Document RaD8gYqdXgLBdrxdqn7x5Npxk
[
H.W.: various
ISSUED: 2/15/8$
OSHA: 0.5 asbestos fibers (> 5 t*a long)/ml
PROPERTIES: solid,
NIOSH: 0.1 asbestos f/mL [1]; 3 glass fibers (>10 pm x <3.5 pm)/mL [2]
fibrous
ACGIH: 0.2 crocidolite; 0.5 amosite; 2 chrysotile and other asbestos, f/mL
SYNONYMS: asbestos (actinolite [CAS *77538-66-4], grunerite (amosite) [CAS *12172-73-5], anthophyl1ite [CAS *77536-67-5], chrysotile [CAS *12001-29-5], crocidolite [CAS *12001-28-4], tremolite [CAS *77536-68-6]); fibrous glass.
SAMPLING
MEASUREMENT
SAMPLER: FILTER (0.8-1.2 pm cellulose ester mentorane, 25-am diameter)
TECHNIQUE: MICROSCOPY, PHASE CONTRAST ANALYTE: fibers (manual count)
FL0U RATE*: > 0.5 L/min
SAMPLE PREPARATION: acetone/triacetin method
VOL-MIN*: 400 L 8 0.1 fiber/mL -MAX*: 1920 L 8 0.1 fiber/mL
Adjust for 100 to 1300 fibers/me3 (step 4)
SHIPMENT: routine
SAMPLE STABILITY: indefinite
BLANKS: 101 of samples (minim* 2) [3]
ACCURACY
~
RANGE STU0IED: 80 to 100 fibers counted
BIAS: see EVALUATION OP METHOD
OVERALL PRECISION (s^: 0.115 to 0.13 [3] (A Rules)
COUNTING RULES: Set A (P4CAH 239 [3,4]) or Set B (modified CRS [5])
EQUIPMENT: 1. phase-contrast microscope 2. Walton-Beckett graticule (100 pm field diameter): A Rules use G-22; B Rules use Type G-24 3. phase-shift test slide (HSE/NPl)
CALIBRATION: phase-shift detection limit about 3 degrees [7]
RANGE: 100. to 1300 fibers/me3 filter area [6]
ESTIMATED LOO: 7 fibers/me3 filter area
PRECISION: 0.10 to 0.12 [3] (A Rules)
APPLICABILITY: The working range is 0.02 fiber/mL (1920-L air saeple) to 1.25 fibers/mL (400-L
air sanple). The method gives an index of airborne asbestos fibers but may be rred
for other materials such as fibrous glass by inserting suitable parameters into the counting ....
rules.- The method does not differentiate between asbestos and other fibers. Asbestos fibers .
less than ca. 0.2S urn diameter will not be detected by this method [71.
INTERFERENCES:-* Any other airborne fiber may Interfere since all particles meeting the counting
criteria are counted. Chain-like particles may appear fibrous. High levels of non-fibrous dust
- .particles may obscure fibers in the field of view and raise the detection limit.
OTHER METHODS: This method introduces changes for ioproved sensitivity and reproducibility and
--replaces P4CAH 239 f3.4l.
--
2/15/84
7400-1
PLAINTIFF'S exhibit AL-1003
FIBERS
METHOO: 7400
REAGENTS:
EQUIPMENT:
1. Acetone.*
1. Sampler: field monitor, 2S mm, three-piece cassette
2. Triacetin (glycerol triacetate),
with 50-nw extension cowl with cellulose ester filter,
reagent grade.
0.8 to 1.2-um pore size and backup pad.
NOTE: Analyze representative filters for fiber
See Special Precautions.
background before use and discard the filter lot if
more than S fibers/100 fields are found.
2. Persona) sanpling purp, > O.S l/min (see step 4 for
flow rate), with flexible connecting tubing.
3. Microscope, phase contrast, with green or blue filter,
8 to TOX eyepiece, and 40 to 45X phase objective (total
magnification ca. 400X); nunerical aperture = 0.65 to
0.75.
4. Slides, glass, single-frosted, pre-cleaned, 25 x 75 me.
5. Cover slips, 25 x 25 ora, no. 1-1/2, unless otherwise
specified by microscope manufacturer.
