Document gmpD66ZYDekk1ywxoj0rmD1V

STANDARD METHOD FOR ANALYSIS OP ENVIRONMENTAL MATERIALS FOR POLYCHLORINATED BIPHENYLS 1. Scope 1.1 These methods cover the determination of the amount and type of polychlorinated biphenyls (PCBs) in en vironmental materials, specifically air, water and soils and sediments. Absolute confirmation of PCB structures is not obtained with these methods. Structure proof can be obtained using additional techniques such as mass spectrometry. 2. Significance 2.1 Since it is possible for minute quantities of PCBs to find their way into the environment, it is de sirable to have a method to identify and measure the quantity of such trace amounts, 3. Sampling 3.1 Water Sampling 3.1.1 Environmental sampling of natural bodies of water and effluents are covered in many methods under the Jurisdiction of Committee D-19. Method D-510, ASTM Part 23, Sampling for Biological Tests, is recommended with the following stipulations. Under 26(a) a final rinse with nanograde hexane should be added. The sample collected should be at least 250 ml. 3.2 Sediment and Soil Sampling 3.2.1 A sediment or soil sample of at least 1000 grams should be taken according to ASTM Method D-1587, ASTM Part 11 or any other ac ceptable procedure, taking due precautions to minimize contamination. 3.3 Air Sampling 3.3.1 Environmental sampling of ambient air or source sampling is under the Jurisdiction of ASTM Committee D-22. There are various methods of sampling from stacks or ambient air available. See Method D-I605, ASTM Part 23. MOWS 031443 -2- 3.3.2 For the measurements of airborne PCBs the air sample is drawn through one or more fritted bubbler absorbers filled with nano grade toluene. The PCBs are absorbed in the toluene. it. Principle 4.1 The PCBs present in environmental samples are Isolated from the bulk of the sample matrix by solvent extrac tion. Interfering matrix components and endogenous re sidues are then removed from the sample by a suitable icombination of clean-up procedures and the amount and type of PCBs present determined by the electron capture gaB chromatography (EC/OC). 5. Reagents 3.1 Hexane - Nanograde 5.2 Acetonitrile - Nanograde 5.3 Sodium Sulfate - Anhydrous, granular: AR grade. Heat 400C for one hour prior to use. 5.4 Alumina Adsorption - (Por chromatographic analysis) 80/200 mesh. Heat at 400C for a minimum period of four hours and deactivate with 5$G (w/w) distilled water. Alumina column preparation: Fill a chromatographic column with hexane up to the point where the reservoir Joins the column and push a glass wool plug to the bottom with a glass rod. In a 50 ml beaker measure 35 ml of deactivated alumina (30G), and pour this slowly into the column. Tap or vibrate the column to settle the alumina and top the alumina with 2-3 cm of anhydrous sodium sul fate. Wash the column with 50-100 ml of hexane prior to the addition of the sample. 5.5 Distilled Water - Extracted with hexane to remove hexane soluble electron capturing impurities. 5.6 Sulfuric Acid - Analytical Reagent Grade, SO 1.84 5.7 Potassium Hydroxide - Analytical Reagent Orade 5.8 Ethanol - Formula 2B 5.9 2.5J6 (w/v) Alcoholic Potassium Hydroxide - Dissolve 12.5 gram3 of AR grade KOH in 500 ml of ethanol. RGI*S 071A44 -3- 5.10 Toluene - Nanograde 5.11 PCB Standards - Production Grade - Aroclor 1221, Aroclor 12*12, Aroclor 1248, Aroclor 1254 and Aroclor 1260 6. Apparatus 6.1 Separatory funnels equipped with ground glass stoppers and Teflon stopcocks: 125, 250, 500, 1000 and 2000 ml capacities. 6.2 Kunderna-Danlsh Evaporative Concentrators, 500 ml capa city equipped with 3-ball Snyder columns and graduated 5 ml capacity vials. 6.3 Chromatographic columns, glass, 10" x 20 mm (OD) with a 5" x 50 mm (OD) reservoir at the top, equipped with Teflon stopcocks. 