Rio Tinto Minerals 8051 E. Maplewood Ave Greenwood Village, CO 80111 (303) 713-5000 Fax: (303) 713-5788
TECHNICAL REPORT
Project No. A11054
Date:
29-Jul-11
To: Desiree Giroux
From:
Carol Slowik
Analyst(s): C. Slowik, J. Sulik, A. Zamurs
Copy:
R. Reilly, H. Bruning, J. Pier, S. Sharma, M. Khan
Title:
TEM ASBESTOS ANALYSIS OF ARGONAUT 2010 INFILL SAMPLES
Request:
Thirteen samples, representing drilling samples from Argonaut, were submitted to the Technical Center for asbestos analysis by transmission electron microscopy (TEM). This analysis is in partial fulfillment of RTM's Fiber Management Framework Program.
Results:
The attached TEM results were obtained using the current ASTM draft method "Standard test method for indirect analysis of talc by transmission electron microscopy for asbestos mass concentration." This method is currently being developed by ASTM committee D22.07 (Sampling and Analysis of Asbestos) and is a modification of ASTM method D 5756 "Standard test method for microvacuum sampling and indirect analysis of dust by transmission electron microscopy for asbestos mass surface loading." The modification is to adapt existing methodology to the specific analysis of talc products and/or ore samples.
A program of laboratory "blank" analysis is in place and has been used to determine the detection limit for this analysis according to ASTM method D 6620, "Standard practice for asbestos detection limit based on counts."
PLAINTIFFS EXHIBIT
CAM-359
Page 1 of 3
ITA-Ratcliff-000871
Table 1. Result Summary
Sample Description
Sample No.
VAA10-005-AB
A11054-1
A11054 TEM Analysis Results
AMPHIBOLE
Number of Structures
Fiber Concentration
>5 um 0
<5 um 0
ppm BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
VAA10-005-CD A 1 1 0 5 4 -2
0
0
BDL
00
BDL
VAA10-006-AB
A 1 1 0 5 4 -3
0
0
BDL
00
BDL
VAA10-006-CD A 11054-4
0
0
BDL
00
BDL
VAA10-007-AB
A 1 1 0 5 4 -5
0
0
BDL
00
BDL
VAA10-007-CD A 1 1 0 5 4 -6
0
0
BDL
00
BDL
VAA10-008-AB
A 1 1 0 5 4 -7
0
0
BDL
00
BDL
VAA10-009-AB
A 1 1 0 5 4 -8
0
0
BDL
00
BDL
VAA10-009-CD A 1 1054-9
0
0
BDL
00
BDL
VAA10-010-AB A 11054-10
0
0
BDL
00
BDL
VAA10-010-CD A 11054-11
0
0
BDL
00
BDL
VAA10-011-AB A 110 5 4 -1 2
0
0
BDL
00
BDL
VAA10-011-CD A 11054-13
0
0
BDL
00
BDL
BDL = Below detection limit, as defined by ASTM D 6620
Page 2 of 3
ITA-Ratcliff-000872
Table 2. Analysis parameters
Sample Description
Sample No.
VAA10-005-AB A11054-1
Starting Weight
g 0.2435
VAA10-005-CD A 1 1 0 5 4 -2
0.2799
VAA10-006-AB A 1 1054-3
0.2602
VAA10-006-CD A 11054-4
0.2770
VAA10-007-AB A 1 1 0 5 4 -5
0.2515
VAA10-007-CD A 1 1 0 5 4 -6
0.2899
VAA10-008-AB A 1 1 0 5 4 -7
0.2695
VAA10-009-AB A 1 1054-8
0.2227
VAA10-009-CD A 1 1054-9
0.2378
VAA10-010-AB A 110 5 4 -1 0
0.2381
VAA10-010-CD A 11054-11
0.2227
VAA10-011-AB A 1 1 0 5 4 -1 2
0.2442
VAA10-011-CD A 110 5 4 -1 3
0.2380
A11054 TEM Analysis Parameters
Aliquot Filtered
mL
Grid Spaces Analyzed
#
Total Sample Analyzed
ng
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm
ppm
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.3 10 70.3 0.013 0.09 0.009 0.06
0.3 10 80.8 0.012 0.08 0.008 0.05
0.3 10 75.2 0.012 0.08 0.008 0.05
0.3 10 80.0 0.012 0.08 0.008 0.05
0.3 10 72.6 0.013 0.09 0.009 0.06
0.3 10 83.7 0.011 0.07 0.007 0.05
0.3 10 77.8 0.012 0.08 0.008 0.05
0.3 10 64.3 0.015 0.10 0.010 0.06
0.3 10 68.7 0.014 0.09 0.009 0.06
0.3 10 68.8 0.014 0.09 0.009 0.06
0.3 10 64.3 0.015 0.10 0.010 0.06
0.3 10 70.5 0.013 0.09 0.009 0.06
0.3 10 68.7 0.014 0.09 0.009 0.06
` Analytical Sensitivity is defined by ASTM D 5756 as the detection of one structure with minimum dimensions.
