Document QXyGoq1MDbRBrwqGMm9Y9oaBE
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Quantitative Determination of PFOS, PFOSA, PFOSAA, POAA, PFHS,
M556 and M570 in Human Serum by LC/MS/MS
Assay Revalidation Addendum Report
Northwest Bioanalytical (NWB) A Division of NWT Inc. 1121 East 3900 South
Salt Lake City, UT 84124
PREPARED FOR:
3M Environmental Technology and Services 935 Bush Avenue St. Paul, MN 55133
AUTHOR:
/
(q('-,3,_d:9/c,._ L tz-yE-_'_zL
Connie O. Sakashita, B.S., NWB Project Manager
APPROVEDF_ P3,ELEASEBY: /X'-'__,_b r _,_..-p_.,_---
Patrick Bennett, M.S., M.B.A., NWB LaboratoryDirector
DATE: II'qVC'
DATE: _" 2 ,..-/
Page l
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
QUALITY ASSURANCE
STATEMENT
LABORATORY:
SPONSOR:
COMPOUND(S): NWB STUDY NUMBER: SPONSOR STUDY NUMBER:
Northwest Bioanalytical (NWB) A Division of NWT, Inc. 1121 East 3900 South
Salt Lake City, UT 84124
3M Environmental Technology 935 Bush Avenue
St Paul, MN 55133
Services
PFOS, PFOSA, PFOSAA, PFHS, M556 and M570 NWBS00-040
Method Validation
NWB STUDY TITLE:
Quantitative Determination of PFOS and Metabolites in Human Serum by LC/MS/MS, Assay Validation
The method validation study descn_oed in this report is not included within the definition of a GLP regulated nonclinical study. However, Northwest Bioanalytical conducts all projects in accordance with the guidelines of the U.S. FDA Good Laboratory Practice Regulations for Nonclinical Laboratory Studies (Title 21 CFR Part 58), the OECD Principles of Good Laboratory Practice and the Japanese MHW Good Laboratory Practice Standard Ordinance for Nonclinical Laboratory Studies on the Safety of Drugs (Ordinance No. 21, PAB Notification No. 424.) The following inspections were performed by the NWB QAU per SOP.
Inspection and Reporting Statement.
Inspection Date 15 May 2000 15-18 May 2000 07-18 Sep 2000
24 Jan 2000
19-30 Iu12001
05-07 Nov 2001
20 Nov 2001
Phase of Study Analytical Plan Assay Validation Report Draft/Raw Data (NWBR00-108) Final Report ('NWBR00-108) Report Drain/Raw Data (NWBR00-122) Report Drain/Raw Data (NWBR00-122) Final Report (NWBR00-122)
Date Inspection Report Issued To
NWB Project Manager iNWB Management
15 May 2000
31 May 2000
18 May 2000
31 May 2000
20 Sep 2000
29 Sep 2000
24 Jan 2000 30 Ju12001
31 Jan 2000 31 Ju12001
08 Nov 2001 20 Nov 2001
30 Nov 2001
.:':
30 Nov 2001
: :'
":"
IReports to NWB Management are issued monthly.
As can reasonably be established, the methods and procedures described and the results incorporated into this final report accurately reflect the raw data.
Sheryl A. Mathews, A.A.S.
Date
NWB QAU Compliance Auditor
Page 2
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
COMPLIANCE STATEMENT
The method validation study described in this addendum report is not included within the definition of a GLP nonclinical regulated study. However, to the best of our knowledge, the stability, extraction efficiency, and plasma abbreviated validation experiments were conducted in accordance with the guidelines of the U.S. FDA Good Laboratory Practice Regulations for Nonclinical Laboratory Studies (Title 21 CFR Part 48) and according to the methods and procedures described within this report. In addition, these experiments followed the guidelines of the OECD Principles of Good Laboratory Practice and the Japanese MHW Good Laboratory Practice Standard Ordinance for Nonclinical Laboratory Studies on the Safety of Drugs (Ordinance No. 21, PAB Notification No. 424.)
This addendum report also included results fi:omresearch experiments conducted outside of the validation study. These experiments were conducted according to the procedures described within section 7 of this report.
Serum sample results obtained using plasma curves for PFOSA, PFOSAA, POAA, PFHS, M556 and M570 did not meet NWB SOP requirements for validation.
Any known circumstances that may have affected the quality or integrity of the study or data are discussed within the report. This report represents an accurate record of the raw data.
:u-j,'L,:-_:.,_//_,-..,__z_.._S Connie0.Sakash_/Bt.aS,. NWB ProjecMtanager
Date: /7-D:-)'::
Patrick K. Bennett,M.S., M.B.A. o;" : NWB Laboratory Director
Page 3
NorthwestBioanalytical
StudyNo. NWBS00-040 Report No. NWBR00-122
TABLE OF CONTENTS
SIGNATURE PAGE ........................................................................................................................
1
QUALITY ASSURANCE STATEMENT ...................................................................................... 2
COMPLIANCE STATEMENT .......................................................................................................
3
TABLE OF CONTENTS .................................................................................................................
4
LIST OF TABLES ...........................................................................................................................
5
1. INTRODUCTION ..... .............................................................................................................
13
2. METHODOLOGY ....................................................................................... _......................... 14
3. STABILITY EVALUATION ................................................................................................
17
3.1. Freeze-Thaw and Room Temperature Matrix Stability ..................................................... ! 7
3.2. Reinjection Stability ...........................................................................................................
18
3.3. Extract Storage Stability .......................................................... .......................................... 18
3.4. Long-Term Matrix Stability in Human Serum ................................................................... 19 4. EXTRACTION EFFICIENCY FROM HUMAN SERUM ................................................... 19
5. COMMENTS AND CONCLUSIONS .................................................................................. 21
6. ABBREVIATED VALIDATION - HUMAN PLASMA ...................................................... 21
6.1. Range of Quantitation ........................................................................................................
23
6.2. Lower Limit of Quantitation .............................................................................................. 23
6.3. Intra-Assay Precision and Accuracy ............................................................. ..... ................ 24
6.4. Proposed Acceptance Criteria for Human Plasma ............................................................. 24 7. ADDITIONAL POST-VALIDATION EXPERIMENTS ...................................................... 25
7.1. Ion Suppression Experiment .............................................................................................. 25 7.2. Dilution Matrix Experiment ............................................................................................... 27 7.3. Serum and Plasma Comparison ......................................................................................... 29 8. DATA MANAGEMENT ............................................ _.......................................................... 31
9. REFERENCES ........................................................ .............................................................. 31
10. DATA RETENTION .............................................................................................................
31
J
Page4
NorthwestBioanalytical
LIST OF TABLES
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table I. Summary of Calibration Curve Parameters for PFOS in Human Serum ........................ 32
Table 2. Summary of Calibration Curve Parameters for PFOSA in Human Serum ..................... 32
Table 3. Summary of Calibration Curve Parameters for PFOSAA in Human Serum .................. 32
Table 4. Summary of Calibration Curve Parameters for POAA in Human Serum ...................... 33
Table 5. Summary of Calibration Curve Parameters for PFHS in Human Serum ........................ 33
Table 6. Summary of Calibration Curve Parameters for M556 in Human Serum ........................ 33
Table 7. Summary of Calibration Curve Parameters for M570 in Human Serum ........................ 33 Table 8. Back-Calculated Concentrations of Calibration Standards for PFOS in
Human Serum ...................................................................................... _...................... 34
Table 9. Back-Calculated Concentrations of Calibration Standards for PFOSA in Human Serum ............................................................................................................. 34
Table 10. Back-Calculated Concentrations of Calibration Standards for PFOSAA in
Human Serum .............................................................................................................
34
Table 11. Back-Calculated Concentrations of Calibration Standards for POAA in Human Serum ................. ............................................................................................ 35
Table 12. Back-Calculated Concentrations of Calibration Standards for PFHS in
Human Serum ............................................................................................................. 35
o;" ;
Table 13. Back-Calculated Concentrations of Calibration Standards for M556 in
." ,
:" ,.it. 4
Human Serum ................................................................ ......................... '......... _......... 35
Table 14. Back-Calculated Concentrations of Calibration Standards for M570 in Human Serum ................................. _........................................................................... 36
Table 15. Quality Control Samples for PFOS in Human Serum .................................................. 37 Table 16. Quality Control Samples for PFOSA in Human Serum ............................................... 37 Table 17. Quality Control Samples for PFOSAA in Human Serum ............................................ 37
Page5
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 18. Quality Control Samples for POAA in Human Serum ................................................. 38 Table 19. Quality Control Samples for PFHS in Human Serum .................................................. 38 Table 20. Quality Control Samples for M556 in Human Serum .................................................. 38 Table 21. Quality Control Samples for M570 in Human Serum .................................................. 38 Table 22. Freeze-Thaw and Room Temperature Matrix Stability for PFOS in
Human Serum ............................................................................................................. 39
Table 23. Freeze-Thaw and Room Temperature Matrix Stability for PFOSA in Human Serum ............................................................................................................. 40
Table 24. Freeze-Thaw Stability and Room Temperature Matrix for PFOSA.A in Human Serum ............................................................................................................. 41
Table 25. Freeze-Thaw and Room Temperature Matrix Stability for POAA in Human Serum ............................................................................................................. 42
Table 26. Freeze-Thaw and Room Temperature Matrix Stability for PFHS in
Human Serum .............................................................................................................
43
Table 27. Freeze-Thaw and Room Temperature Matrix Stability for M556 in
Human Serum .............................................................................................................
44
Table 28. Freeze-Thaw and Room Temperature Matrix Stability for M570 in Human Serum ..................................................................................................
........... 45
Table 29. Reinjection Stability for PFOS ................................................................. ..... ............... 46 Table 30. Reinjection Stability for PFOSA ..................................... ......................... -.......:..-......... 4"7_' Table 3 I. Reinjection Stability for PFOSAA ..................................... ........................ ._.................. 48 Table 32. Reinjection Stability for POAA .................................................................................... 49 Table 33. Reinjection Stability for PFHS ..................................................................................... 50 Table 34. Reinjection Stability for M556 ..................................................................................... 51 Table 35. Reinjection Stability for M570 ..................................................................................... 52
Page6
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 36. Room Temperature Extract Stability for PFOS .......................................................
53
Table 37. Room Temperature Extract Stability for PFOSA .............................................
.54
Table 38. Room Temperature Extract Stability for PFOSAA .............................................
.55
Table 39. Room Temperature Extract Stability for POAA ...............................................
56
Table 40. Room Temperature Extract Stability for PFHS ................................................
.57
Table 41. Room Temperature Extract Stability for M556 ....................................................
58
Table 42. Room Temperature Extract Stability for M570 ......................................................... .59
Table 43. Reduced Temperature (1 to 8 C) Extract Stability for PFOS ............ .. ...................... .60 Table 44. Reduced Temperature (1 to 8 C) Extract Stability for PFOSA ................................... 61
Table 45. Reduced Temperature (1 to 8 C) Extract Stability for PFOSAA .............................. .62 Table 46. Reduced Temperature (1 to 8 C) Extract Stability for POAA .................................... .63
Table 47. Reduced Temperature (1 to 8 0C) Extract Stability for PFHS ...................................... 64 Table 48. Reduced Temperature (I to 8 C) Extract Stability for M556 .................................... .65
Table 49. Reduced Temperature (1 to 8 C) Extract Stability for M570 ..................................... .66 Table 50. Reduced Temperature (-20 C) Extract Stability for PFOS. ...................................... .67
Table 51. Reduced Temperature (-20 C) Extract Stability for PFOSA ......................................
.68
Table 52. Reduced Temperature (-20 C) Extract Stability for PFOSAA ................................... 69
Table 53. Reduced Temperature (-20 C) Extract Stability for POAA ................. _ii.'............ ...70
Table 54: Reduced Temperature (-20 C) Extract Stability for PFHS..........................................71 Table 55. Reduced Temperature (-20 C) Extract Stability for M556... ................ " ............... .72
Table 56. Reduced Temperature (-20 C) Fxtract Stability for M570... ..................................... ..73 Table 57. Long-Term Matrix Stability for PFOS in Human Serum ........................................ .74
Table 58. Long-Term Matrix Stability for PFOSA in Human Serum ................ ..... ................. .74
Table 59. Long-Term Matrix Stability for PFOSAA in Human Serum ...............................
.75
Page 7
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 60. Long-Term Matrix Stability for POAA in Human Serum ............................................75
Table 61. Long-Term Matrix Stability for PFHS in Human Serum .............................................76
Table 62. Long-Term Matrix Stability for M556 in Human Serum .............................................76
Table 63. Long-Term Matrix Stability for M570 in Human Serum .............................................77
Table 64. PFOS Extraction Efficiency from Human Serum.........................................................78 Table 65. PFOSA Extraction Efficiency from Human Serum ......................................................79
Table 66. PFOSAA Extraction Efficiency from Human Serum ...................................................80 Table 67. POAA Extraction Efficiency from Human Serum ...............................:.......................81
Table 68. PFHS Extraction Efficiency from Human Serum .........................................................82 Table 69. M556 Extraction Efficiency from Human Serum.........................................................83
Table 70. M570 Extraction Efficiency from Human Serum .........................................................84 Table 71. Summary of Calibration Curve Parameters for PFOS in Human Plasma.....................85
Table 72. Summary of Calibration Curve Parameters for PFOSA in Human Plasma ..................85 Table 73. Summary of Calibration Curve Parameters for PFOSAA in Human Plasma ...............85
Table 74. Summary of Calibration Curve Parameters for POAA in Human Plasma ...................86
Table 75. Summary of Calibration Curve Parameters for PFHS in Human Plasma ..................... 86
Table 76. Summary of Calibration Curve Parameters for M556 in Human Plasma.....................86 .:,..
Table 77. Summary of Calibration Curve Parameters for M570 in Human Plasma.....................87
Table 78. Back-Calculated Concentrations of Calibration Standards for PFOS in Human Plasma ............................................................................................................ 88
Table 79. Back-Calculated Concentrations of Calibration Standards for PFOSA in Human Plasma ............................................................................................................ 88
Table 80. Back-Calculated Concentrations of Calibration Standards for PFOSAA in Human Plasma ............................................................................................................ 88
Page 8
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 81. Back-Calculated Concentrations of Calibration Standards for POAA in Human Plasma ............................................................................................................89
Table 82. Back-Calculated Concentrations of Calibration Standards for PFHS in Human Plasma ............................................................................................................89
Table 83. Back-Calculated Concentrations of Calibration Standards for M556 in Human Plasma ............................................................................................................89
Table 84. Back-Calculated Concentrations of Calibration Standards for M570 in Human Plasma ............................................................................................................90
Table 85. Lower Limit ofQuantitation for PFOS in Human Plasma ...........................................91
Table 86. Lower Limit of Quantitation for PFOSA in Human Plasma ........................................92
Table 87. Lower Limit of Quantitation for PFOSAA in Human Plasma......................................93
Table 88. Lower Limit of Quantitation for POAA in Human Plasma ..........................................94
Table 89. Lower Limit of Quantitation for PFHS in Human Plasma ...........................................95
Table 90. Lower Limit of Quantitation for M556 in Human Plasma ...........................................96
Table 91. Lower Limit of Quantitation for M570 in Human Plasma ...........................................97
Table 92. Intra-Assay Precision for PFOS Quality Controls in Human Plasma ...........................98 Table 93. Intra-Assay Precision for t;FOSA Quality Controls in Human Plasma ........................99
Table 94. Intra-Assay Precision for PFOSAA Quality Controls in Human Plasma ..............._...100
Table 95. Intra-AssayPrecision for POAA Quality Controls in Human Plasma........................I01
" :
" . .". .o
Table 96. Intra-Assay Precision for PFHS Quality Controls in Human Plasma.........................102
Table 97. Intra-Assay Precision for M556 Quality Controls in Human Plasma ......................... 103
Table 98. Intra-Assay Precision for M570 Quality Controls in Human Plasma .........................104
Table 99. Summary of Calibration Curve Parameters for PFOS in Serum Diluted with Buffer ................................................................................................................ 105
Page 9
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 100. Summary of Calibration Curve Parameters for PFOSA in Serum Diluted
with Buffer ................................................................................................................
105
Table 101. Summary of Calibration Curve Parameters for PFOSAA in Serum Diluted
with Buffer ................................................................................................................
105
Table 102. Summary of Calibration Curve Parameters for POAA in Serum Diluted
with Buffer ................................................................................................................
106
Table 103. Summary of Calibration Curve Parameters for PFHS in Serum Diluted
with Buffer ................................................................................................................
106
Table 104. Summary of Calibration Curve Parameters for M556 in Serum Diluted
with Buffer ................................................................................................................
106
Table 105. Summary of Calibration Curve Parameters for M570 in Serum Diluted
with Buffer ................................................................................................................
107
Table 106. Back-Calculated Concentrations of Calibration Standards for PFOS in Serum
Diluted with Buffer ...................................................................................................
108
Table 107. Back-Calculated Concentrations of Calibration Standards for PFOSA in Serum
Diluted with Buffer ...................................................................................................
108
Table 108. Back-Calculated Concentrations of Calibration Standards for PFOSAA. in Serum
Diluted With Buffer ...................................................................................................
108
Table 109. Back-Calculated Concentrations of Calibration Standards for POAA in Serum Diluted with Buffer ................................................................................. ".:._.............. 109
;l'able 110. Back-Calculated Concentrations of Calibration Standards for PFHS in'Serum
a
Diluted with Buffer ...................................................................................................
109
Table 111. Back-Calculated Concentrations of Calibration Standards for M556 in Serum Diluted with Buffer ............. _..................................................................................... 109
Table 112. Back-Calculated Concentrations of Calibration Standards for M570 in Serum
Diluted with Buffer ...................................................................................................
110
Table 113. PFOS Quality Controls Samples Prepared in Serum Diluted with Buffer ..................
Page I0
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 114. PFOSA Quality Control Samples Prepared in Serum Diluted with Buffer .............. 111
Table 115. PFOSAA Quality Control Samples Prepared in Serum Diluted with Buffer ........... 112 Table 116. POAA Quality Control Samples Prepared in Serum Diluted with Buffer ................ 112
Table 117. PFHS Quality Control Samples Prepared in Serum Diluted with Buffer ................. 113
Table 118. M556 Quality Control Samples Prepared in Serum Diluted with Buffer ................. 113
Table 119. M570 Quality Control Samples Prepared in Serum Diluted with Buffer ................. 114 Table 120. Summary of Calibration Curve Parameters for PFOS (Serum vs. Plasma) .............. 115
Table 121. Summary of Calibration Curve Parameters for PFOSA (Serum vs. Plasma) ........... 115 Table 122. Summary of Calibration Curve Parameters for PFOSAA (Serum vs. Plasma) ........ 115
Table 123. Summary of Calibration Curve Parameters for POAA (Serum vs. Plasma) ............ 116 Table 124. Summary of Calibration Curve Parameters for PFHS (Serum vs. Plasma) .............. 116
Table 125. Summary of Calibration Curve Parameters for M556 (Serum vs. Plasma) .............. 116
Table 126. Summary of Calibration Curve Parameters for M570 (Serum vs. Plasma) .............. 117
Table 127. Back-Calculated Concentrations of Calibration Standards for PFOS
(Serum vs. Plasma) ...................................................................................................
i 18
Table 128. Back-Calculated Concentrations of Calibration Standards for PFOSA
(Serum vs. Plasma) ...................................................................................................
