Document KGz5eoB4QJw3bz3ooZQr6am1r
Tentative Laboratory Performance Specifications Testing Protocol
and Evaluation Criteria for Passive Samplers
3H 001469
Purpose The purpose of this work is to define performance specifications and to recommend testing protocols for passive samplers used in determining personal time weighted average (TWA) exposures. Additionally, evaluation criteria for passive samplers and their associated analytical finish have been developed.
Background Technical information concerning capabilities and limitations is supplied to the user with some_-p arsine, samplers, while other samplers are furnished with little or no technical information. Even with samplers for which technical information is made available, it is often difficult to determine which sampler is best for a given application because different types of experiments may have been conducted in evaluating the different devices, in some cases, information necessary to conduct successful field surveys is not supplied by the manufacturer (e.g,, monitor capacity in multicomponent or humid atmospheres). The intent in developing and making available the attached performance specifications is: for passive monitor suppliers..to realize the parameters and the ranges of these parameters over which their sampling devices must operate in order to be acceptable sampling media; to make potential users of passive monitors aware of the critical parameters and limitations associated with these devices; and to alert testing and evaluating laboratories to the necessary areas of consideration for passive monitors. The attached testing protocol provides a recommended experimental
i
3M 001470
scheme by which suppliers, independent laboratories, and users may evaluate passive monitors to determine their performance characteristics with respect to standard evaluation criteria. The attached evaluation criteria are an adaptation of the sampling and analytical methods validation criteria developed in 1974 for the joint NIOSH/OSHA Standards Completion Program (i). These evaluation criteria are pertinent to passive sampling, unlike the previous criteria which were primarily intended for active (pumped) sampling systems, and did not address some of the factors significant to diffusional or permeation sampling (1,2). Since the two sampling systems, passive and active, function by different mechanisms and therefore have different capabilities and limitations, different performance specifications and evaluation criteria are required.
Passive Samplers Passive samplers rely on molecular diffusion across a quiescent zone or permeation through a membrane as the means of mass transport of the analyte to the sorbent. Personal passive monitors utilizing each of these mechanisms for the rate limiting diffusive step were reported in the literature in 1973. Palmes and Gunnison (3) devised an open-end tube with a quiescent zone across which the analyte diffused to the sorbent, while Reiszner and West (4) employed a permeable membrane into which the analyte must dissolve, diffuse through, and then out of the other side in order to reach the sorbent. Passive monitors currently on the market employ one of these two principles, diffusion across a quiescent zone or permeation through a membrane, as the rate limiting step.
3
3M 001471
The mathematics initially used to describe samplers using diffusion across a quiescent zone were limited to the integrated form of Pick's First Law of Diffusion (3):
0 - (DA/l)(C0- C)
(1)
Where:
J * diffusive flux D diffusion coefficient of the analyte in air A * cross-sectional area of the diffusion zone L * length of the diffusion zone CQ = ambient concentration of the analyte C = analyte concentration at the sorbent surface
If the analyte concentration at the sorbent surface (C) is assumed to be
zero (since the sorbent captures the analyte) and all of the analyte that
enters the monitor is collected (J * quantity collected, Q) for some >
exposure time (t), then the first equation conveniently reduces to:
Q - (DA/L) C0t
(2)
The DA/L term is the sampling rate and can be calculated from the diffusion constant of the analyte in air (D) and the geometry of the monitor (A and L). While this term does define tjig sampling rate, it is probable that
A
3M 001472
draft shields or other mechanisms may inhibit the mass flow somewhat and make the calculated sampling rate only an approximation. Also, the diffusion coefficient is typically^" pxpprim&r^.'y^tleternii.ned value and as such has some error associated with it and therefore may cause the calculated sampling rate to be different from the actual sampling rate. For this reason, sampling rates for passive monitors must be verified in the laboratory.
