyze according to Section 7.4. Be sure to use 100 ml as the dilution volume for the final calculation. 8.6 Acetonitrile Partition 8.6.1 Place 10.0 ml of the previously diluted sample Into a 125-ml separatory funnel with enough hexane to bring the final volume to 15 ml. Extract the sample four times by shaking vigorously for one minute with 30-ml portions of hexane- saturated acetonitrile. 8.6.2 Combine and transfer the acetonitrile phases to a one-liter separatory funnel and add 650 ml of distilled water and 40 ml of saturated sodium chloride solution. Nix thoroughly for 30 to 35 seconds. Extract with two 100 ml portions of hexane by vigorously shaking about 15 seconds. 8.6.3 Combine the hexane extracts In a one-liter separatory funnel and wash with two 100-ml portions of distilled water. Discard the water layer and pour the hexane layer through a MOWS 015561 3 to 4 Inch anhydrous sodium sulfate column Into a 500-mi. K-D flask equipped with a 10 mt ampul. Rinse the separa tory funnel and column with three 10 ml portions of hexane. 8.6.4 Concentrate the extracts to 6 to 10 mL In the K-0 evaporator In a hot water bath, then adjust the volume to 10.0 mL. 8e sure to use the correct dilution volume (8.6.1) for the final calculation. 8.6.5 Analyze according to Section 7.4. 8.7 Florlsll Slurry Cleanup 8.7.1 Place 10 mi. of the diluted sample Into a 20-mL narrow-mouth screw cap container. Add 0.25 g of Florlsll.. Seal with a Teflon lined screw cap and shake for 1 minute. 8.7.2 Allow the Florlsll to settle then decant the treated solu tion Into a second container. Analyze according to Section 7.4. 9. Calibration 9.1 Single Point Calibrations - Prepare calibration standards from standard stock solutions In hexane that are close to the unknown In composition and In concentration. When using an electrolytic con ductivity detector, if the sample response Is In the low level nonlinear detection area, then the calibration point must be within 20X of the sample. The ECO must be operated only within Its linear response range. 9.2 As an alternative to 9.1, prepare a calibration curve for each Aroclor or PCB detected In the sample. The standard curve must HONS 015562 contain at least 3 points, two of which must bracket the sample concentration. When using an electrolytic conductivity detector. If the sample response Is In a low level nonlinear area of the calibration curve, two of the calibration points must be within 20X of the unknown. The calibration curve must be checked dally, using the Quality Check Sample Section 5.7. If the calibration curve Is not within 15 percent of the Quality Check Sample, recalibrate the Instrument. It will be necessary to correct electron capture data for recovery (See Section 3.4). Use the recovery value determined the same day the calibration curve was generated. The corrected value must be within 15X of the spike value, otherwise the instru ment must be recalibrated. 10. Quality Control 10.1 Each laboratory that uses this method Is required to operate a formal quality control program. The minimum requirements of this program consists of an Initial demonstration of laboratory capa bility and the analysis of spiked samples as a continuing check on performance. The laboratory Is required to maintain performance records to define the quality of data that Is generated. After January 1, 1982. ongoing performance checks must be compared with established performance criteria to determine If the results of analyses are within accuracy and precision limits expected of the method. 10.1.1 Before performing any analyses, the analyst must demonstrate the ability to generate acceptable accuracy and precision with this method. This ability Is established as described in Section 10.2. HONS 015563 10.1.2 In recognition of the rapid advances that are occurring In chromatography, the analyst Is permitted certain options to improve the separations or lower the cost of measurements. Each time such modifications are made to the method, the analyst Is required to repeat the procedure In Section 10.2. 