6.4 Sintered glass filter funnels, 600 ml capacity, 90M. 6.5 Plat bottomed boiling flasks, 125 ml capacity 6.6 Liebig Condensers, 200 mm In length 6.7 Hot plates 6.8 Water bath 6.9 Reciprocating variable speed shaker 6.10 10 pi syringes 6.11 32 oz. all glasB mortar and pestle 6.12 8 x 12 x 2" (2-1/2 qt.) Pyrex baking dishes 6.13 U.S. Standard Sieve, No. 30 6.14 Usual laboratory glassware 6.15 Gas scrubbing bottles, high form, ground glass Joint, fritted coarse discs, 250 ml capacity 6.16 Dry Test Meter (0.1 ft^/revolutlon) 6.17 laboratory Vacuum Pump 6.18 Rotatory Vacuum Evaporator HONS 071445 -4- 7. Sample Preparation and Extraction 7.1 water 7.1.1 Where possible, the entire water sample. In cluding the container In which It was collec ted, Is extracted with hexane. With larger samples, where this Is not physically possible, the containers are simply agitated and a 250 ml portion used for analysis. 7.1.2 Extraction of water samples - after agitating, transfer the entire aqueous sample or a 250 ml aliquot Into a graduated glass cylinder. Record the volume of the sample and quantitatively trans fer It to a separatory funnel with distilled water. 7.1.3 Rinse the sample container and graduated cylinder with two 50 ml portions of hexane and add each to the separatory funnel. 7.1.4 Stopper the separatory funnel and hand shake vigorously for at least 1 minute. Allow the layers to separate and transfer the lower aque ous phase to a second separatory funnel. 7.1.5 Extract the water sample a second time with a 50 ml portion of hexane. After the layers have separated, add the first hexane extract to the second separatory funnel and transfer the aque ous layer to the original separatory funnel. 7.1.6 Repeat the extraction with a third 50 ml portion of hexane. Discard the aqueous layer and com bine the hexane extracts. 7.1.7 Filter the combined extracts through a 4" funnel plugged with glass wool which Is covered with sodium sulfate. Collect the filtrate In a Kunderna-Danlsh evaporative concentrator, add a small boiling chip, put the Snyder column In place, and reduce the hexane volume to less than 5 ml by heating the apparatus In an 80-90*C water bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED TO A HOOP.) HONS 071446 -5- 7.1.8 Inject a fraction of a ul of the concentrate Into the gas chromatograph to check for In terferences and determine the approximate level of PCBs present. If no interferences are present dilute or concentrate the sample to a known volume, as determined by the elec tron capture chromatogram, and proceed with the gas chromatographic analysis. 7.1.9 If Interferences are present proceed with the chemical treatment and column chromatographic clean up procedures. 7.2 Sediment and Soil 7.2.1 Any excess water Is decanted and the entire sediment or soil sample transferred to a glass baking dish to air dry at room temperature. The dried material is transferred from the dish Into a mortar and pestle and ground. The ground sediment Is Bleved, remixed, and a 250g portion taken for analysis, 7.2.2 Weigh 25Qg (to the nearest O.Olg) Into a 16 oz. narrow neck screw cap (aluminum foil liner) glass bottle. 7.2.3 Moisten the soil with water (10 ml) and add 150 ml of acetonitrile. Cap the bottle tightly, and mechanically shake for a minimum period of 1 hr. 7.2.4 Quantitatively transfer the acetonitrile extract Into a 600 ml sintered glass filter funnel con taining a 1/4" layer of anhydrous sodium sulfate. Collect the filtrate In a 600 ml beaker (vacuum filtration may be necessary). 7.2.5 After the acetonitrile has completely drained Into the beaker, wash the bottle twice with 50 ml portions of acetonitrile, adding each wash to the funnel after the previous has completely percolated through the sediment. 7.2.6 Quantitatively transfer the extract to a Kunderna- Danlsh evaporative concentrator, add a small boiling chip, put the Snyder column In place, and reduce the solvent volume to less than 5 ml by heating the apparatus In a 80-90*C water bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED INTO A H5SB7) " HONS 071*47 -6- 7.2.7 Carefully evaporate the acetonitrile extracts to dryness with the aid of a gentle stream of dry nitrogen. Re-dissolve the residue In maximum of 5 ml of nanograde hexane. 