6. Knife, #10 surgical steel, curved blade.
7. Tweezers.
8. Flask, Guth-type, insulated neck, 250 to 500 mt (with
single-holed rubber stopper and elbow-jointed glass
tubing, 16 to 22 cm long).
9. Hotplate, spark-free, stirring type; heating mantle; or
infrared lamp and magnetic stirrer.
10. Syringe, hypodermic, with 22-gauge needle.
11. Graticule, Ualton-Beckett type with 100 pm diameter
circular field at the specimen plane (area - 0.0078S
me*) (Type G-22 for A Rules; Type G-24 for B Rules).
Available from Graticules Ltd., Morley Road, Tonbridge
TN9 1RN, Kent, England (Telephone 011-44-732-359061).
NOTE: The graticule is custom-made for each microscope.
Specify disc diameter needed to fit exactly the
ocular of the microscope and the diameter (ma) of
the circular counting area (see step 11).
12. HSE/NPl phase contrast test slide. Nark II. Available
from PTR Optics ltd., 145 Newton Street, Waltham, MA
02154 (Telephone (617) 891-6000).
13. Telescope, ocular phase-ring centering.
14. Stage micrometer (0.01 ma divisions).
SPECIAL PRECAUTIONS: Acetone is an extremely flanxnable liquid and precautions must be taken not to ignite it. Heating of acetone must be done in a ventilated laboratory fune hood using a flameless, spark-free heat source.* 1 2
SAMPLING: 1. Calibrate-each personal sampling punp with a representative sampler in line [3]. 2. Fasten the sampler to the worker's lapel as close as possible to the worker's mouth. Remove the top cover from the end of the cowl extension (open face) and orient face down. Wrap the joint between the extender and monitor body with shrink tape to prevent air leaks.
2/15/84
7400-2
NETHOO: 7400
FIBERS
3. Submit at least two field blanks (or 101 of the total sanples, whichever is greater) for each set of samples. Remove the caps from the field blank cassettes and store the caps and cassettes in a clean area (bag or box) during the sanpling period. Replace the caps in the cassettes when sampling is completed.
4. Sanple at O.S l/min or greater [8]. -Oo not exceed 1 eg total dust loading on the filter. Adjust sampling flow rate, Q (l/mki), and time to produce a fiber density, E (fibers/mm2). of 100 to 1300 fibers/me2 [3.85 104 to 5 10* fibers per 25-ma filter with effective collection area (Ag* 385 me2)] for optima counting precision (sae step 21). Calculate the minima sapling time, t^f, (min), at the action level (one-half the current standard), l (fibers/et), of the fibrous aerosol being saepled:
tmin
(Ae)(E) (Q) (l) 10*
5. Remove the field monitor at the end of sampling, replace the plastic top cover and small end caps, and store the monitor.
6. Ship the saoples in a rigid container with sufficient packing material to prevent jostling or damage. NOTE: Oo not use polystyrene foam in the shipping container because of electrostatic forces
which may cause fiber loss from the saapler filter.
SAHPIE PREPARATION: NOTE: The object is to produce samples with a smooth (non-grainy) background in a mediua with a
refractive index equal to or less than-1.46. The method below- collapses the filter for easier focusing and producas permanent-mounts which are useful for quality control and interlaboratory comparison. Other mounting techniques meeting the above criteria may also be used (e.g., the non-permanent field mounting technique used in P6CAH 239 [1,3,4]). 7. Ensure that the glass slides and cover slips are free of dust and fibers. 8. Place 40 to 60 mL of acetone into a Guth-type flask. Stopper the flask with a single-hole rubber stopper through which a glass tube extends 5 to 8 cm into the flask. The portion of the glass tube which exits the top of the stopper (8 to 10 at) is bent downward in an elbow which makes an angle of 20 to 30* with the horizontal. 9. Place the flask on a stirring hotplate or wrap in a heating mantle. Heat the acetone gradually to its boiling teeperature (ca. 58 *C). CAUTION: The acetone vapor must be generated in a ventilated fuar hood away from all open
flames and spark sources. Alternate heating methods can be used, providing no open flame or sparks are present. 10. Mount either the whole sample filter or a wedge cut frae the sample filter on a clean glass slide. a. Cut wedges of ca. 2Si of the filter area with a curved blade steel surgical knife using a rocking motion to prevent tearing. b. Place the filter or wedge, dust side up, on the slide. Static electricity will usually keep the filter on the slide until it is cleared. c. Hold the glass slide supporting the filter approximately 1 to 2 cm from the glass tube port where the acetone vapor is escaping frae the heated flask. The acetone vapor stream-should cause a condensation spot on the glass slide ca. 2 to 3 cm in diameter. Hove the glass slide gently in the vapor stream. The filter should clear In 2 to S sec. If the filter curls, distorts or is otherwise rendered unusable, the vapor stream is probably not strong enough. Periodically wipe the outlet port with tissue to prevent liquid acetone dripping onto the filter.
