W . G . KRUMMEIOH PLANT
ENVIRONMENTAL CONTROL GUIDELINE SF-'l
ISOKINETIC CAMPLING USING
A FLIP-FLOP SAMFLER
March 1972
SCOPE
.
This sampler is a modification of the ''Isokinetic Waste Sampler
for the Sauget Waste Treatment Plant'' guideline SP-2.
.
To obtain a representative sample of waste stream containing
solids or liquids in suspension with varying degrees of dispersion.
DIFFICULTIES ENCOUNTERED
The suspended materials tend to separate in the sample lines
at low flow rate. .
b. Small diameter sample lines plug easily and cannot be used to
maintain isokinetic velocities in the sample lines.
c. Continuous pumping at isoKinetic rates results in a very
large sample.
d. Peristaltic pumps, while excellent in many respects, have a
short life when operated continuously.
'
e. Suspended liquids separate on to the surface of collection
vessels and cannot be completely re-suspended when it is
necessary to withdraw aliquots of the collected material.
f. When the pumping is intermittent, the sample may not be
representative until the line is flushed clear.
Problems a, b, c, and d were largely overcome by Intermittent pumping
into a 99 gallon drum equipped with a mechanical agitator (Ref. SP-2).
However, problems e.
4
^.serious drawbacks when analysing for trace
amounts "a viscous oily material vrith very low solubility.
f
DSW 295959
STLCOPCB4065903
2
The flip-flop sampler was developed to overcome these difficulties.
It should be noted that it is adaptable to continuous pumping if
a suitable pump can be found and also for proportional to flow
sampling by electronically varying the pulse rate. '
EQUIPMENT
1 - Peristaltic pump "Little Giant" model LG-101 with 110 v AC, /X rf .
1725 PPM and 1/3 HP motor. Use 1/V ID ygon tubing.
2 - Cam Timer Model J3953 with A-l8 gear rack. Meylan Stop
Watch Corp.
,
3-r Pulse timer SV-800-013 modified at factory. Valcor Eng. Corp.
^ - Solenoid valve 5^ P1903b modified at Monsanto. Valcor Eng.
Corp. 5-2 1/2 gal. square jar. PRINCIPLE OF OPERATION
C Uutkt, /V,
'
The peristaltic pump cannot be run continuously due to a tubing
failure and to insure life to the tubing we use a cam timer. Timer
setting can be adjusted according to our needs. Vie normally run
the pump about 30 seconds every 15 minutes.
The cam timer energizes the pump and the pulse timer, the latter
can be set to 1, 2 or 3 pulses during the -45 seconds run. The
frequency of pulses give us the size of the sample. If the sampler is
programmed as described and the pulse timer set at 2 pulses it will
collect about 1 gal. sample in 2^ hours. t
ISOKINETIC PICK-UP TIP
Using a new "Cygon tubing the rate is about 1 GPM.
With this
information and estimating the sewer velocity we can size the tip
of the pick-up tubing to have an isokinetic samplei^,
DSW 295960
STLCOPCB4065904
-3-
Sewer stream velocity = 3 ft./sec.
pumping rate = 1 gal./60 sec.
Q=AxV
A=Q V
A 1 x .1337 3x60
= .742 x 10"3 ft'
D =/.742 x 10"3 x j4 T7T4
= 000945 = .0307 ft
Pick-up tipJ 0.307 ft.[= 1/4 Jfij sch. 40 ss pipe,
SAMPLER BOX
l > o)c
The sampler is housed in a plywood box (see drawing), the rump and
the flip-flop sampler are located in the upper shelf and the sample
jar is located in the lower shelf. The box can be winterized easily.
The overflow line is a 3/4" ID rubber hose and is directed back to
the sewer.
Julio Munoz, Environmental Control Groun