Document ka4OBrwk2kQvjao2Dyyg4OOLb
BACK TO MAIN
Centre Analytical Laboratories, Inc.
3048 Research Drive State C a ? g e ,PA 16801
www.centrelab.com
(814)231-8032
Fax: (814)231-1253or (814)231-1580
Analytical Report
Fluorochemical Characterizationof Surface Water Samples Decatur, Alabama (W1979)
Centre Analytical Laboratory Report No. 023-014 8 (Revision 1)
Revision Date 3/21/01
Testing Laboratory Centre Analytical Laboratory, Inc.
3048 Research Drive State College, PA 1680'1
3M Environmental Laboratory Contact Kent R. Lindstrom
Bldg. 2-3E-09
P.O. Box 33331 St. Paul, MN 55133-3331 Phone: (651) 778-5352!
Requester Kris J. Hansen, Ph.D. 3M Environmental Technology & Safety Services
Bldg. 2-3E-09 P.O. Box 33331 St. Paul, MN 55133-3331
PAGE 1 OF 5
BACK TO MAIN
1 Introduction
Results are reported for the analysis of a series of surface water samples received by Centre Analytical Laboratories, Inc. (Centre) from the 3M EnvironmentalLaboratory. The samples were collectedfrom Decatur, Alabama. The Centrestudy number assigned to the project is 023-014.
Specific fluorochemical characterization by liquid chromatography/ tandem mass spectrometry (LC/MS/MS) was requestedfor all samples. A total of 9 samples were received for analysis.
The samples were prepared and analyzed by LC/MS/MS for the following list of fluorochemicals:
I Table 1:Target Analysis ComDoundName Perfluorooctane Sulfonate Perfluorooctane Sulfonvlamide Perfluorooctanoate
;1;E 1
The analytical method used was validated by Centre. The validation protocol and results are on file with Centre. Data presentedhere is the highest quality data available at this time.
2 Sample Receipt
The samples were submitted in individual plastic containers and were not preserved. Nine individual sample containers were received. Sample hAC-202H was not analyzed as per client request. (E-mail communication from Kris Hansen on 5/16/00). Samples were received on 05/03/00. The sample collection dates were not supplied. Chain-of-custody information is presented in Attachment C.
3 Holding Times
The analytical method used was validated against a maximum holding time of 14 days. The stability of the analytes of interest for longer periods has not been determined. However, it should be noted that field fortifications in water and other matrices have shown acceptable recoveries at 100 and 1000 ngR for periods longer than 14 days.
PAGE 2 OF 5
BACK TO MAIN
4 Methods - Analytical and Preparatory
4.1 LC/MS/MS
4.1.1 Sample Preparationfor LC/MS/MS Analysis
Samples were initially treated with 200 ULof 250 mg/L sodium thiosulfate solution to remove residual chlorine. Solid phase extraction (SPE) was used to prepare the samples for LC/MS/MS analysis. A forty-milliliterportion of sample was transferredto a C18SPE cartridge. The cartridge was first eluted with 5 mL of 40% methariol in water solution. The eluate was discarded and the SPE column was then eluted with 100% methanol. A 5 ml portion of methanol was collected for analysis by LC/MS/MS. This treatment resulted in an eight-fold concentrationof the samples prior to analysis.
4.1.2 Sample Analysis by LC/MS/MS
In HPLC, an aliquot of extract is injectedand passed through a liquid-phasechromatographic column. Based on the affinity of the analyte for the stationary phase in the column relative to the liquid mobile phase, the analyte is retained for a characteristic amount of time. Following HPLC separation, ES/MS provides a rapid and accurate means for analyzing a wide range of organic compounds, including fluorochemicals. Electrospray is generally operated at relatively mild temperatures; molecules are ionized, fragmented, and detected. Ions characteristic of known fluorochemicals are observed and quantitated against standards.
A Hewlett-PackardHPl1 00 HPLC system coupled to a Micromass Ultima MS/MS was usedto analyze the sample extracts. Analysis was performed using selected reaction monitoring (SRM). Samples were extracted on 5/16/00and analyzed by MS/MS between 5/19/00and 5/20/00. The HPLC and MS/MS methods used for analysis and instrument parameters can be found in attachment D.
