Document aKwD8Ke64R04wzrgDDv2rDzN

STANDARD METHOD FOR ANALYSIS OF ENVIRONMENTAL MATERIALS FOR POLYCHLORINATED BIPHENYLS. 1. Scope . 1.1 These methods cover the determination of the amount and . type of polychlorinated biphenyls (PCB's) in environmental ' materials, specifically air, water and soils and sediments. ' Absolute confirmation of PCB structures is not obtained with these methods. Structure proof can be obtained using additional techniques such as mass spectrometry. 2. Significance .' 2.1 Since it is possible for minute quantities of PCB's to ' find their way into the environment, it is desirable to have a method to identify and measure the quantity _ of such trace amounts. 3. Sampling ' 3.1 Water Sampling 3.1.1 Environmental sampling of natural bodies of water and effluents are covered in many methods under . . the jurisdiction of Committee D-19, Any of the methods of obtaining a representative sample are V 'acceptable for this method as long as the sample . .is at least 250 ml. See Method D-510, ASTM . ' Part 23. . - 3.2 Sediment and Soil Sampling .` 3.2.1 A sediment or soil sample of at least 1000 grams should be taken according `to ASTM Method D-1587, ASTM Part 11. `' DSW 362861 STLCOPCB4098296 n V/ 2 3. Sampling (Cont'd) . 3.3 . Air Sampling 3^3.1 Environmental sampling of ambient air or source . sampling is under the jurisdiction of ASTM Committee D-22. The^e are various methods of >/ cW / ?> L / fa - sampling from stacks or ambient air -as.-to--the- . -nttJce=no zzle--and^girs^olname-mea-s uremen-ts . See Method D-1605, ASTM Part 23. ' . 3.3.2 For the measurements of Airborne PCfi's the sampled air is drawn through one or more fritted bubblers or impingers in cylinders . filled with toluene. The PCB's are absorbed in the toluene. * 4. Principle '. 4.1 The PCB's present in environmental samples are isolated . from the bulk of the sample matrix by solvent extraction. ' Interfering matrix components and endogenous residues ' are then removed from the sample by a suitable combina tion of clean-up procedures and the amount and type of ' . PCB's present determined by the election capture gas . .chromatography (EC/GC) . . 5. Reagents . 5.1 Hexane - Nanograde. 5.2 N,N-Dimethyl Formamide - Certified. (Saturated in a separatory funnel with hexane.) -. 5.3 Ethyl Ether - Anhydrous, AR grade. '. . ' DSW 362862 STLCOPCB4098297 n 3 Reagents (Cont'd) .. 5.4 Acetonitrile - Nanograde. 5.5 Sodium Sulfate - Anhydrous, granular: AR grade. Heat 400C for one hour prior to use. 5.6 Alumina Adsorption - (For chromatographic analysis) 800/200 mesh. Heat at 400C for a minimum period . of 4 hours and deactivate with 5% (w/w) distilled water. Alumina column preparation: Fill a chromatographic column with hexane up to the point where the reservoir joins the column and push a glass wool plug to the bottom with a glass rod. In a 50 ml beaker measure '35 ml of deactivated alumina (30g), and pour this slowly into the column. Tap or vibrate the column to settle the alumina and top the alumina with 2-3 cm of anhydrous sodium sulfate ,t Wash the column with . . 50-100 ml of hexane prior to the addition of the sample. 5.7 Distilled Water Extracted-with hexane to remove . hexane soluble electron capturing impurities. 5.8 .Sulfuric Acid -- Analytical Reagent Grade, SG = 1.84. 5.9 . Potassium Oxalate - Analytical Reagent Grade. 5.10 Potassium Hydroxide Analytical Reagent Grade. 5.11 Ethanol - Formula 2B. 5.12 2.5% (w/v) Alcoholic Potassium Hydroxide - Dissolve 12.5 grams of AR grade KOH in 500 ml of ethanol. .' dSnN**863 . STLCOPCB4098298 4 Reagents (Cont'd) . 5.13 5% (w/v) Aqueous Potassium Oxalate - Dissolve 5 grams of AR grade potassium oxalate in 100 ml of distilled water. 5.14 Toluene - Nanograde. - 5.15 9/1 (v/v) Sulfuric Acid-Water - Carefully add 270 ml of AR grade sulfuric acid to 30 ml of distilled water in a 500 ml iced beaker. . 5.16 PCB Standards - Production Grade - Aroclor 1221, Aroclor 1242, Aroclor 1248, Aroclor 1254 and Aroclor 1260. Apparatus . 6.1 Separatory funnels equipped with ground glass stoppers and Teflon stopcocks: 125, 250, 500, 1000 and 2000 ml capacities. , 6.2 Kunderna-Danish Evaporative Concentrators, 500 ml capacity equipped with 3-ball Snyder columns and graduated 5 ml capacity vials. 