Analytical Report
Fluorochemical Characterization of Sediment, POTW Sludge, and
Landfill Leachate Samples Pensacola, FL (W2176)
Centre Analytical Laboratory Report No. 023-014T (Revision 1)
Revision Date 3/28/01
Testing Laboratory Centre Analytical Laboratory, Inc.
3048 Research Drive State College, PA 16801
3M Environmental Laboratory Contact
Kent R. Lindstrom Bldg. 2-3E-09 P.O. Box 33331
St. Paul, MN 55133-3331 Phone: (651)778-5352
Requester Kris Hansen Ph.D. 3M Environmental Technology & Safety Services
Bldg. 2-3E-09 P.O. Box 33331 St. Paul, MN 55133-3331
PAGE 1 0 F 5
BACK TO MAIN
1 Introduction
Results are reported for the analysis of a series of sediment, leachate, and sludge samples received by Centre Analytical Laboratories, Inc. (Centre) f rom the 3M Environmental Laboratory. The samples were collected from Pensacola, Florida and are part of 3M Project W2176. The Centre study number assigned to the project is 023-014.
Specific fluorochemical characterization by liquid chromatography / tandem mass spectrometry (LC/MS/MS) was requested for all samples. A total of 9 samples were received for analysis.
The samples were prepared and analyzed by LC/MS/MS for the following list of fluorochemicals:
Table 1: Target Analysis
Compound Name Perfluorooctane Sulfonate Perfluorooctane Sulfonvlamide Perfluorooctanoate
Acronym PFOS PFOSA POAA
The analytical methods used for originally developed for water samples and were validated by Centre. The water sample validation protocol and results are on file with Centre. The methods were modified for the sediment and sludge samples. The procedures have not been fully validated for these matrices. Data presented here is the highest quality data available at this time.
2 Sample Receipt
The samples were submitted in individual plastic containers and were not preserved. Seven individual sample containers were received. Samples were received on 7/28/00. Sample
collection d a te s w e re not supplied. C h a in -o f-cu sto d y inform ation is p re se n te d in A tta ch m e n t A.
3 Holding Times
The analytical method used was validated against a maximum holding time of 14 days in water samples. Stability after this time period has not been validated. However, it should be noted that field fortifications in water and other matrices have shown acceptable recoveries at 100 and 1000 ng/L for periods longer than 14 days.
PAGE 2 OF 5
BACK TO MAIN
4 Methods - Analytical and Preparatory
4.1 LC/MS/MS
4.1.1 Sample Preparation for LC/MS/MS Analysis
Water samples (leachate and field blanks) were initially treated with 200 uL of 250 mg/L sodium thiosulfate solution to remove residual chlorine. Solid phase extraction (SPE) was used to prepare the samples for LC/MS/MS analysis. A forty-milliliter portion of sample was transferred to a Ci8 SPE cartridge. The cartridge was first eluted with 5 mL of 40% methanol in water solution. The eluate was discarded and the SPE column was then eluted with 100% methanol. A 5 mL portion of methanol was collected for analysis by LC/MS/MS. This treatment resulted in an eight-fold concentration of the samples prior to analysis.
For the sediment and sludge samples, a representative portion of sample (5 grams) was first extracted into 5 ml of methanol. The extracts were filtered and diluted to a final volume of 40 mL with Type I water. The diluted extracts were then treated in the same manner as the water samples, beginning with the solid phase extraction.
4.1.2 Sample Analysis by LC/MS/MS
In HPLC, an aliquot of extract is injected and passed through a liquid-phase chromatographic column. Based on the affinity of the analyte for the stationary phase in the column relative to the liquid mobile phase, the analyte is retained for a characteristic amount of time. Following HPLC separation, ES/MS provides a rapid and accurate means for analyzing a wide range of organic compounds, including fluorochemicals. Electrospray is generally operated at relatively mild temperatures; molecules are ionized, fragmented, and detected. Ions characteristic of known fluorochemicals are observed and quantitated against standards.
A Hewlett-Packard HP1100 HPLC system coupled to a Micromass Ultima MS/MS was used to analyze the sample extracts. Analysis was performed using selected reaction monitoring
(SRM). Sam ples were extracted between 8/22/00 and 8/23/00 and analyzed by MS/MS
between 8/25/00 and 8/28/00. The HPLC and MS/MS methods used for analysis and instrument parameters can be found in Attachments D.
