polychlorinated Biphenyl (PCB) Analysis
lr researching this subject one finds that the method presently recommended
Is a sound technical method based on gas chromatography with either electron
capture or microcoulometric detectors. A method, however, highly sensitive
but not highly specific. This GC method is presently being reviewed by
ASTM Committee D-19. 06 for acceptance as a standard method. This method
and other GC procedures had been used for analyzing samples containing
unknown PCB mixtures but the rise of a more specific yet less sensitive
technique CC/MS (gas chromatography-mass spectrometry) (2)(3) has given
identity to these PCB mixtures. GC/MS can identify and quantitate (4)
the isomers of concern by determining the mass ion of a particular isomer
at its known retention time. Several of the most abundant isomers are
available commercially (5) with more being made available monthly. The
work of Eichelberger et al. of the EPA (4) points out these problems and
proposes a GC/MS procedure that is more sensitive than previous GC/MS '
methods.
The question is whether a regulation should be monitored by a method
10 to 1000 times more sensitive than GC/MS yet lacking the specificity
of mass spectrometry. A combination of these techniques would be more
Ideal. Enough is known now, therefore, to use gas chromatography with
electron capture or microcoulometric detection to analyze for individual
isomers using the standards presently available and GC/MS to setup the
method conditions and confirm selected samples.
'
In fact standardization with individual isomers will be more sensitive than
with PCB mixtures where the standard response per unit weight is spread out over several isomers.
The following conclusions may be made. .
(1) Present PCB method lacks specificity.
(2) CC/MS is specific but 10-1000-fold less sensitive.
(3) Using the present GC method for specific isomers confirmed by GC/MS will exceed the present method sensitivity.
The point, therefore, that must be emphasized is that the analytical technology of the compounds on the toxic pollutants list is a highly dynamic field and like the toxicological data must be kept up to date in our methodology.
HONS 039612
February 13, 1974
1 -2-
References
1. "Method for Polychlorinated Biphenyls (PCB) in Water",
Environmental Protection Agency, Analytical Quality Control Laboratory,
Cincinnati, Ohio 45268.
'
2. Stalling, D. L. and Huckins, J. N., "Gas-Liquid Chromatography-
Mass Spectrometry Characterization of Polychlorinated Biphenyls (Aroclors)
and MC1-Labeling of Aroclors 1248 and 1254", Journal of AOAC, Volume 54
*4, 801 (1971).
3. Fishbein, L., "Polychlorinated Biphenyls", J. Chromatography
68, 345 (1972).
.
4. Eichelberger, J. W., Harris, L. E., and Budde, W. L., "Analysis of the Polychlorinated Biphenyl Problem", Analytical Chemistry, Vol. 46
#2, 227 (1974).
5. Analabs Research Notes, Vol. 12 42, 12 (1972).
MONS 039613
H* tfm f
r. S^tTZA. , /3?y /???,
T TbT'U
rt "i
k l i ii 1 1, X
l KL
m.
R0tfrs 1. Idealised mass spectra of mon* to baxachlorob-phen
yhabovemaaMBO .
..
,, -
f^niniNMi
awte* IM* rt*4*4 H *>
rtummuiM *
. - 0m lore%'Cc,c-a,>t.*v.>*
mami.
.
* Route 1. Typical pas chromatocrams o! standard and a water
' extract Aroclor 1254 in hexane
.
.j
at, v X"t. 'f c,"-
y\_
. " " - - i . Ui i i. 11. .u n ;."i
Rpura 2. Ion abundance ehremategrams of o mixture of 5 no cl
each of seven pure PCS isomers. Lower ehromaterrem ma-ii-
tors matsea 40-400; upper chromatocram monitors mo PCS
aubsst masses
^
__ ;
Solubility of chloiobiphcnyis '
>rc
*"
Compound
Solubility 0 e 7<i
(mg)
(ppm)
MonochlorobiplienvU
8.0
4-__
3.35 1.20
Dkhlorobiphcnyls
PUw :-t=.SA.i503 .90 < 4*........:...-...... :.......- o.os
* Trichlorobiphcnyls
S, 4, 4' ________________ 0.0S5
a, 4, 2*- ............................. 0.075
, Tctrachlorobiphenyls
2, 2*. 5. 5-
2, 0. 2\ 3-------------S, 3. 2', 5*. -------------
0.U4G 0.037 0.170
^f-2, 4, 2*. 4`- ............ ...... 0.070
*o iro tso i*o see sio
2, 4, 3\ 4'- ..............
. flMNSAHTM/C
. -2, 5, 3', 4'- .................
9 Caapesltlon of PCS residue* in fish. Arrava and triangles `1* 4 .....................
designate the r.oee characteristic peaks of the Aroeler. All Puiitnchloiobijilicnyli
0.055 0.011
0.175
Mabenrd peak* In chrcantocrass of fish extracts are PCS#, 2, 3, 4, 2', C' ................
Zhese uusligru deal mace the nuaber or chlorine *eoa in thc2 -1 5 2' 5*-
.1,* u Jrt..,ntntd hr
It-.er.iturc ' WNnchWhiplicuyi"
0.021 0.035
. prORiua r.t 4*C/uln, lie } pei, 50' X 0.020,,-l.d. Sl>i0
scot column,
cc-ns.
2* 2\ 4, 4\ 5, 3*. .
0.00SS
O
(7*
rafj y> Zax
5.8 3.0 1.17
1.40 1.44 1.35 0.07
0.0S5
O.OSO
0.04G
o.oin
0.170
O.Ou." o.ociT
0.041 0.175
0.023 0.02S
0.0355