Additionally, an assumption was made in using Pick's First taw (C=Q) that may not necessarily be true. For solid sorbents, the analyte concentration at the sorbent face during sampling will actually be some equilibrium vapor concentration (Cg) of the analyte above the sorbent (i.e. C= C instead of zero). After substituting C for C and rearranging, equation 1 becomes:
J = (DA/L) C0 - (DA/l) Ce
(3)
Consequently, when the equilibrium vapor concentration (Cg) of the analyte
at the sorbent surface is very low, i.e. the sorbent has a very high
affinity for the analyte, the Cg term drops out and equation 3 can be made
equal to equation 2. The reverse diffusion term [-(DA/L)Cg] becomes
significant for a weakly retained species particularly for the following
cases: when the analyte is competing for sorbent sites (e.g. coadsorption of
other species, including water vapor}j whpn a high
^pn^irp is followed
for a significant sampling time bv no.or-very_.lnw analytf exposure: ana when
5
3M 001473
long ston^ge tildes occur .between exposure and analysis (5, 6). Laboratory verification is necessary to confirm that a passive monitor does not operate under these "reverse diffusion" conditions. For passive monitors incorporating irreversible sorption (the analyte is reacted with the sorbent) the sorbent acts as an efficient collector of the analyte and equation 2 instead of equation 3 applies.
As was already mentioned, passive monitoring may also be accomplished using a permeable membrane as the rate limiting step (4, 7). The flow of a vapor through a membrane can be expressed as:
J - DA (CQ - Cx)/S
(4)
Where:
J diffusive flux D solubility of the analyte in the membrane A = cross-sectional area of the membrane Cq = concentration of the analyte on the exposed side
of the membrane concentration of the analyte on the sorbent side
of the membrane S = membrane thickness
If the sorbent is an efficient collector of the analyte then 0 and Cq is due to the atmospheric concentration of the analyte.
6
3H OOI474
Consequently, the quantity of analyte collected (Q) over an exposure time (t) is:
Q = kCQt
(5)
where k is a constant (sampling rate) dependent on properties of the analyte and membrane and the dimensions of the membrane (4, 7).
Note that the sampling rate term of equation 2, providing the length of the diffusion zone (l ) does not change during sampling, can be set equal to some constant (k) and the resulting expression is the same as that describing permeation sampling (equation b). Thus for empirically determined sampling rates, equation 5 may be used for either permeation or quiescent zone diffusion controlled passive samplers.
There are several factors that affect the sampling rate, and subsequently the precision and accuracy of determinations made following diffusional or permeation sampling. Additionally, the nature of the sorbent and the analytical finish used will add some imprecision to the sampling and analysis regimen. Those factors that affect the amount of analyte collected and consequently the determined air concentration will be discussed in the following sections.
7
3M 001475
General Considerations Several factors will affect the performance of passive samplers. These include; the sampling rate and its related parameters (linearity, response time, and reverse diffusion), ambient conditions during sampling (face velocity, temperature, humidity, and pressure), and direct sampling interferences (coadsorption, etc.). There are also several factors which must be known for all sampling and analytical methods regardless of whether diffusion, permeation, or a mechanical pump is used to effect the analyte/sorbent contact. These include the analytical range (as determined from the sensitivity and capacity), precision, accuracy, analytical interferences, and stability (including shelf-life before sampling and stability of the analyte on the sorbent after sampling).
The performance specifications, recommended experimental protocol and evaluation criteria are summarized in the following table. Following the table, the individual specifications, experiments, and criteria are discussed.
8
3M 001476
Performance Specification Experimental Protocol
Evaluation Criterion
1. Analytical Recovery For monitors utilizing Spike and analyse moni desorption, analytical tors at each of 4 con recoveries should be centration levels (0.1, complete and precise 0.5, 1.0 & 2.0 X PEL )*
The means of the recovery for each of the 3 higher levels must be >75% and the CV^ must be ^0.1**
2. Sampling Rate and Capacity
Sampling rate should Expose monitors for vary- The sampling rate (slope
be empirically veri- ing times (80%R.H.,
ofthe plot) must be
fied and the maximum 100 fpm, 2 X PEl. ) prior constant over the MUST.
recommended sampling to analysis. Plot quan-
time (MR ST) determined, tity determined vs. time
exposed.