7.2.8 Inject a fraction of a ul of the concentrate Into the gas chromatograph to check for Inter ferences and determine the approximate level of PCBs present. If no interferences are pre sent dilute or concentrate the sample to a known volume, as determined by the electron capture chromatogram, and proceed with the gas chromatographic analysis. 7.2.9 If Interferences are present proceed with the chemical treatment and column chromatographic clean up procedures. 7.3 Airborne PCBs 7.3.1 After scrubbing the desired amount of air, re cord the metered volume, pressure and tempera ture. 7.3.2 Quantitatively transfer the scrubbing solvent to a round bottom flask and reduce the volume to approximately 2 ml by rotary vacuum evapora tion. 7.3.3 Quantitatively transfer the concentrate to a 30 ml beaker with the aid of several small por tions of toluene. 7.3.4 Inject a fraction of a ul of the concentrate Into the gas chromatograph to check for Inter ferences and determine the approximate level of PCBs present. If no Interferences are present dilute or concentrate the sample to a known volume, as determined by the electron capture chromatogram, and proceed with the gas chroma tographic analysis. 7.3.5 If Interferences are present proceed with the chemical treatment and column chromatographic clean up procedures. MONS 071448 -7- 8. Supporting Procedures - Sample Clean Up 8.1 Saponification with Alcoholic Potassium Hydroxide and Extraction with Sulfuric Acid' 8.1.1 Quantitatively transfer the concentrated ex tracts to a 125 ml extraction flask with the aid of several small portions of solvent. 8.1.2 Evaporate the extract Just to dryness with a gentle stream of dry filtered air and add 25 ml of 2.5^ alcoholic potassium hydroxide. 8.1.3 Add a boiling chip, put a water condenser In place, and allow the solution to reflux on a hot plate for 45 minutes. 8.1.4 After cooling, transfer the solution to a 250 ml separatory funnel with the aid of 25 ml of distilled water. 8.1.5 Rinse the extraction flash with 25 ml of hexane and add It to the separatory funnel. 8.1.6 Stopper the separatory funnel and shake vigo rously for at least 1 minute. Allow the layers to separate and transfer the lower aqueous phase to a second separatory funnel, 8.1.7 Extract the saponification solution with a second 25 ml portion of hexane. After the layers have separated, add the first hexane extract to the second separatory funnel and transfer the aqueous alcohol layer to the ori ginal separatory funnel. 8.1.8 Repeat the extraction with a third 25 ml portion of hexane. Discard the saponification solution and combine the hexane extracts. 8.1.9 CAREFULLY add 25 ml of concentrated sulfuric acid, in small portions, to the hexane extracts. 8.1.10 Stopper the separatory funnel and shake vigo rously for at least one minute venting the funnel as necessary. Allow the layers to separate and discard the lower aqueous-acid layer. Repeat this step until the acid layer is colorless. HONS 071ViS -8- 8.1.11 Wash the hexane with 25 ml portion of water. Discard the water wash. 8.1.12 Filter the hexane extract through a 4" funnel plugged with glass wool which is covered with a layer of sodium sulfate. Collect the fil trate In a Kundema-Danish evaporative concen trator, add a small boiling chip, put the Snyder column In place, and reduce the hexane volume to less than 5 ml by heating the appara tus in a 80-90*C water bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED TO A HOODTT 8.2 Llquld/Solld Chromatographic Clean-Up With Alumina 8.2.1 After preparation of the alumina column, as described In the "Reagents" section, quantita tively transfer the hexane extracts (5 ml or less) on to the column with the aid of several 5 ml portions of hexane (lowering the solvent level to the adsorbant level after each addition of solvent. Do not allow the top of the eolumn to go dry). 