Document QXm9Yy29qpJXLEQo08zReE70k

November 14, 1949 MEMORANDUM To: Doctor Vorwald From: Mr. Urban Subject: GENERAL ELECTRIC SILICA FUME `CM Our studies of the comparison of performance of the electro static precipitator and filter paper sampler (AEC model pump) have demonstrated a variance in air flows from those indicated by the manufacturers. Based upon three (3) short test runs only, the electro static precipitator was 2.5% off. The filter paper sampler was about 125$ off in a low air flow range. Only one sample point with this instrument was tested. Both these instruments are scheduled for final calibration at Doctor Wright's laboratory this week, Tuesday, November 15, 1949, if possible. All the filter paper samples collected for the G.E. fume test to date were run at the same sample rate of .005 cubic meters per minute. The values recorded to date can be properly adjusted by the correction factor which will be developed before they are pre sented to the General Electric Company. ECJUrbw