3. Stability During Exposure
Reverse diffusion should not be signi ficant
Expose monitors at 80% The different between the
R.H., 100 fpm and 2 XPEL means of the two sets of
for 50% of MRST, re
monitors must not exceed
move and analyze some , 10% at the 951 confi
monitors, analyze others dence level
after exposure at 80%
R.H., 100 fpm and zero
analyte for remainder of
MRST.
3M 001477
9
Performance Specifications Experimental Protocol
Evaluation Criteria
4. Storage Stability Monitors should be stable for a reason able amount of time between exposure and analysis
Expose 3 sets of moni The difference between
tors at 80% R.H.,
the means of monitors
100 fpm, IXPEl and IX analyzed on days 1 and
MRST and analyze one 14 or days 1 and 15 must
set on day after ex not exceed 10% at the
posure, one set after 95% confidence level
14 days storage at
room temperature and
one set after 15 days
refrigerated storage
5. Performance Factors
The following factors Use 16 run fractional Any factor that causes a
(over the ranges shown) factorial to determine statistically significant
should not affect the significant factors
difference in the re
quantity of analyte
and interactions
covery at the 95% con
determined
fidence level must be
defined.
10
3H 001478
Performance Specifications Experimental Protocol
Evaluation Criteria
F actor concentration exposure time face velocity humidity interferrent orientation
Range 0.1-2XPEL 0.1-1XMRST 20-300fpm 10-80XRH 0-lXPEl parallel-perpendicular (to air flow in chamber)
Temperature Effects
Variation in quantity Expose monitors to the The temperature affect
determined with expos- same analyte concen- must be defined and
ure temperature should tration and at 10, 25 correction factors pro
be defined.
and 40*C while main- vided if needed
taining a constant
(+5% absolute humi
dity.)
Precision and
Accuracy
Monitors should be pre- Expose monitors at cise and accurate over 80XR.H., 100 fpm the intended sampling and 1 X MRST at 0.1,
CV-j. must be ac ceptable.
ranges.
O.b, 1.0 and 2.0 X PEl.
Analyze monitors.
11
3H 001479
2 X PEL = two times the Occupational Safety and Health, Permissable Exposure Limit. See Appendix A - Statistical Methodology
12
3H 001480
Analytical Recovery
a. Performance Specification: For monitors utilizing desorption, analytical recoveries should be complete and precise.
b. Experimentation: Spike at least six monitors at each of four concentration levels (0.1, 0.5, 1.0 and 2.0 X PEL). Allow at least twelve hours for equilibration then desorb and analyze the monitors.
c. Evaluation Criterion: The means of the recoveries for 0.5, 1.0 and
2.0 X PEL must not be statistically different (at the 9b%
confidence level), must be greater than 75% and the
must be
0.1.
d. Before investigating the factors affecting analyte collection, it is necessary to verify that the analytical methodology used is accurately determining the quantity of material collected. A means for accomplishing this is by spiking the sorbent material with a suitable solvent containing dissolved quantities of analyte representative of exposures at 0.1, 0.5, 1.0, and 2.0 X PEL, allowing the analyte and sorbent to equilibrate, desorbing and analyzing the sample. Spiking is normally done by pipetting an analyte containing solvent (preferably <10 uL ) dicec.Uy_.onto the sorbent. This method of spiking has been shown to De the most
13
3H 001481
comparable to dynamic--s-amp'l'ing (8). In calculating the quantity ot analyte necessary for spiking, a sampling time of 8 hours for TWA measurements or 15 minutes for ceiling exposure measurements should be used. If previous experimentation has shown that an eight hour exposure time is unrealistic because of monitor capacity, this lime must be adjusted accordingly but should not be less than one hour for TWA measurements. The sampling rate used in calculating the quantity to spike should be known from previous experimentation or furnished by the monitor manufacturer.