5 Analysis
5.1 Calibration
A 7-point calibration curve was analyzed at the beginning and end of the analytical sequence for the compounds of interest. The calibration points were prepared at 0, 25, 50, 100, 250, 500, and 1000 ng/L (ppt) for LC/MS/MS analysis. The instrument response versus the concentration was plotted for each point. Using linear regression with 1/x weighting, the slope, y-intercept and correlation coefficient (r) and coefficient of determination (r2) were determined. A calibration curve is acceptable if r >.0.985 (r2 > 0.970).
Calibration standards are prepared using the same SPE procedure used for samples.
PAGE3 OF5
BACK TO MAIN
Calibration check standards were analyzed periodically (every three to five sample injections) throughout the analysis sequence. Compliance is obtained if the standard analyte concentrations are within +1-20% of the actual value.
For the results reported here, calibration criteria were met.
5.2 Blanks
Extraction blanks were prepared and analyzed with every extraction batch of samples. The extraction blanks should not have any target analytes present at or above the concentration of the low-level calibration standard. For these samples, the extraction blanks were compliant.
Instrument blanks in the form of clean methanol solvent were also analyzed after every highlevel calibration standard, and after known high-level samples. Again, the blanks should not have any target analytes present at or above the low-level calibration standard. For the samples presented here the instrument blanks are compliant.
5.3 Surrogates
Surrogate spikes are not a component of the LC/MS/MS analytical methods.
5.4 Matrix Spikes
Matrix spikes were prepared for every field sample using all compounds of interest. Matrix spike recoveries are given in Attachment B. No field spikes were collected from this site.
5.5 Duplicates
All field samples were analyzed in duplicate. Results are given along with the sample results in Attachment A.
5.6 Laboratory Control Samples
For LC/MS/MS analyses, Milliq water was spiked with all compounds of interest at 25 and 250 ng/L during each extraction set. All recoveries for all compounds were between 70-130% in each LCS. Results are given along with the raw data in Attachments D.
5.7 Sample RelatedComments
Empty containers were submitted as filed blanks. The containers were treated with 40 mL of Milliq water and analyzed as per the method. There are no other sample related comments for this data set.
6 Data Summary
Please see Attachment A for a detailed listing of the analytical results. Field blank water and leachate results are reported in parts per trillion (ppt) (ng/L). Sediment and sludge sample results are reported in parts per billion (ppb) (ug/Kg) on both an as-received and dry-weight basis.
7 Data/Sample Retention
Samples are disposed of one month after the report is issued unless otherwise specified. All electronic data is archived on retrievable media and hard copy reports are stored in data folders maintained by Centre.
PAGE 4 OF5
BACK TO MAIN
8 Attachments
8.1 Attachment A: Results 8.2 Attachment B: Matrix Spike Recoveries 8.3 Attachment C: Chain of Custody 8.4 Attachment D: LC/MS/MS Raw Analytical Data
9 Signatures
iL po
____
^johnMJ^ahert^JOperaionsManageT
Kevin J Lloyd, Vice President
Other Lab Members Contributing to Data Karen Smith
Date
?8 nnA H >o i
Date
PAGE5 OF5
Centre Analytical Laboratories, Inc.
3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253
Analytical Results W2176 Pensacola, Florida (GEN 017)
3M Sample Identification
PFOS (ng/L)
PFOSA (ng/L)
POAA (ng/L)
MC-313K Field Blank
ND ND ND
MC-314K Field Blank
ND ND ND
MC-335K Landfill Leachate
ND
ND
ND
MC-335K Landfill Leachate dup
ND
ND
ND
MC-336K Field Blank
ND ND ND
MC-345K Field Blank
ND ND ND
Limit of Detection (LOD) for the procedure is approximately 2.5 ng/L for PFOS and PFOSA and 7.5 ng/L for POAA
Limit of Quantitation (LOQ) for the procedure is 25 ng/L for all compounds
ND - Compound not detected
NQ - Compound detected at a level between the LOD and LOQ. Result is not quantifiable.
ND < LOD < NQ < LOQ
Please refer to the reverse side for our standard terms and conditions.
0
ANALY
Centre Analytical Laboratories, Inc.
