THI ^.^y^ITATITE DETERMIKATION OF CHLORIDES. Ill PYRANOLS
. ....
Scope
.
This method applies to the quantitative determination of ; chlorides in 1478, 1476, 1482, 1488, 1467 Pyranols and any
fixture of these Pyranols.
Specif ication 4 .-
............... - . ' " ' /
'^-rXr:
, . The chloride content of any Pyranol listed above shall not ; exceed 0.10 parts per million by weight.
Standards
....... ' -
'
Using reagent grade sodium chloride, make .up e I liter solution - containing 100 parts per million by weight of chlorides ms
follows: weigh .1648 grams of reagent grade Na Cl and C quantitatively transfer to a 1 liter volumetric flask, dissolve in distilled water," dilute to mark and shake well. Plaqe solution in 1 liter glass stoppered Pyrex bottle.
100 c.o. of the 100 ppm solution diluted to 100C c.c. - 10 ppm 10 c.c. of the 100 ppm solution diluted to 1000 c.c. - 1 ppm .10 c.c.of the 10 ppm solution diluted to 1000 c.c. - 0.1 ppm / 15 c.c-.of the 10 ppm solution diluted to 1000 c.c. 0.15 ppm i 20 c.o.of the 10 ppm solution diluted to 1000 c.c. 0.2 ppm \30 c.c.of the 10 ppm solution diluted to 1000 c.c. - 0.3 ppm
Allsolutions are stored in glass stoppered Pyrex bottles.
All solutions are freshly made every two weeks.
`
. All bottles are shaken well and 10-25 c.c. discarded every time
any solution is poured out of the bottle.
` The basic standard is the 0.1 ppm solution. When examined in
a 20x150 mm Pyrex teBt tube under e Tyndall beam after adding
a few drops of acidified Ag NO3 solution, the beam intensity
within 30 seconds after the addition of the Ag KO3 solution
shall be Just visible in a darkrooa. If the beam intensity
is not visible at all or is easily visible (I.e. too strong),
the solutions shall be discarded and new standards made.
Apparatus -
. ; ;
. . v -.
(1) Test tubes - Pyrex - 20x150 min. (2) Beakers, separatory funnels graduates, stirring rods
{3) A "Penlite" flashlight (2 batteries) JtLaxlWUA. to 4 ala
PITTSFIELD WORKS
{Cont.on H-6845480-1)
6/26/45 EL Raab 8/21/47 EL Raab Date 5/29/40 prepared by---HWLC&A-rt?--______
CNQINCMINO OCPT.
lab. Sect. 802
Sketch Ho.-U.-6.S4 54-80
fw-i,ro./i of
3309^
dsvm
;.... t
v- . -
STLCOPCB4077555
GENERAL ELECTRIC COMPANY, SCHENECTADY, N. Y.
Title Quantitative, determination cf chici-.tijes i:: pyraijcls
diameter light aperture. Kew batteries must be used frequently in order to be able tc perceive the .1 ppra standard.
3*-'
Solutions
--\ -
/ ' - .' . ' ^ ' '. ' .
(1) Chloride free distilled v.ater.
(2) Chloride standards as described above. .
(3) A 105C solution of Ac NO3 containing approximately 10#
by volume of reagent grade concentrated HNC3.
.
Apparatus Cleanliness
.
..
(1) All gloss apparatus such as beakers, graduates, stirring
rods, separatory funnels, etc. , , shall- be checked with
~
chloride free distilled writer just before using. A small
amount of distilled water swirled in sr.y piece cf glass
apparatus shall show no Tyndall'cone- when several drops of
A `O3 solution are udded to the water in a test tube and
examined as previously described.
y^>'v
-
Extraction of Chlorides from Pyrancls ?y
/V ' (1) 50 c,c. of Pyranol are heated to/90-100 C, placed in a
chemically clean 2h0 c.c. separatory' funnel and extracted
wit): two portions of hot (9G-95cC) chloride free distilled
water. The total weight of water shall be equal to that of the 50 c.c. of Pyranol tested. The`first extraction shall be made by using approximately 2/3 of the water (50 c.c.), shaking violently for 30 seconds, 5 minutes separation and decantation of the bulk of the water layer into a beaker or 9mall Erlenmeyer. The second extraction shall be made using the remaining 1/3 of the water (25 c.c.), shaking violently 30 seconds, 5 minutes separation, discarding the Pyrunol layer, and combining the two water extracts in the beaker or Erlenmeyer. Mix both water extracts thoroughly.
Analysis of V/ater Extract
(1) 10 c.c. of the water extract 1c placed In a chloride free 20x150 min. test tube and is cleared by either of the . alternative methods below.
(Cont. on H-G845480-2)
Dale
8/21/47 G/20/47
5/29/40
E. L. Raab E. L. Raab
Prepared by--EU-j/L--
fW l*?u A err
PITTSFIELD WORKS
CNQINECRINQ DIPT.
Lab. Sect. 002 Sketch No. tf-.68454.80-1......
DSW 330912
V
i:
STLCOPCB4077556
4-
QENERAL ELECTRIC COMPANY, SCHENECTADY, N. Y,
TW* QUANTITATIVE DETERMINATION.. p.F.;.CI^R11^,JN_WBAPIS..
(a) Centrifuge until the water shows & negative Tyndall beam
effect.
.. '
. .. ' ' ' - . . .
'
(b) Shake in the test tube with 3-5 c.c. of chloride free ethyl ether (tested previously with water)until the water shows a negative Tyndbll beam effect Mre than one ether
treatment may be necessary to clear. -'?*
(2) Add 3-5 drops of the acidified silver nitrate solution.
Do not shake. . . '
-.
(3) Test for chloride turbidity in a dark room using the spot beam as previously described with the beam directed across the diameter of the tube.
(4) The chloride content of the Pyranol is determined in parts
per million by weight by comparing the chloride content,of <
the extract with that of a stundard chloride solution in the'
same manner. When comparing to standards, the Ag NO3 solution
is added to the unknown and standards, within several seconds
and the Tyndall beam comparison mad e ml thin 30 seconda-jgggT' :
t&JF- the addi-tion of tho Ag Nftg.U V
'
(5) The chloride content must not cause a greater Tyndall am effect then that of a standard sodium chloride solution ntulnlng 0.1C parts per million by weight. This amount of
chloride as previously mentioned, is the minimum that can be detected by this method as outlined.
(A//me
t^
Jo S
/ r>
y
Dale
6/21/47
E. L. Raab
6/26/47
E. L. Raab
P./P.?/^P.. Prepared by -P W. Calrd . .