Document NZkgJZG4mZbvkdVNNXGkQaNQ
TO:
Distribution
FU *
FROM:
Interoffice
DATE:
Communtoation
SUBJ:
J.L. Cox, R&D, Ponca City April 28, 1987
TCLP - ANALYTICAL CAPABILITIES
At the TCLP meeting held in Aberdeen on April 1, questions were raised as to the legality of using Headspace GC for TCLP testing and the precision of VCM analyses* It was decided that I would check on the legal issue and either Ed Sones or I would prepare VCM standards to run at ETC (Memphis), Aberdeen, Oklahoma City and Lake Charles, as well as go to ETC to see if we could offer advice on improving their accuracy.
Joe Ledvina confirmed EPA Method 5020 (Headspace GC) as an acceptable TCLP method. I also spoke with Gayle Hansen at the EPA who said that Method 5020 was acceptable as well as Methods 5030, 8010 and 8240.
On April 2, additional data was available from Ed Sones which showed
that while ETC might be having some problems with VCM analysis,
there was at least as much error due to loss of VCM from the
prepared standards sent to ETC for analysis. With this information,
Veldon Messick, Larry Kardos and I decided that it would be better
to first determine the precision of Headspace GC for VCM analyses
and then, if necessary, look at purge and trap analyses at ETC.
Modifications would be looked into, if needed, to improve the
precision of the anlysis* Once the precision of the analysis was
acceptable, reproducability of standards in the extraction vessels
would be checked and, finally, the reproducability of analyzing
resin samples.
i . +i
With only minor modifications on the integration parameters, 30 samples prepared at 50 wppb VCM and 30 samples prepared at 10 wppb VCM in TCLP extraction fluid #1 were analyzed. One 50 wppb sample was thrown out due to incorrect integration. The following results were obtained.
Cone.
Ave. Observed Concentration
Standard Deviation
50 wppb
10
48.2 wppb 9.9
3.15 0.33
VAB.0001104322
Joint IOC Page 2 April 28, 1987
Although the range charts show out-of-control points, the precision is better than was originally expected. The 50 wppb samples had 2 points which were low (31.6 and 30.7 wppb). These could have been due to several reasons other than the instrument. If these 2 points are removed, the average observed concentration is 49.5 wppb with a standard diviation of 1.07 and a range of 46.3 wppb to 51.4. With regard to the 10 wppb, the observed range was 8.8 wppb to 11.1. From this we conclude that our instrumental capabilities in Aberdeen are adequate for TCLP testing.
To check the reproducability of using the extraction vessels, a standard solution using TCLP extraction fluid #1 was prepared and used to fill each of the 8 vessels. In addition, the standard solution was also poured into headspace vials for analysis without going through the additional step of the extraction vessels. This was necessary because VCM loss was unavoidable with the mixing and handling procedures. So, although the intent was to prepare a 40 wppb standard, analyses of the headspace vials showed that the actual concentration was 31 wppb.
After the filled extraction vessels sat 18 hours, 4 samples were taken from each vessel for analysis. Three samples were analyzed that day and one was refrigerated for use later, if needed. The extraction vessels were carried through this procedure twice. VCM analysis on 3 samples from each of the 16 extraction procedures averaged 30.5 or 100% recovery. "X-R" charting shows this process to be in control. Standard deviation was 1.29.
The reproducability of analyzing resin samples will begin after the rotators arrive in Aberdeen.
Instrumental conditions and SPC charts are attached.
Joyce Cox Ch emi st
cdh/4-28DIS.JLC Attachments Distribution: Ponca City: SEM HJH JLI ELS Aberdeen: JF VEM FGJ CWT HGC DSC KWB JWW DWH PJK MLN JGC OKC: SHC Houston: JCL LZK LCVCM: JCC PLF
A
VAB.0001104323
Joint IOC - Attachment Page 3 April 28, 1987
Instrumental Conditions - Perkin-Elmer HS-100
Samples conditioned for 60 min, at 90C Prepressurization - 1 min, Inj, Time - 0,1 min. Needle Withdrawal - 1 min, Backflush - Valve on at 1,5 min.
Gas on at 1,3 min. Gas off at 4.9 min.
Vent - none Columns - 18" Carbowax 400/Durapak preceding backflush tee
18" Chromosorb 102 after backflush tee
Column Temp. - 119C Range - 1 Attenuation - 1 Pressure Gauge Readings
Nigrogen 37/33 psi head/backflush
Air
20
Hydrogen 20
I
Integration Paraumenters - Spectra-Physics 4290
Peak Width - 12 Peak Threshold - 12 BL - on at 1.3 min.-
SAMPLE CHROMATOGRAPH
CHANNEL fi
INJECT 04/13/8? 21s36s16 REPLAYED FROM BIN #
A
VCM Retention Time - 2.23 minutes Measured VCM concentration - 18.3 wppb
k
VAB.0001104324
Joint IOC - Attachment Page 4 April 28, 1987
SPECTRA-PHYSICS INTEGRATOR FILE
FILE2
MN- 5
REM FE= 2.
CH= "A* PSTM 1*
m " 2 It
PW 12*
RN = 2726 * PH = 0*
CZ 1 *
NV 0* RC 0* *
PT IX
*
TB LC SI RA
12* 1 a 0. 0* 1* 1.
CA 0.
DD = 2. T1 1*5
LS 0*
sz 1*
BP 0.
TT <1) = TT <2)TM TT ( 3) == TT < 4 ) =
1 1 *3 4*5
4 *9
TF<1)== T4" TF(2)= BL" TF<3>= ER" TF < 4) = T4 "
Ty<i)=
TV(2)= TV<3>= TV<4>=
1* 1* 1* 0.
RT <1)TM
2 *23
RF <1) 4S000
AN *DTECH"
CN(1)= RVCMX"
AN<1>=
n
CM<l)a PPM ANC2)a II
CU TM tt
CU<1)= " PPM "
SN(1) = *R
It
SA(1> =
1*
XF(1>=
1*
A
VAB.0001104325
Joint IOC -- Attachment
rage j
April 28, 1987
PRECISION TESTING--50 wppb TOM
A
m - iMJt
impifiiriTj CBAlf
-------m - 07.
-------usl - na
xn - is
ro uj
VAB.0001104326
Joint IOC - Attachment Page 6 April 28, 1987
PRECISION TESTING--10 wppb VCM
A
VAB.0001104327