BRANDERA AT A1 AT EUOIS2 08/06/97 04:59 PM
To; cc:
Subject:
ZIPPELATA1 ATWWPS CHAPMAOA AT A1 AT WWPS , AL, MOR6ANRA AT A1 AT WWPS , KRONBERG AT A1 AT WWPB , HEERC AT A1 AT EUOIS2 , MAA8JP AT A1 AT WWPS , DAUBEC AT A1 AT WWPS
RE: Can you help explain the attached
Roger, Jason, I presume your numbers are on. dry FEP basis ?
in 1995 and 1996 we analyzed C-8 in FEP 9302 at the V-disH..
yeart
199S
frequency
every hour
C-8 ppm 349, 285, 411, 221 (dry basis ?)
year; frecniency
C-8 ppm solids %
1996 332,
every half hour
284, 218, 249, 481, 555,
88.6, 93.9, 84.7, 78.4,
580, 392 (dj?y basis ) 89.1, 89.4, 89.2, 92.0
After that w improved our analytical methods drastically and found
date
24-6--97
type
100-LG
frequency
every 20 minutes
solids 71.4, 67.8, 59.6, 65.8, 45.1
C-8 ppm Me-bl 3282, 3305, 3711, 3465, 3310
C-8 ppm ECD
3191, 3312, 3590, 3392, 3197
date
type frequency solids C-8 Me-bl C-8 ECO
16-7-97 5100-H
every 70,7, 70.2,
3396, 2967, 3204,
20 minutes 69.8, 72.1, 3499, 3026, 3160, 3152,
71.2 3243, 3276
3267
We found it is VERY VBRY important to be in the linear rang of the
BCD detector. We diluted the extracted solutions 200 times to achieve that. An BCD has a rather short linear range. You should check what
it is by making a calibration line* If you are above the linear
range, you will find far too low results, unless you interpolate in
the non-linear area. Also we had a fouled detector, and C-8 wae blamed by some people.
As you can eea the BCD gives slightly lower resulta, probably ther
are still some in-gitu surfactants in the wet fluff, which are eaan
by He-blue but not by BCD. We don't have the variations anymore that we used to see, I don't know why, but there were quite some process
EID936987
changes since early 1996.
One trick you can do is to reanalyze in a random order, to see if
moment; of analyzing was a factor. good luck, Rik