5 Analysis
5.1 Calibration A -/-point calibration curve was analyzed at the beginning and end of the analytical sequence for the compounds of interest. The calibration points were prepared at 0, 25, 50, 100, 250, 500, and 1000 ngR (ppt) The response of the quantita1:ion ion versus the concentration was
plotted for each point. Using linear regression with l/x weighting, the slope, y-intercept and
correlation coefficient (r) and coefficient of determination (?) were determined. A calibration curve is acceptable if r 20.985 (?2 0.970).
Calibration standards are prepared using the same SPE procedure used for samples.
Calibration check standards were analyzed periodically (every three to five sample injections) throughout the analysis sequence. Compliance is obtained if the standard analyte concentrations are within +/-20% of the actual value.
For the results reported here, calibration criteria were met.
PAGE3OF5
BACK TO MAIN
5.2 Blanks
Extraction blanks were prepared and analyzed with every extraction batch of samples. The extraction blanks should not have any target analytes present at or above the concentration of the low-levelcalibrationstandard. For these samples, the extraction blanks were compliant.
Instrument blanks in the form of clean methanol solvent were also analyzed after every highlevel calibration standard, and after known high-level sa,mples. Again, the blanks should not have any target analytes present at or above the low-level calibration standard. For the samples presentedhere the instrument blanks are compliiant.
5.3 Surrogates
Surrogate spikes are not a component of the LC/MS/MS analytical method.
5.4 Matrix Spikes
Matrix spikes were prepared for every field sample (excluding blanks) at a concentration of 100 ng/L using all compounds of interest. Matrix spike recoveries are given in Attachment 6. All laboratory matrix spikes showed recoveries of all compounds between70-130%
Field spikes were prepared on sample MC-201H at a concentration of 100 ng/L using all compounds of interest. The field spike is identified as FdC-204H. Field spike recoveries are also given in Attachment B. PFOS showed high recovery in the field spike. All other compounds showed recoveries between 70-130%.
5.5 Duplicates
All field samples (excluding blanks) were analyzed in duplicate. Results are given along with the sample results in Attachment A.
5.6 Laboratory Control Samples
Milliq water was spiked with all compound of interest at ;!5 and 250 ngR. All recoveries for all compounds were between 70-130% in each LCS. Recoveries are given with the raw data in Attachment D.
5.7 Sample Related Comments
Field blank samples consisted of empty containers. Forty milliliters of type I water filtered
through a hypercarb cartridge was added to the empty container and analyzed in the same manner as the other samples.
6 Data Summary
Please see Attachment A for a detailed listing of the analytical results.
7 DatdSample Retention
Samples are disposed of one month after the report is i:;sued unless otherwise specified. All
electronic data is archived on retrievable media and hard copy reports are stored in data
folders maintained by Centre.
PAGE4OF5
BACK TO MAIN
8 Attachments
8.1 Attachment A: Results 8.2 Attachment 6:Matrix Spike Recoveries(Fieldand Laboratory Spikes) 8.3 Attachment C: Chain of Custody 8.4 Attachment D: LC/MS/MS Raw Analytical Data
9 Signatures
. 6
Kevin J Lloyd, Vice President
Other Lab Members Contributingto Data Enaksha Wickremesinhe Karen Smith David Bell
27 mMw I Date
PAGE 5 OF5
\ 4 laboratories,,Inc. CEntre Analytical 3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253
BACK TO MAIN
,Analytical Results W1979 Decatur, Alabama
:3MSample Identification
MC-201H MC-203H MC-206H NA MC-207H NA MC-208H MC-284H NA
Sample Description
Site 1 P/N Surface Water Site 1 PIN Surface Water Duplicate Site 2 P/N Surface Water Site 2 P/N Surface Water Duplicate Site 3 P/N Surface Water
Site 3 PIN Surface Water Duplicate
Field Blank P/N Empty Quiet P/N Surface Water Quiet PIN Surface Water Duplicate
PFOS (ng/L.)