8.2.2 Carefully add 125 ml of hexane to the column reservoir and collect the total eluent In a Kunderna-Danlsh evaporative concentrator. 8.2.3 Add a small boiling chip to the Kunderna-Danlsh evaporative concentrator, put the Snyder column in place, and reduce the hexane volume to less 5 ml by heating the apparatus In a 80*-90*C water bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED TO A HOT5PT) 9, Electron Capture Oas Chromatographic Procedure 9.1 Instrument: Hewlett-Packard 5750 Research Chromatograph or equivalent 9.2 Detector: High Temperature Nl^ Electron Capture Cell 9.3 Column: 6 mm x 61 glass column, mesh Chromosorb-W, HP, AW-DMCS XE-60 on 80/100 9.4 Column Temperature: Aroclor 1221, 170*; Aroclor 1242, , .190; Aroclor 1248, 190*; Aroclor 1254, 205*; Aroclor 1260 220 HONS 071450 -9- 9.5 Detector Temperature: 250-300C 9.6 Injection Port Temperature: 220*0 9.7 Pulse: 50 usee 9.8 Plow Rates: Helium Carrier 60 ml/mln 90% Argon- 10% Methane Purge 120 ml/mln 9.9 Using EC/OC as the determinative step. Inject In dupli cate 2 to 8 ul of each solution Into the chromatograph. By comparison with standard solutions Injected, under the same operating conditions, determine the amount and type of Aroclor using the total area method. 10. Quantitative Determination 10.1 Quantitative determinations employing the electron cap ture detector are non-stolchlometrlc measurements made by comparing the total areas for known concentrations with those for unknown concentrations. Pour variations of the area quantification procedure have been employed. Case I EC gas chromatogram of PCB unknown changed with respect to standard PCB with no evi dence of Interferences. Case II EC gas chromatogram of PCB unknown altered with respect to standard PCB with no evi dence of Interference. Case III EC gas chromatogram of PCB unknown unchanged with respect to standard PCB with evidence of Interference. Case IV EC gas chromatogram of PCB unknown altered with respect to standard PCB with evidence of Interference. 10.2 The PCB level In Case I & II samples can be determined by comparison of the total area of the peaks In the un known with a calibration curve prepared by plotting the total area of the peaks In the electron capture chroma togram of the corresponding standard vs. the number of ng of the standard Injected. Case III and IV samples are Initially subjected to the chemical cleanup proce dure followed by chromatography on alumina. If the In terferences are removed by this treatment, the PCB levels are calculated for Case III and IV in the same manner as MCNS 071451 -10- Case I and II. If dominant Interference(a) la (are) still present, the maximum FCB level la eatlmated as In Case I & II after correcting the total area for that of the interfering peak(s). In all cases, the response of the electron capture must be linear for quantitative analysis. 11. Calculations 11,1 Assuming that the calibration curve la prepared by plotting the electron capture response (total peak area) vs. nanograms of the PCB standard Injected, the PCB level present In the air, water, or sediment/ soil sample can be calculated In the following manner: (ng PCBs from) , . .6 (calib. curve) X (ml, volume of X (lO3 ul) X (10 mg) (ul of sample) (final concentrate) ( ml) ng j (Injected) = mg of PCBs present In volume or weight of the sample analyzed where (cu ft of air) (sampled from) X ( meter ) 28,32 L) (m3 ) (760-mrn Hg) cu ft ) X (i63l ) X ( YbO J - volume of air sampled In m3 1 atm and 25*C IwsSrffi) Results are expressed as ng PCBs for air samples or ppm (mg) for water and soll/sedlment samples. (kg) 12. Accuracy & Precision In order to establish the accuracy and precision In the method used for estimating PCB content In the environment, a round robin testing experiment was set up. Samples were prepared by spiking water with PCBs at the level of about 500 ppb. Indi vidual preparations Included blanks (no PCB). Aroclor 1242.only, Aroclor 1254 only, Aroclor 1260 only, all 50-50 two component mixtures, and