Document BR4xe0XvGXy2apeeE2OpzLeq4

Spirometric Standards for Healthy Nonsmoking Adults'" JAMES F. MORRIS, ARTHUR KOSKI, and LFVON C. JOHNSON SUMMARY Previous studies designed to establish normal standards for ventilatory testing have usually included substantial numbers of cigarette smokers and hospital patients. Forced expiratory flows and volumes were measured in 988 healthy nonsmoking men and women who had relatively little exposure to any form of air pollution. Xegative correlations were obtained with age and positive correlations with height. Prediction formulas were derived and nomograms were constructed for 4 standard ventilatory measurements. Comparisons were made with prediction formulas of other workers. There was progressive decline in ventilatory function with age; larger values in men were present at all ages. Wide variation was encountered and was not attributable to age, sex, or height. This was most marked in the two forced expiratory flows, which had low multiple correlations and largc standard errors of estimate. The wide range of normal values seriously impairs the usefulness of spirometry in reliably detecting impairment of ventilatory function in asymptomatic individuals. Other techniques are needed to detect early impairment of pulmonary function. - Introduction Spirometers are being used increasingly for measuring ventilatory function by individual physicians and in pulmonary function screening programs. The increased interest in spirometry is primarily related to the burgeoning incidence of chronic ob- structive pulmonary disease. T h e widespread concern about air pollution and the well publicized harmful effects of inhaling (Received in revised form May 15, 1970) 1 From the Departments of Medicine, University of Oregon Medical School, and Veterans Administration Hospital, Portland, Oregon; and the Department of Health Education, Oregon State University, Corvallis, Oregon. 2Supported by a grant from the Oregon Tuberculosis and Respiratory Disease Association. 8 Requests for reprints should be addressed to Dr. James F. Morris, Veterans Administration Hospital, Portland, Oregon 97207. cigarette smoke have also stimulated public interest i n lung function testing. Lagging behind the wider usage of spirometry has been the development of accurate normal standards. T h e value of spirometric testing is influenced to a large extent by the accuracy of predicted normal standards. Most recognized standards have included nonsmokers, smokers, and former smokers. Other factors such as previous .history of respiratory disease, degree of physical debility, and exposure to atmospheric pollution have not always been critically examined. T h e purpose of this study is to provide reliable normal ventilatory standards in the form of prediction equations and nomograms based upon a healthy nonsmoking population. Materials and Methods Most of the 988 subjects tested were members of the Church of Jesus Christ of the Latter Day Saints (55 per cent) and the Seventh Day Adventist Church (24 per cent). The tenets of both churches forbid tobacco smoking. The subjects ' AMERICAN REVIEW OF RESPIRATORY DISEASE, VOLUME 103, 1971 57 58 MORRIS, BOSKI, AND JOHNSON TABLE 1 DISTRIBUTION OF SUBJECTS BY HALF DECADES IvAega.w) 20-24 25-29 30-34 35-39 4044 4549 50-54 56-59 6064 65-69 70-74 75-79 80%4 Total ~~ Number of Subjects Women Men 53 73 53 54 47 42 44 27 25 16 19 10 -8 47 1 102 67 77 68 52 42 41 26 16 11 9 2 2 517 TABLE 2 MEAN VALUES AND STANDARD DEVIATIONS FOR PHYSICAL CHARACTERISTICS AND PULMONARY FUNCTION MEASUREMENTS Women Mean SO Men Mean SD Age. Yean Height, inches FVC, liter FEV,, liter FEF200- 12 0 0 , Iitedrec FEF25- 75%. litedsec 42.5 64.2 3.47 2.70 5.2 3.1 15.9 2.3 0.74 0.68 37.9 70.1 5.2 3.97 1.4 7.9 1.0 4.1 14.3 2.9 0.97 0.00 1.8 1.3 lived in the mid-Willamette River Valley area Oregon, south of Portland. This area is EOn J ered to be relatively free of significant urban d pollution, and the elevation is less than 500 fa above sea level. The following procedures were used to obt?fi a population of nonsmokers who were free c disease or disability. A questionnaire was filled out immediately bc fore testing and established the following criteri regarding each subject: ( I ) never smoked 4 arettes for a total period longer than six month1 (2) never had asthma, chronic bronchitis, a pneumonia; (3) never had a persistent cough (4) had not been treated recently for any m piratory condition; (>) never had persisten chest wheezing; (6)never had any chest injur aor operation; (7) never worked in a polluted! atomosphere for any extended period. An interview cross-checked the results of the! questionnaire to eliminate subjects not fulfilling the criteria. During the testing, the subjects were observed for coughing or wheezing. The subjects' height was measured without shoes. All tests were done in the standing position, and a nose-dip was used. The tests were performed at various times of the day, but most were done between 7 pm and 9 pm. Two Stead-Wells spirometers with the soda lime cannister and two flutter valves removed from each were used. All tests were conducted by 6 trained technicians. The ventilatory tests consisted of the forced vital capacity (FVC), forced expiratory volume at 1 second (FEVJ, forced expiratory flow (FEF,oo-,200), and forced midexpiratory flow (FEF2,-7,qr,). The terminology of the ventilatory function tests is in accordance with the recommendations of the Committee on Pulmonary Physiology of the TABLE 3 CORRELATION COEFFICIENTS OF PHYSICAL CHARACTERISTICS AND PULMONARY FUNCTION MEASUREMENTS FOR WOMEN Age. w a n Height, inches FVC. liter FEVl, liter FEFZ00-1200, liter/sec FEF25-75%, liter/sec Age (years) 1.o -0.61 -0.47 -0.54 -0.66 -0.25 Height (inches) -0.61 1.o 0.61 0.40 0.82 0.50 FVC (liter) -0.47 0.61 1.o 0.61 0.65 0.35 FEVi (liter) -0.54 0.48 0.61 1.o 0.71 0.27 FEFZ00-1200 (literhec) -0.66 0.82 0.65 0.71 1.o 0.45 FEF25-75% (literhed -0.25 0.50 0.35 0.27 0.45 1.o SPIROMETRIC STANDARDS FOR NONSMOKING ADULTS 59 TABLE 4 CORRELATION COEFFICIENTS OF PHYSICAL CHARACTERISTICS AND PULMONARY FUNCTION MEASUREMENTS FOR MEN Age, years Height. inches FVC, liter FEV,, liter FEFZoo-1200, liter/rec FEFzs-,s%, liter/rec M=517 1.o -0.49 -0.41 -0.52 -0.65 -0.26 -0.49 1.o 0.39 0.36 0.77 0.54 -0.41 0.39 1.o 0.44 0.51 0.27 -0.52 0.36 0.44 1.o 0.73 3.23 -0.65 0.77 0.51 0.73 1.o 0.48 -0.26 0.54 0.27 0.23 0.48 1.o American College of Chest Physicians (1). Each test was performed at least twice to obtain a maximal effort. The best effort was always selected. Lung volumes were corrected to body temperature and pressure, saturated with water vapor (BTPS). The methods of calculation were those of Kory and associates (2). The results were randomly checked for accuracy by an independent pulmonary function laboratory. Results A total of 988 healthy nonsmoking men and women were included in the study. The distribution by half decades of age from 20 to more than 80 years is shown in table 1. T h e means and standard deviations for age and height for each sex are listed in table 2. Weight and body surface area as physical characteristics was not included. Kory and associates (2) reported that weight had lit- tle o r no correlation with any of the spirometric measurements. T h e matrix for product-moment correlation (r) based on the two physical characteristics and four ventilatory function measurements are shown in tables 3 and 4. Negative correlations were obtained between each lung function measurement and age. T h e FEF,,,,,,, showed the largest negative correlation with age for both sexes. Height was directly related to all four lung functions. T h e FEF,,o~,,o, had the largest positive correlation coefficient with height. Because all four pulmonary functions showed correlation with age and height, prediction formulas were derived for all spirometric variables for both sexes. T h e prediction formulas are listed in table 5. To facilitate determination of the predicted values, a nomogram for each sex was constructed which included the four ventilatory functions and the two physical measurements (figures l and 2). By using a TABLE 5 PREDICTION FORMULAS WITH MULTIPLE CORRELATION COEFFICIENT (R) AND STANDARD ERROR OF ESTIMATE (SEE) P r d l a i o n Formula R SEE Women - -FVC = 0.1 15 Hin 0.024A 2.852 - -FEVl 5 0.089 Hin 0.025A 1.932 0.71 0.73 0.52 0.47 FEFzoo-,zoo 00.145 Hin- 0.036A- 2.532 FEF~~-,s%-0.060 Hi,- 0.030A + 0.551 0.53 0.56 1.19 os0 Men FVC = 0.148 Hin- 0.025A- 4.241 - -FEV1 0.092 Hin 0.032A 1.260 -FEF~00-1200= 0.109 Hfn 0.047A + 2.010 0.65 0.73 0.44 0.74 0.56 1.66 - -FEF~s-,5%=0.047 Hin-0.045A+ 2.613 0.63 1.12 0 . . K8y: Hin height in 1nch.S; A ago in yean. MORRIS. KOSKI, AND JOHNSON HEIGHT, in. cm. 145 62 66 FEF25-75%1'/sec A*, 1 16.5 c1 r AGE, Yrs -7." -3.5 - 3.0 - - -6 0 e.570 80 2.0.- 90 1.5.- c L i!--5.5 - 5.0 3.5 - 4 5 3.0 2.0 - 4.0 - 2.5 1.5 3.5 C-I E I R SEE -F E F 2 0 0 . 1 2 0 0 ' 0.145 Hi,- 0.036A 2.532 [Os53 1 . 1 93 11.5 =F E F 2 5 - 7 5 ? ~ 0.060 Hi,- 0 . 0 3 0 A + 0.551 c0.56 0.801 -FEVl.o,,, = 0.089 Hi,- 0.025 A 1.932 [0.73 0.471 -F V C = 0.1 I5 Hi,,- 0.024 A 2.852 [O.? I 0.521 Fig. 1. Prediction nomogram for normal women. SPIROMETRIC STANDARDS FOR NONSMOKIXNG ADULTS 61 HEIGHT, in. cm. * MALES t -FE 5.5 - -5.0 -- 4.5 --4.0 --3.5 - - 3.0 ---2 5 --2.0 - FVC,L R SEE -FEF200-,200 = 0.109 Hi, 0.047 A + 2.010 r0.44 1.661 F E F ~ ~ - T= ~0.%047Hin-0.045A +2.513 [0.53 1.121 -FEV 1.0 scc = 0.092 Hi, - 0.032 A 1.260 [0.73 0.551 - -FVC = 0.148 Hi, 0.025A 4.241 [0.65 0.741 Fig. 2. Prediction nOmOgram for normal men. I 62 MORRIS, KOSKI. A N D JOHNSON 1 .i !. ,!.b.. I I I TABLE 6 COMPARISON OF PREDICTION FORMULAS FOR WOMEN USING MEAN VALUES FOR AGE AND HEIGHT FROM PRESENT STUDY Investigators Test Ilt?erl Prediction Formula SEE Pred Mean Obs Mean Morris et al Berglundet al (15) Ferris et a1 (17)' Ferris et a1 (17)t Goldman et al ( 18) Kory et al (20) Needham et a l ( 1 9 ) Morris et a1 Berglund et al (15) Ferriset al (17). Ferris et sl (17 ) t Kory et al(20) Lindall et al (10) FVC VC FVC FVC VC FVC VC FEVl FEVl FEVl FEVl FEVl FEVl 0.115 Hin - 0.024A- 2.852 -0.040 Hcm - 0.022A 2.351 0.033 Hcm- 0.023A- 1.40 -0.030 Hcm- 0.023A 0.94 - -0.052 Hcm 0.018A 4.36 0.041 Hcm- 0.018A-2.69 0.039 Hun- 0.020A- 2.71 - -0.089 Hin 0.025A 1.932 0.0267 Hcm- 0.027A- 0.54 0.025 Hem- 0.022A- 0.62 -0.022 H, 0 . 0 2 4 ~- 0.00 - -0.028 Hcm O . M l A 0.87 67.6 H i n - 23.OA- 918 0.52 0.40 0.46 0.47 0.43 0.37 0.36 0.47 0.36 0.40 0.42 0.33 0.36 3.51 3.24 3.00 2.98 3.36 3.23 2.87 2.72 2.67 2.52 2.49 2.80 2.44 3.47 3.47 3.47 3.47 3.47 3.47 3.47 2.70 2.70 2.70 2.70 2.70 2.70 Key: Hin = height in inches; H, = height in centimeters; A = age in years. Differ +0.04' -0.231 -0.41. - 0.49 -0.1 1; -0.24: -04 +0.024 -0.03 -0.18 -0.21 +0.10' - 0.26 Included nonsmokers and exsmokers straight edge to connect an individual's age and height, all four values of forced expiration can be obtained at once. Comparisons between the present study and previous