Document GmO761rbja6xykewO76Lw8xLm
' Date: 11/17/43 Mat-aal:
Aroclor
By: HOHiDBH . ijf'
Ref:
Test:
Dielectric Constant
Method No. 10,119-43 |pnt: B No. 246-012
Page: 1
1. Clean the electrodes scrupulously as follows before each sample is tested, being careful not to touch any part except the con tact terminals. (When the electrodes lose their high polish, deliver to the chief analytical chemist, who will consider hav ing them replated.)
a. Scrub thoroughly with hot tap water and trisodium phosphate using the brush reserved exclusively for the purpose.
b. Rinse thoroughly with hot tap water.
c. Next rinse thoroughly with distilled water.
d. Dry the electrodes in the special oven at 120C for 1 hr. standing erect on a clean sheet of lintless paper.
2. Clean a 1000-mli beaker, the glass electrode spacer, and a -10 to 220C. thermometer as follows before each sample is tested:
a. Scrub thoroughly with hot tap water and trisodium phosphate using the brush reserved exclusively for the purpose.
F. 221
Monsanto Chemical Company ANALYTICAL LABORATORY
DSW 330922
STLCOPCB4077563
' Date:
By: Ref:
H/17/43 Mat#sal:
Aroclor
H0H:DBH
Test: Dielectric Constant
JtMethod No. 10,119-43
: B No. 246-012
Page: 2
b. Rinse thoroughly with hot tap water.
( c. Place the spacer and the thermometer in the beaker, fill
with chromic acid cleaning solution (ca. 6oC.), and
allow to stand 15 min. Discard the cleaning solution after it becomes green.
d. Rinse thoroughly with hot tap water.
et Finally rinse thoroughly with distilled water and drain well.
f. Dry tho clootrode-s in the special oven at 120C. for 1 hr* standing erect on a clean sheet of lintless paper.
3. Remove the apparatus from the oven. Immediately place the spacer in the bottom of the beaker, put the two electrodes in the spaoer grooves, and the thermometer outside the outer electrode.
4. Place the beaker in the electric heater which has also been
preheated.
-
5. Balance the dielectric constant apparatus as follows:
f. 221
Monsanto Chemical Company ANALYTICAL LABORATORY
DSW 330923
STLCOPCB4077564
* Date: 11/17/43
By: HOH: DBH Ref:
Mate^l:
Aroclor
v3
Test: Dielectric Constant
Method No. 10,119"43 B No. 246--012
Page: 3
a. Allow electrode leads to hang free. ,
b. Close double-pole, single-throw switch.
c. Turn toggle switch on amplifier to "On" position.
d. Turn frequency dial to "1000
e. Turn amplifier volume dial to give.a volume of sound as experience dictates.
f. Adjust precision condenser to 1600 mmf.
g. Adjust variable condenser to a null point, using the re sistance box if necessary.
6. Connect the red lead of the dielectric constant set to the inner eleotrode end fthe black to the outer electrode.
7. Adjust the precision condenser to a second null point and read (+ 0.5 mmf.).
8
F. 221
Pour a sufficient quantity of preheated sample (97 2C.) to
Monsanto Chemical Company ANALYTICAL LABORATORY
DSW 330924
STLCOPCB4077565
Date: ll/l7/43 Majggal:
By: HOHiDBH
I#
Aroclor
Method No. 10,119-43 >t: B No. 246-012
Ref:________________ Test: Dielectric Constant_____ Page: 4___________
cover the electrodes l/2 In.
9. Stir slowly. Vvhen the sample reaches 100.0 + 0.5C., remove the thermometer and check the electrodes to see that they are properly spaced.
10* Immediately rebalance the precision condenser to a null point end read (+ 0.5 mmf.).
11. Next determine resistivity (Method No. 10,120-43) if required.
Calculg.tiQaa:
1) From the following chart convert the precision condenser readings
to micromicrofarads:
F. 221
Monsanto Chemical Company ANALYTICAL LABORATORY
DSW 330921
STLCOPCB4077566
S.*.
' Date: II/17/43
By: H0H:DBH
Ref:
WMatdal:
Aroclor
Test:
Dielectric Constant
M-sthod No. 10,119-43 ! At: B No. 246-012 Page: 5_________ '
CALIBRATION DATA GENERAL RADIO COMPANY TYPE 222 L PRECISION CONDENSER
(SERIAL NO. 1189)
Scale Reading Divisions
0 100 200 300 400 500 600 700 800 900 1000 1100 1200
F. 221
Capacitance MMF.
Scale Reading Pivi slone
49 57
8
101 43
159 48 218 59
277
59 58
335 La
396 n
455
59 6w0
515 60
575 635 695
60
60 60
1300 1400 1500 1600 1700
1800
1900 2000 2100 2200 2300 2400 2500
Monsanto Chemical Company ANALYTICAL LABORATORY
Capacitance _____ ML------
755 815 875 935 995
1055 1116 1176
1237 1297 1358 1418
1468
60 60
60
60 60 60
61 60 6l 60
&L
60
50
SW 330979
STLCOPCB4077567
<r" ' ' Date: H/17/43 By: HOHlDBE
Ref:
Mat&l:
U
Aroclor
Test: Dielectric Constant
Msfhod No. 10,119-43 |pt: B No. 246-012
Page: 6__________________
, 2) Dielectric Constant = -935.
( 935 - capacity with air
The method is accurate to 0.0$.
f Report the result to the nearest 0.01.