Document nmY1R5x5KpQJ2xp5Yk8Oe4N7w
RECOMMENDED PROCEDURE FOR DETERMINATION OF RECOVERY COEFFICIENTS
FOR 3M ORGANIC VAPOR MONITOR #3500
The 3M Company publishes the Recovery Coefficient for each organic material on the Qualification Data Sheet. This number represents the mean value 2 stan dard deviations from the tests conducted at 3M. A minimum of 5-fold replication (both monitors and standards) of the recovery test is performed at 3M. However, we encourage the user to verify the recovery coefficients, since techniques and presence of multiple contaminants can affect recovery coefficients.
The recovery coefficient is determined by vapor-state spiking of monitors. This is accompanied by the following steps:
-- Remove the white barrier film from the monitor.
-- Place a 2.5 cm diameter Whatman #5 filter paper on the spacer plate.
-- The closure cap is then snapped on.
-- Calculate the amount of material to be injected to correspond to the amount sampled by the monitor at a concentration of 1/2 TLV for 4 hours (approximately a 7.0 liter sample).
Example:
For Toluene with a TLV of 100 ppm or 375 mg/m3, the spike would have:
37~5 (0.007) = 1.31 mg.
-- Inject the known quantity of the organic material onto the filter paper through the center port.
-- The monitor is allowed to sit for 16-24 hours to allow total transfer of the organic material from the filter paper to the sorbent before elution.
-- Remove filter paper from monitor.
-- Proceed with elution and determination of amount recovered by G.C. analysis.
For further assistance, call 3M Occupational Health and Safety Products Technical Service toll free on 1-800-328-1300, or write 3M Company, 3M Center, Occupational Health and Safety Products Division, P.O. Box 33155, St. Paul, MN 55101.
R-OVMRP-A(39,3)R2
Occupational Health and Safety Products Division/3M 3M Center St Paul. Minnesota 55101
3H 105489
Litho in U.S.A.
RECOMMENDED PROCEDURE FOR DETERMINATION OF RECOVERY COEFFICIENTS
FOR 3M ORGANIC VAPOR MONITOR #3500
The 3M Company publishes the Recovery Coefficient for each organic material on the Qualification Data Sheet. This number represents the mean value 2 stan dard deviations from the tests conducted at 3M. A minimum of 5-fold replication (both monitors and standards) of the recovery test is performed at 3M. However, we encourage the user to verify the recovery coefficients, since techniques and presence of multiple contaminants can affect recovery coefficients.
The recovery coefficient is determined by vapor-state spiking of monitors. This is accompanied by the following steps:
-- Remove the white barrier film from the monitor.
-- Place a 2.5 cm diameter Whatman #5 filter paper on the spacer plate.
-- The closure cap is then snapped on.
-- Calculate the amount of material to be injected to correspond to the amount sampled by the monitor at a concentration of 1/2 TLV for 4 hours (approximately a 7.0 liter sample).
Example:
ForToluene with aTLV of 100 ppm or 375 mg/m3, the spike would have:
375 --...(0.007) = 1.31 mg.
-- Inject the known quantity of the organic material onto the filter paper through the center port.
-- The monitor is allowed to sit for 16-24 hours to allow total transfer of the organic material from the filter paper to the sorbent before elution.
-- Remove filter paper from monitor.
-- Proceed with elution and determination of amount recovered by G.C. analysis.
For further assistance, call 3M Occupational Health and Safety Products Technical Service toll free on 1-800-328-1300, or write 3M Company, 3M Center, Occupational Health and Safety Products Division, P.O. Box 33155, St. Paul, MN 55101.
R-OVM RP-A(39,3) R2
Occupational Health and Safety Products Division/3M 3M Centei St Paul, Minnesota 55101
Litho in U.S.A.
3W 105490
RECOMMENDED PROCEDURE FOR DETERMINATION OF RECOVERY COEFFICIENTS
FOR 3M ORGANIC VAPOR MONITOR #3500
The 3M Company publishes the Recovery Coefficient for each organic material on the Qualification Data Sheet. This number represents the mean value 2 stan dard deviations from the tests conducted at 3M. A minimum of 5-fold replication (both monitors and standards) of the recovery test is performed at 3M. However, we encourage the user to verify the recovery coefficients, since techniques and presence of multiple contaminants can affect recovery coefficients.