studies regarding prediction formulas are shown in tables 6 and 7, which indicate the differences between the predicted mean and the observed mean values. The mean values of height and age from the present study were substituted in the form las of previous studies in calculating the pre{ dicted mean. The observed that were calculated from tmheeatnostaalresatmhopslee1f d the present study. The difference between ~ predicted and observed means permitted comparison of the different piediction for- mulas. Of the 24 predicted means shown in tables 6 and 7, only the FEV, of women re- TABLE 7 COMPARISON OF PREDICTION FORMULAS FOR MEN USING MEAN VALUES FOR AGE AND HEIGHT FROM PRESENT STUDY Investigators Test (liter) Prediction Formula SEE Pred Mean Morris et al Berglund et al (15) Cotes (16) Ferris et al (17)' Ferris et a1 ( 1 8 ) t Goldman et al ( 19) Kory et a1 (2) Needham et al (19) Morris et al Anderson ( 2 0 ) Berglund at al (15) Cotes (16) Ferris et al (17). Ferris e t al ( 1 7 ) t Kory et al (2) FVC VC FVC. FVC FVC VC vc VC FEVl FEV1 FEVl FEV, FEV, FEV, FEV, 0.148 Hin- 0.025A- 4.241 0.048 Hem- 0.020A- 2.81 - -0.051 Hcm 0.032A 3.02 -0.051 Hcm 0.025A- 3.55 - - '0.046 Hcm- 0 027A- 2.79 0.064 Hcm 0.031A 5.335 0.052 Hcm- 0.022A- 3.60 0.043 Hcm - 0.035A- 1.91 0.092 Hi,,- 0.032A - 1.260 - -0.005 Hem- 0.029A + 3.614 0.034 H, 0.033A 1.OO 0.035 HCm- 0.033A - 1.12 -0.036 Hcm- 0.027A 1.65 0.029 Hem- 0.028A- 0.70 0.037 Hcm - 0.028A- 1.59 0.74 0.50 0.52 0 59 0.66 0.49 0.58 0.44 0 55 0.61 0.50 0.45 0.49 0.59 0.52 5.19 4.98 4.84 4.59 4.38 4.89 4.83 4.42 3.98 3.41 3.80 3.79 3.79 3.40 3.94 Key: Hin = height in inches; H, = height in centimeters; A = age in years. *Included smokers. tlncluded nonsmokers and exsmokers. Obs Mean 5.20 5 20 5.20 5 20 5.20 5.20 5.20 5.20 3.97 3.97 3.97 3.97 3.97 3.97 3.97 Difference -0.01 - 0.22 - 0.36 -fl 61 -0.82 -0.31 -0.37 -0.78 +0.01 -0.56 -0.17 -0.18 -0 23 -0.57 -0.03 '1 4 SPIRONETRIC STANDARDS FOR NONSMOKING ADULTS 63 I by Kory and associates exceeded the in the present study. T h e decline in lung function by age for sex is illustrated in figures 3 and 4. presumably represented the effects of rather than smoking or apparent, pul- tion of subjects. The mean values of all tests demonstrated larger values for voiue MALES of methods, prediction formulas, and al ranges for various pulmonary func- This resulted from a Fig. 4. Relationship of forced expiratory mea- by the Veterans AdministrationCaoperative Study (2). No data reg the smoking habits of the subjects, r, were provided. I n the second paper same study (4), 337 men performed supine vital capaaty but not the FEV,. T h e questionnaire survey a h revealed that the FEF25-75sawasfrequently used by the 73 responding laboratories from which usable informatian was obtained. Twenty- FEMALES mal. T h e FEF200-1200was used least often of the standard spirometric tests, and the Veterans Administration-Army prediction equations were again reported to be most commonly used although none was given in the reference a t e d (2). .ko i o o: e; o; Age (years) FEVl.~ip`L' A 3. Relationship of forced expiratory meaents to age in noma1 women. T h e present study found regression with age and decreasing height for all four tests of forced expiration. Correlation coefficients for the four tests showed considerable variation, ranging from 0.25 to 0.82. T h e value of forced expiratory flows, especially the FEF25a,Sc, has been both praised and can- demned. T h e test has been recommended by some as an early index of airway obstruction (6) whereas others (7, 8) have found . 64 MORRIS, KOSKI. A N D JOHNSON a 2 a .I , !' .'.' * I .