Document ZOVd9NBqj3vNKMaONko80ov8
BACK TO MAIN
CEntrtz Analytical Laboratories, Inc.
3048 Research Drive State College, PA 16801
www.centrelab.com
(814) 231-8032
Fax: (814) 231-1253 or (814) 231-1580
Analytical Report
Fluorochemical Characterization of Drinking Water Samples Port St. Lucie, FI (W2363)
Centre Analytical Laboratory Report No. 023-007E (Revision 1)
Revision Date 3/20/01
Testing Laboratory Centre Analytical Laboratory, Inc.
3048 Research Drive State College, PA 16801
3M Environmental Laboratory Contact Kent R. Lindstrom
Bldg. 2-3E-09
P.O. Box 33331 St. Paul, MN 55133-3331 Phone: (651) 778-5352
Kris J. Hansen, Ph.D. 3M EnvironmentalTechnology & Safety Services
Bldg. 2-3E-09 P.O. Box 33331 St. Paul, MN 55133-3331
PAGE 1 OF5
BACK TO MAIN
1 Introduction
Results are reported for the analysis of a series of drinking water samples received by Centre Analytical Laboratories, Inc. (Centre) from the 3M Environmental Laboratory. The samples were collected from Port St. Lucie, FI. The Centre study number assigned to the project is 023-007.
Specific fluorochemical characterization by liquid chromatography / tandem mass spectrometry (LC/MS/MS) was requestedfor all samples. A total of 16 samples were receivedfor analysis.
The samples were prepared and analyzed by LC/MS/MS for the following list of fluorochemicals:
0 Table 1:Target Analysis
-
ComDound Name Perfluorooctane Sulfonate Perfluorooctane Sulfonvlamide Perfluorooctanoate
Acronvm PFOS PFOSA POAA
The analytical method used was validated by Centre. The validation protocol and results are on file with Centre. Data presented here is the highest quality data available at this time.
2 Sample Receipt
The samples were submitted in individual plastic containers and were not preserved. Sixteen individual sample containers were received. Samples were received on 02/15/00. The sample collection dates were not supplied. Chain-of-custodyinformationis presentedin Attachment C.
3 Holding Times
The analytical method used was validated against a maximum holding time of 14 days. The stability of the analytes of interest for longer periods has not been determined. However, it should be noted that field fortifications in water and other matrices have shown acceptable recoveries at 100 and 1000 ng/L for periods longer than 14 days.
PAGEZOFS
BACK TO MAIN
4 Methods - Analytical and Preparatory
4.1 LC/MS/MS
4.1.1 Sample Preparation for LC/MS/MS Analysis
Samples were initially treated with 200 UL of 250 mg/L sodium thiosulfate solution to remove residual chlorine. Solid phase extraction (SPE) was used to prepare the samples for LC/MS/MS analysis. A forty-milliliter portion of sample was transferred to a Cla SPE cartridge. The cartridge was first eluted with 5 mL of 40% methanol in water solution. The eluate was discarded and the SPE column was then eluted with 1OOoh methanol. A 5 ml portion of methanol was collected for analysis by LC/MS/MS. This treatment resulted in an eight-fold concentration of the samples prior to analysis.
4.1.2 Sample Analysis by LC/MS/MS
In HPLC, an aliquot of extract is injected and passed through a liquid-phase chromatographic column. Based on the affinity of the analyte for the stationary phase in the column relative to the liquid mobile phase, the analyte is retained for a characteristic amount of time. Following HPLC separation, ES/MS provides a rapid and accurate means for analyzing a wide range of organic compounds, including fluorochemicals. Electrospray is generally operated at relatively mild temperatures; molecules are ionized, fragmented, and detected. Ions characteristic of known fluorochemicals are observed and quantitated against standards.
A Hewlett-PackardHPl100 HPLC system coupledto a Micromass Ultima MS/MS was usedto analyze the sample extracts. Analysis was performed using selected reaction monitoring (SRM). Samples were extracted on 2/25/00 and analyzed by MS/MS on 2/26/00. The HPLC and MS/MS methods used for analysis and instrument parameters can be found in attachment
D.
5 Analysis
5.1 Calibration
A 7-point calibration curve was analyzed at the beginning and end of the analytical sequence for the compounds of interest. The calibration points were prepared at 0, 25, 50, 100, 250, 500, and 1000 ng/L (ppt) The response of the quantitation ion versus the concentration was plotted for each point. Using linear regression with l/x weighting, the slope, y-intercept and
correlation coefficient (r) and coefficient of determination (P) were determined. A calibration curve is acceptable if r 20.985 (?2 0.970).
