Document JJG8ZLbVndjJO8aJNKNxp9yMZ

BACK TO MAIN Centre Analytical Laboratories, Inc. 3048 Research Drive State College, PA 16801 www.centrelab.com ^P hone: (814) 231-8032 Fax: (814) 231-1253or (814) 231-1580 Analytical Report Fluorochemical Characterization of Sediment, POTW Sludge, and Landfill Leachate Samples Pensacola, FL (W2176) Centre Analytical Laboratory Report No. 023-014T (Revision 1) Revision Date 3/28/01 Testing Laboratory Centre Analytical Laboratory, Inc. 3048 Research Drive State College, PA 16801 3M Environmental Laboratory Contact Kent R. Lindstrom Bldg. 2-3E-09 P.O. Box 33331 St. Paul, MN 55133-3331 Phone: (651)778-5352 Requester Kris Hansen Ph.D. 3M Environmental Technology & Safety Services Bldg. 2-3E-09 P.O. Box 33331 St. Paul, MN 55133-3331 PAGE 1 0 F 5 BACK TO MAIN 1 Introduction Results are reported for the analysis of a series of sediment, leachate, and sludge samples received by Centre Analytical Laboratories, Inc. (Centre) f rom the 3M Environmental Laboratory. The samples were collected from Pensacola, Florida and are part of 3M Project W2176. The Centre study number assigned to the project is 023-014. Specific fluorochemical characterization by liquid chromatography / tandem mass spectrometry (LC/MS/MS) was requested for all samples. A total of 9 samples were received for analysis. The samples were prepared and analyzed by LC/MS/MS for the following list of fluorochemicals: Table 1: Target Analysis Compound Name Perfluorooctane Sulfonate Perfluorooctane Sulfonvlamide Perfluorooctanoate Acronym PFOS PFOSA POAA The analytical methods used for originally developed for water samples and were validated by Centre. The water sample validation protocol and results are on file with Centre. The methods were modified for the sediment and sludge samples. The procedures have not been fully validated for these matrices. Data presented here is the highest quality data available at this time. 2 Sample Receipt The samples were submitted in individual plastic containers and were not preserved. Seven individual sample containers were received. Samples were received on 7/28/00. Sample collection d a te s w e re not supplied. C h a in -o f-cu sto d y inform ation is p re se n te d in A tta ch m e n t A. 3 Holding Times The analytical method used was validated against a maximum holding time of 14 days in water samples. Stability after this time period has not been validated. However, it should be noted that field fortifications in water and other matrices have shown acceptable recoveries at 100 and 1000 ng/L for periods longer than 14 days. PAGE 2 OF 5 BACK TO MAIN 4 Methods - Analytical and Preparatory 4.1 LC/MS/MS 4.1.1 Sample Preparation for LC/MS/MS Analysis Water samples (leachate and field blanks) were initially treated with 200 uL of 250 mg/L sodium thiosulfate solution to remove residual chlorine. Solid phase extraction (SPE) was used to prepare the samples for LC/MS/MS analysis. A forty-milliliter portion of sample was transferred to a Ci8 SPE cartridge. The cartridge was first eluted with 5 mL of 40% methanol in water solution. The eluate was discarded and the SPE column was then eluted with 100% methanol. A 5 mL portion of methanol was collected for analysis by LC/MS/MS. This treatment resulted in an eight-fold concentration of the samples prior to analysis. For the sediment and sludge samples, a representative portion of sample (5 grams) was first extracted into 5 ml of methanol. The extracts were filtered and diluted to a final volume of 40 mL with Type I water. The diluted extracts were then treated in the same manner as the water samples, beginning with the solid phase extraction. 