Document YDMyZwjEpODRR5bLJ048or150
r ABD00076804 4
LEAK DETECTION AND ELIMINATION PROGRAM ABERDEEN PVC PLANT ` TABLE OF CONTENTS
A. GENERAL B. VINYL CHLORIDE CONTINUOUS MONITORING SYSTEM C. PORTABLE HYDROCARBON DETECTORS
1. INTRODUCTION 2. OPERATION 3. DETECTION PRINCIPLE AND THEORY D. CONTINUOUS MONITORING SYSTEM - CALIBRATION AND MAINTENANCE PROCEDURES 1. CALIBRATION 2. MAINTENANCE E. PORTABLE HYDROCARBON DETECTOR - CALIBRATION AND MAINTENANCE PROCEDURES 1. CALIBRATION 2. MAINTENANCE F. ROUTINE LEAK PATROL - DATA RECORDING AND HANDLING G. LOCATION OF CONTINUOUS MONITORING POINTS H. ACTION PLAN FOR LEAK ELIMINATION I. DEFINITION OF A LEAK
ABD00076805
LEAK DETECTION AND ELIMINATION PROGRAM ABERDEEN PVC PLANT ` LIST OF TABLES
TABLE 1 TABLE 2 TABLE 3 TABLE 4 TABLE 5 TABLE 6 TABLE 7 TABLE 8 TABLE 9 TABLE 10 TABLE 11 TABLE 12 TABLE 13 TABLE 14 TABLE 15 TABLE 16 TABLE 17
ANALYZER SECTION, OPERATING CONTROLS AND INDICATORS MONITOR MODULE, 'OPERATING CONTROLS AND INDICATORS SEQUENCER MODULE, OPERATING CONTROLS AND INDICATORS AMPLIFIER MODULE, OPERATING CONTROLS AND INDICATORS COMPONENT MODULE, OPERATING CONTROLS AND INDICATORS VALVE MODULE, OPERATING CONTROLS AND INDICATORS DUAL PREAMP MODULE, OPERATING CONTROLS AND INDICATORS SPECIFICATIONS FOR MODEL PI-101 PHOTOIONIZATION ANALYZER BRIEF DESCRIPTION OF INSTRUMENT CONTROLS AND FUNCTIONS VERIFICATION OF ELECTRONIC ZERO FOR PHOTOIONIZATION ANALYZER RELATIVE PHOTOIONIZATION SENSITIVITIES FOR VARIOUS GASES SOME DERIVATIVES OF OLEFINS RELATIVE SENSIVITIES FOR VARIOUS GASES SPECTRUM ANALYSIS TROUBLESHOOTING CHART OVEN HEATING SYSTEM TROUBLESHOOTING CHART LOCATION OF CONTINUOUS MONITORING POINTS ACTION LEVELS FOR LEAK DETECTION REPORTING AND ELIMINATION PROGRAM
ABD00076806
LEAK DETECTION AND ELIMINATION PROGRAM ' ABERDEEN PVC PLANT LIST OF FIGURES
FIGURE 1 FIGURE 2 FIGURE 3 FIGURE 4 FIGURE 5 FIGURE 6 FIGURE 7 FIGURE 8 FIGURE 9 FIGURE 10
CONTROL PANEL FUNCTIONS OF PHOTOIONIZATION ANALYZER
TIME RESPONSE FOR THE PHOTOIONIZATION ANALYZER TYPICAL CALIBRATION CURVE FOR PHOTOIONIZATION ANALYZER BLOCK DIAGRAM OF PORTABLE PHOTOIONIZATION ANALYZER ELECTRICAL BLOCK DIAGRAM OF PHOTOIONIZATION ANALYZER CALIBRATION PROCEDURE FOR PHOTOTONIZATION ANALYZER COMPONENT PARTS OF PROBE COMPONENT PARTS OF LAMP HOUSING COMPONENT PARTS OF READOUT UNIT POWER SUPPLY PC BOARD
ABD00076807
LEAK DETECTION AND ELIMINATION PROGRAM Aberdeen pvc plant
A. GENERAL The emission standard for vinyl chloride,requires the plant to have a formal leak detection program. The Federal Register describes specific criteria which must be met within the program. This document describes the leak detection program. Each of the specific requirements will be met. Namely, this section describes: The continuous VCM detection system which is used by the plant. The portable hydrocarbon detectors which are used by the plant. The calibration and maintenance procedures for the above equipment. /The Leak Patrol Survey and procedure when a leak is found by the V portable hydrocarbon detector. _^/The location of the continuous monitoring points which are used for leak detection. The criteria which will be used to identify the presence of a leak. The action plan which will be used to eliminate a leak after it has been detected.
A
ABD00076808
i
i
LEAK DETECTION AND ELIMINATION PROGRAM ABERDEEN PVC PLANT
4
B. VINYL CHLORIDE CONTINUOUS MONITORING SYSTEM
Introduction
9
The concentration of vinyl chloride in the atmosphere is continuously monitored by four flame ionization chromatographs, each.of which is operated continuously. Each chromatograph is a Honeywell Model 1000 Process Gas Chromatograph manufactured by:
Honeywell Systems, Inc. 6625 McGrew Street Houston, Texas
Each chromatograph sequentially measures, on a continuous basis, the vinyl chloride content in air samples taken from ten different locations. A total of forty (40) locations are monitored throughout the plant continuously and the concentrations at each point are recorded.
Description
Analytical Technique
The Flame Ionization Detector used in the Honeywell 1000 Process Gas Chromatographs has a lower detection limit of 0.1 part per million.
The special-purpose elements of the FID chromatograph are a flame cell, an FID Electrometer Assembly, a Temperature Control/Inverter Assembly, and a plug-in FID Amplifier Module which mounts in the Control Section.
The flame cell consists of an ignitor wire, a hydrogen-fed burner, and an electron collector. Within the cell, hydrogen combustion occurs in an electrical field created by the voltage from the ionization inverter. Because of the carbon-free property of hydrogen, ionization occurs only when a carbon-bearing sample is injected into the flame. Ionization causes current to flow from the collector in direct proportion to the amount of the hydrocarbon elements present in the sample. The- output current from the flame cell is routed through the Electrometer (Preamp) in the Analyzer, and .is. .applied.to the input of the FID Amplifier in the Control Section. The gain is made available for external recording.
The FID Electrometer Assembly is physically divided into two major subassemblies: the Power Supply-Electrometer P.C. Card Assembly; and the Electrometer P.C. Card Assembly. The electrometer power supply accepts +_ 20 volt inputs from the Control Section power supply, and produces regulated plus and minus 15 volt outputs for operation of the electrometer circuits and the inverter located on the Temperature Control/Inverter P.C. Card Assembly. The inverter output is a -100 volt ionization potential which is applied to the base (burner) of the flame cell.
ABD00076809
\
I
i
t
SEQUENCE AND ANALYZER CONTROL
MODULE
ANALYZER SECTION
RECORDER
SAMPLE POINT
ABD00076810
Depressing the IGNITE pushbutton switch applies 115 vac to the primary of a transformer, which in turn produces approximately 2.2 vac across the ignitor coil inside the flame cell. Depressing the IGNITE pushbutton also produces a ground connection which enables the solenoid driver circuit to.open the hydrogen shut-off valve. . With 2.2 vac applied to the ignitor coil, and the hydrogen shut-off valve open, the flame will ignite within five to 10 seconds.
When the flame ignites, current flow from the flame cell collector is amplified by the electrometer circuit, the output of'which is routed to both the Control Section FID Amplifier and to a level detect circuit on the Power Supply-Electrometer P.C. Card Assembly. When the level detect circuit senses an electrometer output of sufficient magnitude to indicate that the flame is lit, the circuit produces a logic level output which lights the FID Amplifier "FLAME ON" indicator. In addition, the circuit de-energizes a relay which extinguishes the analyzer "FLAME OUT" indicator, and provides a ground to the solenoid driver circuit so that the hydrogen shut-off valve will be maintained in the open position when the IGNITE pushbutton is released.
During sustained operation, the level detect circuit continuously monitors the output of the electrometer circuit, and any momentary output drop below a selectable threshold level will cause the FID Amplifier FLAME ON indicator to momentarily extinguish. If the reduced signal level continues beyond a selectable time delay, the level detect relay will energize, thus closing the hydrogen shut-off valve and illuminating the analyzer FLAME OUT indicator. If this condition occurs, the IGNITE pushbutton must again be pressed and held to re-light the flame.
Electrometer P.C. Card Assembly
The functional center of the electrometer circuit is a 3450J high gain integrated circuit amplifier. The flame cell collector current is brought into the circuit via resistor R1 and the divider network at the amplifier input. HIGH/LOW range selection is accomplished by a RANGE control input from the FID Amplifier. When L0 range.is selected, relay K1 is energized and R7 is placed in parallel with R8 to reduce the over-all sensitivity of the circuit.
The signal OUTPUT at pin 21 is routed to the FID Amplifier input in the Control Section, and to the level detect circuit on the Power SupplyElectrometer P.C. Card Assembly.
Power Supply - Electrometer P.C. Card Assembly
The Power Supply - Electrometer P.C. Card Assembly contains both the electrometer power supply and the level detect circuitry associated with monitoring the status of the hydrogen flame. The power supply (IC1, Ql, Q2, etc.) accepts plus 20 vdc and minus 20 vdc from the control section, and produces regulated plus and minus 15 vdc for the electrometer circuits and for the ionization inverter on the Temperature Control/Inverter P.C. Card Assembly.
ABD00076811
The level detect: circuit' consists of the operational amplifiers in IC-2; the -LEVEL DET ADJUST control (R7); the associated decoding logic in lC*s 3, 4, 5,; and the level detect relay, Kl. The electrometer output is brought into the circuit at pin 21, and is routed through operational .amplifiers 1C 2-1 and IC 2-7 to the input of IC 2-10.
The level at IC 2-9 is set by R7, so that if the electrometer output drops below a selected threshold, the flame will be considered extinguished and a positive transition will occur out of IC 2-8. In practice, the threshold level is set just below the "quiescent electrometer output when only air and hydrogen are passing through the detector.
If the output of IC 2-8 goes positive, signifying an inadequate signal level, the logic-0 at IC 4-6 illuminates DS2 on the electrometer front panel, and the logic-1 at IC 4-8 extinguishes the FID Amplifier FLAME ON indicator. In addition, any positive level Out of 1C 2-8 will also trigger adjustable one-shot IC 3, which produces a selectable time delay before energizing relay Kl and illuminating DS1 on the electrometer front panel. This delay, which is adjustable from 2 to 20 seconds, and optionally from 2 to 200 seconds, ensures that the relay will not be energized by transient conditions in the flame cell or electrometer.
If the IC 2-8 output is still positive when the selected delay is completed, the resultant logic-1 levels at IC 5-1 and IC 5-2 produce a logic-0 at IC 5-6 which illuminates DS1 and energizes Kl. In the energized condition, the closed contacts of Kl illuminate the analyzer "FLAME OUT" indicator and the open contacts remove the ground to the solenoid driver circuit, thus closing the hydrogen shut-off valve.
Temperature Control/Inverter Assembly
The temperature control circuit in the FID chromatograph is identical to that of the TC system.
The inverter circuit is of conventional design, and operates from the plus 15 vdc regulated input provided by the electrometer power supply. Transistors Q3 and Q4 and the associated elements form a free-running oscillator, the output of which is coupled through transformer T1 to produce plus and minus 100 vdc at the output. The -100 vdc is the ionization potential which is applied to the base of the flame cell.
Amplifier Module
The FID Amplifier is essentially the same as the TC Amplifier, except for the addition of a HI/LO gain range switch concentrically mounted around the MANUAL RANGE switch, and the addition of a FLAME ON indicator to the front panel. The HI/LO range switch controls relay Kl in the FID electrometer, to change the divider resistance values at the electrometer input, thus changing the electrometer sensitivity.
The FLAME ON indicator is controlled by a logic level input from the Power Supply - Electrometer P.C. Card Assembly ini the electrometer. No time delay is provided between the level detect circuitry and the FLAME ON indicator, thus ensuring that any transient "flame off" conditions will be seen on the indicator.
ABD00076812
<
Sample Analysis and Specific Chromatograph Parameters
Sample analysis is accomplished through a dual column with a backflush attachment. This column configuration insures that the analysis is specific for vinyl chloride. Separation and analysis is accomplished in the follow ing manner:
The sample is injected with the dual column valve energized (light on), and with the backflush valve in the foreflush position. This allows the sample to be swept from the sample valve into column number one. Column number one separates inerts or other light components from the heavier components, but not from each other. The combined light com ponents are allowed to pass into the second column where they are separated.
At a selected time during the analysis, column number two is switched out of the stream and the heavy components from column number one are routed to the detector via a restrictor. The light components, not yet eluted from column number two, remain in that column until the dual column valve is switched back to the dual column position. Then they are eluted through the detector. The resultant reverse flow through column number one causes those components still- in the column to be eluted as a group.
The specific parameters for the Honeywell chromatographic column are as follows:
analysis cycle time: 60 sec. carrier gas: nitrogen carrier pressure: 50 psig column vent flow: 35 cc/rain. backflush vent flow: 45 cc/min. heater air: 50 psig oven temperature: 60C sample loop length: 48 in. sample loop I.D.: 0.043 in. column: 1/16" column material: SS packing: 15% bis(2-ethoxyethyl) adipate on chromosorb P
Parameters for the air and hydrogen are as follows:
hydrogen flow equals 40 cc/min. @ 50 psig air flow equals 350 cc/min. @ 220 psig
ABD00076813
t
4j
The Analyzer Section of the Model 1000 has been designed to permit the high-speed analyses require by the technological advances of the past few years.
The basis of this high-speed analytical system is the Micro Packed Column-High-Speed Micro Detector combination.
The resolution of the Micro Packed Column is enhanced by the High-Speed Micro Detector designed and manufactured by Honeywell with the following features: the cell volume is approximately 0.05 microliter; a semi-diffusion type of detector cell is utilized rendering it re latively insensitive to flow rate; the detector response to the presence of a component is suitable for analysis times of less than 1 second.
The Micro Packed Column, Micro Detector and associated sample and column-switching valves have been packaged in an analyzer designed to yield the high thermal stability and access ibility.
The temperature of the analyzer oven is controlled by an all solid state, propor tional temperature controller. Several modifications to the widely used heated air type of temperature control system have resulted in an analyzer design that permits the full sensitivity of the Micro Detector to be utilized.
The use of each control- of the analyzer is given in Table 1.
,^
... .h
Zina?
_; ^
ABD00076814
TABLE 1 ANALYZER SECTION, OPERATING CONTROLS AND INDICATORS
Control or Device AIR Control AIR Gauge CARRIER Control
CARRIER Gauge REE and AUX 1,2,3 Needle Valves Heater Air Shut-off Valve
Remote Valve Switches and Indicators
Preamplifier Meter and BAL ADJ Control Preamplifier Test Jacks Temperature Set Point Control
Temperature Control Indicator
Purpose or Use
Controls air pressure to the valve actuating solenoid(s) and to the low mass heater.
Indicates air pressure (0-100 lbs.) set by the AIR control.
Controls carrier pressure to the column(s) (ultimately controlling the flow) and to the needle valves for reference, etc.
Indicates the carrier pressure (0-160 lbs.) set by the CARRIER control.
Used to precisely adjust carrier flow rates through the reference side of the detector and through three auxiliary valves.
Allows the heater air to be shut off, while air pressure is maintained on the airactuated valves.
Used to remotely actuate the Analyzer valves. Light emitting diodes are illuminated to indicate the correspond ing' air-control solenoid is energized.
% Used to balances the preamplifier output.
Meter/scope test points for monitoring the indicated preamplifier parameters.
Used to set the oven temperature. Clock wise rotation increases the temperature, and counterclockwise decreases the temper ature .
Indicates the'status of the-oven temperature controller. Due to the proportional temp erature control circuitry, and the overtemp switch on the heater barrel, the light will alternately glow and shut off, until the oven temperature begins to approach the ambient operating temperature. When the operating temperature has been reached, the indicator will remain on, but will be very dimly lighted.
ABD00076815
TABLE 1 (Cont *.) ANALYZER SECTION, OPERATING CONTROLS AND INDICATORS
Control or Device
Heater Barrel OverTemp Control
Purpose or Use
Located behind the brass plug in the front cover of the heater housing, this control permits adjustment of the temperature cut off point of the overtemp switch. This temperature can be monitored by connecting a pyrometer to the I/C thermocouple wire coiled under the explosion-proof housing. This temperature is normally factory adjusted for about 300C.
ABD00076816
Sample Injection Procedure
.41
In this.multi-stream application, the sample system permits continuous purge from the stream switching system, to and through the analyzer. This insures that the sample being analyzed is from the proper stream.