6.3 Chromatographic columns, glass, 10" x 20 mm (OD) with .a 5" x 50 mm (OD) reservoir at the top, equipped with . . Teflon stopcocks. . 6.4 Sintered glass filter funnels, 600 ml capacity, 90M. 6.5 Flat bottomed boiling flasks, 125 ml capacity. 6.6 Liebig Condenser, 200 mm in length. 6.7 Hot plates. . 6.8 Water bath. . . .. . .. DSW 362864 STLCOPCB4098299 y y- frV'S'y ^ -3. a a 6. Apparatus ' 6.9 Reciprocating variable speed shaker. 6.10 10 HI Syringes. 6.11 32 oz. all glass mortars and pestles. 6.12 8 x 12 x 2" (2-1/2 qt.) Pyrex baking dishes. 6.13 U.S. Standard Sieve, No. 30. 6.14 Grinding Mills and Grinding Unit. 6.15 Usual laboratory glassware. ' 6.16 Gas Scrubbing Bottled, High Form, ground glass joint, fritted coarse discs, 250 ml capacity. 6.17 Dry Test Meter (0.1 ft^/revolution)* . 6.18 Laboratory Vacuum Pump. 6.19 Rotatory Vacuum Evaporator. 7. Sample Preparation and Extraction 7.1 Water 7.1.1 Where possible, the entire water sample, including the container in which it was collected, is . extracted with hexane.- With larger samples, . where this is not physically possible, the con- - . .tainers are simply agitated and a 250 ml portion . .. used for analysis. . .. 7..1.2 Extraction of water samples - after agitating, . .. transfer the entire aqueous sample or a 250 ml . aliquot into a graduated glass cylinder. Record the volume of the sample and quantitatively transfer it to a separatory funnel with distilled water. , ' DSW 362865 STLCOPCB4098300 6 Sample Preparation and Extraction (Cont'd) 7.1 Water (Cont'd) 7.1.3 Rinse the sample container and graduated cylinder with two 50 ml portions of hexane and add each - . to the separatory funnel. - 7.1.4 Stopper the separatory funnel and shake vigor ously for at least 1 minute. Allow the layers to separate and transfer the lower aqueous phase to a second separatory funnel. 7.1.5 Extract the water sample a second time with a 50 ml portion of hexane. After the layers have separated, add the first hexane extract to the second separatory funnel and transfer the aqueous - layer to the original separatory funnel. 7.1.6 Repeat the extraction with a third 50 ml portion of hexane. Discard the aqueous layer and combine J_______ " . the hexane extracts. . 7.1.7 Filter the combined extracts through a 4" funnel . plugged with glass wool which is covered with . sodium sulfate. Collect the filtrate in a . Kunderna-Danish evaporative concentrator, add . .a small boiling chip, put the Snyder column in . . .. place, and reduce the hexane volume to less than . 5 ml by heating the apparatus in a 80-90C water . . bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED TO A HOOD.) ." ' DSW 362866 STLCOPCB4098301 O 7 Sample Preparation and Extraction (Cont'd) 7.2 Sediment and Soil 7.2.1 Any. excess water is decanted and the entire . sediment or soil sample transferred to a glass - baking dish to air dry at room temperature. The dried material is transferred from the dish into a mortar and pestle and ground. The ground sediment is sieved, remixed, and a 250g portion . taken for analysis. 7.2.2 Weigh 250g (to the nearest O.Olg) into a 16 oz. . narrow neck screw cap (aluminum foil liner) glass bottle. 7.2.3 Moisten the soil with water (10 ml) and add 150 . ml of acetonitrile. Cap the bottle tightly, and mechanically shake for a minimum period of 1 hour. - 7.2.4 Quantitatively transfer the acetonitrile extract into a 600 ml sintered glass filter funnel, containing a 1/4" layer of anhydrous sodium sulfate. Collect the filtrate in a 600 ml beaker . . .(vacuum filtration may be necessary). 7.2.5 After the acetonitrile has completely drained . . into the beaker, wash the bottle twice with 50 . ml portions of acetonitrile, adding each wash . .to the funnel after the previous has completely . . -percolated through the se.diment. DSW 362867 STLCOPCB4098302 8 Sample Preparation and Extraction (Cont'd) 7.2 Sediment and Soil (Cont'd) 7.2.6 Quantitatively transfer the extract to a Kunderna-Danish evaporative concentrator, add a.small boiling chip, put the Snyder column . in place, and reduce the solvent volume to .. less than 5 ml by heating the apparatus in a 80-90C water bath. (CAUTION: SOLVENT VAPORS MUST BE VENTED INTO A HOOD.) 7.2.7 Carefully evaporate the acetonitrile extracts to dryness with the aid of a gentle stream of dry ' . nitrogen. Re-dissolve the residue in maximum of 5 ml of nanograde hexane. . 