LABORATORY NOTE:
A STUDY IN LABORATORY AUTOMATION
Analysis of PCBs in trans
former oils
8y $.1. Lormon, H. Gordon, and J.P. Hendricks
OLYCHLOftfNATSO biphenyls ling thousands of sample* per year
P(PCBs) have been commercial* had to be developed.
ly manufactured since 1929.
The system that was finally devel
They were, until recently, wideloyped has been in use for more than
used throughout industry because of three yean. More than 1000 samples
their chemical and thermal stability, have been assayed with it. The three-
low or nonflammability, and good part method consists of gas chroma
electrical insulating properties. But tography. sample preparation, and
the very properties that have made automation, both instrumental and
PCBs so useful have resulted in a ban data reduction. These were inte
on their manufacture and strict reg grated into one efficient system for
ulation of their disposal by the Envi speedy handling of many samples
ronmental Protection Agency and rapid analysis of data.
(EPA). PCBs are so chemically inert
that they persist in nature long after many other organic chemicals have
Sample tracking
broken down. Most of the biosphere
Sample preparation consists of
Is contaminnted, although no natural four discrete tasks, allowing fa as
source of PCBs exists. Not only are sembly-line work and minimum crit
PCBs environmental pollutants, but ical path planning. The first task was
they are suspected carcinogens as to log in an identification number for
well.
each sample. Next, each sample was
When the EPA announced the reg prepared and its weight entered into
ulations on disposal of PCBs In the log. The sample was then loaded
1971, the technical personnel at Con into the autosnmpler for injection
solidated Edison were faced with a into the gas chromatograph. The last
major problem. It was necessary to task was to eater the sample's i.d.
determine which of the 23,000 trans and weight into the computer. Two
formers in the Consolidated Edison technician-days were required for 50
systems contained PCB-comami- samples, at a rate of about three
nated transformer oil. and at what samples per technician hour.
levels. The oils were almost certainly
contaminated because at one time fuinollititsd Edison had used PCBs
Oas chromatography
as transformer oil. When the switch
Two Perkia-Elmer Sigma stria
was made, the new oil introduced be gas chromatographs were used. Each
came contaminated by PCB residua. was equipped with a Perkia-Elmer
This meant that method for hand- autosampler. The Sigma 3 gat chro
matograph used is a single-channel
Mr. Leman It Senior Stuff Chemist, Mr. Cordon Is Croup Lender of the Special
unit, equipped with an NlM electron capture detector and capable of tem
Analysis Division, and Mr. Hendricks is
perature programming. The Sigma 4
Senior Technician ofthe Specie/ Analysis Division, ConeoUdated Edison Company of Now York. The authors wish to ac knowledge Meryl Mendhfor drafting this
ankle. This paper is adaptedfrom a prosentetfon made at the S7nd Pittsburgh Conference on Analytical Chemistry and Applied Spoctraecopy, Atlantic City, New ferny, March mi.
is a dual channel, isothermal GC. The chromatography was carried
out under the following conditions:
column packing, Chrpmosorb W
HP; column toad, 4% OV-22S; mesh sin, 60-00; column material, glass; column sise: 1.1 m x 4 nun; carrier gas, argon: methane, 95:5; carrier
flow. 30 mL/min; oven temperature, 225 *C; auxiliary temperature, 373 *C; standing current of etactran capture detector. 0.5.
No data reduction wee pafprmed on the gas chromatographs:Thtout put signals were tai'-dkmQt hno e computer.
Sample cleanup and peaparution
Rapid analysis was repaired for the large numbers of templet. For the first analyea, sample cleanup was performed on a packed Plororil column. But, because using Florastl packing is a very laborious proce dure, at this time only dirty or high samples (greater than 500 ppm of PCB) or samptas giving poor chro matograms were cleaned. This proce dure war changed for later PCB aampta. Ail inter analytes were paformed using Sep-Paks (Waters Associates) which are FlorocB units in small disposable packets, This eliminated the need for cohuna prep aration.
The preparation procadnre was a follows: 100 jiL of sample were dis pensed on e previously weighed ScpPak. The loaded Sep-Pak wee than reweighed. Next the sample was diued with 25 mL of hexane, using a syringe, directly into a 25-mL vol umetric flask.