- SPIROMETRIC STANDARDS FOR NONSMOKING ADULTS 65 FEF2,,,,, and FEF,,,,% have spirometers should be determined to demprediction formulas b b u s e onstrate the usefulness of certain normal rediction formulas for FEF26-76% impractical. alone. T h e results of the present T h e decline of ventilatory function with a nonsmoking population reflect age presumably represents species-specific, +were r = 0.50 for women and 0.54 for men. It should be noted &at gression equations using age and subjects over the age of 70 should be studied to achieve statistically valid prediction equations for the aged. awareness of these limitations. normal population in this study can be mn- tion of the present data with other sidered to be free of bronchitis and other 1y used Collins 13.5-liter spirameter, This results from selection of a healthy non- TABLE 10 COEFFICIENT OF VARIATION FOR LUNG VOLUME MEASUREMENTS 66 MORRIS. KOSKI. AND JOHNSON smoking population covering a wide age velocidad y del volumen del Eujo apiratono range of both sexes. Despite this selection, forzado en 988 hombres y mujeres saludables, no the values demonstrate a steady dedine in fumadores, quienes habian sufrido relativamente pulmonary function with age, and a wide variation from mean values not attributable to age, height, or sex. The regression equations reflect the total sample whereas tables 8 and 9-show the extent of variaiton from poca expwid6n a un ambiente contaminado. Se observ6 una correlaci6n negativa con la edad y positiva con la estatura. Se derivaron f6rmulas pronosticadoras y se construyeron nomogramas para cuatro valores ventilatorios normales. Se compararon las f6rmulas pronosticadoras con las .. the normal mean by age groups of ten years. T h e coefficient of variation for each decade de otros investigadores. Se observ6 una disminucidn progresiva en la up to age 70 is shown i n table 10. By defini- funcidn ventilatoria con la edad; en todas las tion, the mean & one SD includes 68 per cent of the population tested, excluding 32 per cent of the normal individuals falling outside of one SD of the mean. Sob01 has emphasized the difficulties in detecting abnormal pulmonary function in the asymptomatic individual by simple ven- edades, superaron 10s varones. Ocurrieron amplias variaciones en todo el grupo, no atribuibles a la edad. sex0 o estatura. Esto fue m& pronunaado en la3 dos velocidades espiratorias forzadas, con correlaciones mdltiples bajas y con grandes errores en lo estimado. Las amplias variaciones de 1% valores normales afectan seriamente la utilidad de la espirometria tilatory tests (14). To achieve the desirable en descubrir con certeza el deterioro de la funci6n goal of reliable early detection of chronic ob- ventilatoria en sujetos asintomAticos. Se hace structive pulmonary disease, development of testing techniques other than spirometry will necesario emplear diversas tecnicas para descubrir precozmente la funci6n pulmonar impedida. . 4 .' be required. For the present, spirometry will continue t a be of value i n documenta- RESUME tion of the degree of ventilatory insufficiency in the symptomatic patient, and for serial observations of progression of disease or the Standards spirometriques pour le5 adultes non-fumeurs et en bonne sante beneficial effects of therapy. Les etudes qui ont ktk prkcedemment menks en vue d'btablir des valeurs normales d'ktalonnage Acknowledgment The writers thank Dr. Roger Petersen for the statistical consultation and Mr. Larry Janssen for the computer consultation (both of Oregon State University). We are also grateful to Mrs. Geraldine King, Pulmonary Function Laboratory, Portland Veterans Administration Hospital, for technician