Sample inptit is measured by a "gas" valve. The "gas" valve is an externalvolume sample valve, in which the sample volume is a length of tubing external to the valve. This valve is used to inject gas samples, and is also the valve used for column switching.
t
The column configuration by which sample injection is accomplished is the dual column backflush configuration. Normally, a sample is injected with the dual column valve energized and with the backflush valve in the foreflush position. This allows the sample to be swept from the sample valve into column number one. Column number one separates inerts or other light components from the heavier components, but not from each other. The combined light components are allowed to pass into the second column where they are separated.
At a selected time during the analysis, column number two is switched out of the stream, and the heavy components from column number one are routed to the detector via the restrictor. The light components, not yet eluted from column number two, remain in that column until the dual column valve is switched back to the dual column position. The resultant reverse flow through column number one causes those components still in the column to be eluted as a group, and recorded as one peak. Note that the combined components bypass column number two via the column bypass restrictor.
The sample volume is 1346 pi. The injection valve specifications are as follows:
Sample valve: Material: SS Diaphgram: High-Temperature Part Number: 30740492-002
Backflush: Material: SS Diaphragm: High-Temperature Part Number: 30740492-002
Sample Handling Equipment
There are 40 sampling points at the Aberdeen Plant. At each sampling point, sample is drawn into a 1/8" I.D. nylon sampling line through a M.S.A. high efficiency small volume .paper filter. The model number of these filters is DZ-78006. These end of line filters are manufactured by:
Mine Safety Appliance Company 408 Penn Center Blvd. Pittsburgh, PA
ABD00076817
Samples are drawn through each of the sampling locations served by a single Honeywell unit, by a single vacuum pump. The vacuum pump is model number 289-EXP, with trademark "Dia-Vac", manufactured by:
4
Combination Control, Inc. Kulpsville, PA 19443
Sample is drawn into the sampling line at a maximum rate of 1.5 cfm at 0 in. Hg vacuum. Sample inflow is continuous.
The Control Section of the Model 1000 consists of solid state, plug-in modules. Each module is self-contained to permit easy replacement for troubleshooting or maintenance. The modules require front access only for replacement. The contacts are gold-plated over nickel for resistance to corrosive atmospheres.
The timing system is a 240 KHz clock which generates a digital pulse each second.
The anlaysis is programmed as follows:
A digital pulse is chosen just prior to setting the exact time the occurrence of the function is desired (Coarse). An adjustable delay, triggered by the digital pulse, is then set to provide the exact (Fine) time of the function. In the case of setting a component time-span (gate), the digital pulse triggers an adjustable time-window. Gate duration ad justments are direct-setting.
Auxiliary test equipment is not required to make any of the adjustments of the Model 1000 Control Section. Test points, with front panel access, are provided on each module for maintenance trouble-shooting.
The basic programmer (Control Section) is composed of the following sub-systems:
The MONITOR MODULE contains a multiple point selector switch and readout meter to permit checking auto zero range, power supply voltages, detector preamp output, signal amplifier output and digital clock status. The memory output can also be checked with this switch, if this option is included in the basic control section. The Power switch is located on this module, as well as a Status switch (Normal/Bypass) and a Standard Sample switch for remote introduction if a calibration standard is desired. A recorder jack is included on the front panel to permit the use of a service recorder, if a dedicated recorder is not required for bargraph readout. An auto zero function is also included in this module.
THE SEQUENCER MODULE is a solid state, digital timing system with direct setting time adjustments for the sequenced funtions. The adjustments for the sample valve ''Inject Time" are located in this module. The basic sequencer is a 0-100 second clock with a digital pulse generated each second.
THE AMPLIFIER MODULE contains an auto zero status switch, manual range (Spectrum) switch, amplifier zero potentiometer, and a built-in calibration. The calibrator generates synthetic signals which may be used to set, or check, the full-scale range of the measured components. This module gives the basic 0-10 volt output signal.
ABD00076818
THE VALVE MO.DULE is required for multi-column applications. Directsetting switches are adjusted to provide the exact time for the signal required for the valve-switch to occur. An additional auto-zero circuit, and its timing adjustment is provided in-this module.
A COMPONENT MODULE is required for each component to be measured. The basic control section can be supplied with one to six component modules Each module contains the necessary adjustments for setting the fullscale range of the measured component and for setting the time-span (gate). An auto zero is an integral part of the module.
The standard Control Section provides bargraph output signals for each measured component. A trend output of peak height or peak area (integration) for each component is also available. The necessary interface requirements for computer-input of data can be provided. Current or voltage output is available as a standard feature.
The signal from the detector bridge is amplified to a level of approximately 0.1 volt before leaving the analyzer section. Thus no low level signals are switched within the Model 1000 system. Since the electronic auto zero circuit performs the zeroing function in less than fifty (50) milliseconds, it is normally actuated by each component "on gate".
The specific functions of each control option are given in Tables 2 through 7.
ABD00076819
TABLE 2
MONITOR MODULE, OPERATING CONTROLS AND INDICATORS
4) .
Control or Indicator ON/OFF Switch PWR (Power) Indicator DET/OFF Switch DET (Detector) Indicator STD SAMP/OFF Switch
STD SAMP Indicator STATUS NORM/BYPASS .Switch
STATUS Indicator Meter Display
Meter Function Selector Switch
+20 and -20 Positions
SIG IN Position
Purpose or Use
Controls AC input power to the Control Section. .
Indicates the power on/off status of the Control Section.
Controls DC power to the bridge detector circuit in the Analyzer.
Indicates the power on/off status of the bridge detector.
Used to actuate a solenoid in the sample system (if provided) for automatic standard sample introduction. In the STD SAMP position, a contact closure is provided at the Control Section rear panel.
Lamp is illuminated when sample is flowing, and is off for normal operation.
Provides a contact closure which can be used as a maintenance "out of service" switch, etc. Terminal connections are provided on the Control Section rear panel for connection to remote indicators, alarms, or a computer.
Lamp is illuminated to indicate "bypass", and is off for normal operation.
Provides a visual indication of the Control Section operating conditions, as sel.ecte.d by the Meter Function Switch.
Selects the Control Section parameters to be presented on the Meter Display.
Display the respective output levels from the Control Section DC power supply. Normal indication is approximately 50% of fullscale deflection, in the respective polarity direction.
Displays the signal amplifier input level received from the preamplifier in the Analyzer. (+0.1 volt range).
Control or Indicator SIG OUT Posi*tion AUTO ZERO Position CLOCK Position
Cl thru C6 Positions
RECORDER Connector
AUTO ZERO START Switches
ABD00076820
*
TABLE ,2"1' (Cont.-)
V
Purpose or Use
<41 j
Displays.the output level of the signal amplifier. (10 volt range).
Displays the voltage required to bring the "baseline" back to zero. (+5 volt range).
Permits confirmation that the Control Section internal clock is functioning. Meter display will indicate 1/2-scale or better, and will drop to zero one second out of every 10 seconds.
Display the peak height voltages stored in the memory outputs for components 1 through 6. Can also be used for a rough setting of zero and span calibration of each memory.
Provides AC power and chart advance levels to drive a portable "service" recorder for temporary spectra or gate readout. Back panel terminals are also available for permanent connections.
A non-dedicated auto zero function which can be set to auto zero at any point during the.analysis by dialing in the time (in seconds) on the thumbwheel switches. The fine adjustment allows delay of the start of Auto Zero from 0 to 2 seconds.
ABD00076821
TABLE 3
-
\--
SEQUENCER MODULE OPERATING CONTROLS AND INDICATORS
->
Control or Indicator SAMPLE VALVE Indicator MODE Selector Switch AUTO Position
GATE Position
SPEC (Spectrum) Position
MAN (Manual) Position
Purpose or Use
Lamp is illuminated when the valve is in the sample inject'position.
Controls the operation of the Sequencer, and the readout to the recorder.
Analysis events (sample inject, auto zero, component gating, etc.) occur automatically as programmed by Control Section front panel switches. The recorder is driven in the bargraph mode, and the amplifier output is enabled only during the selected "component gate" times.
Analysis events (sample inject, auto zero, component gating, etc.) occur automatically as programmed by Control Section front panel switches. The recorder chart is continuously driven, and the amplifier output is enabled only during the selected "component gate" times. This allows the programming gates to be visually checked.
Valve actuations occur automatically, as programmed by Control Section front panel switches. The recorder chart is driven continuously, and the amplifier output is continuously enabled. COMPONENT lights are gated to provide a visual indication of gate timing.
NOTE The Amplifier module AUTO ZERO switch should be in the OFF position. The AUTO/MAN switch must be set to MAN to select the desired signal range.
All functions must be operated manually. (Valve switches left in AUTO will cause the corresponding valves to go to the "on" condition). The recorder chart is continuously driven, and the amplifier output is continuously enabled.
NOTE The Amplifier module AUTO/MAN switch must be set to MAN to select the desired signal range.
ABD00076822
Control or Indicator CYCLE Selector Switch
'Time (lower) Position
"100" Position
"600" Position
Sample Valve AUTO/MAN Switch
AUTO Position
Middle (off) Position
MAN Position SEQ. (sequence) RESET Pushbutton
TABLE 3 (Cont.)
Purpose or Use
Permits selection of analysis time (length)
Analysis time (in seconds) is determined by the setting of the TIME thumbwheel switches (0-99 seconds).
Analysis cycle time is a fixed 100 seconds, regardless of the TIME switch setting. The analysis will run in the time programmed, and will wait for the remainder of the 100 seconds before starting another analysis.
Analysis cycle time is a fixed 600 seconds, regardless of the TIME switch setting. The analysis will run in the time programmed, and will wait for the remainder of the 600 seconds before starting aftother analysis.
Permits selection of automatic or manual sample valve control.
Sample valve is automatically controlled by the Sequencer module with "sampling length" selected by the VALVE DURATION thumbwheel switches.
Sample valve is in the "bypass" position, and the sample is flowing through the sample volume ports.
Sample valve is in the "on" condition, and the sample is being injected into the column.
Allows the Sequencer to be manually reset to the start of an analysis.
NOTE The SEQ. RESET pushbutton does not reset any auxiliary valves. Therefore, before pushing the SEQ. RESET button, wait for the VALVE lights to return to their initial "program start" conditions, or switch the sample valve off until a complete cycle runs on the sequencer. If any VALVE light is on, then one complete cycle must be run before a satisfactory readout can be obtained.
If the CYCLE switch is in the "600" position, SEQ. RESET advances the timer to 100 seconds, and an additional 500 seconds is required be fore the timer is reset.
ABD00076823
Control or* Indicator
Sample VALVE DURATION Thumbwheel Switches
TABLE 3 (Cont.)
-4
/
Purpose or Use
Permits adjustment of the duration of the sample inject period. At the end of the set duration time (in seconds), the sample valve goes to the off (sample purge) position.
ABD00076824
TABLE 4 AMPLIFIER MODULE, OPERATING CONTROLS AND INDICATORS
Control or Indicator RANGE CALIBRATE Control AUTO/CAL Switch
AUTO/ZERO/OFF Switch
AUTO ZERO Position Middle Position Off Position AUTO/MAN Switch
Purpose or Use
Permits adjustment of an internally-generated calibration voltage so that the full-scale range and peak memory output level for each measured component can be set.
In the AUTO position, which is the normal operational position of the switch, the input to the amplifier is the bridge preamp output from the analyzer. In the CAL position, the input to the amplifier is an internallygenerated calibration voltage, the level of which is controlled by the RANGE CALIBRATE Potentiometer.
Controls the operational mode of the Auto Zero circuitry, which automatically compen sates for any drift in the bridge preamp "zero" (baseline) output level. Auto Zero is normally actuated for 50 milliseconds by each component Mon-gateM, and at other times as determined by the setting of the nondedicated Auto Zero function(s). A moveable jumper on each Component Module allows Auto Zero to be inhibited at the corresponding component gate opening, if so desired.
Auto Zero level is automatically updated at each Auto Zero time to compensate for preamp baseline drift.
Allows the last Auto Zero signal level to be held in the Auto Zero circuitry. This leaves the baseline at recorder zero when switching to-Manual or Spectrum operations.
The Auto Zero function is inhibited with Auto Zero "off". The AMPLIFIER ZERO control must be used to zero (balance) the output to the recorder.
In the AUTO position, the amplifier gain changes during each component gate, as determined by the setting of each component's RANGE potentiometer. In the MAN position, the amplifier gain is controlled exclusively by the setting of the MANUAL RANGE switch.
ABD00076825
Control or Indicator MANUAL RANGE Selector Switch
AMPLIFIER ZERO Control GND. IN and OUT Test Jacks
TABLE 4 (Cont .,,)
Purpose or Use
With the AUTO/MAN switch in the MAN position, this switch permits manual selection of the signal output range by attenuating or ampli fying the-fcridge preamp output as follows:
Position 1 2 5 10 20 100 200 500 1000
Factor 10 5 2 1 X2 X10 X20 X50` X100
Permits manual adjustments of amplifier "zero" output, to balance the recorder.
Meter/scope test points for monitoring the amplifier input and output levels.
ABD00076826
' TABLE 5 COMPONENT MODULE, OPERATING CONTROLS AND INDICATORS
4 -
Control or*Indicator COMPONENT Indicator RANGE Control
MEMORY ZERO AND SPAN Controls AUTO/OFF/CAL Switch AUTO Position
OFF Position
CAL Position
RESET Pushbutton GATE Adjustments COARSE Thumbwheel Switches
Purpose or Use
Lighted during the time the associated component gate is open.
Permits full-scale range (gain) adjust ment for the associated component. The resultant gain setting is in effect only when the associated component gate is open, and then only if the Amplifier Module AUTO/MAN Switch is in the AUTO position.
Used in conjunction with the Amplifier Module calibration voltage to calibrate the memory output signal (if used).
Controls the operational status of the respective Component Module.
This position enables the normal oper ational mode of the module, in which the gate circuitry, the bargraph output, and the memory output are operational.
In this position, the bargraph output and all gate functions are inoperative, and memory holds the last value and is allowed to update.
Used in conjunction with the AUTO/CAL switch on the Amplifier Module to calibrate the Component Module RANGE and MEMORY ZERO AND SPAN.
Use.d in conjunction with the memory calibration to reset the memory output to the baseline.
Permit setting the starting time and duration of the associated component gateJ
Permit selection of gate "start" time in one-second increments from 0 to 99 seconds following sample inject.
Control or Indicator FINE Control
DURATION Control GND and OUT Jacks
ABD00076827
A
TABLE 5 (Cont.)
Purpose or Use
Permits the gate "start" time to be delayed from 0 to.2 seconds beyond the setting of the COARSE thumbwheel switches.
Permits vernier selection of gate duration from 0 to 10 seconds.
Meter/scope test points for externally monitoring the memory output.
ABD00076828
#
9
TABLE 6
...
VALVE MODULE, OPERATING CONTROLS AND INDICATORS
Control or Indicator ON Indicator AUTO/OFF/MAN Switch
AUTO Position
OFF Position MAN Position AUTO ZERO Indicator
AUTO ZERO ON/TIME Thumbwheel Switches
Valve ON/TIME Thumbwheel Switches
VALVE Control
Valve OFF/TIME Thumbwheel Switches
VALVE OFF Control
Purpose or Use
Indicates power is applied to the actuating solenoid
Controls the operational mode of the associated valve.
In AUTO, the valve is actuated automatically at the times set on the ON-OFF/TIME thumbwheel switches.
The valve solenoid is de-energized.
The valve solenoid is energized.
Indicates occurance of the respective Auto Zero function.
Permits selection of an Auto Zero "start" time, in one-second increments, from 0-99 seconds following sample inject.
Permit selection of the ON (solenoid energized) starting time in one-second increments, from 0-99 seconds following sample inject.
Permits the valve.ON starting time to be delayed from 0 to 2 seconds beyond the setting of the valve ON/TIME thumbwheel switches.
Permits selection of the OFF (solenoid de energized) starting time, in one-second incre ments, from 0-99 seconds following sample inject.
Permits the valve OFF starting time to be delayed from 0 to 2 seconds beyond the setting of the valve OFF/TIME thumbwheel switches.
ABD00076829
TABLE 7 DUAL PREAMP MODULE OPERATING CONTROLS AND INDICATORS
'4
Control or Indicator Toggle Switch
Balance Meter
DET. BAL. Controls Preamplifier Test Jacks Terminal Strips
Purpose or Use
$ The toggle switch selects one or the other of two detector bridge preamplifier circuits to be monitored on the balance meter.
A 1--0--1 volt meter which is used in balancing the selected detector- bridge/preamplifier circuit.
Adjusts the balance of the respective detector bridge/preamplifier circuit.
Meter/scope test points for monitoring the indicated preamplifier parameters.