Sampling Rate and Capacity
a. Performance Specification: The sampling rate should be empirically verified and the maximum recommended sampling time (MRST) determined.
b. Experimentation: Expose monitors for varying times (eg. 1, 2, 4, 6, 8, 10 and 12 hours for TWA samplers and 5, 10, 15, 20, 25 and 30 minutes for ceiling samplers) at 80% R.H., 100 fpm face velocity and 2 X PEL. Analyze at least two monitors for each time period. Plot the quantity of analyte determined versus the time exposed. The capacity (at 2 X PEL and 80% R.H.) is defined as the point at which a 5% deviation in linearity (using goodness of fit statistics) is determined. The MRST is defined as 0.67 times the capacity for single substance monitors and 0.33 times the capacity for multiple substance monitors. 14
3H 001482
Evaluation Criterion: The sampling rate (slope of the above described plot) must be constant over the MRST. Also, the capacity in units of time at 80t R.H. and 2 X PEL must be defined.
Although sampling rates may be calculated as previously discussed, it is necessary to verify the actual sampling rate in the laboratory. This laboratory verification is necessary for several reasons; the calculated sampling rate has some error associated with it (particularly in the diffusion coefficient), some assumptions were made in developing the sampling rate equation which may not always be true, the use of draft shields or other mechanisms to prevent face velocity effects may alter tne sampling rate somewhai and the diffusion coefficient or solubility in the permeable membrane may not be known.
The actual analyte capacity of the monitor may depend on the concentration of analyte in the atmosphere being sampled as well as the presence of other gases or vapors that exhibit an affinity for the sorbent. A concentration level of 2 X PEL was chosen since the monitor capacity is generally lower for high exposure concentrations than low exposure concentrations. Consequently, the capacity at low exposure concentrations will be at least (and may actually be greater than) the capacity determined at 2 X PEL. Since relative humidity levels may vary considerably between exposure sites and the
lb
3M 001483
water vapor may compete with the sorbent for sorbent sites, the "worst case" (i.e. high humidity level) was chosen for investigation. Also, since any of several different substances may compete with the analyte for sorbent sites and reduce the actual capacity, a safety factor was included. This is applied by defining the MRST as two-thirds of the capacity. For multiple substance monitors a safety factor of two-thirds may not be sufficient (9), consequently a factor of one-third is used. The MRST is defined in units of time (hours or minutes) for an exposure concentration of 2 X PEI and a relative humidity level of 80%. In several instances the actual monitor capacity will likely be greater than that defined in this manner however, an MRST as described will give the industrial hygienist a valuable, easy to work with reference.
Stability During Exposure:
a. Performance Specification: Reverse diffusion should not be significant.
b. Experimentation: Expose at least 20 monitors at 2 X PEL, 80% R.H., and 100 fpm for 50% of the MRST. Remove and analyze 50% of the monitors and expose the others for the remainder of the MRST (no analyte but at 80% R.H., and 100 fpm). Remove and analyze these remaining monitors.
16
3M 001484
c. Evaluation Criterion: The difference between the means of the two sets of monitors must not exceed 10% at the 95% confidence level.
d. As was previously discussed, reverse diffusion may be significant for weakly retained species. This experiment provides a means of checking to determine if reverse diffusion is occurring for a particular sampling system.
