JU'pnntetl from AMERICAX INDUSTRIAL HYGIENE ASSOCIATION JOURNAL Volume SI, Scptcmber-OctoDcr, 1970
Teclmiqnes for the Detection., Identification' and
Analysis of Fibers
ROBERT G. KEENAN" and JEREMIAH R. LYNCH Bureau of Occupational Safety and Health. Public Health Sendee, Department of Health,
Education, and Welfare, tOhl Broadway, Cincinnati, Ohio 45202
(Si Applicability and limitations of the techniques used for the detection, identifica tion, and analysis of mineral fibers arc discussed witli particular ,emphasis as to
the problem of evaluating health hazards resulting from exposure to fibers and the discovery of the mechanisms of the diseases involved.
If)LOGICALLY ACTIVE FIBERS have L' 'ong been an important public health prob ern. As early as 1900. asbestos fibers were identified as the causative agent in a chro tic pneumoconiosis (asbestosis) occurring among asbestos textile workers. Later inves tigators noted an excess of cancer occurring among workers exposed to asbestos fibers. A relationship between the known excess of can cer in urban environments and asbestos has. been suggested by the presence of "asbestos" or "f trruginous" bodies in the lungs of a high proportion ol urban dwellers.
Fibrous glass has been given considerable attention as a possible hazard of the general type of asbestos, although current vnanufac-. 'curing practices produce few fibers capable of penetrating into the pulmonary air spaces. A trend toward the production of smaller glass fibers and the emergence of small, respirable, graphite, silicon carbide, and other man made fibers suggests that the. problem of the health effects of fibers will increase. These, effects are of concern, not only to the indus trial population, but also to urban dwellers who may breathe libers from many sources: occupants of buildings with fibrous glass-lined
UVescrit rwit'rets: George !>. Chyum and A^ociatrs, luc., 257i 1 SouthlicUJ Roo-tf, S`ui!JivM. Mu'ltii'.tn -lfin75.
Men lion of Ct'Homv.vi.t*. prinhscis or cmici'rns. <Ioo$ not coiischme rt>dm'l-:*n>rm by jhc* i'nblic llrjtih Si:iau-c.
ducts and others who may unknowingly be come exposed.
To protect the health of the popul nion and avoid penalizing a valuable economic material which is causing no barm, it is. im portant to learn as much as possible shout fibers in materials, in the air, and in the l ociy. Fibers occurring in. the environment must be detected and identified so they can be tr iced to their source. Quantitative estimates of fiber concentration in industrial and -.ommunitv air are needed to establish an; re lationship between disease patterns and to determine the effectiveness of control mea sures. The trace chemical constituents of hese fibers are of importance in uncovering the mechanisms of the diseases resulting Tom exposure.
Sample Types
The methods of analysis used and the re sults that may be expected depend to a great extent on the type of sample available. In general, the choice may be either a large sample which is remote from the site of activ ity and which yields analytical data of doubt ful significance or a minute sample too small for useful analysis found at the site of the biological event. The solution to this quandary may lie in the selective use of information re garding the suspect material from point ol origin to final deposition. For example, it is
HWBUI0002865
,`jna Srj'ianbcr-Octoht-r, 1970
not necessary to identify each fiber seen in a dust sample collected in an asbestos textile plant if the plant uses only chrysotile asbestos; and the ferruginous body seen in the lung of a known asbestos worker may safolv be called an asbestos body although ibis would not be true of a filter in the bine; of a non-occupationailv exposed individual without adequate supporting analytical data. Certain trace metals and polynuclear aromatic hydrocar bons in the raw material which, can. be traced to laboratory prepared respirable fractions and possibly to airborne dust may also be. entering the exposed individual.
Bulk Samples
Large samples of fibrous raw materials, products made from fibers, and settled dust in factories are usually readily available. High volume samples of airborne dust weighing several grams arc also easy to obtain and may be regarded as bulk samples. These samples generally also contain materials other than the fibers being considered and in some cases the fibers constitute such a small percentage that the sample should be treated as an air borne dust exposure sample. Generally, all methods of analysis are appropriate to these samples; the results are difficult to interpret by themselves, since it is not readily apparent which of the many materials present arc like ly to become airborne and penetrate into the deep lung.
Airborne Dust Exposure Samples
Any study of the effect of inhaled fibers must include a consideration of the fibrous dust the worker breathes. Since the airborne dust may vary considerably with location, the air sample should be taken in the worker's breathing zone, even though the practical in strumentation available for the task permits the taking of only small (several milligrams) samples. When only clust capable of pene trating the pulmonary air spaces is to be con sidered, a size selective presampler that passes only the "respirable" dust should be used. This reduces the weight of material collected. The fiber-containing sample will again be a mixture of fibers, materials associated natural ly with the fibers, other process materials, and
airborne dusts having nothing to do with the fibers.
Lung Fiber Samples Single fibers may be isolated from lung
samples on autopsy. These may be free fibers, more or less in their natural states, or they may be coated with an iron-protein layer and. considerably altered. The presence of fibers in the lung suggests that they may have been the cause of an existing disease) although other respirable fibers later removed by some clear ance mechanism, may have been responsible for the damage. A typical airborne asbestos, fiber in a textile plant weighs about U)'10 grant.1 Since most of the fibers deposited in the pulmonary air spaces are smaller than this average size, the most sensitive methods are necessary for identification and gross analysis of fibers found in the lung.