3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253
BumTM mm
Analytical Results W2176 Pensacola, Florida (GEN 017)
3M Sample Identification
MC-310K Sediment Site 1 MC-31 OK Sediment Site 1 dup MC-311K Sediment Site 2 MC-311K Sediment Site 2 dup MC-312K Sediment Site 3 MC-312K Sediment Site 3 dup MC-344K POTW Sludge MC-344K POTW Sludge dup
PFOS (ug/Kg) (as received)
ND ND 0.205 ND 0.329 NQ 27.8 28.2
PFOSA (ug/Kg) (as received)
ND ND ND ND ND ND 0.285 0.289
POAA (ug/Kg) (as received)
ND ND NQ ND ND ND 0.564 0.540
3M Sample Identification
PFOS (ug/Kg) (dry weight)
PFOSA (ug/Kg) (dry weight)
POAA (ug/Kg) (dry weight)
MC-31 OK Sediment Site 1 MC-31 OK Sediment Site 1 dup MC-311K Sediment Site 2 MC-311K Sediment Site 2 dup MC-312K Sediment Site 3 MC-312K Sediment Site 3 dup MC-344K POTW Sludge MC-344K POTW Sludge dup
ND ND 0.242 ND 0.408 NQ 124 125
ND ND ND ND ND ND 1.27 1.29
ND ND NQ ND ND ND 2.51 2.40
Limit of Detection (LOD) for the procedure is approximately 0.08 ug/Kg for all compounds (as received)
Limit of Quantitation (LOQ) for the procedure is 0.20 ug/Kg for all compounds (as received)
ND - Compound not detected
N Q - C o m p o u n d d e te c te d a t a le v e l b e tw e e n th e L O D a n d L O Q . R e s u lt is n o t q u a n tifia b le .
ND < LOD < NQ < LOQ
% Solids
78.87 78.87 84.78 84.78 80.66 80.66 22.48 22.48
Please refer to the reverse side for our standard terms and conditions.
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L):
MC-310K 250
PFOS PFOSA POAA
Sample Concentration
(ng/L)
0 0 0
Matrix Spike Result (ng/L)
198 186 196
Matrix Spike Result
(% Recovery)
79.2 74.4 78.4
Lower Recovery Limit:
70
Upper Recovery Limit:
130
Concentrations represent the amount present in the sample extract
Criteria (Pass / Fail)
PASS PASS PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L): |
MC-311K 250
PFOS PFOSA POAA
Sample Concentration
(ng/L)
25.6 0
10.6
Matrix Spike Result (ng/L)
205 191 210
Matrix Spike Result
(% Recovery)
71.8 76.4 79.8
Criteria (Pass / Fail)
PASS PASS PASS
Lower Recovery Limit:
70
Upper Recovery Limit:
130
Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are below the limit of quantitation. Results are given in this table for recovery calculations only.
Concentrations represent the amount present in the sample extract
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L):
MC-312K 250
PFOS PFOSA POAA
Sample Concentration
(ng/L)
41.1 0 0
Matrix Spike Result (ng/L)
624 149 263
Matrix Spike Result
(% Recovery)
233.2 59.6 105.2
Lower Recovery Limit:
70
Upper Recovery Limit:
130
Concentrations represent the amount present in the sample extract
Criteria (Pass / Fail)
FAIL FAIL PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L): |
MC-344K 250
PFOS PFOSA POAA
Sample Concentration
(ng/L)
3477 35.7 70.5
Matrix Spike Result (ng/L)
4062 54.2 249
Matrix Spike Result
(% Recovery)
234.0 7.4 71.4
Lower Recovery Limit:
70
Upper Recovery Limit:
130
Concentrations represent the amount present in the sample extract
PFOSsample concentration is greater than 10X the spiking level
Criteria (Pass / Fail)
FAIL FAIL PASS
BACK TO MAIN
Attachment B: LC/MS/MS Laboratory Spike Recovery
Sample ID: Spiked Amount (ng/L):
MC-335K 100
PFOS PFOSA POAA
Lower Recovery Limit:
Upper Recovery Limit:
Sample Concentration
(ng/L) 0 0 0
70
130
Matrix Spike Result (ng/L)
84.8 69.2 92.7
Matrix Spike Result
(% Recovery)
84.8 69.2 92.7
Criteria (Pass / Fail)
PASS FAIL PASS
3M Environmental Laboratory
Form 3877S - PWO
Shipping Address: 3M Bldfl~2-3E-09 935 Bush Avenue SL Paul. MN 55106
Ta{*nhnn*;
Sample Recalving: (651) 778-4948 Alternata: (651) 778-8753 FAX: (651) 778-8176
BACK TO MAIN
Chain of Custody /Reauest for Laboratory Analytical
15089
--------------------------------------- < ? C t \
____________________________________ __ ___________________
Project ID/Project Name L U (L k ^ 4 < !'itu v )in f-ju ^ a ic S jn ir,/,< l , ... t_e--
_AiA.--------- - --
Template #
-------------------------- ------------------
- ~ ' ^ -- -m iv io m ;(i\...x n i V ^ i>
Final Report Due Date
a u.