119
Table 129. Back-Calculated Concentrations of Calibration Standards for PFOSAA: ... (Serum vs. Plasma) .................................................................................... .....:.......... 120-- -
Table 130. Back-Calculated Concentrations of Calibration Standards for POAA
(Serum vs. Plasma) ...................................................................................................
121
Table 131. Back-Calculated Concentrations of Calibration Standards for PFHS
(Serum vs. Plasma) ...................................................................................................
122
Table 132. Back-Calculated Concentrations of Calibration Standards for M556
(Serum vs. Plasma) ...................................................................................................
123
Page 11
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 133. Back-Calculated Concentrations of Calibration Standards for M570 (Serum vs. Plasma) ...................................................................................................124
Table 134. PFOS Quality Controls (Serum vs. Plasma) .............................................................. 125
Table 135. PFOSA Quality Controls (Serum vs. Plasma) ...........................................................127
Table 136. PFOSAA Quality Controls (Serum vs. Plasma) ........................................................ 128
Table 137. POAA Quality Controls (Serum vs. Plasma) .............................................................129
Table 138. PFHS Quality Controls (Serum vs. Plasma) ..............................................................131
Table 139. M556 Quality Controls (Serum vs. Plasma) ..................................i..........................132
Table 140. M570 Quality Controls (Serum vs. Plasma) .............................................................133
Table 141. Serum QC Results Based on Plasma Curves ............................................................135
: ;
:
.. .
";..,"..o
Page 12
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Quantitative
Determination of PFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 in Human Serum by LC/MS/MS
Assay Revalidation Addendum Report 1. INTRODUCTION
Northwest Bioanalytical (NWB) was contracted by 3M Environmental Technology and Services (3M) to modify and revalidate a liquid chromatography/tandem mass spectrometry method for the measurement ofPFOS, PFOSA, PFOSAA, pOAA, PFHS, M556 and M570 in human serum. The accuracy and precision results of the method revalidation were reported in NWB report NWBR00-108 [9.I]. This report summarizes the results from the stability tests conducted as part of the method revalidation for PFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 in human serum for 3M Environmental Technology and Services. The results from the following tests are also included: extraction efficiency from human serum; ion suppression experiment; and diluted matrix experiment. The suitability of the method for use with plasma was determined by demonstrating intra-assay and LLOQ accuracy and precision and comparing serum and plasma results.
Kris Hansen and James Lundberg at 3M Environmental Technology and Services served
as the Study Monitors. The following is a list of NWB supervisory personnel involved in the completion of this work: Connie O. Sakashita, B.S. (Current NWB Project Manager); Brad I. Coopersmith, Ph.D. (NWB Project Manager until October 28, 2000); Lico.ng . ... Jiang, Ph.D. (NWB Senior Scientist); Patrick Bennett, M.S., M.B.A. (NWB Laboratory Director). The following NWB laboratory personnel contributed to the completion of this work: Suzanne Newman, B.S. (NWB Scientist).
NWB SOPs were used in the conduct of this project and were available to project personnel in both electronic and hard copy formats.
Date Study Initiated: 13-Apr-2000 Date Analyses Completed: 25-Apr-2001
Page 13
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
The method revalidation study described in this report is not included within the definition of a GLP regulated nonclinical study. However, Northwest Bioanalytical conducts all studies within the guidelines of the US FDA Good Laboratory Practice Regulations for Nonclinical Laboratory Studies (Title 21 CFR Part 58), the OECD Principles of Good Laboratory Practice and the Japanese MHW Good Laboratory Practice Standard Ordinance for Nonclinical Laboratory Studies on the Safety of Drugs (Ordinance No. 21, PAB Notification No. 424). Any changes to or deviations from the original protocol (Analytical Plan) were documented through approved protocol amendments or deviation memos and are retained within the raw data.
2.
METHODOLOGY
Reference Materials and Matrices
Analyte PFOS (FC-95)
PFOSA
Lot Number Purity
193
100%
214
100%
Expiration Date
12/31/2010
12/31/2010
Source 3M 3M
PFOSAA (FC-129)
617
53.8%
12/31/2010
3M
POAA (FC- 143)
245
100%
12/31/2010
3M
PFHS*
$398-182
100%
12/31/2010
3M
(used for serum)
PFHS
SE-036
91.1%
1/01/2010
3M
(used for plasma)
M556
NBl13047-
99.89%
12/31/2010
3M
8D
M570
118506-26 99.75%
12/31/2010
3M
THPFOS
59909
90%
12/31/2010
3M
*Received as a 6,200 ppm solution in methanol. **Kept dry.
Storage Conditions
Room Temperature
Room Temperature
Room Temperature
Room Temperature**
-20C .: .:
-20C :
" -" Room Temperature Room Temperature Room Temperature
Page 14
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Matrix
Human serum Human plasma (Northern Chinese)
Lot Numbers BC30399-4; BC061900-1 C5186; C4929; C5517; C5200
The reference material purity for PFOS, PFOSA, PFHS and POAA was not available t
prior to the conduct of this study. Therefore, all serum concentrations included in the report for these analytes are based upon an assumed purity of 100%.
After the revalidation was completed, 3M contracted with Centre Analytical Laboratories, Inc. in State College, Pennsylvania to determine the absolute concentration of PFOS, POAA and PFHS in NWB stock solutions used to prepare the analytical standards and controls used for this revalidation and subsequent analyses. All arrangements for purity determinations and transfers of NWB solutions to Centre Analytical Laboratories, Inc. were performed by 3M. Per agreement with the Sponsor, NWB quantitative results were not corrected for the purity corrections. Therefore, based on the results obtained, the serum concentrations included in this report should be corrected according to the following table (except for the' serum samples included in the serum versus plasma experiments, Section 7.3):
Analyte PFOS POAA PFHS
Correction Factor
0.836
0.909
o: -.
0.855
Principles of the Method
The analytical method consisted of a liquid:liquid extraction procedure followed by evaporation and reconstitution of the extract residue with 30:70 (v/v) 20 mM ammonium acetate in water: 20 mM ammonium acetate in methanol. The samples were analyzed by liquid chromatography/tandem mass spectrometry using a PE Sciex AP13000. The instrument was operated in the multiple reaction monitoring (MRM) mode under
Page 15
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
optimized conditions for PFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 detection.
A detailed description of the methodology is included in NWB report NWBR00-108 [9.1].
Acceptance Criteria for Human Serum Calibration Curve
Each run will include in duplicate calibration standards at six or more concentrations covering the lower to upper limit ofquantitation. For all analytes except PFOSA, at least three-fourths of the calibration standard's back-calculated concentrations must be within + 15% (+ 20% for LLOQ) of their individual target concentrations. For PFOSA, at least three-fourths of the calibration standard's back-calculated concentrations must be within + 20% (+ 25% for LLOQ) of their individual target concentrations. A calibration standard will be considered a statistical outlier if the back-calculated concentration is greater than two times the acceptance criteria for that standard.
Lower Limit of Quantitation
The back-calculated concentrations of at least one of the duplicate lowest points in the
calibration curve must be within + 25% of the target concentration for PFOSA to qualify
as the LLOQ and within _+20% of the target concentration to qualify as the LLOQ for all
other analytes. If this criterion is not met, the next level is-subjected to the "s_-netest and
the LLOQ raised accordingly.
:"
: ="
Quality Control Samples
Each analytical run will include low, medium and high QC samples in duplicate. The measured concentrations of at least two-thirds of all analytical QCs must be within + 20% of their target concentrations (+ 25% for PFOSA), and no two QCs at the same concentration can be outside the limit. If study samples require dilution, a dilution QC will be analyzed in triplicate for each dilution level (except for PFOSA and PFOSAA
'" Page 16
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
which should not be diluted with control matrix). At least two dilution QC at each level must be within _ 20% of its target concentrations in order to accept diluted study samples at that level. The dilution QC acceptance is independent of the undiluted analytical QC acceptance.
3. STABILITY EVALUATION
The revalidated method's final acceptance criteria is used as a basis for evaluating fluorochemical stability results. If the mean concentration obtained for stability test QC samples falls within + 20% ( 25% for PFOSA) of the mean control QC concentrations, then stability is considered acceptable under the specified test conditions. Control QC samples are defined as samples not subjected to test conditions.
The calibration curve parameters summary and individual calibration standard results for the serum stability runs can be found in Tables 1-14. The analytical QC results for the stability runs can be found in Tables 15-21.
3.1. Freeze-Thaw and Room Temperature Matrix Stability
The stability ofPFOS, PFOSA, PFOSAA, POAA, and PFHS in human serum for three freeze-thaw cycles was demonstrated and reported under the original validation study NWBS98-092, NWB report NWBR99-005 [9.2]. Therefore, only freeze-thaw stability for M556 and M570 was needed. Acceptable results for the three freezethaw cycle stability test was obtained on Run 4 for M556 and M570 (data not included).
However, since some analyzed subject samples subsequently underwentmore than three freeze-thaw cycles, the stability ofPFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 was evaluated after seven freeze-thaw cycles. All analytes demonstrated acceptable stability when compared against controls that underwent one freeze-thaw cycle (Tables 22 - 28).
During the seven freeze-thaw cycle test, samples were left at room temperature for up to 6.75 hour periods between each freeze cycle. Since the freeze-thaw samples
Page 17
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
demonstrated acceptable stability, the stability of serum samples left at room temperature for up to 6.75 hours is also demonstrated.
3.2. Reinjection Stability
The quantitative reproducibility of standards and controls after reinjection was determined for PFOS, PFOSA, PFOSAA, POAA, PFHS, M556, and M570. The samples were extracted and analyzed on June 30, 2000 (Run 15- data shown in NWBR00-108). These samples were reinjected on July 3, 2000 after storage at room temperature (Run 23). The results from Run 23 indicate reinjection stability when compared with data obtained from the original injection [Run 15] (Tables 29 - 35).
3.3. Extract Storage Stability
Extract stability was evaluated at different storage temperatures and for various lengths of time. The evaluations were performed using QC lots prepared on June 27, 2000 for all analytes and conditions. Although the same QC lots (prepared on June 27, 2000) were not used for PFOSA and PFOSAA run acceptance determination, they still provided a control value for stability evaluations.
3.3.1. Room Temperature Extract Stability
The extract stability ofPFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 was determined for low and high serum QC extracts stored at room temperature for seven days prior to analysis. These test QC samples were compared against serum control samples that were analyzed immediately following eg_:action.
The extracted test QC samples demonstrated acceptable stability after seven days at normal room temperature (Tables 36 - 42).
3.3.2. Reduced Temperature (1 to 8C) Extract Stability
The extract stability ofPFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 was determined for low and high serum QC extracts stored at 1 to 8C for four
Page 18
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
days prior to analysis. These test QC samples were compared against serum control samples analyzed immediately following extraction.
The extracted test QC samples demonstrated acceptable stability after four days at 1 to 8 C (Tables 43 - 49).
3.3.3. Reduced Temperature (-20 C) ExtractStability
The extract stability ofPFOS, PFOSA, PFOSAA, POAA, PFHS, M556, and M570 was determined for low and high serum QC extracts stored at -20 C for five days prior to analysis. These test QC samples were compared against serum control samples analyzed immediately following extraction.
The extracted test QC samples demonstrated acceptable stability after five days at -20 C (Tables 50 - 56).
3.4. Long-Term Matrix Stability in Human Serum
The long-term matrix stability ofPFOS, POAA, PFHS, M556 and M570 in human serum was determined by storing QC samples for approximately 55 days at -20 C. The long-term matrix stability of PFOSA and PFOSAA in human serum was determined by storing QC samples for approximately 42 days at -20 C. The mean test QC concentrations were compared against theoretical.
The test QC samples demonstrated acceptable stability after approximat.e!y42 days for PFOSA & PFOSAA or 55 days for other analytes, at-20 C (Tables 57 -.63). : ...
4. EXTRACTION EFFICIENCY FROM HUMAN SERUM
The extraction efficiencies ofPFOS, PFOSA, PFOSAA, POAA, PFHS,M556 and M570 were determined by comparing the area ratios obtained for the following three cases:
1. Both the analyte and internal standard added following the extraction (postextract).
Page 19
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
2. The analyte added to serum prior to extraction and the internal standard added following extraction (pre-extract analyte).
3. The internal standard added to serum prior to extraction and the analyte added following the extraction (pre-extract internal standard).
The extraction efficiencies were then determined by the area ratio of the pre-extract samples to the post-extract samples. The extraction efficiency experiments were performed at both low and high concentrations to determine that there was no concentration bias.
The total amount ofanalyte in the sample prior to extraction (pre-extract) is equal to the amount of analyte added plus the persistent amount in the blank matrix. The amount of analyte in the post-extract solution is the amount of analyte added plus the amount of analyte recovered (extracted) from the blank matrix. Therefore, the following formula was used to determine the extraction efficiency for PFOS, PFOSAA, POAA, PFHS, M556 arid M570:
((Spk + Per)* EE)/(Spk + (Per * EE)) = Rpre / Rpost; where
Spk = the amount of analyte added Per = the persistent amount of analyte in the blank matrix EE = the extraction efficiency Rpre -- the ratio of analyte/intemal standard determined for the analyte added prior and internal standard added after extraction Rpost = the ratio of analyte/internal standard determined for the analyte and internal standard added after extraction
Mean extraction efficiency (%)
PFOS 37.3
PFOSA 69.6
PFO.$ .AA 62.7 ; .
" .- _
Mean extraction efficiency (%)
POAA 15.6
PFHS 15.3
M556 39.1
Mean extraction efficiency (%)
M570 57.1
The results are shown in Tables 64 - 70.
THPFOS 5.99
Page 20
NorthwestBioanalytical
5.
COMMENTS AND CONCLUSIONS
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Per agreement with the Sponsor, the regressions were not recalculated based upon the updated purity information for PFOS, POAA and PFHS. Some differences might occur due to the effect of rounding if the regressions were performed with the purity corrected concentrations, but these differences would have a negligible effect on the overall interpretation of the revalidation results.
PFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 in human serum appear to be stable for seven freeze-thaw cycles and demonstrated matrix stability for up to 6.75 hours at room temperature. PFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 exhibit extract stability for seven days at room temperature, four days at 1 to 8 C, and five days at -20 C. The analytes also demonstrated matrix stability for approximately 55 days (for PFOS, POAA, PFHS, M556 and M570) and 42 days (for PFOSA and PFOSAA) at -20 C in human serum.
6.
ABBREVIATED VALIDATION - I:IUMAN PLASMA
The choice of matrix for calibration curve preparation is dependent upon the desired LLOQ. 3M supplied NWB with human plasma identified as "Northem Chinese" plasma (referred to as Chinese plasma or CP). This matrix has very low levels of fluorochemicals and will be used for studies with a desired LLOQ of< 5.00 ppb. The following sections summarize the accuracy and precision of the fluorochemicals in human plasma and the quantitative accuracy of serum sample results obtain,e:d. against plasma calibration curves using the revalidated serum method.
The absolute concentrations ofPFOS, PFHS and POAA in the stock solution used for plasma analyses were determined by Centre Analytical Laboratories, Inc. in State College, Pennsylvania. Based upon the results, the concentrations of plasma standard and quality control samples were corrected using the following factors:
Page21
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Analyte PFOS POAA PFHS
Correction Factor 0.821 0.961 1.043
All plasma experiment concentrations included in this report reflect adjustments based upon the correction factors above. The serum target concentrations for these experiments have also been adjusted based upon the correction factors (listed on page 15 of report).
The tests for the abbreviated human plasma validation were conducted after the serum revalidation was completed. They were performed outside of the revalidation or in sample analysis studies conducted after the serum revalidation. However, all applicable NWB SOPs were followed.
Persistent Levels of Analytes in Chinese Plasma
The persistent levels of the analytes in Chinese plasma varied in each tube of plasma tested. Therefore, several tubes of plasma were pooled and the amounts of persistent analytes were determined. Previous determination of persistent levels was achieved by comparison with rabbit serum curves. For the experiments included in this report, the persistent levels were determined as follows:
Known amounts of the analytes were added to Chinese plasma in the range of 1 to 50 or 1 to 100 ppb. The standards were then analyzed according to the revalidated method on at least two separate occasions. A linear regression was performed using the spiked analyte concentration as the y-value and the ratio of the analyte area to internal starldard area as the x-value (Microsoft Excel 97 formulas). The mean resultant y-intercept determined from multiple runs was then used as an estimate of the persistent level of analyte if it was greater than 0.2 ppb.
Page 22
NorthwestBioanalytical 6.1. Range of Quantitation
StudyNo. NWBS00-040 ReportNo. NWBR00-122
The target calibrator concentrations were approximately 1.00,2.50, I0.0, 25.0, 50.0, i00, 250, 400 and 500 ppb for all analytes except M556. The target calibrator concentrations for M556 were approximately 2.50, I0.0, 25.0, 50.0, I00, 250, 400 and 500 ppb. The final curve range can vary based upon the persistent levels of the analyte in the human plasma pool used. For the validation study, the calibration curve ranges were as follows:
PFOS PFOSA PFOSAA POAA PFHS M556 M570
3.94 to 414 ppb 1.00 to 500 ppb l.d0 to 501 ppb 1.92 to 481 ppb 1.36 to 523 ppb 2,50 to 500 ppb 1.00 to 500 ppb
The calibration curve parameter summaries and individual calibration standard results for the plasma runs can be found in Tables 71 - 84.
6.2. Lower Limit of Quantitation
The precision at the Lower Limit of Quantitation (LLOQ) of the assay in human plasma was determined by using six 100-_L aliquots of the pooled Chinese plasma fortified with PFOS, PFOSA, PFOSAA, POAA, PFHS, M556, and M570. at concentrations equal to the LLOQ. The LLOQ concentration was dependent upon the,= persistent levels found in the plasma.
Lower Limit of Quantitation Precision (%CV) Accuracy (%Bias)
PFOS 3.94 ppb
5.4 -1.5
PFOSA 1.00 ppb
10.1 1.0
PFOSAA 1.60 ppb
5.7 -2.5
POAA. 1.92 ppb
4.4 -8.9
Page23
NorthwestBioanalytical
Lower Limit of Quantitation Precision (%CV)
Accuracy (%Bias)
PFHS 1.36 ppb
9.9 -5. I
The results are shown in Tables 85 - 91.