Sample Preparation
Several recent developments in methods of sample preparation enable the analyst to sep arate the fibers from associated materials and to deposit the material in a suitable receptacle for analysis.
Microsieving A fraction of a sample contain ng the res
pirable fibers may be prepared f-om a bulk sample by microsieving. The bulk sample material is dispersed in 95c/c ethyl-alcohol and placed in a 10-micron sieve n an ultra sonic bath. At 15-minute inter sals, sieved portions are removed from the bat ii and more alcohol is added to the sieve. Fcr one gram of the bulk material, the. separation is com plete in about four hours. X-ra) diffraction analysis of sieved chrysorile asbestos has shown no change in the pattern of the sieved frac tion as compared with that of the bulk sam ple. This technique has been used to demon strate that there is a greater concentration of nickel and chromium in the finer particle fraction.2
Low-Temperaturc Dry Ashing A commercial instrument is now available
for the chy ashing of organic matter without volatilizing trace elements. This technique allows the oxidative reactions to proceed in a very gentle manner, avoiding any disturbance of the structure of crystalline substances or
America.^
sintering appears ; without . posited produced through field ana....' about 1 chamber removed {Figure i
The g oxygen rand abou callv erec tile plasr. oxygen r, a free eh A few mi, most of ti the "act.' downstre;. or by col'
The cc er than p-. ]y contri molecule : cally exc; This high largely dc gases as !: which p: oxygen pf
This i i 150CL 4 experimes 5 of lung a 1 changing I; and parti:;
f Dissoluiiei |
A 20- tf
asbestos ni acid is n.f metallic A samples ;; neither oil olite or si to use Inc tion of of the me
HWBUI0002866
-October, 1.970
to do with the
tecl from lung y be free fibers, states, or they otein layer and ;nc.e of fibers in / have been the although other by some cleareen responsible rborne asbestos ;hs about 10`10 :rs deposited in e smaller than nsitive methods ion and gross .e lung.
s in methods of t analyst to sepd materials and .table receptacle
itaining the resed from a bulk re bulk sample Vo ethyl-alcohol eve in an ultra intervals, sieved 2 bath and more . For one gram jaration is comC-ray diffraction bestos has shown the sieved fracjf the bulk samused to demon concentration of ie finer particle
\g is now available
matter without This technique s to proceed in a snv disturbance re substances or
American Industrial- Hygiene Association Journal
589
sintering of ultrafine particles. The technique appears to be promising for the isolation, without alteration, of. mineral substances, de posited in body tissues. Excited oxygen species produced when molecular oxygen is passed through a high-frequency electromagnetic held attack the sample which is maintained at about 1 rnm Hg pressure in an oxidation chamber -from which the exhaust gases are removed bv a mechanical vacuum pump {Figure 1).
The glow region in the RF discharge of oxygen contains 10% to 20% oxygen atoms and about an equivalent amount of electroni cally excited molecular oxygen.3 The rest of the plasma consists of normal ground-state oxygrn molecules, some ionized species, and a fre: electron concentration of 1 X 10n cm."3 A few millimeters away from the glow region, most of the ionic species have disappeared but the '`active" species persist and are carried downstream before they deactivate reactively or bv collision.
The concentration of oxygen atoms is high er than predicted from the rates of the direct ly contributing electron-molecule reactions, mole :ule ion-electron reactions and electroni cally excited molecular-oxygen dissociation. This high concentration of atomic oxygen is largely due to catalytic effects of such foreign gases as hydrogen, nitrogen, or water vapor, which provide the kinetic pathways to atomic oxygen production.
This low-temperature (approximately 1500, dry-ashing method has been used experimentally to oxidize the organic matrix of lung and other biological tissues without changing the crystal lattice of deposited fibers and particulates.
Dissolution of Trace. Metals
A 20- to 30-ntinute digestion of chrysotile asbestos fibers with 5.V hydrochloric or nitric acid is more than adequate to dissolve the metallic constituents of bulk and airborne samples containing this mineral. However, neither of these acids attacks amosite, crocidolite or glass completely, and it is necessary to use hydrolhiorie acid (TIP) for the diges tion of these types ot fibers prior to analysis of the metallic elements. An evaluation of
LTA-6C0
I (jycS
] gk IFLOWRATOR R. F. -i.-Ts t' y1pniCcWalL
(So,
j!
(Dsample
J
uSUPPLY
; (3T pump ll ^ CONTROL
BLEED rV' VALVE M______ ____________
TO
PUMP
ATMOS HERE
Figure 1. Low-temperature dry ashing appa ratus.
Table I
Evaluation of Acid Digestion Methods for 0.5-Graro Portions of Asbestos Minerals
Type of Asbestos Chrysolite
Amosite
Crocidolite
Element
Xi Cr Co Mn
Xi Cr Co Mn Xi C.r Co Mn
Digestion Procccure
KC!
IJF
(*/g element/gm sample)
365 300 197 220
37 30
270 270
None 20 5
910
50 35
16 4,450
None 10
None 55
20 25 10 180
HC1 and HF digestion methods is presented in Table I.
In view of the findings shown in Table I. all samples for heavy metal analyses should be digested with FIF in platinum dishes to ensure a more complete solubilization of the metallic elements contained within the crystal lattice or between the fibrils of the asbestos materials. The routine application of this digestion technique ensures a more complete dissolution of all samples regardless of what is known or not known about their mineral composition.