10.1.3 The laboratory must spike and analyze a minimum of 10 percent of all samples to monitor continuing laboratory performance. This procedure Is described In Section 10.4. 10.2 To establish the ability to generate acceptable accuracy and precision In the use of this method, the analyst must perform the following operations. 10.2.1 For each coeiserclal PCB mixture or Individual PCB normally measured, select a spike concentration representative of high Interest levels In the samples (25 to 75 mg/kg). Using stock standards, prepare a quality control check sample concentrate In Issoctane 1000 times more concentrated than the selected concentrations. 10.2.2 Using a microsyringe, add 100 uL of the check sample concen trate to each of a minimum of four 100-mL aliquots of PCB-free oil. A representative waste oil may be used In place of the clean oil, but one or more additional aliquots must be analyzed to determine background levels, and the spike level must exceed twice the background level for the test to be valid. Analyze the aliquots according to the method beginning In Section 7. 10.2.3 Calculate the average percent recovery, (R), and the HONS 015564 relative standard deviation of the concentration found. Waste oil background corrections mist be made before R calculations are performed. 10.2.4 Using the appropriate data from Tables 7, 8, and 9, deter mine the recovery and single operator precision expected for the method and compare these results to the values calcu lated In Section 10.2.3. If the data are not comparable, the analyst must review potential problem areas and repeat the test. 10.2.5 After January 1, 1982, the values for R and s must meet method performance criteria provided by the U.S. EPA, Environmental Monitoring and Support Laboratory, Cincinnati, Ohio, before any samples may be analyzed. 10.3 The analyst must calculate method performance criteria and define the performance of the laboratory for each spike concentration and parameter being measured. 10.3.1 Calculate upper and lower control limits for method perform ance: Upper Control Limit (UCL) R + 3 s Lower Control Limit (LCL) R - 3 s where R and s are calculated as In Section 10.2.3. The UCL and LCL can be used to construct control charts^ that are useful In observing trends In performance. After January 1, 1982, the control limits above must be replaced by method performance criteria provided by the U.S. Environmental Protection Agency. HONS 015565 10.3.2 The laboratory must develop and maintain separate accuracy statements of laboratory performance for waste oil samples. An accuracy statement for the method is defined as R s. The accuracy statement should be developed by the analysis of four aliquots of waste oil as described In Section 10.2.2, followed by the calculation of R and s. Alter nately, the analyst may use four waste oil data points gathered through the requirement for continuing quality control In Section 10.4. The accuracy statements should be updated regularly. 10.4 The laboratory Is required to collect a portion of their samples in duplicate to monitor splice recoveries. The frequency of spiked sample analysis must be at least 10X of all samples or one sample per month, whichever Is greater. One aliquot of the sample must be spiked and analyzed as described In Section 10.2. If the recovery for a particular parameter does not fall within the control limits for method performance, the results reported for the parameter In all samples processed as part of the same set must be qualified as described In Section 11.9. The laboratory should monitoring the frequency of data so qualified to ensure that It remains at or below 5%. 