Calibration standards are prepared using the same SPE procedure used for samples.
Calibration check standards were analyzed periodically (every three to five sample injections) throughout the analysis sequence. Compliance is obtained if the standard analyte concentrations are within +/-20% of the actual value.
For the results reported here, calibration criteria were met.
PAGE 3 OF 5
BACK TO MAIN
5.2 Blanks
Extraction blanks were prepared and analyzed with every extraction batch of samples. The extraction blanks should not have any target analytes present at or above the concentration of the low-level calibration standard. For these samples, the extraction blanks were compliant. Instrument blanks in the form of clean methanol solvent were also analyzed after every highlevel calibration standard, and after known high-level samples. Again, the blanks should not have any target analytes present at or above the low-level calibration standard. For the samples presented here the instrument blanks are compliant.
5.3 Surrogates
Surrogate spikes are not a component of the LC/MS/MS analytical method.
5.4 Matrix Spikes
Matrix spikes were prepared for every sample at a concentration of 100 ngR using all compounds of interest. Matrix spike recoveries are given in Attachment B. Field spikes were also prepared on several samples at a concentration of 100 ng/L using all
compounds of interest. Field spike recoveries are also given in Attachment B. 5.5 Duplicates
All samples were analyzed in duplicate. Results are given along with the sample results in Attachment A.
5.6 Laboratory Control Samples
Milliq water was spiked with all compound of interest at 25 and 250 ngR. All recoveries for all compounds were between 70-130% in each LCS.
5.7 Sample Related Comments
Field blank samples consisted of empty containers. Forty milliliters of type I water filtered through a hypercarb cartridge was added to the empty container and analyzed in the same manner as the other samples.
6 Datasummary
Please see Attachment A for a detailed listing of the analytical results.
7 DatdSample Retention
Samples are disposed of one month after the report is issued unless otherwise specified. All electronic data is archived on retrievable media and hard copy reports are stored in data folders maintained by Centre.
PAGE 4 OF5
BACK TO MAIN
8 Attachments
8.1 Attachment A: Results 8.2 Attachment B: MatrixSpike Recoveries(Fieldand LaboratorySpikes) 8.3 Attachment C: Chain of Custody 8.4 Attachment D: LC/MS/MS Raw Analytical Data
9 Signatures
Kevin J Lloid, Vicd President
Date
Other Lab Members Contributing to Data
Enaksha Wickremesinhe Karen Smith
PAGE 5OF5
a\Laboratories, Inc. CEntrE Analytical
3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253
BACK TO MAIN
Analytical Results W2363 Port St. Lucie, FI
3M Sample Identification
MC-615H MC-617H MC-620H MC-621H MC-623H MC-626H MC-628H MC-631H NA MC-632H NA MC-633H
Sample Description
Intake-PIN Intake-P/N duplicate Intake-FieldBlank Empty Outflow-PIN Outflow-PIN duplicate Site 1 PIN Site 1 PIN duplicate Site 2 P/N Site 2 PIN duplicate Site 3 P/N Site 3 P/N duplicate Field Blank-PIN Empty
PFOS (ng/L)
ND ND ND ND ND ND ND ND ND ND ND ND
PFOSA (ng/L)
ND ND ND ND ND ND ND ND ND ND ND ND
POAA (ng/L)
ND ND ND ND ND ND ND ND ND ND ND ND
Limit of Detection (LOD) for the procedure is appoximately 2.5 ng/L for PFOS and PFOSA and 7.5 ng/L for POAA Limit of Quantitation (LOQ) for the procedure is 25 ng/L for all compounds
ND - Compound not detected NQ - Compound detected at a level between the LOD and LOQ. Result is not quantifiable,
ND < LOD e NQ < LOQ
43
Please refer to the reverse side for our standard terms and conditions.