2/15/84
7400-3
FIBERS
HETHOO: 7400
d. Using the hypodermic syringe with a 22-gauge needle, place 1 to 2 drops of triacetin on the filter. Gently lower a clean 2S-ia square cover slip down onto the filter at a slight angle to reduce the possibility of forming bubbles. If too many bubbles form or the amount of triacetin is insufficient, the cover slip may became detached within a few hours.
e. Glue the edges of the cover slip to the glass slide using a lacquer or nail polish [9]. NOTE: If clearing is slow, the slide preparation may be heated on a hotplate (surface temperature 50 *C) for 15 min to hasten clearing. Counting may proceed imediately after clearing and mounting are coopleted.
CALIBRATION ANO QUALITY CONTROL: 11. Calibration of the Ualton-Beckett graticule. The diameter, de (m), of the circular
counting area and the disc diameter must be specified when ordering the graticule. a. Insert any available graticule into the eyepiece and focus so that the graticule lines
are sharp and clear. b. Set the appropriate interpupillary distance and, if applicable, reset the binocular head
adjustment so that the magnification remains constant. c. Install the 40 to ASX phase objective. d. Place a stage micrometer on the microscope object stage and focus the microscope on the
graduated lines. e. Measure the magnified grid length, lg (urn), using the stage micrometer.
f. Ramove the graticule from the microscope and measure its actual grid length, La (am). This can best be accomplished by using a stage fitted with verniers.
g. Calculate the circle diameter, dc OaeK for the Ualton-Beckett graticule:
Exapple: If lg - 108 vie, La 2.93 am and 0 100 tea, then dg 2.71 amt. h. Check the field diaamter, D(accaptable range 100 pm t 2 tmi) with a stage micrometer
upon receipt of the graticule from the manufacturer. Determine field area (me?). 12. Microscope adjustments. Follow the manufacturer's instructions and also the following:
a. Adjust the light source for even illueination across the field of view at the condenser iris. NOTE: Kohler illuaination is preferred, where available.
b. Focus on the particulate material to be examined. c. .Make sure that the field iris is in focus, centered on the saaple and open only enough
to fully illimrinate the field of view. d. Use the telescope ocular supplied by the manufacturer to ensure that the phase rings
(annular diaphragm and phase-shifting elements) are concentric. 13. Check the phase-shift detection limit of the microscope periodically.
a. Remove the HSE/NPL phase contrast test slide from its shipping container and center it under the phase objective.
b. Bring the blocks of grooved lines into focus. NOTE: The slide consists of seven sets of grooves (ca. 20 grooves to each block) in descending order of visibility from sets 1 to 7. The requirements for asbestos counting are that .the microscope optics must resolve the grooved lines in set 3 completely, although they may appear scemwhat faint, and that the grooved lines in sets 6 and 7 must be invisible. Sets 4 and S must be at least partially visible but may vary slightly in visibility between microscopes. A microscope which fails to meet these requirements has either too low or too high a resolution to be used for asbestos counting.
2/15/84
7400-4
METHOD: 7400
FIBERS
c. If the image quality deteriorates, clean the microscope optics and if the problem persists, consult the microscope manufacturer.