uai H
Project Lead tC n 'S J - H d n s e jr ^
Internal Due Date
3M Env. Lab Project #
For Internal Use Only
Contact Name U S O -
Report Results to:
Company
3 HI
Mailing Address fh l L A
3 -~ O Q
(3 1 /lS V n
City, State, Zip
P o ,0 \ , f Y ^ ' V L /U .S 'frx -
Telephone# >*3 (
Special Instruction* and/or Specific Regulatory Requirements: (m ethod, lim it of detection, reporting units, etc.)
*0
FAX#
Item # Client Sample Identification
1.
rnc - 5 1o K. - S 11~. 1T P i n il t o-trA
/ QC H un P aj r~
HP2S
S tA m e n t - P e n ric a . 1 /
2.
-S iFtfZ - PI/OiTnhd PP0 5 /a to n fA->r n P L L .- 5m <3 SeAu'm&rdr - f e r v u t o l a - F-i
3.
ftc .- 3iAX-SVf. 3 -p liv /re + v J P&05/QC j-o n p n .tr HPLC-fr)5 S e d im e n t-P e n tito la . Fl
4.
n e - 3V 5 F -fitJd B lo n t-f/W io n P a ir HPLC-6 /rt5 S c d itn e n t- P e n tito la , F|
5.
r'c .-iJ H K -C ie lc /t a n iL -T b te d fPo< JZ n f e - ir
M U -e st ^ ^ n t -fc n fu ia .. Fl
-- ____
6.
n \C .- i? n ie .- tlP TV-hU tfOShXL lo t i fCbC H P LC -iirnS C AreiP ill Lentia \T .- P trO t n \a . F l
7.
fo C -3 $ u K rP i jd /n/C - Pfr\i J o n ft U r lA p d f ih L ta c h n J t- (knS iclQ , Pt
.... ,
7 /v /O rili rFv*) ( QC 4 o n r * u r hhol. fcsrns
j Pi8. P h rw S ludc\e* *Prk4t_i>la
SWttLr9. PSTU
10.
' H i ci a & tart C.~ m f - J-on you r n ru u -c r^ rn 5
*j F I _______________
Collected by (print):
<-/ ,{ Y\uLtX-
Item #
1
Relinquished by/Affiliation
/ - 9 C<w A f J W
/ irh
3M LIMS#
',
Date Available Date Due
0 7 a 7- 0 0
Contract Lab
P re s e rv a tiv e s :
r) 0z1
O <w/)
X
O1A O >
z0c0
s<ku:. 0
Date
Time
Sampled Sampled
Matrix/ Media
Enter the num ber of containers of each
-_ y
sU /
/ /
/
/
Xu / h p /o .
s/ 1 .,/ L
------ --
a
A\ /A 0 ]in L /
/ /
Total Number'of Containers
c e. n
re
d:pinaly s is Teqiteste
Completa be low. ttach ny as sodate d informador .
.
IX
o_
v-- 1
*0 0 l-L CX
<*-or 0 LX. >~
CE
r 0 0-
/
/
/
1 ><i y *
1 Vy X
/
/
1
1
Vy
X X
X
y
1 y XX
/ 1Af
T Ai -- D 7? ) r-----
1yXy
--f
1 y yt
t
1X yX /
1 XyyT
Time
'-- "
. Date
&Collector's signal u r e : ^
Shipped Via:
erthil
u i SL.
Received by/Affiliation
Time //0 0
Date 7 -j2 F c \
Chain of Custody
Sample Condition Upon Receipt:
^Acceptable O Other:
Temperature:
'C
CTReceived on Ice
Other Associated CoCs:
Copies to:
J ________________________________________
A
K3I)
1 _LPage
Original - Accom panying Sam ples
lA c
Last Page - O riginator
Com m ents: See Reverse Side for Instruclions