M556 2.50 ppb
5.7 -2.0
StudyNo. NWBS00-040 ReportNo. NWBR00-122
M570
1.00 ppb 5.2 3.0
6.3. Intra-Assay Precision and Aeeuracy
The precision and accuracy of the LC/MS/MS method for PFOS, PFOSA, PFOSAA, POAA, PFHS, M556, and M570 in human plasma were determined by analyzing three levels of quality control samples in replicates of five on a single run.
Precision (%CV) Accuracy (%Theoretical)
Accuracy (%Bias)
PFOS Intra-assay 1.5 to 2.8 101.2 to 112.3 1.2 to 12.3
PFOSA Intra-assay 1.8 to 5.7 104.0 to 116.3 4.0 to 16.3
PFOSAA Intra-assay 2.6 to 5.8 102.6 to i 11.7 2.6 to 11.7
POAA lntra-assay 0.8 to 5.8 103.4 to 106.2 3.4 to 6.2
Precision (%CV) Accuracy (%Theoretical)
Accuracy (%Bias)
PFHS Intra-assay 1.7 to 3.1 95.7 to 106.0 -4.3 to 6.0
M556 I.rttra-assay 4.7 to 7.6 98.7 to 111.3 -1.3 to 11.3
M570 Intra-assay 1.9 to 5.6 101.3 to 118.0 1.3 to 18.0
The results are shown in Tables 92 - 98.
6.4. Proposed Acceptance Criteria for Human Plasma Calibration Curve
Each run will include in duplicate calibration standards at six or more concentrations covering the lower to upper limit ofquantitation. For all analytes except PFOSA and PFOSAA, at least three-fourths of the calibration standard's back-calculated concentrations must be within + 15% (+ 20% for LLOQ) of their individual target
Page24
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
concentrations. For PFOSA and PFOSAA, at least three-fourths of the calibration standard's back-calculated concentrations must be within _+20% (_+25% for LLOQ) of their individual target concentrations. A calibration standard will be considered a statistical outlier if the back-calculated concentration is greater than two times the acceptance criteria for that standard.
Lower Limit of Quantitation
The back-calculated concentrations of at least one of the duplicate lowest points in the calibration curve must be within + 25% of the target concentration for PFOSA and PFOSAA to qualify as the LLOQ and within + 20% of the target concentration to qualify as the LLOQ for all other analytes. If this criterion is not met, the next level is subjected to the same test and the LLOQ raised accordingly.
Quality Control Samples
Each analytical run will include low, medium and high QC samples in duplicate. The measured concentrations of at least two-thirds of all analytical QCs must be within + 20% of their target concentrations (+ 25% for PFOSA and PFOSAA), and no two QCs at thesame concentration can be outside the limit.
Based upon the results, the method is suitable for use with human plasma.
7.
ADDITIONAL POST-REVALIDATION EXPERIMENTS
.o: _ ;
Additional experiments were conducted subsequent to completion of the reyalida_.ion to : =-_
further elucidate the applicability of the revalidated method to sample analysis. These tests were performed outside of any revalidation or sample analysis studies.
7.1. Ion Suppression Experiment
During analysis of medical surveillance serum samples, it was noted that the internal standard response for the study samples differed from the standards and quality controls. POAA and the internal standard (THPFOS) have very similar retention
Page25
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBK00-122
times. Therefore, THPFOS and POAA are not chromatographically resolved in extracted samples containing high concentrations of POAA. This co-elution can suppress the internal standard response and introduce a (high) bias in the measured concentration of analytes when compared against the standard curve. The suppression of the internal standard signal is proportional tOthe concentration of POAA in the sample and dependent on the LC resolution of the two compounds.
An experiment was performed to evaluate the magnitude of this effect. Low control samples were fortified with various (high) levels of POAA. The test samples were analyzed in trip.licate in a single run (PFOSMD-41). The mean concentration of the test samples were then compared against the mean of the analytical low control samples. A summary of the results is shown in the following table:
PFOS ControlMean Cone. (ppb) 51.8 POAA added at --59200 ppb % change from control mean 182% POAA added at-11800 ppb % change from control mean 28% POAA added at ---5920ppb % change from controlmean 16% POAA added at -592 ppb % change from control mean -2%
PFOSA PFOSAA PFHS 4.05 6.83 8.04 152% 180% 180% 39% 52% 29% 21% 30% 16% -11% -5% -7%
M556 5.41 216% 52% 36% -8%
11/I570 IS Area 6.87 108857 207% -75% 56% -43% 34% -32% -8% +1 i%
The experiment verified that very high levels of POAA resulted in a lower response for the internal standard, which then results in a falsel3/elevated concentration for the other 6 analytes. No significant bias is introduced when the extracted concen_ation of POAA is within the revalidated range of the analytical method (approXimately 500 ppb). However, some samples can contain POAA at concentrations above the revalidated range of the analytical method, and will likely introduce a bias in the reported concentrations of PFOS, PFHS, PFOSA, PFOSAA, M556, or M570 when the concentration of these analytes are determined prior to sample dilution. The amount of bias could range up to approximately 56%. However, the amount of
J
Page26
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
PFOS, PFHS, PFOSA, PFOSAA, M556, or M570 impurities in the POAA reference material is not known. This prevents accurate determination of the % bias. It should be noted that since the amount of ion suppression is dependent upon the amount of extracted POAA, no concentration bias is expected for an analyte result determined from a pre-diluted sample wherein the extracted POAA concentration is less than approximately 500 ppb. This bias will not significantly effect the interpretation of the analytical results.
7.2. Dilution Matrkx Experiment
The amount of Chinese plasma supplied by 3M is limited. Therefore, it cannot be used on a routine basis for dilution. Human serum (control serum) is used to dilute test samples as needed to obtain measurable concentrations within the range of the calibration curve. The measured concentrations of fluorochemicals (FC) obtained for diluted samples must then be corrected for persistent levels of the analytes in the control serum. Experiments were performed to determine whether a substitute matrix that did not contain persistent levels ofFC analytes could be used when performing dilutions.
Previous experiments had shown that samples diluted with buffer instead of serum did not accurately quantitate against a typical serum curve (data not reported). This
indicated that the matrix to buffer ratio of the sample affected the extraction recovery.
For the experiments described in this report, QC samples were diluted with buffer. The calibration curve was prepared in the same buffer to matrix ratio _'he QC samples to determine if this was a viable method for diluting samples. A different " '_ " dilution ratio was prepared to determine if samples could be quantitated from a curve prepared in a matrix to buffer ratio that was different from the QC samples (e.g., if QC samples diluted with matrix by a factor of 50 (2% matrix in buffer) could be quantitated against a curve prepared in 10% matrix in buffer). For the experiments described below, ammonium acetate buffer was used to dilute the matrix for the QCs and calibration curves.
Page 27
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Calibration Curve and QC Preparation
The experiment (PFOSMD run 39) used a curve prepared in 10% serum in buffer, and three levels of QC samples diluted with buffer by a factor of 10 (10% matrix to buffer ratio). The QC concentrations prior to dilution were 150, 400, and 4000 ppb plus the persistent level for each analyte. QC samples at 4000 ppb (plus persistent level) were also diluted 1:50 with buffer.
The calibration curve parameter summaries and individual calibration standard results for the runs can be found in Tables 99 - 112. The quality control sample results can be found in Tables 113 - 119.
Results for 10% Serum in Buffer (PFOSMD Run 39)
The mean accuracy (% theoretical) for all analytes at all QC concentration levels
diluted with buffer by a factor of 10 (the same serum to buffer ratio as the curve)
ranged from 94.1% to 114.3%. The intra-assay precision (%CV) for all analytes were
(
less than or equal to 5.6% for each QC concentration.
The mean accuracy (% theoretical) for the QC samples diluted with buffer by a factor of 50 ranged from 56.5% to 84.8%. The inter-assay precision (%CV) for all analytes were less than or equal to 4.4%. All analytes were biased low, and all analytes except for POAA and PFHS were not within +_20%of the target concentrations.
QC samples met the acceptance criteria of_+ 20% of target concentration (+_25%for PFOSAA) when diluted with buffer if quantitated against a curve prep_e.d in the same serum to buffer ratio. QC samples diluted with buffer at a different ratio than : ... the calibration curve did not meet the acceptance criteria for all analytes.
It appears that quantitating samples diluted with buffer against a curve prepared in the same serum to buffer ratio is a viable way minimizing the corrections needed due to persistent levels Of analytes in the dilution matrix. However, since most analytes cannot be accurately quantitated against a calibration curve that is prepared in a serum/buffer ratio different from that of the diluted samples, a separate curve with the same buffer to matrix ratio would need to be prepared for each dilution factor needed.
Page28
NorthwestBioanalytical 7.3. Serum and Plasma Comparison
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Because the choice of calibration curve and quality control matrix is dependent upon the desired LLOQ, it is not always possible to match the subject sample matrix with that of the calibration curve and quality control matrix. For instance, it is necessary to use Chinese plasma to prepare the calibration standards and quality control samples when analyzing serum samples where a LLOQ of< 5.00 ppb is desired. Therefore, to evaluate if, and to what extent, the use of plasma calibration standards affects the quantitative analysis of fluorochemicals in serum, a comparison of quantitative results obtained from plasma curves versus serum curves was performed.
Calibration curves were prepared in both human serum and Chinese plasma, and were analyzed along with serum and Chinese plasma quality control samples in replicates of four. The results for the calibration and quality control samples are shown in Tables 120- 140.
The confidence interval around the serum QC mean was determined for serum
samples quantitated against the plasma curve using Microsoi_@ Excel 97. Alpha (a) was set to 0.05 for a 95% confidence level. The confidence interval around the mean
was compared to the analytical acceptance range of the QC which is _ 20% (__2_5%
for PFOSA and PFOSAA) of the target concentration. experiment are shown in Table 141.
The summary results of the
The i0/27/00 serum curve for M570 had a raised LLOQ of 5.20 ppb. Data for
,.t .
PFOSA and PFOSA.A. analyzed on 10/27/00 and data for PFHS and M556 analyzed
"
,4 :. .m
on 4/25/01 are not used because either the serum or plasma calibration _urve br the
QC samples did not meet the acceptance criteria. One h!gh QC replicat_ analyzed on 10/27/00 was not included in the statistical calculations because of an extraction error.
The confidence intervals around the mean serum QC results for PFOS are within the acceptance criteria at all levels.
Page29
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
The confidence intervals around the mean serum QC results for PFOSA are greater than the acceptance range of+_.25% deviation. The confidence intervals ranged up to 78% deviation from target.
The confidence intervals around the mean serum medium QC result for PFOSAA was 33.9% deviation which is greater than the plasma acceptance range of+_25% deviation. The confidence interval for the PFOSAA low and high QC samples are within the acceptance criteria.
The confidence intervals around the mean serum QC results for POAA and PFHS are greater than the acceptance range of+_20% deviation. The confidence intervals ranged up to 27.4% deviation from target for POAA and up to 37.6% deviation from target for PFHS.
The confidence interval around the mean serum low QC results for M556 was -26.9% deviation. This is greater than the acceptance range of+_20% deviation and may be due a low value obtained for one replicate. The confidence intervals for the M556 medium and high QC samples are within the acceptance criteria.
The confidence intervals around the mean serum medium QC result for M570 was
21.9% deviation which is slightly greater than the plasma acceptance range of+_20%
deviation. The confidence interval for the M570 low and high QC samples are within
the acceptance criteria. CONCLUSION
o; ;
:
: ,: .
Serum sample results obtained using plasma curves for PFOSA, PFOSA?A, POAA, PFHS, M556, and M570 did not meet NWB SOP requirements for validation. Results may vary on average up to 26% from QC results obtained using human serum calibration curves at some concentration levels for PFHS, PFOSAA, POAA, M556, and M570. For PFOSA, results may vary on average up to 43% from QC results obtained using human serum calibration curves. However, when a lower limit of
Page 30
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
quantitation less than 5 ppb is required by the Sponsor, it will be necessary to use plasma curves.
8.
DATA MANAGElVlENT
PFOS, PFOSA, PFOSAA, POAA, PFHS, M556, M570, and THPFOS chromatographic peaks were integrated using PE Sciex MacQuan software (version 1.6) with a smooth factor of one. Quantitation was based upon quadratic regression analysis of calibration curves (weighted 1/Xz) using the area ratio vs. concentration calculated by the Watson DMLIMS software (versions 6.1.1.03 and 6.1.1.04).
9.
REFERENCES
[9.1.] C. Sakashita. "Quantitative Determination ofPFOS, PFOSA, PFOSAA, POAA, PFHS, M556 and M570 in Human Serum by LC/MS/MS." Assay Revalidation Report. NWB StudyNWBS00-040. NWB Report NWBR00-108. January 24, 2001.
[9.2.] D. Vollmer. "Quantitative Determination ofPFOS, PFOSA, PFOSAA, N-MeFOSE-OH, N-EtFOSE-OH, POAA and PFHS in Human Serum by LC/MS/MS." Assay Validation Report. NWB Study NWBS98-082. NWB Report NWBR_99-005. May 13, 1999.
10. DATA RETENTION
The raw data and final report for this study will be stored in the NWB Archi;ce, 1121 East 3900 South, Salt Lake City, UT 84124 per regulations and contract agreement. 3M "': " Environmental Technology and Services will be notified concerning final disposition of records at completion of contract obligations.
Page 31
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 1. Summary of Calibration Curve Parameters for PFOS in Human Serum Quadratic weighted 1/x2. All concentrationsare expressed as ppb.
Run Date
03-Jul-2000 21-Aug-2000
Run
A
B
C R-Squared
Number
23
-0.000004 0.0i9854 0.112640 0.9862
18
0.000001 0.028989 0.257857 0.9849
A, B, and C arecoefficients used to define the calibration curve.
LLOQ
48.1 48.1
ULOQ
547 547
Table 2. Summary of Calibration Curve Parameters for PFOSA in Human Serum Quadratic weighted 1/x2. All concentrations areexpressed as ppb.
Run Date
03-Jul-2000 21-Aug-2000
Run
A
B
C
R-Squared LLOQ
Number
23
-0.000017 0.054058 0.006149 0.9952
1.00
18
0.000005 0.082782 0.019970 0.9952
1.00
A, B, and C are coefficients used to define the calibration curve.
ULOQ
502 502
Table 3. Summary of Calibration Curve Parameters for PFOSAA in Human Serum Quadratic weighted 1/x2. All concentrationsare expressed as ppb.
Run Date
03-Jul-2000 21-Aug-2000
Run Number
23
A
B
C R-Squared LLOQ ULOQ
-0.000002 0.013656 -0.020557 0.9903
6.00 505
18
0.000005 0.025511 -0.020269 0.9978
A, B, and C are coefficients used to define the calibration curve.
6.00.,. 505 : :.
:- ..-..
Page 32
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 4. Summary of Calibration Curve Parameters for POAA in Human Serum Quadratic weighted I/x2. All concentrations are expressed as ppb.
RunDate
03-Jui-2000 21-Aug-2000
Run
A
B
C
R-Squared LLOQ ULOQ
Number
23
-0.000002 0.013683 0.010580 0.9979
5.76
505
18
0.000003 0.013704 0.017950 0.9966
5.76
505
A, B, and C are coefficients used to define the calibration curve.
Table 5. Summary of Calibration Curve Parameters for PFHS in Human Serum
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
RunDate
03-Iul-2000 21-Aug-2000
Run Number
23
A
B
C
R-Squared LLOQ
-0.000006 0.018323 0.012882 0.9971
3.15
18
-0.000001 0.022486 0.017181 0.9970
3.15
A, B, and C are coefficients used to define the calibration curve.
ULOQ
502 502
Table 6. Summary of Calibration Curve Parameters for M556 in Human Serum Quadratic weighted l/x2. All concentrations are expressed as ppb.
RunDate
03-Jul-2000 21-Aug-2000
Run
A
B
C IR-Squared LLOQ
Number
23
0.000000 0.002714 0.000585 0.9925
4.30
18
0.000001 0.011156 0.005643 0.9958
4.30
A, B, and C are coefficients used to def'me the calibration curve.
ULOQ
502 502
Table 7. Summary of Calibration Curve Parameters for M570 in Humail Quadratic weighted l/x". All concentrations are expressed as ppb.
Serum
.
" _. --. .o"
RunDate 03-Jul-2000 21-Aug-2000
Run Number
23
A
B
C
R-Squared LLOQ
0.000000 0.003818 -0.004160 0.9963
5.60
18
0.000005 0.036404 -0.017720 0.9977
5.60
A, B, and C are coefficients used to define the calibration curve.
ULOQ 505 505
Page 33
Northwest Bioanalytical Table 8. Back-Calculated
Concentrations
of Calibration Serum
Standards
Study No. NWBS00-040 Report No. NWBR00-122
for PFOS in Human
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date 03-Jul-2000 21-Aug-2000
Run Number
23
18
48.1 49.6 57.1 72.1 97.1 147 297 447 547
50.7 52.1 64.0 77.3 98.4 158 309 472 564 45.8 45.6 53.4 68.2 86.6 138 298 435 508 51.7 54.3 60.3 80.5 101 152 304 489 566 44.1 45.6 50.5 70.5 91.1 139 286 417 521
Table 9. Back-Calculated
Concentrations
of Calibration Serum
Standards
for PFOSA in Human
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
RunDate 03-Jul-2000 21-Aug-2000
Run Number
23
18
1.00 2.51 10.1 25.1 50.3 100 251 402 502
1.10 2.54 10.6 27.3 51.6 103 259 419 534 0.918 2.38 9.36 24.6 46.2 95.4 265 398 436 1.01 2.73 10.4 27.9 52.7 106 253 416 530 0.990 2.30 9.62 25.1 44.7 95.8 229 378 502
Table 10. Back-Calculated
Concentrations of Calibration Standards for PFOSAA in Human Serum
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
:
RunDate 03-Jul-2000 21-Aug-2000
Run Number
23
18
6.00 7.50 15.0 30.0 55.0 105 255 i_ 405 505
6.74 7.73 16.6 31.3 57.6 111 278 446 524 5.49 6.93 13.6 27.3 52.0 99.6 249 382 451 5.83 7.71 14.5 29.8 53.7 99.3 239 397 499 6.13 7.50 14.4 32.2 57.0 108 268 414 513
Page 34
Northwest Bioanalytical Table 11. Back-Calculated
Smdy No. NWBS00-040 Report No. NWBR00-122
Concentrations of Calibration Serum
Standards
for POAA in Human
Quadratic weighted I/x2. All concentrations are expressed as ppb.
.
