Document e9rDXbJB4jB164MBYVoyVLrM

3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Final Report A n a ly s is o f P F O A , P F B S , P F H S , a n d P F O S in S u rfa c e a n d G ro u n d W a te r S a m p le s C o lle c te d a t th e B ro w n s F e rry R o a d S ite L o c a te d in D e c a tu r, A L 3 rd Q u a r t e r 2 0 1 3 L a b o ra to ry R e q u e s t N u m b e r: IS O 1 3 -0 5 -0 3 M e th o d R e q u ire m e n t: 3 M M e th o d E T S -8 -0 4 4 .1 Report Date - Date of Final Signature T e s tin g L a b o ra to ry 3 M E n v iro n m e n ta l H e a lth a n d S a fe ty O p e ra tio n s E n v iro n m e n ta l L a b o ra to ry 3 M C e n te r, B ld g 2 6 0 -0 5 -N -1 7 S t. P a u l, M N 5 5 1 4 4 Requester G a ry H o h e n s te in 3 M B u ild in g 2 2 4 -5 W -0 3 S a in t P a u l, M N 5 5 1 4 4 -1 0 0 0 P ho ne : (6 5 1 ) 7 3 7 -3 5 7 0 The testing reported herein meet the requirements of ANSI/ISO/IEC 17025:2005 "General Requirements for the Competence of Testing and Calibration Laboratories", in accordance with A2LA Certificate # 2052.01. Additionally, the laboratory's quality system has been audited and was determined to be in conformance with the EPA GLPs (40 CFR 792) by an independent assessment. PAGE 1 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 3M Environm ental Laboratory 3M Environmental Laboratory Technical Director: William K. Reagen, Ph.D. 3M Principal Analytical Investigator: Susan Wolf Report Author: Kevin Eich Analytical Report ISO13-05-03 Analysis of PFOA, PFBS, PFHS, and PFOS in Surface and Ground W ater Samples Collected at the Browns Ferry Road Site Located in Decatur, AL 3rdQuarter 2013 Report Date: Date of Final Signature 1 Introduction/Summary The 3M Environmental Laboratory prepared and analyzed water samples collected by Weston Solutions personnel from Decatur, AL. Samples were collected on August 27 and 28, 2013 and returned to the 3M Environmental Laboratory on September 4, 2013 at ambient conditions. All samples were analyzed for perfluorooctanoic acid (PFOA), perfluorobutane sulfonate (PFBS), perfluorohexane sulfonate (PFHS) and perfluorooctane sulfonate (PFOS) under laboratory project number ISO13-05-03. The 3M Environmental Laboratory prepared sample containers for fourteen groundwater well and three surface water locations. Also included was a sample container for the collection of a soil/sediment rinseate blank. The results for the rinseate blank are included herein. The results for the soil/sediment samples are reported under project E13-0618. Each empty container was marked with a "fill to here" line that corresponded to a final volume of 200 mL. Sample bottle sets for each sampling location included a field sample and field sample duplicate. Select locations also included a target analyte field matrix spike. Containers designated for field matrix spikes were fortified with an appropriate matrix spike solution containing PFOA, PFBS, PFHS, and PFOS prior to being sent to the field for sample collection. All sample containers were fortified with internal standards and surrogate recovery standards prior to being sent to the field for sample collection. Samples were analyzed for PFOA, PFBS, PFHS, and PFOS by ETS-8-044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis". Table 1 summarizes the sample results using the analytical method identified above. All results for quality control samples prepared and analyzed with the samples will be reported and discussed elsewhere in this report. ........ Testing Cert. # 2052.01 The testing reported herein meet the requirements of ANSI/ISO/IEC 17025:2005 "General Requirements for the Competence of Testing and Calibration Laboratories", in accordance with A2LA Certificate # 2052.01. Additionally, the laboratory's quality system has been audited and was determined to be in conformance with the EPA GLPs (40 CFR 792) by an independent assessment. PAGE 2 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 1. S am ple R e su lts S um m ary (1) 3M LIMS ID ISO13-05-03-001 ISO13-05-03-002 ISO13-05-03-004 ISO13-05-03-005 ISO13-05-03-006 ISO13-05-03-007 ISO13-05-03-008 ISO13-05-03-009 ISO13-05-03-011 ISO13-05-03-012 ISO13-05-03-013 ISO13-05-03-014 