The recovery coefficient is determined by vapor-state spiking of monitors. This is accomplished by the following steps:
-- Remove the white barrier film from the monitor.
-- Place a 2.5 cm diameter Whatman #5 filter paper on the spacer plate.
-- The elutriation cap is then snapped on.
-- Calculate the amount of material to be injected to correspond to the amount sampled by the monitor at a concentration of 1/2 TLV for 4 hours (approximately a 7.0 liter sample). Example: For Toluene with a TLV of 100 ppm or 375 mg/m3, the spike would have 375 [-=^n] (0.007) = 1.31 mg .
-- Inject the known quantity of the organic material onto the filter paper through the center port.
-- The monitor is allowed to sit for 16-24 hours to allow total transfer of the organic material from the filter paper to the sorbent before elution.
-- Remove filter paper from monitor after total transfer has occurred.
-- Proceed with elution and determination of amount recovered by G.C. analysis.
Please do not hesitate to contact toll free the 3M Occupational Health & Safety Products Tech. Service on 1-800-328-1300 for further assistance
OVMRP
3M 105491
RECOMMENDED PROCEDURE FOR DETERMINATION OF RECOVERY COEFFICIENTS
FOR 3M ORGANIC VAPOR MONITOR #3500
The 3M Company publishes the Recovery Coefficient for each organic material on the Qualification Data Sheet. This number represents the mean value 2 stan dard deviations from the tests conducted at 3M. A minimum of 5-fold replication (both monitors and standards) of the recovery test is performed at 3M. However, we encourage the user to verify the recovery coefficients, since techniques and presence of multiple contaminants can affect recovery coefficients.
The recovery coefficient is determined by vapor-state spiking of monitors. This is accomplished by the following steps:
-- Remove the white barrier film from the monitor.
-- Place a 2.5 cm diameter Whatman #5 filter paper on the spacer plate.
-- The elutriation cap is then snapped on.
-- Calculate the amount of material to be injected to correspond to the amount sampled by the monitor at a concentration of 1/2 TLV for 4 hours (approximately a 7.0 liter sample). Example: For Toluene with a TLV of 100 ppm or 375 mg/m3, the spike would have [~^] (0.007) = 1.31 mg.
-- Inject the known quantity of the organic material onto the filter paper through the center port.
-- The monitor is allowed to sit for 16-24 hours to allow total transfer of the organic material from the filter paper to the sorbent before elution.
-- Remove filter paper from monitor after total transfer has occurred.
-- Proceed with elution and determination of amount recovered by G.C. analysis.
Please do not hesitate to contact toll free the 3M Occupational Health & Safety Products Tech. Service on 1-800-328-1300 for further assistance
R-OVMRP
3M 105492
RECOMMENDED PROCEDURE FOR DETERMINATION OF RECOVERY COEFFICIENTS
FOR 3M ORGANIC VAPOR MONITOR #3500
The 3M Company publishes the Recovery Coefficient for each organic material on the Qualification Data Sheet. This number represents the mean value 2 stan dard deviations from the tests conducted at 3M. A minimum of 5-fold replication (both monitors and standards) of the recovery test is performed at 3M. However, we encourage the user to verify the recovery coefficients, since techniques and presence of multiple contaminants can affect recovery coefficients.
The recovery coefficient is determined by vapor-state spiking of monitors. This is accomplished by the following steps:
-- Remove the white barrier film from the monitor. -- Place a 2.5 cm diameter Whatman #5 filter paper on the spacer plate.
-- The elutriation cap is then snapped on. -- Calculate the amount of material to be injected to correspond to the
amount sampled by the monitor at a concentration of 1/2 TLV for 4 hours (approximately a 7.0 liter sample). Example: For Toluene with a TLV of 100 ppm or 375 mg/m3, the spike would have [^j-] (0.007) = 1.31 mg .
-- Inject the known quantity of the organic material onto the filter paper through the center port.
-- The monitor is allowed to sit for 16-24 hours to allow total transfer of the organic material from the filter paper to the sorbent before elution.
-- Remove filter paper from monitor after total transfer has occurred.
--- Proceed with elution and determination of amount recovered by G.C. analysis.
Please do not hesitate to contact toll free the 3M Occupational Health & Safety Products Tech. Service on 1-800-328-1300 for further assistance
R-OVMRP
3H 105493