Document B8kdVxoN0aVBKB4mEMNx8yBmL

S-76-M-24 METHOD FOR THE QUANTITATIVE DETERMINATION OF POLYCHLOROBIPHENYL IN WORKPLACE AIR SCOPE This method determines the concentration of polychloroblphenyl fPCBs) In air from 0.1 to 1.0 mg/m'). The standard (Threshold Limit Value) of PCS is 0.5 mg/mJ. The procedure can determine average worker exposure over an eight-hour day or for short sampling periods. ' PRINCIPLE This method Is based on the collection of PCB In an adsorption tube containing Tenax SC. In the laboratory, the samples are solvent desorbed with hexane and analyzed by electron capture gas chromatography. The procedure is only valid If the chromatogram of the sample matches the chromatogram of a PCB standard mixture because of the wide range of LC response of different PCB Isomers. PROCEDURE 1. Preparation and Conditioning of Collection Column Collection columns containing about 0.16 g of Tenax GC are Spared and conditioned by the procedure In Spectroscopy Special Study S-76-SS-14. Shortly before sailing, recondition the collection columns In a GC at 250"C for 15-20 minutes. Carrier flow of the GC should be ca 30 ml/mln. Check collection coltams periodically for changes In flow rate due to extended use (see S-76-SS-14). 2. SaimlIno Pumps a. Choosing Punp Flow Rate Breakthrough volume Is the amount of air that can pass through a collection column before the component being collected begins eluting from the coltam. This limits possible sample size. With this size Porapak N collection column at 80*F, the breakthrough volume of PCB Is greater than 30 liters. A typical sample Is 10 liters. Increasing sampling temperature will decrease the breakthrough voluae. Depending on the pump available one may need a sample splitter. The purpose of a splitter Is to reduce the flow rate through the column. This permits sampling below the breakthrough volume. For PCB, a flow rate ca. 20 ml/mln Is required If an 8-hour sampling Is desired (see 2b). We have found the personnel pump SKC-222-451 model with a splitter very satisfactory for collecting saaples at this rate. A very slow flow rate pump such as the SKC-222-3-12 model without a splitter also works well. See Spectroscopy Special Study S-76-SS-10 for a description of sample splitters and their use. b. Determining Breakthrough Volumes The most convenient way we find to measure breakthrough volwes Is described below: HONS 004028 S-76-M-24 Page two (1) Attach a freshly conditioned collection col inn Inside the GC to the Injection port. (2) Connect the other end of the colunn to a piece of tubing, preferably glass, that will connect to the detector. (3) Adjust the flow rate to the flow you will be using when you sample. (4) Inject 1 ul of PCB Into the GC and allow It to run at room temperature You will probably get a very broad peak. The point at which the base line begins to rise Is the breakthrough point. (5) Determine the time from Injection to the breakthrough point and cal culate as below: Breakthrough - flow through colunn X time at breakthrough point X 10'3 (liters) (ml/mln) (min) (liters/ml) c. Calibration of Sampling Pump (1) Take a 2-ft piece of tubing and attach It to the dimpled end of a collection tube. (2) Attach the other end of_the tubing to a personnel monitoring punp. (3) Be sure the tubing fits tightly over all connections. (4) Adjust the flow rate of the punp so breakthrough does not occur in the desired sampling time (see below). The flow rate can be measured with a soap bubble meter of the type used In calibrating flow rates for GC carrier gases. It Is essential when checking flow rates to have the collection colwn ( or columns If two In series are to be used) In the system. If you wish to sample for 8 hours, the pump flow should not exceed 20 ml/mln. This Is calculated below: 8 hrs X 480 min fl0w . 10.000 ml Tfcrw :20 ml/mln d. Column Capacity The capacity of these collection tubes for PCB Is at least 0.01 mg. We have not Investigated higher concentrations. The presence of high (mg) levels of other organic chemicals will probably effect the column capacity for PCB. If over 0.02 mg of organic material will likely be collected, sampling should be carried out with two columns In series. Both columns are then analyzed to Insure capacity has not been exceeded. HONS 004029 S-76-M-24 Page three 3. Sample Collection Replace the tube used for calibration with a freshly conditioned collection colunn. If personnel monitoring Is planned, the pump can be-attached to the worker's belt or put in a pocket. The collection tube should be positioned in the breathing zone. Either wrap the tubing around the back of the neck and let the tube hang on the shoulder or clip it to the collar. Start the pump. Record the time. After the end of the sampling period, turn the punp off and record the final time. Disconnect the collection tube and wrap each end with teflon tape. DO NOT USE any other type of tape because of possible contamination. If you anticipate the data will be presented to OSHA or other regulatory agencies, use of a back-up sampling column Is recommended. Do this by con necting two columns In series with a small piece of tubing (not rubber). It may be convenient to clip the first tube to the clothing and let the second hang over the shoulder (see attached figure). By analyzing both tubes, you can show no sample was lost. NIOSH charcoal tube methods include the guide line that If the back-up tube content exceeds 25* of that found on the front column, the possibility of sample loss exists. We recomnend sampling be repeated if such is the case. The conditions for collecting PCS are: Maximum Sample Size: 30 liters air Pusp Flow: Adjust for desired sampling period (20 ml/mln for 8-hour sampling) A. Analysis a) 1. Attach the undimpled end of the collection tube to system similar to figure 3.2 3 4 5 2. Place 10 ml of hexane In funnel and force through the sample tube with N2. 