7.3 Airborne PCB's - 7.3.1 After scrubbing the desired amount of air, record . the metered volume, pressure, and temperature. ... 7.3.2 Quantitatively transfer the scrubbing solvent . ... . . to a round bottom flask and reduce the volume .............to. approximately 2 ml by.rotary vacuum evaporation. .. . 7.3.3 Quantitatively transfer the concentrate to a 30 .... ml beaker with the aid of several small portions . . . of toluene. . . -7.3.4. Inject a fraction of .a.ul of the concentrate into . . the gas chromatograph, to check for interferences . and determine the approximate level of PCB's . present. If no interferences are present dilute SW 362868 STLCOPCB4098303 O9 7. Sample Preparation and Extraction (Cont'd) 7.3 Airborne PCB's (Cont'd) .or concentrate the'sample to a known volume, as determined by the electron capture chromato gram, and proceed with the gas chromatographic .* 7.3.5 analysis. . If interferences are present proceed with the chemical treatment and column chromatographic clean up procedures. . 7.3.6 Transfer the concentrate to a 125 ml extraction flask with the aid of several small portions of .. solvent. 7.3.7 Evaporate the concentrate just to dryness with a gentle stream of dry filtered air and add 25 . . ml of 2.5% alcoholic potassium hydroxide. 7.3.8 Add a boiling chip, put a water condenser in . place, and allow the solution to reflux for 45 .. . minutes. r . 7.3.9 After cooling, transfer the solution to a 250 . ml separation funnel with the aid of 25 ml of '. distilled water. . -' . 7..3.10 Rinse the extraction flask with 25 ml of hexane . and add it to the separatory funnel. . 7.3.11 Stopper the separatory funnel and shake vigorously . for at least 1 minute. Allow the layers to separate and transfer the lower aqueous phase . to a second separatory funnel. # DSW 362869 STLCOPCB4098304 o^ - 10 - Sample Preparation and Extraction (Cont1d) 7.3 Airborne PCB's (Cont'd) 7.3.12 Extract the saponification solution with a second 25 ml portion of hexane. After the layers have - . separated, add the first hexane extract to the . second separatory funnel and transfer the aqueous alcohol layer to the original separatory funnel. . 7.3.13 Repeat the extraction with a third 25 ml portion of hexane. Discard the saponification solution and combine the hexane extracts. . 7.3.14 Carefully add 25 ml of the sulfuric acid solution (9:1 concentrated sulfuric acid/water) to the . hexane extracts. 7.3.15 Stopper the separatory funnel and shake vigorously . for at least one minute. Allow the layers to separate and discard the lower aqueous acid layer. . ~ Repeat this step until the acid layer is colorless. 7.3.16 Wash the hexane with -25 ml portion of water. Discard the water wash. 7.3.17 .Filter the hexane extract through a 4" funnel plugged with glass wool which is.covered with , . . a layer of sodium sulfate into a Kunderna-Danish . . evaporative concentrator. . 7.3.18 Add a small boiling chip, put the Snyder column . in place and reduce the hexane volume to less than 5 ml by heating the apparatus in a B0-90C water bath. " . '' . DSW 362870 STLCOPCB4098305 on - 11 - Sample Preparation and Extraction (Cont'd) 7.3 Airborne PCD's (Cont'd) 7.3.19 After cooling, remove the 5 ml graduated tube and transfer the hexane extract to an alumina . adsorption column washing it in with several 5 . ml portions of hexane. 7.3.20 Carefully add 100 ml of hexane to the column reservoir and collect the total eluent in either a 250 ml volumetric flask or a Kunderna-Danish . evaporative concentrator. 7.3.21 If the column eluent is collected in a volumetric flask, dilute to volume with hexane and proceed with the gas chromatographic analysis. 7.3.22 If the column eluent is collected in a Kunderna- . Danish evaporative concentrator, reduce solvent . . volume, cool, dilute to volume and proceed with . gas chromatographic analysis. Electron Capture Gas Chromatographic Procedure 8.1 Instrument: F&M 402 Biomedial Gas Chromatograph or .. . . equivalent . 