ISO13-05-03-016 ISO13-05-03-017 ISO13-05-03-018 ISO13-05-03-019 ISO13-05-03-020 ISO13-05-03-021 Sample Description BFRS-GW-MW1R-0-130827 BFRS-GW-MW1R-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW3R-0-130827 BFRS-GW-MW3R-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW4R-0-130827 BFRS-GW-MW4R-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW4L-0-130827 BFRS-GW-MW4L-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW5R-0-130828 BFRS-GW-MW5R-DB-130828 Average %RPD Sample/Sample Dup BFRS-GW-MW5L-0-130828 BFRS-GW-MW5L-DB-130828 Average %RPD Sample/Sample Dup BFRS-GW-MW6R-0-130828 BFRS-GW-MW6R-DB-130828 Average %RPD Sample/Sample Dup BFRS-GW-MW6L-0-130828 BFRS-GW-MW6L-DB-130828 Average %RPD Sample/Sample Dup BFRS-GW-MW7R-0-130827 BFRS-GW-MW7R-DB-130827 Average %RPD Sample/Sample Dup PFOA Concentration (ng/mL) 1.79 1.70 1.75 5.2 22.0 16.8 19.4 27 (2) 9.87 7.59 8.73 26 (2) 0.269 0.264 0.267 1.9 0.212 0.206 0.209 2.9 0.635 0.591 0.613 7.2 2.57 2.65 2.61 3.1 0.195 0.238 0.217 20 8.79 8.24 8.52 6.5 PFBS PFHS PFOS Concentration Concentration Concentration (ng/mL) (ng/mL) (ng/mL) 0.108 0.110 0.109 1.8 4.34 3.47 3.91 22 (2) 0.874 0.869 0.872 0.57 21.8 17.4 19.6 22 (2) 1.69 1.69 1.69 0.0 26.0 20.4 23.2 24 (2) 0.728 0.567 0.648 25 (2) 18.5 14.5 16.5 24 (2) 25.8 20.9 23.4 21 (2) 0.0964 0.0906 0.0935 6.2 0.263 0.245 0.254 7.1 0.524 0.556 0.540 5.9 0.0535 0.0675 0.0605 23(2) 0.131 0.144 0.138 9.5 0.0874 0.0905 0.0890 3.5 0.165 0.170 0.168 3.0 0.249 0.250 0.262 0.284 0.273 8.1 0.665 0.688 1.24 1.29 1.27 4.0 7.64 7.71 0.250 0.40 0.677 3.4 7.68 0.91 0.0536 0.0565 0.0551 5.3 0.0853 0.0857 0.0855 0.47 0.324 0.360 0.342 11 0.533 1.80 24.4 0.497 1.74 25.7 0.515 1.77 25.1 7.0 3.4 5.2 NA = Not Applicable (1) All samples associated with the reported sampling location were analyzed by internal standard calibration. The analytical method uncertainties for the reported results using method ETS-8-044.1 by direct injection are as follows: PFOA 13%, PFBS 13%, PFHS 12%, and PFOS 16%. (2) The RPD value did not meet method acceptance criteria of <20%. PAGE 3 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 1 c o n tin u e d . S am ple R esu lts S um m ary (1) 3M LIMS ID ISO13-05-03-022 ISO13-05-03-023 ISO13-05-03-025 ISO13-05-03-026 ISO13-05-03-027 ISO13-05-03-028 ISO13-05-03-029 ISO13-05-03-030 ISO13-05-03-032 ISO13-05-03-033 ISO13-05-03-034 ISO13-05-03-035 ISO13-05-03-036 ISO13-05-03-037 ISO13-05-03-039 ISO13-05-03-040 ISO13-05-03-041 ISO13-05-03-042 Sample Description BFRS-GW-MW8R-0-130828 BFRS-GW-MW8R-DB-130828 Average %RPD Sample/Sample Dup BFRS-GW-MW9R-0-130827 BFRS-GW-MW9R-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW9L-0-130827 BFRS-GW-MW9L-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW10R-0-130827 BFRS-GW-MW10R-DB-130827 Average %RPD Sample/Sample Dup BFRS-GW-MW11L-0-130828 BFRS-GW-MW11L-DB-130828 Average %RPD Sample/Sample Dup BFRS-SW-ET02-0-130828 BFRS-SW-ET02-DB-130828 Average %RPD Sample/Sample Dup BFRS-SW-ET03-0-130828 BFRS-SW-ET03-DB-130828 Average %RPD Sample/Sample Dup BFRS-SW-ET04-0-130828 BFRS-SW-ET04-DB-130828 Average %RPD Sample/Sample Dup BFRS-SS08-RB-0-130828 BFRS-GW-TRIP01 -0 PFOA Concentration (ng/mL) 0.220 0.216 0.218 1.8 0.112 0.109 0.111 2.7 0.0301 0.0344 0.0323 13 0.644 0.669 0.657 3.8 0.286 0.275 0.281 3.9 0.0314 0.0313 0.0314 0.32 0.147 0.178 0.163 19 0.0507 0.0465 0.0486 8.6 <0.0240 <0.0240 PFBS Concentration (ng/mL) 0.0590 0.0575 0.0583 2.6 <0.0250 <0.0250 <0.0250 NA <0.0250 <0.0250 <0.0250 NA 0.0692 0.0658 0.0675 5.0 <0.0250 <0.0250 <0.0250 NA 0.0350 0.0350 0.0350 0.0 0.0565 0.0560 0.0563 0.89 0.0371 0.0352 0.0362 5.3 <0.0250 <0.0250 PFHS Concentration (ng/mL) 0.121 0.117 0.119 3.4 0.0579 0.0510 0.0545 13 <0.0250 <0.0250 <0.0250 NA 0.293 0.306 0.300 4.3 0.0823 0.0730 0.0777 12 <0.0250 <0.0250 <0.0250 NA 0.0820 0.0792 0.0806 3.5 <0.0250 <0.0250 <0.0250 NA <0.0250 <0.0250 PFOS Concentration (ng/mL) 0.268 0.250 0.259 6.9 <0.0232 <0.0232 <0.0232 NA 0.0446 0.0443 0.0445 0.67 0.536 0.594 0.565 10 0.387 0.378 0.383 2.4 0.0547 0.0544 0.0546 0.55 0.295 0.260 0.278 13 0.113 0.105 0.109 7.3 <0.0232 <0.0232 NA = Not Applicable (1) All samples associated with the reported sampling location were analyzed by