S. Tucker - R&D Laboratories - South Second Street ..St._ .Louis
December 12, 1969
R. A. Baxter - Ruabon H. A. Vodden - Ruabon R. E. Keller - S. 2nd W. R. Richard - GO D. Danna - GO
TO :R* A- Lldgett - Ruabon
I agree with your feeling that there Is definitely a need for higher resolution columns for PCB analysis, especially when dealing with weathered PCB samples In which the original isomer distribution has been altered.
I have recently investigated the feasibility of using support
coated open tubular columns with electron capture detection to
analyze for low levels of PCB's (5 X 10"9g injected). I
obtained fairly good resolution (see attached chromatograms of
1242 and 1248) but ran Into several correctable problems. First,
the sample Injection volume I used was much too large and it
adversely affected both the column life and Its resolution. It
appears that the.on column Injection of 5 uls of hexane washed
the liquid phase, off the column seriously shortening the life of
the column. The large solvent volume also seemed to cause some
peak broadening.
^
These effects can probably be overcome.by using more concentrated solutions in conjunction with a sample Inlet splitter, i.e., 1 ul Injection with ~20/l split. This would, of course, reduce the sensitivity by a factor 20 - 100X.
Currently, I am awaiting the arrival of 100' of 0.02" XE-60
S.C.O.T. column Which I have ordered. When the column Is
delivered and I've had time to try the above suggestions, I will
let you know.
.
I have not observed any decomposition of Aroclor during gas
chromatography. "
'
db Attach.
ii
j ;
t
f.
E. S. Tucker