Page 3 of 3
ITA-Ratcliff-000873
* IM E RYS
Talc
Imerys Talc * 2500 San Miguelito Road Lompoc, CA 93436 (805) 737-1212 Fax: (805) 737-1411
TECHNICAL REPORT
Project No. A12008
Date:
13-Sep-12
To: From: Analyst(s): Copy:
Desiree Giroux Carol Slowik C. Slowik J. Pier, S. Sharma, G. Myers, H. Bruning
Title:
TEM ASBESTOS ANALYSIS OF ARGONAUT 2011 INFILL COMPOSITED SAMPLES
Request:
Twenty-one samples of composited Argonaut infill samples were submitted to the Technical Center for asbestos analysis by transmission electron microscopy (TEM). This analysis is in partial fulfillment of Imerys' Fiber Management Framework Program.
Results:
The attached TEM results were obtained using the current ASTM draft method "Standard test method for indirect analysis of talc by transmission electron microscopy for asbestos mass concentration." This method is currently being developed by ASTM committee D22.07 (Sampling and Analysis of Asbestos) and is a modification of ASTM method D 5756 "Standard test method for microvacuum sampling and indirect analysis of dust by transmission electron microscopy for asbestos mass surface loading." The modification is to adapt existing methodology to the specific analysis of talc products and/or ore samples.
A program of laboratory "blank" analysis is in place and has been used to determine the detection limit for this analysis according to ASTM method D 6620, "Standard practice for asbestos detection limit based on counts."
Page 1 of 5
ITA-Ratcliff-000874
Table 1. Result Summary
Sample Description
VAA11-004-AB VAA11-005-AB VAA11-005-C VAA11-006-AB VAA11-006-CD VAA11-009-A VAA11-010-AB VAA11-011 -A VAA11-013-AB VAA11-014-A VAA11-015-D VAA11-016-AB VAA11-016-CD VAA11-017-AB VAA11-017-CD VAA11-018-AB VAA11-018-CD VAA11-019-AB VAA11-019-CD
A12008 TEM Analysis Results
AMPHIBOLE
Sample No.
Number of Structures
Fiber Concentration
>5 um <5 um
ppm
A12008-1
0
0
BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
A 1 2 0 0 8 -2
0
0
BDL
00
BDL
A 1 2 0 0 8 -3
0
0
BDL
00
BDL
A 1 2 0 0 8 -4
0
0
BDL
00
BDL
A 1 2 0 0 8 -5
0
0
BDL
00
BDL
A 1 2 0 0 8 -6
0
0
BDL
00
BDL
A 1 2 0 0 8 -7
0
0
BDL
00
BDL
A 1 2 0 0 8 -8
0
0
BDL
00
BDL
A 1 2 0 0 8 -9
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 0
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 1
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 2
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 3
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 4
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 5
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 6
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 7
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 8
0
0
BDL
00
BDL
A 1 2 0 0 8 -1 9
0
0
BDL
00
BDL
Page 2 of 5
ITA-Ratcliff-000875
Table 1. Result Summary Sample Description
VAA11-022-AB VAA11-022-C
A12008 TEM Analysis Results
AMPHIBOLE
Sample No.