Polynuclear Aromatic Hydrocarbon Extraction
Cyclohexane extracts of large samples of airborne fibers and dusts, or of bulk asbestoscontaining materials are prepared by the method of Monkman.1 The extraction is con
HWBUI0002867
590 Scplcmhcr-Oclobcr, 1970
ducted overnight in a Soxhlet apparatus with a spectrofluoromo trie grade reagent. The -100 nil extract is concentrated to 1 nil and trans ferred to a silica gel cohmm. (The silica gel is kept activated by storage in an oven at 150C.) The aliphatic components are eluted with 400 ml of iso-octane. The aromatic frac tion is then eluted with 100 ml of benzene. The polycyclic hydrocarbons present in the latter eluatc are then separated by a column of activated 100 to 200 mesh alumina pre pared by drying the alumina at 150G. fol lowed by adding water to provide 1.8% H-O in final alumina bed. For example, if a high activity column of 1.4% H-O is used, the elu tion time, is very long, and if more than 2% I TO is present, the elution is poor. For this separation, the benzene extract is evaporated to dryness with filtered air, and the residue i" dissolved in a small volume of cyclohexane which is then transferred to the alumina column which is 10 cm high by 1.0 cm di ameter. After this column is capped with (\5 cm of deactivated alumina, the hydrocarbcns a:e eluted in turn with cyclohexane. Tie separate eluates are monitored for individual polycyclic hydrocarbons with a recording U V spectrophotometer. .Addition of 1% ether after the decline of the. eluted chrysene aids tiie subsequent elutions of benzo(e)pyrene, hznzo (a) pyrene, and benzo(k) fluoranthene
Mounding of X-ray Diffreclion Samples
The technique of mounting representati ;e portions of dust samples on molecular mem brane filters was described by Talvitie aid Brewer. A known weight of the finely com minuted sample is suspended in a 250 ml MCA volumetric flask with the aid of a wet ting agent. This mixture is then dispersed with a small ultrasonic bath, and an aliquot of the thoroughly agitated suspension contain ing a known fraction of the sample is pipetted out and filtered through a molecular mem brane filter. This technique provides a sample of known weight on the filter. Asbestos-con taining dust samples are amenable to this procedure after a preliminary dry screening of gross particles through a 200-mesh sieve, followed by grinding fibers to lengths of <3.5 microns. The membrane filter contain
ing die sample aliquot is mounted in the x-ray diITractomcler.
Airborne dusts and fibers collected on a membrane filter arc moved by placing the filter in a test tube containing a few drops of a suitable wetting agent and sufficient water to cover n. Immersion of the test tube in the ultrasonic bath for five minutes promotes' the removal of the dust and fiber deposit from the filter. After the original filter is removed from the test tube, the resuspended fibrous dust sample is deposited on a second mem brane filter which is mounted in the x-ray diffractometer.
Sample Rxaminavion
Since fibers are defined by their shape and, generally, some other materials mixed with the fiber have the same chemical and crystal lographic characteristics as the fiber, the only certain way of detecting fibers is by visual examinaion. Furthermore, the only way to insure obtaining an analysis of a single fiber from a mixture of other materials is to visual ly present the fiber to the analytical instru ment. It is also possible to identify some fibers if sufficient fine detail can be resolved.
Light Microscopy
In die early days of counting dust par ticles obtained in impinger samples in asbes tos textile factories, few fibers were seen, and as a result all particles were counted to give an estimate of overall dustiness. The lack of visible fibers in these samples was due in part to the inefficiency of the imp nger in the collection of particles of such sm ill aero dynamic size and in part to the low resolving power of the 10X objective used. As a re sult of recent studies of the asbestos industry, techniques were evolved for detecting many of the airborne fibers previously missed.'0,7 Membrane filters with a pore size of 0.8 micron are known to have surface collection efficiencies close to 100% for particles down to a few hundredths of a micron.. Airborne dust samples collected on these filters may be viewed by high resolution transmitted fight microscopy if the filter can be rendered trans parent or "cleared." Acetone, glyceral triace tate. immersion oil. and other materials have been used to dear membrane filters. How
American, h:
ever, the fib: index of ref the cleared illumination, contrast illutcrferenc.c pnation of tu is capable o: media with )y different sensitivity of necessitates i diethyl oxah clear the fd filter with r cause dispk: rapid solublethe particles chloroform ' acetate (Bid filter materir. dimethyl pk to clear a fii tion (USP). niques, fibertenths of a :
Larger fib sometimes b ing.s If a t in a liquid that is, large with wavck particle mat some waveit persion col-: objective lev, in the back mary imago light deviate The color s. depend on > the index <v of the liqvr person liqu: tion range, i With the p: particles, sm identified, r asbestos, a r< identiricatio: addition ol
HWBUI0002868
October, 1970
sun ted in the
lollected on a >y placing the a few drops of ufficicnC water e. test tube in miles promotes er deposit from ter is removed pended fibrous ; second rnernd in die x-ray
heir shape and, tls mixed with cal and crystal: fiber, the only rs is by visual a only way to f a single fiber ials is to visuallalytic.a! instru-
identify some an be resolved.