internal standard calibration. The analytical method uncertainties for the reported results using method ETS-8-044.1 by direct injection are as follows: PFOA 13%, PFBS 13%, PFHS 12%, and PFOS 16%. (2) The RPD value did not meet method acceptance criteria of <20%. PAGE 4 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 2 Methods - Analytical and Preparatory 2.1 Methods Analysis was completed using 3M Environmental Laboratory Method ETS-8-044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis". Table 2. T arget A n a lyte s Target Analytes Perfluorobutanoic Acid (C4 Acid) Perfluorooctanoic Acid (C8 Acid) Perfluorobutanesulfonate (C4 Sulfonate ) Perfluorooctanesulfonate (C8 Sulfonate) Acronym PFOA PFBS PFHS PFOS Reference Material Structure Linear and Branched Linear Linear Linear and Branched 2.2 Sample Collection All samples were collected in 250 mL NalgeneTM (high-density polyethylene) bottles prepared at the 3M Environmental Laboratory. Samples were collected on August 27 and 28, 2013 and returned to the 3M Environmental Laboratory on September 4, 2013 at ambient conditions. Prior to sample collection, bottles designated for field matrix spikes were spiked in the laboratory with a known volume of an appropriate matrix spiking solution containing PFOA, PFBS, PFHS, and PFOS. Table 3. S am ple D e s c rip tio n Key Code. String Number 1 2 String Descriptor Sample Location Sample Type 3 Well Identifier 4 Well Level 5 Sample Type 6 Sampling Date Example: BFRS-GW-MW4L-0-130827 BFRS= Browns Ferry Road Site GW= Ground Water SW=Surface Water Example: MW4 R = Residuum shallow water-bearing zone L = Bedrock water-bearing zone 0 =primary sample DB =duplicate sample FMS = Field Matrix Spike 130827 - August 27, 2013 2.3 Sample Preparation 9/10/13: Samples were prepared by removing an aliquot of the well mixed sample and placing it in an autovial for analysis. 9/13/13: Samples requiring dilution prior to analysis were prepared by removing a 0.4 mL aliquot of the well mixed sample and diluting it with 0.4 mL of methanol (dilution factor of 2). All samples were analyzed by internal standard calibration. During the preparation of the laboratory control samples, an aliquot of a separate internal standard spiking solution was added to the laboratory control samples (nominal concentration of 1 ng/mL). The sample bottles were spiked with an internal standard mix at a nominal concentration of 1 ng/mL prior to being sent to the field for sample collection. PAGE 5 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 2.4 Analysis All samples and quality control samples were analyzed for the four target analytes using high performance liquid chromatography/tandem mass spectrometry (HPLC/MS/MS). Pertinent instrument parameters, the liquid chromatography gradient program, and the specific mass transitions analyzed are described in the tables below. The following analytical runs were processed and reported herein: 9/10/13: Direct Injection/internal standard analysis PFOA, PFBS, PFHS, PFOS, 13C4-PFOA, and 13C4-PFOS; all locations. 9/13/13: Solvent dilution / internal standard analysis PFOS and 13C4-PFOS; locations BFRSGW-MW3R, BFRS-GW-MW4R, and BFRS-GW-MW7R. Due to the nature of the samples, the wide range of concentrations found in the samples, and the environmental occurrence of multiple isomers of the laboratory's analytes of interest, the software used for processing the analytical results is not able to consistently integrate the analytical peak, manual integration of the analytical peak is necessary. All manual integrations are performed following the procedures outlined in method ETS-12-010. The consistency of the laboratory's integration is ensured through the training of laboratory personnel, the peer review process required for all manual integrations, the review of manual integrations by the QAU, and where necessary the review of manual integrations by laboratory management. Table 4. In s tru m e n t P aram eters Instrument Name Analysis Dates Analytical Method Liquid Chromatograph Guard column Analytical column Injection Volume Mass Spectrometer Ion Source Polarity Software ETS Buster 9/10/13 and 9/13/13 ETS-8-044.1 Agilent 1100 Betasil C18 (4.6 mm X 100 mm), 5 g Betasil C18 (4.6 mm X 100 mm), 5 g 20 ijL Applied Biosystems API 4000 Turbo Spray Negative Analyst 1.6.1 Table 5. L iq u id C h ro m a to g ra p h y G radient P rogram Step 0 1 2 3 4 5 6 7 8 9 Total Time (min) 0 0.5 4.0 6.0 11.0 13.0 13.5 16.0 16.5 19.0 Flow Rate (pL/min) 750 750 750 750 750 750 750 750 750 750 2 mM Ammonium Acetate 90.0 90.0 70.0 70.0 20.0 20.0 10.0 10.0 90.0 90.0 Methanol 10.0 10.0 30.0 30.0 80.0 80.0 90.0 90.0 10.0 10.0 PAGE 6 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 6. M ass T ra n s itio n s Analyte Mass Transition Q1/Q3 Internal Standard Mass Transition Q1/Q3 413/369 PFOA 413/219 [13C8]PFOA 421/376 413/169 PFBS 299/80 299/99 [18O2]PFBS 303/84 PFHS 399/99 399/80 [13Cs]PFHS 402/99 499/99 PFOS 499/80 [13C8]-PFOS 507/80 499/130 [13C4]-PFOA surrogate 417/372 [13C8]-PFOA 421/376 [13C4]-PFOS surrogate 503/80 [13C8]-PFOS 507/80 Dwell time was 50 msec for each transition. The individual transitions were summed to produce a "total ion chromatogram" (TIC), which was used for quantitation. 3 Data Analysis 3.1 Calibration 9/10/13: Direct Injection / internal standard calibration. Samples were analyzed against a matrix-matched stable isotope internal standard calibration curve. Calibration standards were prepared by spiking known amounts of the stock solution into laboratoryprepared reagent water. The calibration standards contained an internal standard mix at a nominal concentration of 1.0 ng/mL. Ten curve points ranging from 0.025 ng/mL to 25 ng/mL (nominal) were analyzed with the samples. Of these ten calibration standards, nine contained the surrogates at concentrations ranging from 0.0250 ng/mL to 10 ng/mL (nominal). A quadratic, 1/x weighted, calibration curve of the ratio of the standard peak area counts over the internal standard peak area counts was used to fit the data for each analyte. The data were not forced through zero during the fitting process. 9/13/13: Solvent Dilution / internal standard calibration. Samples were analyzed against an internal standard matrix-matched calibration curve. Calibration standards were prepared by spiking known amounts of the stock solution into a 50:50 Methanol: Laboratory Reagent Water. The calibration standards contained an internal standard mix at a nominal concentration of 0.5 ng/mL. Twelve standards ranging from 0.0125 ng/mL to 25 ng/mL (nominal) were analyzed with the samples. Of these twelve calibration standards, ten contained the surrogates at concentrations ranging from 0.0125 ng/mL to 10 ng/mL (nominal). A quadratic, 1/x weighted, calibration curve of the ratio of the standard peak area counts over the internal standard peak area counts was used to fit the data for each analyte. The data were not forced through zero during the fitting process. For both analyses, each curve point was quantitated using the overall calibration curve and reviewed for accuracy. Method calibration accuracy requirements of 10025% (10030% for the lowest curve point) were met for all analytes. The correlation coefficient (r) was greater than 0.995 for all analytes. 3.2 System Suitability A calibration standard was analyzed four times at the beginning of the analytical sequence to demonstrate overall system suitability. The acceptance criteria for system suitability samples is less PAGE 7 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 than or equal to 5% relative standard deviation (RSD) for peak area or peak area ratio and retention time criteria of less than or equal to 2% RSD. These criteria were met for all analytes. 3.3 Limit of Quantitation (LOQ) The LOQ as defined in method ETS-8-044.1 is the lowest non-zero calibration standard in the curve that meets linearity and accuracy requirements and for which the peak area ratio (or peak area) is at least two times the average peak area ratio (or peak area) of the method blanks. The LOQs associated with the sample analysis are listed in the table below. Table 7. LOQ Analysis Date Calibration 9/10/13 9/13/13 NA=Not Applicable Internal standard Internal standard Dilution Factor NA 2 PFOA LOQ, ng/mL <0.0240 NA PFBS LOQ, ng/mL <0.0250 NA PFHS LOQ, ng/mL <0.0250 NA PFOS LOQ, ng/mL <0.0232 <0.0232 3.4 Continuing Calibration During the course of the analytical sequence, several continuing calibration verification samples (CCVs) were analyzed to confirm that the instrument response and the initial calibration curve were still in control. The method acceptance criteria of 100% 25% was met for all analytes for all analyses. 