Sumika Chamical Analysis Service Chiba Laboratory
Kitasode 9-1, Sodegaura-shi Chiba 299-0266 Japan
Analysis and Test Report PFOA measurement from PTFE and PFA
Please feel free to ask our sales representative or personnel below if you have any inquiry on the test results.
Supervisor: Uchihara Person in charge: Yamauchi
Phone: +81-(0)438-62-4193 Fax: +81-(0)438-62-4162
April 4, 2019 Order number: 8678598
PFOA Measurement in PTFE and PFA
1. Objective Measurement of PFOA (perfluoro-n-octanoic acid) quantity in test samples respectively.
2. Type and number of the test samples 1) Cross-linked PTFE (atmospheric baking) 2) Cross-linked PTFE (nitrogen baking) 3) PTFE 4) Cross-linked PFA (nitrogen baking) Total 4 different test samples
3. Analysis and Test target Perfluoro-n-octanoic acid (PFOA), CAS NO. 335-67-1 Total 1 test target
4. Analysis and Test method 4.1 Sample preparation Step 1) Added solvent to each test sample Step 2) Performed ultrasonic extraction. Step 3) Sampled a certain amount once the volume of the extract was measured. Step 4) Added standard solution for calibration. Step 5) Used it as liquidus sample for the testing. 4.2 Measuring method Measured the liquidus samples acquired from 4.1 using liquid chromatograph-tandem mass spectrometer (LC-MS/MS).
5 Analysis and Test Results Table 1 shows the analysis and test results.
Table 1. Analysis and Test Results
Test sample 1) Cross-linked PTFE (atmospheric baking) 2) Cross-linked PTFE (nitrogen baking) 3) PTFE 4) Cross-linked PFA (nitrogen baking) Lower limit of quantitation
unit: g/kg (=ppb) PFOA
<10 <10 <10 <10 10