NQ NQ NQ NQ ND ND ND 108 114
PFOSA (ng/L)
ND ND ND ND ND ND ND NQ NQ
Limit of Detection (LOD) for the procedure is appoximately 2.5 ng/L for PFOS and PFOSA and 7.5 ng/L for POAA Limit of Quantitation (LOQ) for the procedure is 25 ng/L for all compounds
ND - Compound not detected NQ - Compound detected at a level between the LOD and LOQ. Result is not quantifiable.
ND e LOD c NQ e LOQ
POAA (ng/L)
NQ ND ND NQ ND ND ND 63.1 57.3
-
Please refer to the reverse side for our standard terms and conditions.
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
!Sample ID:
I MC-205H
!Spiked Amount (ng/L): I
100
1
PFOS PFOSA
Sample Concentration
(ng/L)
23.1 0 9.2
Matrix Spike Result
(nQ/L)
128 107 106
Lower Recovery Limit: I
70
1
Upper Recovery Limit: I
130
Matrix Spike Result
(% Recovery)
104.9 107.0 96.8
Criteria (Pass / Fail)
PASS PASS PASS
Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are below the limit of quantitation. Results are given in this table for recovery calculations only.
Also note that sample MC205-H is a laboratory spike of sample MC-201H
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID:
I MC-206H
Spiked Amount (ng/L): I
100
1
Lower Recovery Limit: IJpper Recovery Limit:
Sample Concentration
14.5 0 0
I
70
I
130
Matrix Spike Result (ng/L) 117 103 113
I
Matrix Spike Result
I:% Recovery)
102.5 103.0 1 13.0
Criteria (Pass / Fail)
PASS PASS PASS
Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are I)elow the limit of quantitation. Results are given in this table for recovery calculations only.
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID:
I
Spiked Amount (ng/L): I
MC-207H 100
Lower Recovery Limit: lJpper Recovery Limit:
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result
(ng/L)
100 101 105
I
70
1
130
I
Matrix Spike Result
([% Recovery)
100.0
101.o
105.0
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Rlecovery
!Sample ID:
I
!Spiked Amount (ng/L): I
MC-284H 100
Lower Recovery Limit: IJpper Recovery Limit:
. Sample
Concentration (ng/L)
108 18.9 63.1
I
70
1
130
Matrix Spike Result (ng/L)
200 110 161
Matrix Spike Result
(% Recovery)
92.0 91.1 97.9
Criteria (Pass / Fail)
PASS
PASS PASS
-
Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are below the limit of quantitation. Results are given in this table for recovery calculations only.
BACK TO MAIN
Attachment B: LC/MS/MS Field Spike Recovery
!Sample ID:
1 MC-204H I
!SpikedAmount (ng/L): I
100
I
[ E A POAA
Sample Concentration
(ng/L)
23.1 0 9.2
Matrix Spike Result (ng/L)
169 85.6 111
Matrix Spike
Result
-(% Recovery) 145.9
85.6 101.8
Criteria (Pass / Fail)
FAIL PASS PASS
Upper Recovery Limit: I
130
1
Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are below the limit of quantitation. Results are given in this table for recovery calculations only.
Also note that sample MC-204H is a field spike of sample MC-201ti
3M Environmental Laboratory
FM 38776 - PWO
Shlpplna Addrass: 3 M B 4 2-3509
935BushAvenue it. Paul. MN 55106
T:!:;hs.?o: Sample Recolving: (651)n 8 4 w Alternate: (651)7786751 F A X (651)7786176
BACK TO MAIN
Chain of Custody /Request for Laboratory Analytical
- -IProject ID-lP~roject Name &EA)0\4/11b, b1(%
Template #
Project Lead -4
e Dept. # (main) 42Q\
ur?Afifi&h~+ ",-
Final Report Due Date
Internal Due Date
ClasslJoblProject # 001
1 4266
3M Env. Lab Project #
For Internal Use Only
- -.
Analysis Requested:
Original - Acmmpanyhg Samples
Last Page - Originator
See Reverse Side for lnstruclions