Run Date
Run
5.76 7.26 14.8 29.8 54.8 105 255 405 505
Number
03-Jul-2000
23
5.93 7.11 15.2 30.6 52.0 109 265 406 524
5.74 6.93 15.6 29.4 53.1 99.6 261 404 474
21-Aug-2000
18
5.98 7.19 13.9 28.2 52.8 102 256 401 481
6.06 6.58 15.2 32.5 55.2 108 270 412 511
Table 12. Back-Calculated
Concentrations of Calibration Serum
Standards
for PFHS in Human
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date 03-Jul-2000
Run Number
23
21-Aug-2000
18
3.15
4.65 12.2 27.2 52.2 102 252 402 502
3.13
4.45 12.8 27.7 49.3 103 256 405 515
3.20 4.55 13.7 26.3 49.1 99.7 253 402 487
2.85 4.72 11.8 28.1 48.8 98.3 245 408 497 3.17 5.23 12.1 27.9 51.9 105 258 404 502
Table 13. Back-Calculated
Concentrations of Calibration Serum
Standards
for M556 in Human .:_
Quadratic weighted [Ix2. All concentrations are expressed as ppb.
:
" '=" "
Run Date 03-Jul-2000 21-Aug-2000
Run Number
23
18
4.30 11.8 26.8 51.8 102 252 _402 502
4.44 12.6 28.9 53.2 113 274 434 528 4.18 11.I 23.8 47.5 95.1 247 381 448 4.36 I 1.0 25.5 49.0 101 234 391 469 4.35 11.8 27.2 55.4 106 284 415 516
Page 35
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 14. Baek-Caleulated Concentrations of Calibration Standards for M570 in Human Serum
Quadratic weighted I/x2. All concentrationsare expressed as ppb.
Run Date 03-J'ul-2000 21-Aug-2000
Run Number
23
18
5.60 7.10 14.6 29.6 54.6 105 255 405 505
6.16 6.90 15.4 29.8 54.0 107 273 414 510 5.54 6.43 14.3 28.7 52.6 104 265 404 464 5.11 7.35 13.7 30.2 53.3 101 249 401 503 5.84 7.43 14.4 31.1 54.5 106 269 406 505
; ;
Page 36
Northwest Bioanalytical
Study No. NWBS00-040 ReportNo. NWBR00-122
Table 15. Quality control Samples for PFOS in Human Serum Mean of replicate QCs reported. All concentrations are expressed as ppb.
Run Date 03-Jul-2000
Run Number
23
2 l-Aug-2000
18
Low QC 51.1 ppb
47.7
46.7
%Dif. -6.65
Medium QC 197 ppb
168
%D/f. -14.7
-8.61
170
-13.7
High QC 446ppb
382
378
%Dif.
-14.3 -15.2
Table 16. Quality Control Samples for PFOSA in Human Serum
Mean of replicate QCs reported. All concentrations are expressed as ppb.,
Run Date
03-Jui-2000 21-Aug-2000
Run Number
23
18
Low QC 4.00 ppb
2.73
4.34
%D/f. Medium QC %D/f. 150 ppb
-31.8
100
-33.3
8.50
150
0.00
High QC 400 ppb
286
395
%Dif.
-28.5 -1.25
Table 17. Quality Control Samples for PFOSAA in Human Serum Mean of replicate QCs reported. All concentrations are expressed as ppb.
Run Date
03-Jul-2000 21-Aug-2000
Run Number
23
18
Low QC 9.00 ppb
7.73
9.95
%Dif. Medium QC %Dif. 155 ppb
-14.1
110
-29.0
10.6
160
3.23
High QC 405 ppb
288
413
%Dif.
-28.9 1.98
,:_
z
" _'. ,;.". _
Page 37
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 18. Quality Control Samples for POAA in Human Serum Mean of replicate QCs reported. All concentrations are expressed as ppb.
Run Date 03-Jul-2000
Run Number
23
2 l-Aug-2000
18
Low QC 8.74 ppb
8.45
7.82
%Dif. Medium QC %Dif. 154 ppb
-3.32
150
-2. 60
-10.5
149
-3.25
High QC 403 ppb
400
372
%Dif.
-0. 744 -7.69
Table 19. Quality Control Samples for PFHS in Human Serum Mean of replicate QCs reported. All concentrations are expressed as ppb.
Run Date
03-Jul-2000 21-Aug-2000
Run Number
23
18
Low QC 6.15 ppb
6.24
6.86
%Dif. Medium QC %D/f. 152 ppb
1.46
157
3.29
11.5
164
7.89
High QC 402 ppb
427
403
%Dif.
6.22 0.249
Table 20. Quality Control Samples for M556 in Human Serum Mean of replicateQCs reported. All concentrationsare expressedas ppb.
Run Date
03-Jul-2000 2 l-Aug-2000
Run Number
23
18
Low QC 5.80 ppb
5.52
5.42
%D/f. -4.83
Medium QC %D/f. 152 ppb
145
-4.61
High QC 402 ppb
381
-6.55
141
-7.24
367
%D/f.
-5.22 -8.71
Table 21. Quality Control Samples for M570 in Human Serum.:': Mean of replicateQCs reported. All concentrationsare expressedas ppb.
Run Date
03-Jul-2000 2l-Aug-2000
Run Number
23
18
Low QC 8.60 ppb
8.20
8.02
%D_ -4.65
Medium QC %Dif. 155 ppb
146
-5.81
-6.74
149 ' -3.87
High QC 405 ppb
390
386
%Dif.
-3.70 -4.69
Page 38
Northwest Bioanalytieal
Study No. NWBS00-040 Report No. NWBR00-122
Table 22. Freeze-Thaw and Room Temperature Matrkx Stability for PFOS in Human Serum
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Freeze-Thaw
46.8 45.1 44.0"
cycles
Mean
45.3
S.D.
1.41
% CV
3.1
Control (ng/m L)
46.7
Mean % Dev. from Control
-3.0
....After
MEDIUM QC 7 Freeze-Thaw cycles
170 158 164
Mean
164
S.D.
6.00
% CV
3.7
Control (ng/mL)
170
Mean % Dev. from Control
-3.5
HIGH QC
After 7 Freeze-Thaw cycles
":".
369 376 361
Mean
369
S.D.
7.51
% CV
2.0
Control(ng/mL)
378
Mean % Dev. from Control
-2.4
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods beo,veen each freeze cycle
Page 39
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 23. Freeze-Thaw and Room Temperature Matrix Stability for PFOSA in Human Serum
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW qc After 7 Freeze-Thaw
3.73 3.53 3.48
cycles
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
3.58 0.132
3.7
4.34 -17.5
MEDIUM QC After 7 Freeze-Thaw cycles
108 126 125
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
120 10.1 8.4
150 -20.0
HIGH QC o:'-. After 7 Freeze:Thaw cycles
:.
_,.'._
348 361
350
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
353 7.00 2.0
395 -10.6
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each freeze cycle
Page 40
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 24. Freeze-Thaw Stability and Room Temperature Matrix for PFOSAA in Human Serum
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Freeze-Thaw
9.08 9.85 9.46
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
9.46 0.385
4.1
9.95 -4.9
cycles
MEDIUM QC After 7 Freeze-Thaw cycles
::
114
147
147
Mean
Mean S.D.
% CV
Control (ng/m L) % Dev. from Control
136 19.1 14.0
160 -15.0
HIGH QC After 7 Freeze-Thaw
383. 403 391
cycles
.:_ :
:_: -
Mean
392
S.D.
10.1
L
% CV
2.6
Control (ng/m L)
413
IVlean % Dev. from Control
-5.1
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each freeze cycle
Page 41
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 25. Freeze-Thaw
and Room Temperature Human Serum
Matrix Stability for POAA in
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW Qc
After 7 Freeze-Thaw
8.06 8.83 8.25
cycles
Mean
Control % Dev. from
Mean S.D.
% CV
(ng/mL) Control
8.38 0.401
4.8
7.82 7.2
MEDIUM QC After 7 Freeze-Thaw cycles
151 149 154
Mean
151
S.D.
2.52
% CV
1.7
Control (ng/m L)
,
149
Mean % Dev. from Control
1.3
HIGH QC After 7 Freeze-Thaw
364' 379 372
cycles
Mean
372
S .D.
7.51
% CV
2.0
Control (ng/m L)
372
Mean % Dev. from Control
0.0
,:_:
:
_
_.._
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each fi:eeze cycle
Page 42
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 26. Freeze-Thaw
and Room Temperature Human Serum
Matrix Stability for PFHS in
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Freeze-Thaw
6.05 6.27 7.12
cycles
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
6.48 0.565
8.7
6.86 -5.5
MEDIUM QC After 7 Freeze-TlJaw cycles
165 157 163
Mean
162
S.D.
4.16
% CV
2.6
Control (ng/mL)
164
Mean % Dev. from Control
-1.2
HIGH QC After 7 Freeze.Thaw
397 417 4O5
cycles
Mean
Mean S.D.
% CV
Control (ng/m L) % Dev. from Control
406
10.1
'
2.5
403 0.7
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each freeze cycle
Page 43
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 27. Freeze-Thaw
and Room Temperature Human Serum
Matrix Stability for M556 in
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Freeze-Thaw
4.95 5.07 5.09
cycles
Mean
Mean S.D.
% CV
Control (ng/m L) % Dev. from Control
5.04 0.0757
1.5
5.42 -7.0
MEDIUM QC After 7 Freeze-Thaw cycles
128 140 135
Mean
134
S.D.
6.03
% CV
4.5
Control (ng/mL)
141
Mean % Dev. from Control
-5.0
HIGH QC After 7 Freeze-Thaw
359. 372 374
cycles
Mean
368
S.D.
8.14
% CV
2.2
Control (ng/m L)
367
Mean % Dev. from Control
0.3
,:_,
"" :'-.-
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each freeze cycle
Page 44
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 28. Freeze-Thaw and Room Temperature Matrix Stability for M570 in Human Serum
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Freeze-Thaw
7.68 8.76 8.51
cycles
Mean
Control % Dev. from
Mean S .D.
% CV
(ng/m L) Control
8.32 0.565
6.8
8.02 3.7
MEDIUM QC After 7 Freeze-Thaw cycles
142 142 143
Mean
Mean S.D.
% CV
Control (ng/m L) e/. Dev. from Control
142 O.577
0.4
149 -4.7
HIGH QC
After 7 Freeze-.Th aw cycles
":_".
356 371 370
M ean
366
S.D.
8.39
% CV
2.3
Control(ng/mL)
386
Mean % Dev. from Control
-5.2
i
Control = mean of analytical QC samples analyzed in the same run NOTE: Samples were left at room temperature for up to 6.75 hour periods between each freeze cycle
Page 45
Northwest Bioanalytical
Table 29. Reinjection Stability for PFOS
All concentrations are expressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOWQC
Rei::jeetion (After 2.5 Days Room Ten:_.)
49.1 48.0 50.0 46.1 45.1
Mean
47.7
S.D.
2.04
% CV
4.3
Control (ag/mL)
47.2
Mean % Dev. from Control
1.1
MEDIUM QC Reinjection
(After 2.5 Days Room Temp.
164 172 170 164 170
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
i 68 3.74 2.2
173 -2.9
HIGH QC Reinjection (After 2.5 Days Room Temp.)
391 369 368 372 411
Mean
382
S.D.
18.6
% CV
4.9
Control (ng/mL)
393
Heart % Dev. from Control
-2.8
Study No. NWBS00-040 Report No. NWBR00-122
':': :"
l.
"- "
Control= meanofanalyticaiOCsamplesanalyzedin run 15(originalinjection)
Page 46
Northwest Bioanalytical
Table 30. Reinjeetion
Stability for PFOSA
All concentrations are expressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOW QC Reinjection (After 2.5 Days Room Temp.)
2.79 2.69 2.85 2.66 2.68
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
2.73 0.0820
3.0
2.82 -3.2
MEDIUM QC Reinjection
(After 2.5 Days Room Temp.)
[02 Z02 97.9 94.4 105
Mean $.D.
% CV
Control (ngtmL) Mean % Dev. from Control
I00 4.14 4.1
103 -2.9
HIGH QC Reinjection (After 2.5 Days Room Temp.)
275 273 290
284
307
Mean S.D.
% CV
Control(ng/mL) Mean % Dev.fromControl
286 13.7 4.8
300 -4.7
Study No. NWBS00-040 Report No. NWBR00-122
s: _..
:
: _;..
Control = mean of analytical QC samples analyzed in run 15 (original injection)
Page 47
Northwest Bioanalytical
Table 31. Reinjeetion Stability for PFOSAA
All concentrations are expressed as ppb.
Study No. NWBS00-040 Report No. NWBR.00-122
Run Number Run Date
23 3-Jul-2000
LOW QC Reh_jection (After 2.5 Days Room Temp.)
7.78 7.58 8.15 7.41 7.74
Mean $.D.
% CV
Control (ng/mL) IVlean% Dev. from Control
7.73 0.276
3.6
7.59 1.8
MEDIUM QC Xeinjection (After 2.5 Days Room Temp.)
109 112 113 104 110
Mean
!I0
S.D.
3.51
% CV
3.2
Control (ng/mL)
113
Mean % Dev. from Control
-2.7
HIGH QC Rehzjectlon
(After 2.5 Days Room Temp.)
287 285
284 278 305
Mean
288
S.D.
10.2
% CV
3.5
Control (ng/mL)
30 I
lVlean% Dev. from Control
-4.3
.:. : :
: :...
Control = mean of analytical QC samples analyzed in run 15 (original injection)
Page 48
Northwest Bioanalytical
Table 32. Reinjection Stability for POAA All concentrationsareexpressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOW QC Reinjection (After 2.5 Days Room Temp.)
8.68
8.43 8.31 8.41 8.41
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
8.45 0.138 1.6
8.10 4.3
MEDIUM QC Reinjection
(After 2.5 Days Room Temp.)
150 152 156 144 147
Mean
!50
S.D.
4.60
% CV
3.1
Control (ng/mL)
144
Mean % Dev. from Control
4.2
HIGH QC Reinjection (After 2.5 Days Room Temp.)
417 408 386 388 401
Mean
];00
S.D.
13.2
% CV
3.3
Control (ng/mL)
384
Mean % Dev. from Control
4.2
Study No. NWBS00-040 Report No. NWBR00-122
,:_: :.
" .=..
Control= mean ofanalytical QC samplesanalyzed in run15 (original injection)
Page 49
Northwest Bioanal_ical
Table 33. Reinjection Stability for PFHS
All concentrations are expressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOW QC Reinjection (After 2.5 Days Room Temp.)
6.30 6.32 6.48 5.98 6.11
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
6.24 0.195 3.1
6.22 0.3
MEDIUM QC Relnjection
(After 2.5 Days Room Temp.
153 156 156 159 159
Mean
157
S.D.
2.51
% CV
1.6
Control (ng/mL)
158
Mean % Dev. from Control
-0.6
HIGH QC Reinjection (After 2.5 Days Room Temp.)
441 416
421 409
450
lVlean
427
S.D.
17.4
% CV
4.1
Control (ng/mL)
445
Mean % Dev. from Control
-4.0
Study No. NWBS00-040 Report No. NWBR00-122
.:_.. :.
"_. ,... -
Control = mean of analytical QC samples analyzed in run 15 (original injection)
Page 50
Northwest Bioanalytical
Table 34. Reinjeetion Stability for M556
All concentrations are expressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOW QC Reinjection (After 2.5 Days Room Temp.)
5.78 5.73 5.45 5.29 5.33
lVlean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
5.52 0.227 4.1
5.61 -I.6
MEDIUM QC Reinjection
.(After 2.5 Days Room Temp.)
144
147 147
,
143
146
/Vlean
145
S.D.
! .82
% CV
!.3
Control (ng/mL)
147
IVlean% Dev. from Control
-1.4
nIGH QC RMnjection (After 2.5 Days Room Temp.)
388 373 381
367 398
Mean
38 I
S.D.
12.2
% CV
3.2
Control (n_mL)
396
/Vlean % Dev. from Control
-3.8
Study No. NWBS00-040 Report No. NWBR00-122
:
; ..:. ,
Control = mean of analytical QC samples analyzed in run 15 (original injection)
Page 51
Northwest Bioanalytical
Table 35. Reinjeetion Stability for M570 All concentrations arc expressed as ppb.
Run Number Run Date
23 3-Jul-2000
LOWQc
Reinjection (After 2.5 Days Room Temp.)
8.31 8.36 8.52 8.10 7.71
Mean S.D.
% CV
Controi (ng/m L) Mean % Dev. from Control
8.20 0.312 3.8
8.3 l -1.3
MEDIUM QC Reinjection
(After 2.5 Days Room Temp.)
145 149 146 144 145
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
146 1.92 1.3
147 -0.7
HIGH QC Reinjection (After 2.5 Days Room Temp.)
391 384
399
380
405
IVlean
390
S.D.
9.51
% CV
2.4
Control (ng/mL)
401
Mean % Dev. from Control
-2.7
StudyNo. NWBS00-040 Report No. NWBR00-122
:.
: .: ..
Control = mean of analytical QC samples analyzed in run 15(original injection)
Page 52
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 36. Room Temperature Extract Stability for PFOS
All concentrations are expressed as ppb.
Run Number Run Date
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
18 21-Aug-2000
LOWQc
After 7Days
51.7 46.0 46.2
48.0 3.23 6.7 46.7 2.8
HIGH QC After 7 Days
373 402 397
391 15.5 4.0 378 3.4
-
:
" :..._...-.
Control -- mean of QC samples analyzed in the same run immediately following extraction
Page 53
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBP-.00-122
Table 37. Room Temperature Extract Stability for PFOSA All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Days
3.48 3.34 3.30
Mean
Mean S.D.
% CV Control (n_mL) % Dev. from Control
3.37 0.0945
2.8 3.13 7.7
HIGH QC After 7 Days
337 362 334
Mean
344
S.D.
15.4
% CV
4.5
j:t_
Control (ng/mL)
322
Mean % Dev. from Control
6.8
g
Control = mean of QC samples analyzed in the same run immediately following extraction
Page 54
NorthwestBioanalytical
StudyNo. NWBS00-040 Report No. NWBR00-122
Table 38. Room Temperature Extract Stability for PFOSAA All concentrationsare expressedas ppb.
Run Number
18
Run Date 21-Aug-2000
LOW QC After 7 Days
8.2 9.2 9.3
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
8.89 0.607
6.8
8.02 10.8
HIGH QC After 7 Days
321 337 343
Mean
334
S.D.
11.4
% CV
3.4
Control (ng/mL)
297
Mean % Dev. from Control
12;5
: ""
"_" i.d-.
