Document kaDN7j5a3Mod3ZO96J97V4JMV
FINAL REPORT
ANALYSIS OF DOW VINYL CHLORIDE
RECEIVED APR 1 9 1968
N. Wheeler
AUTHOR:
R. M. Arnold
GROUP LEADER: J. L. Hockersmith
DATE: April 16, 1968
SUMMARY An exchange of standard samples containing known amounts of expected impurities in Dow vinyl chloride has been made between the UCC labora
tories at South Charleston and Texas City and. the Dow laboratory at Freeport, Texas. Future evaluations by the three laboratories will be based on these standards. Resolution of the disagreement between the Texas City and Dow lab oratories has been accomplished through use of a standard. There still remains resolution of the disagreement as to reported levels of certain impurities be tween the Texas City and South Charleston laboratories.
The Texas City Resins Laboratory has evaluated the Dow analytical methods for vinyl chloride. Implementation of these methods into the monomer contract as referee methods has been found acceptable except for the acidity and phenol tests. UCC methods of analysis for these two specifications will be included in the contract as the acceptable referee method.
INTRODUCTION Preliminary investigation of Dow vinyl chloride in November, 1967, revealed that differences between chromatographic methods
at the Texas City and South Charleston laboratories could be resolved only through the use of response factors to report weight percent impurities in vinyl chloride. A standard sample containing expected impurities in Dow vinyl chloride was to be prepared by the Charleston Technical Center and sent to respective laboratories for calculation of response factors.
A meeting in Freeport, Texas, in December, 1967, between representatives of UCC and Dow indicated general agreement in the reported impurities found in typical production samples of Dow vinyl chloride. Only the reported level of vinyl acetylene remained to be resolved. Chemical methods for analyzing vinyl chloride were exchanged. 'The Texas City Resins Laboratory was to implement a study of these methods to determine their acceptance as referee methods for vinyl chloride specifications.
DISCUSSION In October, 1967, representatives of UCC and Dow agreed on imple menting a test program to define the vinyl chloride specifications
to be incorporated into a monomer purchase contract. The initial phase of this program between the UCC and Dow laboratories indicated the need for not only
PLANT LABORATORIES UNION CARBIDE CHEMICALS DIVISION
TEXAS CITY PLANT
UCC
Ur!Oc /D-j>tf
Dow Vinyl Chloride
2.
reaching agreement in analytical methods between the UCC and Dow laboratories, but also between UCC laboratories at Texas City and South Charleston. Stan dardization of chromatographic methods of analysis for vinyl chloride was de termined to be possible only through the use of common response factors used to convert volume percent impurities to weight percent impurities in vinyl chloride.
Three different standards of vinyl chloride containing known amounts of the expected impurities in the Dow monomer were sent to the UCC and Dow labora tories for implementation into their respective chromatographic test methods. One standard was prepared at the Technical Center, one by the Dow laboratory at Freeport, and one containing a known amount of vinyl acetylene was mutually prepared at Texas City and Freeport. Use of these standards permitted calcula tion of response factors for all expected vinyl chloride impurities.
So that an additional exchange of vinyl chloride samples could be elim inated, the "new" response factors were used on previous vinyl chloride data common to the three laboratories. Results of this phase of the monomer test program are found in Table I. Differences between the Texas City and Dow lab oratories that were previously reported appear to be resolved. The only serious difference appears to be the level of vinyl acetylene reported between the South Charleston laboratory and the laboratories at Texas City and Freeport.
This observed difference in vinyl acetylene content should not be cause for extreme alarm in view of the initial differences between the UCC labora tories. Great progress has been made in a few short months. Resolution of the present differences can be accomplished in a short time by a concerted effort on the part of both laboratories.
The Texas City Resins Laboratory now has the capability of utilizing a vapor injection technique. Routine analyses of vinyl chloride are now being made at attenuation (sensitivity) levels compatible with the South Charleston laboratory. Utilization of these factors in future vinyl chloride analyses will add Impetus to resolving the Internal differences between the UCC laboratories.
Evaluation at Texas City has been made on the Dow chemical methods which were to be designated referee methods for the particular vinyl chloride specifi cation unless serious differences were detected between the Dow and UCC methods. Results of this evaluation axe as follows:
1. Acetylene - The current method in use at Texas City is almost identical to the Dow method. Modification of UCC-22-211-3 for acety
lene was made several years ago to eliminate preparation of standards of known amounts of acetylene and the errors and hazards involved in weighing such a volatile and dangerous compound.
2. Aldehyde - The Dow method was found to be acceptable providing modifica tion be made to Indicate the use of a visible DU instrument
measurement rather than a turbidlmetrlc measurement. UCC does not have a turbidimeter at Texas City. Expense for its purchase is not necessary in view of the results obtained on the DU spectrophotometer.
UCC
067639
Dow Vinyl Chloride
3.
5. Acidity - Both UCC and Dow methods are similar. The only significant differences are in choice of indicators and UCC's use of a
blank. Dow informally indicated preference to the UCC method.