10.5 Before processing any samples, the analyst should demonstrate through the analysis of a PCB-free oil sample, that all glassware and reagent Interferences are under control. Each time a set of samples Is analyzed or there Is a change In reagents, a laboratory reagent blank should be processed as a safeguard against contamin ation. ' HONS 015566 10.6 It Is recoaNnded that the laboratory adopt additional quality assurance practices for usa with this nethod. Tha spaclflc practical that ara most productive depend upon the needs of the laboratory and the nature of the samples. Field duplicates fray be analyzed to fronltor tha precision of the sampling technique. When doubt exists over tha Identification of a peak on the chroaetogara, confirmatory techniques such as gas chronatography with a dlsslallar coluan, specific elcaant detector, or aass spectroneter aust be used. Whenever possible, the laboratory should perform analysis of standard reference aaterlals and participate In relevant performance evaluation studies. 10.7 Analyze the quality control check sanple section 5.7 dally before' any sanples are analyzed. Instruwnt status checks, calibration curve validation and long term precision are obtained free these; data. In addition, response data obtained free the quality control check standard can be used to estleate the concentration of the unknowns. From this Information, the appropriate standard dllu* tlons can be determined for single point calibrations. 10.8 Quarterly, analyze an EKSL-Clnclnnatl Quality Control Sample (Section 5.4.8) of PCBs In oil or whenever new standard dilutions are prepared. 10.8.1 The results of tha EMSl Quality Control Saaple should agree within 20X of the true value. If they do not, tha analyst must check each step In the standard preparation procedure to solve the problem. 11. Calculations HONS 015567 K 11.1 Locate each PCS In the sanple chromatogram by comparing the reten tion tlM of the suspect peak to the rotontlon data gathered fro* analytlpg standards and Intarftranca free Quality Control Se>1es. The width of the retention tine window used to make Identifications should be based upon neasurenents of actual retention ties vari ations of standards over the course of a day. Three tines the standard deviation of a retention tine for each PCB can be used to calculate a suggested window size; however, the experience of the analyst should weigh heavily In the Interpretation of chromatograms. 11.2 If the response for the peak exceeds the working range of the system, dilute according to Section 7.3.5. - 11.3 If the neasuraent of the peak response Is prevented by the presence of Interferences, further cleanup Is required. ' 11.4 If the parent Aroclor(s) or-PCB(s) are Identified In the sanple, calibrate according to Section g. The concentration of the PCBs In the sanpl* are calculated by cohering the sun of the responses for each PCB In the standard to the sun of all of the PCBs In the sanple. This Is particularly Important as sanple concentrations approach within 20 percent of 50 ng/kg or any other EPA regulated concentration. If calculations are based upon a single PCB peak or upon a snail percentage of the total PCB peaks, serious errors nay result. Peaks conprlslng less than 50X of the total can be dis regarded only If 1) Interference problems persist after cleanup, 2) when the source of PCBs Is obvious 3) whan the concentration of PCBs are not within t 20* of an EPA controlled value l.e., 50 ng/kg. HONS 01556b 11.4.1 Measure tht peak height or peak area of each peak Identified as'a PCS In 11.1 In both the saaple and the standard. 11.4.2 Use the following foraula to calculate the concentration of PCBs In the sample: - Concentration ng/Kg 8 x vt mxb a Sao of standard Peak Heights (areas) * ng of standard injected */ng 8 . SB.