3.5 Blanks Three types of blanks were prepared and analyzed with the samples: method/procedural blanks, a field/trip blank, and a single equipment rinseate blank. Method blank results were reviewed and used to evaluate method performance to determine the LOQ for each analyte. 3.6 Lab Control Spikes (LCSs) Low, mid, and high lab control spikes were prepared for the target analytes and analyzed in triplicate, while only low and high lab control samples were prepared for the surrogates. LCSs were prepared by spiking known amounts of the analytes into laboratory reagent water to produce the desired concentration. The LCS samples were then diluted in a similar manner as the samples (if diluted) prior to sample analysis. The method acceptance criteria, average of each level of LCS should be within 100% 20% with an RSD <20%, was met for all LCSs. All batch LCS recovery results were used in the evaluation of the analytical method uncertainty in section 3.7 of the report. The following calculations were used to generate data in Table 8. Calculated Concentration LCS Percent Recovery Spike Concentration % LCS% RSD = standard deviation LCS replicates * 10Q% average LCS recovery PAGE 8 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 8. Lab C o n tro l S pike R esults ETS-8-044.1 Analyzed 9/10/13 Internal Standard Calibration Lab ID LCS-130910-1 LCS-130910-2 LCS-130910-3 Average %RSD LCS-130910-4 LCS-130910-5 LCS-130910-6 Average %RSD LCS-130910-7 LCS-130910-8 LCS-130910-9 Average %RSD PFOA (Linear and Branched) Spiked Concentration (ng/mL) Calculated Concentration (ng/mL) %Recovery 0.190 0.190 0.190 1.90 1.90 1.90 16.2 16.2 16.2 0.179 0.187 0.177 95.4% 3.0% 1.97 1.98 1.99 104% 0.96% 14.4 14.5 14.4 89.2% 0.45% 94.2 98.6 93.3 103 104 105 89.0 89.7 89.0 Spiked Concentration (ng/mL) PFBS Calculated Concentration (ng/mL) 0.198 0.198 0.198 1.98 1.98 1.98 16.9 16.9 16.9 0.192 0.200 0.195 98.8% 2.0% 2.19 2.05 2.07 106% 3.6% 15.7 15.6 15.5 92.3% 0.60% %Recovery 97.1 101 98.3 110 103 104 92.9 92.3 91.8 ETS-8-044.1 Analyzed 9/10/13 Internal Standard Calibration Lab ID LCS-130910-1 LCS-130910-2 LCS-130910-3 Average %RSD LCS-130910-4 LCS-130910-5 LCS-130910-6 Average %RSD LCS-130910-7 LCS-130910-8 LCS-130910-9 Average %RSD Spiked Concentration (ng/mL) 0.198 0.198 0.198 1.98 1.98 1.98 16.9 16.9 16.9 PFHS Calculated Concentration (ng/mL) 0.197 0.211 0.193 101% 5.0% 2.21 2.42 2.21 115% 5.0% 16.8 16.1 16.2 96.8% 2.5% %Recovery 99.7 107 97.3 112 122 112 99.6 95.1 95.6 PFOS (Linear and Branched) Spiked Concentration (ng/mL) Calculated Concentration (ng/mL) %Recovery 0.184 0.184 0.184 1.84 1.84 1.84 15.7 15.7 15.7 0.180 0.180 0.202 102% 6.9% 1.98 1.89 1.81 103% 4.1% 13.7 13.6 13.6 86.8% 0.24% 97.7 97.9 110 107 103 98.6 87.0 86.6 86.9 PAGE 9 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 8 continued. Lab C ontrol Spike Results ETS-8-044.1 Analyzed 9/10/13 Internal Standard Calibration Lab ID LCS-130910-1 LCS-130910-2 LCS-130910-3 Average %RSD LCS-130910-4 LCS-130910-5 LCS-130910-6 Average %RSD Spiked Concentration (ng/mL) 0.198 0.198 0.198 1.98 1.98 1.98 13C4-PFOA Calculated Concentration (ng/mL) 0.203 0.209 0.209 105% 1.6% 2.15 2.01 2.08 105% 3.3% %Recovery 103 106 106 109 102 105 Spiked Concentration (ng/mL) 0.189 0.189 0.189 1.89 1.89 1.89 13C4-PFOS Calculated Concentration (ng/mL) 0.208 0.195 0.219 110% 5.9% 2.11 1.93 1.88 104% 6.4% %Recovery 110 103 116 112 102 99.4 ETS-8-044.1 Analyzed 9/13/13 Internal Standard Calibration Lab ID LCS-130913-1 LCS-130913-2 LCS-130913-3 Average %RSD LCS-130913-4 LCS-130913-5 LCS-130913-6 Average %RSD LCS-130913-7 LCS-130913-8 LCS-130913-9 Average %RSD PFOS (Linear and Branched) Spiked Concentration (ng/mL) Calculated Concentration (ng/mL) %Recovery 0.184 0.184 0.184 1.84 1.84 1.84 32.2 32.2 32.2 0.166 0.158 0.170 89.6% 3.6% 1.80 1.70 1.77 95.6% 2.9% 31.1 31.7 31.4 97.5% 0.92% 90.2 86.1 92.5 98.0 92.6 