iting dust parmples in asbeswere seen, and counted to give ss. Tlie lack of es teas due in he impinger in uch small aerore low resolving used. As a resbestos industry, detecting many ous'ly missed.11,7 ore size of 0.8 trfacc collection particles down licron. Airborne, sc filters may iv. aminitted light rendered iransglyeeral friacer materials leave ne filters. How-
American Industrial Hygiene Association Journal
591
ever, the fiber being sought frequently has an index of refraction very near the index of the cleared filter. Thus, under bright field illumination, the fiber may be invisible. Phase, contrast illumination, which utilizes tire in terference patterns produced by tire combi nation of two out-of-phase columns of light, is capable of revealing objects imbedded in media with an index of refraction only, slight ly different from that of the object. The sensitivity of this method to index differences necessitates the use of glyceral triacetate or a diethyl oxalate-dimethyl phthalate mixture to clear the filter perfectly. Clearance of the filter with either of the liquids alone may cause displacement of the particles by too rapid solution of the filter. To prevent this, the particles are either fixed with perspex in chloroform before clearing with glyceral tri acetate (British practice) or pure membrane filter material is added to the diethyl oxalatedimethyl phthalate mixture before it is used to cli-ar a filter to slow the rate of filter solu tion (USPI-IS method). With these tech niques, fibers whose thickness is only a few tenths of a micron can be seen.
Ltrger fibers (over 1 micron thick) may sometimes be identified by dispersion stain ing. If a transparent particle is immersed in a liquid with a steeper dispersion curve, that is, large variation in index of refraction with wavelength, and if the index of tire particle matches the index of the liquid at some wavelength in the visible range, a dis persion color will be produced. When the objertive lens is equipped with a central stop in the back focal plane of the lens, the pri mary image will be eliminated and only the light deviated by the dispersion will bn seen. The color of the particle thus observed-will depend on the wavelength, of light nt which the index of the particle matches the index of the liquid. A wide variety of high dispe.rson liquids covering the index of refrac tion range from 1.850 to 2.11 are available. With the proper liquid, transparent isotropic particles, such as glass fibers, may be easily identified. For anisotropic minerals, such as asbestos, a mixture of colors is produced and identification becomes more difficult. Tile addition of a polarizer results in the produc
tion of the single dispersion color correspond ing to the index of refraction of the fiber along the axis observed. By rotating the polarizer and noting the "highest" (most red) color observed, the various indices along dif ferent axes may be measured and the particle, identity determined. In this manner asbestos may be differentiated from other minerals al though detennhiation of the type of asbestos is more difficult due to a variation in the in dex of .refraction with composition. By using an objective with a centered stop (centered screening or dark field dispersion), isotropic particles of about 1 micron may be identified.
Electron Microscopy The morphology of certain fibers is suffici
ently distinctive to permit positive identi ication with, enough resolution to reveal fine structures when the fiber is observed. Fo. ex ample, the hollow tube down the middle of a chrysotile asbestos fiber distinguishes it from other forms of asbestos and is clearly visible in a 10,000X electron micrograph. Bulk ma terials may be prepared for electron micros copy in a variety of ways, but the mounting of airborne dust or lung samples is more dif ficult. An electrostatic precipitator has been developed which is capable of depositing the dust on the flat surface of a copper grid pre viously coated with formvar. The effieijney of particle collection by this instrument is not size dependent, and the particles are es sentially undisturbed by the collection process. However, this device is not suitable for breathing zone samples, and a long sample period is often necessary to obtain a heavy enough deposit. Dust collected on membrane filters used in personal breathing zone samples may be transferred directly to coated grids. A small square, slightly larger than the grid is cut from the filter and placed dust side down on a coated grid resting on a support screen in a spot plate. Acetone is added to the plate cavity and allowed to contact the grid and filter from underneath by capillary action. The filter dissolves .through the form var, leaving the. dust lying on the grid. Several treatments with clean acetone are necessary to completely dissolve the filter.
Fibers in lungs must first be located before they can be seen by electron microscopy.
HWBUI0002869
592 Scptcmbcr-Octobcr, 1970
Since the fibers are very sparse, especially in the lungs of people, without' occupational ex posures, Mini since the electron microscope field area is so small, direct, mounting of a king specimen is generally not practical. A better method involves concentrating the lung residue' by ashing or digestion and locating a fiber by means of a light microscope. The fiber is encased in a collodion film and then transferred to the electron microscope grid.
Analytical Melhod
Methods are available to determine either the chemical, or elemental composition or the crystal structure of mixtures oi' materials con taining fibers, pure fiber bundles or single fibers. Their limitations are determined by the amount of sample required and their ability to distinguish the fiber from the as sociated material.
X-Ray Diffraction
Milligram quantities of chrysotile. amosite and crocidolite asbestos may be determined by j--ray diffraction. Both external0 and interntP'12 standard methods for the analysis of these minerals have been developed. In the lattt r case the samples are ground dry in a mix;r mill to an average fiber length of <3.5 microns and passed through a 325-mesh U. S. Standard sieve prior to being mixed with a. prescribed quantity of the interna) standard. Standard samples of the pure asbestos min
eral and the internal standard arc. prepared in a series of known weight, ratios and used to calibrate the method. Quart/ serves as the internal standard for amosite and crocidolite; and aquamarine, for chrysotile. l he ratios of the intensities of the quartz 3.34 A peak to tin* crocidolite 3.11 A peak (or amosite 3.OS Al are plotted against the weight ratios to obtain the standard curve. With aquamarine the 3.28 A peak is used as the internal stan dard.