a l/3 1/3, 1/3 preparation containing all three Aroclors. Six samples of each mixture were prepared, and tested by two analytical laboratories. HONS 071452 -11- Each result, expressed In ppb of the Individual components was converted to <f> of the total PCB added. Analysis of variance was run on all the data taken together and on each lab Independently. Two obviously defective samples were deleted from the data. CONCLUSIONS 12.1 The 95)6 confidence limits for Individual results were as follows: Either Lab Laboratory A Laboratory B 87.5 r 55.2)8 81.3 - 46.4)8 93.7 + 19.2)6 12.2 The repeatability for the Individual labs, based on repeat tests of samples having Identical prepared com positions Is as follows: Laboratory A Laboratory B +42.0)6 f n.6)6 12.3 In general. Laboratory B comes closer to properly Iden tifying the Individual Aroclors, with one exception. They found Aroclor 1254 In a mixture of Aroclor 1242 and Aroclor 1260. See below: Mixture ~~I-------- 2 5 6 7 Present Laboratory A laboratory B T54S 1242,(1254} I24g 1242,1254 1242,1254,(1260) 1242,1254 1242,1254,1260 1242,( ),1260 1242,1254,1260 1254 1254,(1260) 1254 1242,1260 1242.1260 1242,(1254),1260 1260 (1248), 1260 1260 1254,1260 (1248),1254,1260 1254,1260 Numbers In parentheses Indicate mlsldentlflcatlons. 13. Contamination 13.1 In determining PCBs by electron capture gas chromato graphy, laboratory sources of contamination can be a major problem. The samples and extracts should never be allowed to come In contact with materials other than glass. Teflon or metal. laboratory glassware and sampling gear should be thoroughly washed with hot, soapy water, rinsed with distilled water, acetone, and then hexane. All equipment should also be rinsed again with hexane Just prior to use and blanks should be frequently carried through all steps of the proce dure to Insure against the possibility of contamination. HONS 071453 -12- 14. Discussion 14.1 Extraction 14.1.1 The extraction of PCBs from water employing hexane as the extractant was found to be quantitative and to be sufficiently simple and rapid for use as a routine procedure. 14.1.2 The evaluation of this method was based on spiking water samples with standard acetone solutions of PCBs. The spiking method consis ted of adding the PCBs In acetone (25-50ul) to 500 ml of tap water in a 32 oz, narrow neck screw cap Jar. After thoroughly mixing, du plicate 225-250 ml aliquots were taken and sub jected to the proposed sample preparation and work up as outlined. The results were quanti fied by preparing a calibration curve using standard hexane solutions of the PCBs used to spike the water samples. The total area of the electron capture chromatogram was used to construct the calibration plot. 14.1.3 The average recovery and deviation achieved substantiated the applicability of the method for the quantitative recovery and analysis of PCBs from water at the ppb-ppm level, 14.2 Electron Capture Oas Chromatography 14.2.1, Columns 14.2.2 Column performance Is the key to effective gas chromatographic analysis and as such the choice of column materials Is particularly Important. Ideally, the support employed should be Inert, mechanically strong, and of high surface area. For these reasons, Chromosorb W, HP, AW-DMCS was used In all of our work. 14.2.3 A variety of polar and non-polar liquid phases were Investigated. The following columns were found to provide adequate separation, etc., for use In PCB analysis by electron capture: (w/w) DC-200, SF-96, OV-17, SE-30, SE-54, XE-60, Aplezon L, and 6% QF-1. DC-200 and XE-60 or QF-1 have been found to be the most suitable of these liquid phases. . HONS 071^54 -13- 14.2.4 Another Important consideration when working with an extremely sensitive detector and con sequently low levels of materials is column conditioning. With polar phases such as XK-60 and QF-1, we have found that operating a new column overnight at a temperature 25-50*0 higher than to be used during analysis results In a more stable column. A no-flow condition ing technique Is employed to condition non polar columns. The column Is purged with car rier gas, heated for 30 minutes at an elevated temperature without carrier flow and then cooled to room temperature. At the end of this cycle, the carrier flow Is resumed and the con ditioning is completed as In the case of the polar liquid phase. Two precautions: During conditioning, the column should not be connec ted to the detector and one should not exceed the maximum safe temperature of the liquid phase. 