Document M4JGrbrvXgvRVxq69LEj3L59a

fcruirfscc itaasurttiicSiieii (N* 4 and Location) As Noted * -i3T {Nafrne and Location) / E. _S--Ramev - CCCTC B. J. Bland - CCCTC Method for Nickel Carbonyl GEN 55 (Rev.6/74) DATE October 27, 1976 REFERENCE NO. ESR-303-7 6 BJB-693-76 Attached Ls a method for determination of nickel carbonyl at levels of 0. 007 mg/m3 (0. 001 ppm). Please look it over and try it out if you have the opportunity. You should all be aware of the possibility of nickel carbonyl formation when you have CO in the presence of Nickel (either in material of construction or in a catalyst). We are still looking at ways to analyze oil mists. Drager has a new tube for EtzO which looks like it may satisfy our .. ^ requirements. ^ 5- Procedures for Sampling (and counting) Asbestos Fibers as recommended by the joint committee of ACGIH and AIHA on aerosol hazard evaluation is attached for your use. Please note the plastic shroud around the membrane filter holder (Figure 1). This device can be made from a plastic bottle of appropriate diameter and length. The stated purpose of the shroud is found in the second paragraph -Mr. C. J. Duey - Bay City Mr. R. E. Green - Clear Lake Mr. R. M. Guedin - Corpus Christi Mr. J. M. Ramey - NYO Ms. A. E. Retherford - Bishop Mr. M. R. Stenzel - Pampa 003962 \ K. . .i.V. 11 mil AMERICAN INDUSTRlAMIVfJIKNK A.N.S' >C \ !*|(-\ IOIRXAL. Volume Pathos 7*2-7h. J.imun u;h n !! ' Nickel Carbonyl: Its Detection and Potential for Formation RICHARD S. HR IK!'.* FRED S. VENAIlEE.i and ROBERT S. AJKMIANt Piuo ll/waroh and Kniiinrrrlng Company. I.indrn. AVi'r J'lrsty and liumblr Oil Hrftninn Company Rcccm recommendations for control of nickel carbonyl exposures have been con sidered in the design of a simple, sensitive field method for sampling nickel carbonyl in air and process gases. The method involves collection in dilute aqueous HC1. pH adjustment, nickel complex development with alpha-furildioximc, and ex traction with chloroform. Color intensity is compared visually (or for greater accur acy, spcctrophotometrically) with liquid standards. Sensitivities on the order of 0.001 ppm are obtainable. A detailed development of the thermodynamics associated with the formation of nickel carbonyl is also presented to show the maximum concentrations of nickel carbonyl that may be formed over a wide range ol CO concentrations, temperatures and pressures. Introduction determining the liberated mercury spectro- NICKEL carbonyl is a highly toxic graphically. A parallel stream of air is drawn material which causes severe damage to through an oxidizing reagent to convert the lungs. In view of this, practical limits fonr ickel carbonyl to cat bon dioxide and the exposure control have been suggested by stream is then passed over the mercuric Stokinger* and arc shown in Table I. In oxide: the liberated mercury is again deter yu|rr to monitor nickel carbonyl concentra- mined spectrographically. The difference in t the suggested levels, extremely sensi- tiic amounts of mercury vapor obtained i\ ...nnpling and evaluation techniques are corresponds to the nickel carbonyl content of "quired. An instrumental method of analysis tiic air. A sensitivity for this.method of 0.0014 or nickel carbonyl has been described in ppm is reported.1. In still another variation which the sensitivity of the instrument ranged the nickel carbonyl vapor is absorbed in an iroin 0.05 to 4 ppm.7 Another such method acidified solution of chloraminc-B in alcohol. "ports a sensitivity of 0.5 ppm.J The author reports that this absorbent is better than an alcohol solution of iodine or Chemical collection and analytical methods a solution of iodine in carbon tetrachloride. are legion. In one. the air sample is bubbled through a saturated solution of sulfur in trinuortvthvlene. The sulfur reacts with the Hir-icI to form a precipitate which is exam- The nickel determination is accomplished colorimetrically using dimethylglyoximc. For a 30-minutc sample, at the suggested sampling rate of 0.5 litcr/niinutc a sensitivity tied spectrographically for emission in the of 0.01 ppm is obtained.6 A fourth method is `hraviolct range, giving a sensitivity of described using iodine in carbon tetrachloride 'round 0.0003 ppm nickel carbonyl.4 Another as the collection medium and following tip method involves drawing air through a tube with a colorimetric dimethylglyoximc deter remaining red mercuric oxide at 200C, and mination of nickel; a sensitivity of 0.1 ppm nickel carbonyl is claimed.7 Collection in dilute sulfuric acid followed by spectropho *Ibtl rj] kr.r-.ich Di.i.iim, Ewo RrM-.ifli and Kiigmcer''t5*m`P4,,V. 1*. O. bo* 45. Linden, New Jriscy. Medieul l)i\iiun. Humbie Oil Rtfininx Cunipauy, Baton tometry using sodium dicthyldithiocarbamatc has also been reported." Lnui'uite. -Medical Uitbinn. Humble Oil and Refining Company, r- O- Ro* Ja, Linden, New Jeney. In view of the extreme sensitivity required for adequate monitoring at the 'low levels 72 /,/;. t.'t/ij/ //vgiVtir Jounml 73 Tahi.k I ' red Limits for Control of Nickel Carbonyl ln*lMan< (ppm) Single Air Daily .Wragp Out* Plant Sample CConcentration (!>!!