8.2 Detector: High Temperature Ni^3 Electron Capture Cell t 6 mm x 6' Glass Column, 4% XE-60 on 80/100 mesh Chromosorb W, HP, AW-DMCS 8.3 Column Temperature: 160-190C . 8.4 Detector Temperature: 300C 8.5 Injection Port Temperature: 195-215C - . . ' .. - DSW 362871 STLCOPCB4098306 Electron Capture Gas Chromatographic Procedure . (Cont'd) - 8.6 Pulse: 150 8.7 Flow Rates: Helium Carrier 60 ml/min Argon-Methane Purge 120 ml/min 8.8 Using EC/GC as the determinative step, inject in duplicate 1-10 ul of each solution into the chromatograph. By . comparison with standard solutions injected, in duplicate, . under the same operating conditions, determine the amount .. . and type of Aroclor using the, individual or total peak . height and area methods. Supporting Procedures 9.1 Saponification with Alcoholic Potassium Hydroxide and Extraction with Sulfuric Acid - -' 9.1.1 Quantitatively transfer the .concentrated extracts . to a 125 ml extraction flask with the aid of several small portions of solvent. ... 9.1.2' After cooling, remove the 5 ml graduated tube and transfer the concentrate of extracts to a 125 ml extraction flask with the aid of several . small portions of solvent. 9.1.3 Evaporate the extract just to dryness with a gentle stream of dry filtered air and add 25 . ml of 2.5% alcoholic potassium hydroxide. 9.1.4 Add a boiling chip,.put a water condenser in . .. place, and allow the solution to reflux for 45 minutes. . . DSW 362872 STLCOPCB4098307 u 13 Supporting Procedures (Cont'd) - 9.1 (Cont'd) . 9.1.5 After cooling, transfer the solution to a 250 ml separatory funnel with the aid of 25 ml of . . distilled water. 9.1.6 Rinse the extraction flask with 25 ml of hexane and add it to the separatory funnel. 9.1.7 Stopper the separatory funnel and shake vigorously for at least 1 minute. Allow the layers to . . separate and transfer the lower aqueous phase . to a second separatory funnel. 9.1.8 Extract the saponification solution with a second . . 25 ml portion of hexane. After the layers have . ' separated, add the first hexane extract to the . second separatory funnel and transfer the aqueous . alcohol layer to the original separatory funnel. 9.1.9 Repeat the extraction with a third 25 ml portion of hexane. Discard the saponification solution . and combine the hexane extracts. . 9.1.10 CAREFULLY add 25 ml of concentrated sulfuric acid, in small portions, to the hexane extracts. .. .9..1.11 Stopper the separatory funnel and shake vigorously for at least one minute. Allow the layers to .. separate and discard the lower aqueous-acid layer. . . Repeat this step until the acid layer is colorless. 9.1.12 Wash the hexane with 25 ml portion of water. Discard the water wash. OSVM 362873 STLCOPCB4098308 14 9. Supporting Procedures (Cont'd) '' 9.1 (Cont'd) . 9.1.13 Filter the hexane extract through a 4" funnel plugged with glass wool which is covered with a layer of sodium sulfate. Collect the filtrate .. .. . in a Kunderna-Danish evaporative concentrator/ add a small boiling chip, put the Snyder column in place, and reduce the hexane volume to less than 5 ml by heating the apparatus in a 80-90C water bath. (CAUTION; SOLVENT VAPORS MUST BE VENTED TO A HOOD.) .' 9.2 Liquid/Solid.Chromatographic Clean-Up With Alumina 9.2.1 After preparation of the alumina column, described . . in the "Reagents" section, quantitatively transfer . the hexane extracts (5 ml or less)- on to the .. ' column with the aid of several 5 ml portions of . hexane (reducing the solvent level to the adsorb . ant level after each-addition of solvent). .. . ,9.2.2 Carefully add 125 ml.of hexane to the column - . reservoir and collect the total eluent in a - . Kunderna-Danish evaporative concentrator. 9.2.3 Add a.small boiling chip to the Kunderna-Danish - evaporative concentrator, put the Snyder column . in place, and reduce the hexane volume to less . 