14.2.5 Since all liquid substrates bleed to one degree or another and columns eventually degrade, we characterize all new columns with two column performance Indicators - the number of theore tical plates (N)*and a tailing factor (T)t PiP'-DDT is employed to check these parameters because It Is known to degrade on "poor'' columns. In this manner, we can determine If the perfor mance of a new column Is satisfactory and when the column performance begins to fall off. We consider a column good If the number of theore tical plates per foot Is on the order of 400-500 with tailing factors of 1.0-1.3. Calculation of these parameters Is shown In the appendix. Additionally, there should be no significant extraneous peaks upon Injection of a pure p,pDDT standard. 14.2.6 Other chromatographic conditions that can be adjusted are column temperature and flow rates. Although resolution of a mixture Increases with decreasing temperature, a temperature should be chosen that allows the elution of all components within a convenient time period. The flow rates are optimum for our Instrument, column and detector system and, of course, should be adjusted If better results can be achieved. See Figure VII. MCNS 071455 -14- 14.2.7 Two gas chromatographic systems have been used for PCB analysis - HP Model 402 and 5750. We find that any system of Instrument and column suitable for chlorinated pesticides Is satis factory for PCB analysis. 14.2.8 The bulk of analyses In our laboratories was carried out using the systems outlined. The use of the high temperature m3 electron cap ture cell Is highly recommended. The ability to operate at higher temperatures prevents maintenance problems due to contamination from high boiling components. HONS 071456 FIGURE I FIGURE II AROCLOR 1221 HAP 402 Nl-63 EC 1/4"X6' 4* XE-60 on 80/100 Chromosorb V HP Col. Temp. -170*C InJ. Temp. Dec. Temp. -220*C - 250*C Carrier Gas - He 60 ml/mln. Purge Gbb - 10* CH4/Argon 120 ml/mln. Pulse Interval - 50 Msec Amt. Injected - 5 pi Std. Cone. - 1.48 X 10_ g/ml Range - 10 Attn. - 4 " 1 . . % - tjo -3o b mint/ fes MOHS 071458 FIGURE IV AROCLOR 1248 H6P 402 Ni-63 EC 1/4"X61 4% XE-60 on 80/100 Chromosorb W HP Col. Temp. - 190C Inj. Temp. - 220C Det. Temp. - 250C Carrier Gas - He 60 ml/min. Purge Gas - 10% cn4/Argon 120 ml/min. Pulse Interval - 50 usee Amt. Injected - 5 til Std. Cone. - 1.31 X 10-6 g/ml Range - 10 Attn. - 4 4 % $ -:2> Q> % 20 0 U HONS C71460 1 i 1 1 i 1 1 t--i--i-..-..., ........., /y & tt i< to 9 6 7 (, sf 4 \3 I-A IL .i. i j z/ 0 FIGURE V AROCLOR 1254 HAP 402 Nl-63 EC 1/4"X6' 42 XE-60 on 80/100 Chromosorb W HP Col. Temp. - 205*C Inj. Temp. - 220*C Det. Temp. - 250*C Carrier Gas - He 60 ml/mln - Purge Gas - 102 CH4/Argon 120 ml/mln. Pulse Interval - SO usee Amt. Injected - 5 vil Std. Cone. - 1.02 X 10-6 g/ml Range - 10 Attn. - 8 ' 1 i I, |i ./s "T -r----- 1---------1--------1--------1------- 1---------1------ r------1--------1--------1-------X /2 // /O f e 7 6 S V 3 2 t O ' /tttrtvfks i FIGURE VI AROCLOR 1260 H&P 402 Ni-63 EC 1/4"X61 4 2 XE-60 on 80/100 Chromosorb U HP Col. Temp. - 220*C InJ. Temp. - 220*C Det. Temp. - 250*C Carrier Gas - He 60 ml/mln Purge Gas - 102 CHS/Argon 120 ml/mln. Pulse Interval - 50 |isec Amt. Injected - 5 yl Std. Cone. - 0.98 X 10-6 g/ml Range - 10 MQNS 071462 /Htnutes FIGURE VII COLUMN PERFORMANCE INDICATORS Calculating Column Efficiency 1. Theoretical Plates, N N = 16(x/y)2 2. Tailing, T T = a/2b HONS 071463