- mfvL Vt dilution voluae of saaple In nsL ! U weight of the saaple In grass 4 11.5 If the parent Aroclor(s) or source of PCBs Is not apparent, saldi late the concentration according to the procedure of Webb and McCall (5). The concentration of each PC8 Is detenslned Individu ally than added together to detarsrtne the total PCS content of the sessile. Each PCS Identified In the saaple oust be Included In these calculations. 11.5.1 Saall variations between Aroclor batches isake it necessary to obtain standards prepared from a specific source of Aroclors. Prlaary dilutions of these reference Aroclors will be available In 1981 froa the Envlronaental Monitoring and Support Laboratory, Quality Assurance Branch, Cincinnati, Ohio 45268. 11.5.2 Analyze a standard alxture of Aroclors 1242, 1254 and 1260 under the conditions shown In Figures 3, 5, and 6. Analyze the saaple under the sase conditions. Coapare the resulting HONS 015569 f Stamford chromatograms to those shown In Figures 3, 5, and 6. Each PCB peak must be resolved as well or better than those shown In the figures. Oetermlne the relative reten tion time (RRT) of each peak In the standards w.lth respect to p.p'-DDE or assign the RRT shots) In the figures to the corresponding peak In the standard. Identify the RRT of each PCB In the sample by compering the sample chromatogram to the standard chromatograms. 11.5.3 Identify the most likely Aroclor(s) present In the sample, using the Identification Flow Chart Figure 7. . 11.5.4 Analyze standards according to Section 9, using the appro priate Aroclors. 11.5.5 Determine the Instrwent response factor (A) for each Individual PCB, using the following formula: Ngi Ng of Aroclor standard Injected Obtain mean weight percent from Tables 1 through 6. 11.5.6 Calculate the concentration of each PCB In the sample using the following formula: Concentration mg/Kg 8 x Vt AX* A Response factor from 11.4.5 B Peak sample rne/ul Vf dilution volume of sample In ml W weight of sample In grams HONS 015570 . The concentration of each PCS must be calculated and added together to obtain the total amount of PCBs present. 11.6 Report-at1 data In mg/Kg. 11.7 Round off all data to 2 significant figures. ' 11.8 Sue all Arocldrs and report what was used as the standard. For example, 57 mg/Kg measured as Aroclor 1260 or 57 mg/Kg measured as Aroclors 1242 and 1260. 11.9 For samples processed as part of a set where the laboratory spiked sample recovery falls outside of the control limits In Section 10.4, data for the affected parameters must be labeled u suspect. 11.10 For electron capture analyses determine the actual recovery for each sample In the uncorrected 40 to 50 mg/kg concentratlon'range (see Section 3.4). Report the corrected value and the recovery.! 12. Single laboratory accuracy and Precision 12.1 The single laboratory data shown In Tables 7 through 9 were generated using the recamended procedures described In this method to analyze both spiked and nonsplked oil samples of varying degrees of complexity using a halogen specific detector and an electron capture detector. These data were generated by the Physical and Chemical Methods Branch, Environmental Monitoring and Support Laboratory, Cincinnati, Ohio, 45266. References 1. Elcholberger, J. W., L.E. Harris, and H. L. Budde, Anal. Chen.. 46, 227 (1974). " 2. Federal Register, Vol. 41, No. 232, Wednesday. December 1, 1976. 3. White, L. 0., et al., AIHA Journal. 31, 22S, (1970). 4. `Handbook of Analytical quahty Control in Water and Wastewater . Laboratories,* EPA-600/4-79-019, U.S. Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45266, March 1979. 5. Webb, R.S., and A.C. McCall, J. Chron. Scl.. 11, 366 (1973). MQNS 015571 L RRT*> TABLE 1 Conposltlon of Aroclor 1221* Mean Height Percent Relative Std. Oev. c Nunber of Chlorines d 11 14 16 19 21 28 32 37 40 Total 31.8 19.3 10.1 2.8 20.8 5.4 1.4 1.7 93.3 IS.8 9.1 9.7 9.7 9.3 13.9 30.1 48.8 . 