APPENDIX L ANALYTICAL METHODS r fly i j l d i UiL'uti iry Method 73-9 Job No. 1005T22 GAS CHROMATOGRAPHIC DETERMINATION OF M0N0CHL0R0BIPHENYL _________ ISOMERS IN MCS 1547 PURPOSE OF ANALYSIS To determine whether all components in MCS 1547 are within specifica tions. SAMPLE PREPARATION Hone (add benzene to liquify if the sample Is solid), CHROMATOGRAPHIC CONDITIONS Instrument: Varlan Aerograph Series 1200 Type Oetector: Flame Ionization 'Column: 50* X 0.02" I.D. Support Coated Open Tubular (SCOT) coated with Poly-1-110 polyimide. - Column Temperature: 200C Detector Block Temperature: 300C Injection Port Temperature: 300C Sample Volume Injected: 0.05 ut (without spilt) Flow Rates Helium: 6 ml/min. Auxiliary Gas: 40 ml/min. Hydrogen: 45 ml/min. Air: 350 ml/min. RETENTION TIMES fUNCORRECTED) OF COMPONENTS SEPARATED (CHROMATOGRAM ATTACHED! Biphenyl (3.6*); o-Chloroblphenyl (4.5'); 2,2'-D1chioroblpheny1 (6.2'); m-Chlorobiphenyl T6-8'); -Chlorobiphenyl (7.1'). CALCULATION TECHNIQUE Normalization with Area Correction Factors. PRECISION No determination of precision was made. The variability of a single GC analysis at the 95X confidence limits is usually +0.5 to IS rela 11ve. db Monsanto Industrial Chemicals Company Applied Sciences St. Louis, Mo. 1/14/74 - G. M. Gasser, E. M. Emery HONS 02171^ HONS 021715