4.1.2 Sample Analysis by LC/MS/MS In HPLC, an aliquot of extract is injected and passed through a liquid-phase chromatographic column. Based on the affinity of the analyte for the stationary phase in the column relative to the liquid mobile phase, the analyte is retained for a characteristic amount of time. Following HPLC separation, ES/MS provides a rapid and accurate means for analyzing a wide range of organic compounds, including fluorochemicals. Electrospray is generally operated at relatively mild temperatures; molecules are ionized, fragmented, and detected. Ions characteristic of known fluorochemicals are observed and quantitated against standards. A Hewlett-Packard HP1100 HPLC system coupled to a Micromass Ultima MS/MS was used to analyze the sample extracts. Analysis was performed using selected reaction monitoring (SRM). Sam ples were extracted between 8/22/00 and 8/23/00 and analyzed by MS/MS between 8/25/00 and 8/28/00. The HPLC and MS/MS methods used for analysis and instrument parameters can be found in Attachments D. 5 Analysis 5.1 Calibration A 7-point calibration curve was analyzed at the beginning and end of the analytical sequence for the compounds of interest. The calibration points were prepared at 0, 25, 50, 100, 250, 500, and 1000 ng/L (ppt) for LC/MS/MS analysis. The instrument response versus the concentration was plotted for each point. Using linear regression with 1/x weighting, the slope, y-intercept and correlation coefficient (r) and coefficient of determination (r2) were determined. A calibration curve is acceptable if r >.0.985 (r2 > 0.970). Calibration standards are prepared using the same SPE procedure used for samples. PAGE3 OF5 BACK TO MAIN Calibration check standards were analyzed periodically (every three to five sample injections) throughout the analysis sequence. Compliance is obtained if the standard analyte concentrations are within +1-20% of the actual value. For the results reported here, calibration criteria were met. 5.2 Blanks Extraction blanks were prepared and analyzed with every extraction batch of samples. The extraction blanks should not have any target analytes present at or above the concentration of the low-level calibration standard. For these samples, the extraction blanks were compliant. Instrument blanks in the form of clean methanol solvent were also analyzed after every highlevel calibration standard, and after known high-level samples. Again, the blanks should not have any target analytes present at or above the low-level calibration standard. For the samples presented here the instrument blanks are compliant. 5.3 Surrogates Surrogate spikes are not a component of the LC/MS/MS analytical methods. 5.4 Matrix Spikes Matrix spikes were prepared for every field sample using all compounds of interest. Matrix spike recoveries are given in Attachment B. No field spikes were collected from this site. 5.5 Duplicates All field samples were analyzed in duplicate. Results are given along with the sample results in Attachment A. 5.6 Laboratory Control Samples For LC/MS/MS analyses, Milliq water was spiked with all compounds of interest at 25 and 250 ng/L during each extraction set. All recoveries for all compounds were between 70-130% in each LCS. Results are given along with the raw data in Attachments D. 5.7 Sample RelatedComments Empty containers were submitted as filed blanks. The containers were treated with 40 mL of Milliq water and analyzed as per the method. There are no other sample related comments for this data set. 6 Data Summary Please see Attachment A for a detailed listing of the analytical results. Field blank water and leachate results are reported in parts per trillion (ppt) (ng/L). Sediment and sludge sample results are reported in parts per billion (ppb) (ug/Kg) on both an as-received and dry-weight basis. 