Provide connections for respective detector, leads.
ABDQ0076830
' LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN PVC PLANT
PORTABLE HYDROCARBON DETECTORS
1. Introduction
The Aberdeen PVC Plant uses portable hydrocarbon detectors for routine leak patrols, to precisely locate leak sources when leaks are indicated by the continuous monitoring system and to monitor vessels prior to entry. The leak detection program submitted in May, 1977 included Century Organic Vapor Analyzers. The plant has experienced problems with main tenance of these analyzers. Because of these problems, the plant has field tested another portable hydrocarbon analyzer. This analyzer is:
HNU Photoionization Analyzer Model PI-101
The HNU portable analyzer is manufactured by:
HNU Systems 383 Elliot Street Newton Upper Falls, Mass. 02164
>
The plant has found the HNU portable hydrocarbon detector is more reliable, easier to maintain, and less subject to damage than the Century Organic Vapor Analyzer. The plant therefore wants to switch to using the HNU analyzer as a part of the leak detection and elimination program. The plant asks EPA to review the following information on the HNU Photoioni zation Analyzer and approve its use as part of the leak detection program.
The model PI-101 has been designed to measure the concentration of trace gases in many industrial or plant atmospheres. The analyzer employs the principle of photoionization for detection. This process is termed photoionization since the absorption of ultraviolet light (a photon) by a molecule leads to ionization via:
RH + hv --> RH+ + e"
Where RH = trace gas
hv a photon with an > Ionization Potential of RH
The sensor consists of a sealed utlraviolet light source that emits photons, which are energetic enough to ionize many trace species (particularly or ganics) but do not ionize the major components of air such as 02, N2, CO, CO2, or H2O. A chamber adjacent to the ultraviolet source contains a pair of electrodes. When a positive potential is applied to one electrode, the field created drives any ions, formed by absorption of UV light, to the collector electrode where the current (proportional to concentration) is measured.
ABD00076831
To minimize adsorption of various sample gases, the ion chamber is made of an inert fluorocarbon material, is located at the sampling point, and a rapid flow of sample gas is maintained through the small ion-chamber volume.
The analyzer will operate either from a rechargeable battery for more than 10 hours or continuously from the AC battery charger. A solid state amplifier board in the probe and a removable power supply board in the readout module enable rapid servicing of the unit in the field.
The useful range of the instrument is from a fraction of a ppm to about 2,000 ppm. For measurement at levels above 2,000 ppm, dilution of the sample stream with clean air is recommended. Some typical specifications for the model PI 101 Photoionization Analyzer are given in Table 8.
Operation
Turn the function switch to the battery check position. The needle on the meter should read within or above the green battery arc on the scaleplate. If the needle is in the lower portion of the battery arc, the instrument should be recharged prior to making any measurements. If red LED comes on, the battery should be recharged.
Next, turn the function switch to the on position. In this position the UV light source should be on. Look into the end of the probe to see the purple glow of the lamp.
A brief description of the instrument controls and functions is shown in Figure 1.
To zero the instrument, turn the function switch to the standby position and rotate the zero potentiometer until the meter reads zero. Clockwise rotation of the zero potentiometer produces an upscale deflection while counterclockwise rotation yields a downscale deflection. Note: no zero gas is needed, since this is an electronic zero adjustment (see below). If the span adjustment setting is changed after the zero is set, the zero should be rechecked and adjusted, if necessary. Wait 15 or 20 seconds to ensure that the zero reading is stable. If necessary, readjust the zero.
The instrument is now ready for calibration or measurement- by switch ing the function switch to the proper measurement range. The instrument is supplied calibrated to read directly in ppm (v/v) 0-20, 0-200, 0-2000 of benzene with the span position set at 9.8. For additional sensitivity, the span potentiometer is turned counterclock wise (smaller numbers) . to increase the gain. By changing the span setting from 10.0 to 1,0 the sensitivity is increased approximately ten fold. Then, the 0-20, 0-200, and 0-2000 ppm scales become 0-2, 0-20, and 0-200 ppm full scale, respectively. This span control is also utilized to make the instrument scale read directly in ppm of
ABD00076832
the compound being measure. E.g., it is adjusted to match the value of a calibration gas to that same reading on the instrument scale. The*span control can be utilized to calibrate nearly any compound, measured by photoionization, to be direct reading on the 0-20 ppm range. For example, gain settings of 4.5 or 8.9, respectively, will provide direct reading capability (0-20, 0-200 ppm) for vinyl chloride and trichloroethylene, respectively.. For a listing of approximate gain setting values see Table 11.
A small DC operated fan is used to pull air through the photoioniza tion sensor at a flow rate of three to seven hundred cubic centimeters per minute (ca. 0.5 1pm). The fan provides nearly instantaneous response times (Figure 2) while consuming little power. The characteristics of a fan are such that it connot tolerate a significant pressure drop without affecting the flow rate and therefore either the instrument reading or response time. Since photoionization is essentially a nondestructive technique, changes in flow rate do not affect the signal but if a large pressure drop is imposed at the inlet the probe, the sample may not reach the sensor.
The instrument was designed to measure trace gases over a concentration range from less than 1 ppm to 2000 ppm. Higher levels of various gases (to percentage range) can be measured but the recommended procedure is to dilute the sample with clean air to a concentration of less than 500 ppm. This is generally within the linear range of the instrument and if the measured concentration is multiplied by the dilution ratio the correct concentration in the stream can be determined. A typical calibration curve is shown in Figure 5. Note that the calibration curve for benzene (the photoionization standard) is linear (over more than three decades) up to about 600 ppm (v/v).
If the probe is held close to AC power lines or power transformers, an error may be observed. For measurements made in close proximity to such items, their effect on measurements can be determined by the following procedure. Zero the instrument in an electrically quite area, in the standby position, then move the instrument to the questionable area involved. If AC pickup is going to be a problem, the meter (in the standby position) will indicate the magnitude of the error.
The instrument is equipped with an automatic solid state battery protection circuit. When the battery voltage drops below~ll volts, this circuit will automatically turn off the power to the instrument. This .prevents deep discharging of the battery and considerably extends the battery life. If the instrument is unintentionally left on over night, the battery will be unharmed because of the battery protection circuit. If the instrument battery check reads low and the lamp doesn't fire, plug the charger into the instrument. The power to the analyzer should then be returned.
ABD00076833
/
To charge the battery, place the mini phone plug into the jacket on left side of the bezel prior to plugging charger into 120 VAC. When disconnecting charger, remove from 120 VAC. before removing mini phone plug. The battery is completely recharged overnight (ca. 14 hours). To ensure that the charger is functioning, turn the function switch to the battery check position, place phone plug into jack and plug charger into AC outlet. The meter should-go upscale if charger is working and is correctly inserted into the jack.
The instrument can be operated during the recharge cycle. This will lengthen the time required to completely recharge the instrument battery.
Detection Principle and Theory
The detection principle of this portable instrument is photoionization. A wide variety of organic compounds and some inorganic compounds can be measured with this technique. Photoionization (with 10 eV photons) applied to the analysis of trace gases in air can eliminate fragment ion formation (signals) from the major components of air yet still allow the ionization of many impurities of interest in industrial atmospheres. This is demonstrated by the listing of ionization potentials* in Table 12. Note the high (12 eV) ionization potentials for the major components of air. In addition, the choice of a suffi ciently low ionization energy often permits the selective ionization of one or two components in a complex gas mixture.
While the ionization potential serves as a rough guide to whether or not a response is obtained, it does not predict what the quantitive response actually is. In some cases, a species with an ionization potential 10.3 or 10.4 eV will give a response. In these cases, however, the response is usually low because of its low ionization effeciency at 10 eV. A partial list of actual relative sensitivities obtained with a photoionization analyzer is given in Table 13. The use of the tables should allow a determination of the specificity of the instrument in a given application on many industrial processes; this instrument may not respond to the starting materials or by products but will respond to a product. An example of this is seen in the vinyl chloride monomer plants where neither ethylene or dichloroethane is detected but vinyl chloride is detected.
A block diagram of-the major components of the photoionization analyzer is shown in Figure 4. The instrument is separated into two units interconnected by multiconductor electrical cable. The probe contains a fan for moving the air into the sensor, the ultraviolet lamp which is ignited by applying a DC voltage between the anode and cathode, the ionization chamber which contains a pair of electrodes and is adjacent to the lamp, and a signal amplifier. The photons (~10 eV) which are emitted from the lamp pass through a UV radiation by a
* Ionization potential is defined as the energy required to move an electron an infinite distance from the nucleus or more simply, the
- energy required to produce a positive ion and an electron.
ABD00076834
molecule which has an ionization potential of 10 eV or less will lead to ion formation via:
RH + hv RH+ + e-
k positively biased high voltage electrode is used to push any ions formed by absorption of UV to the cpllector electrode where the current (proportional to concentration) is measured. This current is then converted to a proportional voltage by the amplifier in the probe. An electrical diagram of the instrument is depicted in Figure 5. The amplifier is gain stabilized by negative feedback and provides a voltage source output to drive the analog meter readout as well as the gain control network. The sensitivity of the instrument is controlled by changing the loop gain of the amplifier. A 12 volt battery provides the primary power for a high efficiency DC-DC converter which supplies the various potentials required for instrument operation.
ABD00076835
TABLE 8 SPECIFICATIONS FOR MODEL PI 101
PHOTOIONIZATION ANALYZER .
performance (benzene referred)
range 0.1 to 2000 ppm detection limit 0.1 ppm sensitivity (max) 0-2 ppm FSD over 100 division meter scale repeatability + 1% of FSD linear range 0.1 to 600 ppm useful range 0.1 to 2000 ppm response time 4 3 sec to 90% of full scale ambient humidity to 95% RH operating temperature ambient to 40C*
physical
size:
probe 6.3 DIA x 28.5L (cm) readout 21W x 13D x 16.5H (cm) stowed 21W x 13D x 24H (cm) cable 80 cm long (32")
(2-1/2 x 11-1/4") (8-1/4 x 5-3/16 x 6-1/2") (8-1/4 x 5-3/16 x 9-1/2")
weight:
probe .55 kg (20 ounces) readout 3.2 kg (7 pounds) total (shipping) 5.4 kg (12 pounds)
controls and functions
mode switch Off, Battery Check, Standby (zero), 0-2000, 0-200, 0-20 ppm low battery indicator light zero (10 turn +_ 300% FDS max)
span (10 turn counting dial 1.0 to 10 times nominal sensitivity) readout 4-1/2" (11.3 cm) meter Taut Band movement graduated 0-5-10-15-20,
divisions signal output for recorder 0-(-5V) FSD power output for recorder 12 VDC - jack on side of instrument
power requirements of operating times
continuous use, battery>10 hours continuous use. with HNU recorder reduces instrument battery operating -time
to 1/2 normal time recharge time, max< 14 hours, 3 hours to 90% of full charge recharge current, max .4 Amps @ 15 VDC
ABD00076836
TABLE 8 - (Cont.)
construction Designed to withstand the shock and abuse to which portable instruments are often subjected. The readout is housed in a two piece aluminum case, and finished with a solvent resistant baked acrylic textured paint. The probe is fabricated from extruded aluminum sections and machined plastic.
serviceability The probe and readout are of a modular design allowing rapid servicing and/or replacement of mechanical and electrical components. All module interwiring includes quick disconnects.
maintenance The instrument contains only one moving part, and consumes no gases or reagents. The only routine maintenance procedure is cleaning the light source window every several weeks.
calibration check Check instrument calibration at least once per week with HNU calibration standard to ensure that the high sensitivity of the instrument is maintained*
* Instrument is temperature compensated so that a 20C change in temperature corresponds to a change in reading of<+ 2% full scale at maximum sensitivity.
ABD00076837 TABCE 9
{
BRIEF DESCRIPTION OF INSTRUMENT CONTROLS AND FUNCTIONS*
Control Six Position Switch
Zero Potentiometer Span Potentiometer
Function
OFF - Shuts off all power and removes DC voltages.
ON - In any other function position or measuring mode, the electronics are on.
BATTERY CHECK - Indicates the condition of the battery. If needle position is in lower portion of green battery arc, the instru ment should be recharged.
STANDBY - UV lamp is off but electronics are on. This position will conserve power and extern the useful operating time between recharges of the battery. This position is also uti lized to adjust the electronic zero.
RANGES - 0-20, 0-200, 0-2000 direct reading ranges available at minimum gain for benzene. More sensitivity is available by adjusting the span potentiometer.
A ten turn potentiometer is employed to adjust the zero electronically when the instrument is placed in the standby position with the probe attached. This eliminates the need for a hydro carbon free gas.
A ten turn counting potentiometer is utilized for upscale setting of the meter on calibration gas. Counter-clockwise rotation increases the sensiti vity ( 10 times). This pot can increase the sensitivity to make the instrument direct reading for nearly any gas which the instrument responds to.*
* For position of layout controls see Figure 1.
ABD00076838 t
Figure I Control Panel Functions of Photoionization Analyzer
(
Battery Check Position
12 Pin Interface Connector between readout unit and serjsor.
ABD00076839
TABLE 10
VERIFICATION OF ELECTRONIC ZERO FOR
PHOTOIONIZATION ANALYZER* *
Sample
Instrument Reading (ppm)
Room Air
0.7
Room Air Passed Through 6" x 3/4" OD Charcoal Scrubber
0.1
Zero Air
0.25
Zero Air Passed Through 6" x 3/4" OD Charcoal Scrubber
0.04
* Maximum Gain = 2 ppm full scale.
% of F.S. 35 5
12.5 2
ABD00076840
TABLE 11 RELATIVE PHOTOIONIZATION SENSITIVITIES*
FOR VARIOUS GASES
Chemical Grouping
Aromatic Aliphatic Amine Chlorinated Unsaturated
Carbonyl Unsaturated
Sulfide
Paraffin (C5-C7) Ammonia Paraffin (Cj-C^)
Relative Sensitivity
10.0 10,0 5-9
5-7 3-5
3-5
1-3 0.3
0
Examples
Benzene, Toluene, Styrene Diethylamine Vinyl Chloride, Vinylidene
Chloride, Trichloroethylene MEK, MIBK, Acetone, Cyclohexene Acrolein, Propylene, Cyclohexene,
Allyl Alcohol Hydrogen Sulfide, Methyl
Mercaptan Pentane, Hexane, Heptane
Methane, Ethane...
* Sensitivities in ppm (v/v).
ABD00076841
TABLE 12 SOME DERIVATIVES OF OLEFINS
Molecule
vinyl chloride cis-dichloroethylene trans-dichloroethylene trichloroethylene tetrachloroethylene vinyl bromide 1.2-dibromoethylene tribromoethylene 3-chloropropene 2.3- dichloropropene 1-bromopropene 3-bromopropene CF3CCI=CCICF3 n-CsFnCF=CF2
acrolein crotonaldehyde mesityl oxide vinyl methyl ether allyl alcohol vinyl acetate
*
'
IP (eV)
9,995 9,65 9.66 9,45 9.32 9,80 9.45 9.27 10.04 9.82 9.30 9.7 10.36 10.48
10.10 9.73 9.08 8.93 9.67 9.19
ABD00076842
TABLE 13
A`
RELATIVE SENSITIVITIES FOR VARIOUS GASES
(10.2 eV LAMP)
Species
Photoionization y Sensitivity*
p-xylene m-xylene
j 11.4 11.2
benzene toluene diethyl sulfide diethyl amine styrene trichloroethylene carbon disulfide isobutylene
10.0 (reference standard) 10.0 10.0
9.9 9.7 8.9 7.1 7.0
acetone tetrahydrofuran
6.3 6.0
methyl ethyl ketone methyl isobutyl ketone cyclohexanone
5.7 5.7 5.1
naptha (86% aromatics)
5.0
vinyl chloride
5.0
methyl isocyanate iodine methyl mercaptan dimethyl sulfide
4.5 4.5 4.3 4.3
allyl alcohol
4.2
ABD00076843 Figure 2 Time Response for the Photoionization Analyzer.
Instrument Reading
1 Time (seconds)
T 5
ABD00076844 Figure 3 Typical Calibration Curve for Photoionization Analyzer.
Instrument Reading
ABD00076845 READOUT UNIT
Figure 4 Block Diagram of Portable Photoionization Analyzer.