>) v.ilur Discontinue opr ration require respiration protect ton) Slmf(k>*n operation 0.04 0.2-2.0 2.0 ft.OOl 0.001-0.005 0.005 (UXX13 n.not determine tuiemgrams of nickel contained in the 5 ml chlomfoim. For Greater accuracy, die linal color can he compared against reagent blanks in I-cm cells in a spectro photometer set al Cl.') milli-microns. The volume sampled tint- the weight of nickel collected are then used to calculate the nickel carbonyl conivniiation in the gas. `Omcrntration averaged over one inmuh, Discussion suggested in Table 1. the instrumental methods were considered to be inadequate and too expensive. Chemical methods had to he simple, rapid and adaptable to field tech niques for romine monitoring. Some of the methods described above met the sensitivity requirement but failed on some of the other field evaluation needs. The purpose of this paper is to describe a new technique suitable for field use. which is derived in part from a modification of the analytical method of Forrester and Jones. , Sampling and Analysis The air or process gas sample is bubbled into 10-15 ml of 3% aqueous HC1 solution !p > midget impinger bubbler for 30 minutes le hovir at 0.1 cubic feet per minute w.jiil. A filter assembly should precede the bubbler to insure that no nickel solids enter the rollecting solution and that only nickel carbonyl is collected. The liquid is transferred to a 60 ml separatory flask in which the following reagents are added in sequence: . {I) two drops phenolphtbalein: (21 four drops ammonium hydroxide (ap proximately 20-30% aqueous solution) (3) 20% aqueous sodium hydroxide to the phenoiphthalein end-point, plus three drops excess; rh 3.0 ml alpha-furildioxime solution (1 % alpha-furildioxime in 1:1 alcohol/watcr): and (h 5 ml chloroform. At ter shaking for at least one minute and allowing the chloroform layer to separate k"low the aqueous phase, the chloroform is drawn into a test tube. The color is compared with pre-established standards to The reaction lias a critical pH when carried out with sodium hydroxide alone; therefore, ammonium hydroxide is added to increase the pH range. Nickel reacts with alpha-furildioxime to give a yellow color in the presence of chloroform. The maximum color development 01 cuts in about five min utes and is stable over a period of weeks. No serious interferences .ate known, although copper gives a brown precipitate which can be removed by washing the chloroform ex tract twice'with 10r/< ammonium hydroxide. Palladium does not interfere on the basic side but does on tin- acid side. Chromium can be tolerated up w> about 0.1% and iron and vanadium up to about 0.5%. The color standards used for visual comparison in the field were made from aqueous sodium dichromatc solutions and matched in a spec trophotometer to chloroform color develop ment at various microgram levels of nickel. The collection efficiency of 3% HC1 for nickel was checked in a calibration chamber and found to lie greater than 90%. The tech niques for solution transler, pH control, color development and laser separation were all mastered in the field without any difficulty by technicians trained on the spot. Although the sampling lime is limiting, the actual analyti cal procedure takes no more than 5-10 minutes. The sensitivity of the method was checked in both a calibration chamber and in actual field tests. Samples taken from certain process lines suspected of containing nickel carbonyl showed distinct color intensity dif ferences between 0.004 ppm Ni(CO)4 and the blank. It is expected that based on the color intensity noted for 0.004 ppm the meth od would be sensitive visually to at least 0.002 ppm. Other tests in which the color intensi ties were compared in a spectrophotometer 74 indicated a sensiti 30 minute samp sampling rale), yellow due to tlu found in routine X:\OH. The sens be extended if nit 'Fhe entile unit Jet Constant Rn. from Union Inch: tion, Fall River, suction in areas pumps were, not p ence, no positive 1 air atmospheres c: above, positive n samples taken frot pressure in the li sampler. The Uni ed to a field kit is . Potential for Niclct Estimates of the nickel cfj50hyl fr Figure 1. Nickel freon aspirator, b reagents and sealed 1 ,1 ' II m '-T' 003964 1 J,`tiary-Fcbruary, 1065 :i-ii .1 M'lMlkily nl O.OtKll! ppm |fn- 20i nun- sampling' pcrinds (:u 0.1 cfm simp! hi: i.itc'. 'i'lio blank ni:iy lie slightIv wllm* tliii* to 11 traces ol nickel normally Ininin in commercially available I IC'l anil NaOIt. Tlu1 sensitivity of the method would he extended if nickel-free reagents were used. The entire unit has been run using tlu: UniJet Constant Rate Air Sampler (available: from Union Industrial Equipment Corpora tion. Fall River. Mass.) as a source of Miction in areas where electrically operated pumps were not permitted. In actual experieme. no positive results were obtained in the air atmospheres checked. However, as noted above, positive results were obtained with sampli-s taken from process lines in which the pressure in the line teas used to drive the sampler. The Uni-Jet sampling system adapt' til to a iield kit is shown in Figure 1. Potential for Nickel Carbonyl Formation Estimates of the maximum concentration of 'nickel carbonyl from a reacting system may be inhi.ihle in .i.s e.ssiu; tbe