TO: R. A. Conrad Interoffice Communication FROM: DATE: SUBJECT: D. R. Booth October 10, 1989 VISTA Attached is a letter from LDEQ stating that the Honeywell 1000 VCM monitor was operating within the allowable 15% relative accuracy at the time of the June 2, 1989 LDEQ inspection. vvv 000003295 BUDDY ROEMER Governor State of Louisiana Department of Environmental Quality PAUL TEMPLET Secretary September 21, 1989 Mr. David Booth Lake Charles VCM Plant VCM Plant Road P. 0. Box 6C5 Westlake, Louisiana 70669 Dear Mr. Booth: RE: Audit Results of Honeywell 1000 Vinyl Chloride Monitor On June 2, 1989, the above referenced monitor was audited by Emission Testing Services under contract by the Louisiana Department of Environmental Quality, Air Quality Division. The purpose of the audit was to determine the relative accuracy of the monitor by introducing two different concentrations of EPA protocol gases and recording the monitor output. The relative accuracy of the monitor for the lower concentration audit was 1.06% and for the higher concentration audit was 13.54%. Since these results are within the specified limit of 15%, no corrective action will be required. Should you have any questions, please call James Magee at (504) 342-9056. Sincerely, TC:JMspfv Tom Coerver Engineering Supervisor OFFICE OF AIR QUALITY P.O. BOX 44096 F------------------- ~ ---------------,, . AN EQUAL OPPORTUNE