Document G5eay7m3vyOOOm0NR7oBXz9n7
PD'
Monsanto Chemical. Company
W. 0. KRUMMHICH PLANT LABORATORY
STANDARD ANALYTICAL METHOD
DATE ISSUED
______7-20_-71
WRITTEN BY
| PAOE
| NVNioen
1 or U I__ ]AM6vvktff 6V
JHFrancis
HJKorner
HTACHMCNT.1
Sample Chromatograms (3)
--
PHODUC T
MSCiQIo,
Arocler, Monochloroblphenyl (Crude and Finished Goods)
TEST
a:
xarrh I crobiphenyls
THIS FORM
JPCRSLOEC METHOD REVISES INSTRUCTIONS
ISSUED ON
NEW
This procedure determines penta- and hexachlorobiphenyls in KC3 1016 and in 12-2 Aroclor. Approximate Isomer distribution can also be determine,; on. I'CS 1016,. 1242 Aroclor and mcnochlcrobi phenyls.
DO NCT USE THIS PROCEDURE WITH HIGHER CHLORINATED BIPHENYLS. THEY ELUTE FROM THE GLC COLUMN WITH EXTREME DIFFICULTY AND ONLY AT TEMPERATURES THAT DRASTICALLY SHORTEN COLUMN LIFE.
EQ'JIPHEE'T AND SUPPLIES
1. GLC Instrument, Varian-Aerograph Model 144410-00, flame ionisation detector, capillary splitter, 11near-programmed, single column. Varian-Aerograph, 2700 Mitchell Drive, Walnut Creek, CaliP., 94596 (or equivalent).
2. Recorder, any good 1 mv. full scale, l-seccnd or les3 response time equipped with DISC or electronic Integration.
J. Column, 50 ft. X 0.02 inch I.D. S.C.O.T. column with FFAP liquid phase. Perkin Elmer Corp. ,- Norwalk, Conn., 06852.
4. Syringe, 0.2 ul fixed-volume. Precision Sampling'Co . Cat. No. 200002 (or equivalent syringe to deliver 0.2 ul). Fisher Scientific Co., 1241 Ambassador Blvd. , St. Louis, Mo., 63132
SAFETY
Chlorinated biphenyls are now considered environmental contaminante. Handle with care and dispose of waste and spill cleanup in the prescribed containers.
GLC '.rejection ports and ovens are usually very hot. Take care not to touch the hot equipment.
MQNS 072492
Product: MOS 1010 Test: Penta- and KexacnlorobiphenyIs
Method No. T-2302 Page No. 2 of 4
INSTRUMENT CONDITIONS
Injection. Port Detector Column Oven Range (input) Column Helium Flow
Splitter Flow Hydrogen Air Injection
325 + 10C
320C
235C Isothermal*
10-11 Amps/MV
10 cc/min. at room temperature
50 cc/min.(equiv. to a 5-1 split)
30 cc/min.
.
300 cc/min.
0.2 ul
Column temperature may be adjusted downwards as the
column ages, to maintain retention time. Different
column temperature may be necessary for monochloro-
blpbenyl isomer distribution.
.
DETERMINATION
1. Bring the instrument to the conditions described above for at least 5 minutes. The baseline should be stable at 4-1 attenuation. Set the attenuator at 256-1 before
injecting the sample if Disc integration is to be used. Electronic Integration can be at any attenuation but the RANGE setting must not be changed during a run.
2. Load the syringe by Immersing the needle at least 1 inch below the surface of the sample and moving the plunger in and out slowly for at least 5 cycles or until the small
bubbles of air are no longer expelled from the needle. Wipe the needle off with a soft tissue.
3. Depress the computer start button to activate the computer scan and Immediately inject the sample Into the GLC. Leave the needle in tne injection port for 5-6 seconds and then withdraw slowly.
4. If Disc Integration is used the attenuation must be changed to 2-1 or 16-1 after 16 minutes (new column) so that the penta- and hexachlorobiphenyIs elute with a large enough pe.r(s) ta integrate accurately (see sample chromatograms).
5. The chromatogram should be complete in 20 minutes for MCS 1016 and in 30 minutes for 1242 Aroclor. Monochloroblphenyls should be complete in less than 20 minutes, (if the computer program is used it will automatically end at 30 minutes for all samples.)
5. Determine tr.c counts under the MCS 1016 peak envelope (1-16 minutes) arc for the combined penta- and hexachloroblphenyls (lo-JO minutes).
HONS 072493
Hexachloro biphenyls
Method No. T-2302 Page 3 of'* pages
calcuut:
% Penta- and Hexachl-roblphenyls =
( Cij'j 1 o +
's)
Area for PCBP + HCBP X 100 Total Area
Isomer (individual)
Area for Isomer X 100 X ACF* Total Area (Corrected)
Monochloro'olpher.yls will require areacorrection factor for Bl#. See sample carve.
Report the results to the nearest 0.0156.
Duplicate analysis are expected to check to within 0.0A< (actual).
Precision data should be collected aa soon as Is feasible.
DISCUSSION
According to recent Information pentachlorobiphenyls and higher chlorinated Isomers are very difficult to biologically degrade and tend to concentrate In fish and birds. This test is being performed to help keep these Isomers to a minimum In MCS 1016.
Column life Is expected to be about 1 month at 235C. Column life can be extended if when there are. to be no samples for several hours the column temperature is reduced to 200C. It should only take about 10 minutes for the column to restablllze when the temperature Is again increased to 235C. A spent column will be indicated when the peaks show marked tailing and loss of separation of some of the doublet peaks. If In doubt, refer to the sample chromatograms.
Monochlorobiphenyl is a relatively new product and standards are not presently available to prepare a sample chromatogram.
It Is expected that the column temperature will have to be reduced to about 200"C to effect a proper separation of the
isomers. Order of elution of components will be Biphenyl, o-Chlorobipheny. m-chlorobiphenyl, p-chloroblphenyl, then dlchloroblphenyls and
trlchlorobiphenyls. Higher chlorinated isomers are not expected to be present.
We expect production to eventually a3k for analysis of
Recovered Biphenyl for chloroblphenyls. The Recovered Biphenyl will probably be a solid at room temperature. Dilution in benzene (1-1 volume/volume) and injection of 0.5 ul should handle this type sample. (Benzene counts would then be Ignored in the area normalisation.) A 1 ul syringe with the plunger In
the needle she :ld be used for this injection.
MONS 072494
Produce: MCS 1016 Teat: Per.ta- and KexachloroblphenyIs
Method No. T 2302 Page 4 of 4 pages
Hydrogen and air flows can be checked by attaching a bubble meter to the flow diverters on the right side of the Instrument
and pushing In all the way. When finished checking the flows make sure the diverters are pulled all the way out and then re-igr.ite the flame detector. Column flow is checked bv JlTconnecting the column exit end and checking the flow with a bubble meter. Splitter flow is checked at the split exit on the right side of the instrument. These flows should be checked at least once a week or whenever abnormal operation is suspected. Occasionally the flame becomes extinguished for unknown reasons. If after ar. Injection has been made and peaks are overdue or extremely small check to see if the flame
Is still lit by holding in the Igniter button for about 5 seconds. The cap on the detector will usually "pop" if the flame was out. The injection must then finish eluting before an analysis can be restarted.
HONS 072495