potential hazard of this gas in the work enviionmenl. For tbe situation wheie CO is contained in other gases not reacting with nickel, for example, hydrogenation processes over nickel catalyst in refinery operations, a thermodynamic evaluation has been pei formed. In the ccpialioii N'i h4CO = N'i(CO)4 the equilibrium may be expressed in terms of equation (1). "Nitcoid r flT ' 3 (D [C0]4 LPT . where: K --equilibrium co'nstant nxi (eo) 4=moles of nickel carbonyl at equilibrium, gm-moles nco=moles of carbon monoxide at equilibrium, gm-moles n-r--.total moles at equilibrium, gmmoles PT=total pressure at equilibrium, ' ^atm. Kr=ratio of fugacity coefficients, approximately equal to one for the range of temperatures, pressures and CO concentrations shown in Figures 2 through 6. In Figures 2, 3. 4. 5 and 6, the equilibrium concentrations of Ni(CO), are shown as a function of temperature, total pressure, and CO concentration at levels of 0.0005 to 3.0 molc-^c in the feed gas. The equilibrium constant may be calcu lated from free energy considerations. How ever, it should he noted that there is a wide variation among the leported heat of reac tion values used to calculate the standard free energy--from 27,500-52,600 cal/gmniolc. Work at Columbia University10 showed that the value of 38,650 cal/gm-mole report ed by Sykes11 was probably the most reliable. Accordingly, equation (2) was developed to estimate the equilibrium constant: l( 1. Xickcl' carbonyl sampler showing 'c'n aspirator, bubbler, prcfiUcr, glassware, *ttcriU and sealed color standards. (2) -38,650 + 99.3T K*e L RT Vv 1--!"p;i n-K-wr s * V . j.s : 003965 hiiht'tiiiil Ilygicnr Journal ir, Figure 2. Figure 4, k Figure 3. "N 003966 \ v ** * ** /*' /;/ ////; \ . I*th ; Kqiiution f.1 isMihrt! !\ M'lei line suluf-. nl y, 1 and l^j- and assuming a sciirs r *I sabi'1' for x. I lie \alue fif .A is ralrtilulrd and plotted; the \alue of x eonespotuline to i"0 is the correct solution. 1 jittering x in equation M i eiw-s the equili i brium cum conation "I NilCt). . (1) 0.25 xy 100 - 0.75 xy 106 = ppm. Ni (CO lo ease the labors of baud compulation, a computer program was written for the IHM 7090 to solve equations f3'-and (4*. The concentrations of nickel cat bony! obtained are plotted in Figures 2 through 6 as a function of , reaction temperature, total pressure and car bon monoxide concentration in the inlet gas. References whore c=napcrian logarithmic base, 2.718 T=absolute temperature, K R=gas constant= 1.9872 caI/K gitimole To solve the gcncrttl case use 100 gm-n.oles of the original mixture as a basis for fuither computation. Then let y=gm-moIcs of CO (or molc%) in origi nal mixture x=fraction of CO converted at cquilihiiutn The composition of the equilibrium mixture will then be neo=y(l-x) *'xt(co)<:=0.25 xy nT= 100-0.75 xy By substituting these values and equation (2)'in equation (1) and transposing, equa-tion (3) is obtained: STUKIMStK. U. K. 'I lit- Mrt.iK iKtcIikIiim; laa.I. ... Patty, F. A., Kd. Indmtrud Ifygirne and Vol. 2. 2ml Kev. Ed., p. HI2. Jntmcirurr i'uhlish.-rs N. Y. (IOGJi. McCaku.y, J. K.. K. .V S*i iv.'un, and K. If. Ovi7r.it \ vV * Kcrordim; Nirkrl C.irlinu'l Detector. ,4no/. Ch*ut JH T 8110 (May IMG). 3. It'll., K. K., ami K. K, Sur/M'N: Dfimniii.tiioia * Truce' of Tctrartlnllrad and Nirkrl (!aih<mt| il \u Papef. PjUshur.cli Conference of Analytical CIh'Om'Mn and Applied Spectroscopy fpri. 2ft, PHiUi. 4. Pitkt, O.: Detection of Traces >! Nirkrl (!,irlmn\l lit tin Factory Atmo'tilirro. Atrh. Malndiet 21: h74 !Ort.-Nov. 1%0). 5. Vmt.'mxc, N. Sii.: Determination of Vrare .(moo/ifi / .ViVJfccf Carbonyl in Air. U.S.S.R. Patent n5.fH4 f|Vl 15, l%2). 6. lift vaKIiv. A. A.: The Determination of Mrn.\:i,in Quantities of Nickel Tctraearbonvl ami Its .Solid Dcioin* Iftf.ll).position Products in Air. /.mndst. l.ob. 13K lFel 7. Hu.s'ou*. Cr. A., and W. I'ikivi'Ua: *1 lie Rolr of Nirk>*l Catlxmyl and it Dctcnumniion m Air. ,4r/rittJu*t; H: 193 (Aug. 19011. 8. Ai.ijcnMuji. O. K., K. Nf.# (..oo\k. and N. J. l.isnr SjHTtrophoiomctric Determination of Tr.uvs of Nirkrl Jnd. Eng. Chcm., Anal. Ed. 18: 20(1 (Mniih. l94t#. 9. KokuksTi.k, J. S., and L. Jom.n; Kapid Cliriou.il Determination of lion, S'tikrl and Vanadium in Petro leum Oils. Anal. Chern. 32: 1443 (Oci. PJoO). 10. Sckulman, J. H.: Personal Communication. Jul 5. l%3). 11. SvKrs, K. W.: The Heat of Formation of Nirkrl Carbonyl from ComlniMion Colorimetry. Cheat. Sue. J (London) p. 2053 (1958). (3) v 0.25 xy f 100 - 0.75 xy 1 [yli-x)]4 * L PT J -38,650 99.3T I.9872T A 003967 ) /$tyLLS^ 7/Z7/7S'' l Recommended Procedures for Sampling and Counting Asbestos Fibers Procedures for the Evaluation of Occupational Exposures to Airborne Asbestos i Joint AIHA-ACGIH Aerosol Hazards Evaluation Committee ibrous asbestos has been FIDENTIFIED as the causative agent of asbestosis and has been associated with an increased cancer incidence. The American Conference of Governmental Industrial Hy gienists (ACGIH) has established (1974) a Threshold Limit Value (TLV) for asbestos of 5 fibers greater than 5 micrometers in length per milliliter of air. A TLV footnote specifies the determination shall be made by ae membrane filter method at 400 X to 450 X magnification and