3. Record the volume of extract 4. Repeat ___ 5. Analyze by EC/6C as described below. b) Prepare standards by weighing ca 0.17 g PCS Into a 100 ml volumetric flask. Dilute to mark with nanograde hexane. Three pi of this solution Injected Into a 6C Is equivalent to 0.5 ug/1 PCS In a 10 liter air sample (the TLV). Calculate the PCS concentration In mg/ml. Inject the standard Into the EC/GC. EC/SC parameters for PCB are: Electron Capture Gas Chromatographic Procedure Instruments - A gas chromatograph equipped with Isothermal oven tempera ture control. Detector - High-temperature N1 Electron Capture Cell. Colum - A 6-am by 6 ft. (1.8-m) glass column, 4t XE-60 on 80/100 mesh Chromosorb W, high-performance. HONS 004030 S-76-M-24 Page four Column Tenperature - 190C Isothermal. Detector Temperature - The detector temperature shall be 250 to 300C. Injection Port Temperature - The Injection port temperature shall be 220C. Pulse - The pulse rate shall be 50 us. Flow Rates - The flow rates shall be as follows: Helium Carrier - 60 ml/mln 901 Argon-lOX Methane Purge - 20 ml/mln. Using EC/SC as the determinative step. Inject In duplicate 2 to 8 ul of each sample solution Into the chromatograph. By comparison with standard solutions Injected, under the same operating conditions, determine the amount and type of Aroclor using the total area method. Quantitative Determination Quantitative determinations eaploylng the electron capture detector are non-stolchlometrlc measurements made by comparing the total areas for known concentrations with those for unknown concentrations. Four variations of the area quantification procedure have been employed as follows: Case 1 - EC gas chromatogram of PCB unknown unchanged with respect to standard PCB with no evidence of Interferences. Case II - EC gas chromatogram of PCB unknown altered with respect to standard PCB with no evidence of Interference. Case III - EC gas chroamtogram of PCB unknown unchanged with respect to stan dard PCB with evidence of Interference. Case IV - EC gas chromatogram of PCB unknown altered with respect to stan dard PCB with evidence of Interference. Determine the PCB level In Case 1 and II samples by comparison of the total area of the peeks in the unknown with a calibration curve prepared by plotting the total area of the peaks In the electron capture chromatogram of the corresponding standard versus the niaber of nanograms of the standard Injected Initially subject Case III and IV samples to the chemical cleanup procedure and follow by chromatography on alumina. If the Interferences are removed by this treatment, calculate the PCB levels for Cases III and IV In the same manner as Cases I and II. If dominant Interference(s) Is(are) still present, estimate the maximum PCB level In accordance with Cases I and II after cor recting the total area for that of the Interfering peak(s). In all cases, the response of the electron capture must be linear for quantitative analysis. Calculations Assuming that the calibration curve Is prepared by plotting the electron capture response (total peak area) versus nanograms of the PCB standard Injected, calculate the PCB level present In the air, water, or sediment- soil sample as follows: ' HONS 00403k S-76-M-24 Page five Air Samples: f(A/BHD 10' ul)/m) no' mq/ni G(28.32 liter/ft Iter - Pi/Po)(T0/273 T,) where: A * PCB In unknown from calibration curve, ng, B * final concentrate Injected Into the Instrument ul, D volume of final concentrate, m., E sample, weight, kg, F concentration of PCBs In air sample, mg/m', at one atmosphere and 25,C. S air sampled from meter, ft' , Pg- differential pressure of gas stream, mm Hg, Pi* standard pressure, 760 mu Hg, To* standard temperature 298 K, and Ti* temperature of sampled gas, "C, The corrections for the pressure drop through the scrubbers and for the gas temperature are necessary to get an accurate measurement of gas volume smapled. DISCUSSION He validated this method by Injecting the standard directly on the absorption column and solvent desorbed the material with nanograde hexane. Comparison with EC/GC standards gave us this result - VALIDATION CONCENTRATION PCB INJECTEO t RECOVERY PCB ON TUVE (uol EQUIVALENT CONCENTRATION IN 1 LITER AIR (uo/1) 0.2 Standard Standard 2.0 Standard 109> 120/ 951 87/ 91\ 99/ 114 91 96 1 5 10 0.1 0.5 1.0 References: Spectroscopy Special Studies S-76-SS-10, S-76-SS-11, S-76-SS-1 *, S-76-SS-15; GC File #76-234 Attachment Monsanto Industrial Chemicals Co., Applied Sciences, St. Louts, Mo. 12/76 - L. M. Chapman, C. E. Hoffmann, M. H. Dietrich HONS 004032 S-76-HMONS 004033 S-76-M- tubing i MONS 004034 nitrogen line funnel connecting tubing collection tube MQNS 004035