Number of Structures
Fiber Concentration
>5 um <5 um
ppm
A 1 2 0 0 8 -2 0
0
0
BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
A 1 2 0 0 8 -2 1
0
0
BDL
00
BDL
BDL = Below detection limit, as defined by ASTM D 6620
Page 3 of 5
ITA-Ratcliff-000876
Table 2. Analysis parameters
Sample Description
VAA11-004-AB VAA11-005-AB VAA11-005-C VAA11-006-AB VAA11-006-CD VAA11-009-A VAA11-010-AB VAA11-011-A VAA11-013-AB VAA11-014-A VAA11-015-D VAA11-016-AB VAA11-016-CD VAA11-017-AB VAA11-017-CD VAA11-018-AB VAA11-018-CD VAA11-019-AB VAA11-019-CD
Sample No.
A12008-1
A12008 TEM Analysis Parameters
Starting Weight
g
Aliquot Filtered
mL
Grid Spaces Analyzed
#
Total Sample Analyzed
ng
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm
ppm
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.2511
0.3
10 72.5 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -2
0.2743
0.3
10 79.2 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -3
0.2452
0.3
10 70.8 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -4
0.2625
0.3
10 75.8 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -5
0.2702
0.3
10 78.0 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -6
0.2738
0.3
10 79.1 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -7
0.2341
0.3
10 67.6 0.014 0.09 0.009 0.06
A 1 2 0 0 8 -8
0.2643
0.3
10 76.3 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -9
0.2639
0.3
10 76.2 0.012 0.08 0.008 0.05
A 1 2 0 0 8 -1 0
0.2516
0.3
10 72.7 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 1
0.2442
0.3
10 70.5 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 2
0.2592
0.3
10 74.9 0.013 0.08 0.008 0.06
A 1 2 0 0 8 -1 3
0.2449
0.3
10 70.7 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 4
0.2464
0.3
10 71.2 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 5
0.2433
0.3
10 70.3 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 6
0.2329
0.3
10 67.3 0.014 0.09 0.009 0.06
A 1 2 0 0 8 -1 7
0.2440
0.3
10 70.5 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 8
0.2433
0.3
10 70.3 0.013 0.09 0.009 0.06
A 1 2 0 0 8 -1 9
0.2543
0.3
10 73.5 0.013 0.08 0.009 0.06
Page 4 of 5
ITA-Ratcliff-000877
Table 2. Analysis parameters Sample Description
VAA11-022-AB VAA11-022-C
Sample No.
A 1 2 0 0 8 -2 0
A12008 TEM Analysis Parameters
Starting Weight
g 0.2425
Aliquot Filtered
mL
0.3
Grid Spaces Analyzed
#
10
Total Sample Analyzed
ng
70.0
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm 0.013
ppm 0.09
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.009
0.06
A 1 2 0 0 8 -2 1
0.2605
0.3
10 75.2 0.012 0.08 0.008 0.05
` Analytical Sensitivity is defined by ASTM D 5756 as the detection of one structure with minimum dimensions.
Page 5 of 5
ITA-Ratcliff-000878
F&PA Global Research, Applications, and Technology Center 71 Daggett Drive San Jose, CA 95134 PFM.analytical@imerys.com
TECHNICAL REPORT
Project No. A13152
Date:
6-Dec-13
To: From:
Desiree Giroux Julie Pier
Analyst(s): R. Fishell, A. Knorpp, J. Pier
Copy:
S. Sharma
Title:
TEM ASBESTOS ANALYSIS OF 2013 ARGONAUT INFILL SAMPLES
Request: Twelve samples of Argonaut ore from the 2013 infill drilling campaign were submitted to the Analytical Group for asbestos analysis by transmission electron microscopy (TEM).
Results:
The attached TEM results were obtained using the current ASTM draft method "Standard test method for indirect analysis of talc by transmission electron microscopy for asbestos mass concentration." This method is currently being developed by ASTM committee D22.07 (Sampling and Analysis of Asbestos) and is a modification of ASTM method D 5756 "Standard test method for microvacuum sampling and indirect analysis of dust by transmission electron microscopy for asbestos mass surface loading." The modification is to adapt existing methodology to the specific analysis of talc products and/or ore samples.
A program of laboratory "blank" analysis is in place and has been used to determine the detection limit for this analysis according to ASTM method D 6620, "Standard practice for asbestos detection limit based on counts."
ITA-Ratcliff-000879
Table 1. Result Summary
Sample Description
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
VAA13-01 VAA13-02 VAA13-03 VAA13-04 VAA13-05 VAA13-06 VAA13-07 VAA13-08 VAA13-09 VAA13-10 VAA13-11 VAA13-12
A13152 TEM Analysis Results
AMPHIBOLE
Sample No.