ABD00076846 Figure 5 Electrical Block Diagram of Photoionization Analyzer
12V READOUT ASSEMBLY
ABD00076847
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN PVC PLANT
CONTINUOUS MONITORING SYSTEM - CALIBRATION AND MAINTENANCE PROCEDURES
1. Calibration:
a. General
The Honeywell continuous monitors are each calibrated and spanned once per week. This is more often than is necessary to retain accuracy based on our experience with the units and based upon our conversations with the equipment manufacturers.
b. Calibration Procedure
1. Pre-calibration equipment checks (information as noted is re corded on "Daily Checklist" at the end of this section). a. Visually check strip chart recorder for proper operation. b. Visually check the control panels for the following settings: 1. Monitor panel "ON". 2. Sequence panel in "AUTO" setting. 3. Amplifier panel in "AUTO" setting. 4. Dual valve panel in "AUTO" setting. 5. Dual component panel in "AUTO" setting. 6. Stream panel showing the proper sequence of stream sampling. c. Record the following data (all recorded on "Daily Checklist"). 1. Air cylinder pressure, high. 2. Air cylinder pressure, low. 3. Nitrogen cylinder pressure, high. 4. Nitrogen cylinder pressure, low. 5. Hydrogen cylinder pressure, high. 6. Hydrogen cylinder pressure, low. d. Check air, nitrogen, and hydrogen usage by looking at difference between the previous day's reading and present reading. A large difference indicates possible system leaks which would need correction prior to calibration. e. Insure that punp is operating normally (record on "Daily Checklist"). f. Check to see if flame is on (record on "Daily Checklist"). g. Check the oven temperature to insure that it is correct. (Record on "Daily Checklist.") h. Make.sure that gates 1 through 10 are operating correctly (record on "Daily Checklist"). i. Check the oven parameters: temperature, air pressure, air flow.
ABD00076848
Calibration (record data for each of the four units on "Honeywell Continuous Monitor Calibration Report" at end of this section).
a. Shut off regular sample flow. b. Open flow line of standard gas until flow meter indicates proper
flow rate (5 cc/min.). c. Change the sequencer panel setting from "auto" to "gate". d. After the peaks on the strip chart line out for the standard
gas reading, do the following:
1. change the sequencer panel setting to "spectrum". 2. change the amplifier panel setting from "auto" to "auto zero"
to insure correct timing between peaks. 3. when peak timing is seen to be stable, do the following on
the sequencer panel: set the "auto" and "auto zero" switches to "calib"; also do the following to the dual component panel: change the setting to "calib". 4. adjust the FID setting and the range setting. 5. reset the sequencer panel to "auto" and "auto zero". 6. reset the dual component panel to "auto". 7. if the resulting peak registers -5 ppm for the .5 ppm standard gas, the calibration is finished; 8. if the resulting peak does not' register 5 ppm, change the sequencer switches from "auto" to "calib" and from "auto zero" to calib; also reset the dual component panel on "calib". 9. repeat steps (4) through (7) above.
Post-Calibration Procedure
a. Check the panel settings to insure they are as described below:
1. sequence panel in "auto" setting. 2. amplifier panel set on "auto". 3. dual component panel set on "auto".
b. Make the following notations on the strip chart as an additional record that calibration has been accomplished:
1. date 2. time 3. calibration span (must be 10 ppm) 4. calibrator's initials
c. Shut off standard gas flow line.
d. Open normal sample flow line.
A
ABD00076849
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN PVC PLANT ' -
D.. CONTINUOUS MONITORING SYSTEM - CALIBRATION AND MAINTENANCE PROCEDURES
2. Maintenance
a. Preventive Maintenance Schedule and Procedure
Preventive maintenance includes the following procedure:
1. Check the air flow through the sample lines and the vacuum pump operation.
2. Check the gas bottles for adequate pressure (includes bottles of air, H2, N2); make sure backup supply cylinders are always present.
3. Check the pressure in the VCM standard cylinder. Standard is 5 ppm VCM.
4. Check the combustion flame for proper appearance. 5. Make sure oven temperature is correct. 6. Check the air pressure to valves. 7. Examine the recorder to insure that:
a. each sample point is represented on the chart; b. cycle times are correct. 8. Check the sample switching valves operation. 9. Check the sampling station lights to insure correct electronic operation of the system.
The preventative maintenance procedure is followed every week day. The attached preventative maintenance checklist is com pleted at the time of the check.
b. Corrective Maintenance Procedure
When a malfunction occurs, the following troubleshooting procedure is first followed:
1. Check the flame ionization cell for the voltage reading (should be -100 V DC); also check cleanliness of flame tip, and see if igniter is open.
2. Check preamplifier. 3. Check amplifier input and output. 4. Check recorder input and output. 5. Examine power supply. 6. Check sequencer to see if sample streams are being
' sampled in- proper order. 7. Check stream selector to see if sample inlet solenoid
valves are operating properly. 8. Check stream selector cards and dual component cards with
the test module card, to see if threshold voltage of the IC chip cells is correct.
ABD00076850
Corrective Maintenance Procedure (Cont.)
If the above procedure fails to pinpoint the cause of the mal function, a more detailed troubleshooting procedure must be followed. This procedure is now discussed in detail, as is the repair procedure.
The Model 1000 PGC is specifically designed so that significant electrical and electronic levels can be readily monitored at frontpanel test points, in the Control Section and in the Analyzer Section. Preamplifier test points, and the preamplifier "BALANCE" adjustment, are accessible by removing the threaded access Plate in the front of the Analyzer's explosion-proof electronics housing. If required, the entire front of the housing can be removed to expose the Analyzer terminal strips for measurement of Analyzer input and output levels. The terminal strips and the printed circuit card rack can be removed by turning 1/4 turn fasteners.
An additional maintenance feature is the meter and function-selector switch on the front panel of the Monitor Module. This feature is especially valuable in monitoring the operation of the Chromatograph, and in troubleshooting the equipment.
Inasmuch as possible, the troubleshooting procedures are intended to fully utilize the modular design concept of the Honeywell Model 1000. and the front-panel test points provided in both the Control Section and the Analyzer Section. The primary objective of these procedures is to localize the problem, so that where possible, module substitution can be used to quickly return the equipment to proper operation.
Preliminary Checks
Before attempting to troubleshoot the equipment, make certain that the trouble is not caused by improper control settings or by some external condition. Confirm that all operating parameters are as specified. If a discrepancy is noted, refer to the appropriate troubleshooting chart in this section to further isolate the problem and to implement repairs.
Spectrum Interpretation
Operation of the equipment in the Spectrum mode is an effective means of troubleshooting the unit. Run several analyses in the. spectrum mode to establish reproducibility, and interpret the results in accordance with Table 3.
Oven Heating System Troubleshooting
Refer to Table 15 for possible heater malfunctions and probable causes.
V ABD00076851 *
Analyzer Leak Check
Leaks in the Analyzer plumbing are a common cause of many operational prpblems. A leak can cause insufficient sensitivity, poor reproduci bility, excessive valve-switching transients, and overall system instability. If the equipment exhibits any of these malfunctions, the Analyzer should be leak checked as follows. If internal valve leakage is suspected, refer to the valve checkout procedures.
1. Apply carrier pressure to the Analyzer. 2. Plug all carrier gas vent lines. 3. Close valve at carrier gas supply and check that pressure
drop on the carrier supply gauge is less than 0.5 pounds per hour. If pressure drop exceeds this rate, localize the leak by successively plugging the exit from each valve or column.
Preamplifier Troubleshooting
Check that the preamplifier is operational by observing the preamp meter and varying the setting of the BAL (balance) adjustment potentio meter from side to side. The meter should move in a direction opposite the rotation of the BAL adjustment, and 'it should be possible to center the meter with this control.
If the meter indicates full-scale in either direction, adjustment can be simplified by placing a meter across the"+ BAL test points so that the effect of turning the BAL adjustment can be observed while setting the preamp 1-0-1 volt meter back within full scale. If the bridge will not balance, short out both detectors at the terminal strip on the preamp front panel, and again attempt to center the meter with the BAL adjustment. If the meter can be centered, a defective detector element is indicated.
If the meter cannot be centered, check from the ground test point to the + and - test points for +15 volts and -15 volts, respectively. The regulated +15 volts is produced within the preamplifier from the +20 volts supplied by the Control Section. If the +^5 volts is not present, check that the ^20 volts is present at the terminal strip in the upper portion of the explosion-proof housing.
Detector Bridge Checkout
Proper operation of the detector bridge can be confirmed, by monitoring the bridge current at the R-C-S test points on the preamplifier front panel, or at the R-C-M test points on the Dual Preamplifier. Use a 20,000 ohms per volt meter, or better, and measure from the common (c) test point to the R or M test point, and then the S test point. Each voltage reading, multiplied by three, provides the detector current. This value can be compared with the nominal detector current as stated on the Instrument Operational Parameters sheet. Some variation from the specified value may exist due to temperature vari ance, however, the two sides of the bridge should be equal.
ABD00076852
Detector Bridge Checkout' (Cojit.)
As a further check, the resistance of each thermistor can be measured by turning the power off at the Control Section, and disconnecting the detector leads from the terminal stirp on the front of the preamp. Refer to the Instrument Operational Parameters sheet for the nominal resistance value at 25C. Again, some variation from this value is acceptable due to temperature variance, however, the two thermistors should read the same.
Valve Checkout Procedures
1. Standard Valve (External Sample Loop)
a. Connect four-way valve to control ports (outer plates), and set air pressure to .32 psig.
b. Connect supply gas line to port #1. Use of helium is recommended.
c. Connect bubble tubes to ports #2 and #3. d. Plug port #4. e. Submerge bubble tubes in a beaker of water (down about 1/2"). f. Set supply gas pressure to 100 psig. g. Port #2 should flow freely with the four-way valve in one
position, and should not flow more than one bubble every 50 seconds with the valve in the other position. h. Port 3 should not pass more than one bubble every 50 seconds in either control position, however, momentary flow is permissable during the switching transient. i. Repeat steps 7 and 8 with the supply gas pressure set at 2 psig. j. Connect supply gas line to port #8. k. Connect bubble tubes to ports #6 and #7. l. Plug port #5. m. Submerge bubble tubes in a beaker of water. n. Set the supply gas pressure to 100 psig. o. Port #7 should flow freely with the four-way valve in one position, and should not flow more than one bubble every 30 seconds with the valve in the other position. p. Port #6 should not pass more than one bubble every 50 seconds in either control position, however, momentary flow is permissable during.the switching transient. q. Repeat steps 15 and 16 with the supply gas pressure set at 2 psig. r. Check for external leakage by plugging all ports except one and submerging the valve in iso-propyi alcohol with 100 lbs. of helium pressure. s. If the valve fails to meet the above requirements, replace the valve, or recondition the valve.
ABD00076853
2. Internal Sample Loop Valve
*
...
a. Connect four-way valve to control ports (outer plates), and set air pressure to 32 psig.
b. Connect supply gas line .to port #.l.
c. Connect bubble tubes to ports #2 and #3. d. Plug port #6. e. Submerge bubble tubes in a beaker of water (down
about 1/2").
f. Set supply gas pressure to 100 psig. g. Port #2 should flow freely with the four-way valve in
one position, and should not flow more than one bubble
every 30 seconds with the valve in the other position. h. Port #3 should not pass more than one bubble every 30
seconds in either control position, however, momentary
flow is permissable during the switching transient. i. Repeat steps 7 and 8 with the supply gas pressure set
at 2 psig. j- Connect supply gas line to port #8. k. Connect bubble tubes to ports #6 and #7. l. Plug port #3. m. Submerge bubble tubes in a beaker of water. n. Set the supply gas pressure to 100 psig. o. Port #17 should flow freely with the four-way valve in
one position, and should not flow more than one bubble every 30 seconds with the valve in the other position. P* Port #6 should not pass more than one bubble every 30 seconds in either control position, however, momentary
flow is permissable during the switching transient. q- Repeat steps 15 and 16 with the supply gas pressure
set at 2 psig.
r. Check for external leakage by plugging all ports except one, and submerging the valve in isopropyl alcohol with 100 lbs. of helium pressure.
s. If the valve fails to meet the above requirements, repalce the valve, or recondition the valve.
Repair and Replacement Procedures
Module Replacement
As previously mentioned, module substitution can be effectively used to troubleshoot the equipment, and to quickly return the equipment to proper operation. However, in substituting one module for another, the .following points must be considered.
The first three left-hand module positions in the Control Section must always contain the Monitor Module, the Sequencer Module, and the Amplifier Module, in that order. All other modules can be intermixed in the remaining module positions without restriction.
WARNING
Do not install or remove plug-in modules while power is applied to the Control Section. Damage.to the module and the Control Section will result.
.A
ABD00076854 t
In replacing a Monitor Module, a Component Module, an Integrator Module,, or a Valve Module, the green AUTO ZERO jumper must be moved to either the functional position or the "disabled" position, depending on the requirements of the application. Refer to the Instrument Operational Parameters sheet furnished with the equipment.
In replacing a Component Module, a Dual Component Module, an Integrator Moduel, or a Ratio Module, the blue GATE jumper (if GATE is used) must be set to coincide with the Component number for which the module is to be used. In Component Modules, the orange MEMORY jumper (if MEMORY output is used) must also be set to coincide with with the appropriate Component number. In replacing a Valve Module or a Dual Valve Module, the red VALVE jumper must be set to coincide with the corresponding valve number. Note that the VALVE #1 terminal on the Valve Module and Dual Valve Module printed circuit boards corresponds to the Valve #2 connections on the back of the Analyzer Section. In the Analyzer, Valve #1 is the Sample valve.
In replacing the Sequencer Module, it may be necessary to adjust the CHART ADVANCE potentiometer on the Sequencer printed circuit board to achieve the desired bargraph spacing. In addition, some adjustment of the VALVE DELAY potentiometer may be required to "fine turn" the sample inject duration to coincide with the setting of the previously used module.
In replacing a Stream Selector Module which uses analog stream identification, it may be necessary to adjust trimpot R50 on the analog identification printed circuit card to achieve the desired width of the identification bar.
Detector Replacement
1. Turn detector power and Control Section power off by means of the appropriate switches on the Monitor Module. Shut-off all gas flow through the Analyzer.
2. Remove four screws, and remove the cover from the detector housing.
3. Disconnect four detector leads from the terminal strip inside the housing.
4. Remove two screws and bend the terminal strip up and out of way.
5. Remove the four tubing nuts which connect the detector cell to the feed-thru adapters inside the housing.
6. From outside the housing, disconnect the four tubing leads. 7. . Loosen the lock-nut on each tubing adapter, and back out the
four adapters as required to obtain clearance for removal of the cell. 8. Remove the detector cell and mount securely in a vise. 9. Remove the two detector hold-down nuts and remove the two detectors. Replace with two detectors obtained from Honeywell Inc. Part Number for 8K Thermistor - 11600228001; for 100K Thermistor - 1160021003. 10. Re-install the detector hold-down nuts and use a box-end wrench to tighten the nuts securely.
ABD00076855
Detector Replacement (Cont.)
XI. Place the detector cell inside the detector housing, and retighten the four tubing adapters in the housing ends.
12. Re-connect the four tubing leads inside the housing, and tighten the four tubing nuts.
13. Place the terminal strip in position and secure with two screws. 14. Connect the four detector leads to the terminal strip. 15. Install the detector housing cover and secure with four screws.
Valve Overhaul
1. Remove the 3/8-inch bolt through the center of the valve and set aside the bolt, the spring washer, and the bottom plate.
2. Disassemble the valve, being careful not to drop the parts or misplace the two 1/8-inch dowel pins.
*3. Remove and discard the rubber diaphragm and the Teflon diaphragm . and paper backing.
4. Examine the center place for evidence of scratches or chemical etching. If defects are noted, replace with P/N 30740489 (internal volume) or P/N 30740483 (external loop) obtained from Honeywell, Inc.
5. Apply one drop of #550 silicone oil to each piston and reinstall in the respective cylinder. Check each piston for proper fit.
6. Re-assemble the valve by first inserting the two 1/8-inch dowel pins into the center plate. These pins will align the remaining parts during reassembly.
7. Install replacement Teflon diaphragms (Honeywell P/N 30740488-2) on each side of the center plate. The smooth side of the diaphragm must be against the center plate.
8. Install the backing paper, the cage, and the pistons and cylinders. Check that the blank cylinders are positioned over the blank areas on the center plate.
9. Install the replacement rubber diaphragms (Honeywell P/N 50740488-1) 10. Install the head plates, making sure that the four holes in the
lapped side of each plate are toward the center plate. 11. Install the spring washer and 3/8-inch hex head bolt, and
tighten to 10ft.-lbs. of torque. 12. Check the operation of the valve in accordance with the appropriate
procedure. In the event that port-to-port leakage occurs following installation of new diaphragms, increase air pressure to approx imately 100 psig to "wear in" the Teflon. After the valve has been "worn in" at 100 psig, reduce air pressure to 32 psig and repeat the valve checkout procedure.
Standard Sample Calibration
In order to perform a meaningful calibration, all equipment parameters must be as specified in the Instrument Operational Parameters sheet. In addition, allow at least 30 minutes for the oven temperature and the Control Section electronics to stabilize.