1 1 2 2 2 2 asx 3 15* 2 10* 3 90S 3 3 ' .. _ 1 Data obtained fron (*). b Retention tine relative to p,p'-0OE*100. Measured fron first appearance of solvent. Overlapping peaks that are quantitated as one peak are bracketed. c Standard deviation of seventeen results as a percentage of the nean of the results d fron SC-MS data. Peaks containing Mixtures of Isoners of different chlorine nunbers are bracketed. MOMS 015572 L RRT" TABU 2 Composition of Aroclor 1232* Nam Might Percent Relative Std. Oev. c Number of Chlorines d 11 14 16 20 21 28 32 37 40 47 54 SB 70 78 Total 16.2 9.9 17.8 17.8 9.6 3.9 6.8 6.4 4.2 3.4 2.6 4.6 1.7 94.2 3.4 2.5 2.4 2.4 3.4 4.7 2.5 2.7 4.1 3.4 3.7 3.1 . 7.5 1 1 2 2 2 2 408 3 608 3 3 3 4 3 338 5 678 4 4 908 5 108 4 1 Beta obtained from (5). 6 Retention tine relative to p,p'-D0100. Measured from first appearance of solvent. Overlapping peaks that are quantitated as one peak are bracketed. e Standard deviation of four results as a naan of the results. d From BC-MS data. Peaks containing mixtures of Isomers of different chlorine lumbers are bracketed. HONS 015573 RRT*> n 16 21 28 32 37 40 47 54 58 70 78 84 98 104 148 Total TABLE 3 Composition of Aroclor 1242* Mean Weight Percent Relative Std. Dev. c Number of ' Chlorines b 1.1 2.9 11.3 11.0 6.1 11.5 11.1 8.8 6.8 5.6 10.3 3.6 2.7 1.5 1.6 1.0 98.5 35.7 4.2 3.0 5.0 4.7 5.7 6.2 4.3 2.9 3.3 2.8 4.2 9.7 9.4 20.4 19.9 1 2 2 2 25X 3 75X 3 3 3 4 3 33X 4 67X 4 4 90X 5 10X 4 5 5 5 8SX 6 15X 5 75X 6 2SX . > ; .* * Oats obtained from (). b Retention tint relative to p,p'-OOE*100. Measured from first appoaranca of solvant. c Standard deviation of six rasults as a parcantago of the naan of tna rasults. d From SC-MS data. Peaks containing mixtures of Isoamrs of different chlorine numbers are bracketed. HONS 015574 RRT* 21 28 32 47 40 47 54 58 70 78 84 98 104 112 125 146 Total TABU 4 Composition of Aroclor 1248* Mean Height Percent Relative Std. Oev. c Number of Chlorines 1 1.2 5.2 3.2 8.3 8.3 15.6 9.7 9.3 19.0 6.6 4.9 3.2 3.3 1.2 2.6 1.5 103.1 23.9 3.3 3.8 3.6 3.9 1.1 6.0 5.8 1.4 2.7 2.6 3.2 3.6 6.6 5.9 10.0 2 3 3 3 3 8SS 4 15S 4 3 106 4 90S 4 4 80S 5 20S 4 5 5 4 10S 5 90S 5 5 90S 6 10S 5 85S 6 1SS * Data obtained froai ($). b Retention tlM relative to p,p'-00*100. Measured from first appearance of solvent. ' c Standard deviation of six results as a percentage of the wan of the results. d Fros 6C-MS data. Peaks containing Mixtures of Isomers of different chlorine numbers are bracketed. HONS 015575 t RUT* TABU S Composition of Aroclor 1254 Mean weight Percent Relative Std. Oev. c Number of ' Chlorines 4 47 54 $8 70 84 98 104 125 146 160 174 203 232 Total 6.2 2.9 1.4 13.2 17.3 7.5 13.6 15.0 . 10.4 1.3 8.4 1.8 1.0 100.0 3.7 2.6 2.8 2.7 1.9 5.3 3.8 2.4 2.7 8.4 5.5 18.6 26.1 4 4 4 4 25% 5 75% 5 .5 5 5 70% ' . 5 30% 5 30% 6 70% 6 6 6 7 1 Data obtained from ($). b Retention tlmi relative to p,p'-D0E*100. Measured from first appearance of solvent. c Standard deviation of six results as a percentage of the mean of the results. d from SC-MS date. Peeks containing mixtures of Isomers of different chlorine numbers are bracketed. HONS 015576 RRT* 70 84 96 104 117 12S 144 160 174 203 232 244 280 332 372 448 528 Total Composition of Aroclor 1260* Main Height Percent 2.7 4.7 3.8 3.3 12.3 14.1 4.9 12.4 9.3 9.8 n.o 4.2 4.0 .6 1.5 98.6 Relative Std. Oev. ' 6.3 1.6 3.5 6.7 3.3 3.6 2.2 2.7 4.0 3.4 2.4 5.0 8.6 25.3 10.2 Number- of Chlorines 4 5 5 t 5 60S 6 40* 6 5 15* 6 85* 6 6 50* 7 50* 6 6 10* 7 90* f 6 10* 7 90S 7 7 8 8 8 - . - * Otto obtained from (5). b Retention tlat relative to p,p'-D0E100. Measured from first appearance of solvent. Overlapping peaks that are quantitated as one peak are bracketed. c Standard deviation of six results as a man of the results. 