Document ZnG7OR2OeoG6D0nDxYJM6XeL
May 30, 1991
Dow U.S.A.
The Dow Chemical Company P O Box 150
Plaquemine, Louisiana 70765 0150
Dean Walters
Cell Service
Building 5601
On May 30, 1991, AT-500 and conveyor In the Cell Service Block, Building 5601, was checked for face velocity. The Alnor Model 8925 Thermo Anemometer was used to take the face velocity measurements. AT-500 was checked at a(n) 24 " sash opening.
The criteria for hood classification, smoke tube observation, and flow measuring devices used in rating hood ventilation performance is attached.
Also included are recommendation(s), observation(s), testing data sheet(s) and a maintenance checklist which should be followed to maintain effective hood ventilation and appropriate classification.
Please respond in writing to Industrial Hygiene within one month as to action taken or planned in regard to these recommendations.
Please feel free to contact me with any questions.
Thank you,
Darlene S. Creel Industrial Hygiene Department Building 3502-W Ext. 6181
DO A 07F8R0 OONFTDFNT T Al
Page 2
Observations:
The laboratory hood designation, the average face velocity (FV) and minimum measurements (FVm) reported in feet per minute (FPM), and its rated classification is as follows:
DesignationSash Hood Opening
AT-500
24 "
(FPM) FV
163
(FPM)
FVm____ Classification
125
The smoke turbulence was good on AT-500.
Cm/EVOR The smoke turbulence was GOOD.
Recommendations:
1. AT-500 and conveyor should be put on C-9 checklist for annual check and routine maintenance done on them
2. Practice good housekeeping inside the AT-500.
no a OONF T DFNT T A1
Page 3
CRITERIA FOR HOOD CLASSIFICATION. SMOKE TUBE OBSERVATION. AND FLOW MEASURING DEVICES
CLASSIFICATIONS:
CLASS "A" (HIGH TOXICITY USE) CARCINOGENIC, TOXIC HIGHLY FLAMMABLE
125-150 FT/MIN AVERAGE FACE VELOCITY 100 FT/MIN MINIMUM READING
CLASS "B" (MODERATE TOXICITY USE) TOXIC OR FLAMMABLE
100 FT/MIN AVERAGE FACE VELOCITY 80 FT/MIN MINIMUM READING
INADEQUATE DOES NOT MEET MINIMUM REQUIREMENTS
SMOKE TUBE OBSERVATIONS:
""GOOD (AIR CAPTURE BY HOOD IS STRONG AND WELL DEFINED WITH MINIMAL INTERNAL TURBULENCE, AND NO OBSERVABLE SMOKE SWIRUNG BACK TO HOOD FACE WHEN PERSON STANDS AT HOOD FACE.)
""MODERATE (AIR CAPTURE BY HOOD IS POSITIVE BUT WITH SOME INTERNAL TURBULENCE OBSERVED APPROACHING THE HOOD FACE. NO SMOKE ESCAPING THE HOOD.)
""POOR (AIR CAPTURE BY HOOD IS WEAK AND/OR NOT WELL DEFINED. TURBULENCE INSIDE THE HOOD AND SMOKE OBSERVED ESCAPING HOOD FACE.)
ELQW MEASURING. DEVICES:
WHEN THE STATIC PRESSURE DEVIATES (-/+) 10% FROM THE LISTED STATIC PRESSURE, ITEM #20 SHOULD BE FOLLOWED ON THE MAINTENANCE CHECKUST ATTACHED TO THIS REPORT. THE STATIC PRESSURE READING WAS ESTABLISHED AT TIME OF PERFORMANCE CHECK AND INDICATED BY A RED UNE ON THE DEVICE. DOCUMENTATION OF AIR FLOW MONITOR DEVICE PERFORMANCE IS TO BE DONE ONCE PER WEEK. INSPECTION/ZEROING/CAUBRATION OF AIR FLOW DEVICE IS TO BE DONE ONCE PER YEAR. THIS DOCUMENTATION IS TO BE PART OF THE INDUSTRIAL HYGIENE RECORDS FOR YOUR PLANT. PLEASE FORWARD DOCUMENTATION TO THE PLANT INDUSTRIAL HYGIENE CONTACT FOR YQUR PLANT.
DO A OONFTDFNTTAt
PAGE 4
LOUISIANA DIVISION INDUSTRIAL HYGIENE HOOD PERFORMANCE TESTING DATA
PLANT: Cell Service HOOD I.D.: AT-500
DATE:
May 30, 1991
TESTED BY: Darlene Creel
CONVEYOR---smoke observation was GOOD.
HOOD FACE VELOCITY READINGS AT A(N) 24" SASH OPENING.
|125
193|
|160
119|
SMOKE TUBE OBSERVATIONS WITH A(N) 7A" SASH OPENING -
****GOOD (AIR CAPTURE BY HOOD IS STRONG AND WELL DEFINED WITH MINIMAL INTERNAL TURBULENCE, AND NO OBSERVABLE SMOKE SWIRLING BACK TO HOOD FACE WHEN PERSON STANDS AT HOOD FACE.)
(FPM)(FPM)
DesignationSash Hood Opening______________________ EY_____EYm___Classification
AT-500
24 "
163 125
Used the Alnor Model 8925 Thermo Anemometer
FV = AVERAGE FACE VELOCITY FVm = MINIMUM MEASUREMENT FPM = FEET PER MINUTE
DO A 023883 CONFIDFNT T Al
I
CONSEQUENCES OF EXPOSURE TO TOXIC AMOUNTS
f
NOISE
ACUTE:
Short exposure to high noise levels may cause ringing in the ears ana/or temporary hearing loss. Potential for aamage
increases as sound pressure levels increase above 90 dBA.
Repeated short-term exposures may cause permanent loss.
CHRONIC: Repeated exposures may destroy hair cells which transmit sound signals to the brain and may cause permanent hearing loss.
Degree of hearing loss determined by sound level (dBA) & exposure time.
Average dBA
% of exposed workers with hearing loss at end of working lifetime
80 85 90 .
0-5 10-15 21-29
Typical sound pressure levels (dBA) Situation
quiet room
window air conditioner
noise requiring you to raise voice
to be heard at 3 feet,
power lawn mower
saws, grinders
sand blasters
CQNFTDFNTT A(
pneumatic chipper or nail driver
DOW CONFIDENTIAL
{CAS 8/89
Page