96.2 96.6 98.4 97.6 Spiked Concentration (ng/mL) 13C4-PFOS Calculated Concentration (ng/mL) 0.189 0.189 0.189 1.89 1.89 1.89 0.181 0.184 0.181 96.1% 0.99% 1.76 1.85 1.88 96.8% 3.4% %Recovery 95.5 97.2 95.6 93.1 98.0 99.4 PAGE 10 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 3.7 Analytical Method Uncertainty Analytical method uncertainty is based on historical QC data that is control charted and used to evaluate method accuracy and precision. The method uncertainty is calculated following ETS-12012.2. The standard deviation is calculated for the set of accuracy results (in %) obtained for the QC samples. The expanded uncertainty is calculated by multiplying the standard deviation by a factor of 2, which corresponds to a confidence level of 95%. Table 9. A n a ly tic a l M ethod U n ce rta in ty Analyte PFOA PFBS PFHS PFOS Calibration Internal Internal Internal Internal Standard Deviation (%) 6.42 6.72 6.12 8.22 Method Uncertainty 13% 13% 12% 16% 3.8 Field Matrix Spikes (FMS) Field matrix spike sample containers were prepared for select sampling locations to verify that the analytical method is applicable to the collected matrix. Field matrix spikes are generated by adding a measured volume of field sample to a container spiked by the laboratory with the target analytes prior to shipping sample containers to the field for sample collection. Field matrix spikes must be at least 0.5 times the analyte concentration to be considered an appropriate spike level. Field matrix spike recoveries within method acceptance criteria of 10030% confirm that "unknown" components in the sample matrix do not significantly interfere with the extraction and analysis of the analytes of interest. The reference standards for PFOA and PFOS in the field matrix spiking solution consisted of linear and branched isomers. The locations for which FMS sample containers were prepared are listed in Table 10. Field matrix spikes are presented in section 4 of this report. FMS ,,Recovery = 4(-S--a-m--p- -le---C--o--n--c-e--n-t-r-a--t-i-o-n---o--f-F--M---S------A--v-e--r-a-g2-e---C--o--n--c--e-n--t-r-a--t-i-o-n--:-F--i-e-l-d--S--a--m--pL-le---&---F--ie--ld---S--a--m- pL-le---D--u-p*.1)0 0 % Spike Concentraton Table 10. Field M a trix S pike (FMS) C o n ce n tra tio n s Sampling location GW-MW1R, GW-MW4L, GW-MW5L, GWMW8R, GW-MW10R, SW-ET03, and Trip Blank PFOA ng/mL 0.958 PFBS ng/mL 1.00 PFHS ng/mL 1.00 PFOS ng/mL 0.927 4 Data Summary and Discussion Tables 11-18 below summarize the sample results and target analyte field matrix spike recoveries for the sampling locations as well as the Trip Blank. The surrogate recovery standard recoveries are presented in Table 19. Each table provides the average concentration and the relative percent difference (%RPD) of the sample and sample duplicate. Results and average values are rounded to three significant figures. Percent relative difference (%RPD) values are rounded to two significant figures. Field matrix spikes meeting the method acceptance criteria of 30%, demonstrate that the method is appropriate for the given matrix. PAGE 11 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 All target analyte field matrix spikes met method acceptance criteria with the following exception: BFRS-SW-ET03; The 13C4-PFOS results for this location are not reportable due to the analyte area counts were below the LLOQ area counts and less than two times the average area counts for the method blanks. This location had acceptable FMS recoveries for each analyte. Table 11. BFRS-GW -MW 1R 3M LIMS ID Description ISO13-05-03-001 BFRS-GW-MW1R-0-130827 ISO13-05-03-002 BFRS-GW-MW1R-DB-130827 ISO13-05-03-003 BFRS-GW-MW1R-FMS-130827 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 1.79 NA 1.70 NA 2.55 84.0 1.75 ng/mL 5.2% PFBS Concentration (ng/mL) %Recovery 0.108 NA 0.110 NA 1.02 91.1 0.109 ng/mL 1.8% 3M LIMS ID Description ISO13-05-03-001 BFRS-GW-MW1 R-0-130827 ISO13-05-03-002 BFRS-GW-MW1 R-DB-130827 ISO13-05-03-003 BFRS-GW-MW1 R-FMS-130827 Average Concentration (ng/mL) %RPD NA = Not Applicable PFHS Concentration (ng/mL) %Recovery 0.874 NA 0.869 NA 1.92 105 0.872 ng/mL 0.57% PFOS Concentration (ng/mL) %Recovery 1.69 NA 1.69 NA 2.62 100 1.69 