14. Quality control of fiber counts. a. Prepare and count field blanks along with the field sanples. Report the counts on each blank. Calculate the maul of the field blank counts and subtract this value from each sample count before reporting tho results. NOTE 1: The identity of the blank filters should be unknown to the counter until all counts have been completed. NOTE 2: If a field blank yields fiber counts greater than 7 fibers/100 fields, report possible contamination of the samples. b. Perform blind recounts by the same counter on 101 of filters counted (slides relabeled by a person other than the counter).
15. Use the following test to determine whether a pair of counts on the same filter should be rejected because of possible bias. This statistic estimates the counting repeatability at the 951 confidence level. .Discard the saeple if the difference between the two counts exceeds 2.77 (FlSy., where F - average of the two fiber counts and sr relative standard deviation, which should be derived by each laboratory based on historical in-house data. NOTE: If a pair of counts is rejected as a result of this test, recount the remaining samples in the set and test the new counts against the first counts. Discard all
rejected paired counts. 16. Enroll each new counter in a training course which compares performance of counters on a
variety of saeples using this procedure. NOTE: To ensure good reproducibility, ail laboratories engaged in asbestos counting should
participate in an asbestos proficiency testing, program such as the NIOSH Proficiency Analytical Testing (PAT) Program and routinely participate with other asbestos fiber counting laboratories in the exchange of field samples to aspare performance of counters.
MEASUREMENT: 17. Place the slide on the mechanical stag*of the- calibrated microscope with the center of the
filter under the objective lens. Focus the microscope on the plane of the filter. 18. Regularly check phase-ring alignment and Kohler illumination [7]. 19. Select one of the following sets of counting rules:
NOTE: The two sets of rules have been demonstrated to produce equivalent mean counts on a variety of asbestos sample types [5] and must be strictly followed in order to obtain valid results. No hybridizing of the two sets of rules is permitted. The calibration of the aricroscope with the HSE/NPl test slide determines the minimus detectable fiber diaamter (ca. 0.2S pm),
a. A Rules (seem as P6GAH 239 rules [1,3,4]). NOTE: The A Rules are required for monitoring asbestos for compliance purposes under 09* or NIOSH standards. 1. Count only fibers longer than S ten* Measure the length of curved fibers along the curve. 2. Count only fibers with a length-to-width ratio equal to or greater than 3:1. 3. For fibers which cross the boundary of the graticule field, do the following: a. Count any fiber longer than S pa which lies entirely within the graticule-area. b. Count as 1/2 fiber any fiber with only one end lying within the graticule area. c. Oo not count any fiber which crosses the graticule boundary more than once.
d. Reject and do not count all other fibers.
2/1S/84
7400-S
FIBERS
WETHOO: 7400
4. Count bundles of fibers as one fiber unless individual fibers can be identified by observing both ends of a fiber.
5. Count enough graticule fields to yield 100 fibers. Count a minimus of 20 fields. Stop at 100 fields regardless of fiber count.
b. B Rules NOTE: The B Rules are preferred analytically because of their demonstrated ability to improve the reproducibility of fiber counts [51. 1. Count only ends of fibers. Each fiber must be longer than S i and less than 3 i*i diameter. 2. Count only ends of fibers with a length-to-width ratio equal to or greater than 5:1. 3. Count each fiber end which falls within the graticule area as one end, provided that the fiber meets rules b.l and b.2. 4. Count visibly free ends which meet rules b.l and b.2 when the fiber appears to be attached to another particle, regardless of the size of the other particle. 5. Count the free ends of fibers emanating from large cl nips and bundles up to a maxima* of 10 ends (5 fibers), provided that each segaent meets rules b.l and b.2. 6. Count enough graticule fields to yield 200 ends. Count a minimua of 20 fields. Stop at 100 fields, regardless of the fiber count. 7. Oivide the total end count by 2 to yield fiber count. NOTE: Split fibers will normally be counted as more than two ends if the free ends meet the rules b.l. and b.2.
