On 2/12/73 we experienced difficulties in obtaining
consistent and reproducible analytical results on
water determinations of Aroclor samples. We suspected
that our,"electrometric dead stop" instrument was
T
at fault. Our instrument technician made frequent
checks and re-standardizations of the instrument.
The difficulty then would disappear and reoccur at
intervals. A substitute instrument was used with the
same results.
Our next approach was to purchase new Karl Fischer reagent and the benzene and methanol used as a solvent in making our determinations.
We traced the problem to the methanol. We have been purchasing this methanol from the same supplier for several years with no ill effect. It is interesting
to note that whatever contaminate there is in the methanol affects only Aroclor analyses. It does not
affect the standardization or standard (non-Aroclor) water samples. SectionI runs benzene at 10-15 ppm range with no problem.
We have set up a separate job for Aroclor water analyses
using methanol A-^12 (reagent grade). This has eliminated
our problem.
. .
"