4 FroM 6C-KS data. Peaks containing nlxtures of Isomers of different chlorine numbers are bracketed. * Composition determined at the canter of peak 104. f Composition determined at the center of peak 232. HONS 015577 TABLE 7 ACCURACY ANO PRECISION USING SPIKED MOTOR OIL Hlutlon Method Spike Ratio Oatactor Clean-uiI ng/Kg Aroclor Spike Ave. Cone. Found ng/Kg (Precision) Rel. Std. Oevlatlon X (Accuracy) Percent Recovered Nuahar ef 01letlee 00:1 00:1 00:1 00:1 03:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 00:1 30:1 00:1 90:1 00:1 00:1 00:1 X:1 00:1 X:1 HED None ECO None HED None ECO None HED 8.1 ECO 8.1 HED 8.1 ECO 8.1 HEO 8.2 ECO 8.2 HEO 8.2 ECO 8.2 HEO 8.3 ECO 8.3 HED 8.3 ECO 8.3 HED 8.4 ECO 8.4 HED 8.4 ECO 8.4 HED 8.5 ECO 8.5 HED 8.5 ECO 8.5 HED 8.6 ECO 8.6 HED 8.6 ECO 8.6 30.3 X.3 31.1 31.1 X.3 X.3 31.1 31.1 X.3 X.3 31.1 31.1 X.3 X.3 31.1 31.1 X.3 X.3 31.1, 31.1 X.3 X.3 31.1 31.1 X.3 31.1 X.3 31.1 1242 1242 1260 1260 1242 1242 1260 1260 1242 1242 1260 1260 1242 1242 1260 1260 1242 1242 1260 1260 1242 1242 1260 1260 1242 1242 1260 1260 28.2 26.7* 27.2 23.9 28.4 25.4 28.1 24.3 X.7 27.3 X.9 31.0 X.3 28.9* 29.8 X.8 29.4 26.4* 29.4 23.6 31.9 23.4* 33.6 X.9 34.4 23.4 29.1 27.0 4.2 5.7 2.0 2.2 11.5 6.1 8.0 7.8 2.4 10.2 3.6 8.6 8.6 5.0 4.7 6.5 5.8 5.3 5.2 4.5 8.5 3.0 9.2 5.5 3.8 4.4 4.2 4.6 93.1 88.1 87.5 76.8 93.7 83.8 X.3 78. K JOl 90.1 99.4 98.7 100. : 96.4 95.8 99.0 97.0 87.1 94.5 75.9 105 77.2 108 99.4 107 77.2 96.7 86.7 S 3 S 3 3 3 3 3 4 4 4 4 3 3 3 3 3 3 3 3 3 2 3 3 4 4 4 4 Severe Interference probleen in alutlon area of 1242. Measurement based upon only 3 of the 10 nonaally resolved major peaks. Clean-up techniques 8.1. 8.2, 8.3. 8.4. 8.5. and 8.6 did not Inprove the quality of the 1242 chraaatogran. If this Mere an unknown sample. It would be Inpossible to qualitatively Identify the presence of Aroclor 1242 using ECO. The HED provided an Interference-free chroantogran. HONS 015570 TABLE 8 ACCURACY AMO PRECISION USING WASTE TRANSFORMER FLUIDS Jaeple Dilution Ratio 1260 Method Spike Detector Clean-up ng/Kg Av. Cone. Found (Precision) Rel. Std. Oevlitlon t (Accuracy) Percent Recovered of D1letter A 100:1 ECO None -- A HED None A ECO 8.1 -- A HED 8.1 ' -- A ECO 8.2 -- A HED 8.2 -- A ECO 8.3 -- A HED 8.3 -- A A N ECO 8.4 -- MED 8.4 A A Aa ECO 8.5 -- HED 8.5 -- ECO None 27,0 A HED None 27.0 8 1000:1 ECO None 8 HED a -- 8 8 a ECO HED a 455 a 455 c c c c 1000:1 ECO HED ECO HED None a a a -- 300 300 . dart waste oil - black waste oil with suspended solids - clear waste oil - all saeplas contained Aroclor 1260 - duplicate walyses wade at each dilution 22.6 27.0 22.8 29.7 22.4 28.2 22.7 27.8 20.9 30.2 23.8 28.6 45.0 55.2 452 471 875 916 284 300 607 686 3.6 1.7 2.5 1.4 1.0 2.2 1.3 2.8 mm 0.3 4.1 3.3 1.5 0.8 1.2 0.5 2.0 1.2 1.4 3.6 3.9 102 7* __ 7 96 7* 99 7* mm 7 7 104 7* 114 7 HONS 015579 TABLE 9 Accuracy end Precision and Limit of Detaction Date Basalts of Analysts of Shall Transformer Fluid . Spiked with PCBs at S.O and 27 mg/Kg Electron Capture Method (100:1 Dilution) Aroclor 1221 1242 1254 1260 Spike (mg/kg) S.O 5.0 5.0 5.0 Number of Avg Standard X Recovery MOL* Analyses (9/kg) Deviation (9/kg) 7 7.5 14 3.8 7 4.1 14 4.7 0.43 0.18 0.08 0.18 150 1.4 76 0.5 82 0.2 94 0.5 Hall Method (100:1 Ollutlon) Aroclor 1221 1242 1254 1260 Spike (9/kg) 5.0 5.0 5.0 5.0 Number of Avg Standard Analyses (*9/kg) Deviation 6 7.5 7 5.9 6 5.8 7 ' 5.4 0.23 0.17 0.16 0.10 X Recovery MOL* (g/kf) 150 0.7 V ` . 118 0.5 J* . : 116 0.5 . * 108 0.3 - Shell Transformer Oil1 27 ppm Aroclor 1260 (100:1 Dilution) Method -Spike (g/kg) Number of Analyses Avg (mg/kg) Bel. Std. Deviation X X Recovery Electron Capture Hall 700-A 27 27 14 24.0 7 28.3 .70 2.1 89 105 netnoo Detection Limit at Hi confidence that the value Is not zero, note; At these values It would be leposslble to Identify Aroclor patterns with any degree of confidence. 1 mg/Kg appears to be a reasonable KM.. where: KH. *(n-l,.99) S WL * t(n-1..99) (S) the method detection limit the students' t value appropriate for a 99* confidence level and a standard deviation estimate with n-1 degrees of freedom, standard deviation of the replicate analyses HONS 015580