20. Start counting from one end of the filter and progress along a radial line to the-other end, shift either up or down on the filter and continue in the* reverse direction [10]. Select fields randomly by looking away from the eyepiece briefly while advancing the mechanical stage. When an agglomerate covers ca. 1/6 or more- of the field of view, reject the field and select another. Do not report rejected fields in the lumber of total fields counted. NOTE: When counting a field, continuously scan a range of focal planes by moving the fine focus knob to detect very fine fibers which have become embedded in the filter. The small-diameter fibers will be very faint but are an important contribution to the total count.
CALCULATIONS: 21. Calculate and report fiber density on the filter, E (fibers/ma2), by dividing the total
fiber count, F, minus the mean field blank count, B, by the nunber of fields, n, and the field area, Af (0.00785 me3 for a properly calibrated Walton-Oeckett graticule):
E
-11. (ni (Af)
fibers/ima2.
22. Calculate the concentration, C (fibers/mL), of fibers in the air volume sampled, V (L), using the effective collection area of the filter, Ac (385 me2 for a 25-ma filter):
C-
NOTE: Periodically check and adjust the value of Ac, if necessary.
EVALUATION OF METHOO: This method is a revision of NIOSH Method PSCMI 239 [1,3,4]. A suaaary of the revisions is as follows:
2/15/84
7400-6
METHOO: 7400
FIBERS
A. Sampling The change fro* a 37-ma to a 2S-am filter size was incorporated to improve sensitivity and reduce problems associated with non-uniform fiber loading reported on the 37-eia filters [10]. The change in flow rates allows for 2 mJ full-shift saeples to be taken, providing that the filter is not overloaded with non-fibrous particulates. The collection efficiency of the sampler is not affected by changes in flow rate in the range 0.S to 16 l/min [8].
8. Saaple Preparation Technique The acetone vapor-triacetin preparation technique has been incorporated in the method as a faster, more permanent mounting technique than the dimethyl phthalate/diethyl oxalate method of P4CAH 239 [1,3,4,11].
C. Measurement 1. The inclusion of the Walton-Backett graticule in the method was made- to standardize the
field area observed through the eyepiece [6,11]. 2. The introduction of the HSE/NPl test slide was made to standardize microscope optics for
sensitivity to fiber diameter [7,11]. 3. A recent international collaborative study involved 16 laboratories using prepared slides
from the asbestos, cement, milling, mining, textile, and friction material industries [5]. The relative levels of count by different counting rules were:
Sample Tvoe
Nimber of Sancles
Aspect Ratio >3:1 AIA Mod. CRS*
Aspect Ratio >5:1 AIA Mod. CRS*
Mining Milling Asbestos Caemnt Textile Chrysotile Friction Material Others (Insulation, Aaosite)
10 .10 J4 10 10
6
100 100 100 100 100 100
TOTAL: 60 MEAN: ioo
127 74 112 84
146 90
109 89 130 87 127 92
Tis
92 95 137 99 116 118
no
*Arithmetlc means of counts made bv different laboratories relative to the AIA counts.
The modified CRS (MX05H 8) Rules were found to be more precise than the AIA (NIOSH A)* Rules. The ranges of relative standard deviations (s,J which varied with saaple type and laboratory were:
Intra laboratory-
*p Inter laboratory
Overall
AIA (NIOSH A Rules)* Modified CRS (NIOSH 8 Rules)
0.12 to 0.40 0.11 to 0.29
0.27 to 0.85 0.20 to 0.3S
0.46 0.2S
*Under AIA rules, only fibers having a diameter less than 3 pm are counted and fibers attached to particles larger than 3 pm are not counted. NIOSH A Rules are otherwise similar to the AIA rules.
The B Rules have also been favorably received by analysts as less ambiguous and sinpier to use; these rules also showed the least bias relative to AIA rules in the collaborative study. An independent NIOSH laboratory, study using amosite fibers reported a relative standard deviation, including within- and between-*ple
2/15/84
7400-7 -
FIBERS
METHOD: 7400
variability, of 0.157 for the 8 Rules [12]. Adding an estimated sampling punp error, s,., of 0.05 [13] to the within-sanple variability in this study results in an estimate of overall precision, sr, of 0.102 for the B Rules.