4. ^ron " The Dow method is much more sensitive at the 0.15 ppm level than the UCC method, which is simply a color comparison of the unknown
to a standard. The color formation is very faint at this low concentration and is not sensitive to small changes in iron concentrations. The Texas City Resins Laboratory will implement an Atomic Absorption method for determining iron in vinyl chloride in the near future. However, the Dow laboratories do not have this capability.
5. Non-volatiles - Both UCC and Dow methods are the same.
6. Phenol - Both methods are similar. The Dow method scans the phenol sample from 220-320 m/A and determines the absorbance from a
maximum and minimum reading obtained from a potentiometric recorder. The UCC method measures absorbance directly at 271 ffl/a . If the Dow method is to be used as a referee method, it should be modified to a measurement compatible with the UCC method.
7. HgO - The Dow method is less susceptible to interference by moisture in the air. The closed system is more desirable than the open en
vironment inherent in the Carbide method. However, the Dow method specified in the contract should indicate exclusion of the instrumental technique invol ving use of an automatic titrator. The Texas City Resins Laboratory does not have this capability of analysis.
8. Peroxides - The Dow method is the more sensitive method of the two and should be the specified referee method. The UCC method is
quite erratic at peroxide levels <0.1 ppm.
9. Sulfur - UCC does not have a method for determining sulfur in vinyl chloride. The Dow method should be the indicated referee method.
Although the Dow method of analysis for determining oxygen was received, it was not evaluated hy the Texas City Resins Laboratory, Evaluation would have necessitated preparation of oxygen standards and additional chromatographic columns. Several tank cars were analyzed for oxygen(hefore and after unloading) by the Texas City Chemicals Laboratory using a UCC chromatographic method. Re sults of this evaluation are found in Table II. Residual oxygen content would not appear to be a problem when the monomer purchase contract takes effect. Residual oxygen levels, in fact, should he significantly lover since vinyl chloride vapors instead of nitrogen will be used to pump liquid vinyl chloride from the tank cars.
RECOMMENDATIONS
1. The Dow chemical methods of analysis should he accepted as the referee methods except for phenol and acidity. The Dow method for aldehyde should be modified as indicated in the previous section.
UCC
067640
Dow Vinyl Chloride
A
2. Additional samples of vinyl chloride should be exchanged between the South Charleston and Texas City laboratories until mutual agreement as to concentration of reported impurities is reached. The South Charleston laboratory should undertake the responsibility of developing the capability to report the quality of vinyl chloride as defined by the Texas City and Dow laboratories.
RMA/ab Attachments (2)
R. ~M. Arnold
IS CITYrn ,'v
lilwV-
Mr. J. H. Barrett Mr C. E. Bowman Mr. o. T. Carlisle Mr. M. E. Eisenhour Mr. J. H. Field Mr. D. D. Hawthorne Mr. J, L. Hockersmith Mr. F. L. Johnson Mr. J. B. Leverton Mr. D. E. Richardson Mr. J. J. Roeser
NEW YORK OFFICE Mr. G. P. Bigelow Mr. A. C. MacLeod
DISTRIBUTION
SOUTH CHARLESTON Mr. J. W. ChatfieId
Mr. J. W. Kearns Mr. R. N. Wheeler
TECHNICAL CENTER
Mr. W. W. Fink Dr. J. F. Haskin Dr. D. L. McKinley
Mr. M. E. Sutherland Mr. W. E. Whitehurst
UCC 067641
Impurities Isobutane n-Butane Butene-1 Allene Trans-butene-2 Methyl Acetylene 1,3-Butadiene Methyl Chloride Vinyl Acetylene 3-Chloropropene 1,1-Dichloroethane
TABLE II ANALYSIS OF DOW VINYL CHLORIDE USING COMMON RESPONSE FACTORS
TEXAS CITY__________
IA-IB
IIA-IIB
3351
33 6
6.5 1.5 9 ---------
Trace
Trace
Trace
14.6 3 -
25.6 8 2
17 10 -
-
Trace
-
SOUTH CHARLESTON
IB
3.84 7.56 41 <1 1.4 19.8 9-9 -
I IB
4.16 2.16 41 41 1.05 27-5 17.8 -
3351 7.04 9 .55
-
.35 14.3 4.6 -
---
IA-IB
3.31 4.53 0.13 .98 15.45 3.31 -
DOW
IIA-IIB
2.96 1.49 0.09 09 .07 .85 23.3 8.27 -
3351 7.8 8.9 .7 17.1 10.6 -
Total Impurities, ppm 27.1
40.1
42
43.5 53.7 35.8
27.7
37.1
45.1
ZfcQlQO
oon
TABLE II OXYGEN ANALYSIS IN DOW SHIPMENTS OF VOX
Car No.
dowx-3351 GATX-30771 DOWX-3352 DOWX-3351 GATX-30762 GATX-30771 GATX-30762 DOWX-3352 GATX-30771
Date
12-21-67 12-31-67 12-31-67 1-2-68 1-2-68 1-11-68 1-12-68 1-25-68 1-29-68
PPM Oxygen (by volume)
Full
Empty
68 ll4l
94 580
105 558
210 178
220 520
10 600
79 -
- 875 - 300
ucc
087643