1. Check the preamp balance and adjust as required. 2. Set all front panel controls as specified in the Instrument
Operational Parameters sheet and run several bargraph analyses of the sample gas.
ABD00076856
3. Observe each component peak and check for improper component levels. To calibrate a specific component, proceed as follows.
4. Set the Amplifier Module AUTO ZERO/OFF Switch to the centered position. This will hold the baseline at the same position during the calibration.
5. Set the Sequencer Module MODE switch to MAN to inhibit the timer 6. Set all AUTO/OFF/CAL switches to OFF, except for the component
of interest.- Set this switch to the CAL position. 7. Set the Amplifier Module AUTO/CAL switch to the CAL position. 8. Adjust the Amplifier Module RANGE CALIBRATE control to produce
the same' recorder displacement as the component produced during the previous analyses. 9. Adjust the RANGE control on the Component Module of interest to achieve the correct percentage of full-scale deflection on the recorder. 10. Repeat steps 6 through 9 for each remaining component which requires calibration.
Memory Calibration
1. Set all front panel controls as specified in the Instrument Operational Parameters sheet. If desired, calibration time can be substantially shortened by setting the Sequencer Module TIME switches to about 10 seconds, and setting the gate time for the component of interest to about 7 seconds.
2. Set the Sequencer Module sample valve AUTO/MAN switch to the centered (off) position, and the MODE switch to any position except MAN.
3. Set all Component Module AUTO/OFF/CAL switches to OFF, except for the component of interest. Set this switch to the CAL position.
4. Set the Amplifier Module AUTO/MAN switch to AUTO, the AUTO/ZERO/ OFF switch to OFF, and the AUTO/CAL switch to CAL.
5. Monitor the amplifier output level by placing a meter across the Amplifier Module GND and OUT test points.
6. Adjust the Amplifier Module RANGE control to produce the desired full-scale output level from the amplifier. Allow the Sequencer to run through several cycles.
7. Measure the output at the GND and OUT test points of the Component Module of interest, and adjust the corresponding MEMORY SPAN control to bring the memory output to the level of the amplifier output, or to the trend recorder full-scale.
8. Place the Amplifier Module AUTO ZERO/OFF switch in the AUTO ZERO position and run several analyses.
9. Again measure the' output level at the GND and' OUT test points of the Component Module of interest, and adjust the corresponding MEMORY ZERO control to produce a 1 volt (or trend recorder zero) output.
10. Re-set the Amplifier Module AUTO ZERO/OFF switch to OFF, and repeat steps 7, 8, and 9 until the proper zero and full-scale memory outputs are produced.
ABD00076857
Chart Advance Adjustment
.A
Bargraph spacing between successive gates and between successive analyses can be adjusted by placing the Sequencer Module on a board extender and adjusting the CHART ADVANCE potentiometer on the printed circuit board. Because analysis spacing is a function of gate spacing, adjustment of CHART ADVANCE effects both functions.
WARNING
Do not install or remove plug-in modules while power is applied to the Control Section. Damage to the module and the Control Section will result.
Sample Valve Duration Adjustment
Sample valve inject duration can be adjusted up to one second beyond the thumbwheel setting by placing the Sequencer Module on a board extender and adjusting the VALVE DELAY potentiometer on the printed circuit board.
WARNING
Do not install or remove plug-in modules while power is applied to the Control Section. Damage to the module and the Control Section will result.
Heater Pressure Switch
The heater air pressure switch is factory set to automatically shut off electrical power to the oven heater in the event that the applied air pressure drops below 15 psig. No further adjustment is required. If defective, order replacement P/N 212-010-0005 from Honeywell, Inc.
Oven Temperature Adjustment
The optimum oven temperature, as stated in the Instrument Operational Parameters sheet, is factory set, and normally should not require re-adjustment. If adjustment is required, remove the threaded access plate on the front of the Analyzer electronics housing, and adjust the TEMP SET control on the front of the temperature controller. Clock wise rotation increases oven temperature, and counterclockwise rotation decreases oven temperature.
NOTE
If the oven operating temperature is to be changed from that specified on the instrument.
ABD00076858
Adjustment of Oven Over-Temperature Switch
The oven over-temp switch does not require adjustment, except when the Analyzer is re-configured by installation of new column types which require a change in the oven operating temperature. If adjustment is required, proceed as follows:
1. Remove the threaded access plate on the front of the Analyzer electronics housing so that the indicator on the front of the temperature controller can be seen.
2. Remove four screws on the front of the oven over-temperature switch and remove the cover plate.
3. Set the switch adjustment flush with the front of the setscrew color, then turn the adjustment 3 1/2 turns clockwise.
CAUTION
Do not turn the adjustment fully clockwise.
4. Close the oven door and apply air pressure and electrical power to the Analyzer Section. Adjust the TEMP SET control on the temperature controller as required to stabilize the oven at 10C above the specified operating temperature for the columns to be used.
5. Open the oven door and rotate the switch adjustment counter clockwise until the dimly-lit indicator on the controller goes out.
6. Re-adjust the TEMP SET control to stabilize the oven at the proper operating temperature.
7. Re-install the cover over the switch housing, and re-install the threaded access plate.
Adjustment of Heater Barrel Over-Temperature Switch
The heater barrel over-temperature switch does not require adjustment, except following replacement of the heater barrel. To adjust, proceed as follows:
1. Connect a pyrometer to the thermocouple leads which are coiled and stored below the explosion-proof electronics housing in the upper portion of the Analyzer Section.
2. Remove the threaded access plate on the front of the housing so that the indicator on the. front of the temperature controller can be seen.
3. Remove the-brass plug in the front of the heater housing, and turn the adjustment fully clockwise.
4. Apply power to the Analyzer Section and observe the pyrometer. When the barrel temperature reaches 300-350C, slowly turn the adjustment counterclockwise until the light on the temperature controller goes out.
5. Disconnect the pyrometer; coil and stow the leads; and install the brass plug and the threaded access plate.
ABD00076859
Auto Zero Track and Amplifier Offset Adjustments
The Auto Zero Track and Amplifier Offset adjustments are factory set, and field adjustment is not recommended. If the amplifier fails to meet the performance requirements, the Amplifier Module should be replaced and returned to Honeywell, Inc. for servicing. If field adjustment is to be attempted, the following instructions must be fully understood and precisely implemented.
NOTE
Place the Amplifier Module on a board extender and allow at least one hour for the Control Section-electronics to thermally stabilize.
1. Disconnect the preamplifier signal leads at the rear of the Control Section and short the amplifier input.
2. Set the Sequencer Module sample valve AUTO/MAN switch to the centered (off) position, and the MODE -switch to SPEC. Set the TIME selector switches to 02 seconds and the DURATION switches to 01 seconds.
3. Set the Valve Module AUTO/OFF/MAN switches to OFF. 4. Set the Monitor Module Auto Zero COARSE time selector switches
to 01 seconds. This will cause the auto zero function to be repetitively fired during the adjustments.5. Set the Amplifier Module AUTO/MAN switch to the MAN position and set the MANUAL RANGE selector to the "1000" position. 6. Set the Amplifier Module AUTO/CAL switch to AUTO and the AUTO ZERO/OFF switch to OFF. 7. Center the Amplifier Module AMPLIFIER ZERO control to produce zero volts at the wiper arm of the potentiometer. 8. Place a meter across the Amplifier Module GND and OUT test points, and adjust trimpot R42 on the printed circuit board to produce zero volts out. 9. Set the Amplifier Module AUTO/CAL switch to the CAL position, and adjust the RANGE CALIBRATE control to produce one or two volts (positive or negative) at the amplifier output. 10. Place the Amplifier Module AUTO ZERO/OFF switch in the AUTO ZERO position and allow the Sequencer to cycle three or four times. 11. Measure the voltage level at the amplifier output and adjust .trimpot .R18 on-the printed circuit board to produce zero, volts. 12. Turn the Amplifier Module AMPLIFIER ZERO control fully to one end, and allow the sequencer to cycle three or four times. 13. Measure the amplifier output level and note any variance from zero volts. 14. Adjust trimpot R18 to remove one-half of the error, and adjust trimpot R28 to remove the remaining error.
ABD00076860
15. Turn the Amplifier Module AMPLIFIER ZERO control fully in-the opposite direction.
16. Measure the amplifier output level and adjust R18 to remove one-half the error, and then adjust R28 to remove the remaining error.
17. Repeat steps 13, 14, and 15 until the Auto Zero circuitry tracks as close to zero as possible. Any error present must be the same at both extremes of the AMPLIFIER ZERO control.
18. Adjust trimpot R42 on the Amplifier Module printed circuit board to reduce any offset of the amplifier output to zero.
19. Perform the Auto Zero Tracking and Amplifier Offset Checks and -"fine tune" R18 to compensate for any negative baseline shift of the component with the highest gain setting.
20. Remove the short at the amplifier input and re-connect the preamplifier leads.
Stream Selector Analog Identification Adjustments
1. Set the analysis times as specified in the Instrument Operational Parameters sheet. If desired, the times may be set proportionally shorter to speedup the adjustment procedure. Set all Component and Valve modules to OFF.
2. Place the Stream Selector Module on an extender board, apply power to the Control Section, and allow 30 minutes for warmup.
3. Set the Sequencer Module sample valve AUTO/MAN switch to the centered position, and the MODE switch to SPEC.
4. Set all "stream-select" switches to the ON position. 5. Depress the Stream Selector RESET switch, and then the Sequencer
Module SEQ. RESET switch. 6. Center the zero adjustment (R73) and the span adjustment (R78).
Observe the recorder or use a meter at TP2, and adjust R78 to produce a 1-volt difference between each successive output as the Stream Selector cycles. 7. Adjust R73 to produce a 5-volt output when Stream 5 is selected. The resultant output for each of the other stream selections should be nominally: Stream 1=1 volt, Stream 2=2 volts, etc. Alternately adjust R73 and R78 to produce a readily identifiable output for each stream selected. Precise adjustment is not required. 8. Adjust R50 on the analog identification printed circuit board to vary the width of the identifying bar on the recorder..
Integrator Module Adjustments
1. .Place the Integrator Module on an extender board and apply power to the Control Section. Allow 30 minutes for warmup.
2. Set the Sequencer TIME switches to produce a 15-second analysis, and set the sample valve AUTO/MAN switch to the centered (off) position. Set MODE to any position except MAN.
3. Set the Integrator COARSE switches to produce an integrator gate which is ON at 05 seconds and is OFF at 10 seconds.
4. Short the Integrator input at pins 36 and 37 on the integrator printed circuit board. Adjust offset potentiometer R19 to produce 0 volts out of TP2.
ABD00076861
5. Adjust R38 on the integrator printed circuit board so that the voltage at pin 2 of Q4 is equal to the voltage at pin 6 of- Q4. On some models, R38 is a fixed resistor, and must be selected to produce the desired voltage at pin 2.
6. Monitor the front-panel GND and OUT test jacks, and adjust the front-panel ZERO control to produce 1-volt out after the integrator gate is closed.
7. Remove the short at the Integrator input. 8. Adjust the SPAN control, as required to produce an optimum output
waveshape for the peak(s) being integrated.
Ratio Module Adjustments
The five trimpots associated with the ratio circuitry of A1 and IC5 (R42, R44, R45, R48, R49) should not be adjusted in the field. These adjustments effect the operation of the ratio elements of the circuit, and specialized equipment is required for calibration and adjustment. Instead, adjust the input memories and the output memory as follows:
1. Place the Ratio Module on an extender board and apply power to the Control Section. Allow 50 minutes for warmup.
2. Set the Sequencer TIME switches to 60 seconds and set the sample valve AUTO/MAN switch to the centered (off) position.
3. Set one component gate of interest to be ON for 10 seconds duration, starting at 05 seconds into the analysis. Set the other component gate of interest to be ON for 10 seconds duration, starting at 25 seconds into the analysis.
4. Set the Ratio Module COARSE switch to 40 seconds. 5. Set all AUTO/OFF/CAL switches to OFF, except for the two
components of interest and the Ratio Module. Set these switches to AUTO. 6. Set the Amplifier Module AUTO ZERO/OFF switch to AUTO ZERO, and the AUTO/CAL switch to CAL. 7. Adjust the Amplifier Module RANGE CALIBRATE control to produce 5 volts out of the Amplifier Module GND and OUT test jacks during the gate associated with relays K1 and K2 in the Ratio Module. 8. Adjust zero control R22 on the Ratio Module printed circuit board to produce 1 volt at TP6 during the Auto Zero which occurs coincident with the gate opening. Adjust span control R31 to produce 5 volts at TR6 after the Auto Zero is completed. 9. Adjust the RANGE control on the Component Module associated with relays K3 and K4 in the Ratio Module, to produce 5 volts out of the Amplifier Module GND and OUT test jacks during the associated component gate. 10. Adjust zero control R26 on the Ratio Module printed circuit board to produce 1 volt at TP5 during the Auto Zero which occurs coin cident with the gate opening. Adjust span control R37 to produce 5 volts at TP5 after the Auto Zero is completed. 11. Allow the Sequencer to complete several cycles. Adjust the Ratio Module front-panel ZERO control to produce a 1-volt output at TP9.
ABD00076862
12. Adjust the RANGE control on the Component Module associated . with relays K3 and K4 in the Ratio Module, to produce 0 volts out of the Amplifier Module GND and OUT test jacks during the associated component gate.
13. Allow the Sequencer to complete several cycles. Adjust the Ratio Module front-panel SPAN control to produce a 5-volt output at TP9.
14. Re-calibrate the Component Module associated with relays K3 and K4 in the Ratio Module.
Re-Application Data
The preamplifier gain can be increased or decreased by changing the values of resistors R15 and R17 in the Preamplifier Module. The preamp gain is normally a function of the intended application and is thus pre set at the factory. In the event of reapplication in the field, the following factors should be considered.
Resistors R15 and R17 are a precisely matched pair of high-quality wirewound resistors. If the preamp gain is to be changed, refer to Table 8 and order the appropriate matched pair from Honeywell, Inc. The gain of the preamp should be selected so that the largest anticipated peak will not drive the amplifier to saturation. Max imum peak output should be 5 to 6 volts.
In installing the replacement resistors, exercise extreme care in soldering the leads in place. Do not apply prolonged excessive heat which could change the resistance of the components.
Component Module or Integrator Module Gain Change
The gain range of a Component Module or Integrator Module can be increased or decreased by changing the value of resistor R65 on the printed circuit card. Refer to Table 9 and select a resistor value which will produce a slightly less than full-scale output and will allow final full-scale adjustment by means of the Component Module or Integrator Module RANGE Control.
Installation of New Column Types
In the event that field re-application involves installation of columns which differ from those previously used, special attention must be given to establishing the proper oven temperature.
ABD00076863
TABLE 14 SPECTRUM ANALYSIS TROUBLESHOOTING CHART
Indication
Possible Cause or Solution
Baseline drift in one direction
1.
2. 3.
4.
' 5. 6. 7.
Analyzer oven not stabilized at proper temperature. Incorrect column or reference flow. Check for column leak, or uneven column bleeding in two-column system. Check the sample for materials which may be entering the column as a liquid. Check detector bridge current. Replace supply gas cylinder(s). Replace supply gas drier(s).
Wavy or sine-wave baseline
1. Analyzer oven temperature cycling. 2. Check the recorder. 3. Check stability, of column flow. 4. Check carrier pressure and leak check
carrier flow system. 5. Check sample gas for leakage into
carrier gas.
Noisy or erratic baseline
1. Defective Preamplifier Module. 2. Replace Amplifier Module. 3. Excessive power supply ripple. Monitor
power supply output with an oscilloscope at the rear of the Control Section. 4. Check for loose leads, bad ground connections or dirty switch contacts. 5. Check for carrier gas leak. 6. Replace or recondition contaminated columns.
Improper elution times
1. Check that the Analyzer oven has stabilized at the proper temperature.
2. Check the column flow rates. 3. Replace ox recondition the columns.
Incorrect component readout
1. Check that the Analyzer oven has stabilized at the proper temperature.
2. Calibrate the Control Section.3. Check the column flow-rates. 4. Check for proper valve switch timing. S. Check for foreign components in the sample.
Baseline shift at Auto Zero times
1. Auto Zero circuitry defective or improperly adjusted.
ABD00076864
TABLE 14 (Cont.)
. Indica-tion No peaks
Poor Sensitivity
Steady baseline drift in one direction, then in the other direction. "Stairstep" baseline Inverted peaks Incorrect trend output Auto Zero not functioning
Column-switching valve not functioning
Memory (trend output inoperative)
Possible Cause or Solution
1. Check that the Monitor Module DET
. (detector) power switch is on.