5 ml by heating the apparatus in a 80-90C water bath. (CAUTION; SOLVENT VAPORS MUST BE VENTED TO A HOOD.) -DSW 362874 STLCOPCB4098309 10. Quantitative Determination 10.1 Quantitative determinations employing the electron capture detector are non-stoichiometric measurements made by comparing peak heights or areas for known - . concentrations with those for unknown compositions. _ Four variations of the peak height or area quantifica tion procedures have been employed. Case I EC gas chromatogram of PCB unknown unchanged ... with respect to standard PCB with no evidence of interferences. Case II EC gas chromatogram of PCB unknown altered with respect to standard PCB with no evidence . of interference. Case III EC gas chromatogram of PCB unknown unchanged . with respect to standard PCB with evidence . of interference. .Case IV EC gas chromatogram of PCB unknown altered . with respect to standard PCB with evidence . . of interference. 10.2 The PCB level in a Case I sample can be determined by . -comparison of the height of the major peak in the un . known with'a calibration curve prepared by plotting the . height of the major peak in the electron capture chromato gram of the corresponding standard vs the number of ng of the standard injected. With a Case II sample because . of the alteration, the maximum PCB level is extimated DSW 362875 STLCOPCB4098310 16 10. Quantitative Determination 10.2 by comparison of the total area of the electron capture envelope of the unknown to a calibration .curve prepared by plotting the total area of the electron capture envelope of the most similar standard,vs_the number of ng of the standard injected. Case III and IV samples are initially subjected to the chemical clean up procedure followed by chromatography as outlined on . alumina. If the interferences are removed*by this treat- ment/ the PCB levels are calculated for Case III and IV in the same manner as Case I and II, respectively. If dominant interference(s) is (are) still present, then . .. the height of a different analytical peak, free from . obvious interference, is employed to calculate the PCB .. . level in Case III. With Case IV samples,, the maximum PCB level is estimated as in Case II after correcting . the total area for that of the interfering peak(s). - In all cases, the response of the electron capture must ' be linear for quantitative analysis. , 11. Calculations . . 11.1 Assuming that the calibration curve is prepared by . plotting the electron capture response (peak height . or area) vs nanograms of the PCB standard, injected, . the PCB level present in the air sampled can be . ' ..calculated in the following manner: 3g28 STLCOPCB4098311 - 17 11. Calculations (Cont'd) - . - 11.1 (Cont'd) - ' (ng PCB's from) ' (calib. curve ) x (ml, volume of) x (103 ul) x (10~^ mg) (ul of sample ) (final concentrate) ( ml) ng ) ( injected ) ' = mg of PCB's present in volume of air sampled . (cu ft of air) (sampled from) x ( meter ) . , u* ( cu ft ) X (M^____ ) (103L ) y (760 - mm Hg) (760 ) v * ( 298 ) (273+T1C]) = volume of air sampled in 1 atm and 25C Results are expressed as m PCB's. M3 12. Contamination . 12.1 In determining PCB's by electron capture gas chroma- 'tography, laboratory sources of contamination can be - a major problem. The samples and extracts should never be allowed to come in contact with materials other than glass. Teflon or metal. Laboratory glassware should be thoroughly washed with hot, soapy water, rinsed with distilled water, acetone, and then hexane. All equipment should also be rinsed again with hexane just prior to use and blanks should be frequently carried through all steps of the procedures to insure against the possibility of contamination. SW 362877 STLCOPCB4098312 O 13 - 13. Discussion 13.1 Extraction 13.1.1 The extraction of PCB's from water employing ' ... hexane as the extractant was found to be . quantitative and to be sufficiently simple and rapid for use as a routine procedure. . 13.1.2 The evaluation of this method was based on . spiking water samples with standard acetone . solutions of PCB's. The spiking method consisted . ................. of adding the PCB's in acetone (25-50 ul) to . . 