training, and to research assistants Messrs. Edward Ford, Raymond Koski, and Victor Williams and Misses Frances Seuell an4 Colleen Timmerman for testing and record analysis. pour les epreuves ventilatoires, ont habituellement i n c h un nombre important de fumeurs de cigarettes et de malades hospitalis&. Les taux de dCbit expiratoire ford et les volumes ont 6tP mesurks chez 988 hommes et femmes, en bonne santC et non fumeurs, qui avaient eu une exposition relativement faible P la pollution de l'air, sous quelque forme que ce soit. On a obtenu des corrhlations 13gatives avec 1'8ge. et des corrClations positives avec la taille. On en a derive des formules de prediction, et des nomogrammes ont et4 construits pour quatre mesures ktalons de la ventilation. Des comparaisons ont kte effectukes auec les formules de prediction ktablies par d'autres chercheurs. RESUMEN On a observe une diminution progressive de la Valores espirometricos en adultos no fuma- dores saludables fonction ventilatoire avec l'r2ge. La superiorit6 des hornmes h cet (gard est cependant maintenue ?i tous les 2ges. De larges variations ont ktk Prcvios estudios practicados para determinar 10s observces dans la population, qui ne pouvaient valores normales de las pruebas ventilatorias, han pas Ctre attribuks A l'age, au sexe ou ?i la taille. incluido gencralmente un numero considerable Ces variations Ctaient les plus pronondm pour de fumadores de cigarrillos y de pacientes hos- les deux taux de debit expiratoire force, qui pitalizados. Se detcrminaron 10s valores de la thoignaient de correlations multiples faibles, of respiratory impairment, Arch. Environ. xio Health (Chicago), 1966,12,146. no 10. Lindall, A., Medina, A., and Grismer, J. T.: nte A reevaluation of normal pulmonary funcSe tion measurements in the adult female, ! \ Amer. Rev. Resp. Dis., 1967,95,1061. las 11. Kory, R. C., and Hamilton, L. H.: Evaluation ld', of spirometers used in pulmonary function se studies, Amer. Rev. Resp. Dis., 1963, 87, 228. '3s 12. Ericsson, P., and Imell, L.: Spirometric studies of ventilatory capacity in elderly peola ple, Acta Med. Scand., 1969,185,179. as 13. Anthonisen, N. R., Danson, J., Robertson, 11. P. C., and Ross, W. R. D.: Airway closure as es a function of age, Resp. Physiol., 1969, 8, 1- 338. 5. 14. Sobol, B. J.: Assessment of ventilatory abT. normality in the asymptomatic subject: An excerdse in futility, Thorax, 1966, 21, 445. \ 15. Berglund, E.,Birath, G., Bjure, J., Grimby, a G., Kjellmer, I., Sandqvist, L., and Soder- 1 holm, B.: Spirometric studies in normal c r subjects: 1. Forced expirograms in subjects between 7 and 70 years of age, Acta Med. Scand., 1963,173,185. 16. Cotes, J. E., Rossiter, C. E., Higgins, I. T. T., and Gilson, J. C.: Average normal values for the forced expiratory volume in white Cau- casian males, Brit. Med. J., 1966,l, 1016. 17. Ferris, B. G., Jr., Anderson, D. O., and Zick- I mantel, R.: Prediction values for screening tests of pulmonary function, Amer. Rev. Resp. Dis., 1965,91,252. 18. Goldman, H. I., and Becklake, M. R.: Res- piratory function tests: Normal values at median altitudes and the prediction of nor- mal results, Amer. Rev. Tuberc., 1959, 79, 457. 19. Needham, C. D., Rogan, M. C., and Mc- Donald, 1.: Normal standards for lung vol- umes, intrapulmonary gas mixing, and max- imum breathing capacity, Thorax, 1954, 9, 313. 20. Anderson, D. 0.:Smoking and respiratory disease, Amer. J. Public Health, 1964, 54. 1856. 21. Smith, Josef R., and Kory, R. C.: Labora- tory aids in investigating pulmonary dis- eases, in Textbook of Pulmonary Diseases, Gerald L. Baum, ed., Little, Brown & Co., Boston, 1965,p. 25.