Storage Stability:
a. Performance Specification: Monitors should be stable for a reasonable time period between exposure and analysis.
t>. Experimentation: Expose three sets of at least 10 monitors per set at 80%R.H., 100 fpm, 1 X PEl and 1 X MRST. Analyze one set of monitors on the day following exposure, one set after 14 days storage at room temperature (-25 *C) and one set after 14 days refrigerated storage (-5 *C).
c. Evaluation Criterion: The difference betwen the means of monitors analyzed within one day of exposure and monitors analyzed after two weeks storage following exposure (for either room temperature or refrigerated storage) must not exceed 10% at the 95% confidence level. 17
3H 001485
d. Since there is invariably a lag time between exposure and analysis of sampling devices, the stabilty of the exposed monitors during this waiting period must be determined. The monitors are tested with both room temperature storage and refrigerated storage. Storage at room temperature is preferred since it eliminates problems that may result in trying to maintain samples at decreased temperatures during shipment to the laboratory. Provision is made in the protocol to examine refrigerated sample storage if recoveries are inadequate following room temperature storage.
5. Performance Factors
a. Performance Specification: The following factors (over the ranges shown) should not affect the quantity of analyte determined.
F actor concentration exposure time face velocity humidity interferent orientation
Range 0.1 - 2 X PEl 0.1 - 1 X MRST 20 - 300 fpm 10 - 80% R.H. 0 - 1 X PEl parallel - perpendicular (to air flow in chamber)
18
3H 001486
Experimentation: Use a 16 run fractional factorial experimental design to determine significant factors and two factor interactions Use at least 4 monitors per run.
Evaluation Criterion: The operational (functional) range for any factor that causes a statistically significant difference in the recovery (at the 95% confidence level) must be defined.
Ambient conditions during exposure should have minimal effects on results determined with passive monitors. Specifically, there are desired ranges for factors affecting the sampling rate over which the monitors should properly function. These factors and their desired operational ranges are analyte concentration (0.1 - 2 X PEl), exposure time (0.1 - 1 X MRST), face velocity (20 - 300 fpm), temperature (10-40 *C) and interferences (relative humidity, 10-80% R.H. and chemical or analytical,0 - 2 X PEl). These factors may all be evaluated using a.fractional factorial experimental design with a significance level of 0.95 (10).
Knowledge of the analyte concentration range and the exposure time range over which a monitor properly functions is necessary for the industrial hygienist conducting the sampling. Face velocities may vary significantly in the workplace and may lead to undersampling
19
3M 001487
(at low face velocities starvation may occur) or to oversampling . (at ni.gh face velocities turbulence may occur in the quiescent zone). The range of 20-300 fpm should effectively bracket the range of face velocities encountered in a typical workplace. Convection induced currents alone have minimum velocities of 10 to 20 fpm (11) and when combined with other causes of air movement (heating and cooling, ventilation, and personnel and equipment movement) most likely result in minimum face velocities of greater than 20 fpm. The velocity of air striking a worker in an indoor facility is very likely less than 300 fpm. This corresponds to the face velocity created by briskly walking (3.5 miles/hour) in still air. For routine outdoor use, the monitors should be investigated at face velocities of up to 1800 fpm (20 miles/hour).
There are two principal types of interferences that may cause problems in sampling and analysis with passive monitors. These are competition of the analyte with another substance for the sorbent (coadsorption) and substances that react with the sorbent or react during analysis to give an erroneous measurement (chemical interference). Moisture (humidity) is a potential coadsorbent which may change daily and will certainly be different for different processes and facilities. It must be investigated over the range where it will most likely be encountered (10-80%.)
20
3H 0P148S
In addition to moisture, other gasses or vapors may be present during sampling and compete for the soroent in adsorption type monitors. For this case, a material snould be chosen whicn is known to have some degree of affinity for the sorbent which will likely be present during sampling. This material should be investigated at concentrations of 0 and 1 times the PEi to determine if it causes an erroneous measurement. For absorption type monitors, a suspected analytical interference should be chosen as something that would likely be present during sampling and investigated at a concentration range of 0 to 1 times the PEL.