ng/mL 0.0% Table 12. BFRS-GW -MW 4L 3M LIMS ID Description ISO13-05-03-008 BFRS-GW-MW4L-0-130827 ISO13-05-03-009 BFRS-GW-MW4L-DB-130827 ISO13-05-03-010 BFRS-GW-MW4L-FMS-130827 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 0.269 NA 0.264 NA 1.22 99.5 0.267 ng/mL 1.9% PFBS Concentration (ng/mL) %Recovery 0.0964 NA 0.0906 NA 1.10 101 0.0935 ng/mL 6.2% 3M LIMS ID Description ISO13-05-03-008 BFRS-GW-MW4L-0-130827 ISO13-05-03-009 BFRS-GW-MW4L-DB-130827 ISO13-05-03-010 BFRS-GW-MW4L-FMS-130827 Average Concentration (ng/mL) %RPD NA = Not Applicable PFHS Concentration (ng/mL) %Recovery 0.263 NA 0.245 NA 1.32 107 0.254 ng/mL 7.1% PFOS Concentration (ng/mL) %Recovery 0.524 NA 0.556 NA 1.55 109 0.540 ng/mL 5.9% PAGE 12 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 13. BFRS-GW-MW5L 3M LIMS ID Description ISO13-05-03-013 BFRS-GW-MW5L-0-130828 ISO13-05-03-014 BFRS-GW-MW5L-DB-130828 ISO13-05-03-015 BFRS-GW-MW5L-FMS-130828 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 0.635 NA 0.591 NA 1.59 102 0.613 ng/mL 7.2% PFBS Concentration (ng/mL) %Recovery 0.165 NA 0.170 NA 1.11 94.3 0.168 ng/mL 3.0% 3M LIMS ID Description ISO13-05-03-013 BFRS-GW-MW5L-0-130828 ISO13-05-03-014 BFRS-GW-MW5L-DB-130828 ISO13-05-03-015 BFRS-GW-MW5L-FMS-130828 Average Concentration (ng/mL) %RPD NA = Not Applicable PFHS Concentration (ng/mL) %Recovery 0.262 NA 0.284 NA 1.35 108 0.273 ng/mL 8.1% PFOS Concentration (ng/mL) %Recovery 1.24 NA 1.29 NA 2.26 107 1.27 ng/mL 4.0% Table 14. BFRS-GW -MW 8R 3M LIMS ID Description ISO13-05-03-022 BFRS-GW-MW8R-0-130828 ISO13-05-03-023 BFRS-GW-MW8R-DB-130828 ISO13-05-03-024 BFRS-GW-MW8R-FMS-130828 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 0.220 NA 0.216 NA 1.23 106 0.218 ng/mL 1.8% PFBS Concentration (ng/mL) %Recovery 0.0590 NA 0.0575 NA 1.01 95.2 0.0583 ng/mL 2.6% 3M LIMS ID Description ISO13-05-03-022 BFRS-GW-MW8R-0-130828 ISO13-05-03-023 BFRS-GW-MW8R-DB-130828 ISO13-05-03-024 BFRS-GW-MW8R-FMS-130828 Average Concentration (ng/mL) %RPD NA = Not Applicable PFHS Concentration (ng/mL) %Recovery 0.121 NA 0.117 NA 1.15 103 0.119 ng/mL 3.4% PFOS Concentration (ng/mL) %Recovery 0.268 NA 0.250 NA 1.17 98.3 0.259 ng/mL 6.9% PAGE 13 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 15. BFRS-GW-MW10R 3M LIMS ID Description ISO13-05-03-029 BFRS-GW-MW10R-0-130827 ISO13-05-03-030 BFRS-GW-MW10R-DB-130827 ISO13-05-03-031 BFRS-GW-MW10R-FMS-130827 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 0.644 NA 0.669 NA 1.50 88.0 0.657 ng/mL 3.8% PFBS Concentration (ng/mL) %Recovery 0.0692 NA 0.0658 NA 1.15 108 0.0675 ng/mL 5.0% PFHS 3M LIMS ID Description ISO13-05-03-029 BFRS-GW-MW10R-0-130827 ISO13-05-03-030 BFRS-GW-MW10R-DB-130827 ISO13-05-03-031 BFRS-GW-MW10R-FMS-130827 Average Concentration (ng/mL) %RPD Concentration (ng/mL) %Recovery 0.293 NA 0.306 NA 1.26 96.1 0.300 ng/mL 4.3% NA = Not Applicable PFOS Concentration (ng/mL) %Recovery 0.536 NA 0.594 NA 1.42 92.2 0.565 ng/mL 10% Table 16. BFRS-SW-ET03 3M LIMS ID Description ISO13-05-03-036 BFRS-SW-ET03-0-130828 ISO13-05-03-037 BFRS-SW-ET03-DB-130828 ISO13-05-03-038 BFRS-SW-ET03-FMS-130828 Average Concentration (ng/mL) %RPD PFOA Concentration (ng/mL) %Recovery 0.147 NA 0.178 NA 1.16 104 0.163 ng/mL 19% PFBS Concentration (ng/mL) %Recovery 0.0565 NA 0.0560 NA 1.04 98.4 0.0563 ng/mL 0.89% 3M LIMS ID Description ISO13-05-03-036 BFRS-SW-ET03-0-130828 ISO13-05-03-037 BFRS-SW-ET03-DB-130828 ISO13-05-03-038 BFRS-SW-ET03-FMS-130828 Average Concentration (ng/mL) %RPD NA = Not Applicable PFHS Concentration (ng/mL) %Recovery 0.0820 NA 0.0792 NA 1.06 97.9 0.0806 ng/mL 3.5% PFOS Concentration (ng/mL) %Recovery 0.295 NA 0.260 NA 1.36 117 0.278 ng/mL 13% PAGE 14 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 17. Trip Blank 3M LIMS ID ISO13-05-03-042 ISO13-05-03-043 Description BFRS-GW-TRIP01 -0 BFRS-GW-TRIP01-FMS PFOA Concentration (ng/mL) <0.0240 0.874 %Recovery NA 91.2 PFBS Concentration (ng/mL) <0.0250 0.974 %Recovery NA 97.4 3M LIMS ID ISO13-05-03-042 ISO13-05-03-043 NA = Not Applicable Description BFRS-GW-TRI P01-0 BFRS-GW-TRIP01-FMS PFHS Concentration (ng/mL) <0.0250 1.03 %Recovery NA 103 PFOS Concentration (ng/mL) <0.0232 0.856 %Recovery NA 92.3 Table 18. S o il/S e d im e