4. Because of past Inaccuracies associated with low fiber counts, the minima loading has been increased to 100 fibers/me? filter area (80 fibers total count). This level yields an overall sr * 0.13, as indicated in Figure 3 (revised) of PAGAN 239 [3,4] which
corresponds to a measurement sr - 0.12 after removal of pump error [13]. Similarly, at the maxima count of 100 fibers, overall Sf. 0.115 and measurement sr 0.10 are obtained. 0. Evaluation of the method using the A and B counting rules will proceed on a continuing basis through the NIOSH Proficiency Analytical Testing (PAT) Program. The new- PAT reporting form allows for reporting of results by either set of rules as of January, 1984.
REFERENCES: [1] Revised Recomeended Asbestos Standard, U.S. Department of Health and Hunan Services, Ptl. (NIOSH) 77-169 (1976). [2] Criteria for a Reccrnnended Standard...Occupational Exposure to Fibrous Glass, U.S. Department of Health and Hunan Services, Publ. (NIOSH) 77-152 (1977). [3] leidel, N. A., S. G. Bayer, R. 0. Zumwalde, and K. A. Busch. USPHS/NIOSH Hmtorane Filter Method for Evaluating Airborne Asbestos Fibers, U.S. Department of Health and Hunan Services. Publ. (NIOSH) 79-127 (1979). [4] NIOSH Manual of Analytical Methods, 2nd. ed., V. 1., P4CAM 239, U.S. Department of Health and Hunan Services, Publ. (NIOSH) 77-1S7-A (1977). [5] Crawford, N. P., H. 1. Thorpe, and W. Alexander. "A Conparisoa of the-Effects of Different Counting Rules and Aspect Ratios on the level and Reproducibility of Asbestos Fiber Counts," Part I: Effects on Level (Report No. TH/82/23), Part II: Effects on Reproducibility (Report No. TH/82/24), Institute of Occupational Medicine, Edinburgh, Scotland (Oecanber, 1982). [6] Walton, W. H. "The Nature, Hazards, and Assessment of Occupational Exposure to Airborne Asbestos Oust: A Review," Ann. Occuo. Hyg., 25, 115-247 (1982). [7] Roofcer, S. 0., N. P. Vaughn, and J. M. LcGuen. "On the Visibility of Fibers by Phase Contrast Microscopy," Anar. Ind. Hyg. Assoc. J., 43, 505-515 (1982). [8] Johnston, A. H., A. 0. Jonas, and J. H. Vincent. "The Influence of Externa.) Aerodynamic Factors on the Measurement of the Airborne Concentration of Asbestos Fibres by the Mmbrane Filter Method." Am. Ocorn. Hyg.. 25, 309-316 (1982). [9] Asbestos International Association, AIA Health and Safety Recamaended Technical Method #1 (RTHI). "Airborne Asbestos Fiber Concentrations at Workplaces by light Microscopy" (Haafcrane Filter Method), London (1979).
[10] Hooke, M. B., C. E. Feigley, and 0. A. Ludwig. "Interwedge Variation in the Membrane Filter Method for Airborne Asbestos Fibers," Amer. Ind. Hyg. Assoc. J.. 44, 542-546 (1983)
[11] Chatfield, E. J. Measurement of Asbestos Fibre Concentrations in Workplace Atmospheres. Royal Commission on Natters of Health and Safety Arising from +he Use of Asbestos in Ontario, Study No. 9, 180 Oundas Street west, 22nd Floor, Toronto, Ontario, CANADA MSG 128
[12] Taylor, 0. G., P. A. Baron, S. A. Shu 1man and J. W. Carter. "Identification and Counting of Asbestos Fibers," Am. Ind. Hyg. Assoc. J. 4S(2), 84 88 (1984).
[13] Busch, K. A. and 0. G. Taylor. "Statistical Protocol for the NIOSH Validation tests", Chemical Hazards in the Workplace, Measurement and Control, ACS Symposiue Series 149, American Chemical Society, Washington, DC (1981).
HE1HOO REVISED BY: James W. Carter, David G. Taylor, Ph.O., CIH, and Paul A. Baron, Ph.O., NIOSH/DPSE; based on the revised Method P4CAM 239 [1,3.4].
2/15/84
7400-8