This method has been used for the quanti tative determination of each of the three as bestos minerals in bulk or settled dust samples. Recoveries over the 1 mg to 10 mg range extends from 85% to 103%. With resp:rable dust samples of very small particle size, some variability in recovery has been noted, and the method is not as reliable.
Atomic Absorption Spectrophotometry
Atomic absorption spectrophotometry is a method for the elemental analysis of samples in solution in an aqueous or organic, solvent. A portion of this solution is aspirated into the flame (Figure 2) where the vaporized sample constituents are irradiated by a lamp con taining a cathode made from the analysis elcmcnt(s). The ground-state atoms cf the element in the vaporized sample absorb the resonance-line radiation, emitted by this same dement in (he hollow-cathode lamp, thereby
1 34
1 . HOLLOW-CATHODE LAMP
2 . ROTATING CHOPPER
3 . VAPORIZING SYSTEM
4 . MONOCHROMATOR 5 . PHOTODETECTOR 6 . ELECTRONICS AND
READOUT
Figure 2. Atomic absorption spectrometer light path.
Amcric.au Iruh:
effecting a der light passing lb eliminator to t: tage of tiiis ro of the analyst element's conadvantages oi are two-fold:
(1) It protivity, for a : it detects den ergy levels. A: below, at vrid 99.99% of th in their gronv-
(2) It is optical inter!: narrow reson ment. (The ; involved in L its ground sta
Atomic ab.: analysis of c> in samples of bulk asbestos
The magi; lee ted in p ; for workers >. SiO, - 2H-0) samples are cate beakers, minutes, dilu rectly. The line provides of 0.01 fig } sensitivity p filter halves field counts : other half o; samples wh-.too great a : line is used, magnesium . quantity of i.
Goinparis diffraction r of results ;; for a limicc-s. a sufficient!; technique, for magnesh ,
HWBUI0002870
tober, 1970
;e prepared is and used el ves ns the crocidolite; 'he ratios of
A peak to nnosite. 3.03 ht ratios to aquamarine tternal stan-
thc quantihe three aslust samples. I mg range :h respirable le size, some noted, and
nelry tometry is a s of samples anic solvent, ited into the rizecl sample t lamp conthe analysis toms of the ; absorb the by this same imp, thereby
>
American Industrial Hygiene Association Journal
593
effecting a decrease in the intensity of the light passing through the instrument's mono chromator to tiie photodeteclor. The percen tage of this absorption by the atomic vapor of the analysis element is a measure of the element's concentration in the sample. The advantages of the atomic absorption method are-two-fold:
(1) It provides a high degree of sensi tivity, for a direct-reading method, because i.t detects dements at their ground-state en ergy levels. At temperatures of 3,000 K, and below, at which flames operate, more than 99.9990 of the atoms of many elements are in their ground state.
(2) It is highly specific, as it eliminates optical interference by operating on the very narrow resonance lines of the analysis ele ment. (The resonance line is the. transition involved in the drop of an excited atom to its ground state in a single jump.)
Atomic absorption has been used in the analysi; of certain metallic elements present in samples of airborne asbestos fibers and of bulk a.bcstcs minerals.13
The magnesium in the airborne dust col lected in personal breathing zone samples for workers exposed, to chrysotile (3 MgO-2 SiOv 2ITO) may be measured. The filter samples are transferred to separate borosilicate beakers, digested with 5N HC1 for 30 minutes, diluted to volume and analyzed di rectly. The magnesium 2852 A resonance iine provides a sensitivity (1% absorption) of 0.01 ,ug Mg per ml of final solution. This sensitivity permits magnesium analysis of filter halves with deposits so small that light field counts of particles may be made on the other half of the same filter. In the case of samples where the use of this line requires too great a dilution, the magnesium 2026 A line is used. In either event, the estimated magnesium concentration is converted to the quantity of chrysotile in the total filter sample.
Comparison of this method with the x-ray diffraction procedure has shown a variation of results no greater than 109c to 15% for a limited number of samples containing a sufficiently large dust deposit for the x-ray technique. The atomic, absorption method for magnesium has permitted ns to run thou
sands of personal samples, including many whose total dust weighed less than 0.1 mg. The application, of course, is limited to ex posures containing only one. source material of the analysis clement; chrysotile asbestos in this case. However, where, only one etiologic agent containing the analysis element is pres ent in an airborne dust, the technique of fers the recognized advantages of high sensi tivity, accuracy, specificity, and speed com mon to all atomic absorption methods. The. error of the method is < rh 5fo as deter mined from the analysis of known samples in our laboratory.
Atomic absorption has also been used to determine several heavy metals in bulk sam ples of chrysotile, amosite, and crocidolite. the asbestos minerals most used in this cot ntvy- For this purpose, 0.25 gm of a bulk material is digested with hydrofluoric acid in a platinum dish for three to four hours, f, procedure which involves a double evapora tion to dryness. The final residue is dissoh ed in 1:1 redistilled hydrochloric acid and double distilled water and then filtered through a Whatman No. 42 paper. After the filter is washed, the solution is concentrated and then finally diluted to 25 ml. A onehalf portion of this solution is reserved; the remainder is diluted, as required, for the individual analyses. Ammonium chloride is added to provide a concentration of 20 mg/ ml to reduce the interference of iron with chromium. Iron suppresses the chromium response: for example, 200 /rg Fe will sup press the response of 5 ,ug Cr to that of < 3 /xg Cr. Some elements found in asbestos samples arc listed in Table II along with their sensitivities (amounts giving 1% ab sorption) at the stated resonance lines.