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID:
I MC-619H 1
Spiked Amount (ng/L): [
100
PFOS PFOSA POAA
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result (ng/L)
87.1 90.4 96.2
Matrix Spike Result
(% Recovery)
87.1 90.4 96.2
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID:
I MC-625H 1
Spiked Amount (ng/L): I
100
I
PFOS PFOSA POAA
Lower Recovery Limit:
Upper Recovery Limit:
Sample Concentration
(ng/L)
0 0 0
I
70
I
130
Matrix Spike Result (ng/L) 83.6 89.5 93.2
I
Matrix Spike Result
(% Recovery)
83.6 89.5 93.2
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID:
MC-630H 1
PFOS PFOSA POAA
Lower Recovery Limit:
Upper Recovery Limit:
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result (ng/L)
87.4
81 .O
98.3
1
70
J
1
130
I
Matrix Spike Result
(% Recovery)
87.4
81 .O
98.3
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L):
MC-631H I
100
PFOS PFOSA POAA
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result (ng/L)
95.0 87.0 104
Lower Recovery Limit:
~
70
I
Upper Recovery Limit: I
130
Matrix Spike Result
(% Recovery)
95.0 87.0 104.0
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment 9: LC/MS/MS Laboratory Spike Recovery
Sample ID:
I MC-632H I
Spiked Amount (ng/L): I
100
1
PFOS PFOSA POAA
Lower Recovery Limit:
Upper Recovery Limit:
Sample Concentration
(ng/L)
0 0 0
1
70
1
130
Matrix Spike Result (ng/L)
94.0 97.1 95.8
Matrix Spike Result
(% Recovery)
94.0 97.1 95.8
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Field Spike Recovery
Sample ID:
Spiked Amount (ng/L): I
MC-618H
100
I
PFOS PFOSA POAA
Sample Concentration
(m-1
0 0 0
Matrix Spike Result (ng/L)
125 128 125
Lower Recovery Limit: I
70
Upper Recovery Limit: I
130
1
Matrix Spike Result
(% Recovery)
125.0 128.0 125.0
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Field Spike Recovery
Sample ID:
I MC-624H
Spiked Amount (ng/L):
100
I
PFOS PFOSA POAA
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result (ng/L)
128 109 126
Lower Recovery Limit: I
70
I
Upper Recovery Limit: I
130
1
Matrix Spike Result
(% Recovery)
128.0 109.0 126.0
-
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment 9: LC/MS/MS Field Spike Recovery
Sample ID:
I MC-629H I
Spiked Amount (ng/L): I
100
1
PFOS PFOSA POAA
Sample Concentration
(WL)
0 0 0
Matrix Spike Result (ng/L)
130 107 126
Matrix Spike Result
(% Recovery)
130.0 107.0 126.0
Criteria (Pass / Fail)
PASS PASS PASS
Upper Recovery Limit: I
130
I
Envii .mental Laboratory
Form 30770 - PWO
Shipping Address: 3M Bldg 2-3E09
Telephone: Sample Recelvlng: (651) 7784948
BACK TO MAIN
1 7133 Chain of Custody/Reque,,,,>r LaboratoryAnalytical . - Project IDlProject Name 023-077 mL,I.+&
Template #
935 Bush Avenue SI Paul, MN 55106
Alternate: (651) 778-6753 FAX: (651) 778-6176
Project Lead )L,,-&~Q
Dept. # (main) I 124/
Internal Due Date
ClasslJoblProject # (9QlQs3sm
3 Contact Name &+3 Company 3;w\'
Date Available -
Po kqL m331
Date Due
3M Env -I Project #
For I n t e t d Use Only
,pedalInstructions andlor Specific Regulaloiy Requlrements: nethod. limit of detedion. repoiling units. etc )
GLU-014
ua3c03
'30ft- 541 b L i q J'L
Item #
Client Sample identification
1. __ - 2. - 3. - 4. - 5.
- 6. - 7.
- 8. - 3.
- 10.
FAX# /&TI)
Contract Lab
I Preservatives:
&& LIMW
Time Sampled Sampled
Matrix/ .
od
Analysis Requested:
Complete below. Attach any associated Information.
SamDle Condition UDon Receipt:
e Acceptable 0 Other:
Shippinu Address: 3M Bldg 2-3E09
..
Telephone: Sample Rccslvlng: (651) 7784948
- Company 3m A
P)'
3 #,- E MailingAddress
g city, State, ZIP
Telepho:<:n,,e,#,f,
- (QOlri
peclal Inslrucllons rndlor Specific Regulatory Requlremenls:
nelhod. limilof deledim. reportingunils. eIc )
DlProject Name f)a2,-cQ7 ' m
~
FAX# (bg)ixc bl7b
BACK TO MAIN
9-5
J
Date Available
Date Due Contract Lab
II I
Analysis Requested:
Item ~~
~
#
Client Sample Identification
3M LIMS#
Date Sampled
Time
Matrix/
E Media
7.
8.
3.
IO.
Collected by (print): A ]
8
I'
3
\I
Relinquished bylAffiliation
.-C
I
U
c m
I 0
I
SamDle Condition UDon Receipt:
0 Acceptable
0 Other:
c
I I
~~
I
I
-57
Date
I
I I
Page
I
Original- Accompanying Sampler
I
Lasl Page - Origmalor
See Reverse Side for lnslruclions