Number of Structures
Fiber Concentration
>5 um <5 um
ppm
A 1 3 1 5 2 -1
0
0
BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
A 1 3 1 5 2 -2
0
0
BDL
00
BDL
A 1 3 1 5 2 -3
0
0
BDL
00
BDL
A 1 3 1 5 2 -4
0
0
BDL
00
BDL
A 1 3 1 5 2 -5
0
0
BDL
00
BDL
A131526
0
0
BDL
00
BDL
A 1 3 1 5 2 -7
0
0
BDL
00
BDL
A 1 3 1 5 2 -8
0
0
BDL
01
BDL
A 1 3 1 5 2 -9
0
0
BDL
00
BDL
A 1 3 1 5 2 -1 0
0
0
BDL
00
BDL
A 1 3 1 5 2 -1 1
0
0
BDL
00
BDL
A 1 3 1 5 2 -1 2
0
0
BDL
00
BDL
BDL = Below detection limit, as defined by ASTM D 6620
Page 2 of 3
ITA-Ratcliff-000880
Table 2. Analysis parameters
Sample Description
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
Argonaut Infill Composite
VAA13-01 VAA13-02 VAA13-03 VAA13-04 VAA13-05 VAA13-06 VAA13-07 VAA13-08 VAA13-09 VAA13-10 VAA13-11 VAA13-12
Sample No.
A13152-1
A13152 TEM Analysis Parameters
Starting Weight
g
Aliquot Filtered
mL
Grid Spaces Analyzed
#
Total Sample Analyzed
ng
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm
ppm
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.2494
0.3
10 72.0 0.013 0.09 0.009 0.06
A 1 3 1 5 2 -2
0.2607
0.3
10 75.3 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -3
0.2622
0.3
10 75.7 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -4
0.2731
0.3
10 78.9 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -5
0.2672
0.3
10 77.2 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -6
0.2554
0.3
10 73.8 0.013 0.08 0.008 0.06
A 1 3 1 5 2 -7
0.2783
0.3
10 80.4 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -8
0.2638
0.3
10 76.2 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -9
0.2580
0.3
10 74.5 0.013 0.08 0.008 0.06
A 1 3 1 5 2 -1 0
0.2684
0.3
10 77.5 0.012 0.08 0.008 0.05
A 1 3 1 5 2 -1 1
0.2845
0.3
10 82.2 0.011 0.08 0.008 0.05
A 1 3 1 5 2 -1 2
0.2736
0.3
10 79.0 0.012 0.08 0.008 0.05
` Analytical Sensitivity is defined by ASTM D 5756 as the detection of one structure with minimum dimensions.
Page 3 of 3
ITA-Ratcliff-000881
F&PA Global Research, Applications, and Technology Center 71 Daggett Drive San Jose, CA 95134 PFM.analytical@imerys.com
TECHNICAL REPORT
Project No. A15383
Date:
4-Jan-17
To: From:
Desiree Giroux Julie Pier
Analyst(s): A. Boxleiter, R. Fishell, J. Lee, K. Persaud, J. Pier
Copy:
S. Sharma, G. Myers
Title:
TEM ASBESTOS ANALYSIS OF 2015 ARGONAUT INFILL SAMPLES
Request: Thirty-four samples of Argonaut ore from the 2015 infill drilling campaign were submitted to the Analytical Group for asbestos analysis by transmission electron microscopy (TEM).
Results:
The attached TEM results were obtained using the current ASTM draft method "Standard test method for indirect analysis of talc by transmission electron microscopy for asbestos mass concentration." This method is currently being developed by ASTM committee D22.07 (Sampling and Analysis of Asbestos) and is a modification of ASTM method D 5756 "Standard test method for microvacuum sampling and indirect analysis of dust by transmission electron microscopy for asbestos mass surface loading." The modification is to adapt existing methodology to the specific analysis of talc products and/or ore samples.
A program of laboratory "blank" analysis is in place and has been used to determine the detection limit for this analysis according to ASTM method D 6620, "Standard practice for asbestos detection limit based on counts."
ITA-Ratcliff-000882
A15383 TEM Analysis Results
Table 1. Result Summary
AMPHIBOLE
Sample Description
Sample No.
