Document RQOqZnb1DMaEpDbZ2oX3NQj8
Analysis of PCB's
in Transformer Oil
Pamela Edwards, Altex Scientific
Detection of part-per-million levels of polychlorinated biphenyls (PCB's) Is important because of the health hazards
these compounds present. PCB's were used lot many years as cooling-lubricants In transformers without knowledge of their toxic effects ; The Improper handling of PCB's during (hat time has contributed to a
situation where PCB's can be found in waste water, fish products, soil, and in oil which has replaced PCB's in transformers.
The analysis of PCB's in transformer oil depends in part on the limits the U.S. Environmental Protection Agency has es tablished At the time of this writing, the limits dictate that materials containing 50 ppm PCB's or more are subject to strict disposal regulations. Thus, a sensitive analytical method is necessary to deter mine the level ot PCB's. Several recent arti cles describing both gas chromatographic (QIC) and liquid chromatographic (HPLC) methods have been published. However, these reported procedures require time consuming sample preparation17J The HPLC method described in this report eliminates the need lor tedious sample preparation by using a detector which is specific tor chemical species such as PC8's. This method requires only filtration as semple preparation and is sensitive enough to detect 1-2 ppm of PCB's.
Materials and Method*. All HPLC work was done on a Model 332 system, con sisting of two 110A pumps, a 420 mi croprocessor/controller and e 210 injection valve with a 20 pi loop. The detector was a Tracor 965 photoconductivity detector with a zinc lamp. An Ultrasphere ODS (250 mm x 4.6 mm) column was used for analysis. A C-R1A integrator was used to plot and analyze detector output signal.
Acetonitrile (MeCN) and isopropanol (IPA) were obtained from Burdick and Jackson. A mobile phase of 56% IPA, 14% MeCN end 30% water was used tor ah analyses at a flow rate of 1 ml/min. PCB standards, clean transformer oil, and transformer oils contaminated with PCB's were obtained from industry. Sample prep aration consisted of filtering the oil sam ples through a 0.45 micron filter before injection.
ReeuNe and Discussion. A chromato gram of the PCB standard. Aroclor 1242. in MeCN Is shown in Figure 1. In Figure 2 a separation of the same standard at the same concentration, 100 ppm, dissolved in clean transloimer oil is shown. Because it Is the total amount of PCB's which is
6
;l
FiQt/fQ \ Photoconductivity Clvomatogiam ot standard AtocIO' 1242 In acalonilnla Sample volume was 20 /it and total concentialion ol PCB's was 100 ppm
Figuf 2 Photoconductivity ChfomatOQram ot Standard Aioclot 1242 made up in clean ttanslotmei oil to a total PCB concentration ot 100 ppm
IM1 FiQure 3 Photoconductivity chtomalogiam ot 20 clean ttansformei oil
fiQute 4 Photoconductivity ehiometogram of 20 *jl ol diity ttanslotmei oil containing moie than 100 ppm PCB's.
Figure 5 Photoconductivity chromatogram ol 20 pi of transloimer oil containing small amounts of PCB's
required, not amounts of individual compo
nents. complete separation of all compo
nents is not critical "Fingerprint" chromato grams of PCB's which can be quantified by a grouping system is adequate It is apparent that injecting the oil directly reduces resolution of individual compo nents but does not prevent quantifying total PCB's To insure that the transformer orl contained no substances which would show up using the photoconductivity detector, a blank injection was made No significant peaks are present in the chro matogram of clean oil shown in Figure 3 The solvent front may be due to dissolved gas in the sample
Several samples of dirty transformer oil were also analyzed A chromatogram of an oil sample which obviously contains more than 50 ppm PCB's and would be subject to the EPA regulations is shown in Figure 4 Quantification ol the PCB'S In this sample would have been quite simple using a grouped/external standard technique A dirty oil which does not contain high levels ol PCB's is shown in Figure 5. By compar ing this chromatogram to those in Figures 1 and 3, it Is clear that the peaks at 6.53 and 7.66 minutes represent small amounts 01 PCB's.
The photoconductivity detector is also sensitive to nitrosamines, sullonated organics, end polybrominated biphenyls which are on occasion also found in transformer oils. To eliminate concerns that these compounds might be coeluting with the PCB's. fractions were collected Irom a dirty oil analysis and were analyzed by GCMS. The results of the GC-MS indicated the presence of polynuclear aromatics and other organic compounds None of the lat ter compounds would have been detected by the photoconductivity detector.
The instrumentation and methods de scribed here lor analyzing PCB's in transformer oil have advantages compared to other methods in that very little sample preparation is needed for specific detec tion with more than adequate sensitivity (i 2 ppm) for EPA regulations
Acknowledgement. The author wishes to acknowledge Paul Kelly ol Finnegan Instruments lor the assistance with the GCMS analysis, and Michael Minner ol the City ol Tacoma Sewer Utility Division, Washington, for the PCB standards and samples
BIBLIOGRAPHY
i.Kiimisch, hm and (ngeonghlson, DM Anal Chem 52,11,1675-8 (1980)
2 Ogata JN.. Okun, JO, Hylm, JW. and Bevenue, A.. J Chromatogr 189(3). 425-7 (1980)
3 Belliardo, F, Gionchiglia, E and Nano. G M .J Uq Chromatogr 2(1). 77-83 (1979)