Document g2Ee7J8xJnXOvbEK1Gva808y3
Federal Register / Vol. 51, No. 119 / Friday, June 20, 1936 / Rules and Regulations
22741
b. Place the filter or wedge, dust side up. on . the slide. Static electricity will usually keep the filter on the slide until it is cleared.
c. Hold the glass slide supporting the filler approximately 1 to 2 cm from the glaBS tube port where the acetone vapor is escaping from the heated flask. The acetone vapor stream should cause a condensation spot on the gjass slide ca. 2 to 3 cm in diameter. Move the glass slide gently in the vapor stream. The filter should clear in 2 to 5 sec. If the filter curls, distorts, or is otherwise rendered unusable, the vapor stream is probably not strong enough. Periodically wipe the outlet port with tissue to prevent liquid acetone dripping onto the filter.
d. Using the hypodermic syringe with a 22gnuge needle, place 1 to 2 drops of triacetln . on the filter. Gently lower a clean 25-mm square cover slip down onto the filter at a slight angle to reduce the possibility of forming bubbles. If too many bubbles form or the amount of triacetin is insufficient, the cover slip may become detached within a few hours.
e. Glue (he edges of the cover slip to the glass slide using a lacquer or nail polish.
Note.--If clearing is slow, the slide preparation may be heated on a hotplate (surface temperature 50 *C) for 15 min to hasten clearing. Counting may proceed immediately after clearing and mounting are completed.
Calibration and Quality Control
11. Calibration of the Walton-Deckelt graticule. The diameter, dc(mm). of the circular counting area and the disc diameter must be specified when ordering the graticule.
a. Insert any available graticule into the eyepiece and focus so that the graticule lines are sharp and clear.
b. Set the appropriate interpupillary. distance and, if applicable, reset the binocular head adjustment so that the ~ magnification remains constant.
c. Install the 40 to 45 x phase objcclive. d. Place s stage micrometer on the microscope object stage and focus the microscope on the graduate lines. e. Measure the magnified grid length, Lu(min), using the 6tage micrometer. f. Remove the graticule from the microscope and measure its actual grid length, U(mm). This can best be accomplished by using a stage fitted with, verniers. g. Calculate the circle diameter, dc(mro), for the Walton-Beckett graticule:
Uk D dc =------------
to
Example.--If L* = 108 urn, L, = 2.93 mm and D = 100 um, then dc => 2.71 mm.
h. Check the field diameter. D(accepiable range 100 mm 2 mm) with a stage micrometer upon'receipt of the graticule from the manufacturer. Determine field area (mm*).
12. Microscope adjustments. Follow the manufacturer's instructions and also the following:
a. Adjust the light source for even
Note.--To ensul*e good reproducibility, all
illumination across the field of view at the
laboratories engaged in asbestos, tremolite,
condenser Iris.
anthophyllite, and actinolile counting are
Note.--Kohler illumination is preferred,
required to participate in the Proficiency
where available. b. Focus on the particulate material to be
examined. c. Make sure that the field iris is in focus,
centered on the sample, and open only enough to fully illuminate the held of view.
d. Use the telescope ocular supplied by the
Analytical Testing (PAT) Program and should routinely participate with other asbestos, tremolite, anthophyllite, and actinolile fiber counting laboratories in the exchange of field samples to compare performance of counters.
Measurement
manufacturer to ensure that the phase rings
17. Place the slide on the mechanical stage
(annular diaphragm and phase-shifting elements) are concentric.
13. Check the phase-shift detection limit of the microscope periodically.
a. Remove the HSE/NPL phase-contrast test slide from its shipping container and center it under the phase objective.
b. Bring the blocks of grooved tines into focus.
Note.--The slide consists of seven sets of grooves (ca. 20 grooves to each block) in descending order of visibility from sets 1 to 7. The requirements for counting are that the
microscope optics must resolve the grooved lines in set 3 completely, although they may
of the calibrated microscope with the center of the filter under the objective lens. Focus the microscope on the plane of the filter.
18. Regularly check phase-ring alignment and Kohler illumination.
19. The following are the counting rules: a. Count only fibers longer than S um. Measure the length of curved fibers along the curve. b. Count only fibers with a length-to-width ratio equal to or greater than 3:1. c. For fibers that cross the boundary of the graticule field, do the following: 1.- Count any fiber longer lha 5 um that lies
appear somewhat faint, and that the grooved lines in sets 8 to 7 must be Invisible. Bets 4
entirely within the graticule area. 2. Count as Vfc fiber any fiber with only one
and 5 must be at least partially visible but
end lying within the graticule area.
may vary slightly in visibility between .
3. Do not count any fiber that crosses the
microscopes. A microscope which fails to
graticule boundary more than once.
meet these requirements has either too tow or
4. Reject and do not count all other fibers.
too high a resolution to be used for asbestos,
d. Count bundles of fibers as one fiber
tremolite, anthophyllite, and actinolile
unless individual fibers can be identified by
counting.
observing both ends of a fiber;
c. If the Image quality deteriorates, dean
e. Count enough graticule fields to. yield 100
the microscope optics and, if the problem persists, consult the. microscope manufacturer.
14. Quality control of fiber counts. a. Prepare and count field blanks along * with the field samples. Report the counts on each blank. Calculate the mean of the field blank counts and subtract this value from each sample count before reporting the results. Note.!.--The identity of-the blank filters should be unknown to the counter until all counts have been completed. Note 2: If a field blank yields fiber counts greater than 7 fibers/100 fields, report possible contamination of the samples. b. Perform blind recounts by the same counter on 10 percent of filters counted (slides relabeled by a person.other than the counter).
fibers. Count a minimum of 20 fields. Stop at 100 fields regardless of fiber count.
20. Start counting from one end of the filter and progress along a radial line to the other end, shift either up or down on the filler, and continue In the reverse direction. Select fields randomly by looking.away from the eyepiece briefly while advancing the mechanical stage. When an agglomerate covers ca. V or more . of the field of view, reject the field and select another. Do not report rejected fields In the number of total fields counted.
Note.--When counting a field, continuously scan a. range of focal planes by moving the fine focus knob to detect very fine fibers which have become embedded in the filter. The small-diameter fibers wlU be very faint but are an Important contribution to the.total. count
15. Use the following test to determine whether a pair of counts on the same filter should be rejected because of possible bias. This statistic estimates the counting repeatability at the 95% confidence level.
Discard the sample if the difference between the two counts exceeds 2.77{F)Sr, where F=average of the two fiber counts and
8r=relative standard deviation; which should
Calculations
21. Calculate and report fiber density on the filter, E (fibers/mm*); by dividing the total fiber count, F; minus the mean field blank count, B, by the number of fields, n; and the field area, At (0.00785 mm* for a properly.calibrated Walton-Beckett . graticule):
be derived by each laboratory based bn
historical in-house data. Note.--If a pair of counts is rejected as a
result of this test, recount the remaining samples in the set and test the new counts
F-B.
E S3 --fibers/mm* (n)(A,)
against the first counts. Discard all rejected
paired counts.
16. Enroll each new counter in a training
22. Calculate the concentration, C (f/cc), o
course that compares performance of
fibers in the air volume sampled. V (L), using
counters on a variety of samples using this
the effective collection area of the filter, A
procedure.
(385 mm* for a 25-mm filter):
GLEASON-000989