TOLUENE,aBENZENE, XYLENE, AND OTHER VOLATILE COMPOUNDS
ASSOCIATED WITH GLUE SNIFFING Type B Procedure
Nansh C. Jain
APPARATUS
1. Centrifuge 2. Gas chromatograph (Varian Aerograph Hy-Fi Model 6OO-D),equipped with a flame ionization
detector, a 0-I-mv recorder ( k e d s and Northrup Speedomax H or a suitable substitute), and a spud stainless-steel column, 6` long and 34" O.D., packed with 30% Carbowax 20M on acid-washed
60-80-mesh Chromosorb W
OPERATPIC CONDITIONS
i
cas Flow Rates: hnyitdroroggeenn--3258mml/lm/minin
L
-oxygenb 100 ml/min
Column Temperature: 100C or 13OoC,depending upon the compounds to be separated
Table 1 for retention times at these temperatures)
Injection Port Temperam: 160C
Attenuation: X 32`
4
Condition the column overnight at 180C with a flow of nitrogen carrier gas. 3. Shell vials, 2 dram, or suitable tubes, such as culture tubes 4. Blood pipettes, calibrated "to contain" (T.C.), 0.5 ml 5 . Syringes(Hamilton 7101-N), 1 pl
REAGENTS
AU chemicals used should be analytical reagent grade.
1. Sodium tungstate, 10g/100ml Dissolve 10 of sodium tungstate in water and dilute to 100 ml.
2. Propyl acetate! internd-control solution, 20.9 mg/100 mle
:Toluene has been selected to illustrate this method. Other volatile compounds a n be similarly assayed. Oxygen is used for increased sensitivity, but air can also be used.
:This can be wried t o give increased sensitivity.
If propyl acetate is the volatile substance suspected, a different compound, such aa acetone, should be urcd
control. elt is not necessary t o know the exact concentration of the internalcontrol solution M long as the same for both the known and unknown samples.
Toluene and other volatile compounds B
Pipet 0.25 ml of propyl acetate into a 1-liter volumetric flask and dilute to the mark with distilled water. 3. Toluene reference solution, 21.65 mg1100 ml Pipet 0.25 ml of toluene into a 1-liter volumetric flask and dilute to the mark with distilled water. (If any of the other substances are suspected, prepare the appropriate reference solution).
SPECIMEN PREPARATION
If
1. If the specimen is blood or urine, no further preparation is needed.
-7. If the specimen is stomach content, centrifuge it to obtain a workable volume (OS ml or more) of
supernatant. Use the supematant as the test sample.
i 3. If the specimen is tissue, add 10.0 ml of the sodium tungstate solution to 5 g of tissue. Homogenize the tissue, then centrifuge the homogenate. Use the supernatant as the test sample.
PROCEDURE
<
Qualitative Determination
I . Examine the gas chromatograph to be sure that the operating conditions stated above are met. Refer to Table 1 for setting the column temperature (100C or 130C, depending upon the separation needed foc the compounds under investigation).
2 . Draw about 0.5 pl of distilled wate; into the syringe, followed by about 0.5 pl of the test sample. Inject this mixture into the gas chromatograph.
TABLE 1.
P:f RETENTION TIMES OF VARIOUS VOLATILE SUBSTANCES
AT SPECIFIED GAS-CHROMATOGRAPHIC OPERATING CONDITIONS*
Column Temperature, 100C
Column Temperature, 130"C
Compound
Retention Time,
Relative
minutes
Retention T i m e
Retention Time,
Relative
minutes
Retention Timet
Acetddehyde Acetone Benzene Chloroform Ethanol
Formaldehyde Isobutanol Isopropanol Vethanol '(ethyl ethyl ketone Propyl acetate Toluene Xylene
2.56 2.38 4.31 5.62 3.50
1.81 7.25 3.19 3.25
3.38 4.50 6.94 11.38
0.73 1.25 0.60
0.68 1.62 0.76
1.23 2.81
1.33
1.61 3.25 1.00 2.12
1.53 1.00
0.52 1.38 0.65
2.10 3.94 1.86
0.9 1 2.06 0.97
0.93 2.06
0.97
0.97 2.19
1.03
1.29 2.75
1.30
1.98 4.06 1.92
3.25 6.38 3.01
'Stainless-steel column, 6' x @, packed with 30% Carbowax ZOM on Chromosorb w,AW,
60-so m e h ; nitrogen flow rate, 35 ml per minute: hydrogen flow rate, 2 8 ml per minute: Oxygen now rate, 100 ml per minute. iTo ethanol.
339
I I
,
I
CRC Manual of Analytical Toxicology
3. If no peak results from the injection, the test is negative. Make sure that the operating conditions are set for maximum sensitivity before reporting the result as negative. If there is a peak corresponding to the retention time of the compound suspected, perform the quantitative procedure to determine the amount present.
I
I Quantitative Determination
.tf 1*'-.
jII I `t`
I,;!I' ! `i
i/1I!
',:I 4`
1. Pipet 0.5 rnl of the n-propyl acetate internal-control solution into each of two shell vials, one labeled "`known" and the other labeled "unknown".
2. Using a 0.5-ml T.C. blood pipette, transfer O S ml of the toluene (or other appropriate) reference solution to the sheU vial marked "known". Wash out the pipette three times with the intemal-control solution in the shell vial to be certain that all of the known sample is transferred to the vial.
3. Using a second 0.54 T.C. blood pipette, transfer 0.5 ml of the unknown solution to the shell vial marked "unknown". Wash out the pipette three times with the internalcontrol solution III the sheu
vial to be certain that the specimen is quantitatively transferred. 4. Inject duplicate 0.5-/.d aliquots of the known and unknown solutions into the gas chromatograph,
using 1-4Hamilton syringes. 5 . Measure the average peak heights corresponding to the substance to be quantitated and to the
intemal-control solution for both the "known" and "unknown" samples.
,
CALCULATION
-.
d --d For blood, urine, or stomach content, when used directly!
a
and for tissue (5 g) homogenized with 10.0 ml of the sodium tungstate solution,8
where C, = concentration of the specimen;
C, = concentration of the reference solution for the volatile to be assayed;
h, = peak height of the "unknown" mixture at the retention time of the substance to be assayed;
h, = peak height of the substance to be assayed in the "known" solution;
R, = peak height of the internal control in the "known" mixture;
& peak height of the internal control in the "unknown" mixture.
INTERPRETATION (See p. 7.)
The presence of such solvents as benzene, methyl ethyl ketone, toluene, and xylene in blood and lungs is usually associated with glue sniffiig. Most volatile compounds ingested anGor inhaled will be f o k d unchant9d in the biologic material. Isopropanol, however, is metabolized to acetone. If ingestion Of
`If further dilution$ are necwsery because of high concentrations of volatiles present, make sure that thb i s taken into
account in the calculation.
f
i
340
Toluene and other volatile compounds B
isopropanol is suspected, test for both isopropanol and acetone. Acetone is also likely to be present in
blood samples from diabetic patients.
ACCURACY AND PRECISION
The technique of injection is important and should be consistent for both the reference and unknown solutions. Best results are obtained when reference and unknown solutions have approximately the same concentrations, so that both give similar peak heights following injection of equal volumes. Also, for excellent accuracy and reproducibility, the r a t i o s 5 a n d L.! should equal approximately 1.
RS Rx
REFERENCE
Jain, N. C., Direct Blood-Injection Method for Gas-Chromatographic Determination of Alcohols and Other Volatile Compounds. Clin. Chem., 17, 82,1971,