2 Check for proper flow rates and air pressure settings.
3. Check valve switch timing. See Instrument Operations Parameters sheet.
4. Check for defective sample inject valve. 5. Check preamp and detector bridge.
6. Check for defective column-switching valve.
1. Check for improper MANUAL RANGE setting.
2. Check that the Analyzer oven has stabilized
at the proper temperature. 3. Check for insufficient sample size or
partially blocked sample volume chamber. 4. Check for extreme preamplifier imbalance.
1. Check for cycling oven temperature. 2. Check for contaminated carrier or drier.
1. Check recorder for proper operation. Check ground connections.
1. Check for reversed recorder leads at the rear of the Control Section.
1. Memory circuitry defective or improperly adjusted.
1. Check that Amplifier Module AUTO ZERO/OFF
. switch is set to AUTO ZERO position.
2 Check that the corresponding Auto Zero jumper (green) is set to the functional position.
1. Check for proper air pressure. Adjust as required.
2. . Check that the. corresponding valve switch on the Solenoid Driver Card is in the AUTO position.
3. Check that the Valve Module VALVE jumper (red) is set to the proper position.
1. Check that the Component Module MEMORY jumper (orange) is set to the proper position.
ABD00076865
Indication
Incorrect Ratio Output
Incorrect Integrator Output Incorrect Analog Stream Identification Stream Selector Inoperative
TABLE 1.4 (Cont.)
Possible Cause or Solution
1. Ratio memory circuits defective or improperly adjusted.
1. Integrator module defective or integrator memory circuits improperly adjusted.
1. Stream Selector analog identification circuit defective or imporperly adjusted.
1. Stream 1 ON-OFF switch is not set to the ON position.
ABD00076866
TABLE 15 OVEN HEATING SYSTEM TROUBLESHOOTING CHART
Indication No heating
Oven stabilizes at incorrect temperature Oven temperature will not stabilize
Possible Cause or Solution
1. Check AC power to Analyzer Section at terminal strip inside explosion-proof electronics housing.
2. Check fuse on front of temperature controller module.
3. Check that sufficient air pressure (34 psig) is applied, and that the heater air shut-off valve is open.
4. Check for 115 VAC across the heating element.
5. Check for AC power to and through the contacts of the oven over-temperature switch, the low-pressure shut-off switch the heater barrel over-temp switch.
1. Temperature Controller Module defective or improperly set.
2. Oven over-temp switch defective or set below the specified operating temperature.
3. Check for leaks around oven door seals.
1. Check operation of Temperature Controller Module.
2. Check for leaks around oven door. 3. Check for air flow obstructions in the oven. 4.- Heater barrel over-temp switch or oven
over-temp switch defective.
ABD00076867
HONEYWELL VALVE OVERHAUL
Procedure: Disassembly
1. Place the valve into a vise allowing only the mounting fixture to remain in the jaws of the vise.
2. Loosen the 3/8 inch bolt by turning in a counter clockwise direction.
3. Remove entire valve from vise and prepare for disassembly in a clean area.
4. Remove 3/8" bolt, belleville washers and force washer from-valve head plate, and set aside.
5. Carefully slide remaining portion of valve from mounting fixture. 6. Place the valve on clean tissue paper with port number 8 toward you and
at the lower right hand corner. 7. Remove head plate from top portion of valve and inspect surface for
scratches.
8. Leave both dowel pins in place until disassembly leaves only the center plate.
9. Remove the actuating diaphragm and discard.
10. Remove cylinder plate and inspect cylinder walls for wear and galling.
11. Remove pistons, cylinders and cylinder blank from valve and check each piston and cylinder for ease of movement. This is done by gently tapping the bottom of the piston with the finger and observing the free movement of the piston within the cylinder. If the piston does not move freely, check to determine if piston or cylinder is the cause, and replace as necessary.
12. Remove sealing diaphragm and discard.
13. Taking remaining portion of valve, turn it over and repeat above operation
ABD00076868
HONEYWELL VALVE OVERHAUL
Procedure: Assembly
1. Place a clean, lapped and de-burred center plate upon a clean tissue.
2. Having carefully inspected the above plate for scratches, insert two dowel pins on either side of the center plate.
3. Turn the center plate such that port 8 will be at the lower right hand corner facing you.
4. Take one sealing diaphragm and inspect both sides for scratches, cuts, etc
5. Place over, dowel pins and onto center plate such that the dull side faces the center plate and slide down onto the center plate.
6. Slide a clean cylinder plate over dowel pins onto the sealing diaphragm. Cylinder plate must be free of scratches on inside walls that house piston and cylinder assembly.
7. Pistons and cylinder assemblies, and cylinder blanks must be thoroughly cleaned with freon, and cylinder and cylinder and piston land areas inspected for scratches, etc. Each outside wall of cylinders must be inspected for evidence of binding in the cylinder plate.
8. Each piston and cylinder assembly must move freely without any binding.
9. Place the cylinder blank and piston and cylinder assembly into their respective positions and be sure that the vent hole faces the vent area in the cylinder plate.
10. If the valve is to be used in a dual column mode, place a small viton o-ring 1/32" thick over the Diston which controls flow to nort #2. Check assembly drawing. Do not insert Dlue into Dort #2 until valve test is comDlete.
11. Place clean actuating diaphragm over dowel pins and onto top of cylinder plate housing piston-cylinder assemblies.
12. Place top head plate over dowel pins making sure exit holes are down. The actuating air connection port is on the side with ports #6 and #7 for 8 port valves, and on the same side as ports #3 and #6 for 6 port valves. Plug air ports not used with 1/8" s.s. ball bearing and seal with screw socket.
13. Turn the valve over and repeat above procedure exactly except replace step 10 with step 14.
14. If the valve is to be used in the dual column mode, place a small viton o-ring 1/32" thick over the piston controlling flow to port #7. Check assembly drawing. Do not insert plug into port #7 until valve test is complete.
15. Having completed the above, turn the partially assembled valve right side up and place upon the mounting fixture.
ABD00076869
Procedure: Assembly (Cont.)
16. 'Slide the belleville washer stack onto the bolt. Check assembly drawing to be sure the belleville washers have been stacked properly.
17. Slide the force washer onto the bolt below the belleville washer. Stack and insert through the center of the valve and' thread hand tight into the mounting fixture.
18. Grasp mounting fixture only in a holding device and with a torque wrench and 9/16" socket, tighten to 110 inch pounds.
19. The valve should be re-torqued after 24 hours and stored in a protective enclosure when not in use.
A
ABD00076870
VALVE TEST PROCEDURE
- Check torque to be sure it is 110 inch/pounds.
- Check that ports in manifold plates which are not being used are plugged. The actuating air line from each manifold must be connected to a four way actuating valve - or to two two-ways.
- Set actuating pressure to 50 PSIG.
- Actuate the valve 20 times, allowing 5 seconds between actuations.
Flow Rate Test
- Set supply gas (carrier) pressure .to 1.0 PSIG and connect to ports as belov;. A flow rate of at least 300 ml/min should be measured out of the appropriate port. Should proper flows not be obtained, disassemble valve and inspect center section for restricted passage.
(A) Standard 8 Port Valve
Supolv Gas
Flow Meter
Manifold Pressurized
#L.................. #1..................
#3.................. #5.................. #3.................. m..................
.. ..
.. .. ..
#2.................. #4..................
#6.................. m.................. m..................
Internal Loop Sampling Valve
SuddIv Gas
Flo'w Meter
Manifold Pressurized
L.................. m...................
#3.................. #7..................
.. ..
.. ..
#2.................. #6..................
#8.................. m..................
Dual Column Valve Applications
Suoplv Gas
Flow Meter
*
Manifold Pressurized
#5 ...... . . #5 . . : . . . .
#1.................. . . #3.................. -
#6.................. #8 . . . . . #4.................. #4..................
Backflush Valve Applications
Supply Gas
Flow Meter
Manifold Pressurized
#L.................. . . #1..................
. #3.................. #5.................. . . . ..
#2..................
#6.................. #8 . . . t .
ABD00076871 , rR^;jnK
-Bet supply gas pressure at 110 PSIG. Fur charier pressure consult Tabie 1 for the proper air actuation pressure;
;xcess or 110 ltd ;,
-Connect supply ras, plugs, and bubble tube to appropriate ports as described below. When properly connected, no flow must be observed through the babble tube.
-If gas flow is observed, disassemble-the valve and inspect for foreign matter ir the sealing area. Also, inspect the center section, cylinder lands, piston lands for surface
(A) Standard 8 Port Valve
SuddIv Gas
Bubble Tube
PLuered Ports
Manifold Pressurised
#1 ... ... #1 . . . ... #1 . . . m . . . ...
. . . ... #8 . . . ...
#2.................. ` #2 ..................
#6 .................. m ...;. #5 ..................
#3, #4 #4
#2, #3
#5, #7 #5
#6, #7
.
.
.
- Connect jumper tube between ports #4 and#5 for external sampling application. Plug port - Connect 110 PSIG supply- gas to port #3.
- Connect bubble tubes to ports #6 and #2 . - Apply actuation pressure to top manifold and note momentary bubble of gas out of port ?f6. - Apply actuation pressure to bottom manifold. No bubble must be obser/ed at port r`6 or #2.
(B) Internal Loop Sampling Valve 6 Port
SuddIv Gas
Bubble Tube
Pluoe-ed Ports
Manifold Pressurized
#1 . . .
#3, #6, *7,
.......................Top
#1 . . . .... #3 .................. 3^2, rfo, tt7, rr8
#3 . . . .... #1 .................. #6,
#8 .
#3 . . . .... #6 .................. nG7, #8
#7 . - . .... #6 ..... #3, #L ...
#7 . . . .... #8 .................. 0
....
- Connect 110 PSIG supply gas to pert #3- Connect bubble tubes to port #6 and #2. Plug port nS. - Apply acutation pressure to top manifold and note mome ntary bubble out of port 6. - Apply actuation pressure to bottom manifold. No bubble must be observed at port #6 or #2.
(C) Dual Column Valve Application
SuddIv Gas
Bubble Tube
Plumzed Ports
Manifold Pressurised
#L . . - . . .
#1 . . . ...
#1' . . . . t
#8 . . . ... #B . . . ...
m . . . ...
#2 ..... . #4 ....................... #3'. .... #7 ....................... #5 ....................... #6 ......................
. .
#3, #4 . . #3 . . #4 . . .
#5, #6 . . #6 . . .
#5 . - .
i vi';. j n^2 and $7 should now be plugged.
(D) Backflush Valve Application
SuddIv Gas
Bubble Tube
Fluffed Ports
Manifold Prossurlo-d
#1 . . . .... #2 .................. . . #3, #4 . . .......................Top
#1 . . . .... #3 .................. . . #4 . . .......................Top
#1 . . .
. . #2f #3 . . ....................... Bottom
m . . . ....
.................. . . #5, #6 . . ...... Bottom
m #8
.. . . . ................. #6
.m . . ...... Bottom
..... . . m, #n . ............ Top
- Port #7 should now be plugged.
ABD00076872
Electronic Zero Periodic Verification Procedure
Th.e following procedure is followed at the same time it is calibrated.
1. Set the Sequencer Module sample valve AUTO/MAN switch to the centered position, and the MODE switch to SPEC.
t
2. Set the Amplifier Module AUTO/MAN switch to MAN, and the MANUAL RANGE selector to "1000". Place the AUTO ZERO/OFF switch to OFF.
3. Place the Amplifier Module AUTO/CAL switch in the CAL position, and adjust the RANGE CALIBRATE control to produce one or two
volts at the amplifier output.
4. Place the Amplifier Moduel AUTO/MAN switch to AUTO, and the AUTO ZERO/OFF switch to AUTO ZERO.
5. Obsepve the baseline and check for any excessive baseline shift during the gate openings. If excessive baseline shift is noted, the Amplifier Module should be replaced and returned to Honeywell for service.
0A1L1 CHECK LIST
UR PREr^RE
HIGH A
RECORD .0/in2
HIGH B
LOW A
LOW B
NITROGEN PRESSURE
HIGH A
RECORD 0/in2
HIGH B
LOW A
LOW B
HYDROGEN PRESSURE
HIGH A
RECORD 0/xn2
HIGH B
LOW A
LOW B
STANDARD GAS 0/in2
HIGH
PUMP
RUNNING
FLAME ON
OVEN TEMPERATURE C
CHECK GATE AND ADJ. AS NEEDED
check sample flow on #1
02
n
u
05
06
07
08
09
010
ABD00076873
HONEYWELL
%
MON.
TUES.
LOCATION WED. TOURS.
_;______________________
FRI.
SAT.
SUN.
A
----- aBttT UU76874
?\
C o m m e n ts
-
r
C a lib ra tio n Checked by
Reading W ith
Span Gas
After Calibration (ppm )
Reading W ith
Span Gas
Prior to Calibration (ppm)
HONEYWELL CONTINUOUS MONITOR CALIBRATION REPORT
3O. JOS
9
co
6s0 T<J9
V
4pJ
Oyo CC Cfc3>
y o -a
a O n
MO
*3 ~0
*O -ap
s o
.o o
y
<>s> w2C3 cO^
c
<9
.M
in o w
Z O
>
O
*<CP9
o J~US
<
O
uy
o y
J
3 *O3 ^
3 crc3-
5 1 *cc
-CO
> oN tflx>59 3 vi
OOPi--
v) 2^O=
wo~
-o5 --E5cy
<o
2
"wy
r
2 5y
oy
X3<-**>
.52/5
7.
----'
'J 4)
cwQ> O
*(39 --sp
tN oO 23 .S< 00 o
C *3
*0 C
cn
J9
Tr3t
-o3
9 O yi- 2*-
oCt
> *
3 Q>
5 yo
(=9 U
rka_
s>
s
3
^
O
-y3
y J ^ g3j) '-
3 '5 2 5 5
c3
V>5 byy.
"
--O
y
--
J9 5
y
O
~
z<> Q
*e3 0^ >< Q o-Z EC9 C<^
t/> u- O X
2Sg2
U y
y U 3
J CT C
O -- cl
aC
3
O0
2-
OL. CC c. *->
< 2-- w0 "--
"b. --
> y uy.
y y
^5
10
^4) S
o
u
Vi
c3 os
Hoo i--
--
_S**Z-w
CO
y *->
CO
y
00
(A
<
^
<
u. -- to
-(sm^w'Ohco C\ O=i
Span Gas
Concentration (pprii)
Range PPM
-- C'i -- ri -- ri
Gas
Chromatograph Location
Allm ond 213
ABD00076875
* LEAK DETECTION AND ELIMINATION PROGRAM
' ' ABERDEEN PVC PLANT
PORTABLE HYDROCARBON DETECTOR - CALIBRATION AND MAINTENANCE PROCEDURES
1.. Calibration
There are three HNU portable hydrocarbon detectors in the plant. One is kept at the control room to pinpoint leaks found by the continuous monitoring system. The other two are stored in the guardhouse. Once per week, the monitor in the control room is taken to the guardhouse and exchanged for a monitor that has been checked and .calibrated. The used monitor is taken to the instrument shop for calibration and checking. This monitor is returned to the guardhouse (see attached calibration report form). The portable hydrocarbon monitors are also used for routine leak detection monitoring. The person performing the routine leak detection monitoring picks up a monitor from the guardhouse and returns it to the instrument shop. The instrument shop then returns the monitor to the guardhouse.
The first step in the calibration procedure for the portable hydrocarbon detector is to turn the instrument to the standby position and adjust the' electronic z*ero. A pressurized container of a calibration gas is connected to one side of a "T". Another side of a "TV is connected to a rotameter and the third side of the "T" is connected directly to the 8" extension to the photoionization probe (see figure 6). The cylinder has a pressure regulator which is set at about 5-10 psig. The valve on the cylinder is opened until flow is indicated on the rotameter. The instrument draws in the volume of sample required for detection, and the flow in the rotameter indicates an excess of sample. This flow is vented out of the building. Now the span pot is adjusted so that the instrument is reading the exact value of the calibration gas. (If the instrument span setting is changed, the instrument should be turned back to the standby position and the electronic zero should be readjusted, if necessary.)
The calibration gas is traceable to the National Bureau of Standards.
2. Maintenance
If problems occur while using the photoionization analyzer, it is re commended that the following troubleshooting guide be followed before consulting the factory.
General Aid to Fault Determination
Check battery condition. Recharge if necessary. Turn the instrument on. Look into the Sample Inlet of the probe unit. A violet colored glow from the UV light source should be observed in all positions of the mode switch except the standby position. If unstable readings are obtained, *a faulty probe cable or electrical connection could be the problem. To check, hold the probe normally and flex the cable firmly. Watch the meter for fluctuations as the cable is stressed. Individual wires in the readout can be checked in a similar way. Check the coaxial connector on the amplifier board in the probe.
HNU PORTABLE HYDROCARBON DETECTOR
u
..