Tablet H ftnnovnry IraniBgpPtii
ml Heatawe
6 10 18 20 20
%h aBgMl
m
4.0 0.7 0.3 0.2
TM 034 10.7 B4 10.7 90.0
^
176 : FEBRUARY 1962
MGNS 003138 I
21,OM
14 Si
Pttrarch for
organoaillcon
pscialtist
I organoslIIcon 1 loompounds --| ' now Rogiotor [
A Rovlow providos data on ovor 1000
products
Edition S-S ol the Potrarch Organoaillcon Raglatar A Review hat
|uat baan publithad. Thit uaaful documant
pratants structural, physical and
toxicological data on ovar 1000 compounds. An absolute neeeaaity for the research chemist
-- and equally indispensable for the
technologist and engineer engaged in matarlals development, eurfaoe modification, compoeite structures, rheological studiee and other applications. Seven authoritative chapters discuss the uea of organotllanaa in tyntheaia, tailoring surfaces and other
topics of current Interact. For your free
copy of thte uaaful catalog, contact Petrarch today.
i Flpurv 1 PCB tarnp/a before ofeemrp.
pctrarch nam MC. Bertram Road
Bristol. PA 19007 215: 781-9255
in tits Far East: Chtaao Corp., Tokyo
Chela Bsadai Bandas Card Net 97 17B : FEBRUARY 1982
_ 2 Samp* la Fpur* 1 after cleanup. Mod paaSa and We oftromafopranTf are Ifle him, Ii PMIarlal from th cfmn-ufi procedure.
mare la no toe* ol *19 no Mea of PCS
The recovery was found to be ex cellent (rM* /). A sample known to be clean gave tha same chromato gram bosh before and after deenup.
It wu concluded, therefore, that no PCBe were loet. A known dirty sam ple save a well-reeolved ebromatogram aher dsonup (Ffgtrcw 1 ~4j.
Stmdtrds
Originally the transformer oils were identified by trade asms, nab* as standards Askasal, Insrteea. and Pyranol, which era dUfaeeac sUaturea of PCBa km all centals sham 30*h trlchlorobmeane. Later, far
HONS 003139 I
Figure 3 A very poor cnromttogram batora tampla ciaanup.
Figure 4 Sampta f/t Figure 9 attar efaanup. Nota tha hfgtify Improvad raaoludon, and much cfaanar ohromatogram at a raault of tfia ciaanup procedure.
Now, improved Mark III ZJppatts diepaneare are highly roaletant to breakage. Tha boroaiiieata giaaa syringe ia ancaaad In protective
polypropylene. Oeltvery velvet ere positioned within the reservoir for added sefety. The delivery tip ie secured by a rigid platform to eliminate dislodging by accident. Dispensing range fully adluatebie throughout syringe capacity. Five sizes available: 0-0-5 ml, 0-9 mi, 0-10 ml, 0-30 ml, 0-50 ml. Fully
autociavabie without disassembly. Available with low profile, sntMtp, safety reservoir of amber glees, or one gallon amber plastic coaled bottle for the 0-30 ml and 040 ml
sizes. Screw threaded adapters for use with common laboratory con tainers are available.
consistency in reporting, the Aro*
don 1242, 1254, and 1240 ware used. Recently Arodort 1016 end 1221 were added. Now commercial Arodor standard# of 1000 ppm in isoctane art used.
The standard# were prepared by diluting to 2 ppm PCS in hexane.
This is equivalent to taking 0.1 g of oil containing 500 ppm PCB, and di luting to 25 tnL with hexane. Five mkroUters of the standard, contain ing 20 ng PCB, were injected into the
chromatograph. Standardization was performed, and the standard usually hcM with very little drift.
For complete information, sand for the Wheaton Zlppette brochure. Tel: 900425-1400, extension 2232.
WHEATON
INSTRUMENTS
1301 N. 10th at.. MMt Nw Jnq 00338
ciwi,
nmcmifcn
AMERICAN LABORATORY : 17*
RONS 003140
FCBe continue*
Datt reduction
An HP 3354 LAS computer was used for data reduction. The com puter it a real-time on-line system with instrument and autosampler control and It compatible with any lvolt analog instrument output. In (hit cam, it received data directly Tram the gae chromatographs.
Hie computer was programmable in BASIC, allowing for individual data analysis. The system-level lan guage was a user-oriented conversa tional language for peak-integration analysis.
Calculation and Arocior selection were done as follows. The computer searched for 5 key standard peaks for each Arocior of interest. It also calculated a concentration for each Arocior based on total peak area. Each standard peak was calibrated at total concentration, ail 3 peaks averaged, and a relative standard de viation (BSD) calculated. An error factor was calculated using both the BSD and the total-area value. The Arocior with the lowest error factor was choeen. The computer then looked at the % BSD. If it was below 40%, the mean value was reported. If It was above 40%, the total-area value was reported (figuns 3 and 6, reOfrJ).
There are several reasons why this software is so significant. The pro-
0
cgr rine .17*44 t*.74>>3
apoclop u<0
0
PT T[M 1.5
ftf '5CL0S i:e
2
pph :o.*n7 mean
-t?.o?r:
flE"ii
eppop
pet U"E 1,4*63? 2.*47; 2.60l?> 3.13204 4. 7'365
2
pen 5>> ?*><9 T-XJ 3*0. 500
AP0CI.OP
ME All
I500
P1 ET TINE ai;e4ociivP
?*> 5m0e0an
EPPOP ;r: .'#E 0
0
PE f Tirrc :-.i?204 4,r;.5 r. 04-544 *.17044 10.1*443
APOO.0P 1254
2
pcm
1:45.4
51.0,:*5 44.4445 23.3112
MEAH 2*7.514
PET Tint 1.3
AP0CL0P 1234
PPM 2i:.-';5
MEAN 213.- 33
hame *12**-l i2<0-2
peak-: >
l24'-jl 1242-4 1242-5
PEA*?