If the 16 run fractional factorial provides insufficient data to determine the useful (operational) range of significant factors, then these factors must be further investigated to determine this useful range. If more than one factor requires further investigation then a second quarter replicate (16 run fractional factorial) is probably the most efficient experimental scheme and should be used.
Temperature Effects
a. Performance Specification: Tne variation in quantity of analyte determined with exposure temperature should be defined.
21
3M 001489
b. Experimentation: Expose at least 10 monitors at each of 3 temperature levels (10, 2b, and 40 'C) at 1 X PEu, 100 fpm, and 1 X MRST while maintaining a constant (+ 5%) absolute humidity.
c. Evaluation Criterion: The temperature effect must oe defined and a correction factor provided if needed.
\
d. Temperatures will vary in workplace environments and the diffusion
coefficient (for quiescent zone sampling) varies proportionately to
temperature raised to the l.S power. However, the gaseous volume
also varies with temperature such that the concentration on a mole
per unit volume basis varies inversely with temperature (12, 13).
Consequently, the quantity of material sampled varies as a function
of temperature to the 0.5 power. This may be perturoed depending
on the effect of temperature on the material used for draft shields
and on tne sorbent/analyte affinity. For permeation type monitors,
temperature effects are due to tne nature of the permeable membrane^
and may be quite large or small depending on the material used
(4).
Precision and Accuracy
a. Performance Specification: Monitors should be precise and accurate over the desired sampling range.
22
3M 001490
b. Experimentation: Expose monitors (at least 6 per concentration level) to 0.1, 0.5, i.O and 2.0 X PEl , 100 fpm, 80% RH and 1 X MRST. Analyze the monitors and calculate the CVp
c. Evaluation Criterion: CVy for O.b, 1.0 and 2.0 X PEl (for six samplers at each concentration level) must be Jjess than or equal to 0.105 for a non-biased system. The critical ITV^ for a biased system is taken from Figure 1. Bias is determined as the difference between the average of the values obtained for the independent methods and the values obtained with the passive monitors.
d. The precision and accuracy of the sampling and analytical method (over the range of 0.1 to 2.0 times, the PEl) must be defined and should be such that the determined concentration will be within ^25% of the true value 95% of the time. The 0.1 concentration level is included in the experimentation in order to obtain some degree of confidence in measurements taken at this level. It is not included as a requirement in the evaluation criterion since the criterion is concerned primarily with the precision near the standard.
23
3H 001491
the of the Specifications, Protocol and Criteria
Reference values are determined as the average of two independent methods. Analyte concentration in the chamber determined by these two mechanisms must () jree within at least p%. The independent methods should be N10SH Class 13 rr better methods (if they exist)* a calibrated instrument with demonstrated accuracy and stability of calibration, or the calculated chamber concentration. If permeation tubes or certified gas cylinders are used, the permeation rate of the tube or the cylinder concentration should be verified on a routine, periodic basis.
In many cases four concentration levels are utilized in the experimentation, however, successful completion of only the three higher levels are necessary to meet the evaluation criteria. The protocol/criteria were established in this manner so that monitors which perform satisfactorily at these low levels (0.1 X PEL) will be evaluated and consequently usable with some degree of confidence at these concentration levels. Monitors whicn do not perform satisfactorily at these low levels Dut are acceptable at the other three levels (Q.b, 1.0 and 2.0 X PEL) should be also oe considered evaluated arid consequently unable for the indicated concentration levels.
24
3H 001492
References
1. Taylor, D. G., R. E. Kupel and J. M. Bryant, Documentation of the NIOSH Validation Tests, DHEW (NIOSH) Publication no. 77-185.
2. Gunderson, E. C. and C. C. Anderson, Development and Validation of Methods for Sampling and Analysis of Workplace Toxic SuDstances, DHHS (NIOSH) Publication No. 80-133.
3. Palmes, E. D. and A. F. Gunnison, Personal Monitoring Device for Gaseous Contaminants, Am. Ind. Hyg. Assoc. J., 34, (1973).