PORTABLE DETECTOR NUMBER
CALIBRATION GAS CYLINDER
NUMBER
DATE
CALIBRATION
READING
GAS
WITH
CONCENTRATION CALIBRATION
(PPM)
GAS BEFORE
CALIBRATION
(PPM)
READING WITH
CALIBRATION GAS AFTER
CALIBRATION (PPM)
CALIBRATION
.CHECKED BY
COMMENTS ....
ABD00076876
- N-
ABD00076877
In the more sensitive ranges, a fluctuation in the reading may be noted if hand or other large object is placed in very close proximity to the probe. This is normal for the instrument and will not result in an error in the measurement as long as the probe is held stationary while the measurement is being taken.
If the probe is held close to AC power lines or power transformers an error may be observed. If measurements are to be made in close proximity to such items, their effect on measurements can be determined by the following procedure. Zero the instrument in an electrically quiet area in the standby position, then move the instrument to the questionable area involved. If AC pickup is going to be a problem, the meter (in the standby position) will indicate the magnitude of the error.
Disassembly of Instrument
PROBE - Turn the function switch to the OFF position and disconnect the probe connector from the readout unit. Remove the exhaust screw found near the base of the probe (see Figure 7}. Grasp the end cap in one hand and the probe shell in the other,, and gently pull to separate the end cap and lamp housing from the shell. Loosen the screws on the top of the end cap and separate the end cap and ion chamber from the lamp and lamp housing. Care must be taken so that the ion chamber doesn't fall out of the end cap and the lamp doesn't slide out of the lamp housing. Turn the end cap over in your hand and tap on the top of it; the ion chamber should fall out in your hand. Place one hand over the top of the lamp housing and tilt slightly; the light source will slide out of the housing. The amplifier board can be removed from the lamp housing assembly by unsnapping the coaxial connection and them removing the retaining screw.
To reassemble this unit, first slide the lamp back into the lamp housing. Place the ion chamber on top of the lamp housing, checking to make sure that the contracts are properly aligned. Place the end cap on top of the ion chamber and replace the two screws. The screws should be tightened only enough to seal the "0" ring. DO NOT OVERTIGHTEN. Line up the pins on the base of the lamp housing with the pins inside the probe shell. Gently slide the housing assembly into the. probe shell. DO NOT FORCE the assembly into the shell as it only fits one way.
READOUT UNIT - Turn the function switch to the OFF position and disconnect the probe from the readout unit before disassembly is conducted (see Figure 9). Remove the accessory power jack plug. Loosen the screw on the bottom of the case and, holding the instrument by the bezel, remove the case. The power supply board and control panel can be removed by unscrewing two screws and two nuts. The entire panel, including the function switch, zero and span pots is removed in this operation. Electrically disconnecting this module is simple, since all connections are made with Molex connectors.
ABD00076878
Vent A
Figure 6
Calibration Procedure for Photoionization Analyzer
ABD00076879
4
Figure 7 Component Parts of Probe
*-<
ABD00076880 Coaxial Connector
Amplifier Board
.i*. s &)*.& *X?
Figure 8 Component Parts of Lamp Housing.
ABD00076881
Specific Faults
^
1. No meter response in any switch position (including BATT CHK)
A. Broken meter movement
(1) Tip instrument rapidly from side to side. Meter needle should move freely, and return to zero,
B. Electrical connection to meter is broken
(1) Check all wires leading to meter and clean the contracts of quick-disconnects.
C. Battery is completely dead
(1) Disconnect battery and check voltage with a volt-ohm meter.
D. Check 2 amp fuse
E. If none of the above solves the problem, consult the factory.
2. Meter responds in BATT CHK position, but reads zero or near zero for all others
A. Power supply defective
(1) Check power supply voltages per Figure 10. If any voltage is out of specification, consult the factory.
B. Input transistor or amplifier has failed
(1) Rotate zero control; meter should deflect up/down as control is turned.
(2) Open probe. Both transistors should be fully seated in sockets
C. Input signal connection broken in probe or readout
(1) Check input connector on printed circuit board. Should be firmly pressed down.
(2) Check components on back side of printed circuit board. All connections should be solid, and no wires should touch any other object.
(3) Check all wires in readout for solid connections.
ABD00076882
3. Instrument responds correctly in BATT CHK, 'and STBY, but not in measuring mode.
A. Check to see that light source is on (See General Faults section.)
(1) Check high voltage power supply (see Figure 10).
(2) Open end of probe, remove lamp and check high voltage on lamp contact ring.
(3) If high voltage is present at all above points, light source has most likely failed. Consult the factory.
4.* Instrument responds correctly in all positions, but signal is lower than expected.
A. Check span setting for correct value.
B. Clean window of light source
C. Double check preparation of standards
D. Check power supply 180 V output. See Figure 10.
E. Check for proper fan operation. Check fan voltage. See Figure 10.
F. Rotate span setting. Response should change if span pot is working properly.
5. Instrument responds in all switch positions, but is noisy (erratic meter movement).
A. Open circuit in feed back circuit. Consult the factory.
B. Open circuit in cable shield or probe shield. Consult the factory.
6. Instrument response is slow and/or irreproducible.
A. Fan operating improperly. . Check fan voltage. See Figure 10.
B. Check calibration and operation. 7. Low Battery indicator.
t
A. Indicator comes on if battery charge is low.
B. Indicator also comes on if ionization voltage is too high.
Battery
ABD00076883 Power Supply Board
Figure 9 Component Parts Of Readout Unit.
ABD00076884
I. All voltages measured with probe connected and instrument . mode switch in BATT CHK position. *
Z, All measurements referred to ground (pads2,3 and27.) except pad 21 measured to pad 19 and pad 8 to pad 11
Pad 17 + 180 to +181 Volts
Pad 18 +18to+21 Volts
Pad 8 -11 to -14 Volts Battery Voltage
Pad 20 -9.5 to-10.5 Volts
pads 1 2 3 4 5 6 7 8
voltage - 5.7V
GRD GRD -10 7V - 11.3V -12.1V
0 - 12.2V
Pads 21 to 19 (21 low side) (4.5 to 5.5 Volts)
All Voltages Respect to Ground
pads
voltage
pads .
voltage
9 - 12.2V 10 - 12.1V
17 180V 18 - 19.4V
11 0 12 0 13 0 14 .0
19 - 10.6V 20 - 9.7 V 21 -14.5V 22 -400V
15 .0 23 0
16 0 24 0
pads 25 26 27 28 29 30 31
Figure 10 Power Supply PC Board
voltage 0 0
' GRD 0
GRD GRD GRD
ABD00076885
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN PVC PLANT
RODTINE LEAK PATROL - DATA RECORDING AND HANDLING
A leak patrol survey with the portable hydrocarbon detector is conducted once per week. The locations monitored are listed on the attached leak patrol survey form. The inspector records the concentration measured at each point. When a leak is detected during the survey, the date and time leak detected, location, portable monitor reading, and cause of leak are recorded on the leak detection reporting form for portable detector by the person conducting the leak patrol survey. He then gives this form to the shift supervisor for corrective action. When the leak is repaired, the date and time are recorded. Verification that the leak has been re paired is made by taking another reading with the portable hydrocarbon detector. After verification, the leak detection reporting form for portable detector is given to the process engineering technician for filing.
ABD00076886 LEAK PATROL SURVEY
Page 1 of 6
DATE: _______ J_______ _____ _
LOCATION:
INSPECTOR:
I.New Reactor Area - Reactor Module
_______
Cone. ppm
_______ ______
-Cone. . Cone. ppm ppm
A. Ground Level
1. 741 Charge valve
. _______
2. Directly Under 741
_______
3. 742 Charge Valve
4. Directly Under 742 5. 743 Charge Valve
_______
6. Directly Under 743
_______
7. 744 Charge Valve
8. Directly Under 744
_______
Cone. ppm
B. Third Level (Top of Reactors) 1. 741 Lenape Seal 2. 741 Condenser Bottom Seal 3. 741 Recovery Valve 4. 741 Rinse Water Line 5. 742 Lenape Seal 6. 742 Condenser Bottom Seal 7. 742 Recovery Valve 8. 741 Rinse Water Line 9. 743 Lenape Seal
10. 743 Condenser Bottom Seal
ABD00076887
B. Third Level (Top of Reactors) (Cont.) 11. 743 Recovery Valve 12. 743 Rinse Water Line 13. 744 Lenape Seal 14. 744 Condenser Bottom Seal 15. 744 Recovery Valve 16. 744 Rinse Water Line
Cone. ppm
Cone. ppm -
Page 2 of 6
Cone. ppm
Cone, 'ppm
A
Cone. ppm'
Top of Condensors 1. 741 Condenser Seal 2. 741 Condenser Valve 3. 742 Condenser Seal 4. 742 Condenser Valve 5. 743 Condenser Seal 6. 743 Condenser Valve 7. 744 Condenser Seal 8. 744 Condenser Valve New Reactor Area - Recovery Unit A. 2nd Floor 1. 901 (South) Vacuum Pump Seal 2. 901 (South) Vacuum Pump Overflow 3. 902 (North) Vacuum Pump Seal 4. - 902 (North). Vacuum.Pump Overflow 5. 903 Recover Compressor Seal 6. 903 Recovery Compressor Overflow 7. 904 Recovery Compressor Seal 8. 904 Recovery Compressor Overflow 9. 905 Recovery Compressor Seal 10. 905 Recovery Compressor Overflow
ABD00076888
* II. New Reactor Area - Recovery Unit
. Cone. PPTM'
' Cone.
PPm -
Page 3 of 6
Cone.
Cone.
PPm ppm '
A. 2nd Floor (Cont.)
11. VCM Cqllection Tank (Area)
_______
. _______
12. East Recovery Condenser (Area)
_______ _______ _______ _______
13. West Recovery Condenser (Area)
_______ _______ _______ _______
B. Ground Level
1. East VCM Storage Tank Bottom (Area)
_______ _______ _______ _______
2. East VCM Storage Tank Top -(Area) _______ _______ _______ _______
3. Middle VCM Storage Tank Bottom (Area)
4. Middle VCM Storage Tank Top (Area)
5. West VCM Storage Tank Bottom (Area)_______ _______ _______ _______
6. West VCM Storage Tank Top (Area) _______ _______ _____ _ _______
7. East (894) Rec. VCM Return Pump Seal
_______ _______ _______ _______
8. West (893) Rec. VCM Return Pump Seal
_______ _______ _______ _______
9. South (891) Seal Water Separator Pump Seal
_____ _______ _______ _______
10. North (892) Seal Water Separator _______ _______ _______ _______ Pump Seal
11. Seal Water Filter Pack Head
_______ _______ _______ _______
12. East Rec. VCM Charge Pump Seal
_______ _____ __
,
-
13. East Rec. VCM Charge Pump Overflow _______ _______ _______ _______
14'. West Rec. VCM Charge Pump Seal
_______
,
_______ . , .
15. West Rec. VCM Charge Pump Overflow _______ _______ _______ _______
16. Rec. VCM Filter Pack Head
_______ _______ _______ _______
17. East Rec. Charge Pump Seal
_______ _______ _______ _______
18. East VCM Charge Pump Valve
_______ _______ _______ _______
19. West VCM Charge Pump Seal
_______ _______ ________ _______
4 Cone.
ppm-
ABD00076889
*
II. New Reactor Area - Recovery Unit
Cone. ppm -
Cone, ppm
B. Ground Level (Cont.)
20. West VCM Charge Pump Valve
_______ _______
21. North Fresh VCM Filter Pack Head _______ _______
22. South Fresh VCM Filter Pack Head _______ _______
III. Old Reactor Area - Recovery
A. Area West of V-ll
1. Middle VCM Receiver, Top (Area) _______ _______
2. Middle VCM Receiver, Bottom (Area) _______
3. South VCM Receiver, Top (Area)
_______ _______
4. South VCM Receiver, Bottom (Area) _______
.
5. 401 Receiver, Top (Area)
_______ _______
6. 401 Receiver, Bottom (Area)
_______ _______
7. VCM Transfer Pump (Under South Receiver)
_______ _______
8. East VCM Charge Pump Seal
_______ _______
9. West VCM Charge Pump Seal
_______
10. Rec. VCM Filter Pack Head
_______ _______
11. North, Fresh VCM Filter Pack Head _______ _______
12. South, Fresh VCM Filter Pack Head _______ _______
13. Blowdown Tank Relief Vent (Platform)
_______
14. West VCM Return Pump Seal
_____ _ _______
15. East VCM Return Pump Seal
_______ _______
B. 1st Floor - V-ll (Area)
_______ _______
C. 2nd Floor
1. A (East) Vacuum Pump, Seal
_______ _______
2. B (West) Vacuum Pump, Seal
_______ _______
3. East Recovery Compressor Seal
_______ _______
4. West Recovery Compressor Seal
, _______
5. North Recovery Compressor Seal
_______
Page 4 of 6 Cone. Cone.
ppm ppm
^ Cone,
ppm
ABD00076890
Old Reactor Area -.Reactor Module IE of Reactors
.1 East !Steam Jet Exhaust
2. West :Steam Jet Exhaust 3. D-300 Condenser Top Seal 4. D-300 Condenser Bottom Seal 5. D-300 Lenape Manway Seal 6. D-40O Condenser Top Seal 7. D-400 Condenser Bottom Seal 8. D-400 Lenape Manway Seal 9. D-500 Condenser Top Seal 10. D-500 Condenser Bottom Seal 11. D-500 Lenape Manway Seal 12. D-600 Condenser Top Seal 13. D-600 Condenser Bottom Seal 14. D-600 Lenape Manway Seal Bottom of Reactors
.1 D-300 Dump Valve
2. D-400 Dump Valve 3. D-500 Dump Valve 4. D-600 Dump Valve 5. VCM Charge Manifold Line (Area)
Cone. . Cone. ppnr ppm
Page 5 of 6
Cone. ppm
Cone., ppm
Cone, ppm
ABD00076891
Page 6 of 6
VCM Tank Farm 1. #1 Compressor (North) 2. #2 Compressor 3. #3 Compressor 4. #4 Compressor (South) 5. Railcar Connection, North Station 6. Railcar Connection, Middle Station 7. Railcar Connection, South Station 8. North VCM Charge Pump (S. of Sphere) 9. South VCM Charge Pump (S. of Sphere)
Concl ppm
_______ _______ _______ _______ _______ _______ _______ _______ _______
Cone. ppm
_______ _______ _______ _______ _______ _______
Cone. ppm
_______
Cone. ppm
Cone, ppm
ABD00076892
9
LEAK DETECTION REPORTING FORM FOR PORTABLE DETECTOR
z
Cl Cl
Q
UJ d
kH
<
CU UJ d
U.
<
UJ
a z&Cu.
d UJ
E-* U. Q<
UJ d d UJ
CJ
OS
CJ z
o O d cj - Z
UHUoJJ
UaaJ
kH
a<c
cUJ
Z E-
zH
<
kH
O J O -J cu a
k- 3 j-. ZD UJ <
t- CO E- CO d uj
CJ UJ
cj uj
d
<
dd
<d
<d
uc
<d
UJJ
oHU e
d o C/5
uj o
-J t- . O
d .o
CO
UJ UJ UJ kH
Z E- E- < >
UJ UJ Z E-
kH >
H QkH
UJ UJ UJ d'
UJ
kH UJ >- d
a _J UJ CJ E- Q ao UJ
Q
UJ
5Q
Z o
kH
5
CJ
s
U3coJ <
&
o CL
U_Jj C<Oc
O
cu
Uo
u
CO 3
u<
Cu 3 C/5
Eu >--<
CO
z
z UJ E" Z
kH
z
cu a uj a 3 Z -3 UJ co < CO kH
< u- H uj H t-i u. H d d kH
< O uj I o o. > CO
rJ tO
in 'O
oo a o
ABD00076893
LEAK DETECTION AND ELIMINATION -PROGRAM .
ABERDEEN PVC PLANT
LOCATION OE CONTINUOUS MONITORING POINTS
The Plant has four Honeywell Model 1000 Process Gas Chromatographs. Each instrument monitors 10 locations for the presence of vinyl chloride. Thus a total of forty points are monitored.
The continuous vinyl chloride monitoring system serves three primary purposes.
1. Monitoring of plant process equipment to detect vinyl chloride leaks.
2. Protection of employees from exposure to vinyl chloride.
3. Identification of process conditions which precipitate vinyl chloride emissions.
Changing process conditions or newly identified needs may suggest moving the monitoring points. The system used is adaptable to such changes. The plant reserves the right to modify the location of the forty points. The locations will be changed and the location records revised only as desired to optimize the needs of the three defined functions.
The precise location of the forty points is indicated on the attached Table 16. The location is also indicated on the attached Plant Plot Layout drawing.