Control_-meanof QCsamplesa"nalyzedinthesame runimmediatelyfollowingextraction
Page55
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 39. Room Temperature Extract Stability for POAA All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC
After 7 Days
6.92 7.98 8.15
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
7.68 0.667
8.7
7.82 -1.8
HIGH QC After 7 Days
360 395 378
Mean
378
S.D.
17.5
% CV
4.6
Control (ng/mL)
372
Mean % Dev. from Control
1.6
.
"_- .,,..
Control -- mean of QC samples analyzed in the same run immediately following extraction
Page 56
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 40. Room Temperature Extract Stability for PFHS Allconcentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC
After 7 Days
6.31 6.21 5.90
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
6.14 0.214 3.5
6.86 -10.5
mGrl QC
After 7 Days
396 405 412
Mean
404
S.D.
8.02
% CV
2.0
Control (ng/mL)
403
Mean % Dev. from Control
0.2
]
1
Control= mean of QC samplesanalyzedin thesamerun immediately following extraction
Page57
NorthwestBioanalytical
StudyNo.NWBS00-040 ReportNo.NWBR00-122
Table 41. Room Temperature Extract Stability for IV1556 All concentrationsareexpressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7 Days
5.13 5.65 5.50
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
5.43 0.268 4.9
5.42 0.2
HIGH QC After 7 Days
373 376 387
Mean
379
S.D.
7.37
% CV
1.9
Control (ng/mL)
36_/
Mean % Dev. from Control
3.3
o: : ,
_'. ,_
Control=mean of QCsamplesanalyzedin the samerun immediatelyfollowingextraction
Page58
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 42. Room Temperature Extract Stability for M570 All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC After 7Days
8.86 9.13 8.74
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
8.91 0.200
2.2
8.02 11.1
HIGH QC After 7 Days
409 429 423
Mean
420
S.D.
10.3
% CV
2.5
Control (ng/mL)
386
Mean % Dev. from Control
8.8
Control--mean of QCsamplesanalyzedin thesamerun immediatelyfollowingextraction
Page 59
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 43. Reduced Temperature (1 to 8 C) Extract Stability for PFOS All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW Qc
1 to 8C After 4 Days
49.3 45.6 44.8
Mean
46.6
S.D.
2.40
% CV
5.2
Control (ng/mL)
46.7
Mean % Dev. from Control
-0.2
HIGH QC 1 to 8C
After 4 Days
391 385 377
- Mean 384 o;" :
S.D.
7.02
% CV
1.8 :
;.
"_'..2..o
Control (ng/mL)
378
Mean % Dev. from Control
1.6
Control= meanof QCsamplesanalyzedin the samerun immediatelyfollowingextraction
Page 60
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 44. Reduced Temperature (1 to 8 C) Extract Stability for PFOSA All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC I to 8C After 4 Days
3.22 2.79 2.86
Mean S.D.
% CV
Control (n_mL) Mean % Dev. from Control
2.96 0.231
7.8
3.13 -5.4
HIGH QC 1 to 8C
After 4 Days
350 297 304
Mean S.D.
% CV
Control (n_mL) Mean % Dry. from Control
317 28.8 9.1
322 -1.6
,: : :. .
,
i
_" ..;-.
Control= meanof QC samplesanalyzedin thesame run immediatelyfollowingextraction
Page61
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 45. Reduced Temperature (1 to 8 C) Extract Stability for PFOSAA All concentrationsare expressedas ppb.
Run Number Run Date
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
Mean
Mean S.D'.
% CV
Control (ng/mL) % Dev. from Control
18 21-Aug-2000
LOWQC 1 to 8C After 4 Days
7.78 8.55 8.32
8.22 0.395 4.8 8.02 2.5
HIGH QC I to 8C
After 4 Days
312 306 311
310 3.21 1.0 297 4.4
..:
:.
, :'..o
,
Control= meanof QCsamplesanalyzedin the samerun immediatelyfollowingextraction
Page 62
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 46. Reduced Temperature (1 to 8 C) Extract Stability for POAA All concentrationsare expressedasppb.
Run Number Run Date
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
18 21-Aug-2000
LOW QC I to 8C After 4 Days
7.78 8.75 8.39
8.31 0.490
5.9 7.82 6.3
HIGH QC I to 8C
After 4 Days
381 394 385
387 6.66 1.7 372 4.0
' :. ,
" .-. _.,/-. ..
Control-- meanof QC samplesanalyzedin thesamerun immediatelyfollowingextraction
Page63
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 47. Reduced Temperature (1 to 8 C) Extract Stability for PFHS All concentrationsareexpressedas ppb;
Run Number Run Date
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
18 21-Aug-2000
LOW QC 1 to 8C After 4 Days
6.76 6.34 7.74
6.95 0.718 10.3 6.86
1.3
HIGH QC 1 to 8C
After 4 Days
400 411 423
411 11.5 2.8 " 403 2.0
.:_
. .".._
.
Control=meanof QCsamplesanalyzedin the samerunimmediatelyfollowingextraction
Page 64
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 48. Reduced Temperature (1 to 8 C) Extract Stability for M556
All concentrations are expressed as ppb.
Run Number Run Date
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
Mean
Mean S.D. % CV
Control (ng/mL) % Dev. from Control
18 21-Aug-2000
LOW QC 1 to 8C
After 4 Days
5.47 6.57 6.82
6.29 0.718 11.4 5.42 16.1
HIGH QC 1 to 8C
After 4 Days
398 404 403
402 3.9.1 0.8 367 9.5
_.
,
': _:-. _
Control = mean of QC samples analyzed in the same mn immediately following extraction
Page 65
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 49. Reduced Temperature (1 to 8 C) Extract Stability for M570 Allconcentrationsareexpressedas ppb.
Run Number Run Date
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
Mean S.D.
% CV Control (ng/mL) Mean % Dev. from Control
18 21-Aug-2000
LOW QC I to 8C After 4 Days
7.96 .7.90 9.21
8.36 0.740
8.9 8.02 4.2
HIGH QC I to 8C
After 4 Days
396 394 402
397 4.16 1.0 386 2.8
:.
'
_- ,_....
Control= meanof QCsamplesanalyzedinthe samerun immediatelyfollowingextraction
Page66
NorthwestBioanalytical
SmdyNo. NWBS00-040 ReportNo. NWBR00-122
Table 50. Reduced Temperature (-20 C) Extract Stability for PFOS Allconcentrationsare expressedas ppb.
Run Number
18
Run Date 21-Aug-2000
LOW QC -20oC
After 5 Days
47.2 45.5 40.9
Mean
44.5
S.D.
3.26
% CV
7.3
Control (n_mL)
46.7
Mean % Dev. from Control
-4.7
HIGH QC -20oC
After 5 Days
376 373 365
Mean
371
S.D.
5.69
% CV
1.5
Control (ng/mL)
378
Mean % Dev. from Control
-1.9
i
Control= mean ofQC samplesanalyzedin the same runimmediatelyfollowingextraction
Page67
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 51. Reduced Temperature (-20 C) Extract Stability for PFOSA All concentrationsareexpressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC -20oc
After 5 Days
2.95 3.16 3.14
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
3.08 0.116
3.8
3.13 -1.6
HIGH QC -20C
After 5 Days
312 312 296
Mean
307
.:_:
S.D.
9.24
% CV
3.0
Control (ng/mL)
322
.
Mean % Dev. from Control
-4.7
Control= meanof QC samplesanalyzedin thesamerunimmediatelyfollowingextraction
Page68
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 52. Reduced Temperature (-20 C) Extraet Stability for PFOSAA
All concentrations are expressed as ppb.
Run Number Run Date
18 21-Aug-2000
LOW Qc
-20C
After 5 Days
7.52 8.32 7.89
Mean S.D.
% CV Control (n_mL) Mean % Dev. from Control
7.91 0.40O
5.1 8.02 -1.4
HIGH QC -20oC
After 5 Days
294 306 287
Mean
296
S.D.
9.61
% CV
3.2
Control (ng/mL)
297
Mean % Dev. from Control
-0.3
:.
""-., _.-.
Control = mean of QC samples analyzed in the same run immediately following extraction
Page 69
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 53. Reduced Temperature (-20 C) Extract Stability for POAA All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC -20C
After 5 Days
7.84 10.2 6.85
Mean
8.30
S.D.
1.72
% CV
20.7
Control (n_mL)
7.82
Mean % Dev. from Control
6.1
HIGH QC -20oC
After 5 Days
351 373 386
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
370 1_/.7 4.8
372 -0.5
Control= meanof QCsamplesanalyzedinthe samerun immediatelyfollowingextraction
Page70
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 54. Reduced Temperature (-20 C) Extract Stability for PFHS All concentrationsareexpressedas ppb.
Run Number Run Date
18 21-Aug-2000
LOW QC -20oC
After 5 Days
6.68 6.06 7.70
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
6.81 0.828 12.2
6.86 -0.7
HIGH QC -20oC
After 5 Days
403 407 417
Mean
409
.:,:
S.D.
7_21
% CV
1.8
Control (ng/mL)
403
Mean % Dev. from Control
1.5
Control= meanof QCsamplesanalyzedin the same run immediatelyfollowingextraction
Page71
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 55. Reduced Temperature (-20 C) Extract Stability for M556 All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC -20oc
After $ Days
4.51 5.69 5.97
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
5.39 0.775 14.4
5.42 -0.6
mGn Qc
-20C
After 5 Days
358 373 378
Mean
370
S.D.
10.4
%CV
2.8
Control (ng/mL)
367
Mean % Dev. from Control
0.8
""
:'_'"
Control= meanof QCsamplesanalyzedin the samerun immediatelyfollowingextraction
Page72
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 56. Reduced Temperature (-20 C) Extract Stability for M570 All concentrationsare expressedasppb.
Run Number Run Date
18 21-Aug-2000
LOW QC -20oc
After 5 Days
8.43 8.52 7.94
Mean
Mean S.D.
% CV
Control (ng/mL) % Dev. from Control
8.30 0.3 I2
3.8
8.02 3.5
HIGH QC -20C
After 5 Days
366 388 369
Mean S.D.
% CV
Control (ng/mL) Mean % Dev. from Control
374 1"1.9 3.2
386 -3.1
Control= meanof QC samplesanalyzedin thesamerun immediatelyfollowingextraction
Page73
Northwest Bioanalyfical
Study No. NWBS00-040 Report No. NWBR00-122
Table 57. Long-Term Matrix Stability for PFOS in Human Serum All concentrations are expressed as ppb.
Run Date 21-Aug-2000
Run Number
18
Low QC 51.1 ppb 55 Days
52.2 50.1 42.3 43.8 45.0
Medium QC 197 ppb 55 Days
175 171 174 I66 165
High QC 446 ppb 55 Days
381 391 361 371 388
Mean SD
%CV %Bias
N
46.7
170
378
4.26
4.55
12.4
9.1
2.7
3.3
-8.6
-13.7
-15.2
5
5
5
Table 58. Long-Term Matrix Stability for PFOSA in Human Serum All concentrations are expressed as ppb.
Run Date 21-Aug-2000
Run Number
18
Low QC 4.00 ppb 42 Days
4.28 4.56
4.40
4.12 4.32
Medium QC 150 ppb 42 Days
155 158
154.
142 143
High QC 400 ppb 42 Days
376 420
381 ':':
397
..
400
: ,: .-
Mean SD
%CV %Bias
N
4.34
150
395
0.161
7.37
17.4
3.7
4.9
4.4
8.5
0.0
-1.3
5
5
5
Page 74
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 59. Long-Term Matrix Stability for PFOSAA in Human Serum All concentrations are expressed as ppb.
Run Date 21-Aug-2000
Run Number
18
Low QC 9.00 ppb 42 Days
9.64 9.11 10.5 10.2 10.3
Medium QC 155 ppb 42 Days
157 158 164 159 160
High QC 405 ppb 42 Days
403 429 406 413 416
Mean SD
%CVi %Bias
N
9.95
160
413
0.568
2.70
10.2
5.7
1.7
2.5
10.6
3.2
2.0
5
5
5
Table 60. Long-Term MatrLx Stability for POAA in Human Serum All concentrations are expressed as ppb.
Run Date 21-Aug-2000
Run Number
18
,.,
Low QC 8.74 ppb 55 Days
7.74 7.54
7.72
8.05 8.03
Medium QC 154 ppb 55 Days
145 138
155
151 157
High QC 403 ppb 55 Days
364 380
354 .:':
371
.
389
J
Mean SD
%CV %Bias
N
7.82
149
372
0.219
7.76
13.6
2.8
5.2
3.7
-10.5
-3.2
-7.7
5
5
5
Page 75
Northwest Bioanalytical
Study No. NWB$00-040 Report No. NWBR00- !22
Table 61. Long-Term Matrix Stability for PFHS in Human Serum All concentrations are expressed as ppb.
Run Date
Run Number
2 l-Aug-2000
18
Low QC 6.15 ppb 55 Days
7.01 7.06 7.10 6.57
6.56
Medium QC 152 ppb
55 Days
163 158 169 161
168
High QC 402 ppb 55 Days
387 414 375 404
435
Me_ SD
%CV] %Bias
N
6.86
164
403
0.271
4.66
23.4
4.0
2.8
5.8
11.5
7.9
0.2
5
5
5
Table 62. Long-Term Matrix Stability for M556 in Human Serum All concentrations are expressed as ppb.
Run Date 2l-Aug-2000
Run Number
18
Low QC 5.80 ppb 55 Days
6.07 5.67
*4.60
4.97 5.79
Medium QC 152 ppb 55 Days 141 137
149
137 143
High QC 402 ppb 55 Days
355 382
340':
36.2: : . 394
l'
Mean
5.42
141
367
SD
0.612
4.98
21.5
I
%CV
11.3
3.5
5.9
%Bias
-6.6
-7.2
-8.7
N
5
5
5
* > :i:20% deviation from theoretical
: ,_- ,.
Page 76
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 63. Long-Term Matrix Stability for M570 in Human Serum All concentrations are expressed as ppb.
Run Date 2 l-Aug-2000
Run Number
18
Low QC 8.60 ppb 55 Days
7.93 8.07 7.87 8.04 8.20
Medium QC 155 ppb
- 55 Days
147 147 158 144 150
High QC 405 ppb 55 Days
373 391 373 384 408
Mean SD
%CV %Bias
N
8.02
149
386
0.128
5.36
14.6
1.6
3.6
3.8
-6.7
-3.9
-4.7
5
5
5
Page 77
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 64. PFOS Extraction Efficiency from Human Serum
Run 22
Low Concentration = 4.00 ng/mL
Time of Spiking
!. Analyte and IS after extraction
Mean
Ratio AnalyteflS 0.0679 0.0683 0.0735
0.0699
Ratio
IS/Analyte 14.7 14.6 13.6
14.3
2. Analyte prior and IS after extraction
Mean
0.0656 0.0605
0.0649 0.0637
3. IS prior and Analyte after extraction
Mean
0.775 0.726 0.805
0.769
Adjusted mean extraction efficiency = 44.4% Mean extraction efficiency for the internal standard = 5.38%
High Concentration = 400 ng/mL
T.irne of Spiking
1. Analyte and IS after extraction
Mean
Ratio Analyte/IS 1.65 1.55
1.65 1.62
Ratio IS/Analyte 0.605 0.643
0.607 0.618
2. Analyte prior and IS afc_ extraction
Mean
0.522
0.538 0.518 0.526
3. IS prior and Analyte after extraction
Mean
0.0396
:
0.0379
0.0401 :
0.0392
: -" --
Adjusted mean extraction efficiency = 30.1% Mean extraction efficiency for the internal standard = 6.34%
Overall mean e,,ctraction efjTciency for analyte ffi 37.3% Overall mean extraction efficiency for internal standardffi 5.86%
* Due to rounding, calculations based on displayed values may differ from actual.
Page 78
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 65. PFOSA Extraction Efficiency from Human Serum
Run 22
Low Concentration = 4. O0 ng/mL
.Time of Spikin_ I. AnalyteandIS aRerextraction
Mean
Ratio Anal,fte/IS 0.0275 0.0222 0.0214 0.0237
2. Analytepriorand ISafterextraction Mean
0.0I 7 l 0.0147 0.0172
0.0163
Ratio IS/Analvte 36.3 45.0 46.8 42.7
3. 1SpriorandAnalyteafterextraction
2.47
2.25
2.45
Mean
2.39
Mean extraction efficiency for the internal standard -- 5.60%
High Concentration = 400 ng/mL
Time of Spiking 1.AnalyteandIS afterextraction
Mean
Ratio Analvte/IS 2.58 2.47 2.65
2.57
Ratio IS/Analvte 0.387 0.405 0.377
0.390
2. AnalytcpriorandISafterextraction
1.80
1.87
1.75
Mean
1.81
,: _.
3. lSpriorandAnalytcafterextraction Mean
0.0243
:
0.0223
0.0233. :
0.0233
Mean extraction efficiency for the analyte = 70.4%
Overall mean extraction efficiency for analyte = 69.6% Overall mean extraction efficiency for internal standard = 5. 79%
* Due to rounding, calculations based on displayed values may differ from actual.
"_ .,i'..
Page 79
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 66. PFOSAA Extraction Effieieney from Human Serum
Run 22
Low Concentration = 4.00 ng/mL
Time of Spiking
1. Analyte and 1Safter extraction
Mean
Ratio Analyte/IS 0.0246 0.0257
0.0238 0.0247
Ratio IS/Anal_e 40.6 39.0
42.0 40.5
2. Analyte prior and IS after extraction
Mean
0.0206 0.0186 0.0194
0.0195
3. IS prior and Analyte after extraction
Mean
2.71 2.45 2.58
2.58
Adjusted mean extraction efficiency = 62.6% Mean extraction efficiency for the internal standard = 6.37%
High Concentration = 400 ng/mL
Time of Spiking
1. Aaalyt and IS after extraction
Mean
Ratio Analyte/IS 1.50 1.44
1.56 1.50
Ratio IS/Analyte 0.665 0.693
0.641 0.666
2. Analytc prior and I$ after extraction
Mean
3. IS prior and Analyte aRer extraction
Mean
0.962
0.959 0.917
0.946
,:,:
0.0419
:
0.0394
0.0413
0.0409
Adjusted mean extraction efficiency = 62.7% Mean extraction efficiency for the internal standard = 6.14%
Overall mean extraction efficiency for analyte = 62. 7% Overall mean extraction efficiency for internal standard = 6.26%
: -" -
* Due to rounding, calculations based on displayed values may differ from actual.
