Honeywell
December 9, 1974
Conoco
VCM Plant Box 605
Westlake,
La.
70669
Attn: Paul Fetzer
Dear Mr. Fetzer :
We have recently supplied several Model 1000 Chromatographs to YOUR PLANT WHICH ARE APPLIED FOR VlNYL CHLORIDE MEAUSREMENT IN AM BIENT AIR. WE HAVE SPECIFIED THE ACCURACY OF THE MACHINE TO MEET OR EXCEED THE OSHA REQUIREMENT IN A LETTER TO Mr. LEON VERNON, GEN ERAL Manager, Conoco Plastics, Saddlebrook, N.J. dated Octover 23, 1974.
The accuracy of these instruments
is unaffected by other com
ponents in the ambient air unless these components directly interfere
WITH THE SEPARATION CHARACTERISTICS OF THE COLUMN OR ALTER DETECTOR
SENS I TI V I TY.
In the event that other COMPONENTS IN THE AIR ARE SUSPECT OF INTERFERENCE, SUCH AS METHYL CHLORIDE, A SUITABLE SELECTION OF COL UMN CONFIGURATION AND DETECTOR SENSITIVITY MUST BE MADE TO PRESERVE THE ACCURACY OF THE INSTRUMENT.
The accuracy within tolerences, of the Model 1000 Chromatograph,
AS IT IS CURRENTLY APPLIED FOR VCM MEASUREMENT IN AMBIENT AIR WILL
NOT BE AFFECTED BY THE FOLLOWING COMPONENTS LISTED. OF COURSE, SUB JECTING THE INSTRUMENT TO EXCESS CONCENTRATIONS OF ANY COMPONENT WILL IMPAIR THE OPERATION OF THE ANALYZER.
1. Methane, Ethylene, Propylene, Benzene
2. Ethyl Chloride, Chlorobenzene
3. 1,2 Dichloroethane, 1,1 Dichloroethylene 1,1 Dichloroethylene, 1,1 Dichloropropane Dl CHLOROBUTENE, Ci.S & TRANS 1,2 D 1 C H LO R 0 E T H Y L E N E
4. Chloroform, 1,1,1 Trichloroethene 1,1,2 Trichloroethane, Trichloroehtylene
5. Carbon Tetrachloride, 1,1,2,2 Tetrachloroethylene, Tetrachloroethylene
VVC 000002091
HONEYWELL INC., 6625 McGREW, HOUSTON, TEXAS 77017, TELEPHONE 713/645-4911
6, Water, Hydrogen Chloride
Very truly yours
fam
E. General Manager Process Analyzer Center
EM:po
cc:
Leon Vernon, General Manager, Conoco Plastics
Robert Walston, Conoco
Job file
VVC 000002092