Neulron Activation Analysis
This technique is based on the interaction of neutrons with the stable nuclei of chemical elements. These nuclear reactions are similar to ordinary chemical reactions in that they exhibit, a mass change and an energy':of acti vation and they proceed at definite reaction rales which depend upon the experimental conditions. There are several sources of neu tron particles available, but research reactors.
HWBUI0002871
594 Scplcmbcr-Octobsr, 1970
Tabu: I.T
Elements determined by Atomic Absorption Spectrophotometry Applied to HF Digests of Asbestos Minerals
Element
Ck) Cr Fe i'e Mn Ni
Resonance teine, A
:m(>7 3579 3720 24-93 27S5 2320
Scnsteiv/ty,* /My/znl
Absorption)
1 X Seale
3 X Scale
....
0.4 1.0 0.15 0.1 0.4
0.2 0.15
--
-- --
0.15
Table III
Micrograms Element/Gram Asbestos
Element
Ni Cr Co Mn
Atomic Absorption
Method
1690 1090
78 423* 538*
Neutron Activation
Analysis Method
1400 1060
80 noo
*St-par;ue acid teachings.
S'lOcierciincci );i aiuhorA tsbcratoiy.
Figure 3. Electron microprobe schematic. with their high thermal neutron flux, are gen erally used to irradiate samples for trace ele ment analyses. A comparison of the results of neutron activation analysis with those ob tained by atomic absorption spectrophotometry as applied to duplicate portions of a chrysotile sample is given in Table IU. 'Electron Microprobe
The electron microprobe (Figure 3) is an instrument for spot chemical analysis using a finely-focused electron beam to excite char acteristic x-radiation from a small volume at the surface of a solid specimen (.Figure 4). `Die wavelength nnd intensity of the charac teristic x-ray lines are then used to identify the elements present at the target site and to
determine their relative mass concentrations. At the present time, analyses can be per formed for all elements of atomic number 5 and greater.
The commercial electron microprobe usual ly consists of up to three curved crystal, x-ray , monochromators that are used to analyze the characteristic x-ray lines, which have been ex cited by the finely-focused electron bean , im pinging upon a selected spot on the surface of the sample material (Figure 5). The sam ple composition is determined from the wave length and the intensity of these characteristic x-ray lines. The specimen may be moved un der the electron beam for analysis of specific areas located with the aid of an optical mi croscopic viewing system.
The instrument is capable of producing an electron focal spot of the order of 0.5 micron in diameter depending on the beam power and the sample being analyzed. The effective spot obtained with the electron beam is small er when a material of high atomic number is' analyzed than when a. material of low atomic number is analyzed because of the additional penetration of the beam into the low-atomic sample. The use of a low intensity beam will permit the analysis of biological and asbestostype samples.
The electron microprobe can be used to identify metallic impurities located on the sur faces of, or associated within, the lattice of asbestos fibers. When the electron microprobe is applied to fibers obtained from lung sam ples, it has been possible to demonstrate the existence of an arnosite asbestos fiber at the core of an asbestos body taken from the lung of an asbestos worker.14 A pseudo-asbestos body, not otherwise, distinguishable from a true asbestos body, in the lung of a foundry'
American huh-
Dlt: Z'.r
7
Figure 4. Ek source region.
worker was fou at the core.
A ternary-sy used to characi; Fe, Si, and Mg croprobes. Pure certain clearly c: An untreated shift toward the tein coating oi: is removed, the i location as a pw.
The theoretic nique for the 4. in particles are seen, an elemer throughout an tion of a few p: detectable. Ho. of a metal conn fiber bundle ccv
Emission Specif.. Emission spc.
radiations emit; molecular speda thermal or cli radiations lie in and the ultravi. nelic spectrum, analysis make pr, for qualitative purposes. The i used to analyze
HWBUI0002872
October, 1970
Asbestos
Neutron Activation
Analysis Method
1400
1D&S00
1100
:oncen (rations, i can be permic number 5
roprobe usual1 crystal, x-ray to analyze the i have been ex iron beam, imon the surface 5). The sam:rom the wavee characteristic be moved un lysis of specific an optical mi-
f producing art r of 0.5 micron e beam power . The effective ; beam is small>rnic number is l of low atomic ' the additional the low-atomic nsity beam will il and asbestos-
an be used to atecl on the sur, the lattice of ran mic.roprobe rein lung samlemon slrate the tos fiber at the l from the lung pseudo-asbestos ishable from a
of a. foundry'
American Industrial Hygiene Association Journal
Electron beem
595
Figure 5. Electron microprobe analysis of a single particle.
Figure 4, Electron microprobe effective x-ray source region.
Table IV Sensitivity of Electron Microprobe Analyztr
word er was found to have no asbestos fiber at tie core.
A ternary-system phase-diagram may be used to characterize the ratios of the elements Fe, hi, and Mg analyzed by the electron microp -obes. Pure asbestos specimens all lie in certain clearly defined zones on this diagram. An untreated asbestos body, however, will shift toward the iron end due to the iron-pro tein coating on the fiber. When this coating is removed, the. bare fiber is found in the same location as a pure fiber.