Page 80
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 67. POAA Extraction Efficiency from Human Serum
Run 22
Low Concentration = 4. 00 ng/mL
Time of Spiking I. AnalyteandIS afterextraction
Mean
Ratio Analyte/IS 0.0288 0.0291 0.0302
0.0294
Ratio IS/Analyte 34.7 34.3 33.1
34.0
2. AnalytepriorandIS afterextraction Mean
0.00934 0.00760 0.0101
0.00901
3. ISpriorandAnalyteafter extraction
2.02
1.93
2.05
Mean
2.00
Adjusted mean extraction efficiency -- 16.8"/, Mean extraction efficiency for the internal standard -- 5.88/,
High Concentration "_-400 ng/mL
Time of Spiking I.AnalyteandISafterextraction
Mean
Ratio Anal_e/IS 2.61 2.59
2.72 2.64
Ratio IS/Analyte 0.383 0.386
0.367 0.379
2. AanlytepriorandISafterextraction
0.377 0.385
0.383
Mean
0.382
.:'..
3. Is priorandAnalyteafterextraction Mean
0.0248 0.0227
0.0241 -" 0.0239
Adjusted mean extraction efficiency = 14.3"/o Mean extraetion efficiency for the internal standard = 6.31%
Overall mean extraction effieiency for analyte = 15. 6% Overall mean _vtraction efdqciency for internal standard = 6.10%
* Due to rounding, calculations based on displayed values may differ from actual.
Page 81
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 68. PFHS Extraction Efficiency from Human Serum
Run 22
Low Concentration = 4.00 ng/mL
Time of S_iking 1.AnalyteandIS afterextraction
Mean
Ratio Analyte/IS 0.0367 0.0361 0.0366 0.0365
Ratio
IS/Analyte 27.3 27.7 27.3 27.4
2. Ana/ytepriorandISafterextraction Mean
0.00870 0.00778 0.00878
0.00842
3. ISpriorandAnalytealterextraction
1.72
1.63
1.71
Mean
1.69
Adjusted mean extraction efficiency = 16.3% Mean extraction efficiency for the internal standard = 6.17%
High Concentration = 400 ng/mL
Time of Spiking I. Analytand ISal_erextraction
Mean 2. Anal_e priorand IS alterextraction
Mean 3. IS priorandAnalytafterextraction
Ratio Analyte/IS 3.49 3.29 3.58 3.45
0.486 0.514 0.485 0.495
Mean
Ratio IS/Analyte 0.286 0.304 0.279 0.290
,:_..
0.0192 0.0181 0.0192 " 0.0188
Adjusted mean extraction efficiency ffi 14.3% Mean extraction efficiency for the internal standard = 6.48%
Overall mean extraction efficiency for analyte = 15.3% Overall mean extraction effidency for internal standard = 6.33%
:" _:'" "
* Due to rounding, calculations based on displayed values may differ from actual.
Page 82
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 69. MSS6 Extraction Efficiency from Human Serum
Run 22
Low Concentration = 4.00 ng/mL
Time of Spiking i. AnalyteandIS afterextraction
Mean
Ratio Analyte/IS 0.00786 0.00750 0.00810 0.00782
Ratio IS/Analyte 127 133 123 128
2. Analytepriorand IS afterextraction Mean
0.00421 0.00366
0.00310 0.00366
3. ISpriorand Analyteafterextraction
7.54
6.18
6.78
Mean
6.83
Mean extraction efficiency for the analyte = 46.8% Adjusted mean extraction efficiency = 37.7*/,
High Concentration = 400 ng/mL
Time of Spiking 1.Analyteand IS after extraction
Mean 2. Analytepriorand IS afterextraction
Mean
Ratio Analyte/IS 0.788 0.758 0.836 0.794
0.337 0.326 0.303 0.322
Ratio
IS/Analyte 1.27 1.32
1.20 1.26
.:',
3. ISpriorand Anal_e after extraction Mean
0.0753 0.0690
0.0709 0.0717
: "-
Mean extraction efficiency for the analyte = 40.6% Adjusted mean extraction efficiency = 40.4%
Overall mean extraction efficiency for analyte = 39.1% Overall mean extraction efficiency for internal standard = 5.52%
: "" "
* Due to rounding, calculations based on displayed values may differ from actual.
Page 83
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 70. IVI570 Extraction Efficiency Run 22
from Human Serum
Low Concentration = 4. O0 ng/mL
Time of Spiking
I. Analyte and IS after extraction
Mean
Ratio AnalyteflS 0.0333 0.0317
0.0335 0.0328
Ratio IS/Analyte 30.1 31.5
29.9 30.5
2. Analyte prior and IS after extraction
Mean
0.0244
0.0220 0.0244 0.0236
3. IS prior and Analytc after extraction
Mean
1.92 1.72
1.74 1.79
Adjusted mean extraction efficiency ffi 54.4*/, Mean extraction efficiency for the internal standard = 5.87%
High Concentration = 400 ng/mI,
Time of Spiking
I. Analyte and IS after extraction
Mean
Ratio Analyte/IS 2.20 2.16 2.35 2.24
Ratio IS/Analyte 0.454 0.463 0.426 0.448
2. Analyte priorand IS after extraction
Mean
3. IS prior and Analyte after extraction
Mean
1.34
1.37 1.32
1.34
.: ..
0.0294 0.0270 0.0280 " 0.0281
Adjusted mean extraction efficiency = 59.7% Mean extraction efficiency for the internal standard = 6.27%
Overall mean extraction effieiency for analyte= 57.1% Overall mean extraction efficiency for internal standard = 6.07%
i
* Due to rounding, calculations based on displayed values may differ from actual.
: "z:"
Page 84
Northwest Bioanaly'tical
Study No. NWBS00-040 Report No. NWBR00-122
Table 71. Summary of Calibration Curve Parameters for PFOS in Human Plasma
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date Run Number
12-Oct-2000 27-Oct-2000
S00091-2 PFOSMD-40
A -0.000007 0.000000
B
C
R-Squared
0.030663 0.028132
-0_005620 0.010586
0.9986 0.9931
LLOQ ULOQ 3.94 414 3.94 414
A, B, and C are coefficients used to define the calibration curve.
Table 72. Summary of Calibration Curve Parameters for PFOSA in Human Plasma
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
RunDate Run Number
12-Oct-2000 27-Oct-2000
S00091-2 PFOSMD-40
A -0.000011 0.000001
B
C
R-Squared
0.073810 0.052188
0.006421 0.010966
0.9970 0.9912
LLOQ ULOQ 1.00 500 1.00 500
A, B, and C are coefficients used to define the calibration curve.
Table 73. Summary of Calibration Curve Parameters for PFOSAA in Human Plasma
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date
12-Oct-2000 27-0ct-2000
Run Number S00091-2
PFOSMD-40
A
-0.000003 0.000005
B
C
R-Squared
0.032601 0.021209
-0.006279 -0.002450
0.9988 0.9873
L.t,OQ
1.60 {.60
ULOQ :
501
501
A, B, and C are coefficients used to define the calibration curve.
Page 85
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Table 74. Summary of Calibration Curve Parameters for POAA in Human Plasma
Quadratic weighted I/x 2. All concentrations are expressed as ppb.
Run Date
12-Oct-2000 27-Oct-2000
Run Number
S00091-2
PFOSMD-40
A
0.000000 0.000002
B
0.017245 0.014346
C
R-Squared
0.001072 0.002192
0.9985 0.9918
LLOQ
1.92 1.92
ULOQ
481 481
A, B, and C are coefficients used to define the calibration curve.
Table 75. Summary of Calibration Curve Parameters for PFHS in Human Plasma
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date
12-Oct-2000 27-Oct-2000
Run Number
S00091-2
PFOSMD-40
A
-0.000005 -0.000001
B
C
R-Squared
0.025818 0.025534
-0.000970 0.011247
0.9983 0.9953
LLOQ ULOQ
1.36 523 1.36 523
A, B, and C are coefficients used to define the calibration curve.
Table 76. Summary of Calibration Curve Parameters for M556 in Human Plasma
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date
12-Oct-2000 31-Oct-2000
Run Number
S00091-2 24
A
-0.000002 -0.000001
B
0.018315 0.018373
C
0.005561 0.001923
R-Squared LI_()Q ULOQ
"
:" -
0.9954
2.,50 500
0.9972
2.50
500
A, B, and C are coefficients used to define the calibration curve.
Page 86
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 77. Summary of Calibration Curve Parameters for M570 in Human Plasma
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
RunDate Run Number
A
B
C
R-Squared LLOQ ULOQ
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
-0.000006 0.000001
0.048518 0.034031
0:005367 0.011681
0.9978 0.9910
1.00 500 1.00 500
A, B, and C are coefficients used to define the calibration curve.
Page 87
NorthwestBioanalytical
StudyNo. NWBS00-040 Report No. NWBR00-122
Table 78. Back-Calculated
Concentrations of Calibration Plasma
Standards
for PFOS in Human
QuadraticweightedL/x2. Allconcentrationsare expressedas ppb.
Run Date Run Number 3.94 5.17 11.3 23.6 44.2 85.4 209 332 414
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
4.00 5.01 11.2 25.0 42.9 83.7 203 335 390 3.86 5.35 11.1 24.3 43.5 85.3 209 350 424 3.72 5.18 11.3 19.4 45.5 82.1 214 337 369 3.90 5.67 11.7 24.1 45.9 84.4 221 367 407
Table 79. Back-Calculated
Concentrations of Calibration Standards for PFOSA in Human Plasma
Quadraticweighted1/x2. All concentrationsareexpressedasppb.
Run Date Run Number 1.00 2.50 10.0 25.0 50.0 100 250 400 500
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
1.04 2.60 10.4 25.1 49.1 95.9 252 380 462 0.941 2.53 8.93 25.7 50.5 102 255 444 508 0.961 2.59 9.81 20.2 53.4 103 259 356 443 0.999 2.71 10.4 22.1 52.3 101 267 458 515
Table 80. Back-Calculated
Concentrations of Calibration Standards for PFOSAA in Human Plasma
.,. -
Quadraticweightedl/x2. All concentrationsare expressedas ppb.
-.
": ._..--
Run Date Run Number 1.60 3.10 10.6 25.6 50.6 101 251 401 501
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
1.65 3.12 10.7 25.3 50.5 98.2 255 393 469 1.54 3.10 10.4 26.3 51.0 101 245 432 513 1.66 3.50 11.7 21.3 53.1 104 299 406 440 1.47 2.96 10.8 23.5 47.3 90.3 261 441 475
Page88
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBI_00-122
Table 81. Back-Calculated
Concentrations of Calibration Plasma
Standards
for POAA in Human
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date IRunNumber 1.92 3.36 10.6 25.0 49.0 97.1 241 385 481
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
2.00 3.24 10.5 26.1 49.3 97.5 241 386 477 1.96 3.11 10.1 25.2 52.1 95.0 239 392 477 1.73 3.01 10.5 22.9 49.7 99.1 250 367 439 2.30 3.12 10.9 25.1 50.7 96.7 262 422 474
Table 82. Back-Calculated
Concentrations of Calibration Plasma
Standards
for PFHS in Human
Quadratic weighted l]x 2. All concentrations are expressed as ppb.
Run Date Run Number 1.36 2.92 10.7 26.5 52.6 104 262 418 523
12-Oct-2000 S00091-2 27-Oct-2000 PFOSMD-40
1.28 2.74 10.6 27.5 50.2 104 253 441 518 1.44 3.08 10.8 26.9 53.2 104 259 413 524 1.52 2.72 10.6 24.2 54.6 107 259 466 489 1.28 2.78 10.4 28.4 54.1 102 270 425 491
Table 83. Back-Calculated
Concentrations of Calibration Standards for M556 in Human Plasma
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
:.
::-
Run Date Run Number 2.50 10.0 25.0 50.0 100 250 400; 500
12-Oct-2000 31-Oct-2000
S00091-2 24
2.48 11.3 27.9 49.7 101 259 398 521 2.47 9.43 24.5 43.7 96.7 245 384 499 2.33 9.67 24.3 51.9 95.2 255 388 499 2.70 9.81 25.8 49.3 104 249 438 470
Page 89
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 84. Back-Calculated
Concentrations of Calibration Standards for M570 in Human Plasma
Quadraticweightedl/x:. All concentrationsare expressedasppb.
Run Date Run Number 1.00 2.50 10.0 25.0 50.0 100 250 400 500
12-Oct-2000 27-Oct-2000
S00091-2 PFOSMD-40
0.991 1.00 0.993 0.951
2.57 10.2 25.0 50.8 97.0 250 380 451 2.48 9.46 25.4 50.3 101 244 448 528 2.71 10.6 21.2 53.6 104 284 401 444 2.65 10.4 20.9 48.6 93.5 254 443 481
Page90
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBP.00-122
Table 85. Lower Limit of Quantitation for PFOS in Human Plasma All concentrations are expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOS MD 40
LLOQ (3.94 ppb)
3.98 3.51 3.89 4.04 4.06 3.77
Mean S.D.
3.88 0.208
%CV
5.4
%Theoretical
98.5
%Bias
-1.5
N
6
Page 91
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Table 86. Lower Limit of Quantitation for PFOSA in Human Plasma All concentrations are expressed as ppb.
Run Date
Run Number
27-Oct-2000 PFOS MD 40
LLOQ (1.00 ppb)
1.08 1.07 1.11 0.900 1.01 0.865
Mean S.D. %CV
%Theoretical %Bias N
1.01 0.102 10.1 101.0
1.0 6
Page 92
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 87. Lower Limit of Quantitation for PFOSAA in Human Plasma All concentrations are expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOS MD 40
LLOQ (1.60 ppb)
1.63 1.58 1.54 1.39 1.63 1.57
Mean S.D.
%CV %Theoretical
%Bias N
1.56 0.0889
5.7 97.5 -2.5
6
I
Page 93
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 88. Lower Limit of Quantitation for POAA in Human Plasma All concentrationsare expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOS MD 40
LLOQ (1.92 ppb)
1.72 1.84 1.80 1.74 1.79 1.62
Mean S.D. %CV
%Theoretical %Bias N
1.75 0.0776
4.4 91.1 -8.9
6
j:.
Page 94
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Table 89. Lower Limit of Quantitation for PFHS in Human Plasma All concentrations are expressed as ppb.
Run Date Run Number 27-Oet-2000 PFOS MD 40
LLOQ (1.36 ppb)
1.23 1.29 1.36 1.35 "1.06 1.42
Mean
1.29
S.D.
0.128
%CV
9.9
%Theoretical
94.9
%Bias
-5.1
N
6
*> + 20% deviation from theoretical
m; :
Page 95
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 90. Lower Limit of Quantitation for M556 in Human Plasma All concentrations are expressed as ppb.
Run Date 31-Oct-2000
Run Number
24
LLOQ (2.50 ppb)
2.44 2.25 2.50 2.65 2.50 2.34
Mean S.D. %CV
%Theoretical %Bias N
2.45 0.139 5.7 98.0 -2.0
6
Page 96
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 91. Lower Limit of Quantitation for M570 in Human Plasma All concentrations are expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOS MD 40
LLOQ (1.00 ppb)
1.14 1.02 1.01 1.01 1.01 0.997
Mean S.D. %CV
%Theoretical %Bias N
1.03 0.0538
5.2 103.0 3.0
6
Page 97
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 92. Intra-Assay Precision for PFOS Quality Controls in Human Plasma
All concentrations are expressed as ppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 6.40 ppb
7.33 7.38 *7.89 7.07 6.98
Medium QC 126 ppb 132 129 138 132 137
High QC 332 ppb
343 331 334 334 340
Mean S.D.
%CV %Theoretical
%Bias n
7.19 0.195
2.7 112.3 12.3
4
134 3.78 2.8 106.3 6.3
5
336 4.93 1.5 101.2 1.2
5
* Sample deactivated due to unacceptable internal standard response.
:
_.
g
"":" . ...."-
Page 98
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 93. Intra-Assay Precision for PFOSA Quality Controls in Human Plasma
All concentrations are expressed as ppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 4.00 ppb
4.69 4.73 *4.67 4.53 4.66
Medium QC 150 ppb 148 157 172 166 163
High QC 400 ppb
423 409 425 414 409
Mean S.D.
%CV %Theoretical
%Bias n
4.65 0.0866
1.9 116.3 16.3
4
161 9.15 5.7 107.3 7.3
5
416 7.62 1.8 104.0 4.0
5
* Sample deactivated due to unacceptable internal standard response.
,
Page 99
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 94. Intra-Assay Precision for PFOSAA Quality Controls in Human Plasma
All concentrations are expressed as ppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 4.60 ppb
5.20 5.50 *7.43 4.96 4.89
Medium QC 151 ppb 146 148 169 156 155
High QC 401 ppb
411 423 431 404 425
Mean S.D.
%CV %Theoretical
%Bias n
5.14 0.276
5.4 111.7 11.7
4
155 9.04 5.8 102.6 2.6
5
419 11.0 2.6 104.5 4.5
5
* Sample deactivated due to unacceptable internal standard response.
Page 100
Northwest Bioanalyfical
Study No. NWBS00-040 Report No. NWBR00-122
Table 95. Intra-Assay Precision for POAA Quality Controls in Human Plasma
All concentrations are expressed as ppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 4.81 ppb
5.33 5.17 *7.82 4.86 4.69
Medium QC 145 ppb 144 157 163 155 152
High QC 385 ppb
400 395 399 394 401
Mean S.D.
%CV %Theoretical
%Bias n
5.01 0.290
5.8 104.2 4.2
4
154 6.98 4.5 106.2 6.2
5
398 3.11 0.8 103.4 3.4
5
* Sample deactivated due to unacceptable internal standard response.
,.;
Page 101
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 96. Intra-Assay Precision for PFHS Quality Controls in Human Plasma All concentrationsare expressedasppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 4.48 ppb
4.91 4.80 "5.4I 4.72 4.58
Medium QC 156 ppb 162 158 162 150 158
High QC 417 ppb
406 388 399 403 401
Mean S.D.
%CV %Theoretical
%Bias n
4.75
158
399
0.139
4.90
6.88
2.9
3.1
1.7
106.0
101.3
95.7
6.0
1.3
-4.3
4
5
5
* Sampledeactivateddueto unacceptableinternalstandardresponse.
R; _
Page 102
NorthwestBioanalytieal
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 97. Intra-Assay Precision for M556 Quality Controls in Human Plasma All concentrationsare expressedas ppb.
Run Date Run Number
12-Oct-2000
S00091-2
Low QC 4.00 ppb
4.65 4.35 *4.76 4.14 3.89
Medium QC 150 ppb 147 140 150 146 159
High QC 400 ppb
441 **492
459 403 431
Mean S.D.
%CV %Theoretical
%Bias n
4.26 0.322
7.6 106.5 6.5
4
148 6.95 4.7 98.7 " -1.3
5
445 33.1 7.4 111.3 I 1.3
5
* Sampledeactivateddue to unacceptableinternalstandardresponse. **> +__20d%eviationfromtheoretical
g
Page 103
Q
Northwest Bioanalytical
S tudy No. NWBS00-040 Report No. NWBR00-122
Table 98. Intra-Assay Precision for M570 Quality Controls in Human Plasma
All concentrations are expressed as ppb.