with phase contrast illumination. These procedures define a standard of sampling and of processing the collected samples in order to evaluate occu pational exposure to asbestos fibers. Sampling Airborne samples of asbestos must be collected on membrane filters which retain at the surface essentially all of the particles in excess of 0.5 /on in diameter. The filter must be rendered transparent by the mount ing medium. The filter should be packed in a sealed holder which is capable of being readily opened for sampling purposes and of being resealed after the sample has been collected. The filter must be fully exposed during sampling. (Recommended filters and filter holders are listed under Supplies). The filter must be practicallv_dust-free with an average background' ebubi If?'Jess than 25 Reprints of tins article are available for, purchase from either the American' Xmli'-sf-iV Hygiene Association. 66 *ouih Miller Koad.- Af.mil. Ohio 44313 or the American Conference of Govcrnour.lJl ir.ousirial Hygienists, P.O. 1937, Cincinnati, Ohio 43201. The cost is Sl.00 pee *OPT. fibers per square centimeter. At least 2% to 4% of the filters intended for collecting samples should be set aside to determine the background count (See Counting and Cal culations sections). In addition, the filter resistance should not exceed 3 mm of mer cury when filtering air at the rate of 0.3 liters per minute (1pm) per square centimeter of filter. When collecting the sample, care must be taken to prevent dust from falling or from being projected onto the open filter. This may be accomplished by pointing the filter head downward. A guard or shroud should be attached to the filter holder to further protect the open filter from contamination as well as optimize uniform deposition on the filter. Its diameter should be about the same as the holder so that it can be attached tightly to it and it should project at least lVi times the filter diameter in front of it. A protective device is particularly important if the filter head is to rest on contaminated clothing. (See Figure 1.) The filter holder may be stored for a limit of six months once it is resealed. The sample should not be too dense, since samples in which particles overlap must be rejected as uncountable. Similarly, a large enough volume of air, should be sampled so that there are sufficient numbers of fibers in each field. -Experience indicates, that more than 150 particles per field, including nonfibrous background,- may- interfere with counting the'-sAuiple! ; ii is tiosirable that there be no more than.fibers per field. Two or three,samples'should be collected at each sampting location/The .results should i 003968 i D-->*" 84 hi'-3,9 Figure 1. Diagram of shroud for membrane fil ter holder. The shroud may be constructed from a plastic bottle by cutting off the neck and drilling a hole in the bottom. To use, push the filter holder through the hole and reach in through the top to remove the holder cap. To remove the holder, re place the cap and push the filter holder forward through the hole. be averaged to obtain the fiber concentra tion. Breathing zone samples should reflect the worker's entire normal work day. They must be collected as close to the worker's nose and over as long a time period as possible. Calibrated battery powered per sonal pumps (see Appendix III, Pump Cali bration) are satisfactory air movers. They can be worn by the worker and supported at the belt by means of a clip. The sam pling head, which can be pinned to the col lar, is connected to the pump by flexible tubing. Such pumps can be operated continu ously for about eight hours and recharged overnight. If a 37-mm filter is used to collect the sample, the pump should be adjusted tojampie between 1 and 2 liters per minute. The volume of air sampled should be adjusted so that an optimum density of dust is col lected on the filter. - Since the dust concentra tion is. not known before sampling, the op timal sampling period must be determined by trial and error. It is/suggested that, unless experience indicates otherwise., the first sam ple should be collected from 20 liters of air, the secondTfrbnPftjjmd the .third from 80 February, 1975 liters._If a filter other than 37-mm is used, the volume of air sampled should be altered in direct proportion to the open filter area. Samples collected during short periods of exposure should be collected at a high flow rate and if necessary, for the entire exposure period. In all cases the count should be adjusted with regard to time, so that the exposure is expressed, as an 8-hour timeweighted average exposure. For compliance evaluations with short term exposures, 37mm filters samples should be collected for 15 minutes at a minimum of 2 lorn. Sample Preparation A solution one-to-one by volume of dimethyl phthalate and diethyl oxyiate, in which has been dissolved 50 milligrams of membrane filter material per milliliter of solution is the preferred counting medium. The chemicals used to prepare the medium should be examined microscopically to be certain that they are dust free. The fiber material to be dissolved should not be marked and should have a maximum back ground count of 25 fibers per square centi meter of