ABD00076894
LEAK -DETECTION AND ELIMINATION PROGRAM ABERDEEN PVC PLANT
ACTION PLAN FOR LEAK ELIMINATION
The presence of a leak, as detected by the Honeywell Process Chromato
graphs will activate a reporting, leak search, leak identification,
and leak elimination program.
'
If a vinyl chloride process leak occurs, it will be detected and defined
as indicated in Section I. The existence of the leak will generate the following sequence of events.
1. 'The Operator will record the items (1), (2), (3), (-4) and (5) on the attached Leak Detection Reporting Form.
2. The Lead Operator in the area will contact the Shift Supervisor and give him the information that a leak exists.
3. The Operator or Shift Supervisor will then use the portable hydro carbon detector to go to the area to pinpoint the location of the leak. This person is to wear appropriate personnel protective equipment consistent with plant safety rules.
4. When the leak is pinpointed, immediate action will be taken, if possible, to eliminate the source of the emission. In any case, item (6) is to be completed on the Leak Detection Reporting Form.
5. If the leak is eliminated, the Shift Supervisor is to verify this at item (7). The Operations Supervisor should verify this in item (9). The Operator and Shift Supervisor should initial the report under (10). The report is then to be properly filed.
6. If the Operator cannot eliminate the leak source, the Shift Supervisor is to inspect the problem and fill out item (7). If the Shift Supervisor correctes the leak, the Operations Supervisor is to verify under item (9) and the report is to be filed.
7. . If the Shift Supervisor cannot correct the leak, he is still to complete item (7). He is then to generate an Emergency (E) Maintenance Work Order to correct the leak source.
8. When the maintenance work is completed, the Shift Supervisor is to verify under item (8).
9. The Operations Supervisor is to verify this in item (9). The report is to he initialed in item (10) and then appropriately filed.
10. Leak Reporting Forms are to be available at each Honeywell Chromatograph location. The reports when completed will be kept by the Process Engineering Technician. These files will be kept for a minimum of two years.
ABD00076895
LEAK DETECTION REPORTING FORM PORTABLE DETECTOR READING AFTER LEAK
REPAIRED, PPM
OC/5
Z KH
uo<c
o
o
2
3
con
B
Cl.
B
z o
oec
H
aoS
< a:
oZ
E-3
as ' .HM B
os
Z Z E- O `
o
s
O -3 w3
tC--/5(
aE**
C/5 UJ
< OS
S CL CL e-> c/5 >
u
CL CJ UJ os
UJ
_J _J tu
>-
u < os UJ '
)z--1 os
o
*
C/5 C3 Z HH
os
o
E<
CL 3 C/5
H
UJ U3 UJ
a OS
UL
CJ
z
UJ
Ez--
Z o X
H < o
MS E-
zo
s
< UJ OS
UJ
CoL
hH B C/5
hH
H (N to
m '6
00 CTl
OC 8i--i >
UJ
a3.
C/5
Eu.-
O
ABD00076896
#
LEAK DETECTION AND ELIMINATION PROGRAM
ABERDEEN PVC PLANT
DEFINITION OF A LEAK
The plant defines a leak as:
A condition which results in the emission of persistent amounts of vinyl chloride.
The Leak Detection and Leak Elimination Action Plan will be activated by three successive readings in excess of the 5.0 ppm action level. Three successive readings are evidence of persistent quantities of vinyl chloride and evidence of a leak.
The value at each monitoring point will be observed by the operator on the existing recorders and will be transcribed onto a data sheet once per hour. This will be used by management to insure leaks are being properly detected and corrected by plant personnel. The once per hour reading will be recorded on forms similar to the one attached.
Table 17 indicates the 5.0 ppm action levels for.the Leak Detection Program. Also shown are typical background levels at each location.
ABD00076897
&N
2 3
>-
CC
o
K
<
CC
o
s
<
CO UJ
< cc
h
CO
SO
d O co cj
% 3-
S_ *
oc
a g3
j_
5i o
|Iff O
"- m>
ao
i g3
11 E
> O O uo o
3? OM
*2 53
S a 5 CD
>2 TO3 'wC
6>
*-cn^ru>(ora90)o
H
Ul Ui - 8
X
CO
CD
o - = ^= ~
CC
<
u.
~ 55
" oH . =
' . ?> w
J* 2
ccO
CD
*
z
4 H
S <
Ui
o
>
(S H
CO
?.'
= u--. m < a--.
--US
"2o
2
`Sg
SoS
d>
S o
u_
<3
0 <0
3<-
o o
-se
0)
-
je
4>
vO3 ,=tJ63-"o^QCSSu
to 5 2 u cl
o2
>
5 >c C>. SS
a: q o a:
o -CNn^ifiosoowo
H
z
o
CO
ss
i*
s
z>
--Io i" n3 M
zo
4
CC
o
oO
wiiSjS,
a|? c o
|^S205
H
ozo
o
<
s
UJ <
CC
UJ
cc
UJ
<0
>o- fVrN-->O-
Vr*~- w_<C* LhCr-o_-
Oo *5
o
82
fHooflohaoioroKaofZcO^-arooEol>S^^BgSs.
2
*-fNnWtfttOr--C0Ol
o
>
UJ
3 co
Q2
OO P 4 O
o
4 UJ CC H CO
O
ej C/5
a E
(T c>o
I*-?oO
Or C*
8eft
-t Q d
u _ t ..
o a o
w SC ru
i'D> vT5
3E
Z w0
3
a vo
joj
" ow
ao_
w
2 CD K ffl h
J2
vNCOVIftCOr-.COCTO^ c
ABD00076898
TABLE 16
LOCATION OF CONTINUOUS MONITORING POINTS
ABERDEEN PVC PLANT
HONEYWELL CHROMATOGRAPH NO. 1
POINT NO,
LOCATION
1-A 1-B 1-C 1-D 1-E 1-F 1-G 1-H 1-1 1-J
401 Receiver Recovery System, 2nd Floor Dump Sweco 3rd Recovery Compressor Incinerator Off Gas Stack Middle Receiver Bottom of D-300 Top of D-400 Bottom of D-500 Top of D-600
HONEYWELL CHROMATOGRAPH NO. 2
POINT NO.
LOCATION
2-A South Unloading Station 2-B Middle Unloading Station 2-C North Unloading Station
2-D Compressor Shed 2-E Process Control Room 2-F VCM Storage Tanks - 201 $ 202 2-G Respirator Air Filter Outlet 2-H Dryer Line #6, Blend Tanks, 1st Floor 2-1 Dryer Line #6, Activator, 1st Floor 2-J Respirator Air Filter Inlet
HONEYWELL CHROMATOGRAPH NO. 3
POINT NO.
LOCATION
3-A 3-B 3-C '3-D 3-E 3-F 3-G 3-H 3-1 3-J
.Bottom of 741 Top of 742 Bottom of 743 Top of 744 North of Dump Sweco K.O. Tank Pumps, 1st Floor Bottom of Middle VCM Receiver Recovered VCM Transfer Pumps Middle Recovery Compressor, 2nd Floor Fresh VCM Charge Pumps
ABD00076899 TABLE 16 (Cont.)
HONEYWELL CHROMATOGRAPH NO. 4
POINT NO.
4-A 4-B 4-C 4-D 4-E 4-F 4-G 4-H 4-1 4-J
LOCATION
Vinyl Bagging Machines Quality Control Lab Compound Line I Blenders Compound Line III Blenders Development Lab (GC) Welex Deck Dry Blend Control Room V-10 Blend Tank Middle of East Warehouse Shrink Wrap Unit (East Warehouse)
A
ACTION LEVELS FOR LEAK DETECTION REPORTING AND ELIMINATION PROGRAM
ABD00076900
HH [>U &S.
CJ IU.O. <X
OOOO 10 10 10 1-0
ooooo lO lO LO.lO lO
oooooooooo l/lLOLDlfiLOLOLOl/IL/lLA
az OCtf XU4 -S o > aa.
Cfi
oooooooooo
^ LO
lO >--< r-H rH LO *"4 LO
vvvvvvvvvv
LOlOlOLOOlOlolOlOlO OOOOioOOOOO VVVVVVVVVV
0
o
o0
H 03
u. O P
w co
TJ in
C 0 in
CM 0 co
CLCJ
PE
s O <P 0 o o
0 CJ 0h 0 o o
P o > tol O inl O
OX
0 0 0QoQO
0 CO o 0 o 0 'D" \3
zo
CO 0 > p 0 04 1 04 i X 0 O CD co o a O a
2? 0 3 o 0
1--4 Q 0 CO 0 0 0 s 04 E 04
E- > > CO c pH o o O o
<
o Oh H T3 p
p
CJ 0 E -o 0 *3 p CL p a.
3
O0 't CO
3 0 c H o o o a to K-1 2 CO E- CQ
o
0
oO 0
- o u. o
u.
CN .74 p
NO pt) > -(HA
pH in u7 p p: p
c eod o H o
i--4
3 o
c CD
P O
o E- P
H p H
04 0
CD
P <0 P E 0 -a >
3 p CJ
O l p c P
p CO P CO CO
o CO
i-H 0 in H pH
P 0
to p: u. co <c
to c 00 P pH d s rt c 0 O 3 0
Pp
r4 CD
3 o
H 3
a CO
0
Pc
s0
<H<
\D
avOc
O pH O 0 o 04 0 0 0 0
c cpH o CJ 3 o c
o
c o c in
0 P *H rt p
o o in in o 3 X X 3
0 0 in p 0
0
X X 0 0 CO H 0 0 r4
p *3 3 3
p 0
CL E
0 O
s
CL 0 0 CL in X X in
zo H o O 0 CJ 0 0 0 0
CO 2
CJ CL > CO Q a co
E-- *Z--i
Ci fliUiQiWi fciOilHi Oi N<I CNI O(NI U(NIQNIW(INUNI.UMI IhNI NOI
o
a.
A
ACTION LEVEL (PPM)
r ABD00076901
oooooooooo
m ui'i/i ia i/i to ia ia in tn
oooooooooo
Lni/ji/iLninmi/iuii/ii/)
oooooooooo .<-4' *-4 ^4 p-4 44 i--I i-4 44 44 44 vvvvvvvvvv
m in in in in in o i/i in O O O O O O O 44 O O. vvvvvvvvvv
BACKGROUND
LEVEL (PPM)
uoo
*(GN3
tO
rv r<N
Coo43O)
oo iu-4. p
(A
oE3 .-**
O CA
X>0oo3H(4)
3
44
3
*3 H s
(A
aE3&.
3f4
44
(C3A ES C>J
s
O4i
(A
(3UA C0A.
COl. a.
CJ
X*4
3on
u
>oo0)
3
xcj
44 o
E
O P
CPOO
Tj- 44
r* o r*-. 44
4HoC04LC,-04PoE-Q) 4hoQO4-
O
z4Uo-)
Pc
3 E-1
.
o
44 O
EO
P
Po
CO
oo ou >ooc
Q
o X
o
44 *3
XH3
s
u> x u(ouA.
3
(3OA
X3
X2(0cHo3A) on
X XCO 1o-4
u
e H oo
pco
oo CJ
CA
"u3ca3
CQ H4
3c
H P
ca
-^e33u34
CQ
14--444
3
C
H P
cj wCJ X 3
p
p cp 3O
XoO o P(4 co
CJ *3
PC
3 H
"3 C
3 O(3A X23U3 P
IA 3
44
23(4 P UC3JA
P
r4 c x
Cl. 3 U
CO *3 *3 3 C 3 O 2
03 -4
>c*> >H
XC
*cp"C3HHOrc2O30Ej,
G
3O OEC4'
CJ
-oC- .OX
3>
3 -4
a3 s3
3 03
U Q
r*H CQ
O
1 >
3
3 "3
sH
P crt
u
X CO
334-B 44-
< NO
CQ K>
CJ |
to
a1 to
UJ 1
to
u1to
C11J to
X11 to
f* ft
*-5 A
<i rr
CJ
Q1
U1J
u. 11
CJ
X11
4
-5
LOCATION
POINT NO.
ABD00076902
p0fOiMNFIDENM
61,65(b)(8)(Continued) (vi) It contains a definition of leak which is acceptable when compared
with the background concentrations of vinyl chloride in the areas of the
plant to be monitored by the vinyl chloride monitoring system. Measurements of background concentrations of vinyl chloride in the areas of the plant to be monitored by the vinyl chloride monitoring system are to be Included with the description of the program. The definition of leak for a given plant may vary among the different areas within the plant and is also to change over time as background concentrations in the plant are reduced.
STATUS OF COMPLIANCE
The Aberdeen Chemical Plant submitted the Leak Detection and Elimination
Plan to Region IV of the Environmental Protection Agency on May 16, 1977.
Approval of this plan was granted on June 30, 1977. Since this time, the
plan has been revised and approved several times. These are listed below.
DATE OF SUBMITTAL OF PROPOSED REVISION
REQUEST SUBMITTED TO
DATE OF APPROVAL FROM AGENCY
April 6, 1979
EPA Region IV
July 6, 1979
November 21, 1979
EPA Region IV
December 17, 1979
August 15, 1983
Mississippi DNR
August 23, 1983
The plant follows the approved Leak Detection and Elimination Plan. There are five minor modifications to the previously approved Leak Detection and Elimination Plan which the plant is requesting the EPA to approve at this time. These are as follows: 1) The analysis cycle time for the Honeywell chromatographic column has been changed from 60 seconds to 70 seconds to improve sample resolution. 2) The oven temperature for the Honeywell chromatographic column has been changed from 60C to 80C to improve sample resolution. 3) The automatic sample injection of standard gases feature on the gas chromotographs is no longer used. Manual standard gas injection is used. 4) When the 1979 plan was submitted, there were three HNU portable hydrocarbon detectors in the plant. The calibration procedure called for storing checked and calibrated machines in the guardhouse and one for use
-31-
ABD00076903
61.65(b)(8)(Continued)
UONFIDENTIAL
STATUS OF COMPLIANCE(Continued)
In the vinyl area. Once per week, the monitor In the vinyl area was taken
to the guardhouse and exchanged for a machine that had been checked and
calibrated. Now the plant has eight HNU portable hydrocarbon detectors in
the plant. The current calibration frequency for each dectector is at least
once every two weeks regardless of the number of detectors In the plant.
5) The leak detection form that was used In the 1979 submittal has been
changed to allow recording more than one leak per page. A copy of the form
Is attached: (Attachment B)
ADDITIONAL MEASURES TO BE COMPLETED Retraining on the Leak Detection and Elimination Flan will be conducted on an annual basis. Initial retraining required by this Plan will be completed within 270 days of Plan approval.
To further reduce the possibility of leaks, the plant proposes to install a new type of valve position indicator on selected valves. The current position indicator is mounted on the valve operator, not the valve. Thus if the coupling between the operator and the valve fails, a false indication of the valve position may be provided. In addition, the magnetic portion of the valve indicator assembly is not securely fastened to the assembly. The plant has consulted with a limit switch manufacturer and in conjunction with the manufacturer developed a new design whereby a standard proximity switch can be used to provide positive indication of the valve stem rather than the actuator position. The design involves the use of a standard proximity switch with a custom designed indicator bracket. One unit has been tested in the shop and then installed in the field. The performance
-32-
ABD00076904
*
61.65(b)(8)(Continued)
ADDITIONAL MEANSURES TO BE COMPLETED (Continued)
on this test unit has been favorable. This new type valve Indicator will be
installed on 70 valves that have been Identified as having the highest
potential for causing a leak to the atmosphere based on the type of
service or history of leaks.
This project will greatly reduce the possibility of accidental
leakage to the atmosphere due to valve position switch indicator
failure.
The definitive process design for this project will be completed by
sixty days after Plan approval. Operations will be informed of the
operational aspects of this project prior to installation.
Installation of this project will be complete within 300 days after Plan
approval.
61.65 (b) Fugitive emission sources. (9) Inprocess wastewater: Vinyl chloride emissions to the atmosphere
from inprocess wastewater are to be reduced as follows: (I) The concentration of vinyl chloride in each inprocess wastewater
stream containing greater than 10 ppm vinyl chloride measured immediately as it leaves a piece of equipment and before being mixed with any other lnprocess wastewater stream is to be reduced to no more than 10 ppm by weight before being mixed with any other lnprocess wastewater which contains less than 10 ppm vinyl chloride; before being exposed to the atmosphere; before being discharged to a wastewater treatment process; or before being discharged untreated as a wastewater. This paragraph does apply to water which is used to displace vinyl chloride from equipment before it is opened to the atmosphere in accordance with Section 61.64(a)(2) or paragraph (b)(6) of this section, but does not apply to water which is used to wash out equipment after the equipment has already been opened to the atmosphere in accordance with Section 61.64(a)(2) or paragraph (b)(6) of this section.
-33-