Run Date Run Number 12-Oct-2000 S00091-2
Low QC 4.00 ppb
**4.82 4.75 *5.89 4.56 4.74
Medium QC 150 ppb 146 161 170 165 158
High QC 400 ppb
407 407 414 393 403
Mean S.D.
%CV %Theoretical
%Bias n
4.72 0.111
2.4 118.0 18.0
4
160 9.03 5.6 106.7 6.7
5
405 7.69 1.9 101.3 1.3
5
* Sample deactivated due to unacceptable internal standard response. ** > +_20% deviation from theoretical
0..'.. :
" _" ;.;". _
Page 104
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 99. Summary of Calibration
Curve Parameters Buffer
for PFOS in Serum Diluted with
Quadraticweightedl/x:. All concentrationsare expressedas ppb.
Run Date [Run Number
A
B
C
R-Squared
23-Oct-2000 PFOSMD-39 -0.000001 0.009444 -0.004315 0.9954
LLOQ ULOQ 5.30 504
A, B, and C are coefficientsused to definethe calibrationcurve.
Table 100. Summary of Calibration Curve Parameters for PFOSA in Serum Diluted with Buffer
Quadraticweighted1/x2. Allconcentrationasreexpressedas ppb.
Run Date Run Number
A
23-Oct-2000 PFOSMD-39 0.000000
B
C
R-Squared LLOQ ULOQ
0.014752 0.002265
0.9946
1.00 500
A, B,and Care coefficientsusedto definethe calibrationcurve.
Table 101. Summary of Calibration Curve Parameters for PFOSAA in Serum Diluted with Buffer
Run Date 23-Oct-2000
Quadraticweightedl/x2. Allconcentrationsare expressedas ppb.
Run Number
PFOSMD-39
A 0.000002
B 0.009880
C
.R-Squared
0.004993
0.9960
m: :
L,LOQ: ULOQ - ..
{.00 500
A, B, and C are coefficientsused to definethecalibrationcurve.
Page 105
Q
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 102. Summary of Calibration
Curve Parameters Buffer
for POAA in Serum Diluted with
Quadratic weighted l/x _-. All concentrations are expressed as ppb.
Run Date 23-Oct-2000
Run Number
PFOSMD-39
A -0.000001
B 0.011351
C
R-Squared
-0.000137 0.9967
LLOQ ULOQ 1.70 501
A, B, and C are coefficients used to define the calibration curve.
Table 103. Summary of Calibration Curve Parameters for PFHS in Serum Diluted with Buffer
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date 23-Oct-2000
Run
A
B
C
R-Squared
Number
PFOSMD-39 -0.000003 0.018220 0.007453
0.9980
i,
A, B, and C are coefficients used to define the calibration curve.
LLOQ ULOQ 1.00 500
Table 104. Summary of Calibration Curve Parameters for M556 in Serum Diluted with Buffer
QuadratwiecighteId/2x.Allconcentratairoeenxspressaespdpb. .:'..
Run Date 23-Oct-2000
Run Number
PFOSMD-39
A 0.000000
B
C
R-Squared LLOQ uLoQ :
0.008729 -0.000146 0.9936
1.00 500
A, B, and C are coefficients used to define the calibration curve.
Page 106
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Table 105. Summary of Calibration
Curve Parameters Buffer
for M570 in Serum Diluted with
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date RunNumber 23-Oct-2000 PFOSMD-39
A 0.000000
B
C
R-Squared LLOQ ULOQ
0.015628 0.005711 0.9969
1.00 500
A, B, and C are coefficients used to define the ealibration curve.
Page 107
Northwest BioanalyticaI
Study No. NWBS00-040 Report No. NWBR00-122
Table 106. Back-Calculated
Concentrations of Calibration Diluted with Buffer
Standards
for PFOS in Serum
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date Run Number 5.30 6.80 14.3 29.3 54.3 104 254 404 504
23-Oet-2000
PFOSMD-39
5.32 6.18 14.0 25.6 52.3 5.45 7.23 14.8 30.1 54.6
102 264 410 476 113 265 417 488
Table 107. Back-Calculated
Concentrations of Calibration Standards for PFOSA in Serum Diluted with Buffer
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date Run Number 1.00 2.50 10.0 25.0 50.0 100 250 400 500
23-Oct-2000 PFOSMD-39
1.03 2.21 9.29 22.6 50.2 98.2 265 410 455 1.03 2.45 I0.1 25.8 50.6 112 271 423 468
Table 108. Back-Calculated
Concentrations of Calibration Standards for PFOSAA in Serum Diluted with Buffer
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date !RunNumberi 1.00 2.50 10.0 25.0 50.0 100 250 400 500
23-Oct-2000 PFOSMD-39 1.12 2.50 9.73 23.6 50.0 104 267 .408 485 0.901 2.41 9.41 24.1 50.6 111 254 399 .479 ::....
Page 108
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 109. Back-Calculated
Concentrations of Calibration Diluted with Buffer
Standards for POAA in Serum
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date Run Number 1.70 3.20 10.7 25.7 50.7 101 254 401 501
23-Oct-2000
PFOSMD-39
1.56 3.22 10.4 24.7 50.9 99.8 260 401 458 1.88 3.10 10.6 25.3 50.8 109 263 432 491
Table 110. Back-Calculated Concentrations of Calibration Standards for PFHS in Serum Diluted with Buffer
Quadratic weighted I/x 2. All concentrations are expressed as ppb.
Run Date Run Number 1.00 2.50 10.0 25.0 50.0 100 250 400 500
23-Oct-2000
PFOSMD-39
1.01 2.40 I0.I 24.2 49.6 99.3 254 424 450 1.01 2.53 I0.0 24.3 49.0 I06 260 433 473
Table 111. Back-Calculated
Concentrations of Calibration Standards for M556 in Serum Diluted with Buffer
QuadratiwceightedI/x2.Allconcentratioanrseexpresseadsppb.
Run Date Run Number 2.50 10.0 25.0 50.0 100 250 400 500
23-Oct-2000 PFOSMD-39
2.68 I0.6 22.9 51.3 109 2.36 9.33 22.9 48.6 i02
283 432 505 242 379 45i
"" ,,;.
Page 109
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 112. Back-Calculated
Concentrations of Calibration Diluted with Buffer
Standards for M570 in Serum
Quadratic weighted l/x z. All concentrations are expressed as ppb.
Run Date Run Number 1.00 2.50 10.0 25.0 50.0 100 250 400 500
23-Oct-2000
PFOSMD-39
1.07 2.48 9.93 22.5 51.6 0.964 2.34 9.59 24.6 50.8
103 268 408 489 110 254 393 476
o; ;
Page 1I0
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 113. PFOS Quality Controls Samples Prepared in Serum Diluted with Buffer
Run Date 23-Oct-2000
Run Number PFOSMD-3 9
193 ng/mL DF=10
204 219 205 202
443 ng/mL DF=10
464 423 451 477
4040 ng/mL DF=10
4240 4420 4320 4320
4040 ng/mL DF=50
*3180 *3020 *3150 "3100
Mean S.D. %CV
%Theoretical %Bias n
208 7.77 3.7 107.8 7.8
4
454 23.1 5.1 102.5 2.5
4
* >+_20%deviation from theoretical
4330 73.7 1.7 107.2 7.2
4
3110 69.9 2.2 77.0 -23.0
4
Table 114. PFOSA Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date
Run Number
23-Oct-2000 PFOSMD-39
150 n_mL DF=10
147 147 145 136
400 ng/mL DF=10
388 360 370 409
4000 n_mL DF=10
4020 3850 3980 3960
4000 ng/mL DF=50
*2380 *2240 "2210 "2210
Mean S.D.
%CV %Theoretical
%Bias n
144
382
5.25
21.5
3.6
5.6
96.0
95.5
-4.0
-4.5
4
4
* >+20*/odeviation from theoretical
3950 72.7 1.8 98.8 -1.3
4
o
2260 =81.2
3.6 56.5 -43.5
4
Page 111
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 115. PFOSAA Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date 23-Oct-2000
Run Number PFOSMD-39
153 ng/mL DF=10
144 146 142 142
403 ng/mL DF=10
405 379 389 417
4000 ng/mL DF=10
3920 4100 3970 4010
4000 ng/mL DF=50
*2750 *2560 *2620 *2550
Mean S.D. %CV
%Theoretical %Bias n
144
398
1.91
16.8
1.3
4.2
94.1
98.8
-5.9
-1.2
4
4
* >+_20%deviation from theoretical
4000 76.2 1.9 100.0 0.0
4
2620 92.0 3.5 65.5 -34.5
4
Table 116. POAA Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date
Run Number
23-Oct-2000 PFOSMD-39
157 n_mL DF=10
163 174 164 166
407 ng/mL DF=10
410 387 418 411
4010 ng/mL DF=10
4230 4190 4280 4140
4010 n_mL DF=50
3610 3310 3330 ': !3350
Mean S.D.
%CV %Theoretical
%Bias n:
167 4.99 3.0 106.4 6.4
4
407 13.5 3.3 100.0 0.0 4
4210 59.4 1.4 105.0 5.0
4
:3400 141 4.1 84.8 -15.2 4
Page 112
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 117. PFHS Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date
Run Number
23-Oct-2000 PFOSMD-39
154 ng/mL DF=10
157 163 161 166
404 ng/mL DF=10
421 389 416 419
4000 n_mL DF=10
4200 4210 4330 4300
4000 ng)mL DF=50
3490 3300 3410 3260
Mean S.D. %CV
%Theoretical %Bias n
162 3.77 2.3 105.2 5.2
4
411 15.0 3.6 101.7 1.7
4
4260 64.8 1.5 106.5 6.5
4
3370 105 3.1 84.3 -15.8 4
Table 118. M556 Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date 23-Oct-2000
Run Number PFOSMD-39
151 ng/mL DF=10
156 157 145 148
401 ng/mL DF=10
421 385 403 417
4000 ng/mL DF=10
4480 4600 4710 4490
Mean S.D. %CV
%Theoretical %Bias n
152 5.92 3.9 100.7 0.7
4
407 16.3 4.0 101.5 1.5
4
* >+_.20%deviation from theoretical
4570 108 2.4 114.3 14.3 4
4000 nffmL DF=50 *2980 *2740 *2780 *27 I0
o; :
. 2800_ _ 122
4.4 70.0 -30.0
4
" =" -
Page 113
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 119. M570 Quality Control Samples Prepared in Serum Diluted with Buffer
Run Date 23-Oct-2000
Run Number . 157 ng/mL DF=10
PFOSMD-39
159
167
158
160
407 ng/mL DF=10
430 401 408 441
4010 ng/mL DF=10
4240 4390 4290 4330
4010 ng/mL DF=50
"2910 *2730 *2700 *2680
Mean S.D. %CV
%Theoretical %Bias n
161 4.08 2.5 102.5 2.5
4
420 18.7 4.5 103.2 3.2 4
* >+_20%deviation from theoretical
4310 63.4 1.5 107.5 7.5
4
2760 105 3.8 68.8 -31.2 4
,:'.
Page 114
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 120. Summary of Calibration Curve Parameters for PFOS (Serum vs. Plasma) Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date
27-Oct-2000 27-Oet-2000 25-Apr-2001 25-Apr-2001
Run Number
PFOSMD-40
PFOSMD-46
Matrix
Plasma Serum
A
0.000000 -0.000003
B
0.028132 0.031081
C
R-Squared LLOQ ULOQ
0.010586 0.084427
0.9931 0.9506
3.94 414 36.8 454
25
Plasma -0.000003 0.048192 0.045435
0.9841
26
Serum -0.000020 0.057680 -0.301033 0.9934
A, B, and C are coefficients used to define the calibration curve.
3.28 413 36.8 454
Table 121. Summary of Calibration Curve Parameters for PFOSA (Serum vs. Plasma) Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date
25-Apt-2001 25-Apr-2001
Run
Matrix
A
B
C
R-Squared
Number
25
Plasma -0.000005 0.120662 0.028023 0.9887
26
Serum -0.000032 0.166445 0.051786 0.9873
A, B, and C are coefficients used to define the calibration curve.
LLOQ
1.00 1.00
ULOQ
500 502
Table 122. Summary of Calibration Curve Parameters for PFOSAA (Serum vs. Plasma) Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date
25-Apt-2001 25-Apt-2001
Run Number
,,1
25
Matrix
A
B
C
R-Squared LLOQ ULOQ
Plasma 0.000009 0.033608 0.035470
0.9938 .:'1:.30 500
26
Serum 0.000007 0.044621 -0.040847 0.9947
4.20. 503 -
A, B, and C are coefficients used to define the calibration curve. a
Page 115
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00o122
Table 123. Summary of Calibration Curve Parameters for POAA (Serum vs. Plasma) Quadratic weighted l/x _. All concentrations are expressed as ppb.
RunDate
27-Oct-2000 27-Oct-2000 25-Apr-2001 25-Apr-2001
Run Number
PFOSMD-40
PFOSMD-46
Matrix
Plasma Serum
A
0.000002 0.000003
B
C
R-Squared LLOQ ULOQ
0.014346 0.016839
0.002192 0.003246
0.9918 0.9894
1.92 481 6.82 461
25
Plasma -0.000003 0.011215 0.005552 0.9929
26
Serum 0.000001 0.011414 0.000898 0.9931
A, B, and C are coefficients used to define the calibration curve.
1.44 481 6.82 461
Table 124. Summary of Calibration Curve Parameters for PFI:IS (Serum vs. Plasma) Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date Run Number MatrLx
A
B
C
R-Squared LLOQ ULOQ
27-Oct-2000 27-Oct-2000
PFOSMD-40 Plasma -0.000001 0.025534 0.011247 0.9953 PFOSMD-46 Serum -0.000012 0.033256 0.018733 0.9890
A, B, and C are coefficients used to define the calibration curve.
1.36 523 3.85 431
Table 125. Summary of Calibration Curve Parameters for M556 (Serum vs. Plasma) Quadraticweighted I/x2. All concentrationsare expressedas ppb.
RunDate
27-Oct-2000 27-Oct-2000
Run
Matrix
A
B
C JR-Squared LLOQ
Number
PFOSMD-40 Plasma 0.000002 0.016888 0.007835 0.9949 ": 2.50
PFOSMD-46 Serum 0.000005 0.014650 0.008960
0.9903 '3.8_3:
A, B, and C are coefficients used to define the calibration curve.
"
ULOQ
500 501"
Page 116
NorthwestBioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 126. Summary of Calibration Curve Parameters for M570 (Serum vs. Plasma )
Quadratic weighted I/x2. All concentrations are expressed as ppb.
RunDate
27-Oct-2000 27-Oct-2000 25-Apr-200I 25-Apr-2001
Run Number
PFOSMD-40
PFOSMD-46
Matrix
Plasma Serum
A
0.000001 0.000003
B
0.034031 0.034699
C
R-Squared
0.011681 0.028660
0.9910 0.9919
25
Plasma 0.000009 0.039312 0.045599 0.9964
26
Serum 0.000002 0.046817 0.021413 0.9931
A, B, and C are coefficients used to define the calibration curve.
LLOQ
1.00 5.20 1.00 3.70
ULOQ
500 503 500 503
o; ;
Page I 17
Northwest Bioanalytical
StudyNo. NWBS00-040 Report No. NWBR00-122
Table 127. Baek-Caleulated
Concentrations of Calibration Standards for PFOS (Serum vs. Plasma)
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date 27-Oct-2000
Run Date 27-Oct-2000
Run Date 25-Apr-2001
Run Date 25-Apt-2001
Run Number PFOSMD-40
Run Number PFOSMD-46
Run Number 25
Run Number 26
Matrix Plasma
Matrix Serum
Matrix Plasma
Matrix Serum
3.94 5.17 11.3 23.6 44.2 85.4 209 332 414 3.72 5.18 11.3 19.4 45.5 82.1 214 337 369 3.90 5.67 11.7 24.1 45.9 84.4 221 367 407 36.8 38.0 44.3 56.8 77.7 120 245 370 454 34.0 31.7 37.3 49.3 70.7 114 247 374 392 50.5 39.0 45.8 60.7 78.7 125 282 418 436 3.28 4.52 10.7 23.0 43.5 84.6 208 331 413 2.70 _3.90 9.65 21.4 42.2 80.3 190 324 392 4.15 4.63 11.3 25.0 47.6 90.3 217 336 441 36.8 38.0 44.3 56.8 77.7 120 245 370 454 38.3 35.5 44.2 54.7 71.1 124 234 363 420 37.4 39.0 46.1 56.1 75.9 128 256 400 459
=
Page 118
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 128. Back-Calculated Concentrations of Calibration Standards for PFOSA (Serum vs. Plasma)
QuadraticweightedI/x2. Allconcentrationsareexpressedas ppb.
Run Date Run Number Matrix 1.00 2.50 10.0 25.0 50.0 100 250 400 500
25-Apr-2001
25
Plasma 0.935 2.52 7.89 23.0 45.4 102 223 387 477
*1.77 2.99 10.3 27.3 52.5 113 254 402 546
25-Apr-2001
26
Serum 0.951 1.96 9.62 25.6 48.1 109 240 403 422
1.18 2.21 10.7 24.3 52.4 111 279 450 491
* Sampledeactivatedas anoutlier
o; *.
Page 119
Northwest Bioanalytieal
Study No. NWBS00-040 Report No. NWBR00-122
Table 129. Back-Calculated
Concentrations of Calibration (Serum vs. Plasma)
Standards
for PFOSAA
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date 25-Apr-2001
Run Number 25
Run Date 25-Apro2001
Run Number 26
Matrix Plasma
Matrix Serum
1.30 2.80 10.3 25.3 50.3 100 250 400 500 1.28 2.65 8.27 24.7 50.7 102 253 399 480 *2.23 3.22 9.79 24.9 54.4 110 257 391 509 4.20 5.70 13.2 28.2 53.2 103 253 403 503 4.27 5.31 12.8 26.9 50.7 119 259 391 477 4.38 5.78 13.6 25.0 51.9 111 260 407 511
* Sample deactivated as an outlier
_..
Page 120
NorthwestBioanalytical
StudyNo. NWBS00-040 ReportNo. NWBR00-122
Table 130. Back-Calculated
Concentrations of Calibration (Serum vs. Plasma)
Standards for POAA
Quadraticweighted1/x2. Allconcentrationsareexpressedasppb.