filter. The counting medium may be stored in a wide-mouthed Wheaton Bal sam bottle and should be applied with a glass rod. The counting medium has a re fractive index of 1.461 at 25 C. The me dium should be made in small quantities since it has a 6 month shelf life. Asbestos fibers are counted on wedge shaped sections cut from the filter. The wedge should be reasonably sized so that it can be mounted on a 25 x 75 mm (1x3 inch) glass microscope slide. The slide should be cleaned of dust before it is used. A drop or two of counting medium is placed on the slide and the filter wedge is placed dust side up on top of the medium. The wedge is covered, by .No. VA covenlip. Care must be taken while louring the coverslip to avoid trapping air under it. Air bub bles may be forced out by exerting slight pressure on the cover slip with a pencil eraser. Extreme care must be taken to avoid 003969 i i American Industrial Hygiene Association Journal S5 too much pressure, since this causes distor- * tion and stretching in the filter. The wedge is-usually cleared within 15 minutes; however, a residual background granularity may be noted. This will disap pear within a day. Counts must be com pleted within two days since there is a tendency for fiber migration and crystal growth. The prepared medium should be checked periodically for a background count. (See Counting and Calculations sections.) An alternate medium may be used when the samples must be counted immediately. It is prepared in the same manner as the preferred viscous medium, except that the added membrane filter is omitted. Since this medium is much less viscous it should be' stored in narrow-mouthed bottles. This me dium is applied with a glass dropper after the filter wedge has been placed dust side up on the slide. However, a drop should be placed on the slide adjacent to but not on top of each corner of the wedge. This me dium acts very rapidly and the No. I Vz coverglass should be lowered within a min- ute. Slides made up with this "alternate medium" are ready for counting within five minutes and the count must be completed within two hours after preparation. Filters which have been used to collect samples and which have been resealed in a holder should be removed only in a clean environment. They should not be held with the fingers. They may be held with tweezers and cut with either a scalpel or scissors.Bottles containing media must be tightly closed except during use. All bottles, tools used for handling and cutting filters, and containers used for storing tools must be thoroughly cleaned' and dried. Tweezers, scissors, scalpels, etc. should be set aside for this purpose 'exclusively. Although slides and coverslips arc purchased prc-clcaned, they should be wiped lightly with clean lens tisuc to remove traces of dust. Do not at tempt to reuse slides or coverslips. trast substage condenser, a 4 mm "'high dry" phase-contrast objective (40 X to 45 X) and a 10 X eyepiece is used to count the sample. It should have either a pro-focused built-in illuminator supplied with an iris diaphragm or be lighted by a separate bright illuminatorflat mirror combination. If a separate illumina'or is used it1 should be equipped with a condensing lens, iris diaphragm, and means for adjusting both the illuminator level and the distance between the condensing lens and the bulb. Zoom microscopes may be used for counting asbestos fibers, provided that the total magnification is within the 400-450 X ranem. The illuminator should be equipped with an adjustable constant voltage transformer. Both the microscope condenser and stage require close adjustment and both should be rack and pinion mounted. The stage must move along two perpendicular axes in a horizontal plane, while the condenser must move up and down in a vertical plane. Lighting must be adjusted for optimum bal ance. The Kohler method of illumination is recommended. (See Apper. Ji". I for pro cedures required to establish Kohler illu minator.) The counting field is defined by a reticle mounted at the level of the field-limiting di aphragm in the 10 X eyepiece. The Porton reticle (Figure 2) is recommended. However, any reticle which contains markings which, in addition to outlining an area of aboul 0.005 square millimeters, as determined by ooO'Ot I > 1O O' / * ta tt 12 II 1 i i ii i ii J L_ Microscope A microscope equipped with a phase con Figure 2. Porton reticle. 003970 I 11' 86 the combination of eyepiece and objective, and which defines linear distances that may serve as measuring aids, is satisfactory. The reticle must be calibrated prior to use and recalibrated each time it is removed from the microscope eyepiece or when the micro scope tube length or interpupillary distance is changed. Zoom microscopes must be re calibrated each time the zoom position is changed. (See Appendix II for recommended calibration procedures.) Counting A fiber is defined as a particle whose length is at least three times greater than its diameter. All fibers longer than 5 fim within the area delineated by the reticle, or which enter it from either of two adjacent sides, are counted. Fibers entering the area from either of the other two sides, i.e., those not arbitrarily chosen as "counting" sides are not counted, nor fibers in excess of 5 fim in diameter. Touching fibers, i.e., one of whose ends touch another fiber regardless of the resulting angle, are considered as one. Fibers that cross each other are counted in dividually. Fibers that pass through the de lineated area are counted provided that they cross at least one of the arbitrarily chosen "counting" sides. (See Figure 3.) Fiber length should be determined by measuring against standards such as a circle or a space defined by markings on the cali brated reticle. If the fiber is curved or wavy the total length should be estimated by meas uring along the curve. Once sufficient ex perience has been gained in judging length, only fibers whose length or diameter are in question need be measured. A routine for choosing counting fields must be selected so that fields are not count ed more than once and that a representative fraction of the total filter area is viewed. A convenient procedure whicl^may be used is to select a seriei of microscope viewing fields along a radial line extending from the apex of the filter to the outer edge of the sample wedge. The viewing fields must be February, 1975 chosen without preference to a particular area and should be approximately equally spaced along the line. The first field should be located a little bit in from the point at which matter is deposited on the filter. If the required number of viewing fields can not be located on a single radial line, addi tional lines parallel to the first should be chosen. Preferably 100 fibers should be counted but all the fibers in 20 fields should be counted even if there are more than 100 fibers. The counting may be terminated after 100 fields have been searched even if 100 fibers have not been counted. Calculations The concentration of asbestos fibers in air can be expressed as: Asbestos Cone. = Fibers x R Fields x Vol Figure 3. Illustration of fiber counting with a Porton reticle. The top and left side of the large box have been chosen as "counting sides". Circle 6 is 5 pm in diameter. (A) Fiber A crosses the top of the box and is counted, therefore. One end of Fiber At touches Fiber A and thus is considered as part of it and is not counted separately. .(B) Fi ber B is a bundle containing many spikes. It passes through the box crossing bottom and top. It is counted as one fiber. (C) Fiber C passes through the box but it crosses the right side and bottom. It is not counted. (D) Fiber D is entirely within the box and crosses over Fiber A. It is not part of Fiber A, therefore it is counted as a separate fiber. (E) Fiber E is less than 5 pm in length. It is not count ed... I 003971 1 l ! 1 \ American Industrial Hygiene Association Journal where: Fibers = total number of fibers counted Fields = total number of fields counted r_ filtration area area of a counting field Vol = sampling rate (1pm) x time (min) xllP = sample volume in ml of air Since both the filtration area on the mem brane filter and the area of the counting field are normally constant, the ratio of areas, R, is constant for the given conditions. To illustrate: Air is sampled at 1 liter/ minute for 50 minutes on a 37-mm mem brane filter (filtration area 855 mm2). A total of 60 fibers each longer than 5 fim are counted in 100 fields each 0.003 mm2 in area: R= 855 mm2 = 2.8 x I03 3 x 10_J mm2 and Asbestos 2.8 x 103 x 60 Fibers Cone. =- 1in0n0 fFielidis x-1---x--m-1m0:-3--m---l-x 5c0n mm. = 3.4 Fibers/ml Since the precision of this method is limited, the averaged results of the several samples should be reported to the nearest significant "figure. Appendix I Recommended Procedure for Establishing Kohler Illumination The microscope is set on a level surface at a convenient height and in such position that sighting through the eyepieces is possible without undue strain or discomfort. The illuminator is placed directly in front of and aligned with the microscope. If the illuminator is equipped with a coil filament bulb, its iris should be ten inches from the mirror. If a ribbon filament bulb is used, the front of the filter holder should be at least seven inches from the mirror. Only the plane surface of the mirror should be used. All filters should be removed from 87 the system except that if desired, neutral density filters may be used to reduce the light intensity. If the microscope is equipped with a built-in illuminator, proper illuminator positioning has already been accomplished and need not be of further concern. Fix a mounted sample on the stage and recheck the alignment. Raise the substage condenser until it nearly touches the bottom of the slice. Turn on the illuminator and use the tilt controls to direct the light to-the center of the mirror. Tilt the mirror, direct ing the beam upward into the stage con denser. Close the microscope substage iris and the illuminator iris. Use the illuminator focus control to focus the image of the fila ment on the bottom of the substage iris. The reflection of the iris may be viewed in the mirror by leaning over the microscope. After adjusting the illuminator condenser so that there is a sharp image of the filament on the substage iris, open it about halfway. Recheck proper position of the mirror