7 Data/Sample Retention Samples are disposed of one month after the report is issued unless otherwise specified. All electronic data is archived on retrievable media and hard copy reports are stored in data folders maintained by Centre. PAGE 4 OF5 BACK TO MAIN 8 Attachments 8.1 Attachment A: Results 8.2 Attachment B: Matrix Spike Recoveries 8.3 Attachment C: Chain of Custody 8.4 Attachment D: LC/MS/MS Raw Analytical Data 9 Signatures iL po ____ ^johnMJ^ahert^JOperaionsManageT Kevin J Lloyd, Vice President Other Lab Members Contributing to Data Karen Smith Date ?8 nnA H >o i Date PAGE5 OF5 Centre Analytical Laboratories, Inc. 3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253 Analytical Results W2176 Pensacola, Florida (GEN 017) 3M Sample Identification PFOS (ng/L) PFOSA (ng/L) POAA (ng/L) MC-313K Field Blank ND ND ND MC-314K Field Blank ND ND ND MC-335K Landfill Leachate ND ND ND MC-335K Landfill Leachate dup ND ND ND MC-336K Field Blank ND ND ND MC-345K Field Blank ND ND ND Limit of Detection (LOD) for the procedure is approximately 2.5 ng/L for PFOS and PFOSA and 7.5 ng/L for POAA Limit of Quantitation (LOQ) for the procedure is 25 ng/L for all compounds ND - Compound not detected NQ - Compound detected at a level between the LOD and LOQ. Result is not quantifiable. ND < LOD < NQ < LOQ Please refer to the reverse side for our standard terms and conditions. 0 ANALY Centre Analytical Laboratories, Inc. 3048 Research Drive, State College PA 16801 814-231-8032 FAX 814-231-1253 BumTM mm Analytical Results W2176 Pensacola, Florida (GEN 017) 3M Sample Identification MC-310K Sediment Site 1 MC-31 OK Sediment Site 1 dup MC-311K Sediment Site 2 MC-311K Sediment Site 2 dup MC-312K Sediment Site 3 MC-312K Sediment Site 3 dup MC-344K POTW Sludge MC-344K POTW Sludge dup PFOS (ug/Kg) (as received) ND ND 0.205 ND 0.329 NQ 27.8 28.2 PFOSA (ug/Kg) (as received) ND ND ND ND ND ND 0.285 0.289 POAA (ug/Kg) (as received) ND ND NQ ND ND ND 0.564 0.540 3M Sample Identification PFOS (ug/Kg) (dry weight) PFOSA (ug/Kg) (dry weight) POAA (ug/Kg) (dry weight) MC-31 OK Sediment Site 1 MC-31 OK Sediment Site 1 dup MC-311K Sediment Site 2 MC-311K Sediment Site 2 dup MC-312K Sediment Site 3 MC-312K Sediment Site 3 dup MC-344K POTW Sludge MC-344K POTW Sludge dup ND ND 0.242 ND 0.408 NQ 124 125 ND ND ND ND ND ND 1.27 1.29 ND ND NQ ND ND ND 2.51 2.40 Limit of Detection (LOD) for the procedure is approximately 0.08 ug/Kg for all compounds (as received) Limit of Quantitation (LOQ) for the procedure is 0.20 ug/Kg for all compounds (as received) ND - Compound not detected N Q - C o m p o u n d d e te c te d a t a le v e l b e tw e e n th e L O D a n d L O Q . R e s u lt is n o t q u a n tifia b le . ND < LOD < NQ < LOQ % Solids 78.87 78.87 84.78 84.78 80.66 80.66 22.48 22.48 Please refer to the reverse side for our standard terms and conditions. BACK TO MAIN Attachment B: LC/MS/MS Laboratory Spike Recovery Sample ID: Spiked Amount (ng/L): MC-310K 250 PFOS PFOSA POAA Sample Concentration (ng/L) 0 0 0 Matrix Spike Result (ng/L) 198 186 196 Matrix Spike Result (% Recovery) 79.2 74.4 78.4 Lower Recovery Limit: 70 Upper Recovery Limit: 130 Concentrations represent the amount present in the sample extract Criteria (Pass / Fail) PASS PASS PASS BACK TO MAIN Attachment B: LC/MS/MS Laboratory Spike Recovery Sample ID: Spiked Amount (ng/L): | MC-311K 250 PFOS PFOSA POAA Sample Concentration (ng/L) 25.6 0 10.6 Matrix Spike Result (ng/L) 205 191 210 Matrix Spike Result (% Recovery) 71.8 76.4 79.8 Criteria (Pass / Fail) PASS PASS PASS Lower Recovery Limit: 70 Upper Recovery Limit: 130 Note: Sample results less than 25 ng/L are reported as NQ in the results section as they are below the limit of quantitation. Results are given in this table for recovery calculations only. Concentrations represent the amount present in the sample extract BACK TO MAIN Attachment B: LC/MS/MS Laboratory Spike Recovery Sample ID: Spiked Amount (ng/L): MC-312K 250 PFOS PFOSA POAA Sample Concentration (ng/L) 41.1 0 0 Matrix Spike Result (ng/L) 624 149 263 Matrix Spike Result (% Recovery) 233.2 59.6 105.2 Lower Recovery Limit: 70 Upper Recovery Limit: 130 Concentrations represent the amount present in the