Number of Structures
Fiber Concentration
Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15)
VAA15-003-H VAA15-006-H VAA15-024-H VAA15-025-EF VAA15-025-GH VAA15-026-EF VAA15-026-GH VAA15-027-EF VAA15-027-G VAA15-029-H VAA15-031-AB VAA15-031-C VAA15-032-BC VAA15-032-DE VAA15-032-FG VAA15-033-BC VAA15-033-DE VAA15-033-FG VAA15-034-AB
A 1 5 3 8 3 -1 A 1 5 3 8 3 -2 A 1 5 3 8 3 -3 A 1 5 3 8 3 -4 A 1 5 3 8 3 -5 A 1 5 3 8 3 -6 A 1 5 3 8 3 -7 A 1 5 3 8 3 -8 A 1 5 3 8 3 -9 A 1 5 3 8 3 -1 0 A 1 5 3 8 3 -1 1 A 1 5 3 8 3 -1 2 A 1 5 3 8 3 -1 3 A 1 5 3 8 3 -1 4 A 1 5 3 8 3 -1 5 A 1 5 3 8 3 -1 6 A 1 5 3 8 3 -1 7 A 1 5 3 8 3 -1 8 A 1 5 3 8 3 -1 9
>5 um 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
<5 um 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
ppm BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
Page 2 of 5
ITA-Ratcliff-000883
A15383 TEM Analysis Results
Table 1. Result Summary
AMPHIBOLE
Sample Description
Sample No.
Number of Structures
Fiber Concentration
Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15)
VAA15-034-CD VAA15-035-AB VAA15-035-CD VAA15-036-AB VAA15-036-BC VAA15-037-A VAA15-037-D VAA15-038-AB VAA15-038-CD VAA15-039-AB VAA15-039-CD VAA15-040-AB VAA15-040-CD VAA15-041-AB VAA15-041-CD
A 1 5 3 8 3 -2 0 A 1 5 3 8 3 -2 1 A 1 5 3 8 3 -2 2 A 1 5 3 8 3 -2 3 A 1 5 3 8 3 -2 4 A 1 5 3 8 3 -2 5 A 1 5 3 8 3 -2 6 A 1 5 3 8 3 -2 7 A 1 5 3 8 3 -2 8 A 1 5 3 8 3 -2 9 A 1 5 3 8 3 -3 0 A 1 5 3 8 3 -3 1 A 1 5 3 8 3 -3 2 A 1 5 3 8 3 -3 3 A 1 5 3 8 3 -3 4
>5 um 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
<5 um 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
ppm BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL BDL
CHRYSOTILE
Number of Structures
Asbestos Concentration
>5 um <5 um
ppm
00
BDL
00
BDL
01
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
00
BDL
BDL = Below detection limit, as defined by ASTM D 6620
Page 3 of 5
ITA-Ratcliff-000884
Table 2. Analysis parameters
Sample Description
Sample No.
Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15)
VAA15-003-H VAA15-006-H VAA15-024-H VAA15-025-EF VAA15-025-GH VAA15-026-EF VAA15-026-GH VAA15-027-EF VAA15-027-G VAA15-029-H VAA15-031-AB VAA15-031-C VAA15-032-BC VAA15-032-DE VAA15-032-FG VAA15-033-BC VAA15-033-DE VAA15-033-FG VAA15-034-AB
A15383-1 A 1 5 3 8 3 -2 A 1 5 3 8 3 -3 A 1 5 3 8 3 -4 A 1 5 3 8 3 -5 A 1 5 3 8 3 -6 A 1 5 3 8 3 -7 A 1 5 3 8 3 -8 A 1 5 3 8 3 -9 A 1 5 3 8 3 -1 0 A 1 5 3 8 3 -1 1 A 1 5 3 8 3 -1 2 A 1 5 3 8 3 -1 3 A 1 5 3 8 3 -1 4 A 1 5 3 8 3 -1 5 A 1 5 3 8 3 -1 6 A 1 5 3 8 3 -1 7 A 1 5 3 8 3 -1 8 A 1 5 3 8 3 -1 9