and then put the objective in place and focus sharply on the sample. It may he necessary to open the substage iris during focusing. After the sample is in focus, fully close the iris. Readjust the condenser height so that the edge of the iris leaves are in sharp focus and there is a bright spot within the iris. Readjust the mirror to obtain maximum brightness. If the color around the iris is not uniform, recheck the illuminator tilt and focus. If the color is uniform and maximum brightness has been achieved, open the field iris until the blades just pass out of the field of view. The proper phase stop is inserted or ro tated into place in the substage condenser. The phase objective is rotated into place. The eyepiece is replaced with the phase ring centering telescope. The telescope is ad justed for sighting. The location of the phase ring is observed through the telescope. It may be adjusted by using the two rotating shafts provided with the phase condenser. When the ring is centered it appears as a bright annular ring mounted on top of a 003972 ) 88 grey ring. The center is black. There should be no overlap of the bright ring into the center. When the phase stop is centered the telescope is removed and the eyepiece is re placed. The microscope is now ready for use. Appendix II Recommended Procedures for Calibrating Reticle The reticle must be slightly smaller in diameter than the eyepiece in which it is to be used. It fits into the tube and is held in place at the eyepiece focal point by spring pressure. The focal point is marked by an interna! collar which prevents the reticle from passing further into the tube. Some manufacturers locate the focal point just below the collar while others prefer a point immediately above" it. Thus it may be nec essary to insert the reticle between the lens and the collar on one microscope and below the collar on another. In either case the projected image of the reticle on the field must be clear and sharp. The reticle must project a constant count ing area as well as provide markings for sizing. Many types of reticles are satisfac tory. The Porton reticle illustrated in Figure 2 is recommended. It outlines a large rec tangle that fits easily within the field of vision. The rectangle is divided into two equal squares. The left square is further divided into six rectangles. Each side of each square has a length equal to 100 L, then the large rectangle is 200 L units long and the small ones SOL units long and 33VS L units high. To the left of the left square is a scale which divides the side into 20 equal increments of 5 L. units each. The right square is divided horizontally in thirds corresponding to th small rectangles and vertically into a scries of increasing spaces. The distance between each vertical line and the center of the large rectangle corresponds to the diameter of a circle. Circles of sizes 1 through 9 are located above and below the large rectangle. The diameter of each circle February, 1975 is related in terms of L units of length in accordance with the formula: D -- L vW where D is the diameter of the circle or the width of the space and N is the number of the circle, or space. The unit L is determined by measurement. A stage micrometer, is the primary ruler. Any one or more of the definitive linear di mensions may be measured. If the length of the large rectangle is measured then L equals 1/200th of that value. If the stage micrometer is marked in hun dredths of millimeters the value of L is in hundredths of millimeters. If the left square is used as a counting area, its area is equal to (100x0.01 L)2mm2 The filtering area of a 37-mm filter is 855 mm2 and thus R - 855 R " L2 M If other areas are used as counting fields or other sized filters are used, the value of R must be calculated for the individual con dition. It is wise to prepare a chart showing each linear dimension and area for ready reference while counting. Appendix III Pump Calibration Personal sampling pumps should be cali brated for flowrate frequently. They may be calibrated against a primary meter such as a spirometer or against a previously calibrat ed secondary standard such as a wet gas meter or dry gas meter. The train should be arranged so that the pump is operating against the same resistance it would normally operate against, and that the calibrating meter is at atmospheric pressure. Figure 4 illustrates the recommended arrangement. Air flow through the filter may be limited by both restricting the air flow to the pump as well as by admitting secondary air into v ( 003973 I i \ American Industrial Hygiene Association Journal 89 Type AA filters can be used in any di ameter in any non-leaking open-type filter `holder, as long as proper adjustment is made in the calculations for the change in filtering area, and periodic background checks arc made. Membrane filters of other types and man | ufacturers cannot be recommended at this i i time, either because they do not become lets* Figure 4. Diagram of set-up for calibration of optically clear in the mounting medium or because of excessive particle background. pump. Personal Battery Powered Sampling Pumps I the pump. Most personal pumps are con (1) Casella1' MK II or MK III