Document NExRVDzxrgrGmObXqZ02wLDry

Spectroscopy Method 70-1 Job No. 16303 IDENTIFICATION OF POLYCHLORINATED BIPHENYLS BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY ' ""T-- - ' --" ' " SCOPE This write-up describes a procedure for the identification of poly chlorinated biphenyls (PCB) via gas chromatography/mass spectrometry (GC/MS). The method can be used to identify components in chromatographic fractions which have the number of chlorines and the molecular weight combinations typical of PCB's. Aromatic chlorine compounds having similar make-up can interfere with this identifica tion. ' PRINCIPLE A solution to be examined for PCB's is injected into a gas chromato graph coupled to a rapid scan mass spectrometer. As peaks elute from the GC, they are directly introduced into the mass spectrometer. Mass spectra are obtained for GC peaks of interest. The mass spectra are used to determine the rumber of chlorines and molecular weight. INSTRUMENTAL DETAILS Gas chromatographic separations are carried out with an F&M 5751-B gas chromatograph using a flame ionization detector and 1/8 inch stainless steel columns containing SE-52 packing on 80-100 mesh Chromosorb W. The.injection and detector blocks are maintained at 280C. The column temperature is programmed linearly from 170 to 280C at 4 per minute. Programming the temperature sharpens the peaks with longer retention times and facilitates their identification via mass spectrometry. A helium flow rate through the column of 30 ml per minute and a 1:10 split between the flame detector and the mass spectrometer are maintained. A 0.027 inch i.d. stainless steel capillary connects the chromatograph outlet to a Biemann type helium separator. This line is heated to 300C and the separator temperature is 280C. The helium separator is connected to the ion source of an Atlas CH 4B mass spectrometer. Remote actuated valves permit the venting of unwanted components such pas solvent. The electron impact source is operated at 70 ev and 60 amp filament emission at a source temperature of 280C. The ion Source also contains a total ionization monitor whose analog output is proportional to the concentration of ionizable material in the source. This recorded signal produces a trace which resembles the . original chromatogram. .' PROCEDURE A 20-30 ^u liter aliquot of a solution containing about 0.1% solute is injected into the chromatograph and the temperature program ' started. The solvent quickly passes through the column and is vented before reaching the Biemann separator. The vent valve is then closed 2 DEPOSITION 1 | EXHIBIT 1 I Papaqeorge | II Method 70-1 Page 2 and the column effluent passes directly into the ion source. A mass spectrum is scanned each time a peak is observed on the total ion monitor trace. , SPECTRAL INTERPRETATION 35 37 The mass spectra are examined for characteristic Cl and Cl muitiplet peaks. An analysis of these multiplets gives the number of chlorines per molecule. The mass of the peak containing only Cl^S is measured. This gives the molecular weight of the PCB being examined. The relative intensity of the parent ion peak used for the molecular determination provides information about aromaticity of the compound. Mass data of this type make it easy to differentiate between pesti- cides and PCB's. This technique does not permit easy identification of PCB isomers having the same chlorine number. NOTES The flame ionization detector instead of an electron capture detector is used for GC/MS work. Work involving GC only is based on use of the electron capture detector. The latter provides increased sensitivity and specificity for trace determinations. This detector system is not easily compatible with the mass spectrometer because of the small sample sizes used. In order to provide 1000X increase in sample size for the mass work, it is preferable to use the less sensitive flame ionization detector given in this procedure. In addition, the flame ionization detector provides a more complete picture of interfering components being separated and introduced into the mass spectrometer. db Monsanto Company Organic Chemicals Division Research and Development Laboratories Applied Sciences Section St. Louis, Missouri 2/70-M. W. Dietrich, J. P. Mieure, R. E. Keller WATER PCB-SD0000040353 ORGANIC CHlkUCAlt DIVItlOM WATER PCB-SD0000040354 i i I I I <* a--g---- --I / x o 7.0 6.0 5.0 ffM(l) 4.0 1.0 i.O 1.0 ORGANIC CHEMICALS DIVISION WATER PCB-SD0000040355 8-- --3--3--f 1.0 1.0 4.0 5.0 rnw(r) 4.0 r.o s.o 0.0 ib >-> e cn snu E-k ...... . > tiam.. AR0CL0R 125* Source: Monsanto MITM --- ' -- II MM .OJ5. - - - _N . 5 - - .JU *~ 8.0 7.0 6.0 5.0 PfM |i) !o 1.0 2.0 ORGANIC CHEMICALS DIVISION P^*\/---- 1 ,ff | L[ * vJ4 WATER PCB-SD0000040356 * IjO I T 7 7 3.0 f s.o ORGANIC CNMtCALS DIVISION 7.0 4.0 so rm(r) ao 7.0 0.0 0.0 6.0 ' 5.0 H>m(4| Xo 3.0 20 1.0 10 o cw jm'bo eraufea. E*"5 .... e<g i/xaj%& AHOCLOR 1260 Source: Monsanto -FU-- - . twsuMa -S.J. _ m vmMwmaim ft _ --- ftr p.o". twer nut _ t<Bf WWW 1IW13) OWtfl 500 . - 04. _ _ _ 1 B- ftF-vCMua *4TS*A4 J.OIi* ---- WATER PCB-SD0000040357 AROCLOR 1242 CAPILLARY FILM WATER PCB-SD0000040358 AtfSOKtfANLt AROCLOR 1248 CAPILLARY FILM WATER PCB-SD0000040359 ABSORBAN AROCLOR 1254 CAPILLARY FILM WATER PCB-SD0000040360 ABSORBANC 4000 3000 0.0 LUJiilJI Lj , JLJL 2000 ` 1500 CM-i 1000 900 800 700 * / .30 .40 .50 .60 .70 1.0 oo 3 WAVELENGTH (MICRONS) AROCLOR 1260 CAPILLARY FILM WATER PCB-SD0000040361