Document mmk4O4b38gqZxKEXMdJDmw1ZZ
Federal Register !/ Vol:' Si,' No. 1197 Friday, June 20.' 196 /' Rules and Regulations
.22763
comparisons between microscopists to monitor the variability of counting by each microscopist and.between microscopists. In a company with more than one laboratory, the program shall include all laboratories, and shall also evaluate the laboratory-to-' laboratory variability.
2. Interlaboratory program. Each laboratory analyzing asbestos,.tremolite. anthophyliite. and actinolite samples-for compliance determination shall implement an interlaboratory quality assurance program that as a minimum includes participation of at least two other independent laboratories. Each laboratory shall participate in round robin testing at least once every 6 months with al least all the other laboratories in its interlaboratory quality assurance group. Each laboratory shall submit slides typical of its own workload for-use in this program. The round robin shall be designed and results analyzed using appropriate statistical methodology.
3. All individual performing asbestos, tremolite. anthophyliite, and actinolite analysis must have token Ihe NIOSH course for sampling and evaluating airborne asbestos, tremolite. anthophyliite, and actinolile dust or an equivalent course.
4. When the use of different microscopes contributes to differences between counters and laboratories, the effect of the different microscope shail.be evaluated and the microscope shall be replaced, as necessary.
5. Current results of these quality assurance programs shall be posted in each laboratory to keep the microscopists informed.
Appendix B to 1926.58--Detailed Procedure for Asbestos Tremolite, Anthophyliite, and Actinolite Sampling and Analysis--'NonMandatory
This appendix contains, a detailed procedure for sampling and analysis and includes those critical elements specified in Appendix A. Employers are not required to use this procedure, but they are required to use Appendix A. The purpose of Appendix B is to provide a detailed step-by-step sampling and analysis procedure that conforms to the elements specified in Appendix A. Since this procedure may also standardize the analysis and reduce variability, OSHA encourages .employers to use this appendix.
Asbestos. Tremolite, Anthophyliite, and Actinolite Sampling and Analysis Method
.'Technique:Microscopy, Phase Contrast. Analyte: Fibers (Manual count). Sample Preparation: Afeetone/tria.cetin
method! . - Calibration: Phase-shifi detection limit ' about 3 degrees.
Range: 100 to 1300 fibers/mm2 filter area. Estimated Limit of Detection: 7 fibere/nun1 filter area. Sampler: Filter (0.8-1.2 um mixed cellulose ester membrane. 25-mm diameter). .. Flow Rale: 0.5 l/min to 2.5 l/min.(25-tnm cassette); 1.0 l/min to 2.5 l/min-(37-mm .. cassette). Sample Volume: Adjust to obtain 100 to 1300.nbers/mm*. ' ShipmentVRoutine. Sample Stability: Indefinite. Blanks: 10% of samples (minimum 2).
Standard.Analytical Error: 0.25. Applicability: The working range is 0.02 f/ cc (1620-L air sample) to 1.25 f/cc (400-1. air
sample). The method gives an index of airborne asbestos, tremolite. anthophyliite! and actinolite fibers but may be used for other materials such as fibrous glass by inserting suitable parameters into the counting rules. The method does not differentiate between asbestos, tremolite, anthophyliite. and actinolite and other fibers.
Asbestos, tremolite, anthophyliite. and uctinolite fibers less than ca. 0.25 um diameter will not be detected by this method.
Interferences: Any other airborne fiber may interfere since all particles meeting the counting criteria are counted. Chain-like particles may appear fibrous. High levels of
nonfibrous dust particles may obscure fibers in the field of view and raise the detection
limit.
Reagents
1. Acetone. 2. Triacotin (glycerol triacetate), reagent grade.
Special Precautions
Acetone is an extremely flammable liquid and precautions must be taken not to ignite it.
Heating of acetone must be done in a ventilated laboratory fume hood using a nameless, spark-free heat source.
Equipment
1. Collection device: 25-mm cassetle with
50-mm extension cowl with cellulose ester
filter. 0.6 to 1.2 mm pore size and backup pad.