The theoretical sensitivities of this tech nique for the determination of trace elements in particles are given in Table IV. As can be seen, an element such as a metal distributed throughout an asbestos fiber in a concentra tion of a few parts per million would not be detectable. However, concentrated particles of a metal compound contained in. or on, a fiber bundle could be identified.
Emission Spectroscopy
Emission spectroscopy is the study of the radiations emitted by atoms, ions, and a few molecular species, after suitable excitation by a thermal or electrical source of energy. These radiations lie in the near infrared, the visible, and the ultraviolet regions of the. electromag netic spectrum. Spectrochcmical methods of analysis make practical use of these radiations for qualitative and quantitative analytical purposes. The emission spectrograph may be used to analyze more than 70 metallic or
Element
Be Ca Co Cr Fc Mn NPb
Mirt. Detect. Min. Detect. Mass, em % in lp Particle Spectrometer
1 X 10-" 8 X 10-"
4 X 10-" 5 X 10-'1
4 X 10-" 5 X 10-" 2X
5 X 10-*
20.0 0.1 0.03 0,03 0.03 0.03 0.0*
0.01
Grating ADPLiF LiF
LiF LiF LiF LiF LiF-/J>P
metal-like elements in any type of solid or liquid sample.
Most spectrographic analyses in the field of industrial hygiene require the sample :o be burned in spectroscopically pure graphite electrodes in a D. C. arc. This source o' ex citation provides the greatest sensitivity for the detection and determination of trace quantities of the metallic elements. Quantita tive methods developed for specific determi nations are based upon the employment of optimal excitation conditions and the use of spectroscopic buffer salts which minimize the production of the background or "continu um" and, at the same time, increase the in tensity of atomic spectra.
Mineral substances may be ground and then analyzed directly by spectrographic pro cedures. However, some of the metallic im purities. in such minerals as chrysotile, amosite or crocidolitc types of asbestos may be quite low in concentration or may not be homo geneously dispersed throughout the fibrous matrix-. In such cases, the analyst will be re quired to digest the mineral by means of the hydrofluoric arid procedure to reduce the
I. i:
HWBUI0002873
5% St'ptiunbcr-Oclober, 1970
Table V Summary of Methods and Th,ir Application
Method
Eight Microscope Utrhi Field (FXiX)
Flip.se Contrast f-JOOX )
Application
Detection and Enumeration
Detection and Enumeration
Characteristic Determined
Total dust Fibers by shape
I)ispvrsio;s St?.ining Electron Microscope X-rav Diffraction
Atomic Absorption
Emission Spectroscopy Electron Microprobe
Identification
Refractive Index
Detection and Identification
Identification and Analysis
Fiber by shape Fine structure
Crystal structure
Analysis '
Elemental com position
Analysis
Detection. Identilication 2nd Analysis
Elemental com position
Elemental com position
Type Sample
Continents
Impmger
Membrane filter and lung residue
Bulk or airborne dust
Bulk, airborne dust and lung Bulk, airborne
dust (>3.0 mg)
Bulk and airborne dust
Bulk
Traditional standard method for evaluating asbestos exposure.
New method ` for tentative as bestos Tf,V. Do not sec fibers smaller than about 0.1 fi di ameter.
Distinguish asbestos from nonasbestos fibers.
May identify fibers with unique fine structure.
Cannot distinguish fibrous Irorn non-fibrous minerals of same crystal structure.
Sensitive, enough for trace ele ments in mg samples and gross composition of count samples.
Gross and trace constituents.
Bulk, airborne dust and lung
May determine gross compt sit (on of selected fibers and grains.
bulk of the mineral diluent and to provide a sol ition of the silica-free residue. The re sulting sample solution is diluted or concen trated, as required, for the analysis of the elements of interest. Sample solutions arc rneas.ired precisely into the craters of graphite clecUodes with micropipets. After the addi tion of an internal standard solution, the elec trode is dried at 105C, cooled and burned in the If. C. arc--just like other solid samples.
The technique is sensitive to approximately 0.01 rnicrograni of cobalt, nickel, vanadium, chromium, and other elements of similar vol atility contained in a 2-milligram charge of the sample residue introduced into the elec trode crater.
Application of Methods
A summary of the methods discussed and their use is given in Table V. The light mi croscope is the easiest and most direct meth od of fiber detection and identification, but only fibers with diameters greater than about 0.1 micron can be seen in this way. It is used in the standard methods for the counting of dust and fibers to determine compliance with the Threshold Limit Values (TLV's). It may also be used to count fillers found in lung resi dues although lunp; samples examined, should not be assumed to be representative since dust
deposits are not uniformly distributed through out the lung. If the entire lung were ashed or digested, a sample taken of a uniform disper sion of the residue, and. an adequate number of fibers observed, the analyst might Lave an accurate estimate, of the number of fibers in the lung.
The electron microscope may be used tr see fibers smaller than those visible under the light microscope and to identify some typos of fibers. Counting fibers is difficult since the field area observed is very small and many plates must be examined to find enough fibers for an accurate count. Lung fibers may also be examined although it is generally necessary to concentrate the lung dust and may be de sirable to first locate the fibers by light micro scope.
X-ray diffraction will identify the presence of certain minerals, except where, strong inter ferences from associated materials are present. This method cannot distinguish between a fibrous mineral and the parent non-fibrous mineral with the same crystal structure. For example, in analyzing for asbestos, the result will include the non-fibrous serpentine with the fibrous chrysotile. This method also ap pears to be somewhat dependent on particle size and has a sensitivity of about 1 mg of the mineral in question, which is not generally
Anurican Jud:.
low enough i pies.
