Monsanto INDUSTRIAL CHEMICALS CO. 5y . mom .khe . location) e S. Tucker - South Second Street
OATE April 19, 1972
subject
REFERENCE
pCB METHODOLOGY -..........
ACCURACY AND PRECISION
: "*
r
TO
: H. J . Horner - WGK (1740) J . T. Bell - Anniston (1060)
CC; R . E. Keller - TIB
W. B . Papageorge - B2NK
W. R . Richard - T3A
__
R. H. Munch - 1760^m=iC-I
R . A . Stohr - A3SB W. C . Engman - 1740 E. E. Stewart - 1740 P. E. Heisler - 1740 .......... A . E . Leisy - 1740 J. D. Hinchen - R3B
Accompanying this memo are two Statistics special studies which are in tended to give to all concerned a feeling for the precision and accuracy of the "estimated PCB levels" currently being reported by both internal and external analytical laboratories.
The first special study, #72-5, illustrates the accuracy and precision attainable by two experienced Monsanto laboratories analyzing the simplest matrix (pure water) encountered via equivalent methods under conditions eliminating any potential sampling problems.
Analysis of the laboratory data indicates that on the average, two exper ienced laboratories analyzing any equivalent individual sample contain ing a relatively high PCB level (500 ppb) would report 87. 5% of the level really pr.esent and that 95% of the time the value reported could be ex pected to fall in the range of 32. 3-142. 7% of the actual level present.
It should be anticipated that if these standard samples had been spiked at say the 1 ppb level that the precision and accuracy would be poorer.
The data also illustrates that it is 10 times more difficult to identify the specific Aroclor present than it is to estimate the total PCB level in the sample, especially if the laboratory personnel involved are not highly experienced in interpreting electron capture chromatograms. Because of this fact we have, for some time now, recommended that the analytical results be reported as the "estimated total PCB level", referencing the appropriate Aroclor or Arodors used as the standard or standards.
.
It also should be pointed out that all laboratories doing PCB analysis have a definite reagent or laboratory background of apparent PCB's. As such, if the level found in any sample is not at least twice the blank level calcu lated on the basis of the sample size in question, it is unrealistic to report a positive PCB level. Instead it should be reported as none detected, less than twice the apparent background or blank concentration calculated on the basis of the sample size used.
r * - .r V II ?.
DSW 346557 STLCOPCB4084579
2- -
The second special study, #72-6, illustrates that when one is dealing with the analysis of a more complex matrix by different laboratories with varying levels of experience, each using similar but different analytical methods, that the precision and accuracy become poorer, pointing out our urgent need for standardization of the methods currently being used to analyze environmental samples for PC'B's.
js attachments
E. S. Tucker