5
n1am2e42-r PEAk 3
i-*m| *1254-1 41234-2 *1254-3 11234-4 1254-3 PEAK? 3
1254-T PEAk i
. #50
4.47';4C- Pj0
. P'itl 33.044
Tablet AralIw alfrtlhm
Areatar 1MB
All peaks? Vaa
naan 37J
%mo
10M
Total man. V M*.2 area
*ohm
si.a
(% HB0) x 4638
IBM Yba SU 7M m2 11.4
0.1 rag
1MB Yea 36.4 28.4 8&4 28
aa.i M3
BmaptaieAroaiorim ijec mean vetue
., (Mena. - ip* anay(mmw *
seeai.
'
.
total
EPP0P 3C0*E 24.34U
for 1949, aa expected, gram is not complex. It is merely simple statistical process written in BASIC. The computer program wee United to identifying rocognlxabie Arocior chromatograms, t.g., trans former oils. The data were still re viewed by a technician, who used the printouts to write reports and trou bleshoot the instrumentation.
The results of all analyses were stored in e data file, and specific runs (standards and pertinent runs) could be stored la tbrir entirety. Numerous utility programs were also available.
For example, files could be mwiipulated for system operatloa aad tr-
When this project began, PCBs were classified into two ripnfleam concentration level*. Samples wbh
poenl, stmpt for milsnnmnel m entry. Any eamplm greninr than SB ppm wart dmriftad ee PCImaterial*
1B0 : PUmjARY 19B2
MOMS 003141
t
KIT TIWI
AROCLOR 1261
9 RET TINE l.S
AROCLOR i2se
2
PPM
hern
e
2 PPM 138.431
MEAN 139.431
NAME PEAKS e
NAME 1260-T PEAKS l
RSD 9
< RSD tee
ERROR SCORE 9
0
RET TIME 1.3430* 2.K024
3.39462 5.05479
1242
0 RET TIME l
AROCLOR 1242
2
PPM 15.3950 9.99*14 14.3320 13.9122 46.2421
MEAN 19.7176
2
PPM 193.;
MEAN 103.
ERROR SCORE 54.9765
NAME 1242-1 1242-2 1242-3 1242-4 1242-5
PEAKS 5
NAME 1242-T
PEAKS 1
\ RSD 69.2299
>, RSD 80.5755
REAGENTGRADE
SOLVENTS
0
RET TIME 3.33462 5.95479 9.93997 19.6215
AROCLOR 1254
0
RET TIME 1.5
AROCLOR 1254
2
PPM 36.5069 49.1013 64.9035 59.1655
MEAN 47.9143
2
PPM 147.577
MEAN 147.977
ERROR SCORE 11.6913
NAME 41254-1 1254-3 1254-4 1254-8
4 PEAKS 4
NAME 1254-T
PEAKS 1
:. RSD 22.9693
'< RSD 51.9997
Ptfure t Printout foe sompto.
oils with over SO ppm PCS, a third
class of concsntratfcm had to bs oh ktlBt.^1 tMiaa. Mfbk MUMMtpa.
done greater than 500 ppm were re ferred to as PCB material. Samples that were between SO and S00 ppm are now labeled PCO-contamlnaied.
In terms of disposal, this dassl-
flcatlon of the tamplee meant that noo-PCB material was dlspoeedof as
required severs rsstrictions.
Conclusion
Analytical systems designed for very large numbers of temples must be carefully designed. AH aspects of analysis, sample preparation, Instru* meat automation, and data reduc tion must be considered, both sep arately and as a system. Even where
Users oI Corco chromatographic, nine-
tronic and spoctrophomatrie reagent-
grade solvents Know they can count on
Corco for highest quality. Reliability, And
delivery. In pints, gallons, frvo-gaaon*. or
drums.
At Corco, we've bean specialists in
reagent-grade solvents since 1969-
end have bmlt our reputation on meeting
the requirements and speeifteabont of
our customers. ACS and ASTM
Corco can also provide reagent-
grade acids, bases, and specialty chem
icals.
Write or call regarding your specific
solvent requirements and our complete
solvent listing in Bulletin 10. Or circle the
number.
'
rial required some restrictions regardlag disposal, and PCB material
sary, the savings in labor alone make such a system profitable.
Ctwile Eseder tarries CardWe. tie
HONS 003142
AMERICAN LABORATORY : IB*
r