Personal F structed with adjustment valves that permit Sampler both actions. Calibration should be repeated (2) MSAC Monitaire Sampler, Model G. after each 24 to 36 hours of pump use. If (3) Bendixd VM 22 or Micronair Per the pump has not been used for a prolonged sonal Sampler or C-l 15. period of time it should be recalibrated. Microscopes: Phase Equipped I Temperature and altitude corrections should be made. (a) microscope body with a fine focus accuracy of 0.006 mm; Personal sampling pumps should not be (b) 10 X eyepiece; used to such an extent that the batteries are (c) mechanical stage; allowed to run down below normal operating (d) illuminator (preferably built in and voltage. Since battery life varies, it is wise having provisions for adjusting light to occasionally check the life of the battery intensity); by allowing the pump to run down to the (e) 40 X to 45 X (0.65 N.A. at least) first indicated change in voltage under timed positive (bright field) phase-contrast simulated conditions. objective; Supplies Filters and Filter Holders Millipore Corp.* Field Monitor filter hold ers preloaded with 37-mm Type AA filters, either white plain (catalog #MAWP037A0) or gridded (#MAWG037AO) are recom mended. These are sold with a guaranteed average particle background and can be used without additional background checks. It is less expensive to purchase unloaded field monitors (#MABG037A0) to be loaded with plain or gridded filters pur chased separately in packages of 100 (#AA\VP03700rand AAWG03700). The field monitor can then be reloaded after each use. However, it is necessary to care (f) annual ring condenser diaphragm (corresponding to the objective); and (g) phase ring centering telescope. (Note: Most manufacturers sell a basic body unit and built-in illumination system as a unit. Phase-contrast accessories can us ually be purchased as a kit consisting of 10 X, 40 X and 90 X phase objectives, a phase condenser containing appropriate an nular ring diaphragms, a phase ring center ing telescope, and a green filter. It is to the microscopist's advantage to purchase the kit.) . A list of manufacturers of phase-contrast microscopes" and reticles' is given below to aid in selecting a proper instrument. fully clean the empty monitor and to check the background count on filters in the re loaded monitor. * Millipore Corp., Bedford. Mass. 07130. * Willson* Products D*.. P.O. Box 622. Reading. Penna. 19603. C. F. Casella & Co. Ltd.. Repent House. Britannia 003974 t 90 Walk, London, N.l, England. c Mine Safety Appliance Co., 20t Braddock Ave., Pitts* burgh, Penna. 15208. Bendix--NEI, P.O. Box 590, Fall River, Mass. 02722. * Bausch and Lomb, Scientific Instrument Division. 72624 Bausch Street, Rochester, N.Y. 14602. Olympus Micros opes. Micro Optics Company, 28165 Greenfield, Southfield, Mich. 48075. American Optical Corporation, Reichert Products, Buf falo, N.Y. 14215. February, 1975 E. Leiti Inc., Rockkigh, NJ. 07647. Nikon Inc., Instrument Division, Garden City, N.Y. 11530. Unitron Instrument Company, Microscope Sales Divi sion, 66 Needham Street, Newton Highland; Mass. 02161. Carl Zeiss, 444 Fifth Avenue, New York, N.Y, 10018. 2 Edmund Scientific Co., 701 Edsorp Building, Barring ton, N.J. 07007. BGI, Incorporated, 5ft Guinan St., Waltham, Mass. 02154. ; j ., 4' Joint A1HA-ACG1H Aerosol Hazards Evaluation Committee AIHA ACGIH Morton Lippmann, Ph.D., Chairman J. LeRoy Balzer, Ph.D. Douglas K. Craig, Ph.D. Graham W. Gibbs, Ph.D. William C. Janes William H. Krebs, Ph.D. Carl A. Mangold Eric B. Sansone, Ph.D. Marvin Tillery Thomas F. Tomb Russell W. VanHouten Wesley R. VanPelt, Ph.D. Donald L. Webster Howard E. Ayer George Carson, Ph.D. Harry J. Ettinger Murray Jacobson Geoffrey Knight Jeremiah R. Lynch G. Major Owen Moss Milton Scheinbaum Gien W. Sutton '---------------- :------------------------------------- Industrial Hygiene Training in Israel Medical and engineering students at Tel Aviv University will be ob ligated in the future to undertake training in occupational safety, hygiene and health to inspire them with an awareness of their obligations for the maintenance of the health of the worker and for the prevention of ill effects from work on man. This has been made possible by the establishment in 1974 by the Faculties of Medicine and Engineering of the University, in collaboration with the Ministry of Labor, of the National Insurance Institute and the Workers Health Insurance (Kupat-Holim) of a Center for Occupational Safety, Hygiene and Health, with four major departments--Physiology of Work and Rehabilitation, Occupational Toxicology, Occupational Hygiene, and-Safety Engineering. It comes to fill long-standing gaps in the promotion of safety and health of the working population in Israel and the absence of training facilities for specialists in these fields. The new Center will immediately provide refresher courses for physicians and nurses, but curricula for the training of Occupational Toxicologists, . Hygienists and Physiologists are in preparation. .Ultimately the Occupa tional Health Physician will also be trained at the Center. Due to lack of suitable personnel in the country the Center is seeking to recruit specialists from abroad. 003975