4. Reiszner, K. D. and P. W. West, Collection and Determination of Sulfur Dioxide Incorporating Permeation and West-Gaeke Procedure, Environ. Sci. and Tech. 7, 526 (ly73).
- "\
5. ' Posner, J. CC. Letter to the Editor, Am. Ind. Hyg. Assoc. J., 42, A-28 (1981).
5. Moore, G., Letter to the Editor, Am. Ind. Hyg. Assoc. J., 42, A-26 (1981).
7. Thain, W., Monitoring Toxic Gases in the Atmosphere for Hygiene and Pollution Control, Pergamon Press, New York, New York, (1980).
! 8. Krajewski,J., J. Gromiec and M. Dobecki, Comparison of Methods for Determination of Desorption Efficiencies, Am. Ind. Hyg. Assoc. J., 41, 531 (1980).
9. Densley, M. D., Charcoal Tube Air Sampling of Organic Vapor Mixtures, Doctoral Dissertation, Department of Environmental Health, College of Medicine, University of Cincinnati (1981).
10. Strategy of Experimentation, 2nd edition, E. I. duPont de Nemours and Co., Wilmington, Delaware (1975).
11. Hemeon, W. C. L., Plant and Process Ventilation, 2nd edition. Industrial Press Inc., New York, New York (1963).
12. Palmes, E. D., A. F. Gunnison, J. DiMattio and C. Tomczyk, Persona! Sampler for Nitrogen Dioxide, Am. Ind. Hyg. Assoc. J. 37, 570 (1976).
13. Mullins, H. E. and L. W. Anders, A New Innovative Uiffusional Monitor for Sampling Ethylene Oxide in Air, 3 M Technical Report R-AlHAI (71.1 )R (1981).
25
3M 001493
BIAS (%)
FIGURE 1, COMBINATIONS OF CVX AND BIAS WHICH YIELD NET ERRORS (Sampling and * Analysis! OF 25% OR LESS AT LEAST 95% OF THE TIME
Thu critical estimates curve shown is for three sets of six samples. This curve will be slightly different for larger sample sixes.
3H 001494
Appendix Mathematics and Statistics Used in Evaluation
The statistics used are essentially those described in references 1 and 2 adapted for passive monitors. These statistics are based on the assumption that the total sampling and analytical error should be less than 25% for at least 95% of the samples analyzed at levels of 0.5 - 2.0 times the PEL. In order to meet this accuracy requirement the experimentally
determined total coefficient of variation (CVj) must be less than 0.105 for an unbiased method (1). For a biased method, the critical rVj is
reduced according to the graph in Figure 1.
Definitions:
Mean = x = arithmetic average n = number of observations
standard deviation = s
ji n-1
Coefficient of variation = CV = s/x
CV| = CV of the 6 spiked samples at each concentration level. CV2 - CV of the 6 generated samples at each concentration level.
Pooled coefficient of variation = CV
where:
f-j
CVi i f
rv\
cvT
degrees of freedom (n-1) at each concentration level CV of the observations at the it*1 concentration level index for the three concentration levels
tU
pooled CV based on the spiked samples at 0.5, 1.0 and 2.0 X PEL (6 samples at each level) pooled CV based on the generated samples at 0.5, 1,0 and 2.0 X PEL ( 6 samples at each level) pooled CV for sampling and analysis and desorption efficiency *
CVT + 0.1667 (CV;)2
Grubbs test (1,2) should be used to determine if an observation should be rejected as an outlier and Bartlett's test for homogeneity of CV's
(1,2) should be applied to determine the feasibility of pooling the CV's. Chi-squared must be less than 9.21 to pass Bartlett's test at the 1% significiance level (1,2).
*Note that an assumed pump error of 5% is not used in these calculations
sinrp niwim; are not uort with nacsivp monitors anti sAmnl-ino error will ho j
part Of
3H 001495