500 ml of tap water in a 32 oz. narrow neck . . screw cap jar. After thoroughly mixing, dup licate 225-250 ml aliquots were taken and . . subjected to the proposed sample preparation . . and work up as outlined. The results were quantified by preparing a calibration curve . .... using standard hexane solutions of the PCB's . .used to spike the water samples. The total . . area of the electron capture chromatogram was .' .......... used to construct the calibration plot. 13.1.3 The average recovery and deviation achieved . . substantiated the applicability of the method . ........... for the quantitative recovery and analysis of ". * PCB's from water at the ppb-ppm level. . DSW 362878 STLCOPCB4098313 o -: - 19 - o ' . 13. Discussion (Cont'd) . 13.2 Electron Capture Gas Chromatography 13.2.1 Columns . 13.2.2 Column performance is the key to effective gas ' chromatographic analysis and as such the choice .of column materials is particularly important. . . . Ideally, the support employed should be inert, . mechanically .strong, and of high surface area. For these reasons, Chromosorb W, HP, AW-D2*1CS was used in all of our work. . 13.2.3 A variety of polar and non-polar liquid phases were investigated. The following columns were found to provide adequate separation, etc., for use in PCB analysis by electron capture: 4% * . (w/w) DC-200, SF-96, OV-17, SE-30, SE-54, XE-60, ' . .. Apiezon L, and 6% QF-1. DC-200 and XE-60 or QF-1 have been found to be the most suitable of these liquid phases. 13.2.4 Another important consideration when working .. with an extremely sensitive detector and . consequently low levels of materials is ... coiumn conditioning. With polar phases _. such as XE-60 and QF-1, we have found that operating a new column overnight at a tempera . ture 25-50C higher than to be used during . analysis results in a more stable column. A . ** . . no-flow conditioning technique is employed to ' .. DSW 362879 STLCOPCB4098314 o 20 13. Discussion (Cont'd) 13.2 Electron Capture Gas Chromatography (Cont'd) condition non-polar columns. The column is purged with carrier gas, heated for 30 minutes at an elevated temperature without carrier ' .flow and then cooled to room temperature. At - . the end of this cycle, the carrier flow is . . resumed and the conditioning is completed as in the case of the polar liquid phase. Two . precautions: During conditioning, the column . should not be connected to the detector and one should not exceed the maximum safe tempera . . . ture of the liquid phase. 13.2.5 Since all liquid substrates bleed to one degree ' . .. or another and columns eventually degrade, we characterize all new columns with two column performance indicators - the number of theoret- . ical plates (N) and a tailing factor (T). p,p'-DDT is employed to check these parameters . .. because it is known to degrade on "poor" * columns. In this manner, we can determine . t. .if the performance of a new column is satis- . factory and when the column performance begins ' . . . to fall off. We consider a column good if the .. . . number of theoretical plates per foot is on the order of 400-500 with tailing factors of . . ... . 1.0-1.3. Calculation of these parameters is DSW 362880 STLCOPCB4098315 21 13. Discussion (Cont'd) .- ' 13.2 Electron Capture Gas Chromatography (Cont'd) shown in the appendix. Additionally, there . should be no significant extraneous peaks . . upon injection of a pure p,p-DDT standard. 13.2.6 Other chromatographic conditions that can be . ' * . ' adjusted are column temperature and flow rates. Although resolution of a mixture increases with decreasing temperature, a temperature . should be chosen that allows the elution of ' - . . all components within a convenient time period. . . The flow rates are optimum for our instrument, column and detector system and, of course, . . should be adjusted if better results can be *. achieved. . 13.2.7 Two gas chromatographic systems have been used . . for PCB analysis - HP Model 402 and 5750. We . find that any system of instrument and column . suitable for chlorinated pesticides is satis .. factory for PCB analysis. -13.2.8 The bulk of analyses in our laboratories was . . . carried out using the systems outlined. The . use of the high temperature Ni*>3 electron ..- capture cell is highly recommended. The . . .ability to operate.at higher'temperatures . .... prevents maintenance problems due to contamina- ....... . tion from high boiling components. * ' DSW 362881 STLCOPCB4098316