sample extract Criteria (Pass / Fail) FAIL FAIL PASS BACK TO MAIN Attachment B: LC/MS/MS Laboratory Spike Recovery Sample ID: Spiked Amount (ng/L): | MC-344K 250 PFOS PFOSA POAA Sample Concentration (ng/L) 3477 35.7 70.5 Matrix Spike Result (ng/L) 4062 54.2 249 Matrix Spike Result (% Recovery) 234.0 7.4 71.4 Lower Recovery Limit: 70 Upper Recovery Limit: 130 Concentrations represent the amount present in the sample extract PFOSsample concentration is greater than 10X the spiking level Criteria (Pass / Fail) FAIL FAIL PASS BACK TO MAIN Attachment B: LC/MS/MS Laboratory Spike Recovery Sample ID: Spiked Amount (ng/L): MC-335K 100 PFOS PFOSA POAA Lower Recovery Limit: Upper Recovery Limit: Sample Concentration (ng/L) 0 0 0 70 130 Matrix Spike Result (ng/L) 84.8 69.2 92.7 Matrix Spike Result (% Recovery) 84.8 69.2 92.7 Criteria (Pass / Fail) PASS FAIL PASS 3M Environmental Laboratory Form 3877S - PWO Shipping Address: 3M Bldfl~2-3E-09 935 Bush Avenue SL Paul. MN 55106 Ta{*nhnn*; Sample Recalving: (651) 778-4948 Alternata: (651) 778-8753 FAX: (651) 778-8176 BACK TO MAIN Chain of Custody /Reauest for Laboratory Analytical 15089 --------------------------------------- < ? C t \ ____________________________________ __ ___________________ Project ID/Project Name L U (L k ^ 4 < !'itu v )in f-ju ^ a ic S jn ir,/,< l , ... t_e-- _AiA.--------- - -- Template # -------------------------- ------------------ - ~ ' ^ -- -m iv io m ;(i\...x n i V ^ i> Final Report Due Date a u. uai H Project Lead tC n 'S J - H d n s e jr ^ Internal Due Date 3M Env. Lab Project # For Internal Use Only Contact Name U S O - Report Results to: Company 3 HI Mailing Address fh l L A 3 -~ O Q (3 1 /lS V n City, State, Zip P o ,0 \ , f Y ^ ' V L /U .S 'frx - Telephone# >*3 ( Special Instruction* and/or Specific Regulatory Requirements: (m ethod, lim it of detection, reporting units, etc.) *0 FAX# Item # Client Sample Identification 1. rnc - 5 1o K. - S 11~. 1T P i n il t o-trA / QC H un P aj r~ HP2S S tA m e n t - P e n ric a . 1 / 2. -S iFtfZ - PI/OiTnhd PP0 5 /a to n fA->r n P L L .- 5m <3 SeAu'm&rdr - f e r v u t o l a - F-i 3. ftc .- 3iAX-SVf. 3 -p liv /re + v J P&05/QC j-o n p n .tr HPLC-fr)5 S e d im e n t-P e n tito la . Fl 4. n e - 3V 5 F -fitJd B lo n t-f/W io n P a ir HPLC-6 /rt5 S c d itn e n t- P e n tito la , F| 5. r'c .-iJ H K -C ie lc /t a n iL -T b te d fPo< JZ n f e - ir M U -e st ^ ^ n t -fc n fu ia .. Fl -- ____ 6. n \C .- i? n ie .- tlP TV-hU tfOShXL lo t i fCbC H P LC -iirnS C AreiP ill Lentia \T .- P trO t n \a . F l 7. fo C -3 $ u K rP i jd /n/C - Pfr\i J o n ft U r lA p d f ih L ta c h n J t- (knS iclQ , Pt .... , 7 /v /O rili rFv*) ( QC 4 o n r * u r hhol. fcsrns j Pi8. P h rw S ludc\e* *Prk4t_i>la SWttLr9. PSTU 10. ' H i ci a & tart C.~ m f - J-on you r n ru u -c r^ rn 5 *j F I _______________ Collected by (print): <-/ ,{ Y\uLtX- Item # 1 Relinquished by/Affiliation / - 9 C<w A f J W / irh 3M LIMS# ', Date Available Date Due 0 7 a 7- 0 0 Contract Lab P re s e rv a tiv e s : r) 0z1 O <w/) X O1A O > z0c0 s<ku:. 0 Date Time Sampled Sampled Matrix/ Media Enter the num ber of containers of each -_ y sU / / / / / Xu / h p /o . s/ 1 .,/ L ------ -- a A\ /A 0 ]in L / / / Total Number'of Containers c e. n re d:pinaly s is Teqiteste Completa be low. ttach ny as sodate d informador . . IX o_ v-- 1 *0 0 l-L CX <*-or 0 LX. >~ CE r 0 0- / / / 1 ><i y * 1 Vy X / / 1 1 Vy X X X y 1 y XX / 1Af T Ai -- D 7? ) r----- 1yXy --f 1 y yt t 1X yX / 1 XyyT Time '-- " . Date &Collector's signal u r e : ^ Shipped Via: erthil u i SL. Received by/Affiliation Time //0 0 Date 7 -j2 F c \ Chain of Custody Sample Condition Upon Receipt: ^Acceptable O Other: Temperature: 'C CTReceived on Ice Other Associated CoCs: Copies to: J ________________________________________ A K3I) 1 _LPage Original - Accom panying Sam ples lA c Last Page - O riginator Com m ents: See Reverse Side for Instruclions