Run Date 27-Oct-2000
Run Number PFOSMD-40
Matrix Plasma
Run Date 27-Oct-2000
Run Number PFOSMD-46
Matrix Serum
Run Date 25-Apr-2001
Run Number 25
Matrix Plasma
Run Date 25-Apr-2001
Run Number 26
Matrix Serum
1.92 1.73 2.30 6.82 7.32 "18.3 1.44 1.35 "2.14 6.82 7.98 6.23
3.36 10.6 25.0 49.0 97.1 241 385 481 3.01 10.5 22.9 49.7 99.1 250 367 439 3.12 10.9 25.1 50.7 96.7 262 422 474 8.18 15.0 28.6 51.4 97.3 234 370 461 6.86 13.9 27.2 51.9 105 252 399 421 9.23 14.9 29.4 46.2 94.9 246 392 424 2.88 10.1 24.5 48.5 96.6 241 385 481 2.65 10.5 24.1 43.4 91.1 225 396 470 3.35 11.2 26.6 46.5 90.9 241 405 498 8.18 15.0 28.6 51.4 97.3 234 370 461 8.28 15.6 28.8 50.2 103 242 383 457 7.23 15.0 26.9 48.0 100 235 378 433
* Sampledeactivatedas an outlier
Page 121
Northwest Bioanalytical
S tudy No. NWBS00-040 Report No. NWBR00-122
Table 131. Back-Calculated
Concentrations of Calibration (Serum vs. Plasma)
Standards for PFHS
Quadratic weighted 1/x2. All concentrations are expressed as ppb.
Run Date 27-Oct-2000
Run Date 27-Oct-2000
Run Number PFOSMD-40
Run Number PFOSMD-46
Matrix Plasma
Matrix Serum
1.36 2.92 10.7 26.5 52.6 104 262 418 523 1.52 "_ 2.72 10.6 24.2 54.6 107 259 466 489 1.28 2.78 10.4 28.4 54.1 102 270 425 491 3.85 5.13 11.5 24.4 45.7 88.9 217 345 431 4.08 4.36 11.2 21.8 47.5 89.1 235 371 363 "15.4 5.80 10.6 23.5 49.0 89.7 230 388 389
* Sample deactivated as an outlier
o_
Page 122
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 132. Back-Calculated Cohcentrations of Calibration Standards (Serum vs. Plasma)
Quadratic weighted I/x2. All concentrations are expressed as ppb.
for M556
Run Date 27-Oct-2000
Run Date 27-Oct-2000
Run Number PFOSMD-40
Run Number PFOSMD-46
Matrix 2.50 10.0 25.0 50.0 100 250 400 500
Plasma 2.64 I0.7 23.4 52.9 I01 276 396 456
Matrix
2.35 9.95 23.3 48.3 94.1 260 440 481
=.,
3.80 11.3 26.3 51.3 101 251 401 501
Serum 3.89 II.I 25.3 53.7 116 280 441 480
3.91 10.0 25.7 42.9 102 258 396 450 _
; ; :.
_ ,4""
Page 123
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 133. Back-Calculated
Concentrations of Calibration (Serum vs. Plasma)
Standards for M570
Quadratic weighted l/x 2. All concentrations are expressed as ppb.
Run Date 27-Oct-2000
Run Date 27-Oct-2000
Run Date 25-Apr-2001
Run Date 25-Apr-2001
Run Number PFOSMD-40
Run Number PFOSMD-46
Run Number 25
I
Run Number 26
Matrix Plasma
Matrix Serum
Matrix Plasma
Matrix Serum
1.00 2.50 10.0 25.0 50.0 100 250 400 500 0.993 2.71 10.6 21.2 53.6 104 284 401 444 0.951 2.65 10.4 20.9 48.6 93.5 254 443 481 3.70 5.20 12.7 27.7 52.7 103 253 403 503 *5.54 4.92 11.8 27.4 51.0 116 277 432 464 "16.4 5.75 12.4 26.9 45.5 106 271 404 468 1.00 2.50 10.0 25.0 50.0 100 250 400 500 1.01 2.43 8.50 24.6 52.8 105 246 420 483 "1.90 2.66 9.29 25.4 51.9 108 247 387 504 3.70 5.20 12.7 27.7 52.7 103 253 403 503 4.07 4.60 12.5 27.4 50.3 118 262 395 478 3.94 4.60 13.3 26.0 49.7 109 264 405 502
* Sample deactivated as an outlier
Page 124
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 134. PFOS Quality Controls (Serum vs. Plasma)
All concentrations are expressed as ppb.
Run Date Run Number Low QC Medium QC High QC
_
(Plasma)
(Plasma)
(Plasma)
6.40 ppb
126 ppb
332 ppb
27-Oct-2000 PFOSMD-40
6.44
115
316
5.99
136
326
5.91
118
298
Mean SD
%CV %Bias
n Run Date Run Number
27-Oct-2000 PFOSMD-46
6.11 0.286 4.7 -4.5
3
Low QC (Serum) 39.3 ppb
35.0 35.6 35.5 35.8
123 11.4 9.3 -2.4
3
Medium QC (Serum) 161 ppb 164 151 149 152
313 14.2 4.5 -5.7
3
High QC (Serum) 370 ppb
336 "199 343 346
Mean SD
%CV %Bias
n
35.5
154
342
0.340
6.78
5.13
1.0
4.4
1.5
-9.7
.4.3
-7.6
4
4
3
* Sample deactivated due to a sample preparation error. ** > +_20%deviation from theoretical
'; ":
: =" "
Page 125
Northwest Bioanalytical
Study No. NWBS00-040
Report No. NWBR00-122
r
Table 134. PFOS Quality Controls (Serum vs. Plasma) Continued
All concentrations are expressed as ppb.
Run Date 25 -Apr-2001
Run Number 25 '
Low QC (Plasma) 5.75 ppb
5.44 5.62
5.59
5.90
Medium QC (Plasma) - 126 ppb
143 139
136
132
High QC (Plasma) 331 ppb
333 325
347
333
Mean SD
%CV %Bias
5.64
138
335
0.192
4.65
9.15
3.4
3.4
2.7
-I .9
9.5
1.2
;'
n
4
4
4
Run Date
Run Number
Low QC (Serum) 39.3 ppb
Medium QC (Serum) 161 ppb
High QC (Serum) 370 ppb
25-Apt-2001
26
41.3
174
377
45.4
182
388
45.8
172
408
45.6
169
381
Mean SD
%CV %B ias
n
44.5
174
2.16
5.56
4.9
3.2
13.2
8.1
4
4
* > +_20% deviation fi'om theoretical
389
13.8
': ':
3.5
:
5.1
:
4
Page 126
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 135. PFOSA Quality Controls (Serum vs. Plasma)
All concentrations are expressed as ppb.
Run Date 25-Apr-2001
Run Number 25
Low QC (Plasma) 4.00 ppb
4.06 3.99
4.49
4.73
Medium QC (Plasma) 150 ppb
158 175
170
159
High QC (Plasma) 400 ppb
395 412
438
424
Mean SD
%CV %Bias
n Run Date Run Number
25-Apr-2001
26
4.32 0.353
8.2 8.0 4
Low QC (Serum) 4.00 ppb
3.52 4.75 4.55 **4.85
166 8.35 5.0 10.7
4
Medium QC (Serum) 150 ppb
153 163 155 151
417 18.2 4.4 4.3
4
High QC (Serum) 400 ppb
390 431 417 377
Mean SD
%CV %Bias
n
4.42
156
404
0.611
5.26
24.6
13.8
3.4
6.1
10.5
4.0
1.0
4
4
4
* Sample deactivated due to a sample preparation error. ** > +9_0%deviation from theoretical
":': "
Page 127
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 136. PFOSAA Quality Controls (Serum vs. Plasma) All concentrations are expressed as ppb.
Run Date Run Number
25-Apr-2001
25
Low QC (Plasma) 4.30 ppb
3.59 3.85 3.74
4.96
Medium QC (Plasma) 150 ppb
** 187 175 172
169
High QC (Plasma) 400 ppb
429 412 440
429
Mean SD
%CV %Bias
n Run Date Run Number
25-Apr-2001
26
4.04 0.626 15.5 -6.0
4
Low QC (Serum) 7.20 ppb
6.74 8.11 8.29 8.37
176 7.89 4.5 17.3
4
Medium QC (Serum) 153 ppb
150 162 159 157
428 11.6 2.7 7.0
4
High QC (Serum) 403 ppb
363 409 396 376
Mean SD
%CV %Bias
n
7.88
157
386
0.766
5.10
20.5
9.7
3.2
5.3
9.4
2.6
-4.2
4
4
4
Sample deactivated due to a sample preparation error. * > +_20%deviation from theoretical
":' "
g
: "" -
Page 128
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 137. POAA Quality Controls (Serum vs. Plasma)
All concentrations are expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOSMD-40
Low QC (Plasma) 4.81 ppb
4.57
4.31
4.52
Medium QC (Plasma) .145 ppb 146
167
147
High QC (Plasma) 385 ppb
373
419
393
Mean SD
/oCV %Bias
n Run Date Run Number
27-Oct-2000 PFOSMD-46
4.47 0.138
3.1 -7. I
3
Low QC (Serum) 9.54 ppb
9.42 9.17 8.53 9.62
153
395
11.8
23.1
7.7
5.8
5.5
2.6
3
3
Medium QC (Serum) 143 ppb
158
High QC (Serum) 370 ppb
373
146
I *253
148
375
150
388
Mean SD
%CV %Bias
n
9.19
151
379
0.474
5.26
8.14
5.2
3.5
2.1
-3.7
5.6
2.4
4
4
3
Sample deactivated due to a sample preparation error. * > +_20%deviation from theoretical
': ""
:--= ....
Page 129
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR.00-122
Table 137. POAA Quality Controls (Serum vs. Plasma) Continued All concentrations are expressed as ppb.
Run Date 25-Apt-2001
Run Number 25
Low QC (Plasma) 4.32 ppb
5.13 4.47 4.35
4.44
Medium QC (Plasma) . 145 ppb
143 145 145
137
High QC ('Plasma) 385 ppb
394 383 412
404
Mean SD
%CV %Bias
n Run Date Run Number
25-Apt-2001
26
4.60 0.359
7.8 6.5 4
Low QC (Serum) 9.54 ppb
10. I 11.4 *'13.2 9.82
143 3.79 2.7 -1.4
4
Medium QC (Serum) 143 ppb
157 152 149 152
398 12.6 3.2 3.4
4
High QC (Serum) 370 ppb
374 389 385 391
Mean SD
%CV */,Bias i
n
11.1
153
1.54
3.32
13.9
2.2
16.4
7.0
4
4
** > +_20% deviation from theoretical
385
7.59
2.0
"" "
4.1
:
" '- -
4
g
Page 130
Northwest Bioanalytieal
Study No. NWBS00-040 Report No. NWBR00-122
Table 138. PFHS Quality Controls (Serum vs. Plasma) All concentrations axe expressed as ppb.
Run Date Run Number 27-Oct-2000 PFOSMD-40
Low QC (Plasma) 4.48 ppb
4.62
4.35
4.39
Medium QC -(Plasma) 156 ppb
157
173
151
High QC (Plasma) 417 ppb
397
407
384
Mean SD
%CV %Bias
n Run Date Run Number
27-Oct-2000 PFOSMD-46
4.45 0.146
3.3 -0.7
3
Low QC (Serum) 6.41 ppb
5.65 5.70 6.87 5.71
160 11.4 7.1 2.6
3
Medium QC (Serum) 132 ppb 144 136 129 135
396 11.5 2.9 -5.0
3
High QC (Serum) 345 ppb
343 "221 395 367
Mean SD
%CV %Bias
n
5.98
136
368
0.592
6.16
26.0
9.9
4.5
7.1
-6.7
3.0
6.7
4
4
3
Sample deactivated due to a sample preparation error. * > .+.20%deviation from theoretical
._' : :
_- _
Page 131
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 139. M556 Quality Controls (Serum vs. Plasma)
All concentrations are expressed as ppb.
Run Date 27-Oct-2000
Run Number PFOSMD-40
Low QC (Plasma) 4.00 ppb
3.73
3.45
3.34
Medium QC (Plasma) 150 ppb
133
168
139
High QC (Plasma) 400 ppb
389
420
387
Mean
SD
%CV
%Bias
'
n
3.51
147
399
0.201
18.7
18.5
5.7
12.7
4.6
-12.3
-2.0
-0.3
3
3
3
Run Date Run Number 27-Oct-2000 PFOSMD-46
Low QC (Serum) 5.30 ppb
5.03 5.23 *'4.15
5.01
Medium QC (Serum) 151 ppb 166 152 152
155
High QC (Serum) 401 ppb
383 **243 371
388
Mean SD
%CV */,Bias
n
4.86
156
0.480
6.65
9.9
4.3
-8.3
3.3
4
4
** > +_20%deviation from theoretical
381
8.74
2.3
"""
-5.0
:
3
: '--" "
Page 132
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 140. M570 Quality Controls (Serum vs. Plasma)
All concentrations are expressed as ppb.
Run Date 27-0ct-2000
Run Number PFOSMD-40
Low QC (Plasma) 4.00 ppb
3.95
3.54
3.38
Medium QC (Plasma) 150 ppb 133
161
136
High QC (Plasma) 400 ppb
379
403
360
Mean SD
%CV %Bias
n Run Date Run Number
27-Oct-2000 PFOSMD-46
3.62 0.294
8. I -9.5
3
Low QC (Serum) 6.70 ppb
6.36 6.35 **5.33 6.39
143 15.4 I 0.8 -4.7 3
Medium QC (Serum) 153 ppb 169 150 156 155
381 21.5 5.6 -4.8
3
High QC (Serum) 403 ppb
397 *227 367 400
Mean SD
%CV %Bias
n
6.11
158
388
0.519
8.10
18.2
8.5
5.1
4.7
-8.8
3.3
-3.7
4
4
3
* Sample deactivated due to a sample preparation error. ** > +90% deviation from theoretical
': "
: .:"
Page 133
Northwest Bioanalytical
Study No. NWBS00-040 Report No. NWBR00-122
Table 140. M570 Quality Controls (Serum vs. Plasma) Continued
All concentrations are expressed as ppb.
Run Date 25-Apt-2001
Run Number 25
Low QC (Plasma) 4.00 ppb
3.24 3.49
3.58 4.05
Medium QC (Plasma) . 150 ppb
180 175
164 171
High QC (Plasma) 400 ppb
420 417
440 432
Mean SD
%CV %Bias
n
Run Date Run Number
25-Apr-2001
26
3.59 0.339
9.4 -10.3
4
Low QC (Serum) 6.70 ppb
5.90 7.49 7.03 8.03
173 6.76 3.9 15.3
4
Medium QC (Serum) 153 ppb
166 167 167 161
427 10.7 2.5 6.8
4
High QC (Serum) 403 ppb
392 445 437 409
Mean SD
%CV %Bias
n
7. I 1
165
0.906
2.87
12.7
1.7
6.1
7.8
4
4
* > +__20%deviation from theoretical
421
24.6
5.8
":" "
4.5
:
4
: '= "
Page 134
Northwest Bioanalytical "_,
Study No. NWBS00-040 Report No. NWBR00-122
i
Table 141. Serum QC Results Based on Plasma Curves
Sample ID
PFOS (ppb)
Low (analyzed 10/27100)
41.2
Low (analyzed 10/27/00)
41.8
Low (analyzed 10/27/00)
41.8
Low (analyzed 10/27/00)
42.1
Low (analyzed 4/25/01)
41.7
Low (analyzed 4/25/01)
46.5
Low (analyzed 4/25/01)
46.7
Low (analyzed 4/25/01)
47.0
mean * std. dev. *
%CV * %dev *
43.6 2.61 6.0% 10.9%
acceptance range * 31.4to47.2 95*,/, confidence range * 41.8 to 45.4
PFOSA (ppb)*
5.04 6.47 6.75 6.88 6.29 0.847 13.5% 57.1% 3.00to5.00 5.45 to 7.12
PFOSAA (ppb)*
6.68 8.5 8.72 8.83 8.18 1.01 12.4% 13.6% 5.40to9.00 7.48 to 8.88
POAA (ppb)
PFHS ,(ppb)*
11.1
7.64
10.8
7.71
10.I
9.22
I 1,4
7.72
9.62
9.93
11.3
13.0
10.9 1.07 9.9% 14.3%
8.07 0.766 9.5% 25.9%
7.63to 11.4 5.13to7.69 10.2 to 11.7 7.32 to 8.82
M556 (ppb)*
4.44 4.61 3.67 4.42
4.29 0.419 9.8% -19.2% 4.24to6.36 3.87 to 4.7
M570 (ppb)
6.99 6.98 5.93 7.02 6.41 7.74 8.29 8.94
7.29 0.986 13.5% 8.8%
5.36to8.04 6.60 to 7.97
Meal Med Med Med Med Med Med Med
(analyzed (analyzed (analyzed (analyzed (analyzed (analyzed (analyzed (analyzed
10/27/00) 10/27/00) 10/27/00) 10/27/00) 4/25/01) 4/25/01) 4/25/01) 4/25/01)
mean *
std. dev. * %CV * %dev *
acceptance range* 95% confidence range*
182 168 165 168 185 189 191 199
181 12.5 6.9% 12.3%
129 to 193 172 to 190
205 207 210 220
211 6.66 3.2% 40.3%
113to188 204 to 217
192 200 203 207
201 6.4 3.2% 31.0%
115to191 196 to 205
186 172 174 177 160 164 164 170
171 8.37 4.9% 19.5%
114to172 165 to 177
180 170 162 169
170 7.41 4.4% 29.0% 106 to 158 163 to 178
149 136 137 139
140 5.97 4.3% -7.1% 121 to 181 134 to 146
175 155 161 160 184 190 191 191
176 15.3 8.7% 15.0%
122 to 184 165 to 186
High (analyzed 10/27/00) High (analyzed 10/27/00) High (analyzed 10/27/00) High (analyzed 4/25/01) High (analyzed 4/25/01) High (analyzed 4/25/01) High (analyzed 4/25/01)
mean * std. dev. *
%CV * %dev *
364 370 374 396 399 406 424 390 21.8 5.6% 5.5%
493 509 543 560 526 30.7 5.8% 31.6%
451 467 490 506 479 24.3 5.1% 18.7%
441 443 459 453 470 476 478 460 15.2 3.3% 24.3%
400 452 424
425 26.0 6.1% 23.3%
361 349 366
-.
; 359 8.74 2.4% -10.6%
416 384 420 433 450 ,,-. 478 486 438 36.0 8.2% 8.7%
acceptance range* 95%eonfideneerange*
296 to 444 374 to 407
300 to 500 492 to 561
302 to 504 460 to 497
296 to 444 449 to 471
276 to 414 396 to 455
321to481 349 to 369
322 to 484 411to465
Due to rounding, calculations based on displayed values may differ from actual. NOTE: Only data from acceptable runs included in the Table (see report text).
(Data obtained on 10/27/00 from run PFOSMD-40. Data obtained on 4/25/01 from run S00040-25.)
Page 135