A15383 TEM Analysis Parameters
Starting Weight
g
Aliquot Filtered
mL
Grid Spaces Analyzed
#
Total Sample Analyzed
ng
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm
ppm
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.2480
0.5
5 59.7 0.016 0.10 0.010 0.07
0.2515
0.5
5 60.5 0.015 0.10 0.010 0.07
0.2574
0.5
5 62.0 0.015 0.10 0.010 0.07
0.2597
0.5
5 62.5 0.015 0.10 0.010 0.07
0.2598
0.5
5 62.5 0.015 0.10 0.010 0.07
0.2516
0.5
5 60.6 0.015 0.10 0.010 0.07
0.2470
0.5
5 59.5 0.016 0.10 0.011 0.07
0.2521
0.5
5 60.7 0.015 0.10 0.010 0.07
0.2531
0.5
5 60.9 0.015 0.10 0.010 0.07
0.2589
0.5
5 62.3 0.015 0.10 0.010 0.07
0.2586
0.5
5 62.2 0.015 0.10 0.010 0.07
0.2554
0.5
5 61.5 0.015 0.10 0.010 0.07
0.2511
0.5
5 60.4 0.016 0.10 0.010 0.07
0.2591
0.5
5 62.4 0.015 0.10 0.010 0.07
0.2445
0.5
5 58.9 0.016 0.11 0.011 0.07
0.2596
0.5
5 62.5 0.015 0.10 0.010 0.07
0.2577
0.5
5 62.0 0.015 0.10 0.010 0.07
0.2540
0.5
5 61.1 0.015 0.10 0.010 0.07
0.2365
0.5
5 56.9 0.016 0.11 0.011 0.07
Page 4 of 5
ITA-Ratcliff-000885
Table 2. Analysis parameters
Sample Description
Sample No.
Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15) Argonaut 2015 Infill (VAA15)
VAA15-034-CD VAA15-035-AB VAA15-035-CD VAA15-036-AB VAA15-036-BC VAA15-037-A VAA15-037-D VAA15-038-AB VAA15-038-CD VAA15-039-AB VAA15-039-CD VAA15-040-AB VAA15-040-CD VAA15-041-AB VAA15-041-CD
A 1 5 3 8 3 -2 0 A 1 5 3 8 3 -2 1 A 1 5 3 8 3 -2 2 A 1 5 3 8 3 -2 3 A 1 5 3 8 3 -2 4 A 1 5 3 8 3 -2 5 A 1 5 3 8 3 -2 6 A 1 5 3 8 3 -2 7 A 1 5 3 8 3 -2 8 A 1 5 3 8 3 -2 9 A 1 5 3 8 3 -3 0 A 1 5 3 8 3 -3 1 A 1 5 3 8 3 -3 2 A 1 5 3 8 3 -3 3 A 1 5 3 8 3 -3 4
A15383 TEM Analysis Parameters
Starting Weight
g 0.2460
Aliquot Filtered
mL
0.5
Grid Spaces Analyzed
#
5
Total Sample Analyzed
ng
59.2
AMPHIBOLE
Analytical Detection Sensitivity* Limit
ppm 0.016
ppm 0.10
CHRYSOTILE
Analytical Detection Sensitivity* Limit
ppm
ppm
0.011
0.07
0.2547
0.5
5 61.3 0.015 0.10 0.010 0.07
0.2524
0.5
5 60.8 0.015 0.10 0.010 0.07
0.2508
0.5
5 60.4 0.016 0.10 0.010 0.07
0.2431
0.5
5 58.5 0.016 0.11 0.011 0.07
0.2487
0.5
5 59.9 0.016 0.10 0.010 0.07
0.2497
0.5
5 60.1 0.016 0.10 0.010 0.07
0.2386
0.5
5 57.4 0.016 0.11 0.011 0.07
0.2467
0.5
5 59.4 0.016 0.10 0.011 0.07
0.2481
0.5
5 59.7 0.016 0.10 0.010 0.07
0.2443
0.5
5 58.8 0.016 0.11 0.011 0.07
0.2564
0.5
5 61.7 0.015 0.10 0.010 0.07
0.2517
0.5
5 60.6 0.015 0.10 0.010 0.07
0.2534
0.5
5 61.0 0.015 0.10 0.010 0.07
0.2565
0.5
5 61.7 0.015 0.10 0.010 0.07
` Analytical Sensitivity is defined by ASTM D 5756 as the detection of one structure with minimum dimensions.
Page 5 of 5