n t Rinseate B lank 3M LIMS ID ISO13-05-03-041 Description BFRS-SS08-RB-0-130828 PFOA PFBS PFHS PFOS Concentration (ng/mL) <0.0240 Concentration (ng/mL) <0.0250 Concentration (ng/mL) <0.0250 Concentration (ng/mL) <0.0232 Table 19. S u rro g a te R ecovery 3M LIMS ID ISO13-05-03-001 ISO13-05-03-002 ISO13-05-03-003 ISO13-05-03-004 ISO13-05-03-005 ISO13-05-03-006 ISO13-05-03-007 ISO13-05-03-008 ISO13-05-03-009 ISO13-05-03-010 ISO13-05-03-011 ISO13-05-03-012 ISO13-05-03-013 ISO13-05-03-014 ISO13-05-03-015 Description BFRS-GW-MW1R-0-130827 BFRS-GW-MW1R-DB-130827 BFRS-GW-MW1R-FMS-130827 BFRS-GW-MW3R-0-130827 BFRS-GW-MW3R-DB-130827 BFRS-GW-MW4R-0-130827 BFRS-GW-MW4R-DB-130827 BFRS-GW-MW4L-0-130827 BFRS-GW-MW4L-DB-130827 BFRS-GW-MW4L-FMS-130827 BFRS-GW-MW5R-0-130828 BFRS-GW-MW5R-DB-130828 BFRS-GW-MW5L-0-130828 BFRS-GW-MW5L-DB-130828 BFRS-GW-MW5L-FMS-130828 13C4-PFOA %Recovery 100 89.9 84.7 91.2 90.6 97.6 90.8 84.3 90.1 89.9 95.4 91.3 96.6 92.0 94.9 13C4-PFOS %Recovery 91.9 95.8 106 97.8, 94.7 (12) 100, 85.7 (2) 99.7, 93.3 (2) 101, 90.3 (2) 99.5 99.1 94.5 88.7 91.9 98.6 95.0 106 (1) NR=Not Reportable; Area counts were below the LLOQ and were not two times the method blank area counts. (2) Result from analysis on 9/13/13 for PFOS and 13C4-PFOS surrogate. PAGE 15 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. ISO13-05-03 Table 19 co n tin u e d . S u rro g a te R ecovery 3M LIMS ID ISO13-05-03-016 ISO13-05-03-017 ISO13-05-03-018 ISO13-05-03-019 ISO13-05-03-020 ISO13-05-03-021 ISO13-05-03-022 ISO13-05-03-023 ISO13-05-03-024 ISO13-05-03-025 ISO13-05-03-026 ISO13-05-03-027 ISO13-05-03-028 ISO13-05-03-029 ISO13-05-03-030 ISO13-05-03-031 ISO13-05-03-032 ISO13-05-03-033 ISO13-05-03-034 ISO13-05-03-035 ISO13-05-03-036 ISO13-05-03-037 ISO13-05-03-038 ISO13-05-03-039 ISO13-05-03-040 ISO13-05-03-041 ISO13-05-03-042 ISO13-05-03-043 Description BFRS-GW-MW6R-0-130828 BFRS-GW-MW6R-DB-130828 BFRS-GW-MW6L-0-130828 BFRS-GW-MW6L-DB-130828 BFRS-GW-MW7R-0-130827 BFRS-GW-MW7R-DB-130827 BFRS-GW-MW8R-0-130828 BFRS-GW-MW8R-DB-130828 BFRS-GW-MW8R-FMS-130828 BFRS-GW-MW9R-0-130827 BFRS-GW-MW9R-DB-130827 BFRS-GW-MW9L-0-130827 BFRS-GW-MW9L-DB-130827 BFRS-GW-MW10R-0-130827 BFRS-GW-MW10R-DB-130827 BFRS-GW-MW10R-FMS-130827 BFRS-GW-MW11L-0-130828 BFRS-GW-MW11L-DB-130828 BFRS-SW-ET02-0-130828 BFRS-SW-ET02-DB-130828 BFRS-SW-ET03-0-130828 BFRS-SW-ET03-DB-130828 BFRS-SW-ET03-FMS-130828 BFRS-SW-ET04-0-130828 BFRS-SW-ET04-DB-130828 BFRS-SS06-RB-0-130628 BFRS-GW-TRIP01 -0 BFRS-GW-TRIP01-FMS 13C4-PFOA %Recovery 103 98.4 94.5 109 101 94.2 96.0 94.3 94.1 91.5 86.5 91.9 101 88.4 92.6 91.3 89.7 76.0 85.1 93.7 91.3 87.2 94.5 98.2 92.4 90.8 98.9 93.6 13C4-PFOS %Recovery 102 95.8 94.7 91.8 98.8, 95.4 (2) 90.1,92.2 (2) 105 90.3 97.4 99.2 85.3 92.9 93.0 90.4 105 103 104 107 97.0 110 NR (1) NR (1) NR (1) 99.8 94.6 95.7 109 93.3 (1) NR=Not Reportable; Area counts were below the LLOQ and were not two times the method blank area counts. (2) Result from analysis on 9/13/13 for PFOS and 13C4-PFOS surrogate. PAGE 16 OF 17 3M ENVIRONMENTAL LABORATORY REPORT NO. IS013-05-03 5 Conclusion Laboratory control spikes were used to determine the analytical method accuracy and precision for all analytes. The accuracy and precision were then used to estimate the method uncertainty for the results. Field and laboratory matrix spike recoveries demonstrated that the analytical method was appropriate for the given sample matrix. Method ETS-8-044.1 "Method of Analysis for the Determination of Perfluorinated Compounds in Water by LC/MS/MS; Direct Injection Analysis" was used for analysis. Analytical results are reported in Table 1 and 11-19 of this report. 6 Data / Sample Retention All remaining sample and associated project data (hardcopy and electronic) will be archived according to 3M Environmental Laboratory standard operating procedures. 7 Signatures Susan T. Wolf, 3M Pmnjcipal Analytical Investigator & Date William K. Reagen, Ph.D., 3M Environmental Technical Director 33 Date The 3M Environmental Laboratory's Quality Assurance Unit has audited the data and report for this This test report shall not be reproduced except in full, without written approval of the 3M Environmental Laboratory. PAGE 17 OF 17