Both the at . spectroscopy tal composition Atomic absqqr. applicable since dont from ini;has been used in airborne du with deposits .also be used ft. fraction, howes and serpentine.
The electron being one of ! analytical instr ment that pent, analysis of a s;; ent it is not set sis of trace c; throughout a in design and : ble.
Conclusion
Virtually all to the deteclk of fibers have problem of ev; ing from expo.of the mecha Frequently, du and the neces;
HWBUI0002874
"ctoberx 1970
tnt*
idrd method for 'n-stos exposure. for tentative asDo not see fibers
about 0.1 n di-
icstos from non-
sbers with unique
uish fibrous from minerals of same ore. th for trace elesomples and gross of count samples. e constituents.
gross composition bers and grains.
uted throughivere ashed or iiform clisper:j.uate number : might have nber of fibers
be used to see Je under the some types of :u!t since the ail and many enough fibers bers may also rally necessary d may be de>y light micro-
y tire presence e. strong inter>!s are present, s'n between a at non-fibrous structsire. For stos. the result .rpentino with etiaod also ap:nt on particle ut 1 nag of the not generally
American Industrial Hygiene Association Journal
591
low enough for airborne dust exposure sam- fibers mixed with other materials, indirect
pies.
Both the atomic absorption and emission spectroscopy methods determine the elemen tal composition of the total material analyzed. Atomic absorption is preferred wherever it is
sources of data have had to be combined to obtain an overall characterization of the fiber. Work is proceeding, however, in several lab oratories to develop new and improved meth ods.
applicable since it is easier and has more free dom from interferences. Atomic absorption References
has been used for the analysis of magnesium in airborne dust samples of chrysotile asbestos with deposits so small that the samples may also be used for counting. As with x-ray dif fract on, however, the total of both chrysotile and serpentine is obtained.
The electron microprobe shows promise of being one of the most sensitive and versatile analytical instruments. It is the only instru ment that permits the selection and elemental analysis of a single bundle of fibers. At pres ent it is not sensitive enough to permit analy sis of trace elements uniformly distributed throughout a fiber, although improvements in design and. technique may make this possi ble.
Conclusion
'
Virtually all available methods applicable to the detection, identification, and analysis of fibers have been brought to bear on the prob em of evaluating health hazards result ing from exposure to fibers and the discovery of the mechanisms of the diseases involved. Frequently, due to limitations in sample size and the necessity of dealing with samples of
1. Lynch, J. R., and H. E. Aver: Measurement of Dust Exposures in the Asbestos Textile Industry. Amer. Ind. Hyg. Assoc. J. 27: 431 (3966).
2. KupL, R. E., R. E. Kinser. and P. A. Maurr: Separa tion and Analysis of the Less Than 10-Micron Fractions of Industrial Dusts. Amer. Ind. Hyg. Assoc. /. 29: 364 (1968).
3. Hou.ahan. J. R-: Topics in Chemical Instrumentation. XXVII. Analytical Applications of Electindrlessly DIs* charged Gases. J. Chcnx. Ed. 43: A40l (1966).
4. Monkman, J. L.: Personal communication (1955).
5. Talvitie, X. A., and L. \V. Brewer: Separation and Analysis of Dust in Lung Tissue. A mcr. Ind. Hyg. Assoc. ]. 23: 58 (1962).
6. British Occupational Hygiene Society, Committ :e on Hygiene Standards: Hygiene Standards for Chx-sotilc Asbestos. Ann. Occup. Iiyg. 11: 2 (1908).
7. F.dw.muw. G. II- 3rid J. R. Lynch: The Method Used by the U. S. Public Health Sen-ice for Enumeraron of Asbestos Dust on Membrane Filters. .Inn. Occup. Hyg. 11: 1 (1908).
8. McCroxr. W. C., R. .G. Drafty, and G. .T- Inaxv: The Particle Atlas Ann Arbor Science Publishers; Inc., Ann Arbor. Michigan (1967).
9. Lin, B. Y. H., and K. T. Whitby: Electrostatic Aero sol Sampler for Light and Electron Microscopy. 2 he Rcviezv of Scientific lust. 3fi( 1): 100 (1907).
10. Ckal'LE, J. V.: Quantitative Determination of Chryrotile, Amositc and Crocidolite bv X-Ray DiiTriction. Amer. 2nd. Hyg. Assoc. }. 27: 293 (3956).
11. Crasle, J. V., and M. J. Know Application of X-Ray Diffraction to the Determination of Chrysotile ir Bulk or Settled Dust Samples. Amer. Ind. Hyg. Assoc. ] 27: 383 (1966).
12. Oracle, J. V., and M. J. Knott: Quantitative X-Ray Diffraction Analysis of Crocidolite and Amositc in Bulk or Settled Dust Samples. -4mcr. Ind. Hyg. Assoc. J. 27: 449 (1966).
13. Cralley, L. J., R. G. K.f.?.nas, and J. R. Lynch: Ex posure to Metals in the Manufacture of Asbestos Textile Products. Amer. Ind. Hyg. Assoc. J. 28: 452 (1967).
14. Lancer, A. M.: Personal communication (1968).
Received October 3, 1969