Note.---Analyze representative filters for
fiber background before use and discard the
. filter lot if more.thanS fibers/100 fields arc
found!
2. Personal sampling pump, greater than or
equal to 0.5 L/min, with flexible connecting
tubing.
3. Microscope, phase contrast, with green
or blue filter, 0 to 10X eyepiece, and 40 to 45X
phase objective (total magnification ca 400X):
numerical aperlure=0.65 lo 0.75.
- 4. Slides, glass, single-frosted, pre-cleaned.
25X75mm.
. 5. Cover slips, 25x25 mm. no. l`/ unless
otherwise specified by microscope
manufacturer.
6. Knife; #1 surgical steel, curved blade.
7. Tweezers.
&. Flask. Gulh-type, insulated neck, 250 to
500 mL (with single-holed rubber stopper and
elbow-jointed glass tubing, 16 to 22 cm long).
9. Hotplate, spark-free, stirring type;
heating mantle: or. infrared.lamp arid
magnetic stirrer.
10. Syringe, hypodermic, with 22-gauge
needle.
'
11. Graticule, Walton-Beckett type with 100
um diameter circular field el the specimen
, plane.(area^0.00785 mm3), (Type G-22).
Note.--The graticule is custom-made for each microscope.
12. HSE/NPL phase contrast test slide. Markli:
13. Telescope, ocular phase-ring centering. 14. Stage micrometer (0.01 mm divisions).
Sampling
1. Calibrate each personal,sampling pump with a representative sampler in line.
2. Fasten the sampler to the worker's lapel as close as possible to the worker's mouth. Remove the top cover from the end of the cowl extension (open face) and orient face down. Wrap the joint between the extender and the monitor's body with shrink tape to prevent air leaks.
3. Submit at least two blanks (or 10% of the total samples, whichever is greater) for each set of samples. Remove the caps from the field.blank cassettes arid store the cups and cassettes in a dean area (bag or box) during the sampling period. Replace the caps in the cassettes when sampling is completed. v
4. Sample at 0.5 L/min or greater. Do not exceed 1 mg total dust loading on Ihe filter. Adjust sampling flow rate. Q (L/min), and time to produce a fiber density. E (fibers/ mm8), of 100 to 1300 fibers/m8 |3.05X10* to 5X10 fibers per 25-mm filter with effective collection area (Ae=305 mm8)) for optimum counting precision (see step 21 below). Calculate the minimum sampling time, tfflinimum (min) at the action level (one-holf of the current standard), L (f/cc) of the fibrous aerosol being sampled:
tpUn --
{Q)(!.)10"
5. Remove the field monitor at the end of sampling, replace the plastic top cover and small end caps, end store the monitor.
6. Ship the samples in a rigid container w ith sufficient packing material to prevent jostling or damage. NOTE: Do not use polystyrene foam in the shipping container because of electrostatic forces which may cause fiber loss from the sampler filter.
Sample Preparation
Note.--The object is to produce samples with a smooth (non-gralny) background in a medium with a refractive iiidex equal to or less than 1.46. The method below collapses, the filter for easier focusing and produces permanent mounts which are useful for quality control'and interlaboratory comparison. Other mounting.techniqucs meeting the above criteria muy also be used, e.g.. the nonpermanent field mounting . technique used in P ft CAM 239.
. 7. Ensure that the glass slides and cover slips are free of dust and fibers. . 6. Place 40 to 60 ml of acetone into a.Gutii- type flask. Stopper the flask with a single hole rubber stopper through which a glass tube extends 5 to B cm into the n&sk.-The portion of the glass tube that exits the top of the stopper (8 to 10 cm) is bent downward in an elbow that makes an angle of 20 to 30, degrees with the horizontal.
9. Place the flask in a stirring hotplate or wrap in a heating mantle. Heat the acetone gradually to its boiling temperature (ca. 58*C)
Caution.--The acetone vapor must be., generated in a ventilated fume hood away . from all open flames and spark sources.. Alternate heating methods can be used. - providing no open flume or spArks arc- ' present.
GLEASON-001011